agilent technologies 70429ak95
TRANSCRIPT
Agilent Technologies 70429AK95
User’s and Service Guide
Use this manual in conjunction with the following document: E5500 Phase Noise Measurement System
Manufacturing Part Number: 70429-90043
Printed in USA October 2007Supersede: September 2007
© Copyright 2006, 2007 Agilent Technologies, Inc. All rights reserved.
Warranty StatementTHE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
ii
Safety Notes The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document.
WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
Definitions
• Specifications describe the performance of parameters covered by the product warranty (temperature –0 to 55 °C, unless otherwise noted.)
• Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specification that 80% of the units exhibit with a 95% confidence level over the temperature range 20 to 30 °C. Typical performance does not include measurement uncertainty.
• Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.
• Characteristic Performance describes performance parameter that the product is expected to meet before it leaves the factory, but is not verified in the field and is not covered by the product warranty. A characteristic includes the same guard bands as a specification.
iii
iv
Contents
70429AK95Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Verifying the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Recommended Divider Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Manual Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6External Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7DC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Performance Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Typeface Key Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Equipment Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Divided Frequency and Output Level Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12600 MHz Residual Phase Noise Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Residual Phase Noise Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Making the Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1810 MHz 70429AK95 Phase Noise Test Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
10 MHz LNSG Test Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Making the 10 MHz Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Safety and Regulatory Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Connector Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Statement of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Shipping Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Compliance with Canadian EMC Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Compliance with German Noise Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Electrostatic Discharge Protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Instrument Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Contacting Agilent Sales and Service Offices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Contents-1
Contents
Contents-2
70429AK95
User’s and Service Guide 1
70429AK95Introduction
Introduction This User’s and Service Guide Supplement contains information necessary for use with the 70429AK95 Frequency Divider. The 70429AK95 is an accessory used in conjunction with the Agilent E5505A Phase Noise System or an E8257D/E8267D Synthesized Signal Generator to improve the phase noise performance. An RF frequency range of 25.6 MHz to 10 GHz is input to the 70429AK95 from the signal generator. The output frequency of the 70429AK95 is the result of the divide number; set by configuring the appropriate switches on the 70429AK95 or by connecting to an external switch unit, such as an Agilent 34970A. The 70429AK95 has two output paths. One has an output frequency range of 100 kHz to 5 GHz. The second output goes through an amplifier and has a frequency range of 100 kHz to 1.6 GHz. A differential output with a frequency range of 100 kHz to 5 GHz is available by removing the jumper to the input of the amplifier.
This Supplement contains the following information:
• “Introduction” on page 2
• “Verifying the Shipment” on page 3
• “Specifications” on page 4
• “Operation” on page 5
• “Performance Test” on page 11
• “Replaceable Parts” on page 24
• “Safety and Regulatory Information” on page 25
• “Contacting Agilent Sales and Service Offices” on page 30
2 User’s and Service Guide
70429AK95 Verifying the Shipment
Verifying the Shipment After the test set has been unpacked, keep the original packaging materials so they can be used if you need to transport the instrument.
Verify and inspect the items received in Table 1 for any signs of damage that may have occurred during shipment. If your accessories appear to be damaged or missing, refer to “Contacting Agilent Sales and Service Offices” on page 30.
Table 1 Content List
Description AgilentPart Number Qty
70429AK95 User’s and Service Guide 70429-90043 1
Power Cable 83020-60004 1
User’s and Service Guide 3
70429AK95Specifications
Specifications These are the performance standards at which the 70429AK95 can be tested.
Table 2 Specifications
Electrical Characteristics Performance Limits Conditions
Frequency:Input Range 25.6 MHz to 10 GHzOutput Range 100 kHz to 5 GHzOutput Waveform Square wave AC Coupled (setup to drive
50 Ohm load)AmplitudeInput Range:
25.6 MHz to 10 GHz @ 25 °C +3 dBm to +7 dBm
Output Range:100 kHz to 5 GHz @ 25 °C +5 dBm Nominal Divider Output100 kHz to 1600 MHz @ 25 °C +15 dBm Nominal Amp Output
Spectral Purity:Residual SSB Phase Noise in 1 Hz bandwidth. (Phase Noise specifications at any offset can be determined by drawing a line, on a log−log plot, between the given specification points).
