Capnography 102Capnography 102

June 2011Bill Streett-Training Section Chief

Page 2: Capnography 102 - SCFR

TerminologyTerminology

Capnography: the measurement and numerical display of end-tidal CO2 concentration, at the patient’s airway, during a respiratory cycle.

Page 3: Capnography 102 - SCFR

What is Capnography?What is Capnography?

� Definition:

The measurement of carbon dioxide (CO2) in an exhaled breath.in an exhaled breath.

-The vital sign of ventilation.

Page 4: Capnography 102 - SCFR

TerminologyTerminology

Capnograph: the graphical representation of the

concentration or partial pressure of expired CO2

during a respiratory cycle in a “waveform” format.

Page 5: Capnography 102 - SCFR

RespiratoryRespiratorySystemSystem

Page 6: Capnography 102 - SCFR

The respiratory system’s main function:The respiratory system’s main function:

ExhalationExhalation

InhalationInhalation

Page 7: Capnography 102 - SCFR

Child vs. Adult AirwayChild vs. Adult Airway

Page 8: Capnography 102 - SCFR

Benefits of Capnography UseBenefits of Capnography Use

� Verifying endotracheal vs. esophageal intubations

� Continuous monitoring of the airway, post intubation and during transportContinuous monitoring of the airway, post intubation and during transport

� Accessing effectiveness of CPR

� Decision tool for terminating resuscitation efforts in PEA and/or Asystole.

� Assessing cardiac output in pt.s in PEA

Page 9: Capnography 102 - SCFR

Benefits of Capnography Use cont.Benefits of Capnography Use cont.

� Help assess the severity of respiratory distress (CO2 retention)distress (CO2 retention)

� Help assess the degree of circulatory failure from shock (of any cause)

Page 10: Capnography 102 - SCFR

2010 ACLS Standards regarding 2010 ACLS Standards regarding

Waveform CapnographyWaveform Capnography

-Assessing CPR Effectiveness…� It is reasonable to consider using quantitative

waveform capnography in intubated patients to monitor CPR quality, optimize chest compressions, and detect ROSC during chest monitor CPR quality, optimize chest compressions, and detect ROSC during chest compressions or when rhythm check reveals an organized rhythm (Class IIb, LOE C )

-Predicting ROSCIf PETCO2 abruptly increases to a normal value

(35 to 40 mm Hg), it is reasonable to consider that this is an indicator of ROSC (Class IIa, LOE B).

Page 11: Capnography 102 - SCFR

Types of CO2 MonitorsTypes of CO2 Monitors

� Colorimetric Monitors

� Infrared Monitors

Page 12: Capnography 102 - SCFR

Colorimetric Monitoring TechnologyColorimetric Monitoring Technology

� Uses a hygroscopic disc coated with a substance that reacts with CO2 to form H+ ions.

� This occurs when CO2 is exhaled, causing the pH to decrease changing the disc from purple to tan.

CO2 concentration changes from .03-.05%, when disc is purple,

to 2.0-5.0% with exhaled air from the trachea – turning the disc tan.

Page 13: Capnography 102 - SCFR

Colorimetric Monitoring TechnologyColorimetric Monitoring Technology

cont.cont.

� Color change should be evaluated after the sixth ventilation.

� Does not work if it becomes wet. becomes wet.

Be careful to monitor how well device is working if fluid in the ET tube or in pt.’s with pulmonary edema!

Page 14: Capnography 102 - SCFR

Infrared Monitoring Infrared Monitoring

� Uses a variety of different monitoring technology.

� Measures the percentage of CO2 that is present through the third phases of expiration cycle.through the third phases of expiration cycle.

� Based on CO2 diffusion from pulmonary arterial blood carried to the pulmonary capillary beds of the alveoli – where gas exchange occurs.

� Dependant upon adequate circulation and pulmonary perfusion.

Page 15: Capnography 102 - SCFR

Infrared Monitoring TechnologyInfrared Monitoring Technology

Mainstream� Sensor located

directly in pt.’s breathing circuitbreathing circuit

� Used primarily on intubated patients.

� Sensor has a longer warm up time before gas sample is analyzed.

Page 16: Capnography 102 - SCFR

Infrared Monitoring TechnologyInfrared Monitoring Technology

Sidestream

� Sample is removed from pt.’s airway and delivered to a distant delivered to a distant sensor.

� Can be used on non-intubated patients.

� Min. sample volume 100cc – 150cc.

Page 17: Capnography 102 - SCFR

Infrared Monitoring TechnologyInfrared Monitoring Technology

� Microstream� Sample is obtained

from center of the lumen of airway adapter.adapter.

� Can be used on intubated and non-tubated patients

� Min. sample vol. 50 cc

� Sampling device is disposable.

