agilent technologies 1
TRANSCRIPT
April 20, 2015
Agilent Technologies
1
April 20, 2015
Agilent Technologies
2
Improving HPLC Characterization of Biomolecules –Agilent Solutions to Separation Challenges
Paul Dinsmoor
Technical Specialist, Bio-
Columns
Agilent Solutions for Therapeutic Protein Characterization
3
Titer determination and purification (Affinity)
Protein A Bio-Monolith
Protein identification and impurity profiling (RP)
120A Advancebio Peptide Mapping Column and 450A RP mAb Column
Glycan analysis (HILIC)
Advancebio Glycan Mapping Column
Charge variant analysis (IEX)
Bio-IEX 1.7um Column
Aggregation analysis (SEC)
Bio-SEC 3 Column
Agilent Bio-LC Column Portfolio
Agilent Bio-LC Columns
Affinity
Bio-Monolith Protein A
Multiple Affinity Removal System
Reversed Phase
AdvanceBio Peptide Mapping
ZORBAX RRHD 300A 1.8um
Poroshell 300
AdvanceBio
RP mAb
ZORBAX 300SB
ZORBAX Amino Acid Analysis
PLRP-S
HILIC
AdvanceBio Glycan Mapping
ZORBAX RRHD 300-HILIC
Size Exclusion
Bio SEC-3
Bio SEC-5
ProSEC 300S
ZORBAX GF-250
ZORBAX GF-450
Ion Exchange
Bio-Monolith
(QA, DEAE, SO3)
Bio mAb
Bio IEX (SAX, SCX, WAX, WCX)
PL SAX
PL SCX
April 20, 2015
Agilent Technologies
4
Products in red are new!
REVERSED-PHASE BIOCHROMATOGRAPHY
April 10, 2015
Agilent Technologies
5
April 20, 2015
Agilent Technologies
6
April 20, 2015
Agilent Technologies
7
• RP chromatography separates proteins through the interaction of the “hydrophobic foot” of the protein with a non-polar surface of the particle
• RP columns are nearly always based on silica particles
– Mechanical stability, easy to make, surface can be modified, excellent peak shape & efficiency
• Solvent
– Organic modifier: Acetonitrile, isopropanol, methanol
– Ion pair additive: Trifluoroacetic acid (TFA), formic acid
• Gradient separation
Strategy for RP-HPLC Method Development of Proteins and Peptides
April 20, 2015
Agilent Technologies
8
• Start at low pH (acidic mobile phase) and choose the initial column and
conditions
• Initial selection parameters include: pore size (120A, 300A, 450A),
mobile phase, bonded phase, particle size, column length, and internal
diameter
• Obtain best resolution by optimizing: gradient steepness, bonded
phase, temperature, column configuration
• Obtain best recovery by optimizing: Bonded phase, temperature,
sample solubility
• Evaluate alternative columns/ technologies for improved selectivity
and efficiency
Common Challenges with RP Bioseparations
• Poor characterization in the separation results in poor identification
- Need best stationary phase to perfect the separation
- Need high resolution/ efficiency
• Reduced lifetime
• Lack of analytical consistency/ method robustness
• LC system pressure limitations
• Long run times for peptide maps
April 19, 2015
Agilent Technologies
9
Primary Structure Characterization Workflows - mAb
April 20, 2015
Agilent Technologies
10
Improve accuracy andresolution
April 20, 2015
Agilent Technologies
11
Protein/Peptide Separations by Reversed-phase
Larger Molecules = Slower Diffusion
So, a need to decrease the diffusion time for macromolecules!
To improve, we can increase the Diffusion Rate by:
elevating operating temperature decreasing solvent viscosity
and, orCCCCC..
April 20, 2015
Agilent Technologies
13
April 20, 2015
Agilent Technologies
14
ZORBAX 300SB RRHD for Proteins
• Stablebond 300 silica/bonding
• C18, C8, C3, and Diphenyl bonded phase
• 1.8 µm particle size for high resolution
• 1200 Bar pressure limit for UHPLC
• 2.1 x 50 mm and 2.1 x 100 mm
April 10, 2015
Agilent Technologies
15
April 20, 2015
Agilent Technologies
16
Four Rapid Resolution High Definition Phases
17
C18 C8 C3 Diphenyl
Peptide Maps Increasing protein size/hydrophobicitySmall Proteins
Ligand Application
C18 Small intact proteins/peptide maps
C8 Intact proteins
C3 Larger /hydophobic proteins, including MAbs
Diphenyl Unique selectivity
ZORBAX300Å, 1.8 µm
Agilent 1290 Infinity
Family of Four Reversed-Phase Ligands
C3 versus Diphenyl
The Separation Challenge - Does Size Matter?
