legendplex™ · legendplextm assay buffer 1 bottle 25 ml 77562 legendplextm wash buffer, 20x 1...
TRANSCRIPT
LEGENDplex™Mul�-Analyte Flow Assay Kit
Enabling Legendary Discovery™
For Accurate Quantification of Multiple Human Th(T helper Cell) Cytokines from Cell Culture Supernatant,
Serum, Plasma and Other Biological Samples
Please read the entire manual before running the assay
BioLegend.com
LEGENDplex™Mul�-Analyte Flow Assay Kit
Enabling Legendary Discovery™
Cat. No. 740102
Human Cytokine Panel 2(13-plex)
Please read the entire manual before running the assay.
BioLegend.com
For Research Purposes Only. Not for use in diagnostic or therapeutic procedures. Purchase does not include or carry the right to resell or transfer this product either as a stand-alone product or as a component of another product. Any use of this product other than the permitted use without the express written authorization of BioLegend is strictly prohibited.
It is highly recommended that this manual be read in its entirety before using this product. Do not use this kit beyond the expiration date.
biolegend.com1
LEGENDplex™ Human Cytokine Panel 2
Table of Contents Page
Chapter 1: KIT DESCRIPTION..................................................
Introduction……………………………………………..........................
PrincipleoftheAssay……………………....……………....….…......
BeadsUsage...........................................………..……………...
StorageInformation…………………………………….......…..........
MaterialsSupplied………………….....……………….................…
MaterialstobeProvidedbytheEnd-User……...........……...
Precautions.................................……………………................
Chapter 2: ASSAY PREPARATION.............................................
SampleCollectionandHandling…………………………............
ReagentsPreparation………………………………………...............
StandardPreparation........................................................
DilutionofSamples……...........……......................................
Chapter 3: ASSAY PROCEDURE..................................................
PerformingtheAssayUsingaFilterPlate……………….........
PerformingtheAssayUsingMicrotubes……………….............
Chapter 4: FLOW CYTOMETER SETUP.......................................
Setup Procedure for FACSCalibur with Dual Laser..............
Setup Procedure for FACSCalibur with a Single Laser........
Setup Procedure for BD FACSAriaTM, FACSCantoTM and LSR Series..........................................................................
Setup Procedure for Other Flow Cytometers ...................
Chapter 5: DATA ACQUISITION AND ANALYSIS.........................
DataAcquisition..................................................................
Data Analysis......................................................................
3
3
3
4
6
6
7
8
9
9
9
10
11
12
12
15
18
19
23
30
34
35
35
36
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
2
Chapter 6: FREQUENTLY ASKED QUESTIONS...........................
Chapter 7: ASSAY CHARACTERIZATION..........................................
TypicalData……………….……………………………………………........
37
41
41
41
42
42
43
44
45
46
48
51
52
AssaySensitivity...……………………………………………………..…..
Cross-Reactivity……………………………………………………..........
Accuracy.............................................................................
LinearityofDilution………………………………………………..........
Intra-AssayPrecision……………………………………...................
Inter-AssayPrecision……………………………………...................
BiologicalSamples…………………………………………….………....
TROUBLESHOOTING...............………………………………………………....
PLATE MAP....................................................................................
RACK MAP....................................................................................
biolegend.com3
LEGENDplex™ Human Cytokine Panel 2
Chapter 1: KIT DESCRIPTION
Introduction
Cytokines, which are soluble extracelluar proteins or glycoproteins, are key in-tercellular regulators and mobilizers of cells. They are secreted by immune cells and act on other cells to coordinate appropriate immune responses. They are crucialtoinnateandadaptivehostdefenses,cellgrowth,celldeath,differentia-tion,aswellasdevelopmentandrepairprocesses.
The Human cytokine panel 2 isamultiplexbead-basedassaypanel,usingfluorescence–encodedbeadssuitableforuseonvariousflowcytometers.Thispanelallowssimultaneousquantificationof13humancytokines,including TSLP, IL-1α,IL-1β,GM-CSF, IFN-α2,IL-23,IL-12p40,IL-12p70,IL-15,IL-18,IL-11,IL-27,and IL-33.ThisassaypanelprovideshigherdetectionsensitivitiesandbroaderdynamicrangesthantraditionalELISAmethods.Thepanelhasbeenvalidatedbydetectingexpectedchangesinbiologicalsamples.
The Human Cytokine Panel2isdesignedtoallowflexiblecustomizationwithinthe panel. Please visit www.biolegend.com/legendplex formoreinformationonhow to mix and match within the panel.
The Human Cytokine Panel2anditscustomsubpanelsarespecificallydesignedfortheaccuratequantificationofmultiplehumancytokines from cell culture supernatant, serum, plasma and other biological samples.
This assay is for research use only.
Principle of the Assay
BioLegend’s LEGENDplexTM assays are bead-based immunoassays using the same basic principle as sandwich immunoassays.
Beadsaredifferentiatedbysizeandinternalfluorescenceintensities.Eachbeadsetisconjugatedwithaspecificantibodyonitssurfaceandservesasthecap-turebeadsforthatparticularanalyte.Whenaselectedpanelofcapturebeadsismixedandincubatedwithasamplecontainingtargetanalytesspecifictothecaptureantibodies,eachanalytewillbindtoitsspecificcapturebeads.Afterwashing,abiotinylateddetectionantibodycocktailisadded,andeachdetec-tionantibodyinthecocktailwillbindtoitsspecificanalyteboundonthecap-turebeads,thusformingcapturebead-analyte-detectionantibodysandwiches.Streptavidin-phycoerythrin (SA-PE) is subsequently added, which will bind to thebiotinylateddetectionantibodies,providingfluorescentsignalintensitiesinproportiontotheamountofboundanalytes.
Sincethebeadsaredifferentiatedbysizeandinternalfluorescenceintensity
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
4
onaflowcytometer,analyte-specificpopulationscanbesegregatedandPEfluorescentsignalquantified.Theconcentrationofaparticularanalyteisdeter-mined using a standard curve generated in the same assay.
Beads Usage
The Human Cytokine Panel 2 uses two sets of beads. Each set has a unique size thatcanbeidentifiedbasedontheirforwardscatter(FSC)andsidescatter(SSC)profiles(BeadsAandBeadsB,Figure1).Eachbeadsetcanbefurtherresolvedbasedontheirinternalfluorescenceintensities.Theinternaldyecanbede-tectedusingFL3,FL4,orAPCchannel,dependingonthetypeofflowcytometerused.ThesmallerBeadsAconsistsof6beadpopulationsandthelargerBeadsBconsistsof7beadpopulations(Figure2-3).
Usingatotalof13beadpopulationsdistinguishedbysizeandinternalfluo-rescent dye, the Human CytokinePanel2allowssimultaneousdetectionof13cytokinesinasinglesample.Eachanalyteisassociatedwithaparticularbeadset as indicated (Figures 2-3 and Table 1).
Figure1.BeadsDifferentiatedbySize
Beads A = Smaller beads
Beads B = larger beads
Figure2.BeadsAClassificationbyFL4
biolegend.com5
LEGENDplex™ Human Cytokine Panel 2
Figure3.BeadsBClassificationbyFL4
For Beads usage in the panel, please refer to Table 1 below:
Table1.PanelTargets,BeadIDandTopStandardConcentration
Target Bead ID*Top Standard Concentration
(ng/mL)
TSLP A4 10
IL-1α A5 10
IL-1β A6 10
GM-CSF A7 10
IFN-α2 A8 10
IL-23 A10 10
IL-12p40 B2 50
IL-12p70 B3 10
IL-15 B4 50
IL-18 B5 10
IL-11 B6 50
IL-27 B7 10
IL-33 B9 50
*BeadIDisusedtoassociateabeadpopulationtoaparticularanalyteintheLEGENDplexTMDataAnalysisSoftware.TheassociationofanalyteandbeadIDwillbedefinedduringthegatingstepofthedataanalysis.
WhenenteringanalyteandbeadIDinfomationduringthegatingstep,alwaysenterinthesequentialorderofthebeadID(e.g,A4,A5,A6...B2,B3,B4...). Please refer to the LEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelp for details (www.biolegend.com/legendplex).
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
6
StorageInformation
Recommended storage for all original kit components is between 2°C and 8°C. DONOTFREEZEPre-mixedBeads,DetectionAntibodiesorSA-PE.
• Oncethestandardshavebeenreconstituted,immediatelytransfercon-tents into polypropylene vials. DO NOT STORE RECONSTITUTED STAN-DARDS IN GLASS VIALS.
