measurement of vibration

Upload: uma-rao

Post on 05-Apr-2018

238 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/2/2019 Measurement of Vibration

    1/19

    Measurement of Vibration

    A.Rama Rao

    Vibration Laboratory SectionReactor Engineering Division

    Bhabha Atomic Research CentreMumbai-400085

    E-Mail: [email protected]

    There are three quantities, which are of interest in vibration studies. In the case of

    sinusoidal motion, the three quantities are related by the frequency of oscillation. The

    instantaneous position or displacement of a body can be mathematically described as,

    X = Xpeaksin( 2t/T) = Xpeaksin (2ft) = Xpeaksin(t)

    where = 2f = angular frequency.

    Xpeak=Maximum displacement.

    f = 1/T and T = period of oscillation

    As the velocity of motion is the time rate of change of motion, it can be expressed as,

    v = dx/dt = Xpeakcos (t) = Vpeakcos(t) = Vpeaksin(t+ /2)

    and acceleration which is again time rate of change of velocity, it can be express as,

    a = dv/dt = dx2

    /dt2

    = - 2

    Xpeaksin(t) = - Apeaksin(t) = Apeaksin(t+ )

    It can be seen that the form and period of vibration remain the same except that the

    velocity leads the displacement by a phase angle of 90 o and acceleration leads velocity

    again by 90o

    1.0 Amplitude quantification

    Common sense says that vibration severity is indicated by correct measurement of

    vibration amplitude. Figure 1 shows a sinusoidal signal with amplitude, period and phase.

    The peak amplitude is used to describe for example the displacement of a shaft, which

    helps in estimating the position of the shaft with respect to the clearance of the bearing

    supporting the shaft. Whenever there is vibration of a body within a given clearance, peak

    value measurement is advisable.

  • 8/2/2019 Measurement of Vibration

    2/19

    Figure-1 A Periodic signal

  • 8/2/2019 Measurement of Vibration

    3/19

    Figure-2 Amplitude Characteristics of a Periodic Signal

    Measurement of Root Mean Square value of a signal has important significance in

    expressing the power content of the vibration and hence the damage potential of

    vibration. It gives the DC equivalent of oscillating or alternating signal. It is

    mathematically expressed as

    PEAKRMSXX

    2

    1

    Other amplitude quantifications are Form factor (Ff) and Crest factor (Fc).

    Ff= RMS/AVERAGE

    Fc = PEAK/RMS

    These factors give some indication of the wave shape. For example for pure harmonic

    motion Ff= 1.11 and Fc = 1.41. Deviation from these numbers indicates that the signal is

    not pure harmonic. Of the two factors, crest factor has important diagnostic importance.

    For example, while monitoring health of bearings, if the crest factor shows an increasing

    trend then it is considered to be an indication of impending damage to the bearing. This

    happens for example when there is single vibration component caused by a faulty

    bearing. This defect goes undetected by RMS measurement but crest factor easily picks

    up the in defect in the bearing. Figure 2 illustrates the relative significance of these

    measurement quantities.

    2.0 Transducer specification

    Transducers help us in making dependable measurement when their selection for a

    particular type of measurement is carefully followed. A very common mistake made in

    this regard is that one transducer is used for all types of measurement. During normal

    operating condition this mistake may not be too costly but when vibration severity

    increases, the error may be too big. For example, transducer tuned to measure machinery

    vibration strictly cannot be used for assessing piping vibration. The error will be highly

    non-conservative.

  • 8/2/2019 Measurement of Vibration

    4/19

    2.1 Eddy Current Probes

    Eddy current probes, which measure displacements with out contacting the vibrating

    surface, have been standardized in terms of gap between the target and the probe, its

    measurement range, sensitivity and the connecting electronics. The probe design that

    follow standard API 610 give an out put of 8 V/mm for a three meter cable between the

    probe and the proximeter that supplies to the probe the carrier signal for modulation

    during vibration of the target. The proximeter also demodulates the measured signal and

    gives voltage out put. The proximeter is supplied with external DC voltage of 18 to 24 V.

    Figure 3 shows typical mounting details of the probe inside a bearing housing. Care must

    be taken while positioning the tip of the probe with respect to the shaft. As shown in

    Figure 4, no metal should be present within 4 mm of the probe face except the target

    surface.

