thermo scientific safe 2020 and maxisafe 2020 … · thermo scientific safe 2020 and maxisafe 2020...

16
Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment . Comfort . Convenience Revolutionizing Clean Environments

Upload: truongdang

Post on 30-Aug-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Thermo Scientific SAFE 2020 and MAXISAFE 2020Class II Biological Safety Cabinets

Part of Thermo Fisher Scientific

Containment . Comfort . ConvenienceRevolutionizing Clean Environments

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:36 Uhr Seite 2

Page 2: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Thermo Scientific Safe and Maxisafe 2020 Setting new standards for safety

• EN 12469The 1.2 and 1.8 m Safe and Maxisafe 2020 cabinetsare independently tested and certified to EN 12469standard for Class II biological safety cabinets by TÜVNord, the leading testing agency in Europe.

Decades of MarketLeadership

We

laun

ch o

ur fi

rst l

ine

of c

lean

air

prod

ucts

We

laun

ch th

e w

orld

's fi

rst s

afet

y ca

bine

t

with

mot

oriz

ed fr

ont w

indo

w a

nd a

eros

ol

tight

win

dow

sea

ling

NSF

pub

lishe

s fir

st A

NSI

49

stan

dard

for

Clas

s II

biol

ogic

al s

afet

y ca

bine

ts

Ther

mo

Scie

ntifi

c bi

olog

ical

safe

ty c

abin

et e

volu

tion

We

intr

oduc

e th

e w

orld

’s fi

rst T

ÜV

cert

ified

biol

ogic

al s

afet

y ca

bine

t

We

laun

ch o

ur fi

rst L

NE

cert

ified

safe

ty c

abin

et

Firs

t NSF

-cer

tifie

d Th

erm

o Sc

ient

ific

biol

ogic

al

safe

ty c

abin

et is

laun

ched

We

laun

ch o

ur fi

rst D

IN 1

2980

cer

tifie

d sa

fety

cabi

net d

esig

ned

for p

repa

ring

and

han

dlin

g

cyto

toxi

c dr

ugs

Thermo Scientific biological safety cabinets are backed by our global reputation and commitment to provide the safest and most reliable biologicalsafety cabinets available for every lab and customer application.

We deliver decades of market leadership and technology innovations withunmatched quality.

Our philosophy is grounded in compliance with the most stringent regulatorystandards in the world that ensure your safety.

Thermo Scientific – moving science forward.

• DIN 12980Our 1.2 and 1.8 m Maxisafe 2020 cabinets with singlepiece worktops are independently tested and certifiedto the DIN 12980 standard “Safety Cabinets forHandling Cytotoxic Substances”, requiring the highestpersonnel, product and environmental protection.

1993 19951985197819761965 1996

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:36 Uhr Seite 3

Page 3: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Our experience brings innovation to you

Safe and Maxisafe 2020 Class II biological safety cabinets deliver maximum safety, comfortand convenience. Featuring an advanced design for easy operation and maximum safety, ourSafe and Maxisafe 2020 cabinets may be used wherever protection is of paramount importance in safety levels 1 to 3, including research, pharmaceutical quality control labs,pharmacies, and clinics/hospitals.

1.2 m model shown with optional electric stand, gasand service valves in side and rear walls.

2009

Euro

pe h

arm

oniz

es E

N 1

2469

, “Pe

rfor

man

ce

Crite

ria

for M

icro

biol

ogic

al S

afet

y Ca

bine

ts”

We

laun

ch th

e w

orld

's fi

rst s

afet

y ca

bine

t

with

inno

vativ

e br

ushl

ess

DC

mot

or te

chno

logy

We

laun

ch th

e ad

vanc

ed T

herm

o Sc

ient

ific

Safe

and

Max

isaf

e 20

20 b

iolo

gica

l saf

ety

cabi

nets

• Superior Personnel and Product Protection with best-in-class air flow and filter technology

• Exceptionally Comfortable ergonomic design prevents user fatigue and promotessafe working habits

• Maximized Energy Efficiency for environmental protection and significant cost savings

3

2000

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:36 Uhr Seite 4

Page 4: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Motorized front window offers comfortable working conditions at a200 mm working aperture for the Safeand Maxisafe 2020. The windowheight can be easily adjusted with onlya finger tip.

