4-22-13-ams2750e-update
TRANSCRIPT
-
8/16/2019 4-22-13-AMS2750E-Update
1/39
AMS2750 E
Changes from AMS2750D
Feb 2013
-
8/16/2019 4-22-13-AMS2750E-Update
2/39
2 HONEYWELL - CONFIDENTIAL File Number
Overview on Changes
• AMS2750 Specification Updated to Revision E- New Revision of Spec – 2750E revised July 2012;
- Replaces Spec - 2750D (revised Sept 2005)
- Updated to include wireless sensor technology
- Added requirements to document “ As found” instrumentcalibration and offset condit ions (procedural impacts)
- Calibration changes l ike check min. of 6 simulated sensor @min. 4 points, calibrated for each type input/output used
- Provides more details on Temperature Uniformity SurveyTesting (TUS) procedures, furnace repairs and modifications,data collection requirements, survey frequency, number andlocation of sensors, Pass/Fail requirements
- Added Refrigeration & Quench Furnace class to tables
General Update
-
8/16/2019 4-22-13-AMS2750E-Update
3/39
3 HONEYWELL - CONFIDENTIAL File Number
Standards
• AMS2750E Pyrometry
- Covers Temperature Sensor, Instrumentation, Thermalprocessing equipment, System Accuracy Tests, TemperatureUniformity Survey requirements
- Ensures that parts or raw material are heat treated inaccordance to applicable specifications
• AS7102 Nadcap Requirements for Heat Treating
- Documents the requirements for the accreditation program- Industry managed program for continuous improvement
Standard industry audit program for special processes
Focuses on overall supplier performance
Auditing overseen by PRI (Performance Review Institute)
-
8/16/2019 4-22-13-AMS2750E-Update
4/39
4 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Section 2 Definitions Changes
- Par. 2.2.15 DIGITAL INSTRUMENT:
Clarification was added to definition of digital instruments withdigital temperature display and with temperature trace on
paper chart:
An instrument that presents process measurements in a digital
(numeric) display format. Typical examples are mechanical
recorders with digital displays, digital indicators, controllers withdigital displays, etc. In the case of a recorder with digi tal
temperature display and a recorder trace (line) the record
(recorder trace) is analog while the display is digi tal. A
recorder trace on a paper chart is an analog record.
- Par. 2.2.25 INTERVAL
Interval definition was added to AMS2750E to avoid confusion
by some suppliers.
-
8/16/2019 4-22-13-AMS2750E-Update
5/39
5 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Section 2 Definitions Changes
- Par. 2.2.27 LOAD SENSORS :
Clarif ication regarding correct usage of load sensors/thermocouples. (in rev. D it was not included in the load
sensor definit ion but stated in par. 3.1.8.1):
Sensors that are attached to or in contact with product ion
material, a representation of production material, or are buried
in the load of production material and which supply temperaturedata of the production material to process instrumentation.
-
8/16/2019 4-22-13-AMS2750E-Update
6/39
6 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Temp Sensor Changes
AMS2750E
Section
Reference AMS2750E Change Comment
3.1.1.5.1 Allows use of Wireless transmitters as alternative to
extension wire
New
3.1.2.2.1 Certs for sensors now require “Date of Calibration” New
3.1.2.2.6 Correction factor for each calibration temperature
traceable to NIST or other recognized National Standard.
Alternatively, deviation (error) for each calibration
temperature may be listed. The certificate shall clearly
state if it shows deviation or correction factor.
Added this
3.1.2.5.2 Extrapolation of calibration correction factors above the
highest calibration temperature and below the lowest
calibration temperature is prohibited by any calibration
source except NIST or other National Standards
calibration agency.
Added “except NIST or other
National Standards calibration
agency.”