  • 8/2/2019 Measurement of Vibration

    5/19

    Figure-3 Eddy Current Probe Mounting within the Bearing Housing

    Figure 4 Schematic of Proper installation of Eddy Probe.

    Eddy probes are also available to measure the nano micron deflection of the outer race of

    the ball bearings. These probes are highly sensitive and also require special mounting

    within the bearing housing. Figure 5 shows high sensitivity probe to measure nano level

    deflection of outer race of a bearing.

  • 8/2/2019 Measurement of Vibration

    6/19

    Figure-5. High Sensitivity Eddy current Probe

    The probe tip diameter basically decides the range of measurement. They normally are in

    the range of 6mm to 12mm. The magnetic field generated at the tip of the probe need to

    be directed towards the target with out much of loss. If 12 mm probe is used for

    measuring vibration of 30 mm shaft then there is a possibility of error in measurement.

    The orbit plot has very high potential in machinery diagnosis. For example, figure 6

    shows the orbit plot of a shaft measured in x-y direction as shown in figure 3.

  • 8/2/2019 Measurement of Vibration

    7/19

    Figure-6 Orbit plot or X-Y from Eddy Probes

    Figure 6 shows some of the orbit shapes indicating severity of misalignment between to

    coupled rotors and figure 7 shows 3-D plot along with orbit plot used for recognizing

    resonance during start trial of a machine.

  • 8/2/2019 Measurement of Vibration

    8/19

    Figure-7 Orbit plots indicating severity of misalignment

    Figure-8 3-D plot and orbit plot.

    2.2 Accelerometers

    As brought out earlier, accelerometers are widely used for measuring all the three

    vibration quantities. Piezoelectric based accelerometers are widely used type of sensors.

    The piezo material develops charge when subjected to force. Quartz and Rochelle salt are

  • 8/2/2019 Measurement of Vibration

    9/19

  • 8/2/2019 Measurement of Vibration

    10/19

    when excited from low frequency up to its resonance frequency. Normally for

    general-purpose accelerometers the resonance frequency is in the range of 30

    to 90 KHz. As a thumb rule the upper limit of the useful range is taken as 1/3rd

    the resonance. That is 10 to 30 KHz. Figure shows the band useful frequency

    range. Accelerometers are not capable of a true DC response. The

    preamplifier decides the lower limit of the frequency.

    Figure-10 Frequency response of an Accelerometer

    Up to 1 Hz is normally considered acceptable without any amplitude or phase

    distortion. If a measurement involves higher frequencies than 30 KHz, then

    accelerometers with resonance frequency higher than 100 KHz must be used

    and if low frequency measurement is to be carried out involving low

    displacements, then accelerometers are not suitable. Double integration of

    acceleration signal for estimating low displacement gives unreasonably high

    displacement, which is due to noise in the integration. This is one commonly

    made mistake and so there is a need to exercise caution.

    2.2Sensitivity: It is the magnitude of voltage developed across its outputterminals when subjected to certain acceleration. This is specified as V/m/s

    2

  • 8/2/2019 Measurement of Vibration

    11/19

    or pico coulomb/ m/s2. Ideally, the general feeling that higher the sensitivity

    the better suits low level of acceleration. However for measurements

    involving vibration and intermittent shock for example in automobile on a

    bumpy road, high sensitivity could lead to saturation of input signal and hence

    distortion. High sensitivity accelerometer normally has limited high frequency

    range and accelerometers suited for high frequency measurement are of lower

    sensitivity. What is important is to make an estimation of what is the likely

    level of voltage that may be received from the accelerometer. Based on this

    the electronics, and other condition for storage device may be adjusted.

    2.3Dynamic Range: When it is required to measure abnormally low and highvibration levels, the dynamic range needs to be considered. Now-a-daysaccelerometers are available with dynamic range from 60dB to 100 dB.

    Dynamic range of 60 dB measures from says 1mm/s2 to 1000 mm/s2and 100

    dB range measures up to 100,000 mm/s 2. The dynamic range must be seen in

    combination with sensitivity to assess the performance of the sensor in the

    lower or upper range of the amplitude. Figure 11 shows the illustration of

    dynamic range.

    Figure-11 Illustration of dynamic range of accelerometer

  • 8/2/2019 Measurement of Vibration

    12/19

    Upper limit of the dynamic range is determined by structural strength of the

    accelerometer. Accelerometers are available to measure up to 100,000 g,

    which is well into the range of mechanical shock.