19
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14
0.16
0.18
0.2
300SB-C18 (300Å)
SB-C18 (80Å)
PW
1/2
Proper pore size selection results in sharper peaks for large molecules
Effect of Pore Size and Molecular Size on Peak Width
0 min 12 0 min 11
NEW! AdvanceBio Peptide Mapping Column for HPLC and UHPLC:
• 2.7 µm Superficially Porous
• 120Å pore size
• 600 bar pressure limit
• 2 µm frit to reduce clogging
Greater analytical confidence: Each batch is tested with a rigorous peptide mix to ensure suitability and reproducibility
Save Time: 2 to 3 times faster than fully porous particles
Increased Flexibility: Highly compatible with TFA and formic acid mobile phases for efficient LC UV and LC/MS analysis
April 10, 2015
Agilent Technologies
20
2.7 um1.7 um
0.5 um
April 20, 2015
Agilent Technologies
21
Quality Assurance Testing with Agilent Peptide Mix
Peptide Mapping Standard P/N# 5190-0583
Each batch of AdvanceBio Peptide Mapping media is
tested with the Agilent Peptide Standard (PN 5190-0583)
to ensure batch to batch reproducibility
min0 2.5 5 7.5 10 12.5 15 17.5 20
mAU
0
10
20
30
40
50
60
70
2.1 x 150mm AdvanceBio Peptide Mapping ColumnMobile phase: A-water (0.1%TFA), B- ACN (0.08%TFA), 40 C, flow: 0.52mL/min
Hydrophilic peptide retention
Narrow Peaks w baseline resolution
Hydrophobic peptide retention
Reduced and fast analysis time
Critical and desired peptide mapping
components to achieve fast, selective and
highly efficient peptide separations across
a wide dynamic range.
Peptide MappingAdvanceBio Peptide Mapping Column Highlights
BSA tryptic digest
Peptide Mapping by LC/MS
April 10, 2015
Agilent Technologies
24
8x10
0
0.75
1.75
2.75
3.75
+ESI TIC Scan Frag=200.0V igg023.d
Counts vs. Acquisition Time (min)1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
40 min. Run, 2.1 x 100 mm
Native peptide
Deamidated form 2
Deamidated form 1
Critical Post Translational Modifications (PTM) Identified in Fast and Slow Analyses
Native peptide
Deamidated form 2
Deamidated form 1
8x10
0
0.6
1.6
2.6
3.6
4.6
433 bar0.6 mL/min10-40% B
+ESI TIC Scan Frag=200.0V igg011.d
Counts vs. Acquisition Time (min)0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13
140 bar0.2 mL/min10-40% B
Heavy Chain Peptide
357-366 and its two
deamidated forms
conserved
14 min. Run, 2.1 x 100 mm
Poroshell 300
• 300Å pore size
• StableBond and Extend chemistry
• Available in SB-C3, SB-C8, SB-C18, and Extend-C18
• 5 µm particle size
April 1, 2015
Agilent Technologies
25
April 20, 2015
Agilent Technologies
26
Column Internal DiameterPorous Particle
Flow Rate RangePoroshell
Flow Rate Range
2.1 mm 0.1 – 0.3 mL/min 0.3 – 3 mL/min
1.0 mm 30 – 60 µL/min 0.08 - 0.75 mL/min
• Very high flow rates can be used effectively with Poroshell columns
High Flow Rates with 2.1 mm id Poroshell for High Resolution and Fast Separations
April 1, 2015
Agilent Technologies
27
12
34
5
67
8
0 0.5 1.0Time (min)
Columns: Poroshell 300SB-C182.1 x 75 mm, 5 µm
MP: A: 0.1% TFAB: 0.07% TFA in ACN
Gradient: 5 – 100% B in 1.0 min.
Flow Rate: 3.0 mL/min.
Temperature: 70 °C
Pressure: 250 bar
Detection: UV 215 nm
Sample:1. Angiotensin II2. Neurotensin3. Rnase4. Insulin5. Lysozyme6. Myoglobin7.Carbonic Anhydrase8.Ovalbumin
Pub No# 5989-9899EN
for complete app note
High Flow Rate, High Sensitivity LCMS
min0 0.5 1 1.5 2 2.5 3 3.5
0
20000000
40000000
60000000
80000000
1E8
pmoles of protein0.5
2.5
5
0.75
1
Mobile Phase Gradient: 20 -
100% B in 5.5 min.