• Uponreconstitution,leftoverstandardandMatrixBshouldbestoredat≤-70°Cforusewithinonemonth.Avoidmultiple(>2)freeze-thawcycles.Discardanyleftoverdilutedstandards.
Materials Supplied
The LEGENDplexTM kit contains reagents for 100 tests, listed in the table below. Whenassayedinduplicate,thisisenoughforan8-pointstandardcurveand40samples.
Kit Components Quantity Volume Part #Setup Beads 1: FITC Beads 1 vial 1 mL 77840Setup Beads 2: PE Beads 1 vial 1 mL 77842Setup Beads 3: Raw Beads 1 vial 2 mL 77844Human Cytokine Panel 2 Premixed Beads 1bottle 3.5 mL 76101
Human Cytokine Panel 2 DetectionAntibodies 1bottle 3.5 mL 76115
Human Cytokine Panel 2 Standard Cocktail, Lyophilized 1 vial lyophilized 76859
LEGENDplexTM SA-PE 1bottle 3.5 mL 77743LEGENDplexTM Matrix B, Lyophilized 1 vial lyophilized 77549LEGENDplexTMAssayBuffer 1bottle 25 mL 77562LEGENDplexTMWashBuffer,20X 1bottle 25 mL 77564Filter plate 1 plate 76187Plate Sealers 4sheets 78101DataAnalysisSoftwareDongle 1 21217Human Cytokine Panel 2 Manual 1 76119
biolegend.com7
LEGENDplex™ Human Cytokine Panel 2
Materials to be Provided by the End-User
• Aflowcytometerequippedwithtwolasers(e.g.,a488nmbluelaseror532nmgreenlaseranda633-635nmredlaser)capableofdistinguishing575 nm and 660 nm oraflowcytometerequippedwithonelaser(e.g.,488nmbluelaser)capableofdistinguishing575nmand670nm.
Partiallistofcompatibleflowcytometers:
Flow Cytometer
Reporter Channel
ChannelEmission
ClassificationChannel
Channel Emission
Compensa-tionneeded?
BD FACSCaliburTM
(single laser)FL2 575 nm FL3 670 nm Yes
BD FACSCaliburTM
(dual laser)FL2 575 nm FL4 660 nm No*
BD FACSArrayTM Yellow 575 nm Red 660 nm No*
BD FACSCantoTM
BD FACSCantoTM IIPE 575 nm APC 660 nm No*
BDTM LSR, LSR IIBD LSRFortessaTM PE 575 nm APC 660 nm No*
BD FACSAriaTM PE 575 nm APC 660 nm No*
*Compensationisnotrequiredforthespecifiedflowcytometerswhenset up properly, but is recommended for consistent results.
Forsettingupoftheaboveflowcytometers,pleasefollowtheFlow Cytom-eter Setup guide in this manual or visit: www.biolegend.com/legendplex.
Forflowcytometersnotlistedhere,theend-userneedstosetupthemachine following similar guidelines. Please refer to Setup Procedure for Other Flow CytometerssectioninChapter4.
• Multichannelpipettescapableofdispensing5μLto200μL
• Reagentreservoirsformultichannelpipette
• Polypropylene microfuge tubes (1.5 mL)
• Laboratory vortex mixer
• Sonicator bath (e.g., Branson Ultrasonic Cleaner model #B200, or equiva-lent)
• Aluminum foil
• Absorbent pads or paper towels
• Plateshaker(e.g.,Lab-LineInstrumentsmodel#4625,orequivalent)
• Tabletopcentrifuges(e.g.,Eppendorfcentrifuge5415C,orequivalent)
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
8
Iftheassayisrunusingthefilterplateprovided(recommended),
• Avacuumfiltrationunit(MilliporeMultiScreen®HTSVacuumManifold,catalog#MSVMHTS00orequivalent).Instructionsonhowtousethevacuum manifold can be found at www.biolegend.com/legendplex.
• A vacuum source (mini vacuum pump or line vacuum, e.g., Millipore VacuumPump,catalog#WP6111560,orequivalent)
Iftheassayisruninmicrotubes,V-orU-bottom96-wellplate(optional),
• 1.1mLpolypropylenemicroFACStubes,in96-tuberack(e.g.,NationalScientificSupplyCo,catalog#TN0946-01R,orequivalent).
• Centrifugewithaswingingbucketadaptorformicrotiterplatesormicro-tube racks (e.g., Beckman Coulter AllegraTM 6R Centrifuge with MICROPLUS CARRIERadaptorforGH3.8andJS4.3Rotors).
• 96-well,V-bottom,clearpolypropylenemicroplate(lowproteinbinding,e.g., greiner bio-one, Catalog # 651201 or equivalent).
Precautions
• All blood components and biological materials should be handled as poten-tiallyhazardous.FollowuniversalprecautionsasestablishedbytheCenterforDiseaseControlandPreventionandbytheOccupationalSafetyandHealthAdministrationwhenhandlinganddisposingofinfectiousagents.
• Sodiumazidehasbeenaddedtosomereagentsasapreservative.Al-thoughtheconcentrationsarelow,sodiumazidemayreactwithleadandcopperplumbingtoformhighlyexplosivemetalazides.Ondisposal,flushwith a large volume of water to prevent azide build-up.
• Matrix B for LEGENDplexTM kits contains components of human origin and shouldbehandledaspotentiallyhazardous.TherawmaterialhasbeenscreenedforinfectiousdiseasesandisnegativeforHIV,HBVandHCVusingFDA-approved test methods.
• Donotmixorsubstitutereagentsfromdifferentkitsorlots.Reagentsfromdifferentmanufacturersshouldnotbeusedwiththiskit.
• Donotusethiskitbeyonditsexpirationdate.
• SA-PEandPremixedBeadsarelight-sensitive.Minimizelightexposure.
biolegend.com9
LEGENDplex™ Human Cytokine Panel 2Chapter 2: ASSAY PREPARATION
SampleCollectionandHandling
PreparationofSerumSamples:
• Allow the blood to clot for at least 30 minutes and centrifuge for 10 min-utes at 1,000 x g.
• Remove serum and assay immediately or aliquot and store samples at ≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
• Whenusingfrozensamples,itisrecommendedthatsamplesarethawedcompletely,mixedandcentrifugedtoremoveparticulatespriortouse.
PreparationofPlasmaSamples:
• PlasmacollectionusingEDTAasananti-coagulantisrecommended.Centri-fuge for 10 minutes at 1,000 x gwithin30minutesofbloodcollection.
• Remove plasma and assay immediately, or aliquot and store samples at ≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
• Whenusingfrozensamples,itisrecommendedthatsamplesarethawedcompletely,mixedwellandcentrifugedtoremoveparticulates.
PreparationofTissueCultureSupernatant:
• Centrifuge the sample to remove debris and assay immediately, or aliquot andstoresamplesat≤-20°C.Avoidmultiple(>2)freeze/thawcycles.
ReagentsPreparation
PreparationofAntibody-ImmobilizedBeads
SonicatePre-mixedBeadsbottlefor1minuteinasonicatorbathandthenvortex for 30 seconds prior to use. If no sonicator bath is available, increase thevortexingtimeto1minutetocompletelyresuspendthebeads.
PreparationofWashBuffer
• Bringthe20XWashBuffertoroomtemperatureandmixtobringallsaltsintosolution.
• Dilute25mLof20XWashBufferwith475mLdeionizedwater.Storeun-usedportionsbetween2°Cand8°Cforuptoonemonth.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
10
PreparationofMatrixB(forSerumorPlasmaSamplesOnly)
•Add 5.0 mL LEGENDplexTMAssayBuffertothebottlecontaininglyophilized MatrixB.Allowatleast15minutesforcompletereconstitution.Vortexto
mixwell.LeftoverreconstitutedMatrixBshouldbestoredat≤-70°Cforupto one month.
StandardPreparation
1. Priortouse,reconstitutethelyophilizedHumanCytokine Panel 2 Standard Cocktailwith250µLAssayBuffer.
2. Mix and allow the vial to sit at room temperature for 10 minutes, and then transfer the standard to an appropriately labeled polypropylene microfuge tube. This will be used as the top standard C7.
Note:Analytesinthispanelmayhavedifferenttopstandardconcentra-tions.(SeeTable1fordetails)
3. Label6polypropylenemicrofugetubesasC6,C5,C4,C3,C2andC1,re-spectively.
4.Add75µLofAssayBuffertoeachofthesixtubes.Prepare1:4dilutionofthe top standard by transferring 25 µL of the top standard C7 to the C6 tube and mix well. This will be the C6 standard.
