medical physics imaging pet ct spect ct lecture

44
Lecture 14: Emission Tomography III Shahid Younas NUCLEAR IMAGING Emission Tomography III Single Photon Emission Computed Tomography (SPECT)

Upload: shahid-younas

Post on 17-Aug-2015

69 views

Category:

Healthcare


2 download

TRANSCRIPT

Page 1: Medical Physics Imaging PET CT SPECT CT Lecture

Lecture 14: Emission Tomography III

Shahid Younas

NUCLEAR IMAGING

Emission Tomography III

Single Photon Emission Computed Tomography (SPECT)

Page 2: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction

Lecture 14: Emission Tomography III

X- or gamma rays that must traverse long paths through the patient produce

fewer counts, due to attenuation, than those from activity closer to the near

surface of the patient.

Page 3: Medical Physics Imaging PET CT SPECT CT Lecture

Introduction-Attenuation correction

Lecture 14: Emission Tomography III

Images acquired with SPECT has,

Poor spatial resolution

Apparent decrease in activity

Page 4: Medical Physics Imaging PET CT SPECT CT Lecture

Introduction-Attenuation correction

Lecture 14: Emission Tomography III

Transverse image slices of a phantom with a

uniform activity distribution will show a

gradual decrease in activity toward the center.

Page 5: Medical Physics Imaging PET CT SPECT CT Lecture

Introduction-Attenuation correction

Lecture 14: Emission Tomography III

The primary mechanism for attenuation in tissue is Compton

Scattering.

This changes photon direction with loss of energy.

The change of direction results in missed count.

Page 6: Medical Physics Imaging PET CT SPECT CT Lecture

Introduction-Attenuation correction

Lecture 14: Emission Tomography III

The effects of attenuation

are more intense at lower

energies but are still

significant at the highest

energy value.

Page 7: Medical Physics Imaging PET CT SPECT CT Lecture

Introduction-Attenuation correction

Lecture 14: Emission Tomography III

Summing two planar

projection images

separated by 180.

Page 8: Medical Physics Imaging PET CT SPECT CT Lecture

Introduction-Attenuation correction

Lecture 14: Emission Tomography III

The magnitude of

attenuation effect depends

on the tissue type.

Page 9: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction

Lecture 14: Emission Tomography III

Thus, to accurately represent the activity distribution measured with SPECT,

it is necessary to accurately correct for the effects of attenuation.

Page 10: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Approximate methods are available for attenuation correction.

Change Method, assumes a constant attenuation coefficient

throughout the patient.

Over-undercompensate-as attenuation is not uniform

Page 11: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Constant Attenuation Coefficient

A1 A1 A1

A1 A1 A1

A1 A1 A1

Page 12: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Some SPECT cameras have radioactive sources to measure the

attenuation through the patient,

After acquisition, the transmission projection data are reconstructed

to provide maps of tissue attenuation characteristics across transverse

sections of the patient, similar to x-ray CT images.

Page 13: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Some SPECT cameras have radioactive sources to measure the

attenuation through the patient,

Finally these attenuation maps are used during SPECT image

reconstruction to provide attenuation-corrected SPECT images.

Page 14: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Transmission sources are available in several configurations,

Scanning Collimated Line Sources

Fixed Line Sources

Page 15: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Transmission data usually acquired simultaneously with the

acquisition of the emission projection data,

Performing the two separately poses significant problems in

the spatial alignment of the two data sets.

Page 16: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Radionuclide used for transmission measurements is chosen to have

primary gamma-ray emissions that differ significantly in energy from

those of the radiopharmaceuticals.

Separate energy windows are used

Page 17: Medical Physics Imaging PET CT SPECT CT Lecture

Attenuation correction Techniques

Lecture 14: Emission Tomography III

Scattering of the higher energy photons in the patient and in the

detector causes some cross-talk in the lower energy window.

AC using transmission sources is used Myocardial perfusion

imaging.

