sm capnography principles and clinical application
TRANSCRIPT
SM
SM
SM
SM
SM
SM
Capnography
Principles and Clinical Application
Objectives
Describe the advantages of mainstream vs. sidestream CO2 technology.
Discuss normal and abnormal V/Q relationships. Identify a normal capnogram and discuss phase
I thru IV. Discuss the ETCO2/PaCO2 gradient and its
clinical application. Interpret abnormal capnograms and their clinical
intervention.
. .
Capnography - Technology Capnographs utilize infrared (IR) technology
CO2 molecules absorb IR light energy of a specific wavelength
Amount of energy absorbed = CO2
concentration
Infrared is particularly appropriate for measuring CO2
CO2 has a strong absorption band in the infrared spectrum
• In the ICU, the CO2 band is distinct enough from other gases to minimize interference
Capnography – Technology Capnography vs. Capnometry
Capnography
Measurement & display of ETCO2 and the CO2 capnogram
Measured by a capnograph
Capnometry
Measurement & display of the ETCO2
value Measured by a
capnometer
ETCO 2
R R
ETCO 2
R R
CapnographyQuantitative vs. Qualitative ETCO2
Quantitative ETCO2
Provides actual numeric value
Found in capnographs and capnometers
Qualitative ETCO2
Only provides range of values
Termed CO2 detectors - Easy Cap
ETCO 2
R R
mmH g
ETCO 2
0-10
11-20
21-30
31-40
over 40
0-4
5-20
>20
Capnography
Mainstream
vs. SidestreamE TC O 2
R R
E TCO 2
R R
Capnography - Mainstream Sensor placed in ventilator
circuit
Measurement made at the patient’s airway
IR sensor can not be contaminated by patient secretions!
Fast response time
No water traps or tubing needed - hassle free
Sensor
E TC O 2
R R
Capnography – Sidestream
Sensor located away from the airway
IR sensor can be contaminated by patient secretions!!
Measurement made by pump inside the monitor
Slower response time
Water traps and tubing required troubleshooting and maintenance
E TC O 2
R R
Sample measuredinside monitor
CO2 sampleAcquired here
CapnographySolid State vs. Chopper Wheel
Solid State CO2 Sensors No moving parts =
durability Uses a beam splitter to
measure IR light at two wavelengths
IR light source electronically pulsed
Chopper Wheel CO2 Sensors Spinning wheel = very fragile Spins to change parameter
measured by photodetector Gas sample to be measured
(data) Sample plus sealed gas
reference cell No light at all
Capnography
What Are We Measuring?
CapnographyRespiration - The Big Picture
1Cellular Metabolism of food into energy - O2 consumption & CO2 Production
2 Transport of O2 & CO2 between cellsand pulmonary capillaries
3 Ventilation between alveoli and pulmonary capillaries
Capnography Depicts Respiration
Capnography
Ventilation
O2
Transport
CO2
CO
2
Metabolism
CO2
E TC O 2
R R
Capnography
Arterial CO2 (PaCO2)
from ABGETCO2
from Capnograph
Normal Arterial & ETCO2 Values
Normal PaCO2 Values:
35 - 45 mmHg
Normal ETCO2 Values:
30 - 43 mmHg
E TCO 2
R R
Capnography
Arterial - End Tidal CO2 Gradient
In healthy lungs the normal PaCO2 to ETCO2 gradient is 2-5 mmHg
In diseased lungs, the gradient will increase due to ventilation/perfusion mismatch
Ventilation- Perfusion Relationships
Ventilation-Perfusion RelationshipsRelationship between ventilated alveoli and blood flow
in the pulmonary capillaries
Shunt perfusionAlveoli perfused but not ventilated
Deadspace VentilationAlveoli ventilated but not
perfused
CO2 O2
NormalVentilation and
perfusion is matched
Normal V/Q
CO2 O2
ETCO2 / PaCO2
Gradient =2 to 4 mmHg
. .
Shunt Perfusion – Low V/Q
No exchange of O2 or CO2
ETCO2 / PaCO2
Gradient =4 to 10 mmHg
. .
Shunt Perfusion – Low V/Q
Disease processes that may cause Shunt Perfusion: Mucus plugging ET tube in right or left main stem bronchus Atelectasis Pneumonia Pulmonary edema
In short anything that causes the alveoli to collapse or is alveolar filling
. .
Dead Space Ventilation
High V/Q
Perfusion is the problemNo exchange of O2 or CO2occurs
ETCO2 / PaCO2
Gradient is large
Ventilation is not the problem!
. .
Dead Space
Why is understanding Alveolar Dead Space important? As Alveolar Dead Space increases, the
gradient between ETCO2 and PaCO2
increases
Why does increased Alveolar Dead Space create a gradient?
