quick startfrankshospitalworkshop.com/equipment/documents...quick start 1 logiq 200pro series quick...

36
GE Medical Systems Training In Partnership T i P TM LOGIQ200 PRO Series A Training in Partnership Program Quick Start 2234974–100

Upload: nguyendung

Post on 09-Jun-2018

232 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

GE Medical SystemsTraining In Partnership

TiPTM

LOGIQ 200 PRO Series

A Training in Partnership Program

Quick Start

2234974–100

Page 2: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides
Page 3: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

������� �� ���� �� ������ ������ ���

�� ����� �� ��

������� �

����� �����

����������

�������� �

LOGIQ 200 PRO Series

Quick Start Guide

Page 4: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

GE Medical Systems

GE Medical Systems: Telex 3797371P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.(Asia, Pacific, Latin America, North America)

GE Medical Systems–EuropeBeethovenstraβe. 239 Postfach 11 05 60 D–42655 Sol-Solingen GERMANY

Regulatory RequirementThis product complies with regulatory requirements of the following European Directive93/42/EEC concerning medical devices

This Quick Start Guide is a reference for the LOGIQ�200 PRO Series. It applies to allversion of the 0.0 software for the LOGIQ�200 PRO Series.

Page 5: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Revision History ALOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

REV DATE REASON FOR CHANGE

0 AUG, 8 1999 Initial release

��� �� ��������� ����

��� �������� ��� ��������

����� ����� ����� �����

TiP Cover Page N/A

Title Page 0

Revision History A and B 0

Quick Start 1 thru Quick Start 30 0

Page 6: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Revision History BLOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Please verify that you are using the latest revision of this document. Informationpertaining to this document is maintained on GPC (GE Medical Systems GlobalProduct Configuration). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE UltrasoundClinical Answer Center at 1-800-682-5327 or 414-524-5698.

Page 7: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 1LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Introduction

The Quick Start Guide (TRANSLATED) provides a step-by-step description of the basic features andoperation of the LOGIQ�200 PRO Series. It is intended to be used in conjunction with the User Manual inorder to provide the information necessary to operate the system safely.

The Quick Start Guide takes the user from system familiarization through power on, patient data entry, examcategory/preset selection, scan modes/adjustments, basic measurements, report pages, recording imagesand power off.

The LOGIQ�200 PRO Series manuals are written for users who are familiar with basic ultrasound principals andtechniques. They do not include sonography training or clinical procedures.

Prescription Device (for USA only)

CAUTION: United States law restricts this device to sale or use by or on the order of a physician.

Page 8: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 2 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Front View

The following are major features of the LOGIQ�200 PRO Series system console. Most features come with the standardconfiguration, while other features are options to the standard console.

8. Black & White Monitor

9. Release Button to raise and lower the monitor

10. B/W Video Page Printer (Optional)

11. Power Switch

12. Keyboard

13. Probe Connector

14. Probe Holder

15. MOD Driver (Optional)

1

2

7

4

5

6

38

Page 9: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 3LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Back View1. Rear Handle

2. Peripheral/Accesory Connector Panel

3. Black & White Monitor

4. Release Button to raise and lower the monitor

Optional Freeze Foot Switch

1

2

3

4

ÍÍÍ

ÍÍÍÍÍÍ

ÍÍÍ

FREEZE

Page 10: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 4 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Power On

To connect the system to the electrical supply:

� Ensure that the wall outlet is a minimum 20 amp dedicated circuit for 120 VAC (USA) or 7.5amp dedicated circuit for 220–240 VAC (Europe).

� Make sure that the power switch is turned off.

� Unwrap the power cable. Make sure to allow sufficient slack in the cable so that the plug isnot pulled out of the wall if the system is moved slightly.

� Push the power plug securely into the wall outlet.

To power on the system press the Power switch, located to the left of the probe connectors, tothe on(I) position.

IO

Page 11: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 5LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Control Panel Layout

KeyboardKeyboard controls have been arranged according to function and usage.This helps minimize operator movement while scanning.