A: water + 0.1% formic acid
B: ACN + 0.1% formic acid
Flow Rate: 0.6 mL/min
Temperature: 80°C
Injection volume: 1 uL
LC/MS: Pos. Ion ESI
Peakwidth: 0.06 min
Sample:Mixture of insulin, lysozyme,
cytochrome C, myoglobin, BSA,
carbonic anhydrase
Column: Poroshell 300SB-C18, 1.0 x 75 mm
NEW! AdvanceBio RP mAb Column for HPLC and UHPLC
Particle
• 3.5um SP
• 450 Å pore diameter
• 600bar pressure limit
• 2um frit to reduce clogging
The optimum high speed, large molecule resolution for use with both HPLC and UHPLC systems
3.5 um3.0 um
0.25 um
Phases
• C4
• SB-C8
• DiphenylThe most popular phases for proteins, plus a unique selectivity
Fast Intact mAb Analysis
April 10, 2015
Agilent Technologies
31
AdvanceBio RP-mAb provides superior peak shape at a lower pressure and resolves more fine detail than a UHPLC protein column from competitor
Method Parameters
Column dimensions: 2.1 x 100 mmMobile phase A: 0.1% TFA in water:IPA (98:2)Mobile phase B: IPA:acetonitrile:MPA* (70:20:10)Flow rate: 1.0 mL/min
Gradient: 10-58% B in 4 min, 1 min wash at 95% B, 1 min re-equilibration at 10% BSample: 5 µL injection of Humanized Recombinant Herceptin Variant IgG1 Intact from Creative Biolabs (1 mg/mL) Temperature: 80 °CDetection: UV @ 254 nm
AdvanceBio RP-mAb C4, 450Å, 3.5 µm490 bar
Competitive C4, 300Å, 1.7 µm910 bar
* MPA = Mobile Phase A
Fast, High Resolution mAb Fragment Analysis
min2 2.25 2.5 2.75 3 3.25 3.5 3.75 4 4.25
mAU
0
200
DAD1 A, Sig=220,8 Ref=off (AEM_PS450_...C_MD\AEM_PS450_FAB-FC_MD_4 2014-09-10 09-02-53\1443508-69-0005.D)
min2 2.25 2.5 2.75 3 3.25 3.5 3.75 4 4.25
mAU
0
200
DAD1 A, Sig=220,8 Ref=off (AEM_PS450_...B-FC\AEM_PS450_IGG1_FAB-FC 2014-09-11 13-54-40\USRIT001297-003.D)
min2 2.25 2.5 2.75 3 3.25 3.5 3.75 4 4.25
mAU
0
200
DAD1 A, Sig=220,8 Ref=off (AEM_PS450_...B-FC\AEM_PS450_IGG1_FAB-FC 2014-09-11 14-37-19\706785-1-000003.D)
min2 2.25 2.5 2.75 3 3.25 3.5 3.75 4 4.25
mAU
0
200
DAD1 A, Sig=220,8 Ref=off (AEM_PS450_...B-FC\AEM_PS450_IGG1_FAB-FC 2014-09-10 11-45-32\CD4F123-0000003.D)
April 10, 2015
Agilent Technologies
32
Method Parameters
Column dimensions: 2.1 x 100 mmMobile phase A: 0.1% TFA in waterMobile phase B: n-propanol/acetonitrile/MPA (80/10/10)Flow rate: 0.8 mL/min
Gradient: 5-40% B in 5 min, 1 min wash at 95% B, 1 min re-equilibration at 10% BSample: 1 µL injection of Fc/Fab, Papain Digested Humanized Recombinant Herceptin Variant IgG1 from Creative Biolabs (2 mg/mL)Temperature: 60 °CDetection: UV @ 220nm
AdvanceBio RP-mAb provides superior peak shape and resolution than other columns designed for protein separations
AdvanceBio RP-mAb C4, 450Å, 3.5 µm
Competitor A Protein C4, 400Å, 3.4 µm
Competitor B WIDEPORE C4, 200Å, 3.6 µm
Competitor C C4-30, 300Å, 2.6 µm
AdvanceBio RP-mAb C4 Separtes Proteins with small differences: Biosimilars in development
33
Using 1.0 % B/mL gradient
RemicadeRemicade clone
Using 2.1 % B/mL gradient
RemicadeRemicade clone
Shift of Fc
Annotated shift of the Fc part implicates differences in hydrophobicity of the Fc part due to a 2-point mutation in the AA sequence of the biosimilar compared to the originator.The shift is observed with either a fast or slow gradient.
Shift of Fc
Fast Intact mAb Analysis
AdvanceBio RP-mAb Diphenyl resolves additional fine detail - the Diphenyl phase is unique to Agilent
Method Parameters
Column dimensions: 2.1 x 100 mmMobile phase A: 0.1% TFA in water/IPA (98/2)Mobile phase B: IPA/acetonitrile/MPA* (70/20/10)Flow rate: 1.0 mL/min
Gradient: 10-58% B in 4 min, 1 min wash at 95% B, 1 min re-equilibration at 10% BSample: 5 µL injection of Humanized Recombinant Herceptin IgG1 Intact from Creative Biolabs (1 mg/mL) Temperature: 80 °CDetection: UV @ 254nm
April 10, 2015
34
min1.7 1.8 1.9 2 2.1 2.2 2.3 2.4 2.5
mAU
0
20
40
60
80
100
120
140
