COMPARING CENTRAL CORNEAL THICKNESS MEASUREMENTS USING VISANTE OCT, PENTACAM AND GALILEI. Bernilla E; Izquierdo Jr. L; Miranda M. Authors have  no  financial interest Instituto de Ojos OFTALMOSALUD Lima-Perú. OBJECTIVE. - PowerPoint PPT Presentation

TRANSCRIPT

  • COMPARING CENTRAL CORNEAL THICKNESS MEASUREMENTS USING VISANTE OCT, PENTACAM AND GALILEIBernilla E; Izquierdo Jr. L; Miranda M.Authorshavenofinancialinterest

    Instituto de Ojos OFTALMOSALUDLima-Per

  • OBJECTIVETo compare de central corneal thickness measurements found using non contact methods as are Visante OCT, Pentacam and Galilei.

  • MATERIALS AND METHODSProspective, comparative, and double blinded study measuring central corneal thickness on 50 eyes of 25 patients being evaluated for refractive surgery. The measurements were taken by three different researchers without knowing the previous results obtained by the other devices.We analyzed the thickness averages, maximum and minimum values, and their standard deviations. The statistical analysis was made using ANOVA testing and SPSS v 15.

  • Our subjects mean age was 30.8 (SD=5.736). The central thickness average for Visante OCT was 530.02 30.682, for Pentacam was 554.76 32.028, and for Galilei was 562.38 29.21. When comparing the three devices we found that theres a statistical difference between the measurements (p=0.000), and when comparing Pentacam vs Galilei that difference is not found (p=0.648).

    RESULTS

  • RESULTSThe central thickness average for Visante OCT was 530.02 30.682, for Pentacam was 554.76 32.028, and for Galilei was 562.38 29.21

  • RESULTSMean of corneal thickness measurement value was not statistically significant comparing the three devices.

  • Our subjects mean age was 30.8 (SD=5.736). The central thickness average for Visante OCT was 530.02 30.682, for Pentacam was 554.76 32.028, and for Galilei was 562.38 29.21. When comparing the three devices we found that theres a statistical difference between the measurements (p=0.000), and when comparing Pentacam vs Galilei that difference is not found (p=0.648).

    CONCLUSION