Aerosol-tight window sealing offers additional personnel and sample protection when the Safe or Maxisafe2020 biological safety cabinet is set tostand-by mode.

H14 HEPA filters guarantee maximum filtration efficiency of 99.995% at themost penetrating particle size (MPPS)for Safe and Maxisafe 2020 models.The supply and exhaust HEPA filtersensure superior contamination prevention while maximizing filter life.

The most important factor in selecting a biological safety cabinet is the security of particulate containment. Unique technologies adapted for Thermo Scientific safety cabinets ensure safe air filtration, avoiding samplecontamination or environmental hazards.

Best-in-class containment

Second to None in Containment

1.2 m model shown with optional electricstand, gas and service valves in side and rear walls.

4

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:36 Uhr Seite 5

Page 5: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

5

Interlocking H14 HEPA filters under theMaxisafe 2020 work surface provideeasy and safe HEPA filter changes inthe work area under safe negativepressure.

SmartFlowTM uses digital technology forSafe and Maxisafe 2020 models tomaintain constant airflow during normalfilter loading or temporary airflowobstruction from foreign objects.

Pressure sensor monitoring ensures safe airflow across the entire work surfaceof Safe and Maxisafe 2020 cabinets.Independent alarm systems immediatelyalert user to any airflow obstruction.

1.2 m model shown with optional gas and service valves in side and rear walls.

Exceptional design for personnel, product and environmental protection

Easy to read display allows convenient“at a glance” monitoring of safety parameters, including:• Performance Factor “PER”*• downflow air velocity• cabinet malfunctions• time of day • operating time of optional UV-C lights• total operating hours• timer for delayed start• stop-watch function

*Performance factor “PER” provides important information on the cabinets’current safety performance: • 60-100, cabinet is safe• 30-59, cabinet is safe but time for

service• 0-29, cabinet should not be used

The Maxisafe 2020 meets the latest European and worldwide standards forapplications with stringent safety requirements. Robust construction andexcellent ergonomic design are combined with significantly more effectiveair filtration, resulting in exceptional protection for personnel, product andenvironment.

100,000 times better air filtrationAn additional segmented H14 HEPA filter system is used directly below thework area of the Maxisafe 2020, protecting inner air ducts, plenums and fansfrom contamination and providing a cumulative filtration efficiency of99.99999975% to the downflow and exhaust airflow.

The Maxisafe 2020 is ideal for pharmacies preparing cytotoxic drugs and in pharmaceutical control labs.

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:36 Uhr Seite 6

Page 6: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

The optional electric stand for Safe 2020 isadjustable with an easy to use controller. Youcan change the height infinitely from a seatedto a standing working position.

The full working height of 950 mm provides avery comfortable standing working position,even for tall workers.

The electric stand with cable remote controlprovides comfortable height adjustment.

Unmatched ergonomic design

A quiet, comfortable cabinet design promotes user safety. The 10° slopedfront allows an ergonomic, fatigue-free working posture. With its framelessfront sash and side glass windows, the Safe and Maxisafe 2020 offer an unobstructed view of the complete work area from all angles.

Adjustable stand options for convenient working positionsAvailable with electric or manualadjustment, our stands areequipped with levelling feet tosecurely position the cabinet.

Conveniently located display and control panel for ease of useLocated on the outside of thebiosafey cabinet, the large display is easy to see from a distance, and also easy to reachfrom a seated working position.

Second to None in Comfort

The optional manual adjustable stand for Safe 2020 can be set from 750 - 950 mm working height in 50 mm increments. Theoptional foot rest allows a comfortable seatedworking position.

6 COMFO

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:37 Uhr Seite 7

Page 7: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Secure and comfortable armrests enhanceyour comfort during extended work sessionsand prevent carpal- tunnel syndrome. Locatedjust above the intake grill, they also ensurethat the air vents are not inadvertentlyobstructed.

The transparent side glass windowsmaximize light and visibility inside thecabinet, providing a bright and openworking environment.

780 mm interior height of the working chamberprovides high flexibility for your work insidethe cabinet.