AC using transmission sources is promising but it is still under

development.

Page 18: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Collimator

Lecture 14: Emission Tomography III

Most commonly used is the high-resolution parallel-hole collimator

Fan-beam collimators mainly used for brain SPECT

FOV decreases with distance from collimator

Page 19: Medical Physics Imaging PET CT SPECT CT Lecture

Multihead SPECT Cameras

Lecture 14: Emission Tomography III

Two or three scintillation camera heads reduce limitations imposed

by collimation and limited time per view.

Y-offsets and X- and Y-magnification factors of all heads must be

precisely matched throughout rotation.

Page 20: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Performance

Lecture 14: Emission Tomography III

Spatial resolution

X- and Y-magnification factors and multi-energy spatial registration

Alignment of projection images to axis-of-rotation

Uniformity

Camera head tilt

Page 21: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Spatial resolution

Lecture 14: Emission Tomography III

Can be measured by acquiring a SPECT study of a line source

(capillary tube filled with a solution of Tc-99m, placed parallel to

axis of rotation).

FWHM of the line sources are determined from the reconstructed

transverse images (ramp filter).

Page 22: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Spatial resolution

Lecture 14: Emission Tomography III

National Electrical

Manufacturers

Association (NEMA)

specifies a cylindrical

plastic water-filled

phantom, 22 cm in

diameter, containing

3 line sources

Page 23: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Spatial resolution

Lecture 14: Emission Tomography III

NEMA spatial resolution measurements are primarily determined by

the collimator used.

Tangential resolution 7 to 8 mm FWHM for LEHR

central resolution 9.5 to 12 mm

radial resolution 9.4 to 12 mm

Page 24: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Spatial resolution

Lecture 14: Emission Tomography III

NEMA measurements not necessarily representative of clinical

performance

Studies can be acquired using longer imaging times and closer orbits

than would be possible in a patient.

Page 25: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Spatial resolution

Lecture 14: Emission Tomography III

NEMA measurements not necessarily representative of clinical

performance

Studies can be acquired using longer imaging times and closer orbits

than would be possible in a patient.

Page 26: Medical Physics Imaging PET CT SPECT CT Lecture

SPECT Spatial resolution

Lecture 14: Emission Tomography III

NEMA measurements not necessarily representative of clinical

performance

Studies can be acquired using longer imaging times and closer orbits

than would be possible in a patient.

Filters used for clinical studies have lower spatial frequency cutoffs

than the ramp filters used in NEMA measurements.

Page 27: Medical Physics Imaging PET CT SPECT CT Lecture

Comparison with conventional planar scintillation camera imaging

Lecture 14: Emission Tomography III

In theory, SPECT should produce spatial resolution similar to that of

planar scintillation camera imaging.

In clinical imaging, its resolution is usually slightly worse.

Camera head is closer to patient in conventional planar imaging than

in SPECT.

Page 28: Medical Physics Imaging PET CT SPECT CT Lecture

Comparison with conventional planar scintillation camera imaging

Lecture 14: Emission Tomography III

Short time per view of SPECT may mandate use of lower resolution

collimator to obtain adequate number of counts.

In planar imaging, radioactivity in tissues in front of and behind an

organ of interest causes a reduction in contrast.

Page 29: Medical Physics Imaging PET CT SPECT CT Lecture

Comparison with conventional planar scintillation camera imaging

Lecture 14: Emission Tomography III

Main advantage of SPECT is markedly improved contrast and

reduced structural noise produced by eliminating the activity in

overlapping structures.

SPECT also offers promise of partial correction for effects of

attenuation and scattering of photons in the patient

Page 30: Medical Physics Imaging PET CT SPECT CT Lecture

Magnification factors

Lecture 14: Emission Tomography III

The X- and Y-magnification factors, often called X and Y gains,

related distances in the object being imaged, in the x and y directions, to

the numbers of pixels between the corresponding points in the resultant

image.