1 2 3 4 5

6

7

8

9

1 : Patient Entry Keys2 : Alphanumeric Keys3 : Special Keys4 : Preset Keys5 : Probe Select Keys6 : User Define Keys7 : Mode, Display/Record Keys8 : Mesurement/Annotation Keys9 : TGC Keys

Page 12: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 6 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Probe Controls

A probe is activated by pressing the Probe Select key.

Probe Ports

All imaging probes can be plugged into either of the two probe ports.

NOTE: Ensure that the probe port is deactivated before disconnecting the probe.

Probe Select1

1

Probe Select2

2

LOCKUNLOCK

Page 13: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 7LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Starting an Exam

Patient Entry Menu

Press New Patient key at the beginning of each patient study to reset the system and enter newpatient data. When the New Patient key is pressed, the Patient Entry menu appears.

Use the Trackball to move the highlight cursor to the ExamCategory Selection line. Enter the desired exam number and fill inthe appropriate patient data.

Highlight EXIT and press Return or press New Patient again whenthe patient entry menu data has been completed.

The system takes a few seconds to load new exam categoryparameters and calibrate the attached transducers before enteringthe scan mode.

New Patient

[ PATIENT ENTRY MENU ]EXAM CATEGORY SELECTION : 1

1: RAD/ABDOMEN 2: OBSTETRICS 3: GYNECOLOGY4: CARDIOLOGY 5: UROLOGY 6: SMALL PARTS7: USER DEFINE1 8: USER DEFINE2PT NAME :PT ID :NOTE : Oper ID :

EXAM INFORMATION AGE: yrs

Ref MD:

COMMENT :

EXIT

Page 14: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 8 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Scan Mode Controls

Select the desired display mode or combination of display modes (B-Mode or M-Mode).

If the dual display mode (split screen) is desired, the L and R keys activate the Left or Rightdisplayed image.

Scan the desired anatomy. Change modes as desired. Use the mode controls to adjust the imageas necessary.

B Mode

M Mode

R

Page 15: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 9LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

B-Mode Controls

The keyboard controls that effect B-Mode are as follows:

1. TGC—Controls the gain of the received echoes at a specific depth.

2. Dynamic Range —Controls how echo intensities are converted toshades of gray scale that can be adjusted.

3. Map—Determines how the echo intensity levels received arepresented as shades of gray.

4. Frame Average —Averages previous frames of image data with thecurrent frame.

5. Edge Enhance —Brings out subtle tissue differences and boundariesby enhancing the gray scale differences corresponding to the edges ofstructures.

6. Multi Freq —Controls the penetration.

7. Scan Width —Used to widen or narrow the scan width to maximize theimage’s field of view.

8. Scroll —Used to scroll the image presentation up or down in order todisplay deeper parts of the body.

1 8

7

2 3 4 5 6

Page 16: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 10 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

B-Mode Controls (cont’d)

The keyboard controls that effect B-Mode are as follows:

9. Combi Focus —Controls the Focus Combination Value.

10. ATO—Used to optimize the image based upon a specified region ofinterest or anatomy within the image.

11. Rotation —Changes the depth at which the selected number of focalzones are optimized.

12. Reverse —The GE logo at the top of the sector wedge correspondsto the orientation mark on the probe body. When Reverse is active, theimage flips and the GE logo switches to correspond to the probe mark.

13. Depth —Controls the distance over which the B–Mode imagesanatomy.

14. B/M Gain —Controls the amount of echo information displayed in animage.

15. Biopsy —Enables the electronic biopsy zone(s) available for theactive probe.

15 9

10 11 12 13 14

Page 17: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 11LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

B-Mode Controls (cont’d)

B-Mode Optimization

Adjustments for... Do the Following... Adjustments for... Do the Following...

Image too grainy 1. Increase Dynamic Range .

2. Increase Frame Average .

3. Decrease Edge Enhance .

4. Change Map.

Image too soft 1. Decrease Dynamic Range .