5. Inthesamemanner,performserial1:4dilutionstoobtainC5,C4,C3,C2and C1 standards (see the table below).AssayBufferwillbeusedasthe0pg/mLstandard(C0).
Tube/Stan-dard ID
Serial Dilution
Assay Buffertoadd (µL)
Standard to add
Final Conc. (pg/mL)*
Final Conc. (pg/mL)**
C7 -- -- -- 10,000 50,000
C6 1:4 75 25 µL of C7 2,500 12,500
C5 1:16 75 25 µL of C6 625 3,125
C4 1:64 75 25 µL of C5 156.3 781.3
C3 1:256 75 25 µLofC4 39.1 195.3
C2 1:1024 75 25 µL of C3 9.8 48.8
C1 1:4096 75 25 µL of C2 2.4 12.2
C0 -- 75 -- 0 0
*TopstandardconcentrationofTSLP, IL-1α,IL-1β,GM-CSF,IFN-α2,IL-23,IL-12p70,IL-18andIL-27is10ng/mL;**TopstandardconcentrationofIL-12p40,IL-15, IL-11 and IL-33 is 50 ng/mL.
biolegend.com11
LEGENDplex™ Human Cytokine Panel 2
DilutionofSamples
• In general, serum or plasma samples need to be diluted 2-fold with Assay Bufferbeforetesting(e.g.dilute50µLofsamplewith50µLofAssayBuf-fer).
Iffurthersampledilutionisdesired,dilutionshouldbedonewithMatrixBto eusure accurate measurement.
Addingserumorplasmasampleswithoutdilutionwillresultinlowas-sayaccuracyandpossibly,cloggingofthefilterplate.
• For cell culture supernatant samples, the levels of analyte can vary greatly fromsampletosample.Whilethesamplecanbetestedwithoutdilutions,a preliminary experiment may be required to determine the appropriate dilutionfactorforsamples.
Ifsampledilutionisdesired,dilutionshouldbedonewithcorrespondingfreshcellculturemediumorAssayBuffertoensureaccuratemeasure-ment.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
12
Chapter 3: ASSAY PROCEDURE
The LEGENDplexTM assaycanbeperformedinafilterplate,inmicrotubes,orinaV-orU-bottommicroplate.
• Thein-filterplateassayprocedureisrecommendedduetoitsgoodsampleto sample consistency, assay robustness and ease of handling. This pro-cedurerequiresavacuumfiltrationunitforwashing(seeMaterials to be Provided by the End-User).IfyouhaveperformedaLuminex®-basedmultiplexassaybefore,yourlabshouldalreadyhavethevacuumfiltrationunit set up.
• Ifthein-filterplateassayprocedureisnotpossibleorifyouprefer,theas-saycanbeperformedinmicrotubesorinaU-orV-bottommicroplate.Forin-tube assay, we recommend using micro FACS tubes (see Materials to be Provided by the End-User).
Performing the Assay Using a Filter Plate
• Allow all reagents to warm to room temperature (20-25°C) before use.
• Setthefilterplateonaninvertedplatecoveratalltimesduringassaysetupandincubationsteps,sothatthebottomoftheplatedoesnottouchanysurface. Touching a surface may cause leakage.
• Keeptheplateuprightduringtheentireassayprocedure,includingthewashing steps, to avoid losing beads.
• The plate should be placed in the dark or wrapped with aluminum foil for allincubationsteps.
• Standards and samples should be run in duplicate and arranged on the plateinaverticalconfigurationconvenientfordataacquisitionandanalysis(asshowninattachedPLATEMAP).Besuretoloadstandardsinthefirsttwocolumns.Ifanautomationdeviceisusedforreading,theorientationandreadingsequenceshouldbecarefullyplanned.
1. Pre-wettheplatebyadding100μLofLEGENDplexTM1XWashBuffertoeach well and let it sit for 1 minute at room temperature. To remove the excess volume, place the plate on the vacuum manifold and apply vacuum. Donotexceed10”Hgofvacuum.Vacuumuntilwellsaredrained(5-10seconds).BlotexcessWashBufferfromthebottomoftheplatebypress-ing the plate on a stack of clean paper towels. Place the plate on top of the inverted plate cover.
For measuring cell culture supernatant samples:
• Add25µLofAssayBuffertoallwells.• Add 25 µL of each standard to the standard wells.
biolegend.com13
LEGENDplex™ Human Cytokine Panel 2• Add 25 µL of each sample to the sample wells (See Dilutionof
Samples)
For measuring serum or plasma samples:
• Add 25 µL of Matrix B to the standard wells.• Add25µLofAssayBuffertothesamplewells.• Add 25 µL of each standard to the standard wells. • Add 25 µL of each diluted serum or plasma sample to the sample
wells (See DilutionofSamples).
2. Vortexmixedbeadsfor30seconds.Add25μLofmixedbeadstoeachwell.Thevolumeshouldbe75μLineachwellafterbeadsaddition.(Note:Dur-ingadditionofthebeads,shakemixedbeadsbottleintermittentlytoavoidbeadsettling).
3. Seal the plate with a plate sealer. To avoid plate leaking, do not apply posi-tivepressuretothesealerwhensealingtheplate.Wraptheentireplate,including the inverted plate cover, with aluminum foil. Place the plate on a plate shaker, secure it and shake at approximate 500 rpm for 2 hours at room temperature.
4. Do not invert the plate! Place the plate on the vacuum manifold and apply vacuumasbeforeinStep1.Add200µLof1XWashBuffertoeachwell.RemoveWashBufferbyvacuumfiltration.BlotexcessWashBufferfromthebottomoftheplatewithanabsorbentpadorpapertowels. Repeat this washing step once more.
5. Add25µLofDetectionAntibodiestoeachwell.
6. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinverted plate cover, with aluminum foil. Place the plate on a plate shaker and shake at approximately 500 rpm for 1 hour at room temperature.
7. Do not vacuum! Add 25 µL of SA-PE to each well directly.
8. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinverted plate cover, with aluminum foil. Place the plate on a plate shaker and shake at approximate 500 rpm for 30 minutes at room temperature.
9. Repeatstep4above.
10. Add200µLof1XWashBuffertoeachwell(or 150 µL if plate is to be read withanautoampler.Highervolumemayresultinleakingofthefilterplateduringreading.Besuretosetthesamplevolumetobeanalyzedto70uLor less so that sample can be read again if needed). Resuspend the beads on a plate shaker for 1 minute.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
14
11. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay (Note: Prolonged sample storage can lead to reduced signal).
Iftheflowcytometerisequippedwithanautosampler,readtheplatedi-rectly using the autosampler. The probe height might need to be adjusted when using an autosampler.
If an autosampler is not available, the samples need to be transferred from thefilterplatetoFACStubesandreadmanually.Inthiscase,thesamplevolume may need to be increased from 200 µL to 300 µL by adding extra 100µLof1XWashBuffertoeachtubetoavoidsamplerunningdrywhenreadonaflowcytometer.
Assay Procedure Summary for Filter PlateAdd 100 μL 1X Wash Bu�er to �lter plate wells
Vacuum to remove excess bu�er
Incubate 2 hours, RT, shaking
Capture beads
Biotinylated Detection Antibody
Analytes
Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking
Wash 2 times using vacuum �ltration unitAdd 25 μL Detection AntibodiesIncubate 1 hr, RT, shaking
For cell culture supernatant samples, Add to the plate:25 μL Assay Bu�er to all wells25 μL diluted standard to standard wells25 μL sample to sample wells25 μL mixed beads to all wells
Wash 2 times using vacuum �ltration unitAdd 200 μL (or 150µL) of 1x Wash Bu�er Read on a �ow cytometer
For serum and plasma samples,Add to the plate:25 μL Matrix B to standard wells25 μL Assay Bu�er to sample wells25 μL diluted standard to standard wells25 μL sample to sample wells25 μL mixed beads to all wells
BA
C
A B C
A B C
biolegend.com15
LEGENDplex™ Human Cytokine Panel 2
PerformingtheAssayUsingMicrotubesora96-WellU-orV-bottomMicroplate
1. Allow all reagents to warm to room temperature (20-25°C) before use.
2. Determine the number of assays to run. Standards and samples should be runinduplicateandarrangedintheorderconvenientfordataacquisitionandanalysis(seebelow).Ifanautomationdeviceisusedforreading,theorientationandreadingsequenceshouldbecarefullyplanned.