Page 31: Medical Physics Imaging PET CT SPECT CT Lecture

Magnification factors

Lecture 14: Emission Tomography III

Magnification factors determined from a digital image of two point

sources placed against the camera’s collimator

If X- and Y-magnification factors are unequal, the projection images will

be distorted in shape, as will coronal, sagittal, and oblique images.

Page 32: Medical Physics Imaging PET CT SPECT CT Lecture

COR calibration

Lecture 14: Emission Tomography III

The axis of rotation (AOR) is an imaginary reference line about

which the head or heads of a SPECT camera rotate.

If a radioactive line source were placed on the AOR, each projection

image would depict a vertical straight line near the center of the

image.

Page 33: Medical Physics Imaging PET CT SPECT CT Lecture

COR calibration

Lecture 14: Emission Tomography III

This projection of the AOR into the image is called the center of

rotation (COR).

Ideally, the COR is aligned with the center, in the x-direction, of each

projection image.

Page 34: Medical Physics Imaging PET CT SPECT CT Lecture

COR calibration

Lecture 14: Emission Tomography III

Misalignment may be mechanical or electronic.

Camera head may not be exactly centered in the gantry.

Page 35: Medical Physics Imaging PET CT SPECT CT Lecture

COR calibration

Lecture 14: Emission Tomography III

COR Degradation and Sinogram

Page 36: Medical Physics Imaging PET CT SPECT CT Lecture

COR calibration

Lecture 14: Emission Tomography III

COR misalignment causes a loss of spatial resolution in the resultant

transverse images.

Large misalignment cause a point source to appear as “doughnut”.

Doughnut are not centered in the image so can be distinguished from

“ring” artifacts produced by non-uniformities.

Page 37: Medical Physics Imaging PET CT SPECT CT Lecture

COR calibration

Lecture 14: Emission Tomography III

COR alignment is assessed by placing a point source or line source

in the camera field of view.

Projected imaged and or sinogram is analyzed by the camera’s

computer.

Page 38: Medical Physics Imaging PET CT SPECT CT Lecture

COR calibration

Lecture 14: Emission Tomography III

Misalignment may be corrected by shifting each image in the x-

direction by the proper number of pixels prior to filtered back-

projection

If COR misalignment varies with camera head angle, it can only be

corrected if computer permits angle-by-angle corrections.

Page 39: Medical Physics Imaging PET CT SPECT CT Lecture

Uniformity

Lecture 14: Emission Tomography III

Nonuniformities that are not apparent in low-count daily uniformity

studies can cause significant artifacts in SPECT.

Artifact appears in transverse images as a ring centered about the

AOR.

Page 40: Medical Physics Imaging PET CT SPECT CT Lecture

Uniformity

Lecture 14: Emission Tomography III

Cylinder filled with a uniform

radionuclide solution showing

a ring artifact due to non-

uniformity.

Page 41: Medical Physics Imaging PET CT SPECT CT Lecture

Uniformity

Lecture 14: Emission Tomography III

Primary intrinsic causes of non-uniformity are,

a. Spatial non-linearities

stretch the image in some areas

reducing the local count density

compress other areas of the images

Increasing the count density

a. Local variation in the light collection efficiency

Page 42: Medical Physics Imaging PET CT SPECT CT Lecture

Uniformity

Lecture 14: Emission Tomography III

Lookup table can not correct,

Local variations in detection efficiency such as dents or

manufacturing defects in the collimators.

Page 43: Medical Physics Imaging PET CT SPECT CT Lecture

Uniformity

Lecture 14: Emission Tomography III

High-count uniformity images used to determine pixel correction

factors,

At least 30 million counts for 64 x 64 images

At least 120 million counts for 128 x 128 images

Collected every 1 or 2 weeks; separate images for each camera head

Page 44: Medical Physics Imaging PET CT SPECT CT Lecture

Camera head tilt

Lecture 14: Emission Tomography III

Camera head or heads must be exactly parallel to the AOR.