Silence without compromising comfort and safetyUser safety and comfort is paramount with all Thermo Scientific biological safety cabinets. The comfortable, 200 mm front working aperture on the Safe and Maxisafe 2020 does not compromise safety or sample containment while significantly reducing noise level. All of our cabinets have been tested at nominal airflow velocities for added security.

7ORT

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:37 Uhr Seite 8

Page 8: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Innovative design for convenient sample preparation

Second to None in Convenience

Customized work surfaces

Scratch-free, high quality stainless steel work surfacesare available as single or segmented modules and areeasy to clean or autoclave. The indented work moduleseasily trap spilled liquids.

Optional vibration damped work surface section with granite plate for maximized stability.

Standard worksurface is suitable for sensitive weighing applications.

8*The standard Safe 2020 and Maxisafe 2020 configuration is proven to work with microbalances down to 10-5 readings, and depending upon installation

environment, may work down to 10-7 readings. Optional vibration-dampened worksurfaces are available to provide additional stability.

Thermo Scientific safety cabinets are the perfect choice for labs seeking thehighest levels of work efficiency. Our standard cabinets offer extremely lowvibration, making them perfect for use in sensitive weighing applications.*Additional options and accessories provide flexibility to configure the workspace to exactly match your application for enhanced work efficiency.

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:37 Uhr Seite 9

Page 9: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Efficient disinfection

Powerful cross beam UV-irradiation in both side walls(optional) illuminates the entire working area withoutcausing shadows. This ensures thorough disinfection ofthe complete chamber for maximum sample protection.

The fully opened front window allows convenient posture for thorough cleaning and disinfection.

Optional UV-irradiator for chamber disinfection can be mounted on the optional hanging bar.

Schematic of Unique Cross Beam UV-irradiation.

The large working aperture provides fullaccess to the work chamber for loading andunloading of large items and accessories likeour optional UV light on stand.

9

Easy cleaning

The hinged front window can be fully opened for easyand thorough cleaning of the unobstructed rear walland all interior surfaces. Rounded edges inside thecabinet facilitate effective disinfection and maximize contamination prevention. The display with its easy toclean flush surface is located outside the cabinet.

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:37 Uhr Seite 10

Page 10: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Advanced technology for maximumenergy efficiency

Superior DC motor technologyWe introduced the first innovative DC motor technology to the biological safety cabinet market. DC motors offer significant environmental advantages over traditional AC motors.

Minimized heat emmitance to the labDC motors are designed to use energy in an efficient manner, not only consuming less energy to operate the unit, but emitting less heat back into the lab. Lower heat output alsomeans a more comfortable working environment and lower air conditioning costs.

Maximized cost savingsCombined with the Night-Set-Back mode (reduced blower speed when the window is closed),the Safe and Maxisafe 2020 consume hundreds of kilowatts less energy than competingcabinets. Not only is our Safe and Maxisafe 2020 the preferred cabinet for enhanced containment, comfort, and convenience – but it saves your lab money and reduces your lab’s overall carbon footprint.

Safe 2020 Alternative A Alternative BOperational Energy (W) 200 810 1120Reduced Flow (W) 40 710 1020Operational Hours per year 2080 2080 2080(5 days x 8 hours x 52 weeks)Non-operational hours 6656 6656 6656Operational kW-hours 416 1685 2330Reduced Flow kW-hours 266 4726 6789Total Annual kW-hrs 682 6411 9119* Data provided above is based on internal performance testing and has been independently

verified in customer applications.