    2.4Mass: Normally accelerometers are lightweight and compact in design.However when they are used in large numbers say on aeroplane wings or

    panels, the mass loading could be an issue. Additional mass changes the level

    response and the frequency of the structure thus invalidating the measurement

    results. The change in the response can be corrected with the following

    formula.

    as = am (ms+ma)/ms and fs =fm {(ms+ma)/ms }

    1/2

    where

    as = acceleration with out mass loadingfs = frequency with out mass loading

    ms = effective mass of the structure

    ma = mass of accelerometer

    As a general rule, mass of the accelerometer should not be greater than one

    tenth of the effective (dynamic) mass of the structure.

    2.5Transient Response: Accelerometers are also used for measuring shortduration shocks caused by impact of two bodies or say during drop testing of

    packages etc. The two shock parameters that need to be accurately measured

    are time duration of shock and the amplitude of shock. The shock duration

    could be smaller than the natural period of the system. The most likely error

    could be in estimating the time duration due to retention of charge by the

    accelerometer when shock is suddenly applied and in estimating the amplitude

    when the resonance of the accelerometer is excited. The supplier provides the

    safe lower and upper limit of shock measurement.

    2.6Temperature limits: Since property of the piezo material is dependent on thetemperature, it is advisable to stick to the specified limit. Beyond the limit the

    piezo material depolarizes causing permanent loss or change in sensitivity.

  • 8/2/2019 Measurement of Vibration

    13/19

    Some times heat sinks (finned washer) or mica washers are used to reduce

    heat transmission. Sensitivity versus temperature curves is provided by the

    suppliers to estimate the error in measurement due to temperature effect.

    2.7Phase Response: Any shift in the phase between mechanical input andresulting electrical output of the accelerometer indicates time delay between

    input and output and hence distortion of the mechanical input. At frequencies

    below the mounted resonance there should be no phase shift. Figure 12

    illustrated the phase and amplitude relationship.At resonance frequency, the

    motion of the seismic mass inside the accelerometer lags that of the base

    resulting in phase distortion

    Figure-12 Phase and amplitude relation

    2.8 Environmental condition: Most of the accelerometers withstand accumulated

    gamma radiation dose of 2 M Rad with out significant degradation. For

    permanent installation under nuclear radiation, accelerometers to withstand

    accumulated dose of 100 M Rad is available. Magnetic sensitivity of piezo

    sensors is very low. Measurements on power generators and magnetic devices

  • 8/2/2019 Measurement of Vibration

    14/19

    are possible with out distortion. Acoustic noise present in the machine is not

    sufficient to induce any error in the measurement.

    3.0 Piezoelectric Accelerometers

    Since accelerometers are widely used in the industry, let us understand its

    working. Figure 13 shows a general arrangement of measurement using an

    accelerometer.

    Figure-13 General scheme of measurement using accelerometer

    The output impedance of a piezo element is very high and so cannot be directly

    connected to a read out device, which normally have low input impedance. When

    connected, it can greatly reduce the sensitivity as well as its frequency range. To

    eliminate this connectivity problem, the accelerometer out put is connected to a

    pre amplifier, which has high input impedance and low output impedance. Figure

    14 shows the equivalent circuit diagram of an accelerometer, which is basically a

    charge generator q coupled with an internal capacitance C. e is the voltage

    across the capacitor. Figure 15 show the equivalent diagram of accelerometer plus

    cable plus charge amplifier.

    Figure-14 Equivalent circuit for an Accelerometer

    Vibration Preamplifier Analyzer Recorder

    Pick-up

  • 8/2/2019 Measurement of Vibration

    15/19

    Figure-15 Equivalent diagram for an Accelerometer + Cable + Charge Amplifier

    The pre amplifiers are either voltage type in which the out put voltage is

    proportional to input voltage from the sensor or charge type in which the out put

    voltage is proportional to input charge. When voltage amplifier is used, the

    overall system is very sensitive to changes in cable capacitance that is changes in

    the cable length between the sensor and the pre amplifier. However, with charge

    pre amplifiers the out put remains unaffected by the length of the cable. Hence

    one has to be careful in knowing the type of preamplifiers before the

    measurement or even while specifying for purchase of vibration equipments.

    Some preamplifier includes integrators to convert acceleration signal to velocity

    or displacement proportional signal.