2. Increase Edge Enhance .

3. Decrease Frame Average .

4. Change Map.

Image too noisy 1. Decrease B/M GAIN .

2. Decrease Dynamic Range .

3. Increase Frame Average .

4. Increase Edge Enhance .

Improve Uniformity 1. Increase the number of focal zones.

2. Decrease Scan Width .

3. Adjust TGC to compensate forattenuation.

Cystic Imaging 1. Decrease the B/M GAIN .

2. Decrease Dynamic Range .

3. Use Scan Width to reduce imagewidth.

4. Change Combi Focus to increasenumber of focal zones.

5. Optimize focal zone placement.

Technically DifficultPatients

1. Select the proper probe.

2. Maintain a lower Dynamic Range .

3. Decrease Scan Width for fasterframe rates.

Page 18: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 12 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

M-Mode Controls

Since M-Mode is basically a single B-Mode scan vector displayed over time, basic controls that affect the B-Mode displayalso affect the M-Mode display. See Quick Start 9 for comments on these controls.

TGC and Depth affect both the M-Mode and B-Mode displays.

If the scan area size is reduced and the position changed, the M-Mode cursor will follow the position change to stay withinthe displayed scan area.

The Dual Format keys work the same as in dual B-Mode, but both B-Mode and M-Mode are displayed on the left and rightsides of the screen.

The keyboard controls that effect M-Mode are as follows:

M-Mode Optimization

Adjustments for... Do the Following... Adjustments for... Do the Following...

Improve M-Mode 1. Increase/decrease B/M GAIN .

2. Increase Dynamic Range .

Increase Size of Areaof Interest

1. Use M-Mode ZOOM.

Page 19: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 13LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

M-Mode Controls (cont’d)

The keyboard controls that effect B-Mode are as follows:

1. Zoom —Enables zoom function.

2. B/M Gain —Controls the gain of the displayed timeline echoes. Toincrease/decrease, turn the control clockwise/ counterclockwise.

3. Sweep Speed —Controls the speed at which the timeline updatesacross the speed.

3

1 2

Page 20: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 14 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Reviewing Cine Images

To display cine images:

Press the Freeze key to stop image acquisition.

When Freeze is activated, rotate the cine Scroll control to display individual image frames whichwere stored by the system.

NOTE: Cine frame number is the most current image. The lower the cine frame number, theolder the image. The dynamic range and gain parameters displayed on the screen are valid forcine frame number zero only.

NOTE: One click of the Cine Scroll knob erases displayed measurements.Cine ScrollB.M.Gain

FWBW

Page 21: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 15LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Image Annotation

Keyboard Annotation

Pressing the Comment key, assigns the trackball function to controlling the cursor.

The Trackball and Arrow keys on the keyboard are used to move the cursor during theComment function. When the blinking cursor is in the desired position, comments may betyped in. Annotations are entered in the type–over, not insert mode. Be careful not to writeover text when editing.

Comment

Page 22: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 16 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Special Keys–Symbols

Some special annotation symbols can be used by activating the Blue Shift or GreenShift keys. Green Shift does not function while Blue Shift is active.

Activating Blue Shift will cause the arrow, female and male symbols to be printed onthe screen during the comment function.

Special Keys–Language Symbols

The Green Shift key enables the special symbols shown in green on the keyboard.

The green symbols shown in the lower right portion of a key will print when the GreenShift is active.

Green symbols for foreign languages can only be used with the proper designatedletters.

The Green symbols shown in the lower right portion of a key print when the GreenShift is active and that key is pressed.

The Green symbols shown in the upper half of a key will print if the normal shift keyand the appropriate key are pressed while the Green Shift is active.

The Green Shift key acts as a lock function (similar to Caps Lock key). Press again toexit the Green Shift function.

R T Y

D F G H

C V B

Shift

Æ

Ç

Shift

Page 23: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 17LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Body Patterns

Body patterns are a simple graphic of a portion of the anatomy that is frequently scanned.Press the Body Pattern key to activate Body Patterns.