600 MHz output frequency (4.8 GHz input divided by 8, +9V,−9V supplied by Agilent 6236B or E3630A Power Supply or equivalent)
–100 dBc 1 Hz offset from carrier–150 dBc 100 kHz offset from carrier–154 dBc 1 MHz offset from carrier–154 dBc 100 MHz offset from carrier
General:Operating Temperature Range +15 °C to +35 °CPower Requirements +9 Vdc (±0.2 Vdc) 150 mA Nominal
−9 Vdc (±0.2 Vdc) 570 mA Nominal
4 User’s and Service Guide
70429AK95 Operation
OperationThis User’s and Service Guide Supplement contains information necessary for use of the 70429AK95 Frequency Divider. The 70429AK95 is an accessory used in conjunction with the Agilent E5505A Phase Noise System or an E8257D/E8267D Synthesized Signal Generator to improve the phase noise performance. With the input connected to the RF output of the signal generator, the 70429AK95 divides the frequency down to a lower frequency based on the divide number determined by configuring the appropriate switches A1, A2, or A3, B1, B2 or B3. The divided RF output frequency may have less phase noise than it would if it were output directly from the signal generator.
General Information
The 70429AK95 Divider Accessory accepts RF frequencies in the 25.6 MHz to 10 GHz frequency range at a minimum level of +3 dBm, but not greater than +7 dBm. In order to provide the best possible signal to noise ratio at the input to the divider, it is recommended that the signal generator output level be set to +16 dBm and a 10 dB passive attenuator be placed between the signal generator output and the divider input. If the signal generator can not be set to +16 dBm, use a lower output power and a correspondingly smaller attenuator value such that the power at the divider input is approximately +6 dBm. The output signal will be at a frequency equal to the input frequency, divided by a divide number set by configuring the appropriate switches (see Table 2) at a level of +5 dBm nominal (RF Out − and RF Out + Ports) and +15 dBm nominal (1.6 GHz amplifier RF Out Port).
The 70429AK95 has 2 independent dividers; A and B. The output of divider A is connected internally to the input of divider B. Each divider can be programmed independently to divide by 1,2,4,8 or 16, for a maximum division ratio of 256.
Recommended Divider SettingsTable 3 Manual Control (A and B Dividers)
Output Frequency Signal Generator Frequency
Divider A Divider B
2 GHz to 5 GHz 4 GHz to 10 GHz 2 1
1 GHz to 2 GHz 4 GHz to 8 GHz 2 2
500 MHz to 1 GHz 4 GHz to 8 GHz 2 4
250 MHz to 500 MHz 4 GHz to 8 GHz 2 8
125 MHz to 250 MHz 4 GHz to 8 GHz 2 16
62.5 MHz to 125 MHz 4 GHz to 8 GHz 4 16
32.25 MHz to 62.5 MHz 4 GHz to 8 GHz 8 16
100 kHz to 32.25 MHz 25.6 MHz to 8 GHz 16 16
User’s and Service Guide 5
70429AK95Operation
Manual Control
Set the I/E switch to 1 to manually control the 70429AK95.
The RF Output can be enabled or disabled by setting the E/D switch to 1 (enabled) or to 0 (disabled).
The A and B dividers are programmed according to the following table.
Table 4 Manual Control (A and B Dividers)
Switch /1 /2 /4 /8 /16
Divider A A1 0 0 0 0 1
A2 0 0 1 1 0
A3 0 1 0 1 0
Divider B B1 0 0 0 0 1
B2 0 0 1 1 0
B3 0 1 0 1 0
6 User’s and Service Guide
70429AK95 Operation
External Control
To control the 70429AK95 externally, the manual control I/E Switch is set to 0, this enables the external control interface and disables all other manual switches.
The 70429AK95 has an 15 pin D-Connector external interface. This interface has 3 control lines for each divider and 6 control lines for a fine output power adjustment. The default state for the 12 control lines are 5V (1). The external switch control unit must provide either an open (1) or a ground (0) for each control line.
Set either or both dividers control lines to 1 to disable the RF Output.