If using tubes, label all tubes and arrange tubes on a rack (as shown in at-tached RACK MAP, assuming using micro FACS tubes and a 96-microtube rack,e.g.,NationalScientificSupplyCo,catalog#:TN0946-01R).A96-mi-crotuberackisrecommendedbecauseitallowspipettingwithamultichan-nelpipette.
If using a microplate, arrange the standard and samples according to the thePLATEMAPattached.
3. For measuring cell culture supernatant samples:
• Add25µLofAssayBuffertoalltubes/wells.• Add25µLofeachstandardtothestandardtubes/wells.• Add25µLofeachsampletothesampletubes/wells(See Dilutionof
Samples).• Add25µLofmixedbeadstoalltubes/wells.• Add25µLDetectionAntibodiestoalltubes/wells.
For measuring serum or plasma samples:
• Add25µLofMatrixBtothestandardtubes/wells.• Add25µLofAssayBuffertosampletubes/wells.• Add25µLofeachstandardtostandardtubes/wells.• Add25µLofeachdilutedserumorplasmasampletosampletubes/
wells (See DilutionofSamples).• Add25µLofmixedbeadstoalltubes/wells.• Add25µLDetectionAntibodiestoalltubes/wells.
Note: Thevolumeshouldbe100μLineachtube/well.Shake beads bottleintermittentlyduringtheadditiontoavoidbeadsettling.
4. Covertheentirerack/platewithaluminumfoiltoprotectthetubes/platefrom light. Shake (approximate 1,000 rpm for in-tube assay and 600 rpm for in-plate assay) on a plate shaker for 2 hours at room temperature.
5. Withoutwashingthetubes/plate,add25µLSA-PEtoeachtube/well.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
16
6. Covertheentirerack/platewithaluminumfoiltoprotectthetubes/platefrom light. Shake (approximate 1,000 rpm for in-tube assay and 600 rpm for in-plate assay) on a plate shaker for 30 minutes at room temperature.
7. Centrifugethetubes/plateat1,000xg for 5 minutes, using a swinging bucket rotor with microplate adaptor (Please refer to Materials to be Pro-vided by the End-User).
8. Removethesupernatantusingamultichannelpipette.Becarefulnottoremove beads, but to remove as much liquid as possible.
9. Add200µLof1XWashBuffertoalltubes/wells.Resuspendthebeadsbyvortexing (for in-tube assay) or shaking on a plate shaker for 1 minute (for in-plate assay). Centrifugethetubes/plateat1,000xg for 5 minutes, using a swinging bucket rotor with microplate adaptor. Remove the supernatant.
10. Add200µLof1XWashBuffertoeachtube/well(or 150 µL if plate is to be readwithanautoampler.Highervolumemayresultinleakingofthefilterplateduringreading.Besuretosetthesamplevolumetobeanalyzedto70 uL or less so that sample can be read again if needed). Resuspend the beadsbyvortexing(forin-tubeassay)orpippeting(forin-plateassay).
11. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay (Note: Prolonged sample storage can lead to reduced signal).
Iftheflowcytometerisequippedwithanautosampler,thesamplescanbereadeitherdirectly(forin-plateassay)orafterbeingtransferredtoa96-well plate ((for in-tube assay). The probe height might need to be adjusted when using an autosampler.
If an autosampler is not available, the samples can be read directly in FACS tubesorafterbeingtransferredfrommicroplatetomicroFACStubes.Inthis case, the sample volume may need to be increased from 200 µL to 300 µLbyaddingextra100µLof1XWashBuffertoeachtube,toavoidsamplerunningdrywhenanalyzedonaflowcytometer.
biolegend.com17
LEGENDplex™ Human Cytokine Panel 2
Assay Procedure Summary for Tubes (or V-Bottom Plate)
Arrange the number of tubes needed on a rack (or set up the plate)
Incubate 2 hours, RT, shaking
Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking
For cell culture supernatant samples, Add to the tubes/wells:25 μL Assay Bu�er to all tubes/wells25 μL diluted standard to standard tubes/wells25 μL sample to sample tubes/wells25 μL mixed beads to all tubes/wells25 μL Detection Antibodies to all tubes/wells
Spin down beadsWash one timeAdd 200 μL (or 150µL) of 1x Wash Bu�erRead on a �ow cytometer
For serum and plasma samples, Add to the tubes/well:25 μL Matrix B to standard tubes/wells25 μL Assay Bu�er to sample tubes/wells25 μL diluted standard to standard tubes/wells25 μL sample to sample tubes/wells25 μL mixed beads to all tubes/wells25 μL Detection Antibodies to all tubes/wells
A B
C
A
B
C
Capture beads
Detection Antibody
Analytes
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
18
Chapter 4: FLOW CYTOMETER SETUP
Inordertogeneratereliabledata,theflowcytometermustbesetupproperlybeforedataacquisition.Thefollowingsectionswilladdressmachinesetupfortheflowcytometerslistedbelow.
ListofFlowCytometersandPossibleConfigurations
Flow CytometerReporterChannel
Channel Emission
ClassificationChannel
ChannelEmission
Compensationneeded?
BD FACSCaliburTM
(single laser) FL2 575 nm FL3 670 nm Yes
BD FACSCaliburTM
(dual laser) FL2 575 nm FL4 660 nm No*
BD FACSCantoTM,BD FACSCantoTM II PE 575 nm APC 660 nm No*
BDTM LSR, LSRII,BD LSRFortessaTM PE 575-585
nm APC 660 nm No*
BD FACSAriaTM PE 575 nm APC 660 nm No*
*Compensationisnotrequiredforthespecifiedflowcytometerswhensetupproperly, but is recommended for consistent results.
Forflowcytometersnotlistedhere,theend-userneedstosetupthemachinefollowing similar guidelines. Please refer to Setup Procedure for Other Flow Cytometerssectioninthischapter.
The setup process typically includes the following steps. Please see the detailed setupprocedurethatfollows,regardingyourspecificinstrument.
1).Startuptheinstrumentfollowingthemanufacturer’srecommendations.
2).Createatemplatefordataacquisitionusingyourinstrument’sdata acquisitionsoftware.Atemplateisadocumentorworksheetwithdensityplotsthatallowstheusertoperformmachinesetupanddataacquisition.
3).SetupthePMTvoltagesofeachchanneltobeusedfordataacquisition using the Setup Beads provided in the kit.
4).DeterminewhethercompensationisneededbasedontheconfigurationofyoursystemasshowninTable1.Ifcompensationisneeded,perform compensationusingtheSetupBeadsprovidedinthekit.
biolegend.com19
LEGENDplex™ Human Cytokine Panel 2
Setup Procedure for FACSCaliburTM with Dual Lasers
For a dual laser FACSCaliburTM,useFL2forreporterandFL4forbeadsclassifica-tion.Ingeneral,thereisnoneedforcompensationbetweenthesechannelsifthe machine is set up properly, following the setup procedure described below.
1. Start up the Instrument
Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.
2. ObtainaTemplateforDataAcquisition
A template for FACSCaliburTM is a document with density plot that allows theusertoperformmachinesetupanddataacquisition.
Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplate and proceed to Step 3.
If a template is not yet available, create a new template by following the instructionsbelow:
2.1 FromtheBDCellQuestdataacquisitionsoftware,gotoFile→new document.
2.2 CreateadotplotwithFSC(forwardscatter)forX-axisandSSC(side scatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create twogatesandlabelthemBeadsAandBeadsB(Figure4).
Figure 4.
2.3 Create4dotplotsasshownbelow(Figure5)withFL2forX-axis,FL4 or FL1 for Y-axis. For these dot plots, gate on Beads A (dot plots on theleftpanelbelow)andBeadsB(dotplotsontherightpanelbelow). The dot plots should be in log mode.
02
00
40
06
00
80
01
00
0S
SC
-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
10
01
01
10
21
03
10
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
Page 1
Gated on Beads BGated on Beads A
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
20
2.4 Savethenewdocumentas“LEGENDplexTemplateforFACSCalibur Dual Laser” and proceed to the next step of the setup.
Figure 5.
3. Set up the PMT Voltages
The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannel(FL4/APC)andchannelFL1.TheSetupBeads2:PE
BeadsareusedtosetupthePMTvoltageofthereporterchannel(FL2/PE)The Setup Beads 1: FITC Beads are not needed for this setup.