Good for you, good for the environment

Less energy consumption delivers

• reduced energy costs

• lower heat output

• reduced air conditioning costs

• comfortable working environment

• environmental protection

Comparative Energy Efficiency

annu

al k

ilow

att-

hour

s co

nsum

ed

10

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:37 Uhr Seite 11

Page 11: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

2

SPECIFICATIONS AND ORDERING INFORMATION

Options and Accessories

3

5 6

8

1

Safe 2020 and Maxisafe 2020 Options and Accessories

Order No. Description Factory Installed Option Accessory Image Country versions with local sockets and plugs 51900300 Country version Switzerland x51900303 Country version UK x51900306 Country version Italy x51900448 Country version France x51900449 Country version Australia x51900481 Country version Denmark x51900771 Country version B/CZ/SK/PL x51900900 Country version China xChamber equipment51900648 Powder-coated rear wall instead of stainless steel x51900644 Left side wall stainless steel plate instead of glass x51900645 Right side wall stainless steel plate instead of glass x -51900646 Left side wall polycarbonate instead of glass x51900647 Right side wall polycarbonate instead of glass x51900651 Safe/Maxisafe 2020 0.9 without worksurface (required for installation of non-standard worksurface options) x51900652 Safe/Maxisafe 2020 1.2 without worksurface (required for installation of non-standard worksurface options) x51900653 Safe/Maxisafe 2020 1.5 without worksurface (required for installation of non-standard worksurface options) x51900654 Safe/Maxisafe 2020 1.8 without worksurface (required for installation of non-standard worksurface options) x51900786 Battery backup for motorized front window xExhaust transitions50078791 Direct Duct Exhaust Transition for 0.9 m cabinet. Connection diameter 200 mm. x 150073771 Direct Duct Exhaust Transition for 1.2 and 1.5 m cabinet. Connection diameter 200 mm. x 150078790 Direct Duct Exhaust Transition for 1.8 m cabinet. Connection diameter 200 mm. x 150078796 Thimble Duct Exhaust Transition for 0.9 m cabinet. Connection diameter 200 mm. x 250073770 Thimble Duct Exhaust Transition for 1.2 and 1.5 m cabinet. Connection diameter 200 mm. x 250078795 Thimble Duct Exhaust Transition for 1.8 m cabinet. Connection diameter 200 mm. x 250078873 Exhaust cooling device for 1.2 m and 1.5 m cabinets x 350078874 Exhaust cooling device for 1.8 m cabinets x 3UV irradiation51900642 Cross-beam UV Light Option in side walls (comes with protective cover) x 450048696 Mobile UV Light on stand (must also order UV-socket 51900643 or 50109996) x 550073962 Mobile UV Light, hanging from rear wall (may not be used with 0.9 m cabinet) x 6

(must also order UV-socket 51900643 or 50109996 and hanging bar 50073944)51900643 UV socket for mobile UV light option x50109996 UV socket for mobile UV light option, for retro-fit in exisiting cabinet x

4

11

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:37 Uhr Seite 12

Page 12: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

SPECIFICATIONS AND ORDERING INFORMATION

Options and Accessories

Safe 2020 and Maxisafe 2020 Options and Accessories

Order No. Description Factory Installed Option Accessory Image Utilities50046015 Combustible gas tap for side wall, yellow (we recommend to also order 50074905 solenoid valve) x 1050045959 Noncombustible gas tap for side wall, black x 1050044678 Vacuum tap for side wall, grey x 1050044679 Water tap for side wall, green x 1050059017 Hose connection to fit through side wall access port, fits 10-13 mm diameter tubing x 750066668 Hose connection to fit through side wall access port, fits 3 mm diameter tubing x 851900649 Rear wall 3 service taps, left side of cabinet x 1151900711 Rear wall 3 service taps, right side of cabinet x 1151900722 Rear wall 6 service taps total, 3 on left and 3 on right side of cabinet x 11