Table 5 External Control
Switch /1 /2 /4 /8 /16 Disable External Interface Connector
Divider A A1 0 0 0 0 1 1 Pin 1
A2 0 0 1 1 0 1 Pin 2
A3 0 1 0 1 0 1 Pin 4
Divider B B1 0 0 0 0 1 1 Pin 6
B2 0 0 1 1 0 1 Pin 8
B3 0 1 0 1 0 1 Pin 9
Table 6 Output Power Fine Tune
Dac Bit
External Interface Connector
1 Pin 10 MSB
2 Pin 11
3 Pin 12
4 Pin 13
5 Pin 14
6 Pin 15 LSB
User’s and Service Guide 7
70429AK95Operation
DC Power
The 70429AK95 requires the following DC power from an external Agilent 6236B or E3630A (or equivalent power supply) to meet the noise performance noted in the specifications listed in Table 2 on page 4.
Connect the power supply as follows using the supplied DC cable (83020-60004).
Connect the power supply to the 15 pin D-connector as follows, without using the DC cable supplied.
+9.0 Vdc (± 0.2 Vdc) 150 mA
−9.0 Vdc (± 0.2 Vdc) 570 mA
From To
+9.0 Vdc green wire
−9.0 Vdc yellow wire
Ground/common black wire
From To
+9.0 Vdc Pin 3
−9.0 Vdc Pin 7
Ground/common Pin 5
8 User’s and Service Guide
70429AK95 Operation
Figure 1 70429AK95 Frequency Divider (top view)
Figure 2 70429AK95 Frequency Divider (end view)
Manual Control Switch
Divider RF Inputs50Ω, SMA-f
Divider RF Outputs50Ω, SMA-f
1.6 GHz Ampl.RF Output50Ω, SMA-f
1.6 GHz Ampl.RF Input 50Ω, SMA-f
DC Power Input andExt.Control Interface15-pin DConnector-f
User’s and Service Guide 9
70429AK95Operation
Figure 3 70429AK95 Frequency Divider Dimensions
10 User’s and Service Guide
70429AK95 Performance Test
Performance TestThe procedures in this section are used to test the70429AK95 Divider Accessory's specified electrical performance.
Typeface Key Conventions
The following key conventions are used throughout this document.
• [HARDKEYS] are labeled front panel keys.
• SOFTKEYS are unlabeled key whose function is indicated on the instrument display
Key sequences for setting up the 70429AK95 use the standard Softkey and [Hardkey] conventions. Softkey represent keys whose label is determined by the instrument’s firmware. [Hardkey] represent keys located on the instrument. For a complete definition of the different keys, refer to your standard 70429A User’s Guide.
Equipment Required
The equipment required for the performance test is listed below:
Equipment Description
Phase Noise Measurement System Agilent E5504A, E5505A or equivalent
Spectrum Analyzer Agilent E4404B or equivalent (with at least a 5 GHz frequency range)
Signal Source Agilent 70427A or N5507A
Signal Generator Agilent E8257D UNX or any PSG with low phase noise option providing +16 dBm at 600 MHz to 8 GHz
Power Supply Agilent E3630A or 6236B
Power Splitter (2 required) Agilent 11667B or mini-circuits ZFSC-2-4 or equivalent
Line Stretcher ARRA DN3448B (Agilent part number 0955-0755)
10 dB Attenuator Agilent 8493A Option 010 or equivalent
6 dB Attenuator Agilent 8493A Option 006 or equivalent
SMA 50 Ω Termination 1810-0118 or equivalent
User’s and Service Guide 11
70429AK95Performance Test
Divided Frequency and Output Level Test
This test measures the frequency and output level of a divided signal. A microwave signal from the Agilent E8257D is input to the divider 70429AK95. The 70429AK95 is set to divide by each of the available divide settings and the output is measured on the Agilent E4404B Spectrum Analyzer.
Figure 4 Divided Frequency and Output Level Test Setup
Step 1. Connect the equipment as shown Figure 4. Set the output of the Agilent E8257D to 4 GHz and +16 dBm. On the spectrum analyzer, press [Instr Preset] > [Center Freq] > 2.5 GHz > [Span] > 5 GHz.
12 User’s and Service Guide
70429AK95 Performance Test
p el
dBm
pply
pply
Step 2. Set the switches and the E8257D as indicated below and record the levels.
