capnography vs plethysmography

81
CAPNOGRAPHY VS PLETHYSMOGRAPHY PRESENTED BY DR. SOURAV MONDAL MODERATED BY DR. ARVIND RATHIYA, MD DEPT OF ANAESTHESIOLOGY, S.S.M.C , REWA , M.P.

Upload: sourav-mondal

Post on 05-Apr-2017

62 views

Category:

Health & Medicine


1 download

TRANSCRIPT

Page 1: Capnography vs plethysmography

CAPNOGRAPHY VS

PLETHYSMOGRAPHY

PRESENTED BY DR. SOURAV MONDAL

MODERATED BY DR. ARVIND RATHIYA, MD

DEPT OF ANAESTHESIOLOGY, S.S.M.C , REWA , M.P.

Page 2: Capnography vs plethysmography

CAPNOGRAPHY

Page 3: Capnography vs plethysmography

• Capnography was first introduced by KARL LUFT, a German bioengineer, in 1943 with an infrared CO2 measuring device he called URAS, or “Ultra Rot Absorption Schreiber”

• It was big, heavy and very impractical to use

• Today capnography is successfully used in almost all areas of health care.

HISTORY

Page 4: Capnography vs plethysmography

CAPNOGRAPHY(Quantitative ETCO2 Detectors)

• Capnography is a recording of CO2 concentration versus time.

• It is a form of noninvasive monitoring of the end-tidal carbon dioxide (ETCO2) levels in the patient’s exhaled breath.

Page 5: Capnography vs plethysmography

CAPNOGRAPHY PROVIDES INSTANTANEOUS INFORMATION ABOUT

• VENTILATION– How effectively CO2 is being eliminated by the pulmonary

system

• PERFUSION– How effectively CO2 is being transported through the

vascular system

• METABOLISM– How effectively CO2 is being produced by cellular metabolism

Page 6: Capnography vs plethysmography

STANDARD REQUIREMENTS OF CAPNOMETER

• The CO2 reading shall be within ±12% of the value or ±4mm of Hg (0.53kPa) , whichever is greater .

• The manufacturer must disclose any interference caused by ethanol, acetone , methane , helium , tetrafluoroethane, and dicholorodifluromethane as well as commonly used halogenated anesthetic agents.

• The capnometer must have a high CO2 alarm for both inspired and exhaled CO2.

• An alarm for low exhaled CO2 is required.

Page 7: Capnography vs plethysmography

METHODS OF MONITORING Mass Spectrometry

Infrared Spectrography – Most commonly used in the hospital setting

Chemical Colorimetric Analysis – Most commonly used in the pre-hospital setting

Page 8: Capnography vs plethysmography

MASS SPECTROMETRY• It is a technique by which

concentrations of gas in a sample can be determined according to charge- mass ratio.

• A gas sample is passed through an ionizer and molecules become positively charged ions. Because all of the ions generated carry the same positive charge , this allows separation based solely on mass.

• A detector then counts the no.of ions of each mass & the results are translated into concentrations.

Page 9: Capnography vs plethysmography

INFRARED SPECTROGRAPHYPRINCIPAL: Carbon Dioxide selectively absorbs a known amount of infrared light of a specific wavelength (4.26 µm). The amount of light absorbed is directly proportional to the concentration of carbon dioxide molecules.A predetermined amount of infrared light is sent from the emitting side of the sensor through a gas sample and collected on the receiving side of the sensor. The infrared light received is compared to the infrared light transmitted. The difference is then converted by calculations into either partial pressure or percentage of total gas concentration that we see on the monitor.

Infrared Spectrography

Page 10: Capnography vs plethysmography

CHEMICAL COLORIMETRIC ANALYSIS

CHEMICAL COLORIMETRY ANALYSIS

CO2 NOT PRESENT CO2 PRESENT

Page 11: Capnography vs plethysmography

•Consists of pH-sensitive indicator enclosed in housing. When the indicator is exposed to carbonic acid it becomes more acidic & changes colour. During inspiration the colour returns to resting state.

•The inlet and outlet ports are 15mm, so the device can be placed between patient and the breathing system or resusciation bag

TECHNOLOGY

HYGROSCOPIC – contains hygroscopic filter paper that is impregnated with a colourless base & an indicator that changes colour as a function of pH. A purple/mauve colour indicates low (<0.5%) CO2 level, beige colour indicates a moderate (0.5%- 2%) level while a yellow colour indicates high(>2%) level

HYDROPHOBIC – shows a colour change from blue to green to yellow when exposed to CO2. Liquid water may cause improper functioning of the device.