FollowtheinstructionsbelowforsettingupthePMTsettings:
3.1 Vortex the vial of the Raw Beads for 30 seconds to resuspend the beads.
3.2 Transfer400μLoftheRawBeadstoafreshFACStube.
3.3 Settheflowcytometerflowratetolow.Insetupmode,runthe RawBeads.AdjustthesettingsforFSCandSSCsothatbothbead populationsarevisible(Figure6).
Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadpopulationsafteradjustingsettings.
biolegend.com21
LEGENDplex™ Human Cytokine Panel 2 Figure 6.
3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>200.
3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate
(Figure 6).
3.6 AdjusttheFL1settingsothattheFL1signalsforallbeadsare between 1x100 and 1x101 (Note: This step is not required, but is
recommended).
3.7 AdjusttheFL4settingsothattheFL4signalsforallbeadsarebe- tween 1x101 and 5x103 (Figure 7).
Figure 7.
3.8 Vortex the vial of Setup Beads 2: PE Beads for 30 seconds to resus- pend the beads.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
22
3.9 Transfer400μLofthePEBeadstoafreshFACStube.
3.10ReplacetheRawBeadstubefromtheflowcytometerwiththe PE beads tube.
3.11Settheflowcytometerflowratetolow.Insetupmode,runthePE Beads. Note: PE beads are only of small size, falling in the Beads A
gate (Figure 8).
3.12AdjusttheFL2settingsothatthemedianfluorescenceintensityof thePEbeadsfallsbetweenthelot-specificrangelabeledonthePE Beads vial (Figure 8, gate R3).
Figure 8.
3.13 Save the document again for future use.
3.14Theflowcytometerisnowreadyforsampleanalysis.
biolegend.com23
LEGENDplex™ Human Cytokine Panel 2
Setup Procedure for FACSCaliburTM with a Single Laser
For a single laser FACSCaliburTM,useFL2forreporterandFL3forbeadsclassifi-cation.Compensationisneededtoproperlysetuptheinstrument.
1. Start up the Instrument
Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.
2. ObtainaTemplateforDataAcquisition
A template for FACSCaliburTM is a document with density plot that allows theusertoperformmachinesetupanddataacquisition.
Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplate and proceed to Step 3.
If a template is not yet available, create a new template by following the instructionsbelow:
2.1 FromtheBDCellQuestdataacquisitionsoftware,gotoFile→new document.
2.2 CreateadotplotwithFSC(forwardscatter)forX-axisandSSC(side scatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create two gates and label them Beads A and Beads B (Figure 9).
Figure9.
0200
400
600
800
1000
SS
C-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
Page 1
Gated on Beads BGated on Beads A2.3 Create4dotplotsasshownbelow(Figure10)withFL2forX-axis,and FL3 or FL1 for Y-axis. For these dot plots, gate on Beads A (dot plots on theleftpanelbelow)andBeadsB(dotplotsontherightpanelbelow). The dot plots should be in log mode.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
24
Figure 10.
0200
400
600
800
1000
SS
C-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
010
110
210
310
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
Page 1
Gated on Beads BGated on Beads A
2.4 Setallcompensationstozero.
2.5 Savethenewdocumentas“LEGENDplexTemplateforFACSCalibur Single Laser” and proceed to the next step of the setup.
biolegend.com25
LEGENDplex™ Human Cytokine Panel 23. SetupPMTVoltagesandCompensation
The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannelFL3andchannelFL1.TheSetupBeads2:PEBeadsare used to set up the PMT voltage of the reporter channel FL2. All three setupbeadsareneededforsettingupcompensation.
FollowtheinstructionsbelowforsettingupthePMTsettingsandcompen-sation:
3.1 Vortex the vial of the Raw Beads for 30 seconds to resuspend the beads.
3.2 Transfer400μLoftheRawBeadstoafreshFACStube.
3.3 Settheflowcytometerflowratetolow.Insetupmode,runtheRaw Beads.AdjustthesettingsforFSCandSSCsothatbothbeads populationsarevisible(Figure11).
Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadpopulations,afteradjustingsettings.
Figure 11.
3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>200.
3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate (Figure 11).
3.6 AdjusttheFL1settingsothattheFL1signalsforallbeadsarebe tween 1x100 and 1x101 (Figure 12).
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
26
Figure 12.
3.7 AdjusttheFL2settingssothattheFL2signalsforallbeadsarebe- tween 1x100 and 1x101 (Figure 12).
3.8 AdjusttheFL3settingsothattheFL3signalsforallbeadsarebe- tween 1x101 and 5 x 103 (Figure 12).
3.9 Vortex the vial of the Setup Beads 1: FITC Beads for 30 seconds to resuspend the beads.
3.10Transfer200μLoftheFITCBeadstoafreshFACStube.Add200μL ofRawBeadsandmixwell(ThiswillgenerateFITC-positiveand FITC-negativepopulationsofbeadsandisneededforproper compensation).
3.11 In setup mode, run the mixed FITC and Raw Beads.
3.12 On the FL1 vs FL2 dot plot (Figure 13), the beads will display as FITC- negativeandFITC-positivepopulations(indicatedbyanarrow).
10
01
01
10
21
03
10
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L3-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
10
01
01
10
21
03
10
4F
L1-H
100 101 102 103 104
FL2-H
Acquisition Dot Plot
02
00
40
06
00
80
01
00
0S
SC
-H
00 200 400 600 800 1000FSC-H
Acquisition Dot Plot
Beads A
Beads B
Page 1
Gated on Beads BGated on Beads A
biolegend.com27
LEGENDplex™ Human Cytokine Panel 2Note: FITC beads are only of small size, falling in the Beads A gate.
Figure 13.
3.13AdjusttheFL2-%FL1compensationsetting(e.g.,FL2-FL1=20%)so thattheFITC-negativeandFITC-positivepopulationshavesimilar meanFL2fluorescenceintensities(Figure14).
Figure 14.
3.14VortexthevialofPEBeadsfor30secondstoresuspendthebeads.
3.15Transfer200μLofthePEBeadstoafreshFACStube.Add200μLof RawBeadsandmixwell(ThiswillgeneratePE-positiveand PE-negativepopulationsofbeadsandisneededforpropercompen- sation).
3.16 In setup mode, run the mixed PE and Raw Beads.
3.17 On the FL1 vs FL2 dot plot (Figure 15), the beads will display as PE- negativeandPE-positivepopulations(indicatedbyanarrow).Note: PE beads are only of small size, falling in the Beads A gate.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
28
Figure 15.
3.18AdjusttheFL2settingsothatthemedianfluorescenceintensityofthe PEbeadsfallsbetweenthelot-specificrangelabeledonthePEBeads vial (Figure 16).
3.19AdjusttheFL1-%FL2compensationsetting(e.g.,FL1-FL2=1.5%)so thatthePE-negativeandPE-positivepopulationshavesimilarmean FL1fluorescenceintensities(Figure16).
Figure 16.
3.20 On the FL3 vs FL2 dot plot (Figure 17), the beads will display as PE- negativeandPE-positivepopulations(indicatedbyanarrow).
3.21 AdjusttheFL3-%FL2compensationsetting(e.g.,FL3-%FL2=40%)so thattheFL3-lowpopulationandthePE-positivepopulationhave similar median FL3 signal (Figure 18).
biolegend.com29
LEGENDplex™ Human Cytokine Panel 2 Figure 17.
Figure 18.
3.22 Save the document again for future use.
3.23Theflowcytometerisnowcompensatedandreadyforsample analysis.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
30
Setup Procedure for BD FACSAriaTM, FACSCantoTM and LSR Series
ThispartoftheguideappliestoBDdigitalflowcytometersusingFACSDivaTM softwareversion6.0andabove.
For the BD FACS machines running FACSDivaTM, use the PE channel for reporter andtheAPCchannelforbeadsclassification.Ingeneral,thereisnoneedforcompensationbetweenthesechannelsifthemachineissetupproperlyfollow-ing the setup procedure described below.
Thissetupprocedureisrequiredunderthefollowingsituations:
• YouarerunningtheLEGENDplexkitforthefirsttime.
• It has been over a month since the procedure was last performed.
• Yourflowcytometerhasbeenservicedsinceyoulastperformedthis procedure.
This setup process is not needed if you have run this experiment before and haveaccesstoasavedexperimenttemplate(Thesettingswillbesavedinthefinalstepofthissetupprocedureandanysettingssavedcanbeimportedtoanew experiment. Please refer to Step 2 and Step 3.9 below).
1. Start up the Instrument
Performinstrumentstartupandverificationcheckfollowingthemanufac-ture’srecommendations.