A3 A2 A1 B3 B2 B1 Divide By
E8257D Frequency
Output Frequency
Divider Output Level (dBm) ≥ 4 dBm
1.6 GHz AmOutput Lev(dBm) ≥ 14
0 0 0 0 0 0 1 4 GHz 4 GHz Does not a
1 0 0 0 0 0 2 8 GHz 4 GHz Does not a
1 0 0 1 0 0 4 6.4 GHz 1.6 GHz
1 0 0 0 1 0 8 8 GHz 1 GHz
1 0 0 1 1 0 16 8 GHz 500 MHz
1 0 0 0 0 1 32 8 GHz 250 MHz
0 1 0 0 0 1 64 8 GHz 125 MHz
1 1 0 0 0 1 128 8 GHz 62.5 MHz
0 0 1 0 0 1 256 8 GHz 31.25 MHz
User’s and Service Guide 13
70429AK95Performance Test
600 MHz Residual Phase Noise Test
This test measures the residual phase noise of a 70429AK95 Divider. Before proceeding with this measurement, it would be helpful to review the measurement techniques of “Phase Noise Without Using a PLL” in Measurement Definitions, in the Reference Manual supplied with your system.
Refer to the Figure 5 as you step through the “Residual Phase Noise Test Procedure” on page 15. Figure 5 provides a block diagram of the signal paths and external connections and should be used for reference when performing the following procedures.
Figure 5 600 MHz Residual Phase Noise Test
14 User’s and Service Guide
70429AK95 Performance Test
Residual Phase Noise Test Procedure
Step 1. Left-click on System in the drop-down menu and select Server Hardware Connections.
Step 2. Left-click on Define in the drop-down menu and select Limit Lines.
Step 3. Enter the Spec data listed in Table 7 under the Upper Noise Limit tab.
Step 4. Left-click on Define in the drop-down menu and select Measurement.
Step 5. Enter the following settings under the appropriate tab.
Table 7 Offset from the Carrier (dBc/Hz)
1 GHz 1 Hz 100 kHz 1 MHz 100 MHz
Spec −100 −150 −154 −154
Measured Data
User’s and Service Guide 15
70429AK95Performance Test
Table 8 Phase Nose System Setup
Parameter Setting
Type and Range Tab:
Measurement Type Residual Phase Noise (without using a phase locked loop)
Start Offset Frequency 1 Hz
Stop Offset Frequency 100 MHz
FFT Averages 10
FFT Quality High ResolutionSwept Quality Fast
Source Tab:Carrier Frequency (0utput frequent of DUT)
600 MHz
Residual Source Power +16 dBmDetector Input Frequency = Carrier Frequency
Checked
Carrier Frequency = Residual Source Frequency multiplied by (1/1)
Checked
Cal Tab:Derive detector constant from single- sided spur
Select
Known Spur ParametersAmplitude −30 dBcOffset Frequency 50 kHz
Block Diagram Tab:Residual Source ManualDUT in Path No Selection NecessaryPhase Detector AutoPhase Shifter ManualEstablish Quadrature by adjusting the
Phase Shifter
Test Set Tab:Input Attenuation 0 dB
LNA low Pass Filter AutoLNA Gain AutoPLL Integrator Attenuation 0 dB
Downconverter Tab:Input Attenuation AutoIF Gain AutoMicrowave/Millimeter Band Microwave (0 to 26.5 GHz)Reference Chain Reference 10 MHz
Graph Tab:
16 User’s and Service Guide
70429AK95 Performance Test
Title 600 MHz Agilent 70429AK95 SN USxxxxxxxx Residual Phase Noise
Graph Type Single-sideband Phase Noise (dBc/Hz)
X Scale Minimum 1 HzX Scale Maximum 100E+6Y Scale Maximum 0Y Scale Minimum −170Normalize Trace Data to 1 Hz Bandwidth
Table 8 Phase Nose System Setup
Parameter Setting
User’s and Service Guide 17
70429AK95Performance Test
Making the Measurement
Step 1. Left-click on Measure in the drop-down menu and verify that Pause at Connect Diagram is checked. If not, select it.
Step 2. Left-click on Measure in the drop-down menu and select New Measurement.
Step 3. Click Yes in the Do you want to perform a New Calibration and Measurement popup window.
Step 4. The Agilent E5500 Instrument Connections window will appear.