Page 12: Capnography vs plethysmography

TYPES OF CAPNOMETERS

The 2 Types of Capnometers

Mainstream Sidestream

The infrared sensor is in the direct path of the gas source, and connected to the monitor by an electrical wire.

The sample of gas is aspirated into the monitor via a lightweight airway adapter and a 6ft length of tubing. The actual sensor is inside the monitor.

Page 13: Capnography vs plethysmography

MAINSTREAM CAPNOMETERS ADVANTAGES

Mainstream – Advantages

Mainstream

No sampling tube to become obstructed.

No variation due to barometric pressure changes.

No variation due to humidity changes.

Direct measurement means waveform and readout are in ‘real-time’. There is no sampling delay.

Suitable for pediatrics and neonates.

Page 14: Capnography vs plethysmography

MAINSTREAM CAPNOMETERSDISADVANTAGES

Mainstream – Disadvantages

Mainstream

The airway adapter sensor puts weight at the end of the endotracheal tube that often needs to be supported.

In older models there were minor facial burns reported.

The sensor windows can become obstructed with secretions and water rainout.

Sensor and airway adapter can be positional – difficult to use in unusual positions (prone, etc).

Page 15: Capnography vs plethysmography

SIDESTREAM CAPNOMETERSADVANTAGES

Sidestream – Advantages

Sampling capillary tube and airway adapter is easy to connect.

Can be used with patient in almost any position (prone, etc).

Can be used in awake patients via a special nasal cannula.

CO2 reading is unaffected by oxygen flow through the nasal cannula.

Page 16: Capnography vs plethysmography

SIDESTREM CAPNOMETERSDISADVANTAGES

Sidestream – Disadvantages

The sampling capillary tube can easily become obstructed by water or secretions.

Water vapor pressure changes within the sampling tube can affect CO2 measurement.

Delay in waveform and readout due to the time it takes the gas sample to travel to the sensor within the unit.

Page 17: Capnography vs plethysmography

PHYSIOLOGY OF CAPNOGRAPHY• During cellular respiration, small

amounts of CO2 produced , is excreted via exhalation

• When no cellular respiration is occurring, even if ventilation is, there will be no CO2 exhaled– In poor perfusion states

(cardiac arrest) no CO2 is transported to the lungs to be exhaled, so a low reading will occur

– In poor ventilation states (hypoventilation) CO2 is retained, so a high reading will occur

Page 18: Capnography vs plethysmography

END-TIDAL CO2 The peak partial pressure of CO2

during exhalation (the highest level of expired CO2 reached during exhalation) is known as the end-tidal CO2 (ETCO2).– Normally occurs at the end

of the alveolar plateau

ETCO2 is a reflection of alveolar ventilation, CO2 production and pulmonary blood flow.

Normal value is 35-45 mmHg

Page 19: Capnography vs plethysmography

CLINICAL APPLICATION OF ETC02

• Verification of endotracheal tube placement

• Continuous monitoring of tube location during transport

• Gauging the effectiveness of resuscitation and prognosis during cardiac arrest

• Titrating ETC02 levels in patients with suspected increases in intracranial pressure

• Determining adequacy of ventilation

Page 20: Capnography vs plethysmography

WAVEFORM DISPLAYS(Quantitative Device)

The waveform is divided into 4 phases.

Phases I, II and III occur during and reflect the three phases of exhalation.

Phase IV occurs during and reflects inspiration

Page 21: Capnography vs plethysmography

CAPNOGRAM: PHASE I

Phase I (A-B) occurs during exhalation of air from the anatomic dead space, which normally contains no CO2.

This part of the curve is normally flat, providing a steady baseline.

Page 22: Capnography vs plethysmography

CAPNOGRAM: PHASE II

Phase II (B-C) occurs during alveolar washout and recruitment, with a mixture of dead space and alveolar air being exhaled.

Phase II normally consists of a steep upward slope.

Page 23: Capnography vs plethysmography

CAPNOGRAM: PHASE III Phase III (C-D) is the alveolar

plateau, with expired gas coming from the alveoli.

In patients with normal respiratory mechanics, this portion of the curve is flat, with a gentle upward slope.

The highest point on this slope (D) represents the ETCO2 value.