2. ObtainaTemplateforDataAcquisition
A template for FACSDivaTM is a worksheet with density plots that allows the usertoperformmachinesetupanddataacquisition.
If a template is not yet available, create a new template by following the instructionsinstep2.Afteratemplateiscreated,savethefileinD:\BDEx-port\Templates\Experiment.DonotchangethenameoftheTemplatesfolder.
Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplateandproceedtoStep3.Toopenanexistingtemplate,selectEx-periment>NewExperiment.AlistoftemplatessavedinD:\BDExport\Tem-plates\Experimentwillpopup.Selectthedesiredtemplatefromthelist.
Tocreateanewtemplate,followtheinstructionsbelow:
2.1 From the BD FACSDivaTMsoftware,gotoExperiment>NewExperi- ment.
biolegend.com31
LEGENDplex™ Human Cytokine Panel 2 2.2 In the global worksheet, open the worksheet. Create a dot plot with
FSC(forwardscatter)forX-axisandSSC(sidescatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create two gates and label them Beads A and Beads B (Figure 19).
Figure19.
2.3 CreatetwodotplotswithPEforX-axis,APCforY-axis(Figure20), gatedonBeadsA(leftpanelbelow)andBeadsB(rightpanelbelow, respectively.CreateonedotplotwithFITCforX-axis,APCforY-axis, gated on Beads A and Beads B (graph not shown). The plots should all be in log mode.
Figure 20.
2.4 Savethedocumentas“LEGENDplexTemplateforFACSDiva” inD:\BDExport\Templates\Experimentandproceedtothenextstep of setup.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
32
3. Set up PMT Voltages
The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannelAPC,reporterchannelPE,andFITCchannel.TheSetup Beads 1: FITC Beads and 2: PE Beads are not needed for this setup becausenocompensationisrequiredifthesetupproceduredescribedhereis closely followed.
FollowtheinstructionsbelowforsettingupthePMTsettings:
3.1 Vortex the vial of Raw Beads for 30 seconds to resuspend the beads.
3.2 Transfer400μLoftheRawBeadstoafreshFACStube.
3.3 Settheflowcytometerflowratetolow.RuntheRawBeads.Adjust thesettingsforFSCandSSCsothatbothbeadpopulationsarevisible (Figure 21).
Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadspopulationsafteradjustingsettings.
Figure 21.
3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>50(x1000).
3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate (Figure 21).
3.6 AdjusttheFITCsettingsothattheFITCsignalforthemajorityof beads is between 1x101 and 1x102 (Figure 22).
biolegend.com33
LEGENDplex™ Human Cytokine Panel 2 Figure 22.
Beads Beads A+ Beads B
3.7 AdjustthePEsettingsothatthePEsignalforthemajorityofbeadsis between 1x101 and 1x102 (Figure 23).
Figure 23.
3.8 AdjusttheAPCsettingssothatthetheAPCfluorescenceintensitiesof allbeadpopulationsarebetween1x102 and 5 x 104 (Figure 23).
3.9 Save the document again for future use.
Tosaveyourassay-specificsettings,inthebrowser,right-click CytometerSettingsandselectSavetoCatalog.Namethefile,and thenclickOK.Toimportthesavedsettingforanewexperiment,right clickoncytometersettingsandselectimportsettings.
3.10Theflowcytometerisnowreadyforsampleanalysis.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
34
Setup Procedure for Other Flow Cytometers
Forflowcytometersnotaddressedabove,thesetupprocedurewilldifferfromone to another. It is very important for the end-user to set up the machine following similar guidelines. Inthiscase,machinecompensationbetweenthereporterandbeadsclassificationchannelsisstronglyrecommended.
Theflowcytometersetupinstructionswillalsobepostedonourwebsite:www.biolegend.com/legendplex. Newflowcytometersetupinstructionswillbeadded to on our website when they become available.
biolegend.com35
LEGENDplex™ Human Cytokine Panel 2
Chapter 5: DATA ACQUISITION AND ANALYSIS
DataAcquisition
1. Beforereadingsamples,makesurethattheflowcytometerissetupprop-erly.Forflowcytometersetup,pleasefollowtheFlowCytometerSetupguide in this manual or visit: www.biolegend.com/legendplex.
2. Createanewtemplateoropenanexistingtemplate(fordetailsonhowtocreateacytometer-specifictemplate,pleaserefertotheFlowCytometerSetup Guide).
3. Vortex each sample for 5 seconds before analysis.
4.Settheflowratetolow.Setthenumberofbeadstobeacquiredtoabout300peranalyte(e.g.,acquire4,000beadsfora13-plexpanel).Donotacquiretoomanyevents(e.g.>10,000totalevents).
Note: Do not acquire too few or too many beads. Too few beads acquired may result in high CVs and too many beads acquired may result in slow data analysis later.
5. Read samples.
Whenreadingsamples,settheflowcytometertosetupmodefirstandwaituntilbeadpopulationisstabilizedbeforeswitchingtoacquisitionmode.
To simplify data analysis using the LEGENDplexTMDataAnalysisSoftware,read samples in the same order as shown on the PLATE MAP or RACK MAP attachedattheendofthemanual.Foranin-plateassay,readcolumnbycolumn (A1, B1, C1...A2, B2, C2...). For an in-tube assay, read row by row (A1, A2, A3,...B1, B2, B3...).
Whennamingdatafiles,trytousesimplenameswithaconsecutivenum-bering for easy data analysis (e.g. for standards, C0.001, C0.002, C1.003, C1.004,C2.005,C2.006,C3.007,C3.008,...C7.015,C7.016;forsamples,S1.017,S1.018,S2.019,S2.020,S3.021,S3.022…)
StoreallFCSfilesinthesamefolderforeachassay.Ifrunningmultipleas-says, create a separate folder for each assay.
6. Proceed to data analysis using LEGENDplexTMDataAnalysisSoftwarewhendataacquisitioniscompleted.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
36
Data Analysis
• TheFCSfilegeneratedonaflowcytometershouldbeanalyzedusingBioLegend’s LEGENDplexTMDataAnalysisSoftwareorothercompatibledataanalysissoftware.TheLEGENDplexTMDataAnalysisSoftwarecanbedownloaded here: www.biolegend.com/legendplex.
• Afterdownloading,installitonaPC(runningWindows7orWindows8)anduseitinconjunctionwiththeDataAnalysisSoftwareDongleincludedin this kit. The dongle has a license key stored in it and is needed to run the software.Tousethedongle,simplyplugitintheUSBportofthecom-puteronwhichthedataanalysissoftwareisinstalled,priortolaunchthesoftware.
• TheSoftwareDonglehasafixednumberofpoints.Eachanalysiswillcon-sume a certain number of points from the dongle. The number of points consumeddependsonassayplexsizeandnumberofsamples.Afterthepoints on a dongle are consumed, a new dongle will be needed to run moredataanalysis.Althoughthesoftwarewillcontinuetobefunctional,thedatawillnotbesaveduntilanewdongleisavailable.Anewdongleisprovided in each LEGENDplexTM kit. A saved analysis can always be re-ana-lyzedusingthesoftwareregardlessofdonglestatus.
• Follow the LEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelptousethesoftware(www.bioLegend.com/legendplex; or press F1 for online help at any step of the data analysis).
biolegend.com37
LEGENDplex™ Human Cytokine Panel 2
Chapter 6: FREQUENTLY ASKED QUESTIONS
Q. WhatisthedifferencebetweenLEGENDplexTM and Luminex®Assays?
BioLegend’s LEGENDplexTM assays and Luminex®-basedmultiplexassaysare both bead-based immunoassays using the same basic principle of sandwichimmunoassays.Bothsystemsusefluorescence-codedbeadstoachievemultiplexing.Themajordifferenceishowthedataisacquired.LEGENDplexTMassaysusecommonlabflowcytometersandtheirrespec-tivesoftwarefordataacquisition,whereasLuminex®-based assays use dedicatedmachinesandsoftwarefordataacquisition.Therefore,oneofthe advantages of LEGENDplexTM assays is that they can be run on common flowcytometersandnospecializedmachineisneeded.
Q. Can I use LEGENDplexTM kits on Luminex®machines?
No. LEGENDplexTMkitsshouldbeusedonaregularflowcytometerandthedatacanbeanalyzedusingtheuser-friendlydataanalysissoftwareavail-ablefordownloadatwww.biolegend.com/legendplex.