Step 5. Select the Downconverter button.
Step 6. Open the Microwave/Millimeter Band dropdown and select A (26.5 to 40 GHz).
Step 7. Enter 3E10 in the Input Frequency field and click Enter.
Step 8. After several seconds, 3.6 GHz will be displayed in the LO Frequency field and the LO power will be set to +16 dBm.
Step 9. Open the LO Frequency drop-down and select 4.8 GHz. After several seconds the new frequency will be set and the power will be +16 dBm.
Step 10. Close the E5500 Downconverter window by clicking the X in the upper right corner.
Step 11. Connect all the components shown in Figure 5 on page 14. Use short, high quality cables with SMA or Type N connectors as much as possible. BNC cables may be used for the signal generator and spectrum analyzer.
Step 12. Turn On the K95 divider power, set divider A to divide by 2 and divider B to divide by 4.
Step 13. Set the spectrum analyzer to [Center Frequency] > 1 GHz > [Span] > 2 GHz.
Step 14. Verify that a 600 MHz signal appears on the spectrum analyzer at a level in the approximate range of −10 to +10 dBm and that no lower frequency signals appear.
Step 15. Set the spectrum analyzer [Center Frequency] > 600 MHz > [Span] > 200 kHz.
Step 16. Set the Signal Generator to 600.05 MHz and adjust it's level so that the spectrum analyzer displays a spur approximately 50 kHz above the 600 MHz divider output at exactly −30 dBc ± 0.2 dB.
Step 17. Set Signal Generator RF to OFF.
Step 18. Select Meter in the Agilent E5500 Instrument Connections window.
Step 19. Adjust the line stretcher and verify that a reading of +100 mV or −100 mV can be achieved. Then adjust the line stretcher for a reading of 0 v ±2 mV.
Step 20. Select Meter in the Agilent E5500 Instrument Connections window to turn off the meter.
18 User’s and Service Guide
70429AK95 Performance Test
Step 21. Select Continue.
Step 22. When Apply modulation to carrier signal popup appears, turn the Signal Generator RF ON and select Continue.
Step 23. When Remove modulation from carrier signal and connect the DUT popup window appears, turn the Signal Generator RF Off. Set the spectrum analyzer to Single Sweep, attenuation to its maximum and select Continue in the popup displayed by the E5500 software.
Step 24. Allow phase noise measurement to complete.
Step 25. Compare the estimated mean value of the trace with the specification line. If the estimated mean value of the trace falls on or below the specification line, the DUT is regarded as having met the specification.
Step 26. Print out the E5500 phase noise plot.
User’s and Service Guide 19
70429AK95Performance Test
Figure 6 Test Setup Block Diagram
10 MHz 70429AK95 Phase Noise Test Limits
10 MHz LNSG Test Procedure
Using the E5504B Phase Noise Measurement System.
Step 1. Left-click on Define in the drop-down menu and select Limit Lines.
Step 2. Enter the Noise Spec data (Table 9) in the Upper Noise Limit tab.
Step 3. Left-click on Define in the drop-down menu and select Measurement. Refer to Table 10.