Page 24: Capnography vs plethysmography

CAPNOGRAM: PHASE IV

Phase IV (D-E) occurs during inspiration, where the ETCO2 level normally drops rapidly to zero.

Unless CO2 is present in the inspired air, as occurs when expired air is rebreathed , this part of the waveform is a steep, downward slope.

Page 25: Capnography vs plethysmography
Lenovo
Page 26: Capnography vs plethysmography

VOLUME CAPNOGRAM It is a graphic display of CO2 concentation/ partial pressure

versus exhaled volume. The inspiratory phase in not defined in volume capnogram.

ADVANTAGES

Allows estimation of relative contributions of anatomical dead space and alveolar dead space

More sensitive Allows for determination of total mass of CO2 exhaled during a

breath & provides estimation of VCO2

Page 27: Capnography vs plethysmography

ABNORMAL CAPNOGRAMS

Page 28: Capnography vs plethysmography

CAUSES OF AN ELEVATED ETCO2

• ↑CO2 PRODUCTION & DELIVERY TO LUNGS

↑metabolic rate Fever Sepsis Seizures Malignant hypothermia Thyrotoxicosis ↑Cardiac output (during

CPR) Bicarbonate administration

• ↓ALVEOLAR VENTILATION Hypoventilation Respiratory centre depression Partial muscular paralysis Neuromuscular disease COPD

• EQUIPMENT MALFUNCTION Rebreathing Exhausted C02 absorber Leak in ventilator circuit Faulty inspiratory/ expiratory valve

Page 29: Capnography vs plethysmography
Page 30: Capnography vs plethysmography

CAUSES OF A DECREASED ETCO2

• ↓ CO2 PRODUCTION AND DELIVERY TO LUNGS

Hypothermia Pulmonary hypoperfusion Cardiac arrest Pulmonaryembolism Haemorrhage Hypotension

• ↑ALVEOLAR VENTILATION Hyperventilation

• EQUIPMENT MALFUNCTION Ventilator disconnect Esophageal intubation Complete airway obstruction Poor sampling Leak around ETT cuff

Page 31: Capnography vs plethysmography
Page 32: Capnography vs plethysmography
Page 33: Capnography vs plethysmography
Page 34: Capnography vs plethysmography
Page 35: Capnography vs plethysmography
Page 36: Capnography vs plethysmography
Page 37: Capnography vs plethysmography

INCOMPETENT INSPIRATORY UNIDIRECTIONAL VALVE

• The waveform shows a prolonged plateau & a slanting inspiratory downstroke

• The inspiratory phase is shortened & the baseline may or may not reach zero.

Page 38: Capnography vs plethysmography

LEAK IN SAMPLING LINE DURING PPV

• Will result in upswing at the end of Phase III.

• The brief peak is caused by the next inspiration , when positive pressure pushes undiluted end-tidal gas through the sampling line.

Page 39: Capnography vs plethysmography
Page 40: Capnography vs plethysmography

PULSE OXIMETRY

(PHOTOPLETHYSMOGRAPHY+ OXIMETRY)

Page 41: Capnography vs plethysmography

PULSE OXIMETRY

Pulse oximeters combines the principle of oximetry and plethysmography to noninvasively measure oxygen saturation in arterial blood

Page 42: Capnography vs plethysmography

HISTORY

• In early 1940 GLEN MALKIKAN coined the term oximeter.

• MATHEES- father of oximetry 20 papers in1934 –1944• HERTZMAN 1937 –use of photoelectric finger

plethsmography• 1975 –concept of pulse oximetry –Japan• In 2008 modification continued and term High

Resolution Pulse Oximetry come into existence

Page 43: Capnography vs plethysmography

INTRODUCTION• Also called the fifth vital sign

• Low SpO2 provide warning of hypoxemia before other signs such as cyanosis or a change in heart rate are observed.

• Until the 1980s, noninvasive oximeters, known oximeters, were large, expensive, and cumbersome. They required “arterialization”.

• Technical advances, including LEDs, miniaturized photodetectors,

and microprocessors, allowed the creation of a new generation of oximeters.

• These differentiate the absorption of light by the pulsatile arterial component from the static components, so they are called Pulse Oximeters.

Page 44: Capnography vs plethysmography

OXYGEN SATURATION

• Saturation is defined as ratio of O2 content to oxygen capacity of Hb - expressed as percentage.

• Desaturation leads to Hypoxemia – a relative deficiency of O2 in arterial blood. (PaO2 < 80mmHg – hypoxemia)

• (SaO2< than 76% is life threatening.)