DAD1 H, Sig=254,8 Ref=off (AEM_PS450_...\AEM_PS450_IGG-INTACT_MD_4 2014-08-21 08-02-26\1443508-52-0006.D) DAD1 E, Sig=254,8 Ref=off (AEM_PS450_...\AEM_PS450_IGG-INTACT_MD_4 2014-08-20 09-06-44\1435601-25-0037.D) DAD1 E, Sig=254,8 Ref=off (AEM_PS450_...\AEM_PS450_IGG-INTACT_MD_4 2014-08-19 15-36-09\DIP143501-3-047.D)
min1.7 1.8 1.9 2 2.1 2.2 2.3 2.4 2.5
mAU
-4
-2
0
2
4
6
8
10
12
14
DAD1 H, Sig=254,8 Ref=off (AEM_PS450_...\AEM_PS450_IGG-INTACT_MD_4 2014-08-21 08-02-26\1443508-52-0006.D) DAD1 E, Sig=254,8 Ref=off (AEM_PS450_...\AEM_PS450_IGG-INTACT_MD_4 2014-08-20 09-06-44\1435601-25-0037.D) DAD1 E, Sig=254,8 Ref=off (AEM_PS450_...\AEM_PS450_IGG-INTACT_MD_4 2014-08-19 15-36-09\DIP143501-3-047.D)
AdvanceBio RP-mAb C4AdvanceBio RP-mAb SB-C8AdvanceBio RP-mAb Diphenyl
* MPA = Mobile Phase A
35
Large fibrous proteins
Exploiting chemical stability
PLRP-S:
Features
pH 1-14
Extreme buffer concentrations
High temperature stability
Durable and Resilient
Inherently hydrophobic so does not require a bonded alkyl chain to confer hydrophobicity
Benefits
Acid and base cleanup
Typically < 8M
200°C
Long Lifetimes
Avoids typical silica problems of silanol group
Analysis of very large biomolecules or high speed separations1000A & 4000A pores
RP Summary
Agilent Column Positioning
ZORBAX RRHD 300SB• High resolution UHPLC analysis of proteins, including intact mAbs, and protein
fragments
AdvanceBio Peptide Mapping
• Fast, high resolution HPLC and UHPLC analysis of peptides
Poroshell 300 • Fast, HPLC analysis of large intact proteins, including intact mAbs
AdvanceBio RP mAb• Designed for mAb separations• Fast, high resolution HPLC and UHPLC analysis of intact mAbs and mAb
fragments
PLRP-S • Polymeric for high pH stability, alternate selectivity, and large pores
April 10, 2015
39
AdvanceBio RP-mAb is an addition to the market leading Agilent reversed-
phase bio-column portfolio and complements existing columns
How the New Columns Fit: Agilent Positioning
April 15, 2015
40
Match The Column To System Pressure Capabilities
Agilent Column Particle Pressure Rating Phases
AdvanceBio RP-mAb SPP, 3.5 µm, 450Å 600 bar SB-C8, C4, Diphenyl
ZORBAX RRHD 300SB TPP, 1.8 µm, 300Å 1200 bar SB-C18, SB-C8, SB-C3, Diphenyl
Poroshell 300 SPP, 5 µm, 300Å 400 bar SB-C18, SB-C8, SB-C3, Extend-C18
ZORBAX 300SB TPP, 3.5 & 5 µm, 300Å 400 bar SB-C18, SB-C8, SB-C3, SB-CN
PLRP-STPP, 3, 5, 8 um 100, 300, 1000, and 4000A
400 bar NA
Benefits
April 20, 2015
Agilent Technologies
41
Pain Features and Advantages Benefit
Insufficient resolution
Superficially porous particles and sub2um fully porous particles increase diffusion rate, mass transfer, and efficiency for biomolecules
Improved confidence in analysis results (accuracy)
Long analysis timesAnalysis time can be decreased using fasterflow rates and shorter superficially porous & sub2um fully porous particle columns without loss in efficiency
Improved throughput - reduced costs
Short column lifetimeColumn with robust Poroshell packed bed and with 2 um frit decreases chances of bed-collapse or inlet blockage
Improved resource use - reduced costs
HPLC pressure limitations
Superficially porous particles maintain compatibility with all LC instruments System flexibility
HYDROPHILIC INTERACTION LIQUID CHROMATOGRAPHY(HILIC)
April 10, 2015
Agilent Technologies
42
Why are glycans important to profile? Bio-Therapeutics Are Glycosylated
� Changes in glycosylation may affect drug safety and efficacy.
Reducing Risk, Variation through controlling glycosylation
� The structural characterization and quantitative estimation of glycans is highly essential in biopharmaceutical projects.
FDA Classification & Regulations
� It is tremendously challenging to comprehensively characterize glycan profiles and determine the structures of glycans.