• Temperature Sensor - Section 3.1 Changes
- This section deals with the requirements for the temperaturesensors used in these heat treatment processes
-
8/16/2019 4-22-13-AMS2750E-Update
7/397 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Temp Sensor Changes
• Temperature Sensor - Section 3.1 ChangesAMS2750E
SectionReference
AMS2750E Change Comment
3.1.2.6 When calibrating thermocouples rolls, AMS2750 Rev. E
requires sampling from both ends regardless of roll
length. (1000 feet/305 meters limit from AMS2750D for
one end sampling only is no longer acceptable). Other
requirements for rolls calibration remains the same.
Added “Rolls shall be sampled at
both ends of the roll”
3.1.3.5 Nonexpendable base metal TUS TCs used below 1200F
(650C) do not need to be permanently installed on TUS
rack between its usage like it was stated in Rev. D
TUS T/C’s can be reinserted for each
TUS through ports in the furnace;
need to be used at the same location
and depth of insertion.
3.1.5.2 Added time of 90 days
“Expendable base metal load sensors may be used for
up to 90 calendar days after first use and for no morethan 30 uses”
Previously was just 30 uses
3.1.5.3 Changed the requirement on the life of nonexpendable
base metal load thermocouples
2200 °F (1205 °C) to 2299 °F (1260 °C) –
3 months or 10 uses
1801 °F (980 °C) to 2199 °F (1205 °C) –
3 months or 90 uses
Previously for 2200˚Fwas just 10
uses, no time period specified
1801˚F was 30 days or 90 uses;
now 90 days or 90 uses
-
8/16/2019 4-22-13-AMS2750E-Update
8/398 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Temp Sensor Changes
• Temperature Sensor - Section 3.1 Changes
AMS2750ESection
Reference AMS2750E Change Comment
Figure 1 Added Table of Reuse and Recalibration of Test & Load
Sensors
Summarizes Reuse & Recalibration
Primarily for clarification
Figure 1 Sensor - Secondary Standard – Max. Permitted Error
changes
Increased max. permitted error
Figure 1 Sensor – System Accuracy Test –
• Max. Permitted Error changes
• Recalibration permited if Types E & K are used below
500F (260C) only (Note 11)
Increased max. permitted error
3.4.5.2.1 Type B TCs are allowed as resident SAT TCs fortemperatures above 500F (260C)
Type B T/C’s were not allowed byRev D.
-
8/16/2019 4-22-13-AMS2750E-Update
9/399 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E InstrumentsControlling, Monitoring or Recording Instruments
AMS2750E
Section
Reference AMS2750E Change Comment
3.2.1 AMS2750D allowed to use instruments meeting
requirements of AMS2750C (but not meeting rev D
requirements) if these instruments were purchased
before September 2006 (one year after rev. D issue). If
you have such instruments, now you have 3 years you
upgrade your equipment to meet Rev. E requirements
Exemption will run out for older
instrumentation, now have to
upgrade within 3 years of issuing of
AMS2750E (issued July 2012)
3.2.1.1.1 Temperature resolution requirements for chart
recorders shall be in accordance with Table 4. (See next
slide for table)
New requirements for refrigeration &
quench system recorders added
3.2.1.1.2 Process Recorder Print and Chart Speeds shall be in
accordance with Table 5. (See next slide for table)Added metric speeds and updated
notes
• Instrumentation - Section 3.2 Changes
- This section deals with the requirements of theinstrumentation used in these heat treatment processes
-
8/16/2019 4-22-13-AMS2750E-Update
10/3910 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E InstrumentsControlling, Monitoring or Recording Instruments
New - requriements for refrigeration
and quench system recorders were
not defined in rev D.
-
8/16/2019 4-22-13-AMS2750E-Update
11/3911 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E InstrumentsControlling, Monitoring or Recording Instruments
AMS2750E
SectionReference
AMS2750E Change Comment
3.2.1.5.4 Calibration of secondary standard instruments and field
test instruments shall be performed in accordance with
the manufacturer’s instructions. These instruments shall
be tested at a minimum of 6 simulated sensor inputs.