(may not be used with 0.9m cabinet)50068235 Combustible gas tap for rear wall, yellow (we recommend to also order 50074905 solenoid valve) x 950068234 Noncombustible gas tap for rear wall, black x 950068232 Vacuum tap for rear wall, grey x 950068233 Water tap for rear wall, green x 951900650 Magnehelic pressure gauge x50074905 Solenoid valve for combustible gas tap x51900641 Drain pan valve xFilter51900657 Additional exhaust HEPA filter for all models, factory installed x 1651900658 Additional exhaust Carbon filter for all models, factory installed x 1650078857 Additional exhaust HEPA filter for 0.9 m cabinet, field installed x 1450073772 Additional exhaust HEPA filter for 1.2 and 1.5 m cabinet, field installed x 1450078856 Additional exhaust HEPA filter for 1.8 m cabinet, field installed x 1450078783 Additional exhaust Carbon filter for 0.9 m cabinet, field installed x50073773 Additional exhaust Carbon filter for 1.2 and 1.5 m cabinet, field installed x50078784 Additional exhaust Carbon filter for 1.8 m cabinet, field installed x50076846 Grid pre-filter for 0.9 and 1.2 m (Safe 2020 only) x50076921 Grid pre-filter for 1.5 and 1.8 m (Safe 2020 only) x51900781 Activated carbon filter that fits under 0.9 m cabinet worksurface (Maxisafe 2020 only) x 1551900782 Activated carbon filter that fits under 1.2 m cabinet worksurface (Maxisafe 2020 only) x 1551900783 Activated carbon filter that fits under 1.5 m cabinet worksurface (Maxisafe 2020 only) x 1551900784 Activated carbon filter that fits under 1.8 m cabinet worksurface (Maxisafe 2020 only) x 15User convenience equipment50073943 Hanging bar with set of 6 hooks (for 0.9 m cabinet only) x 1350073944 Hanging bar with set of 12 hooks (may not be used with 0.9 m cabinet) x like 1350078900 Utility basket for use with hanging bars 50073943 or 50073944 x 1250078901 2 Utility baskets including additional hanging bar x like 1250073663 Replacement stainless steel armrest (set of 2) x 17

7

8

16

11

10

9

17

14 15

12 13

12

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:37 Uhr Seite 13

Page 13: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

SPECIFICATIONS AND ORDERING INFORMATION

Options and Accessories

21

26

2523

2019

18

25

26

27

22

Safe 2020 and Maxisafe 2020 Options and Accessories

Order No. Description Factory Installed Option Accessory Image Work surfaces50073666 One-piece flat worksurface module 150 mm wide x 1850072685 One-piece flat worksurface module 300 mm wide x 1950073667 One-piece flat worksurface module 600 mm wide x 2050076398 One-piece flat worksurface module 900 mm wide x 2150073668 One-piece indented worksurface module 600 mm wide x like 2250073669 One-piece indented worksurface module 900 mm wide x like 2250073670 One-piece indented worksurface module 1200 mm wide x 2250073671 One-piece indented worksurface module 1500 mm wide x like 2250078750 One-piece indented worksurface module 1800 mm wide x like 2250073757 Vibration damped work surface 300 mm width. Constructed in stainless steel with x 24

granite plate. For weighing equipment and microscopy (not for use with Maxisafe 2020)50076162 Vibration damped work surface 600 mm width. Constructed in stainless steel with x like 24

granite plate. For weighing and microscopy equipment (not for use with Maxisafe 2020) 50084339 Vibration damped granite block 300 mm width. For weighing equipment and microscopy x like 24

equipment for triple filter cabinets only. (To be placed on top of standard worksurface)50084350 Vibration damped granite block 600 mm width. For weighing and microscopy equipment for x like 24

triple filter units only. (To be placed on top of standard worksurface)50078875 Mini sink tabletop insert 300 mm, requires drain pan valve 51900641 to also be fitted. Holds ~3 liters. x 23Stand options for Safe 2020 (Manual adjustable floorstand is included with Maxisafe 2020)50116441 Fixed height floor stand 0.9 m, 750 mm working height x 2550109309 Fixed height floor stand 1.2 m, 750 mm working height x 2550116442 Fixed height floor stand 1.5 m, 750 mm working height x 2550109311 Fixed height floor stand 1.8 m, 750 mm working height x 2550116443 Manual adjustable floor stand 0.9 m, 750 to 950 mm working height in 50 mm increments x like 2550109312 Manual adjustable floor stand 1.2 m, 750 to 950 mm working height in 50 mm increments x like 2550116444 Manual adjustable floor stand 1.5 m, 750 to 950 mm working height in 50 mm increments x like 2550109313 Manual adjustable floor stand 1.8 m, 750 to 950 mm working height in 50 mm increments x like 2550117449 Electric adjustable floor stand 0.9 m with cable remote control, 750 to 950 mm working height x 2750109314 Electric adjustable floor stand 1.2 m with cable remote control, 750 to 950 mm working height x 2750117450 Electric adjustable floor stand 1.5 m with cable remote control, 750 to 950 mm working height x 2750109325 Electric adjustable floor stand 1.8 m with cable remote control, 750 to 950 mm working height x 2750051982 Footrest bar 0.9 m (not to be used with electric stand) x 2650051983 Footrest bar 1.2 m (not to be used with electric stand) x 2650051984 Footrest bar 1.5 m (not to be used with electric stand) x 2650051985 Footrest bar 1.8 m (not to be used with electric stand) x 2650109977 Castors, for fixed height floor stands only (not to be used with electric stands) x