Early stages
Glycans
Fab
Fc
Monoclonal Antibody Therapeutic
Protein
Glycans
April 20, 2015
Agilent Technologies
44
Principles of HILIC
(glycans)
• Glycans are very hydrophilic
• They may be neutral or posses charges that will interact with silica and bonded phase
• Column bonded phases will greatly impact the separation performance of glycans
April 20, 2015
Agilent Technologies
45
Glycan Analysis – Common Challenges
• Very long analysis times
• Instrument limitations
• Difficulty achieving reproducible results
April 19, 2015
Agilent Technologies
46
N-Glycan Mapping
Glycoprotein
N-Glycan
2-AB labelled glycan
HILIC column
LC/MS
HILIC column
LC/FLD (-MS)
Majority process
April 10, 2015
Agilent Technologies
47
Glycan Analysis Workflow
April 10, 2015
48
Glycoprotein
N-Glycans
2-AB Labelled N-glycans
Deglycosylation
Work-up
2-AB Labelling
Work-up
HILIC FLD / MS
Deglycosylation Kit
Description (24 or 96 samples)
Reaction buffer 5X
Denaturant
Detergent
PNGase F
Deglycosylation Work-up
Description (24 or 96 samples)
SPE cartridges
2-AB Labeling Kit
Description (24 or 96 samples)
2-AB solution
Reductant solution
2-AB Labeling Work-up
Description (24 or 96 samples)
SPE cartridges
Glycan Analysis Workflow
April 10, 2015
49
Deglycosylation
Work-up
2-AB Labelling
Work-up
HILIC FLD / MS
Glycoprotein
N-Glycans
2-AB Labelled N-glycans
Unlabelled
Standards
Labelled
Standards
50
April 20, 2015
Agilent Technologies
51
NEW! AdvanceBio Glycan Mapping Column for HPLC and UHPLC:
Glycan Analysis by LC-FLD
April 10, 2015
Agilent Technologies
53
1260 Infinity Bio-inert HPLC
AdvanceBio Glycan Mapping,
2.7 um
AdvanceBio Glycan Mapping,
2.7 um+
AdvanceBio Glycan Mapping,
1.7 um+
1290 Infinity UHPLC
Superficially Porous Glycan Mapping Column –High Speed and Resolution with Less Backpressure
AdvanceBio glycan column delivers faster separations with 1.8 µm particles
April 20, 2015
min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5
LU
0
0.2
0.4
0.6
0.8
1
1.2
FLD1 A, Ex=260, Em=430 (GLYCANAMIDE_RRHD\GLYCANS_RRHD040.D)
16
.00
01
6.0
06
17
.00
5
18
.24
01
8.2
47
18
.56
7
18
.92
6
19
.55
0
20
.58
7
23
.66
1
min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5
LU
0
0.2
0.4
0.6
0.8
1
1.2
FLD1 A, Ex=260, Em=430 (GLYCANAMIDE_RRHD\GLYCANS_RRHD050.D)
9.6
12
10
.53
5
11
.16
9
11
.66
8
12
.01
7
12
.34
8
12
.98
4
13
.39
0
13
.96
6
14
.37
3
17
.07
6
Column RT (min) Rs 2,1 Rs 3,2 Rs 4,3 Rs 6,5 avg. PW (min) Peak Capacity
Waters BEH Glycan 20.2 1.77 1.94 3.39 2.10 0.1085 214
AdvanceBio Glycan 12.7 2.60 2.90 5.43 2.81 0.0741 221
Agilent AdvanceBio
Glycan Mapping, 1.8 um
Waters Acquity
UPLC BEH Glycan
1.7um
Time %B ml/min
0 80 0.50
25 60 0.50
26 0 0.50
27 80 0.50
0.50mL/min, 55° C
Inj = 2uL in 70/30 ACN/water
Mobile phase A: 100mM NH4 Formate pH 4.5,
Mobile phase B: ACN
FLD: Ex=260 nm, Em=430 nm
Sample: Agilent human IgG glycan library
2
1
4
3
5
6
2
1
4
3
5
6
Same condition:• 40% faster• Better Rs• Narrower peaks• Same elution order
Time %A %B Flow rate
(mL/min)
0 25 75 1.0
12 40 60 1.0
12.15 60 40 0.5
12.5 60 40 0.5
12.9 25 75 0.5
13.05 25 75 1.0
15 25 75 1.0
Column Rs 2,1 Rs 3,2 Rs 4,3 Rs 6,5 avg. PW (min) Peak Capacity
AdvanceBio Glycan
Mapping Column, 1.8um
1.63 1.70 3.05 2.09 0.059 135
Rapid Separation of 2-AB labeled N-linked Human IgG glycan Library (Agilent):2.1 x 150mm 1.8um AdvanceBio Glycan Mapping Column
1 2 3 4 5 6 min0 1 2 3 4 5 6 7 8
9
LU
0
0.2
0.4
0.6
0.8
1
0.3
45
0.3
63
1.9
65
2.0
93
2.4
76
2.8
40
2.9
92
3.3
04
3.4
60
3.6
24
3.7
70
3.9
23
4.4
11
4.6
36
5.2
14
5.6
80
6.1
67
6.4
80
7.3
82
7.8
19
7.9
91
2-AB labelling reagent
1
2
4
3
5
10
+
Separation< 10 min
Monoclonal Antibody Glycan Mapping
April 20, 2015
56
mobile phase:A: 100mM NH4 formate
pH 4.5: B: ACNInj.: 2ul 70:30 ACN: waterTemp: 55°CFLD: Excitation 260nm
Emission 430nm
Isomer
separation
HILIC Summary
Characterization Product Features Advantage Benefit Pain Point
Glycan MappingAdvanceBio
Glycan Mapping
1.8um FPP,bonded phase
High efficiency, right selectivity
Improved accuracy and
reproducibility of data – reliable results, cost
saving
Resolution2.7um SPP,
bonded phase
1.8um FPP,bonded phase Efficiency at
higher flow rate, fast gradients
Improved analysis efficiency, cost
savingThroughput
2.7um SPP, bonded phase
FPP: Fully Porous Particle, SPP: Superficially Porous Particle
April 19, 2015
Agilent Technologies
57
SIZE EXCLUSION CHROMATOGRAPHY
April 10, 2015
Agilent Technologies
58
April 20, 2015
Agilent Technologies
59
Aggregation Studies
• ProteinsCaggregate!