These shall include the minimum and maximum of the
operating range used for test or calibration and a
minimum of 4 points in between either representing
areas of normal operation or spaced at approximately
equal intervals across the range in which the instrument
is used for test or calibration.
New requirements regarding
secondary standard instruments and
field instruments calibration
3.2.1.5.4.1 The instrument shall be calibrated for each type of input
and/or output used (e.g., each thermocouple type in useand for mV, mA, etc. if the instrument is used for these
methods).
New requirements regarding
secondary standard instruments andfield instruments calibration
3.2.1.5.4.2 Calibration shall be performed on each channel in use
that can be altered or adjusted individually, or on each
group of channels that can be altered or adjusted as a
group. It is recommended that channels not in use be
blocked or tagged to stop unintentional use.
New requirements regarding
secondary standard instruments and
field instruments calibration
• Secondary Standard & Field Inst. Calibration Changes
-
8/16/2019 4-22-13-AMS2750E-Update
12/3912 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E InstrumentsControlling, Monitoring or Recording Instruments
AMS2750E
SectionReference
AMS2750E Change Comment
3.2.5.5 Calibration shall be performed on instrument(s) defined
under the Instrument Type column in Table 3.
Regardless of the calibration procedure used, it shall
comply with the requirements of Table 3. Prior to any
instrument alteration or calibration, document the
instrument(s) “as-found condition”
Not documenting “as found”
condition is often an audit failure
point during Nadcap audits
3.2.5.5.1 Check as-found condition at a minimum of three
simulated sensor inputs which represent the minimum,
maximum and at least one point in the middle 1/3rd of
the furnace Qualified Operating Temperature Range. If
necessary, calibrate the instrument in accordance with
the manufacturer’s instructions.
New requirements regarding
calibration of controlling, monitoring
and recording instruments
3.2.5.5.2 Following any adjustments test the instrument at a
minimum of three simulated sensor inputs which
represent the minimum, maximum and at least one
point in the middle 1/3 rd of the furnace Qualified
Operating Temperature Range to document the as-left
condition.
New requirements regarding
calibration of controlling, monitoring
and recording instruments
• Calibration Controlling/Monitoring/Recording Inst. Changes
-
8/16/2019 4-22-13-AMS2750E-Update
13/3913 HONEYWELL - CONFIDENTIAL File Number
Table 3 – Instruments and Instrument Calibration –updates compared to rev. D:
AMS2750E Instruments and Instrument Calibration
Table 3 Notes
(1) Instrum ents of equiv alent or greater accuracy are acceptable.
(2) Comparison checks among the cells shall be performed monthly.
(3) Cells having temperature c ontro l o f ±0.02 °F (±0.01 °C).
(4) Furnace class 1 & 2 instr uments shall have a minimum sensitivi ty of 1
°F (1 °C). Furnace class 3 through 6
instru ments shall have a minim um sensi tivity o f 3 °F (2 °C).
(5) Controlling, monitoring and recording instrument calibration interval
shall be:
Digital Electromechanical
Furnace Class 1 - Monthly Furnace Class 1 - Monthly
Furnace Class 2 - Quarterly Furnace Class 2 - Monthly
Furnace Class 3 - Quarterly Furnace Class 3 - Quarterly
Furnace Class 4 - Quarterly Furnace Class 4 - Quarterly
Furnace Class 5 - Semiannually Furnace Class 5 - QuarterlyFurnace Class 6 - Semiannually Furnace Class 6 – Quarterly
Refrigeration and quenchant monitor (digital or analog) – semiannually.
Exception: If the quenchant/refrigerator sensor is connected to a channel
in a furnace process recorder, the interval for c alibration of that channel
in the instrument shall be the same interval as for the other channels of
the instrument.
(6) Furnace Control, Recording, Monitoring or Data Acquisition
Instruments.
(7) Field Test instruments meeting accuracy requirements of secondary
standards may be used to calib rate SAT and TUS test instruments in t he
field.
(8) Over-temperature instruments used solely for fu rnace over
temperature protection do not need to be calibrated.