Page 18: Capnography 102 - SCFR

Analyzing the CO2 WaveformAnalyzing the CO2 Waveform

The CO2 waveform can be analyzed from five basic characteristics:

1. Height

2. Frequency

3. Rhythm

4. Baseline

5. Shape

Page 19: Capnography 102 - SCFR

The Normal / End Tidal ValueThe Normal / End Tidal Value

35 - 38 mm HG or 5% ETPCO2

1. Height is determined by end tidal CO2 value

2. Frequency determined by respiratory rate

35 - 38 mm HG or 5% ETPCO2

1. Height is determined by end tidal CO2 value

2. Frequency determined by respiratory rate2. Frequency determined by respiratory rate

3. Rhythm determined by state of the respiratory

center

4. The baseline should be at zero level

5. Shape is compared to that of normal respiration

2. Frequency determined by respiratory rate

3. Rhythm determined by state of the respiratory

center

4. The baseline should be at zero level

5. Shape is compared to that of normal respiration

Page 20: Capnography 102 - SCFR

The Normal CapnographThe Normal Capnograph

Page 21: Capnography 102 - SCFR

Abnormal Normal CapnographAbnormal Normal Capnograph

ETCO2 Less Than 35 mmHg = "Hyperventilation

or Hypocapnia"

ETC02 Greater Than 45 mmHg = "Hypoventilation

or Hypercapnia"

Page 22: Capnography 102 - SCFR

Hypoventilation Hypoventilation –– CapnographyCapnography

When a person hypoventilates, their CO2 goes up

Page 23: Capnography 102 - SCFR

Hyperventilation Hyperventilation –– CapnographyCapnography

When a person hypoventilates, their CO2 goes down

Page 24: Capnography 102 - SCFR

Interpretations of CapnographsInterpretations of Capnographs

� Esophageal Intubation

Page 25: Capnography 102 - SCFR

Interpretations of CapnographsInterpretations of Capnographs

� Esophageal Intubation

Page 26: Capnography 102 - SCFR

Interpretations of CapnographsInterpretations of Capnographs

� Hyperventilation or Hypoventilation?

Page 27: Capnography 102 - SCFR

Interpretations of CapnographsInterpretations of Capnographs

� Hyperventilation or Hypoventilation

Page 28: Capnography 102 - SCFR

Gradually increasing EtCO2Gradually increasing EtCO2

� Possible causes:HypoventilationRising body temperature/malignant hyperthermiaIncreased metabolismPartial airway obstructionAbsorption of CO2 from exogenous source

Page 29: Capnography 102 - SCFR

Interpretations of CapnographsInterpretations of Capnographs

� Leakage between sample tube and capnograph.

Page 30: Capnography 102 - SCFR

Interpretations of CapnographsInterpretations of Capnographs

Pediatric Respirations

Page 31: Capnography 102 - SCFR

Sudden loss of EtCO2 to zero Sudden loss of EtCO2 to zero or near zeroor near zero

Possible causes:Airway disconnectionDislodged ET tube/esophageal intubationTotally obstructed/kinked ET tubeComplete ventilator malfunction

Page 32: Capnography 102 - SCFR

Sustained low EtCO2 with good Sustained low EtCO2 with good alveolar plateaualveolar plateau

Possible causes:HyperventilationHypothermiaSedation, anesthesia

Page 33: Capnography 102 - SCFR

Sustained low EtCO2 without Sustained low EtCO2 without

alveolar plateaualveolar plateau / “Shark Fin”/ “Shark Fin”

� Possible causes:Incomplete exhalationPartially kinked ET tubeAsthma / COPD

� Brochospasm Mucous pluggingPoor sampling techniques

Page 34: Capnography 102 - SCFR

Exponential decrease in EtCO2Exponential decrease in EtCO2

� Possible causes:Cardiopulmonary arrestPulmonary embolismSudden hypotension; massive blood lossCardiopulmonary bypass

Page 35: Capnography 102 - SCFR

Pitfalls to CapnographyPitfalls to Capnography

Things to Remember

Page 36: Capnography 102 - SCFR

Pitfalls to CapnographyPitfalls to Capnography

� In and by themselves, end-tidal CO2 detectors (colormetric or infrared) should not be used solely as a means to confirm proper ET placement.

� They are a tool to be used in addition to � They are a tool to be used in addition to visualization of vocal cords, bilateral breath sounds and chest rise, tube condensation.

� All forms of capnography have a potential for inaccurate readings.

Page 37: Capnography 102 - SCFR

Pitfalls to CapnographyPitfalls to Capnography

Potential sources of inaccuracies:

1. Cap expired

2. Cap got wet from medications

3. Excessive PE3. Excessive PE

4. CO2 in the esophagus or stomach

- Pt.s who ingested large amounts

carbonated liquid prior to arrest.

5. Extended down time.

Page 38: Capnography 102 - SCFR

Pitfalls to Capnography:Pitfalls to Capnography:

Potential sources of inaccuracies:

1. Mechanical malfunction.

2. Sensor not heated or out of calibration.2. Sensor not heated or out of calibration.

3. Sample tubing kinked or blocked

4. Residual CO2 in stomach or esogphogus.

5. Extended down time.

6. Extremely low initial readings

during cardiac arrest.