• Protein aggregates can greatly enhance the immune responses to the protein of interest
• Protein aggregates can also enhance immune responses and may cause adverse event
• These responses may impact safety and efficacy
• Protein aggregates can be induced under forced conditions and can occur simultaneously during biopharmaceutical manufacturing
April 20, 2015
60
Manufacturing ProcessesSteps and Products• Fermentation• Purification• Formulation• Storage• Shipping• Administration
Stress Conditions• Heat• Freeze-thaw• Cross-linking• Protein concentration• Formulation change – pH, salt• Chemical modification• Mechanical stress / surface
Size Exclusion Process
61
Larger molecules spend less time in the pores and elute sooner.
Smaller molecules spend longer in the pores and elute later.
Mechanism of SEC Separation – Pore Size Determines Linear Separation Range
Some General Guidelines for SEC
1. When methods are to be validated, test for ruggedness with several different column lots, mobile phase preparations, and operators.
2. As a rule of thumb, SEC will only provide baseline separation of molecules with more than a 2 fold difference in MW.
3. Sample volume should be limited to below 5% of the total column volume. Max resolution .5 – 2% CV.
Common SEC challenges
• Insufficient/incorrect pore sizes can reduce resolution
• Non-specific interactions contribute to loss of sample, lead to inconsistent results, rework
• SEC is typically slow
• Consistent and robust results
• High salt conditions puts excessive wear on instrument, parts
April 19, 2015
Agilent Technologies
64
Pore Choice for Antibody Analysis
65
300Å
150Å
100Å
1
2
3
4
5
1. Dimer
2. Monomer
3. Monomer Fragment
4. Azide
5. Retained Molecule
Eluent: 50mM NaH2PO4 + 0.15M NaCl, pH6.8
Columns: Agilent Bio SEC-3 ,4.6x300mm
Flow: 0.35ml/min
Detector: UV@220nm
System: Agilent 1260 Infinity Bio-Inert LC
Sample: Mouse IgG
SEC Column Choice: Resolving Ranges
April 10, 2015
Agilent Technologies
66
Size Exclusion Columns
• 5µm Particle
• 100Å, 150Å, 300Å, 500Å, 1000Å, 2000Å pore sizes
• High stability and long lifetime
• Great reproducibility
• Unique, 3µm particle
• 100Å, 150Å, 300Å pore sizes
• Highest resolution
• Highest efficiency
• Faster SEC separations
Fast, High resolution SEC characterization
Step 1: Improve Resolution
Technology: Small Particle SEC ColumnsResults: Higher Resolution SEC Separations
Use Higher Resolution to Achieve Faster Separations
69
Column: Agilent Bio SEC-3, 7.8 x150mm Sample: mAb (2mg/ml)Injection: 5ulFlow rate: 1.0, 1.5 and 2ml/min (56 bar , 75 bar, 105 bar)Eluent: 150mM sodium phosphate + 100mM Na-sulfateDetection: 220nm
Flow Rate Resolution
Monomer/Dimer
Monomer
Efficiency
Percentage
Dimer
1.0ml/min 1.53 3,510 0.64
1.5ml/min 1.43 2,502 0. 47
2.0ml/min 1.13 1,917 0.64
4 Minutes
monomer
dimer
2.0ml/min 1.5ml/min 1.0 ml/min
Increase flow rate and use a shorter column for faster separations.
Technology:Small Particle SEC Columns
Results:Faster SEC separations
Step 2 – Go Faster
Fast SEC – TB5990-8613EN
April 20, 2015
Agilent Technologies
70
Examples of Additives to reduce non specific interactions
• 100-150mM NaCl
• 100-150mM NaSO4
• 5-10% AcN
• 5-10% MeOH
• 50-100mM urea or Guanidine
• .1% SDS
71
Size Exclusion Summary
Characterization Product Features Advantage Benefit Key Pain Point
Aggregation Bio SEC
3um particles, porosities
High efficiency, right selectivity
Improved accuracy and
reproducibility of data – reliable results, cost
saving
ResolutionHigh efficiency, right MW range
3um particles High efficiencyImproved analysis
efficiency, cost saving
Throughput
April 19, 2015
Agilent Technologies
72
AFFINITYCHROMATOGRAPHY
April 20, 2015
Agilent Technologies
73
AFFINITYCHROMATOGRAPHY
April 19, 2015
Agilent Technologies
74
April 20, 2015
Agilent Technologies
75
What is a Monolith – General Monolith Characteristics
• Continuous material with porous channels that connect to other channels
• Mass transfer is convective (not diffusive) therefore faster -important for biomolecules
• Low back pressure
• High flow rates
April 20, 2015
Agilent Technologies
76
Analytical Bio-Monolith Protein A Columns
Used for:
• Fast screening of harvest cell culture samples for IgG – process optimization
• Accurate analysis of mAb quantities to determine protein harvest
• Capture and purification of protein for further characterization
Features and benefits:
• Bio-Monolith Protein A (immunoaffinity)
• Monolith type material for fast, flow rate independent separations
• Monolith material does not clog easily with cell debris
• Attaches easily to all LCs with standard fittings
April 20, 2015
Agilent Technologies
77
2 Minute Analysis – Antibody Titer from Cell Culture Supernatant
April 20, 2015
Agilent Technologies
78
min0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25
mAU
0
50
100
150
200
250
Lysate proteins containing mAb
wash elute re-equilibrate
IgG1 (2.5 µg)Flow through (8 µg)
Abso
rban
ce (
Abso
rban
ce (
Abso
rban
ce (
Abso
rban
ce (m
AU
mA
Um
AU
mA
Uat
280 m
m)
at 2
80 m
m)
at 2
80 m
m)
at 2
80 m
m)
Time (min)
IgG1
Column Agilent Bio-Monolith Protein ASample: Cell lysate spiked with IgG1Equilibration buffer: 50 mM NaPO4, pH 7.4Elution buffer: 0.1 M citric acid, pH 2.8Flow Rate: 1.0 mL/minDetector: UV 280nmSystem: 1200 Infinity
April 20, 2015
Agilent Technologies
79
ION-EXCHANGE CHROMATOGRAPHY
April 10, 2015
Agilent Technologies
80
Charge variants – What we know?