(9) NIST or o ther equiv alent National Standard.
(10) Includes wireless transmission systems
(11) Instruments are operated in a controlled environment and are not
used on the shop floor.
New
New
New
-
8/16/2019 4-22-13-AMS2750E-Update
14/3914 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E InstrumentsControlling, Monitoring or Recording Instruments
AMS2750E
SectionReference
AMS2750E Change Comment
3.2.5.5.3 Chart recorder (circular and strip) speed(s) shall be
verified annually and shall be accurate within 3 minutes
per hour. Result of chart speed measurement shall be
documented.
Added Chart Speed documentation
requirement
3.2.5.5.5 For furnaces used only at a single temperature, the
calibration of controlling, monitoring or recording
instruments may either be performed per
manufacturer’s instructions at the single temperature
of use, or with a minimum of 3 points: the temperature
of use, and two (or more) points bracketing the
temperature of use.
New requirement
3.2.5.5.6 Wireless equipment which performs the analog to
digital conversion at the furnace and transmits a digitalsignal to the recorder are allowed, but require use of
the complete wireless system (wireless sender,
wireless receiver and associated controlling,
monitoring and recording instruments) during the
calibration process (See 3.1.1.5.1).
New requirement
• Calibration Controlling/Monitoring/Recording Inst. Changes
-
8/16/2019 4-22-13-AMS2750E-Update
15/3915 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E InstrumentsControlling, Monitoring or Recording Instruments
AMS2750E
SectionReference
AMS2750E Change Comment
3.2.7.1 An “electronic record” is any combination of text,
graphics, data, audio, pictorial, or other information
representation in digital form that is created, modified,
maintained, archived, retrieved, or distributed by a
computer system. When using a system (furnace
control, recording, monitoring or data acquisition) that
creates electronic records, systems purchased after
Sept 2006 or any electronic record created after 2015
(3 years after release of this revision) shall meet the
following requirements detailed in section 3.2.7.1.3
Added additional requirement,
3.2.7.1.3.1, to provide a means of
providing evidence that the electronic
record was reviewed.
This can be accomplished in the
Trendview recorder by the mark on
chart feature.
3.2.7.1.1.3.1 The system shall be capable of providing evidence the
record was reviewed – such as by recording anelectronic review, or a method of printing the record for
a physical marking indicating review.
New requirement
This can be accomplished in theTrendview recorder by the mark on
chart feature.
• Electronic Records Changes
-
8/16/2019 4-22-13-AMS2750E-Update
16/3916 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Instruments - Thermal Processing
AMS2750E
SectionReference
AMS2750E Change Comment
3.3.2 Table 3 The temperature indicated by the control sensor in each
control zone shall be recorded by a recording
instrument.
Alternatively, the recording instrument may be
connected to a second sensor contained in the same
sheath or holder as the control sensor, and separatedfrom the control sensor by no more than 0.38 inch (10
mm).
For Instrument Types A, B, C, D
Added alternative means of
connecting a control sensor, this was
not allowed in AMS 2750 D
3.3.2 Table 3 At least 1 recording load sensor in each zone
During production in multi-zone furnaces, empty zones
(i.e., no material is placed in or intrudes into the zone)
do not require a load sensor. However, a notation
must be made to the furnace load record that the zone
was entirely empty.
For Instrument Types A, B, C, D
Added clarification for multi-zone
furnaces
• Furnace Instrumentation Type Classification Changes
-
8/16/2019 4-22-13-AMS2750E-Update
17/3917 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Instruments - Thermal Processing
AMS2750E
SectionReference
AMS2750E Change Comment
3.3.4 Additional Sensors - There is no limit on additional
recording or load sensors in any control zone, but their
use must be defined in controlled operating instructions.
Added clarification for use of
additional sensors
3.3.5 High and low temperature sensors, when required
(instrument types A and C – See Figure 3), may be
inserted in heat sinks (See 2.2.24), positioned at thehigh and low temperature locations of the control zone
based on the most recent TUS when heat sinks of
similar configuration are used during the TUS.