Table 9 Phase Noise Test Limits
10 MHz 1 Hz 10 Hz 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz 2 MHz
Noise Spec
−88 −118 −142 −152 −158 −158 −158 −158
Ext 1 Input
70427A
E8257D
10 MHz Oven OutRemove Jumper
50 KHz to 1600 MHzRF Out
100 kHz to 1.6 GHz
70420A
Signal Input
Tune Voltage
Div 16,16
K95DUT Jumper10 dB
20 User’s and Service Guide
70429AK95 Performance Test
Table 10 10 MHz Phase Nose System Setup
Parameter Setting1
Type and Range Tab:
Measurement Type Absolute Phase Noise (without using a phase locked loop)
Start Offset Frequency 1 Hz
Stop Offset Frequency 2 MHz
FFT Averages 10
FFT Quality High Resolution
Swept Quality Fast
Source Tab:
Carrier Source Connected Test Set
Carrier Source Frequency 10 MHz
Carrier Source Power +13 dBm
Detector Input Frequency Auto
Reference Source Power 16 dBm
VCO Tune Constant 0.1 ppm x Carrier0.1 ppm x 10 MHz = 1 Hz/Volt
VCO Tune Range ±5 Volts
VCO Input Resistance 600 Ohms
Cal Tab:
Measure Phase Detector Constant Check
Calculate VCO Tune Constant Check
Verify calculated phase lock loop suppression
Check
If Limit is exceeded Use Theoretical Value
Block Diagram Tab:
Carrier Source Manual
Downconverter No Selection Necessary
Reference Source Manual
Phase Detector Mode Auto
Test Set Tune Voltage Destination Reference Source
VCO Tune Mode DCFM
Test Set Tab:
Input Attenuation 0 dB
LNA Low Pass Filter Auto
LNA Gain Auto
PLL Integrator Attenuation 0 dB
Downconverter Tab:
Input Attenuation Auto
IF Gain Auto
Microwave/Millimeter Band Microwave (0 to 26.5 GHz)
User’s and Service Guide 21
70429AK95Performance Test
Reference Chain Reference 10 MHz
Graph Tab:
Title 10 MHz Agilent 70429AK95 SNxxxxxxxx Phase Noise
Graph Type Single- sideband Phase Noise (dBc/Hz)
X Scale Minimum 1 Hz
X Scale Maximum 100E+6
Y Scale Maximum 0
Y Scale Minimum −170
Normalize Trace Data to 1 Hz Bandwidth
1. Delta settings are indicated with a bold font.
Table 10 10 MHz Phase Nose System Setup
Parameter Setting1
22 User’s and Service Guide
70429AK95 Performance Test
Making the 10 MHz Measurement
Step 1. Left-click on Measure and verify that the Pause at Connect Diagram is checked. If not, select it.
Step 2. Left-click on Measurement in the drop-down menu and select New Measurement.
Step 3. Click Yes in the Do you want to perform a new Calibration and Measurement? popup. The Agilent E5500 Instrument Connections window will appear.
Step 4. On the 70427A rear panel, remove the jumper from the 10 MHz Oven Out to 10 MHz In. Refer to Figure 6 on page 20.
Step 5. Connect the 10 MHz Oven Out on the 70427A rear panel to the Signal Input on the 70420A.
Step 6. Set E8257D to 2.56 GHz at +16 dBm. Select RF ON.
Step 7. Select FM/ Phi M on the E8257D. Set FM Dev to 256 Hz. Select FM Source > Ext 1. Verify Ext Coupling, DC.
Step 8. Select More > Ext Impedance > 600 Ohm. Select More > FM ON.
Step 9. Connect E8257D RF Output through a 10 dB attenuator to 70429AK95 RF IN +.
Step 10. Set both divider sections in 70429AK95 to divide by 16.
Step 11. Connect the jumper from 70429AK95 RF OUT+ to 1.6 GHz Amplifier RF IN. Connect the 70420A Amplifier RF OUT to 50 kHz to 1600 MHz Ref Input.
Step 12. Connect 70420A Tune Voltage Output to E8257D Ext 1 Input.
Step 13. Verify that a beat note is present, using an oscilloscope or counter connected to the Monitor port of the 70420A.
Step 14. Select Continue and allow the Phase Noise Measurement to complete.
Step 15. If the E5500 Manual Tune Pause Point dialog box appears, adjust the E8257D frequency in 10 Hz steps until the beat note frequency is at a minimum. The best way to observe this is to use an oscilloscope on the Monitor port of the 70420A. Select Continue.
Step 16. Compare the estimated mean value of the trace with the noise specification line. If the estimated mean value of the trace falls on or below the specification line, the 70429AK95 is regarded as having met the noise specification. A small noise peak near 1 MHz may be above the specification line. This is acceptable as the peak is not due to the 70429AK95 Divider.
Step 17. Print the E5500 Phase Noise Plot.
Step 18. Connect the jumper that was removed from the rear panel of the 70427A in Step 4.
User’s and Service Guide 23
70429AK95Replaceable Parts
Replaceable PartsThe 7042AK95 Frequency Divider is not serviceable by the customer or the Agilent Service Center. It is set up for repair or replacement by the factory only. However, verification of performance can be provided by your nearest Agilent Service Center. Refer to the list of sales and service offices on Page 30.