    4.0 Cables

    The signal carrying cables are one of the loose links in the chain of vibration

    measuring setup. They have to be correctly and carefully used for dependable

    diagnostics. There are many instances wherein bad cables have given wrong

    reading. In spite of standardized connecting cables, over the years of use,connecting errors creep in. Since accelerometers are high impedance device,

    certain problem may also arise due to cable noise. The noise can originate either

    from mechanical motion of the cable, also called tribo-electric effect or from

    the ground loop. Mechanical motions of the cable during the measurements result

    in change in the charge. Such an effect disturb cause disturbance in the low

  • 8/2/2019 Measurement of Vibration

    16/19

    frequency range. This can be avoided by the practice of clamping the cable firmly

    to arrest relative movement between several layers of insulation in the cable as

    shown in figure 16.

    Figure-16 Illustration of fastening cables to eliminate triboelectric noise

    The second source of noise is the hum picked up from the mains supply by

    ground loop in the chain of measuring device. Ground loop currents flow in the

    shielded layer of the cables because of slight difference in the electrical potential

    of grounding points such as at the accelerometer base and the electrical ground of

    the read out device. The best way to eliminate ground loop is to ensure that the

    entire system is grounded at one point meaning electrically isolate the base of the

    accelerometer as shown in figure 17.

    Figure-17 Proper connection to avoid Ground loop

  • 8/2/2019 Measurement of Vibration

    17/19

    Electromagnetic noise can also pose problem when the cable is laid in the vicinity

    of running electrical machinery. If this cannot be avoided, then at least double

    shielded cable must be used to reduce the pick up noise.

    5.0 Mounting the accelerometer

    Selection of the correct mounting arrangement plays a significant roll in the use of

    vibration measurement as its mounting decides the mounted resonance frequency

    and hence its useful frequency range of measurement.The popular methods are

    stud mounted, wax mounted magnet mounted, adhesive mounted and hand probe

    type. The mounting type also depends on the size of the accelerometers and

    position of measurement. For example big size accelerometer cannot be mounted

    with wax when it has to be used for measuring on inclined or vertical plane of a

    machine for the fear of dislodging or slipping downward. However miniature

    sensors can be mounted with wax in any plane. The most popular method is probe

    type mounting because it needs minimum preparation and is quick to use.

    However it is most error prone especially when the measurement involves high

    frequency. Probe type mounting drastically reduces the useful frequency range.

    Figure-18 Illustration of Probe type and Stud Mounting

  • 8/2/2019 Measurement of Vibration

    18/19

    Figure-19 Change of Response of Accelerometer for different type of mounting

    Figure 18 and 19 illustrate the two extreme mounting methods of accelerometer

    and its effect on the frequency response. Looking at the reduction of high

    frequency response during probe type of mounting of accelerometers, one has to

    be careful. Due to ease of measurement with probe type of mounting, there is a

    tendency to make error especially while monitoring high frequency vibration say

    in a bearing.

    6.0 Accelerometer Calibration

    Accelerometer calibration is to ensure the accuracy, reliability and repeatability of

    measurement. Calibration helps to trace measured vibration values to the physical

    standard. There are several reasons for performing a calibration. The most

    important being the contractual reasons as an evidence of the accuracy of the

    sensor, possibly with reference to international standards. At times calibration is

    required to when performance of the sensor in a particular environment has not

  • 8/2/2019 Measurement of Vibration

    19/19

    been documented. System checking also forms an important part of the calibration

    process, particularly in measurement systems consisting of many instruments.

    The most common question raised by users is how often the sensor should be sent

    for calibration. While the manufacturers say that ideally accelerometers do not

    need calibration for several years if they are used as per the manufactured

    specification. However, we know that the accelerometers are subject to rough

    handling in shop floor. If not the sensor per se, at least the connected cables,

    connectors, switches in the readout meters get bad. At least for the system as a

    whole recalibration is required. Then the question is how often the calibration

    should be done. This is best judge is the user himself. Depending on the usage and

    duty cycle of the accelerometer and its electronics, recalibration can be carried at

    least once in year or earlier.

    There are different levels of calibration possible such as primary, secondary, and

    working reference type. In primary calibration, the sensitivity of the

    accelerometer is established in terms of fundamental or derived units for physical

    quantities of SI systems. Such transducers are kept in National Standard Institutes.