Improvement to aid the heat treater
Also refer to Section 2.2.24
definition of a heat sink
3.3.6 For multiple zone vacuum furnaces with 225 cubic feet
(6.4 m3) or less in working volume that are instrumented
with Types A, B, or C instrumentation, it is acceptable to
treat the furnace working volume as a single work zone
for the purpose of locating high and low temperaturerecording thermocouples and determining the number of
load thermocouples required regardless of the number of
control thermocouples or instruments within the vacuum
furnace. The minimum sensors required are shown in
Figure 4. Each furnace shall have over-temperature
protection in the work zone.
Added clarification multi-zone
furnaces
• Furnace Instrumentation Type Changes
-
8/16/2019 4-22-13-AMS2750E-Update
18/39
-
8/16/2019 4-22-13-AMS2750E-Update
19/39
19 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Instruments - Thermal Processing
AMS2750E
Section
Reference AMS2750E Change
Comment
3.3.3.1 Refrigeration equipment shall have a temperature
controller. This temperature controller requirement is not
applicable to liquid nitrogen, dry ice and dry ice/liquid-
cooled containers. All refrigeration equipment shall be
equipped with a temperature recorder where time-at-
temperature (minimum or maximum) is required. The
above requirements are not applicable during
transportation of materials at sub-ambient temperatures.
AMS2750D did not require to have
temperature recording instrument
for quench systems that were set up
before AMS2750D
3.3.3.2 Quench systems used for heat treatments that include a
quenchant temperature requirement (minimum,
maximum or both) shall be equipped with a recording
instrument. Existing installations installed prior to
September 2005 do not require recording instruments
until 3 years after the release of this revision.
AMS2750D did not require a
temperature recording instrument
for quench systems that were set up
before (Sept. 2005). If you still have
a quench system without a
recording instrument, you have 3
years to upgrade per AMS2750E to
include a recorder.
• Instrumentation for Refrigeration & Quench Systems
-
8/16/2019 4-22-13-AMS2750E-Update
20/39
20 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E – System Accuracy Test
• System Accuracy Test Changes – Section 3.4
- Defines what equipment requires SAT, the frequency, waiverconditions and the Test Procedure to use
- The SAT is an on-site comparison of the instrument/leadwire/sensor readings or values, with the readings or values of a
calibrated test instrument/leadwire/sensor to determine if themeasured temperature deviations are within applicable
requirements. The SAT is performed to assure the accuracy of
the furnace control and recording system in each control zone.
-
8/16/2019 4-22-13-AMS2750E-Update
21/39
21 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E – System Accuracy Test
AMS2750E
SectionReference
AMS2750E Change Comment
3.4.2.1 Two sensors in each control zone are Type B, N, R, or S. “B” Type T/C added
3.4.5.2.1 Resident SAT thermocouples shall be restricted to Type B,
N, R or S at temperatures exceeding 500 °F
(260 °C) and shall be nonexpendable if exposed to
temperatures above 1000 °F (538 °C).
“B” Type T/C added
• System Accuracy Test Changes
-
8/16/2019 4-22-13-AMS2750E-Update
22/39
AMS2750 rev. E updates Tables 6 and 7 for SAT requirements:
New
In rev D there was
mistake and
Monthly/Quarterly
was stated
New
AMS 2750E – System Accuracy Test
-
8/16/2019 4-22-13-AMS2750E-Update
23/39
AMS 2750E T U if it S
-
8/16/2019 4-22-13-AMS2750E-Update
24/39
24 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E – Temp Uniformity Survey
AMS2750E
SectionReference
AMS2750E Change Comment
3.5.13.2 Once data collection begins, temperature data shall be
recorded from all TUS sensors at a frequency of at least
one set of all readings every 2 minutes for the duration of
the survey. Data from furnace sensors required by the
applicable instrumentation type (See 3.3) shall be
recorded as follows. (Sensors whose only function is over-temperature protection do not need to be recorded.)
