rfid pressure sensor system for hospital mattresses to prevent infections and bedsores

40
RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores Bikram Virk Sujoy Sanyal Ryan Eiswerth Wasif Khawaja Mushfiq Saleheen Dr. Koblasz, ECE4884 L05 April 22, 2008

Upload: shaine-cunningham

Post on 01-Jan-2016

29 views

Category:

Documents


1 download

DESCRIPTION

RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores. Bikram Virk Sujoy Sanyal Ryan Eiswerth Wasif Khawaja Mushfiq Saleheen. Dr. Koblasz, ECE4884 L05 April 22, 2008. OVERVIEW. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Bikram Virk Sujoy Sanyal

Ryan EiswerthWasif Khawaja

Mushfiq Saleheen

Dr. Koblasz, ECE4884 L05 April 22, 2008

Page 2: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Develop an inexpensive hospital bed mattress that can reduce the risk of bedsores and spread of Nosocomial infections.

Cost of prototype – $1,290

($1,300 less than predicted)

OVERVIEW

1/21

Page 3: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

2/21

Bedsores

• Definition: Unrelieved pressure to part of body for extended period of time, cutting off oxygen and blood flow

• Statistics: - 60,000 deaths/years in America - 1 million patients/year in America - 2/3 of bedsores occur in people over 70 years old

• Costs – over $1 billion/year

Life-Threatening

Page 4: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Life-Threatening - All Hospitalized Patients at Risk• Definition:

Hospital-Acquired Infections caused by bacteria,

viruses, etc. Includes Staph Infections• Statistics:

- more than 2 million cases/year in America

- 17,500 – 70,000 deaths/year in America• Causes:

- Sloppy hygiene: 36% preventable (CDC report)

- Aging hospitals: old mattresses

• Costs - $4.5 - $5.7 billion/year

Nosocomial Infections

3/21

Page 5: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Main Components of Project

4/21

Page 6: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

5/21

Page 7: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Movement Monitoring System

6/21

Interface Kit 8/8/8 Voltage Divider

Trossen Robotics 24” Force Sensing Resistor

Page 8: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

WARNING: PLEASE MOVE

TIMER: 30 Min

7/21

Page 9: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

WARNING: EGRESS DETECTED

8/21

Page 10: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

• Monitor patient contact• System:

- Operating Frequency: 13.56MHz - Target Range: 24” with 1 Watt reader - Tags worn on wrist

RFID

9/21

Page 11: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

RFID System

10/21

Page 12: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Identification Process

11/21

Page 13: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

WARNING: DR. SMITH PLEASE STAND BACK

12/21

Page 14: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Multiplexer Design

Relays

Antennas

13/21

Page 15: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Total Care SpO2RT Hospital Bed $28,500

Secure II Med/Surg Hospital Bed

$10,000

Competing Products

14/21

Page 16: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Product Quantity Unit Price

Price

Movement Detection System

Trossen Robotics 24” FSR 4 $24 $80

HP Pavillion Slimline s7700e PC

1 $400 $400

Cables 1 $10 $10

RFID System

TI-FEIG RFID Reader [10] 1 $350 $350

Copper Plating 25 ft. $20 $100

Antenna Board 1 pair $100 $100

PIC Kit II Microprocessor 1 $250 $250

Total Equipment Cost $1,290

15/21

Page 17: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Selling Point

16/21

• No need to buy new beds

• Installed as add-on to existing beds

• Money saved

Page 18: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Current StatusSpecification Proposed Actual

RFID Reader

- Operating Frequency 13.56 MHz 13.56 MHz

- Current Draw Max. 0.5 A Max. 0.5 A

- Detectable Range 2 ft (24 in.) 1.5-1.75 ft. (18-21 in.)

Multiplexer

- Power Supply Not Proposed 3.3 V

Trossen Robotics 24” FSR

- Force Sensitivity Range <100 g to > 10 kg <100 g to > 10 kg

- Pressure Sensor Range < 1.5 psi to > 150 psi < 1.5 psi to > 150 psi

- Sensitivity to VibrationNot Significantly

AffectedNot Significantly

Affected

Allowed Patient Rest Time 30 minutes 10 seconds*

Senor Time Monitoring Every minute Every second*

17/21

Page 19: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Obstacle: Proximity to Metal

18/21

Page 20: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Alternative Solutions • Dynasys 13.56 MHz Antenna

• 2 foot range

• De-Q feature

• $400

19/21

Page 21: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

For Future Groups…

20/21

• Buy Professionally

made antennas

• Bedside Falls

Page 22: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

For More Information Visit Us At

21/21

Page 23: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Bikram Virk Sujoy Sanyal

Ryan EiswerthWasif Khawaja

Mushfiq Saleheen

Page 24: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores
Page 25: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores
Page 26: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Proximity to Metals

• Difficult to find Matching Network

Page 27: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

• Nearby PC monitors

• PROXIMITY WITH METALS Altered performance of the RFID system can be expected when metals are in close proximity of the radiation field. Proximity to metals effectively lowers the antenna's inductance. Lower inductance causes an increase in resonant frequency and also a reduction in "Q". Designers may consider starting with a higher-than-needed "Q" "in the lab" to start with, expecting it to be lowered when it is installed in its intended location. External capacitance may be required to tune the modified antenna system back to resonance. These types of situations can be experienced around conveyor belt structures and also embedding antenna loops on concrete driveways where metal re-bar absorbs some of the radiated energy.

• http://rfidusa.com/superstore/product_info.php?products_id=121

Page 28: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores
Page 29: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Project Schedule

Acquire Parts 2/11/08

Construct Antennas 3/17/08

Movement Detection 3/19/08

Implement Comp. Interface 4/1/08

Implement the Full System 4/22/08

14/21

Page 30: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

• Bring in computer to show pressure sensors

Page 31: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Antenna Designs

Obstacles

18/21

Page 32: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Our Design

19/21

Page 33: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores
Page 34: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

WARNING: EGRESS DETECTED

Page 35: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores
Page 36: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores
Page 37: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Desired and Actual SpecificationsSpecification Proposed Actual

RFID Reader

- Transmitting Power 1W ± 2 dB 1W ± 2 dB

- Operating Frequency 13.56 MHz 13.56 MHz

- Current Draw Max. 0.5 A Max. 0.5 A

Head-Section Antenna Dimensions

17” x 30”17” x 30”

- Detectable Range 2 ft (24 in.) 1.5-1.75 ft. (18-21 in.)

Mid-Section Antenna Dimensions

9” x 30”9” x 30”

- Detectable Range 2 ft (24 in.)

Foot-Section Antenna Dimensions

15” x 30”15” x 30”

- Detectable Range 2 ft (24 in.)

Multiplexer

- Power Supply Not Proposed

Not Proposed

Not Proposed

Trossen Robotics 24” FSR

- Force Sensitivity Range <100 g to > 10 kg <100 g to > 10 kg

- Pressure Sensor Range < 1.5 psi to > 150 psi < 1.5 psi to > 150 psi

- Sensitivity to VibrationNot Significantly

AffectedNot Significantly

Affected

Allowed Patient Rest Time (No motion)

30 minutes 10 seconds*

Pressure Senor Time Monitoring Every minute Every second*

Page 38: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

Trossen Robotics FSR

• How it works

Page 39: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

5/21

Page 40: RFID Pressure Sensor System for Hospital Mattresses to Prevent Infections and Bedsores

In your final oral report, you should describe similar patented and

commercially available devices • Explain details about your design and

let us know why your design is better