• Surface charge changed mainly by post-translational modifications- Directly, as a change in the number of charged residues
- Indirectly as a chemical or physical alteration that changes surface-charge distribution
• May result during manufacturing or under storage conditions
• Originates from- Deamidation
- Oxidation
- Glycosylation or glycation
- Proteolytic degradation
- Amino acid substitutions
- Amino acid deletions
81
82
Ion Exchange ChromatographyCharge Isoform Analysis of Monoclonal Antibodies
AU
0.00
0.05
0.10
0.15
0.20
0.25
0.30
0.35
0.40
0.45
0.50
0.55
0.60
Minutes0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00
Column: Agilent Bio MAb, NP5, 4.6mm x 250mmBuffer A: 10 mM Sodium Phosphate, pH 7.50Buffer B: A + 100 mM NaCl, pH 7.50Gradient: 15-95% B in 60 minFlow rate: 0.8 mL/min.Sample: 5 µL, 5 mg/mL, mAb
(Acidic Isoforms)weakly bound
(Basic Isoforms)Strongly bound
Ion Exchange ChromatographySeparation based on charge
Some Guidelines for IEX
1.The General Rule for choosing a Bio IEX column- Acidic proteins: SAX or WAX- Basic proteins: SCX or WCX2. Consider the isoelectric point (pI) of your protein when choosing the pH
of your mobile phase: - If pH>pI, your protein will have a net negative charge- If pH<pI, your protein will have a net positive charge3. The pH of your starting buffer should be 0.5 to 1 pH unit from your pI- Above pI for anion-exchange- Below pI for cation-exchange4. If your pI is unknown- Start with pH 6 for cation-exchange- Start with pH 8.0 for anion-exchange
April 19, 2015
Agilent Technologies
85
Ion-exchange sorbents
Weak ion-exchange sorbents have a narrower operating pH range than strong ion-exchange sorbents
Charge Variant Analysis
Common challenges
• Resolution can be limited and inconsistent – can require troubleshooting and rework
- Need capability to handle complex analyses consistently
- mAbs present special challenges, due to their complexity
• Column contamination can lead to early column failure and produce incomplete sample recovery
• Method development is time-consuming- Need capability for faster, systematic method development with
different buffer strengths
April 19, 2015
Agilent Technologies
87
Ion Exchange – Product Families
88
Particle Porosity Functionalities Particle Sizes Pore Size Application
Agilent Bio-IEX Polymer Non-porous SAX, WAX, SCX, WCX
1.7um, 3um, 5um 10um
N/A Peptides, proteins
Agilent Bio MAb Polymer Non-Porous WCX 1.7um, 3um, 5um 10um
N/A IgG
PL-SAX PS/DVB Fully Porous SAX 5um, 8m, 10um, 30um
1000A, 4000A Peptides, oligos, proteins. Larger column sizes
PL-SCX PS/DVB Fully Porous SCX 5um, 8m, 10um ,30um
1000A, 4000A Peptides, proteins. Larger column sizes
Bio-Monolith IEX Polymer Monolith QA, DEAE, SO3 N/A N/A BioMacromoleucles
Page 46
1. Non-porous particles for high efficiency analytical separations
2. Porous particles for scale up to purification
3. Monoliths for high speed separations
Ion Exchange Columns
• Non-porous PS/DVB particles
• Uniform polymeric coating and WCX layer, specifically designed for antibody separations
• Available in 10 µm, 5 µm, 3 µm, 1.7 µm particle sizes
• Non-porous PS/DVB particles
• Uniform polymeric coating with SCX, WCX, SAX, WAX layers, designed for protein and peptide separations
• Available in 10 µm, 5 µm, 3 µm, 1.7 µm particle sizes
• High surface area
• High capacity
89
SpecificationsParameter Agilent Bio IEX
Particle Size 1.7, 3, 5, and 10um
Binding Capacity > 100 ug protein on column
pH Range 2-12
Temp Limit 80°C
Hardware/Pressure 600 bar for SS
400 bar for PEEK
Pressure Limit 10um – 275 bar (4000 psi)
5um – 413 bar (6000 psi)
3um – 551 bar (8000 psi)
1.7um - 689 bar (10000 psi)
Recommended Flow Rate: 0.1 – 1 mL/min
Mobile phase Compatibility Acetonitrile/acetone/methanol mixtures
90
Optimum Flow Rate for NP3 and NP1.7
NP1.7
Proteins are not small molecules !
Proteins are sluggish !