Q. WhatarethecompatibleflowcytometersfortheLEGENDplex™assays?
In general, LEGENDplexTMassayscanbeusedonmostcommonflowcytom-eters, such as:
BD FACSCalibur™ BD FACSCanto™ BD FACSCanto™ II BD TM LSR I BD TM LSR II BD LSRFortessaTM
BD FACSAria™ BD FACSArrayTM
PleaserefertotheMATERIALSTOBEPROVIDEDBYEND-USERsectionfordetailsonthechannelconfigurationsofthecytometer.
Forflowcytometersnotlistedabove,theenduserneedstomakesure
that the machine is set up properly before use (refer to Setup Procedure for Other Flow CytometerssectioninChapter4).In this case, machine compensationbetweenthereporterandbeadsclassificationchannelsisstrongly recommended.
TheFCSorListmodefilesfromthefollowinginstrumentsarecompatiblewith the LEGENDplex TMDataAnalysisSoftware.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
38
Vendor Instrument Model Vendor Instrument
Model
BD
FACScan™ Sony iCyt Eclipse™
FACSCalibur™ Life Technologies Attune®
FACSCanto™ II Miltenyi Biotec MACSQuant®
LSRFortessa™ Partec PAS
LSR II Stratedigm S1400
FACSAria™
Beckman Coulter
CyAn™ ADP
FACSAria™ II FC 500
Accuri™ C6 Gallios™
ORFLO Moxi Flow™ MoFlo®Astrios™
CyteK DxP10™ MoFlo®XDP
Q. Whatistherightprocedureforrunningtheassay?
1. Perform the assay as instructed in this kit manual.
2. Determinethetypeofflowcytometeryouhaveandperformmachinesetup as instructed on Flow Cytometer Setup guide in this manual or visit: www.biolegend.com/legendplex.Openanexistingorcreateamachine-specifictemplatefordataacquisition.
3. AnalyzethesamplesontheflowcytometerandsavetheFCSfilesina new folder.
4. Install the LEGENDplexTM DataAnalysisSoftwarealongwiththesoft-waredongleonaPCwithoperatingsystemWindows7orWindows8(32bitor64bit).Makesuretoinstallthecorrectversion(32bitor64bitversion),dependingonyourcomputer’soperatingsystem.
5. TransferthefoldercontainingtheFCSfilesofyourexperimenttothecomputerwherethedataanalysissoftwareisinstalled.
6. PerformdataanalysisasinstructedintheDataAnalysisSoftwareUserGuide.
Q. WhendoIneedtodomachinecompensation?
Ifaflowcytometerequippedwithasinglelaserisusedforbothreporter(FL2/PE)andclassification(FL3),thencompensationisabsolutelyrequiredto compensate the signal spill over between the two channels, especially from FL2 to FL3. Please use the Setup Beads provided and follow the FLOW CYTOMETER SETUP guide in this manual.
biolegend.com39
LEGENDplex™ Human Cytokine Panel 2 Ifaflowcytometerislistedinthecompatibleflowcytometerlistinthis
manualandisequippedwithtwolasers,oneforreporter(FL2/PE)andtheotherforclassification(FL4/APC),thencompensationisnotrequirediftheFLOW CYTOMETER SETUP guide in this manual is closely followed. How-ever,compensationisrecommendedforconsistentresults.
Forotherflowcytometersnotlistedinthecompatibleflowcytometerlist, the end-user needs to set up the machine following similar guidelines (refer to manufacturer’s manual for proper instrument setup). In this case, machinecompensationbetweenthereporterandbeadsclassificationchannels is strongly recommended.
Q. Isthereaspecialsoftwarerequiredfordataanalysis?
Theflowcytometerrawdatafiles(FCS2.0,3.0)canbeanalyzedusingtheLEGENDplexTMDataAnalysisSoftwareorotherequivalentanalysissoft-ware. Download the LEGENDplexTM DataAnalysisSoftwareforfreehere:www.biolegend.com/legendplex.Asoftwaredonglewithlicensekeyisprovided in the kit.
Fordataacquisition,nospecialsoftwareisneeded.Justusethedataac-quisitionsoftwarethatcomeswiththeflowcytometer,aslongasthedatageneratedisinFCSformat,whichmeetsFCSconvention2.0,3.0and3.1.
Q. CanIselecttomeasureonlysomeanalyteswithinapanel?
Yes. The targets within a panel are fully customizable. The customer can selectanycombinationoftargetswithinapanelandorderacustomizedproduct.Pleaseuseourwebsitetargetselectiontoolfororderingcustom-ized product (www.biolegend.com/legendplex).
Q. CanIselecttomeasureanalytesacrosspanels?
Cross-panelcustomizationisalsopossible,aslongasthetotalplexsizeisno more than 13 and there is no overlapping beads region among targets selected.Forcross-panelcustomization,[email protected].
Q. I have run out of a component from the kit. Can I use a similar compo-nentfromadifferentkit?
TheLEGENDplex™Beads,Standards,DetectionAntibodies,MatrixB,andSA-PEarelot-specificandmustbeusedincombinationwitheachother.Donotmixthesecomponentsfromdifferentkitsorlots.
Othercomponents,suchasAssayBuffer,WashBuffer,Plate,andPlateSealerarenotlot-specificandcanthereforebeexchangedbetweendiffer-ent kits and lots.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
40
Q. My standard curve is not linear; can I use the non-linear part of the stan-dardcurveduringanalysis?
Yes. It is possible to use the non-linear part of the standard curve for calculatingtheresults.TheLEGENDplexTMDataAnalysisSoftwareusesafive-parametercurvefittingalgorithm,whichdeterminestheminimumandmaximumdetectionconcentrationsofeachtargetandreportsthem.Ifsampleconcentrationsfallabovethemaximumdetectableconcentration,thesamplewillhavetobedilutedandreanalyzed.Ifsampleconcentrationsfallbelowtheminimumdetectionconcentration,itisconsiderednon-detectablebytheparticularassay.
Q. Duringdataacquisition,whydothebeadpopulations(definedbyFSCandSSC)sometimesappeartobedispersedorshifted?
Thisisusuallycausedbyafastflowrateorasuddenchangeinflowrate.Therearethreeflow-throughsettings:low,medium,andhigh.Ifyouareusingalowflowrateandthenchangetomediumandhighflowrate,thesuddenchangeofflowratemaysometimesresultinadispersedorshiftedbeadpopulation.Therefore,itisnotrecommendedtochangeflowrateduringdataacquisition.Thebestwayistorunthesampleinsetupmodeusinganidealflowrate,andoncethepopulationisstable,thenchangeittoacquisitionmode.
Q. DoestheLEGENDplex™DataAnalysisSoftwarerunonAppleMacintosh?
Notyet.ThecurrentversionofthesoftwarehastorunonaPCwithoperatingsystemWindows7orWindows8(32bitor64bit).IfyourflowcytometerisconnectedtoaMac,afterdataacquisition,theentirefoldercontainingthedata(FCSfiles)shouldbetransferredtoaPCandanalyzed.
Q. Whydoeseachkitincludeasoftwaredongle?
Thedongleallowsyoutousethedataanalysissoftware.However,thereisalimitednumberofpointswitheachdongle.Whenthetotalnumberofpoints is consumed for a dongle, the dongle needs to be replaced with a new one. Each kit includes a dongle to ensure the data analysis will not be impactedbythelimitedusageofadongle.Eachdongleissufficientforatleast 192 tests or two full 13-plex assay panels, equivalent to 2600 points
Q. Iranoutofpointsonmydongle.CanIgetanewoneseparately?Or,canIuseadonglefromonekitforanotherkit?
Dongle is not sold separately. Contact BioLegend if you need a new dongle. All dongles are the same. The dongle can be used across kits.
biolegend.com41
LEGENDplex™ Human Cytokine Panel 2
Chapter 7: ASSAY CHARACTERIZATION
RepresentativeStandardCurve
This standard curve was generated using the LEGENDplexTM Human Cyto-kinePanel2fordemonstrationpurposeonly.Astandardcurvemustberun with each assay.
1.0
10.0
100.0
1000.0
10000.0
1 10 100 1000 10000 100000
MFI
Concentration (pg/mL)
TSLPIL-1α IL-1β GM-CSFIFN-α2 IL-23IL-12p40IL-12p70IL-15IL-18IL-11IL-27IL-33
AssaySensitivity
Theassaysensitivityorminimumdetectableconcentration(MDC)isthetheoreticallimitofdetectioncalculatedusingtheLEGENDplexTM Data AnalysisSoftwarebyapplyinga5-parametercurvefittingalgorithm.