13

24

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:38 Uhr Seite 14

Page 14: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

Schematic drawing Safe 2020

SPECIFICATIONS AND ORDERING INFORMATION

Schematic drawing Maxisafe 2020

371 14,61

100 3,94 100 3,94371 14,61

100 3,94100 3,94

236

9,29

531

20,9

1

411 16,18

D

F

25 0,98 70 2,76

602,

36

552,

17

552,

17

30,50 1,20

30,50 1,20

B C

E

200

7,87

43,2

71,

70

200

7,87

power supply4900mm /192,91"

135

5,31

153,50 6,04 90 3,54

A15

0059

,06

50 1,97 710 27,95

40 1,57

160

6,30

420

16,5

4

803,

15

799 31,45

863,

39

702,

76

min

. 200

7,87

ceiling

power supply4900mm /192,91"

D

371 14,61

100 3,94 100 3,94

236

9,29

411 16,18

531

20,9

1

371 14,61

100 3,94100 3,94

F

552,

17

552,

17

129

5,08

25 0,98 70 2,76

B C

30,50 1,20

30,50 1,20

E

112,

34,

4220

07,

87

200

7,87

power supply4900mm /192,91"

222,50 8,76 90 3,54

135

5,32

A15

0059

,06

119 4,69 710 27,95

40 1,57

122

4,80

420

16,5

4

401,

57

868 34,17

334 13,15

863,

39

702,

76

min

. 200

7,87

ceiling

power supply4900mm /192,91"

A

seated work position minimum mm 680

seated work position minimum inch 26,77

standing work position maximum mm 880

standing work position maximum inch 34,65

Attention: With additional gas valves in the side walls, you must plan with more width (dimension D).

B C D E F

Safe 0.9 mm 81 491 1000 430 900

inch 3,19 19,33 39,37 16,93 35,43

Safe 1.2 mm 228 644 1300 583 1200

inch 8,97 25,35 51,18 22,95 47,24

Safe 1.5 mm 478 644 1600 583 1500

inch 18,82 25,35 62,99 22,95 59,05

Safe 1.8 mm 474,5 951 1900 890 1800

inch 18,68 37,44 74,80 35,04 70,86

A

seated work position minimum mm 680

seated work position minimum inch 26,77

standing work position maximum mm 880

standing work position maximum inch 34,65

Attention: With additional gas valves in the side walls, you must plan with more width (dimension D).

B C D E F

Maxisafe 0.9 mm 81 491 1000 430 900

inch 3,19 19,33 39,37 16,93 35,43

Maxisafe 1.2 mm 228 644 1300 583 1200

inch 8,97 25,35 51,18 22,95 47,24

Maxisafe 1.5 mm 478 644 1600 583 1500

inch 18,82 25,35 62,99 22,95 59,05

Maxisafe 1.8 mm 474,5 951 1900 890 1800

inch 18,68 37,44 74,80 35,04 70,86

14

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:38 Uhr Seite 15

Page 15: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

SPECIFICATIONS AND ORDERING INFORMATION

Thermo Scientific Safe 2020, Class II Biosafety Cabinet

Model Unit Safe 2020 0.9 Safe 2020 1.2 Safe 2020 1.5 Safe 2020 1.8

Order No. 51026637 51026638 51026639 51026640

Description: Biological Safety Cabinet Class II Class II Class II Class II

0.9 m (3 ft.) 1.2 m (4 ft.) 1.5 m (5 ft.) 1.8 m (6 ft.)