Page 45: Capnography vs plethysmography

FRACTIONAL SATURATION

This is the ratio of oxygenated Haemoglobin to sum of all haemoglobin species in blood.

Fractional saturation = HbO2 - --------------------------

HbO2+ Hb+ Met Hb +CO Hb

Page 46: Capnography vs plethysmography

FUNCTIONAL SATURATION

This is the of ratio between HbO2 to all the functional haemoglobin

Page 47: Capnography vs plethysmography

PLETHYSMOGRAPHY• Pulse oximeters show pulsatile change in

absorbance in a graphical form. This is called the “plethysmographic trace” or “pleth”

• Because absorption of light is proportional to the amount of blood between the transmitter & the photodetector , changes in the blood volume are reflected in pulse oximetry trace. Hence pulse oximeter can also be used as a photoplethysmograph

Page 48: Capnography vs plethysmography

PRINCIPLES • All atom and molecules absorb specific wavelength of light. This

property is the basis for an optical technique known as SPECTROPHOTOMETRY.

BEER-LAMBERT LAW It states that if a known intensity of light illuminates a chamber of

known dimensions , then the concentration of a dissolved substance can be determined if the incident and transmitted light is measured :

It = Iie –dcα

Solved for C,C= (1/dα)ln[Ii/It]

Page 49: Capnography vs plethysmography

Substances have a specific pattern of absorbing specific wavelength – EXTINCTION COEFFICIENT

Uses two lights of wavelengths660nm –deoxy Hb absorbs ten times as oxy Hb940 nm – absorption of oxyHb is greater

Lab oximeters use 4 wavelengths to measure 4 species of haemoglobin

Page 50: Capnography vs plethysmography

OPERATING PRINCIPLES

• The pulse oximeter computes the ratio between the two signals and relates this ratio to the arterial oxygen saturation, using an empirical algorithm.

• Pulse oximeters discriminate between arterial blood and other components.

• The oximeter pulses the red and infrared LEDs ON and OFF several hundred times per second

Page 51: Capnography vs plethysmography

• The rapid sampling rate allows recognition of the peak and trough of

each pulse wave.

• At the trough, the light is transmitted through a vascular bed that contains mainly capillary and venous blood as well as intervening tissue.

• At the peak, it shines through all these plus arterial blood.

• A photodiode collects the transmitted light and converts it into electrical signals.

• The emitted signals are then amplified, processed, and displayed on the monitor.

Page 52: Capnography vs plethysmography

ACCURACY

• A clinically acceptable level of arterial oxygenation(SaO2 above 70%),the oxygen

saturation recorded by pulse oximeters (SpO2) differs by less than 3% from the actual saturation.

• Pulse oximetry also show a high degree consistency of repeated measurements.

Page 53: Capnography vs plethysmography

PHYSIOLOGY OF PULSE OXIMETRY• Pulse Oximetry uses light to work

out oxygen saturation. Light is emitted from light sources which goes across the pulse oximeter probe and reaches the light detector.

• If a finger is placed in between the light source and the light detector, the light will now have to pass through the finger to reach the detector. Part of the light will be absorbed by the finger and the part not absorbed reaches the light detector.

Page 54: Capnography vs plethysmography

The amount of light absorbed depends on the following:

concentration of the light absorbing substance.

length of the light path in

the absorbing substance

Page 55: Capnography vs plethysmography

TYPES OF OXIMETRYTRANSMISSION PULSE OXIMETRY• light beam is transmitted through

a vascular bed and is detected on opposite side of that bed

REFLECTANCE PULSE OXIMETRY• relies on light that is reflected to

determine oxygen saturation. The probe have the emitters & detectors on the same side.

advantage - its signal in low perfusion is better.

limitations -The probe design must eliminate light that is passed directly to the probe or is scattered in the outer surface of the skin. The signals are weaker.

Page 56: Capnography vs plethysmography

EQUIPMENT

PROBES

• The probe (sensor, transducer) comes in contact with the patient.

• It contains one or more LEDs (photodiodes) that emit light at specific wavelengths and a photodetector.

• The LEDs provide monochromatic light. • Probes may be reusable or disposable.

Page 57: Capnography vs plethysmography

• Self-adhesive probes are less likely to come off if the patient moves.

• Probes are available in different sizes.

• The photocell should be aligned with the probe

• Contamination should be reduced.