No change with this part of the
requirement
If the normal frequency of temperature data recording in
production is 2 minutes or less, or is continuous as in the
case of analog recorders, the data shall be documented
in the normal production format. If the normal
frequency of temperature data recording in production is
greater than 2 minutes, the recording frequency interval
during the TUS shall not exceed 6 minutes.
Added statement for those
processes that record at less than 2
minutes
• TUS Data Collection Changes
-
8/16/2019 4-22-13-AMS2750E-Update
25/39
AMS 2750E – Temp Uniformity Survey Records
HONEYWELL - CONFIDENTIAL
AMS2750 rev. E par. 3.5.21: Temperature Uniformity Survey Report3.5.21.1 The following items shall be included in the temperature uniformity survey report:
a. Furnace identification name or number
b. Survey temperatures
c. TUS sensor and location identification including a detailed diagram, description or photograph(s)
of any load or rack used
d. Time and temperature data from all recorded sensors required for furnace
inst rumentation type for all zones tested (3.5.13.3.2)
e. Correction factors for TUS sensors at each survey temperature.
f. As found and as left TUS offsets (if used in production)
g. Corrected or uncorrected readings of all TUS sensors at each survey temperature. Readings shall
be identified as corrected or uncorrected.
h. Testing company identification (if not performed in-house)
i. Signature for the testing company (if not performed in-house)
j. Identification of technician performing survey
k. Survey start date and time
l. Survey end date and timem. Survey test instrument identification number
n. Indication of test pass or test fail
o. When required, documentation of furnace survey sensor failures (See 3.5.16)
p. Summary of corrected plus and minus TUS readings at each test temperature after stabilization.
q. Quality Organization approval.
New
-
8/16/2019 4-22-13-AMS2750E-Update
26/39
AMS 2750E – Temp Uniformity Survey Records
HONEYWELL - CONFIDENTIAL
AMS2750 rev. E par. 3.5.21: Temperature Uniformity
Survey Report – (cont inued)
3.5.21.2 Although not a required part of the uniformity survey report, the following shall beaccessible on site:
a. Control instrument tuning parameters
b. TUS sensor calibration report
c. Control and recording sensor calibration report
d. Diagrams of control and recording sensors, load and TUS sensor locations in
three dimensional space.
-
8/16/2019 4-22-13-AMS2750E-Update
27/39
AMS 2750E I t t
-
8/16/2019 4-22-13-AMS2750E-Update
28/39
28 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Instruments3.2.4 Controll ing, Monitoring or Recording Instruments
Requirements for Controlling, Monitoring or Recording Instruments
AMS 2750E Ref. Requirement Comment
Readability
3.2.3.1
At least one recording and/or
control ling instrument for each zone
shall have a minimum readability of
1°F or 1°C
Honeywell Recorders (X-Series Paperless,
DR4500, DPR180, DPR250), HC900 OI and
Controllers (UDC2500, UDC3200 & UDC3500)
have a readabil ity o f at least 1°F or 1°C, most
read down to 0.1°.
Installation
3.2.3.2
Installation shall conform to
manufacturer’s recommendation
Manuals provide detailed information on
proper installation for each product .
Offsets
3.2.3.4
If offsets are used, a documentedprocedure shall exist describing
when and how to perform offsets.
Honeywell Recorders, Controllers and HC900allow offsets to be easily entered and shows
the offsets as part of the configurations.
Honeywell Instruments meet the requirements
for readability, installation, offsets & signal input
Signal Inputs
3.2.3.3
Instruments shall receive an
unmodif ied signal from sensors except
for A to D and D to A conversions, or a
digitally p rocessed, error-checked
equivalent representation of a direct
measured value.
Honeywell Recorders, Controllers and HC900
all accept direct sensor inputs of the T/C
lis ted in Table 2 of AMS 2750D, many of the
recorders can accept and document input
values over communications.