Dimensions

Exterior Dimensions H/W/D mm 1586/1000/800 1586/1300/800 1586/1600/800 1586/1900/800

Exterior Dimensions H/W/D inches 62,2/39,4/31,5 62,2/51,2/31,5 62,2/63/31,5 62,2/74,8/31,5

Usable Work Area/Interior H/W/D mm 780/900/465 780/1200/465 780/1500/465 780/1800/465

Usable Work Area/Interior H/W/D inches 30,7/35,4/18,3 30,7/47,2/18,3 30,7/59,1/18,3 30,7/70,9/18,3

Work Surface Height (with adjustable stand) mm/inches 750-950/30-37 750-950/30-37 750-950/30-37 750-950/30-37

Work Surface Height (with fixed stand) mm/inches 750/30 750/30 750/30 750/30

Working Height of front window mm/inches 200/8 200/8 200/8 200/8

Maximum Opening Height of front window mm/inches 773/30,4 773/30,4 773/30,4 773/30,4

Shipping Dimensions H/W/D1) mm 1760/1110/925 1760/1410/925 1760/1710/925 1760/2010/925

Shipping Dimensions H/W/D1) inches 69,3/43,7/36,4 69,3/55,5/36,4 69,3/67,3/36,4 69,3/79,1/36,4

Weight

Net Weight kg/lbs 170/375 200/441 230/507 280/617

Shipping Weight1) kg/lbs 190/419 225/496 260/573 315/694

Max. load capacity of one-piece work surface kg/lbs 50/110 50/110 50/110 50/110

Max. load capacity of single segmented kg/lbs 25/55 25/55 25/55 25/55

worksurface per 30 cm segments4) (Total load) (50/110) (50/110) (75/165) (75/165)

Ventilation system

Exhaust/inflow air volume m3/h/CFM 288/170 386/227 483/284 580/341

Exhaust volume, thimble connection m3/h/CFM 375/221 501/295 628/369 754/444

Heat emission at 25 °C room temperature

Operation mode (non-vented) W/BTU/hr 160/546 200/682 265/904 340/1161

Reduced flow mode (non-vented) W/BTU/hr 40/136 40/136 70/239 70/239

Filter Specification

Supply/exhaust air filter H14 HEPA EN 1822, 99.995% at the most penetrating particle size (MPPS)

Ergonomics

Noise pressure level EN2) dB (A) 56 56 58 58

Lighting power lx/fc >800/74 >850/79 >1250/116 >1300/121

Electrical data

Voltage V 230 230 230 230

Frequency Hz 50/60 50/60 50/60 50/60

Energy Consumption, W 160 200 265 340operating set point3)

Energy Consumption, reduced W 40 40 70 70flow mode

Max Current Consumption A 7,3 7,3 8,7 8,7

Protection Class I / IP 20 I / IP 20 I / IP 20 I / IP 20

Features

Receptacles One duplex on each side wall. The receptacles have a load capacity of up to 5 A and are protected with

T 5 A fuses. When all receptacles are in use simultaneously they must not exceed the maximum total load capacity of 5 A

Access Ports 4 standard (2 on each side wall)

Service Valves Up to 4 may be installed through access ports (2 on each side wall)

Up to 6 more may be factory installed on rear wall (optional)1) export packaging dimensions and weight on request2) tested according to EN 12469, at setpoint of cabinet and with front window at 200 mm working height 3) Clean filters, fans operating at nominal set point, interior lighting activated4) The total load on overall working area must not exceed the indicated weight

15

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:38 Uhr Seite 16

Page 16: Thermo Scientific SAFE 2020 and MAXISAFE 2020 … · Thermo Scientific SAFE 2020 and MAXISAFE 2020 Class II Biological Safety Cabinets Part of Thermo Fisher Scientific Containment

SPECIFICATIONS AND ORDERING INFORMATION

Thermo Scientific Maxisafe 2020, Class II Biosafety Cabinet

Model Unit Maxisafe 2020 0.9 Maxisafe 2020 1.2 Maxisafe 2020 1.5 Maxisafe 2020 1.8

Order No. 51026651 51026652 51026653 51026654

Description: Biological Safety Cabinet Class II, with Class II, with Class II, with Class II, with

triple HEPA-filter system triple HEPA-filter system triple HEPA-filter system triple HEPA-filter system

Dimensions

Exterior Dimensions H/W/D mm 1586/1000/870 1586/1300/870 1586/1600/870 1586/1900/870

Exterior Dimensions H/W/D inches 62,2/39,4/34,3 62,2/51,2/34,3 62,2/63/34,3 62,2/74,8/34,3