Page 58: Capnography vs plethysmography

CABLE• The probe is connected to the oximeter by an

electrical cable.

CONSOLE• Many different consoles are available . Most

oximeters that are used in the operating room are part of a physiologic monitor.

• Most stand-alone units are line operated but will work on batteries, making them useful during transport.

Page 59: Capnography vs plethysmography

• Most instruments provide an audible tone whose pitch changes with the saturation.

• Alarms are commonly provided for low and high pulse rates & for low and high saturation.

• ASA standards for Basic Anesthetic Monitoring require that the variable pitch pulse tone and low threshold alarm be audible.

Page 60: Capnography vs plethysmography

OXIMETER STANDARDS

• There must be a means to limit the duration of continuous operation at temperature above 41°C .

• The accuracy must be stated over the range of 70% to 100% SpO2.

• There must be an indication when the SpO2 or pulse rate data is not current.

Page 61: Capnography vs plethysmography

• It must be provided with an alarm system

• There must be an alarm for low SpO2 that is not less than 85% SpO2 .

• An indication of signal inadequacy must be provided if the SpO2 or pulse rate value displayed is potentially incorrect.

Page 62: Capnography vs plethysmography

SITES OF PROBE PLACEMENTFingers ToeEar Nose Tongue Cheek Esophagus Forehead

Page 63: Capnography vs plethysmography

MISC• Pharyngeal pulse oximetry by using a pulse oximeter

attached to a laryngeal mask may be useful in patients with poor peripheral perfusion.

• Flexible probes may work through the palm, foot, penis, ankle, lower calf, or even the arm in infants

• Pulse oximetry may be used to monitor fetal oxygenation during labor by attaching a reflectance pulse oximetry probe to the presenting part . A disadvantage is that the probe has to be placed blindly and may be positioned over a subcutaneous vein or artery, which will affect the reliability of the readings

Page 64: Capnography vs plethysmography

USES• Monitoring oxygenation:

PACU, transport.• Detect inadvertent bronchial

intubation • Managing one lung

anaesthesia• Weaning from artificial

ventilation• Controlling 02 administration• Monitoring peripheral

circulation• Determining systolic blood

pressure

• Locating vessels ( eg: axillary artery)

• Avoiding hyperoxaemia• Monitoring vascular

volume• Monitoring sympathetic

tone(dicrotic notch)• Pulse rate• Arrhythmias• Neonatal care(one element

of suggesting congenital heart diseases)

Page 65: Capnography vs plethysmography

ADVANTAGES

• Accuracy• Independence from gases and vapours• Fast response time• Non invasive• Continuous Measurements• Separate Respiratory and circulatory variables• Convenience• Fast start time

Page 66: Capnography vs plethysmography

• Tone modulation• User –friendliness• Light weight & compactness• No heating required• Battery operated• Probe variety• Economy

Page 67: Capnography vs plethysmography

LIMITATIONS AND DISADVANTAGESFailure to Determine the Oxygen Saturation Factors that are reported to contribute to higher failure rates

include ASA physical status III , IV or V patients, orthopedic, vascular, and cardiac surgery; electrosurgery use; hypothermia; hypotension; low hematocrit and motion

Poor Function with Poor Perfusion Readings may be unreliable or unavailable if there is loss or

diminution of the peripheral pulse (proximal blood pressure cuff inflation, external pressure, improper positioning, hypotension, hypothermia, Raynaud's phenomenon, low cardiac output, hypovolemia, peripheral vascular disease).

Page 68: Capnography vs plethysmography

Erratic Performance with Dysrhythmias Double- or triple-peaked arterial pressure

waveform that confuses the pulse oximeter, so it may not provide a reading

Carboxyhaemoglobin COHb has an absorption spectrum similar to

that of 02Hb at 660nm, so most pulse oximeters give falsely elevated readings.

Page 69: Capnography vs plethysmography

Methaemoglobin Methaemoglobin absorbs light equally at the red and infrared

wavelengths that are used by most pulse oximeters. When compared with functional saturation, most pulse oximeters give falsely low readings for saturations above 85% and falsely high values for saturations below 85% .

Sulfhaemoglobin• Sulfhaemoglobinemia may be caused by drugs such as

metoclopramide , phenacetin, dapsone and sulfonamides. Sulfhaemoglobin causes the pulse oximeter to display artifactually low oxygen saturation

Page 70: Capnography vs plethysmography

Mixing Probes SpO2 measurements may not be accurate if one

manufacturer's probe is used with a different manufacturer's instrument

Bilirubinemia Severe hyperbilirubinemia can cause an

artifactual elevation of metHb and carboxyhaemoglobin when using in vitro oximetry but does not affect pulse oximetry readings.