AMS 2750E Instr ments
-
8/16/2019 4-22-13-AMS2750E-Update
29/39
29 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Instruments3.2.5 Instrument Calibration
Requirements for Instrument Calibration AMS 2750E Ref. Requirement Comment
Calibrationcompliance
3.2.5.1
Calibration shall be performed on theinstruments in the Instrument Types
listed in Table 3 & comply w ith the
Table 3 requirements. Prior to any
instrument alteration or calibration,
document the instrument(s) “ as-found
condition”
Honeywell Recorders (X-Series Paperless,DR4500, DPR180 & DPR250), HC900 and
Controllers (UDC2500, UDC3200 & UDC3500) are
considered control, monitoring and recording
instruments.
The recorders can also be used as Field Test
instruments for doing Temp Uniformity Surveys.
Accuracy & Freq.
3.2.5.2
Calibration accuracy & frequency
shall be in accordance with Table 3
Recorders, HC900 & Contro llers can meet the 2°F
(1.1°C) for Digital Instruments or 2°F (1.1°C)
/0.3% of Max Survey Temp for elect ro-mech. inst.
Calibration
Performance
3.2.5.4
Calibration shall be to Mfg’s
instructions or a minimum of 6
simulated sensor inputs at the Min. &
Max. of the Operating Range used for
test or calib. and a minimum of 4 points
in between …...
Honeywell Recorders, HC900 & Controllers
are typically calibrated at the zero and span
values for each range. The calibration
procedure is documented as part of the
product manual.
We can meet the requirements
Honeywell Recorders (X-Series Paperless
and DPRs) provide easy annotation o f the
chart to aid the documentation process.
Calibration
Performance
3.2.5.5.4
Calibration may be performed with a
load in process if furnace temp remains
within the processing tolerance and the
record is appropriately annotated to
indicated that a calibration occurred.
-
8/16/2019 4-22-13-AMS2750E-Update
30/39
AMS 2750E Instruments
-
8/16/2019 4-22-13-AMS2750E-Update
31/39
31 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Instruments3.2.6 Instrumentation Records
Requirements for Documentation of Instrument Calibration
AMS 2750E Ref. Requirement Comment
InstrumentSticker
3.2.6.1
A st icker aff ixed to the instrumentshall indicate the most recent
successful calibration. As a
minimum the sticker shall include:
Date Cal performed, Due date of next
Calibration, Technician who
performed calibration, and any limits
or restrictions of the calibration
The user must establish adequate traceabilitythat the inst rumentation has been calibrated.
The X-Series Paperless Recorder Status
Maintenance & Diagnostic screens can provide
an aid for doing this .
Calibration
Results
3.2.6.2
The results of calibration shall be
documented. Minimum requirements:
Inst or Furnace No., Make & Model of
instrument, Standard used, method of
calibration, Required Accuracy, As Found &
as Left data for each cal point, Offset as
found & as left, Sensitivity, Statement of
acceptance or rejection, Any lim itations or
restr ict ion, Date of Cal, Next Due Date,
Technician name, Calibration Co Name (if
not done in house, Signature of Calibration
Co , Quality Organization approval
The user must establish a report format to
satisfy these requirements.
The user must document instrument calibrations
-
8/16/2019 4-22-13-AMS2750E-Update
32/39
AMS 2750E Instruments
-
8/16/2019 4-22-13-AMS2750E-Update
33/39
33 HONEYWELL - CONFIDENTIAL File Number
AMS 2750E Instruments3.2.7 Electronic Records
Requirements for Electronic Data Recording
AMS 2750E Ref. Requirement Comment
TMP provides secure analysis & data archiving
The X-Series Paperless Recorders provides
full password system to comply wi th 21CFR
Part 11. 50 Users wi th un ique ID & Password,
time outs, expirations of passwords ,
password definitions and audit trail info. The
HC900 OI provides for secured access via a
User specified numerical code.
