The purpose of this exercise is to gain a better understanding of the practical application of the FilterLine™ sampling line, the etCO2 value and the waveform and documentation with the Physio-Control LifePak 12 or LifePak 15 monitor/defribrillators. The exercises will provide you with hands on experience with Microstream™ capnography and the responsiveness of the technology with changing patient conditions.

TRANSCRIPT

  • MICROSTREAM CAPNOGRAPHY EXERCISESPHYSIO-CONTROL LIFEPAK* 12 AND LIFEPAK* 15 MONITOR/DEFIBRILLATORS

    The purpose of this exercise is to gain a better understanding of the practical application of the FilterLine sampling line, the etCO2 value and the waveform and documentation with the Physio-Control LifePak 12 or LifePak 15 monitor/defribrillators. The exercises will provide you with hands on experience with Microstream capnography and the responsiveness of the technology with changing patient conditions.

    NOTE: The following exercises should be performed while wearing Smart CapnoLine Adult non-intubated sampling line and a pulse oximetry sensor. Please perform exercises for hypoventilation, hyperventilation and apnea while sitting in a chair.

    BASELINE ASSESSMENT1. Turn on the Physio-Control monitor.

    2. Plug in the Smart CapnoLine Plus non-intubated oral-nasal sampling line.

    3. Apply the Smart CapnoLine Plus sampling line to your face.

    4.Putthepulseoximetersensoronyourindexfinger.

    a. Monitor your breathing for 60 seconds or until a respiratory rate appears on the Physio-Control LifePak 12 or LifePak 15 monitor/defibrillator.

    b. What is your etCO2 value? __________mmHg

    c. What is your RR? ____________

    d. What is your SpO2%? ____________%

    These etCO2 and SpO2 values are considered your baseline and are important to capture as the starting point with each patient you are monitoring.

    IMPORTANT NOTE: If the Physio-Control LifePak 12 or LifePak 15 monitor/defibrillatordoesnothaveacapnographywaveformonChannel2or3,usetheselector knob to choose the capnography waveform.

    5. Observe the capnography waveform on the screen and press print to collect a CO2 waveform strip.

    HYPOVENTILATION SIMULATION 1. Decrease respiratory rate in order to increase your baseline etCO2 level

    by 5 mmHg.

    2. What is your etCO2 value? __________ mmHg

    3. Observe waveform and capture the event by pressing print.

    Smart CapnoLine sampling line with O2

  • HYPERVENTILATION SIMULATION1. Increase your respiratory rate by breathing rapidly and trying to decrease your

    baseline etCO2 level by 5 mmHg.

    2. What is your etCO2 value? __________ mmHg

    3. Observe the CO2 waveform and capture the event by pressing print.

    APNEA SIMULATION1. With the SpO2probeonfinger,holdbreathtotriggerCO2 apnea alarm (30 seconds).

    2. Observe CO2 waveform, etCO2 value and SpO2 value. Capture the event by pressing print.

    3. Observation: __________________________________________________________

    SAMPLING LINE OCCLUSION SIMULATION1. Kink FilterLine sampling line until Physio-Control LifePak 12 or LifePak 15 monitor/

    defibrillatorindicatesthattheCO2 module is performing a purge. To attempt to clear an occlusion, these monitors purge whenever an occlusion in the line or airway adaptor is detected, and will display PURGING while this function is performed. If the CO2modulecannotcleartheocclusionwithin30seconds,themonitor/defibrillatorwill issue an equipment alert and display the message: FILTERLINE OCCLUSION.

    2. Observe message, then remove the kink from FilterLine sampling line.

    3. Do you observe a CO2 waveform and etCO2 value? If not, disconnect the FilterLine sampling line and reconnect to reactivate the CO2 pump.

    4. If reconnecting fails to produce a CO2 waveform and etCO2 value, replace the FilterLine sampling line.

    5. Observation: __________________________________________________________