Usable Work Area/Interior H/W/D mm 780/900/465 780/1200/465 780/1500/465 780/1800/465

Usable Work Area/Interior H/W/D inches 30,7/35,4/18,3 30,7/47,2/18,3 30,7/59,1/18,3 30,7/70,9/18,3

Work Surface Height (adjustable) mm/inches 750-950/30-37 750-950/30-37 750-950/30-37 750-950/30-37

Working Height of front window mm/inches 200/8 200/8 200/8 200/8

Maximum Opening Height of front window mm/inches 780/31 780/31 780/31 780/31

Shipping Dimensions H/W/D1) mm 1760/1110/925 1760/1410/925 1760/1710/925 1760/2010/925

Shipping Dimensions H/W/D1) inches 69,3/43,7/36,4 69,3/55,5/36,4 69,3/67,3/36,4 69,3/79,1/36,4

Weight

Net Weight kg/lbs 200/441 240/529 280/617 330/728

Shipping Weight1) kg/lbs 220/485 265/584 310/683 365/805

Max. load capacity of one-piece work surface kg/lbs 50/110 50/110 50/110 50/110

Max. load capacity of single segmented kg/lbs 25/55 25/55 25/55 25/55

worksurface per 30 cm segments4) (Total load) (50/110) (50/110) (75/165) (75/165)

Ventilation system

Exhaust/inflow air volume m3/h/CFM 288/170 386/227 483/284 580/341

Exhaust volume, thimble connection m3/h/CFM 375/221 501/295 628/369 754/444

Heat emission at 25 °C room temperature

Operation mode (non-vented) W/BTU/hr 200/682 240/819 305/1041 420/1433

Reduced flow mode (non-vented) W/BTU/hr 40/136 40/136 70/239 70/239

Filter Specification

Primary/Supply/exhaust air filters H14 HEPA EN 1822, 99.995% at the most penetrating particle size

Ergonomics

Noise pressure level2) dB (A) 56 56 58 58

Lighting power lx/fc >800/74 >850/79 >1250/116 >1300/121

Electrical data Voltage V 230 230 230 230

Frequency Hz 50/60 50/60 50/60 50/60

Energy Consumption, operating set point3) W 200 240 305 420

Energy Consumption, reduced flow mode W 40 40 70 70

Max Current Consumption A 7,3 7,3 8,7 8,7

Protection Class I / IP 20 I / IP 20 I / IP 20 I / IP 20

Features

Receptacles One duplex on each side wall. The receptacles have a load capacity of up to 5 A and are protected with

T 5 A fuses. When all receptacles are in use simultaneously they must not exceed the maximum total load capacity of 5 A

Access Ports 4 standard (2 on each side wall)

Service Valves Up to 4 may be installed through access ports (2 on each side wall)

Up to 6 more may be factory installed on rear wall (optional)1) export packaging dimensions and weight on request2) tested according to EN 12469, at setpoint of cabinet and with front window at 200 mm working height3) Clean filters, fans operating at nominal set point, interior lighting activated4) The total load on overall working area must not exceed the indicated weight

© 2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

www.thermo.com/bsc

North America: USA/Canada +1 866 984 3776 (866-9-THERMO) Europe: Austria +43 1 801 40 0, Belgium +32 2 482 30 30, France +33 2 2803 2180, Germany national toll free 08001-536 376,Germany international +49 6184 90 6940, Italy +39 02 02 95059 434-254-375, Netherlands +31 76 571 4440, Nordic/Baltic countries +358 9 329 100,Russia/CIS +7 (812) 703 42 15, Spain/Portugal +34 93 223 09 18, Switzerland +41 44 454 12 12, UK/Ireland +44 870 609 9203Asia: China +86 21 6865 4588 or +86 10 8419 3588, India toll free 1800 22 8374, India +91 22 6716 2200, Japan +81 45 453 9220,Other Asian countries +852 2885 4613 Countries not listed: +49 6184 90 6940 or +33 2 2803 2180

BRO-LECA-SAFE2020-1208

BRO-LECA-SAFE2020-0409.qxd:HERACELL_e.qxd 21.04.2009 16:36 Uhr Seite 1