System Access –
(Passwords)
3.2.7.1.5
The system shall provide methods (e.g.
passwords) to limi t system access to
only individuals whose authorization is
documented
The TrendManager Software does not alter the
source information and supports the
protection, retention and retrieval of records. A data base Management Tool provides the
means to move, copy and mange the files for
storage on secure servers. TrendManger
Software is compatible with current PC
Operating Systems and can read data from all
generations of Trendview Paperless recorders
Retention &
retrieval
3.2.7.1.4
The system shall support protection,
retention, and retrieval of accurate
records throughout the record retentionperiod. The hardware and/or software
shall operate throughout the retention
period as specified in section 3.7
Review
evidence
3.2.7.1.4
The system shall be capable ofproviding evidence the record was
reviewed – such as by recoding an
electronic review, or a method of
printing the record for a physical
marking indicating review
The X-Series Paperless Recorders providesmark on chart, ability to create a report or data
can be imported to the AMS2750 Report Tool
to c reate a report.
-
8/16/2019 4-22-13-AMS2750E-Update
34/39
34 HONEYWELL - CONFIDENTIAL File Number
• Furnace Class Uniformity & Recorder Resolution
- Classes based on Temperature Uniformity requirements- Resolution requirements affect the recorder and chart used- Paperless recorders record actual values eliminating issues
with the chart and the chart resolution
Furnace
Class
Temp
Uniformity
°F
Temp
Uniformity
°C
Max ° per Inch
(CM) of Chart
°F/Inch
Max ° per Inch
(CM) of Chart
°C/CM
Max Chart
Recording
Increment
°F/Line
Max Chart
Recording
Increment
°C/Line
1 ±5 ±3 50 11 2 1
2 ±10 ±6 150 33 5 3
3 ±15 ±8 150 33 5 3
4 ±20 ±10 250 55 10 5
5 ±25 ±14 250 55 10 5
6 ±30 ±28 350 77 25 15
Section 3.3 Thermal Processing Equipment
-
8/16/2019 4-22-13-AMS2750E-Update
35/39
S
-
8/16/2019 4-22-13-AMS2750E-Update
36/39
36 HONEYWELL - CONFIDENTIAL File Number
Recorders That Meet AMS 2750E
• Honeywell has Paper and Paperless Recorders that meet
AMS 2750E requirements for Controlling, Monitoring and
Recording Instruments & Field Test - Key requirement for Controlling, Monitoring & Recording instruments is a
calibrated accuracy of (± 2°F/1.1°C) or 0.2% of reading whichever is greater for
dig ital instruments, ±2 °F (±1.1 °C) or ±0.3% of the maximum survey temperature of
the equipment, whichever is greater for electromechanical
-Key requirement for Field Test ins truments is a calibrated accuracy of (±1°F/0.6°C)
DR4500 Circular
Chart X-Series Paperless
TrendManager
Software
DPR 180 DPR 250
UDC C ll Th M AMS 2750E
-
8/16/2019 4-22-13-AMS2750E-Update
37/39
37 HONEYWELL - CONFIDENTIAL File Number
UDC Control lers That Meet AMS 2750E
• Honeywell has Controllers that meet AMS 2750E
requirements for Control ling, Monitoring andRecording Instruments
- Key requirement for Controlling, Monitoring & Recordinginstruments is a calibrated accuracy (± 2°F/1.1°C) or 0.2% of
reading whichever is greater for digital instruments
UDC 3500
UDC 3200
UDC 2500
HC900 M t AMS 2750D
-
8/16/2019 4-22-13-AMS2750E-Update
38/39
38 HONEYWELL - CONFIDENTIAL File Number
HC900 Meets AMS 2750D
• The Honeywell HC900 and associated 559 & 1042
Operator Interfaces (OI) meet AMS 2750Drequirements for Control ling, Monitoring and
Recording Instruments
- Key requirement for Controlling, Monitoring & Recording
instruments is a calibrated accuracy (± 2°F/1.1°C)
-
8/16/2019 4-22-13-AMS2750E-Update
39/39
www.honeywell.com