Proteins are complex three dimensional structures !
Higher efficiency, higher pressure
NP3 NP5 NP10
Bio-WCX Working Flow Rates/ Pressures
Column Dimension Flow Rate Agilent Bio IEX
Bio-WCX NP10 4.6x250mm .8 ml/min 50-60bar
Bio-WCX NP5 4.6x250mm .8 ml/min 150-250bar
Bio-WCX NP 3 4.6x50mm .5 ml/min 80-120 bar
Bio-WCX NP 1.7 4.6x50mm .5 ml/min 110-320 bar
92
Reasons for High Column Pressure ?
93
���� CAUTION: Risk of blockage with NP1.7 is significantly higher.
Agilent Bio IEX ColumnsComparing Separations on Each Particle Size
April 10, 2015
Agilent Technologies
94
Min0 2 4 6 8 10 12 14 16 18 20 22
1.7 µm
3 µm
5 µm
10 µm
Column: Bio WCX, 4.6 x 50 mmBuffer A: 20 mM PBSBuffer B: A+1.0 M NaClGradient: 0-100%B (20 min)Flow rate: 1.0 mL/min for 10 µm, 5 µm, 3 µm
0.75 mL/min for 1.7 µmSample: 1) Ribonuclease A
2) Cytochrome C3) Lysozyme
Concentration: 1.0 mg/mLDetector: 280 nm
Average N ~80,000 for WCX 1.7 µmPeak N
Peak N
Fast, High Resolution IEX Separations
95
Bio WCX NP3, 4.6 x 50 mm Bio WCX NP1.7, 4.6 x 50 mm1.0 mL/min 1.0 mL/min
Technology Used: small particle columns (3um, non porous)Result: 5 minute, high resolution IEX
Standard Protein Separation: Bio WCX
96
Conventional HPLC (400 bar)
1. Ovalbumin pI 4.52. RNase A pI 9.43. Cytochrome C pI 9.84. Lysozyme pI 11
Gradient: 20mM sodium phosphate buffer, pH 6.50-800mM NaCl (0-20 mins)
Flow rate: 1.0mL/minSample: 10µl inj.Detection: UV, 220nm
Column: Bio WCX NP3 (4.6x50mm SS)
Column: Bio WCX NP5 (4.6x250mm SS)1
2
3
4
min0
2.5 5 7.5 10 12.5 15 17.5 20 22.5
mAU
0
100
200
300
400
500
min0
2.5 5 7.5 10 12.5 15 17.5 20 22.5
mAU
0
50
100
150
200
250
300
350
Change a 5um standard to 3um short column:• saves time • maintains resolution
Selectivity Comparison
April 20, 2015
97
- SCX
- WCX
- MAb
1
2
3
1
2
3
1
3
Columns: Agilent Bio SCX, NP 3, 4.6x50 mm, SSAgilent Bio WCX, NP 3, 4.6x50 mm, SSAgilent Bio MAb, NP 3, 4.6x50 mm, SS
Eluent A:10 mM NaH2PO4.2H2O, pH 5.75Eluent B: A + 1 M NaCl Gradient: 0 to 100% b in 25 minsFlow rate: 0.5 mL/minTemperature : AmbientDetector: UV at 220 nm
Sample: 1. Ribonuclease2. Cytochrome C3. Lysozyme
Sample concentration: 2 mg/mLInjection volume: 2 uL
Analytical Bio-Monolith Ion Exchange Columns
• Polymer based monolithic discs
• Fast, high resolution ion-exchange
• Key applications are for large proteins and biomolecules (virus particles, pDNA, antibodies [IgG and IgM])
• Agilent Bio-Monolith QA (strong anion-exchanger)
• Agilent Bio-Monolith DEAE (weak anion-exchanger)
• Agilent Bio-Monolith SO3 (strong cation-exchanger)
April 10, 2015
Agilent Technologies
98
www.agilent.com/chem/getbioguidesApplication-focused Brochures
“How To” Guides
Keys for Enabling Optimum Peptide Characterizations: A Peptide Mapping “How To” Guide5991-2348ENProtein Identification
and Impurity Profiling using Reversed-Phase HPLC/UHPLC5991-0625EN
Resolve Protein Aggregates and Degradants with Speed and Confidence5991-2898EN
Reversed-Phase
Affinity
Ion-Exchange
Characterize Charged Variants of Proteins with Speed and Confidence5991-2449EN
Size Exclusion Chromatography for Biomolecule Analysis: A “How To” Guide5991-3651EN
Selection Guide
Your Reference Guide to the Analysis of Biopharmaceuticals and Biomolecules5990-9384EN
Ion-Exchange Chromatography for Biomolecule Analysis: a “How to” Guide5991-3775EN
Characterize Charged Variants of Proteins with Speed and Confidence5991-2449EN
Size Exclusion
Resources for More Information
April 20, 2015
Agilent Technologies
99
BioHPLC Columns on the Agilent Website
April 20, 2015
Agilent Technologies
100
To learn more and order online visit www.agilent.com/chem/biocolumns
NEED ASSISTANCE?
LC Column help desk1-800-227-9770
• orders• customer service(option 1,1)
• technical support• applications assistance(option 3,3,2)
101
Thank you !
Questions?