Dead Space Ventilation
0 0 0
0
0
0 0
PaCO2 = 53 mmHg
ETCO2 = 33 mmHg
53
53
53 Alveoli that do not take part in gas exchange will still have no CO2 –Therefore they will dilute the CO2 from thealveoli that wereperfused
The result is a widened ETCO2 to PaCO2 Gradient
A Gradient is a Good Thing
Why? Lets clinicians know when patient status
improves
• PaCO2/ETCO2 gradient narrows
Aids in determining what caused a drop in ETCO2
• If ventilation hasn’t changed a sudden and large drop in ETCO2 usually indicates a change in perfusion.
Dead Space Ventilation Disease processes that may cause Dead
Space Ventilation: Pulmonary embolism Hypovolemia Cardiac arrest Shock
In short anything that causes a significant drop in pulmonary blood flow
Capnography
Clinical Application of Capnography
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Capnography
Clinical utility of the CO2 Waveform or capnogram
Provides validation of ETCO2 value
Visual assessment of patient airway integrity Verification of proper ET tube placement Assessment of ventilator, and breathing circuit
integrity
SM
SM
SM
SM
SM
SM
The Normal Capnogram
Normal Capnogram - Phase I
50
0
25
CO2 mmHg
Beginning of expiration =anatomical deadspace with no measurable CO2
A B
Anatomical Dead Space
Anatomical Dead Space Internal volume of the
upper airways• Nose • Pharynx• Trachea• Bronchi
Anatomical DeadspaceConducting Airway - No Gas Exchange
Normal Capnogram - Phase II
50
0
25
CO2 mmHg
Mixed CO2, rapid rise in CO2 concentration
B
C
Normal Capnogram - Phase III
50
0
25
CO2 mmHg
Time
Alveolar Plateau, all exhaled gas took part in gas exchange
End Tidal CO2 value
C D
Normal Capnogram - Phase IV
50
0
25
CO2 mmHgInspiration starts,
CO2 drops off rapidly
E
D
Capnogram – Valuable Tool
CO2 (mmHg)
0
25
50
Alveolar Plateau established
No Alveolar Plateau
SM
SM
SM
SM
SM
SM
Abnormal
CO2 Waveforms
Capnography
Endotracheal Tube in Esophagus
Possible Causes: Missed Intubation A normal capnogram is the best evidence that
the ET tube correctly positioned. When the ET tube is in the esophagus, little or
no CO2 is present
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Capnography
Obstruction in Airway or Breathing Circuit
Possible Causes: Partially kinked or narrowed artificial airway Presence of foreign body in the airway Obstruction in expiratory limb of breathing circuit Bronchospasm
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Capnography
Muscle Relaxants (curare cleft)
Possible Causes: Patient attempts to take a breath Appear when muscle relaxants begin to subside Depth of cleft is inversely proportional to degree
of drug activity
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Capnography
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Cardiac Oscillations
Characteristics: Rhythmic and synchronized to heart rate
Capnography
Inadequate Seal Around ET Tube
Possible Causes: Leaky or uncuffed endotracheal or trach tube Artificial airway that is too small for patient
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Capnography
Hypoventilation - Increase in ETCO2
Possible Causes:Decrease in respiratory rate Decrease in tidal volume Increase in metabolic rateRapid rise in body temperature
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Capnography
Hyperventilation - Decrease in ETCO2
Possible Causes: Increase in respiratory rate Increase in tidal volume Decrease in metabolic rate Fall in body temperature
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Capnography
Rebreathing
Possible Causes: Expiatory filter that is saturated or clogged,
expiratory valve that is sticking Inadequate inspiratory flow, or insufficient
expiratory time Anything that causes resistance to expired flow
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Case Study
A 29 year old male with head injury, and a compound fracture of his femur sustained in a motorcycle accident
2 weeks post trauma on mechanical ventilation with the following philological values:
PaCO2 – 42 mmHg PaO2 – 95 mmHg
ETCO2 – 38 mmHg Total Rate – 14 bpm
Minute Ventilation – 7 L/Min
Case Study
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Normal capnogram, stable trend ETCO2/PaCO2 gradient 4 mmHg
Case Study
CO (m m Hg)2
0
37
50 Real-Tim e Trend
Sudden decrease in ETCO2 from 38 mmHg to 20 mmHg and remains there
RR – increases to 24 bpmMinute Volume increases to 12 Lpm
Case Study
CO (m m Hg)2
0
37
50 Real-Tim e Trend
ABG was drawn with the following results:PaCO2 38 mmHgPaO2 59 mmHgPaCO2/ETCO2 gradient 18 mmHg
Case Study
Ventilation /perfusion lung scan was consistent
with a pulmonary embolism A sudden drop in ETCO2
Associated with a large increase in the PaCO2/ETCO2 gradient
Often is associated with pulmonary embolism
Summary
Capnography affords the clinician breath by breath trending of ETCO2 and thus a non- invasive look at ventilation
Provides an objective reason for ABG’s Trend ETCO2/PaCO2 gradient to observe
patient improvement
Changes in ventilation and perfusion are are often observed by trending the gradient