A body pattern, generally displayed in the lower left corner of the screen, appears.

Cycle through the body patterns by pressing the Body Pattern key repeatedly.

A Probe Orientation Marker is used to illustrate the probe position.This marker can be placed with the Trackball and rotated with the Rotation/Focus control. Press Set to complete the body pattern selection.

Body Pattern

Focus Rotation

ECG Gain

Probe Orientation

Marker

Page 24: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 18 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Performing Basic Measurements

Basic measurements are displayed on the monitor and not automatically entered into a report page unless a specificmeasurement/calculation is selected from the sub–menu. The specific measurement/calculation can be chosen before orafter the basic measurement has been performed.

Scan the desired anatomy and press Freeze.

Press Measurement key. Type of the measurement can be chosen by pressing theMeasurement key.

MODE

Key Pressed(After Freeze key is pressed)

B M(with B)

Once DistanceEllipse

Tissue Depth

Tissue Depth(Distance)

Depth Difference

Twice Trace/Circle Time Interval

Three Times Gray Scale Echo Level Time Slope

Measurement

Page 25: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 19LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Distance (B–Mode) and Tissue Depth (M–Mode)

Perform a distance measurement such as BPD, FL, CRL, LVIDd and LVPWs.

Press Measurement once.

Use the Trackball to move the “�” cursor to the measurement start point. PressSet to fix the measurement start point cursor and to display a second cursor.

Use the Trackball to move the second “�” cursor to the measurement end point.Press Set to complete the measurement and fix the Distance value displayed onthe bottom part of the screen.

Measurement

Set

Set

Page 26: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 20 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Circumference/Area - Ellipse (B–Mode)

Perform a circumference/area using the ellipse method for calculations such asAC, HC, LVAMd, LVAd and Volume.

Press Measurement once.

Use the Trackball to move the “�” cursor to either end of the major axis of thearea to measure. Press Set to fix the start point cursor.

Use the Trackball to move the second cursor to the major axis measurement endpoint. Press the Ellipse up/down arrow key. An ellipse having an initial circleshape appears.

Use the Trackball to position the ellipse, as necessary, and to size the measuredaxis. Press Set to complete the ellipse measurement and record thecircumference and area.

Measurement

Set

Ellipse

Set

Page 27: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 21LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Circumference/Area - Trace (B–Mode)

Perform a circumference/area using the trace method for calculations such as AC,HC, LVAMd, LVAd and %stenosis.

Press Measurement twice to display a dot “�” cursor on the screen. The displayon the bottom of the screen shows the circumference in cm.

Use the Trackball to move the dot “�” cursor to the measurement start point.Press Set to change the dot “�” cursor to a “�” cursor.

Use the Trackball to trace the measurement area. The circumference displayedon the bottom of the screen will change with the tracing. Press Set to fix theCircumference value displayed on the bottom of the screen and calculate thearea.

Measurement

Set

Measurement

Set

Page 28: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 22 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Echo Level Measurement

Press Freeze to stop image acquisition.

Press Measurement three times to enable the echo level function. A box “�”cursor appears.

Use the Trackball to move the box “�” cursor over the measurement area.

Press Set to fix the Echo value displayed.Set

Measurement Measurement Measurement

Page 29: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 23LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Time Interval

Perform a time interval for calculations such as HR, PHT, and ET.

Press Measurement twice. A “�” cursor with a vertical dotted line appears whenthe cursor is in the M-Mode portion of the display.

Use the Trackball to move the cursor to the measurement start point. Press Setto fix the first cursor.

Use the Trackball to move the end cursor to the measurement end point. PressSet to complete the measurement.

Set

Measurement Measurement

Set

Page 30: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 24 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Slope

Press Freeze to stop image acquisition.

Press Measurement three times. With the cursor in the M-Mode timeline area, a“�” cursor with a vertical dotted line is displayed.

Use the Trackball to position the measurement start point.

Press Set to fix the start point.

Use the Trackball to position the measurement end point.

Press Set to complete the measurement.