24 User’s and Service Guide
70429AK95 Safety and Regulatory Information
Safety and Regulatory Information
Introduction
The 7042AK95 Frequency Divider requires no periodic maintenance. Service of the instrument is done only at the factory and is subject to repair or replacement as necessary upon receipt of the defective 7042AK95.
Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Before Applying Power
Verify that the product is configured to match the available main power source. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.
Connector Care and Cleaning
If alcohol is used to clean the connectors, the power cord to the instrument must be removed. All cleaning should take place in a well ventilated area. Allow adequate time for the fumes to disperse and moist alcohol to evaporate prior to energizing the instrument.
WARNING To prevent electrical shock, disconnect the Agilent Technologies model product from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Declaration of Conformity
For a copy of the manufacturer’s Declaration of Conformity for this apparatus, contact your local Agilent Technologies office or sales representative. Refer to “Contacting Agilent Sales and Service Offices” on page 30.
Statement of Compliance
This instrument has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
User’s and Service Guide 25
70429AK95Safety and Regulatory Information
Shipping Instructions
You must always call the Agilent Technologies Instrument Support Center to initiate service before retuning your instrument to a service office. See “Contacting Agilent Sales and Service Offices” on page 30. Always transport or ship the instrument using the original packaging if possible. If not, comparable packaging must be used. Attach a complete description of the failure symptoms.
Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES-001.Cet appareil ISM est conforme a la norme NMB du Canada.
Compliance with German Noise Requirements
This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3. GSGV Deutschland).
Acoustic Noise Emission/Geraeuschemission
LpA<70 dB Lpa<70 dB
Operator Position am Arbeitsplatz
Normal Operation normaler Betrieb
per ISO 7779 nach DIN 45635 t. 19
26 User’s and Service Guide
70429AK95 Safety and Regulatory Information
Warnings
WARNING The WARNING notice denotes a hazard. It calls attention to a procedure which if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Warnings applicable to this instrument are:
WARNING If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
WARNING This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010-1: 2001.
WARNING No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock do not remove covers.
Cautions
CAUTION The CAUTION notice denotes a hazard. It calls attention to an procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met
Cautions applicable to this instrument are:
CAUTION This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010-1:2001.
User’s and Service Guide 27
70429AK95Safety and Regulatory Information
Electrostatic Discharge Protection
Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the network analyzer. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. To prevent damage to the instrument:
• always have a grounded, conductive table mat (9300-0797) in front of your test equipment.
• always wear a grounded wrist strap (9300-1367) with grounding cord (9300-0980), connected to a grounded conductive table mat, having a 1 MΩ resistor in series with it, when handling components and assemblies or when making connections.
• always wear a heel strap (9300-1126) when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
• always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
• always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port and remove the short from the cable.
Figure 7 shows a typical ESD protection setup using a grounded mat and wrist strap.
Figure 7 ESD Protection Setup
28 User’s and Service Guide
70429AK95 Safety and Regulatory Information
Instrument Markings
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
This symbol indicates that the instrument requires alternating current (ac) input.
This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005. All electric and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE Directive, 2002/96/EC).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is in the STANDBY position.
This symbol indicates that the power line switch is in the OFF position.
Safety Earth Ground. This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.
The CE mark is a registered trademark of the European Community. (If accompanied by a year, it is when the design was proven.)
The CSA mark is a registered trademark of the Canadian Standards Association.
This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.
This is a marking to indicate product compliance with the Canadian Interference-Causing Equipment Standard (ICES-001).
Direct Current.
This is a required mark signifying compliance with an EMC requirement. The C-Tick mark is a registered trademark of the Australian Spectrum Management Agency.
China RoHS regulations include requirements related to packaging, and require compliance to China standard GB18455-2001.
This symbol indicates compliance with the China RoHS regulations for paper/fiberboard packaging.
ICES/NMB-001
User’s and Service Guide 29
70429AK95Contacting Agilent Sales and Service Offices
Contacting Agilent Sales and Service OfficesAssistance with test and measurement needs, and information on finding a local Agilent office are available on the Internet at:http://www.agilent.com/find/assist
If you do not have access to the Internet, contact your field engineer.
NOTE In any correspondence or telephone conversation, refer to the signal generator by its model number and full serial number. With this information, the Agilent representative can determine whether your unit is still within its warranty period.
30 User’s and Service Guide