Page 71: Capnography vs plethysmography

Low Saturations Pulse oximetry becomes less accurate at low oxygen

saturations . This inaccuracy is greater in patients with dark skin. It should be used with caution in patients with cyanotic heart disease.

Malpositioned Probe• Prominent pulsations of venous blood may lead to

underestimation of the SpO2. • High airway pressures during artificial ventilation may cause

phasic venous congestion, which may be interpreted by the oximeter as a pulse wave.

Page 72: Capnography vs plethysmography

Nail polish and coverings• All colour of nail polishes especially black, dark blue & purple may cause

significantly lower saturation readings

Electrical Interference• Electrical interference from an electrosurgical unit can cause the oximeter

to give an incorrect pulse count . • Steps to minimize electrical interference include – keeping the oximeter

probe & console as far from the surgical field as possible; locating electrosurgery ground plate as close as possible to the surgical site ;not plugging in the electrosurgical apparatus and pulse oximeter into the same power circuit

Severe Anaemia -The pulse oximeter may overestimate SpO2, especially at low saturations, in patients with severe anemia.

Skin Pigmentation -pigmentation does not make a significant difference in pulse oximeter accuracy

Page 73: Capnography vs plethysmography

PROBLEM OF MOVEMENT

• Pulse oximeters are very vulnerable to motion, such as a patient moving his hand. As the finger moves, the light levels change dramatically. Such a poor signal makes it difficult for the pulse oximeter to calculate oxygen saturation.

• Lengthening the averaging time will increase the likelihood that enough true pulses will be detected to reject the motion artifacts.

Page 74: Capnography vs plethysmography

PROBLEM OF OPTICAL SHUNTING

• If the probe is of the wrong size or has not being applied properly, some of the light , instead of going through the artery, goes by the side of the artery (shunting).

• This reduces the strength of the pulsatile signal making the pulse oximeter prone to errors. It is therefore important to select the correct sized probe and to place the finger correctly in the chosen probe for best results.

Page 75: Capnography vs plethysmography

PROBLEMS OF TOO MUCH AMBIENT LIGHT

• If the ambient light is too strong, the LED light signal gets "submerged" in the noise of the ambient light. This can lead to erroneous readings. Therefore, it is important to minimise the amount of ambient light falling on the detector.

Page 76: Capnography vs plethysmography

PROBLEM OF NOT DETECTING HYPEROXIA

• Haemoglobin is not the only way oxygen is carried in blood. Additional oxygen can also be dissolved in the solution in which red blood cells travel (plasma).

• The problem is that the pulse oximeter cannot "see" the extra dissolved oxygen. So even if the patient’s blood contains extra oxygen, the saturation will still show 100 %.

Page 77: Capnography vs plethysmography

METHODS TO IMPROVE SIGNALS

• Application of vasodilating cream• Digital nerve block• Administration of intraarterial vasodilators• Placing a gloves filled with warm water over patient

hand.• Warming cool extremities• Trying an alternative probe site.• Trying a different probe• Trying a different machine

Page 78: Capnography vs plethysmography

PATIENT COMPLICATIONS

• Corneal Abrasions

• Pressure and Ischemic Injuries

• Burns

• Electric Shock

Page 79: Capnography vs plethysmography

CAPNOGRAPHY VS PLETHYSMOGRAPHY –

a comparison & conclusion

Page 80: Capnography vs plethysmography

• Oxygenation is monitored by pulse oximetry whereas Ventilation is monitored with capnography

• Oximeters measure saturated haemoglobin in peripheral blood and provide additional information about the adequacy of lung perfusion and oxygen delivery to the tissues. However, pulse oximetry is a late indicator of O2 supply, and is less sensitive than capnography. It does not afford a complete picture of ventilatory status.

• Capnography continuously and nearly instantaneously measures pulmonary ventilation and is able to rapidly detect small changes in cardio-respiratory function before oximeter readings change.

• Hypoventilation & hypercarbia may occur without a decrease in Hb O2 saturation, so pulse oximeter cannot be relied on to detect leaks, disconnections or esophageal intubations ; whereas capnography can be reliably used in these conditions.

Page 81: Capnography vs plethysmography

THANK YOU