Set

Measurement Measurement Measurement

Set

Page 31: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 25LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Erasing Measurements

Pressing Clear or Unfreezing the image erases all measurements and calculations on thedisplay. Measurements and calculations, however, remain on the report pages.

Pressing New Patient erases all measurements and calculations on the display and clears thereport pages.

Rotating the Cine Scroll control erases measurements.

Adding a new measurement that exceeds the maximum number of allowable measurementserases the first or oldest measurement.

Clear

New Patient

Cine ScrollB.M.Gain

FWBW

Page 32: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 26 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Report PagesSpecific diagnostic exam categories have report pages to summarize measurements andcalculations performed while in that exam category. These categories are Obstetrics, Urology and Cardiology.

A report page can be displayed by pressing Report Page key.

Report Page

Hospital Name Oper ID: mm/dd/yyID: Patient ID Name: Patient Name AGE:###LMP(GA):mm/dd/yy G## P## A## E#Ref MD: NOTE: POS: PLAC: Page: 1/4

FLUID: PREVIA?: GRADE:

MEASUREMENT CUA GP% CALCULATIONBPD(HADLOCK) Y ###.#mm ##W#D ±#D CI (70–86) ###%HC(HADLOCK) Y ###.#mm ##W#D ±#D FL/BPD (71–87) ###%OFD(HC) ###.#mm ##W#D ±#D FL/AC (20–24) ###%AC(HADLOCK) Y ###.#mm ##W#D ±#D FC/HC (??.?–??.?) ###%TAD(AC) ###.#mm ##W#D ±#D HC/AC (???–???) ####APD(AC) ###.#mm ##W#D ±#D FL(HADLOCK) Y ###.#mm ##W#D ±#D EFW(FL,AC,HC,BPD) CRL(HADLOCK) Y ###.#mm ##W#D ±#D ####g ±###g ( #lb ##oz )GS(HELLMAN) Y ###.#mm ##W#D ±#D GP>##% (HADLOCK)

AFI(mm) ### HEART RATE(BPM) ###

BIOPHYS ##/## GA(OPE) :##W#D EDD(GA) :mm/dd/yyCUA: :##W#D EDD(CUA):mm/dd/yy

COMMENTS:

EXIT

Page 33: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 27LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Recording Images

Image Memory

Press Image Memory to temporarily store the maximum of 10 images in system memory.

Pressing Image Memory after the warning deletes the oldest image in system memory and savesthe current displayed image.

Press Image Recall to display the last image stored in system memory.

Use Trackball or Ellipse key to select the desired image to recall.

After recall, comments, measurements and body patterns can be added to the image. The image isthen available for printing and archiving to MOD(option).

Printing

Use the Record1 or Record2 keys to print the image/report page a B/W Video Page Printer.

The function of these keys was established during installation.

Image Memory

Image Recall

Record1

Record2

Page 34: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 28 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

VCR Functions

The Ext Video key is used to display the External Video on the LOGIQ�200 PRO Seriesscreen.

Probe Cleaning/Disinfection

Probes should be cleaned and disinfected after each use. The Probe Immersion Levels are provided with the probe butare also pictured in the Probes chapter of the User Manual

Ext.Video

Page 35: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 29LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

Power Off

To power down the system, press down on the Power switch.

While the system is running during this power down process, do NOT turn off thecircuit breaker in the back of the machine and do NOT unplug the system from thewall outlet.x

If the system has not turned off five minutes after pressing the power switch off,listen for hard drive activity. If there is no hard disk drive activity, the circuitbreaker on the bottom of the power supply can be used to turn off the system. DoNOT turn off the circuit breaker while the hard disk is working.

IO

Page 36: Quick Startfrankshospitalworkshop.com/equipment/documents...Quick Start 1 LOGIQ 200PRO Series Quick Start 2234974–100 Rev. 0 Introduction The Quick Start Guide (TRANSLATED) provides

Quick Start 30 LOGIQ�200PRO Series Quick Start2234974–100 Rev. 0

This page left blank intentionally.