Analyte MDC in Cell Culture Medium (pg/mL)
MDC in Serum (pg/mL)
Human TSLP 1.2 1.1
HumanIL-1α 1.1 1.5
HumanIL-1β 0.8 0.8
Human GM-CSF 1.1 0.6
HumanIFN-α2 1.1 1.0
Human IL-23 0.7 1.4
HumanIL-12p40 6.0 5.1
Human IL-12p70 0.9 0.9
Human IL-15 7.4 5.7
Human IL-18 0.7 0.7
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
42
Human IL-11 8.0 4.6
Human IL-27 1.3 1.0
Human IL-33 7.2 5.0
Cross-Reactivity
Thefollowinghumanrecombinantproteinsweretestedat50ng/mLusingthe LEGENDplexTM Human Cytokine Panel 2. IL-23 showed 5.0% cross-reactivitywithIL-12p40assay,IL-27showed0.3%cross-reactivitywithIL-11assay.Noornegligiblecross-reactivitywasfoundamongallotheranalytes.
IL-2 IL-4 IL-5 IL-6 IL-9 IL-10 IL-13
IL-17A IL-22 IFN-γ TNF-α IL-17F IL-21 IL-8
IP-10 CCL11 TARC MCP-1 RANTES MIP-1α MIG
ENA-78 MIP-3α GROα I-TAC MIP-1β IFN-α2 TSLP
IL-1α IL-1β IL-3 IL-7 IL-11 IL-12p40 IL-12p70
IL-15 IL-18 IL-23 IL-27 IL-33 GM-CSF
Accuracy (Spike Recovery)
For spike recovery in cell culture medium, target proteins with known concentrationswerespikedintocellculturemedium(RPMIandDMEMwith10%FCS)atthreedifferentlevelswithintheassayrange.Thespikedsampleswerethenassayed,andthemeasuredconcentrationswerecom-pared with the expected values.
Forspikerecoveryinserum,asamplewithknownhighconcentrationsoftargetproteinswasspikedintounknownserumsamples(n=10).Thespikedsampleswerethenassayed,andthemeasuredconcentrationswerecom-pared with the expected values.
Analyte % of Recovery in Cell Culture Medium
% of Recovery in Serum
Human TSLP 86% 153%HumanIL-1α 103% 132%HumanIL-1β 104% 88%Human GM-CSF 101% 119%HumanIFN-α2 106% 140%
biolegend.com43
LEGENDplex™ Human Cytokine Panel 2
Human IL-23 100% 103%HumanIL-12p40 117% 163%Human IL-12p70 93% 90%Human IL-15 78% 80%Human IL-18 104% 97%Human IL-11 109% 117%Human IL-27 108% 83%Human IL-33 100% 79%
LinearityofDilution
Fortestinglinearityofdilution,serumsamples(n=6)werefirstdilutedtwo-foldwithAssayBuffer,thenseriallydiluted1:2,1:4,1:8withMatrixBandassayed.Themeasuredconcentrationsofseriallydilutedsampleswerethen compared with that of the two-fold diluted samples.
Analyte Linearity of Dilution Analyte Linearity of
DilutionHuman TSLP 100% Human IL-12p70 111%HumanIL-1α 86% Human IL-15 110%HumanIL-1β 105% Human IL-18 110%Human GM-CSF 92% Human IL-11 104%HumanIFN-α2 64% Human IL-27 125%Human IL-23 99% Human IL-33 116%HumanIL-12p40 58%
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
44
Intra-Assay Precision
Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedin one assay with 16 replicates for each sample. The intra-assay precision was calculated as below.
Analyte Sample Mean (pg/mL) STDEV %CV
Human TSLPSample 1 627.0 42.3 7%Sample 2 39.7 2.6 6%
HumanIL-1αSample 1 667.4 56.9 9%Sample 2 37.1 2.1 6%
HumanIL-1βSample 1 630.8 28.9 5%Sample 2 37.5 3.2 8%
Human GM-CSFSample 1 679.7 57.0 8%Sample 2 38.6 2.3 6%
HumanIFN-α2Sample 1 646.4 46.2 7%Sample 2 37.4 2.3 6%
Human IL-23Sample 1 634.3 53.7 8%Sample 2 42.3 2.1 5%
Human IL-12p40
Sample 1 626.4 48.4 8%Sample 2 40.0 3.1 8%
Human IL-12p70
Sample 1 703.5 58.0 8%Sample 2 42.9 3.9 9%
Human IL-15Sample 1 623.8 26.1 4%Sample 2 37.5 2.0 5%
Human IL-18Sample 1 662.4 45.3 7%Sample 2 41.8 3.3 8%
Human IL-11Sample 1 622.2 34.5 6%Sample 2 41.6 3.8 9%
Human IL-27Sample 1 639.9 40.1 6%Sample 2 42.8 3.8 9%
Human IL-33Sample 1 674.1 58.8 9%Sample 2 38.0 2.9 8%
biolegend.com45
LEGENDplex™ Human Cytokine Panel 2
Inter-Assay Precision
Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedin three independent assays with 3 replicates for each sample. The inter-assay precision was calculated as below.
Analyte Sample Mean (pg/mL) STDEV %CV
Human TSLPSample 1 633.6 42.4 7%Sample 2 39.6 3.0 8%
HumanIL-1αSample 1 640.2 55.3 9%Sample 2 37.1 3.3 9%
HumanIL-1βSample 1 619.7 44.6 7%Sample 2 37.4 3.7 10%
Human GM-CSFSample 1 697.9 53.1 8%Sample 2 38.1 2.8 7%
HumanIFN-α2Sample 1 655.2 45.3 7%Sample 2 36.7 2.6 7%
Human IL-23Sample 1 632.3 45.2 7%Sample 2 40.8 3.0 7%
Human IL-12p40
Sample 1 595.1 31.7 5%Sample 2 39.2 2.4 6%
Human IL-12p70
Sample 1 722.7 54.9 8%Sample 2 42.9 4.1 10%
Human IL-15Sample 1 619.1 40.1 6%Sample 2 36.0 2.5 7%
Human IL-18Sample 1 673.4 52.2 8%Sample 2 41.8 3.4 8%
Human IL-11Sample 1 662.0 52.2 8%Sample 2 40.6 3.8 9%
Human IL-27Sample 1 646.6 42.6 7%Sample 2 42.8 4.0 9%
Human IL-33Sample 1 633.4 49.7 8%Sample 2 37.8 3.7 10%
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
46
Biological Samples
Serum and Plasma (Samples are paired.)
Normalhumanserumsamples(n=20)weretestedforendogenouslevelsofthecytokines.Theconcentrationsmeasuredareshownbelow:
Analyte Range (pg/ml)
No. of Detectable
% of Detectable
Mean* (pg/mL)
Human TSLP ND-152.1 10 50% 25.7
HumanIL-1α ND-251.8 9 45% 40.0
HumanIL-1β ND-91.2 3 15% 39.3
Human GM-CSF ND-173.6 2 10% 93.4
HumanIFN-α2 ND-352.6 7 35% 103.9
Human IL-23 ND-204.4 4 20% 112.3
HumanIL-12p40 ND-801.8 15 75% 121.8
Human IL-12p70 ND-41.6 3 15% 20.5
Human IL-15 ND-196.8 11 55% 25.7
Human IL-18 33-406.3 20 100% 133.3
Human IL-11 ND-127.1 5 25% 44.4
Human IL-27 ND-541.1 4 20% 51.0
Human IL-33 ND-536.4 4 20% 161.8
ND=Non-detectable*The mean is calculated based on detectable samples only.
Normalhumanplasmasamples(n=20)weretestedforendogenouslevelsofcytokines.Theconcentrationsmeasuredareshownbelow:
Analyte Range (pg/mL)
No. of Detectable
% of Detectable
Mean* (pg/mL)
Human TSLP ND-104.9 10 50% 14.1
HumanIL-1α ND-171.9 6 30% 39.3
HumanIL-1β ND-64.2 4 20% 19.0
Human GM-CSF ND-113.7 2 10% 58.6
HumanIFN-α2 ND-256.8 9 45% 49.4
Human IL-23 ND-201.5 3 15% 79.1
HumanIL-12p40 ND-475.5 17 85% 63.2
Human IL-12p70 ND-26.0 5 25% 7.6
biolegend.com47
LEGENDplex™ Human Cytokine Panel 2
Human IL-15 ND-140.4 14 70% 20.7
Human IL-18 29.9-409.9 20 100% 142.7
Human IL-11 ND-58.3 4 20% 21.4
Human IL-27 ND-323.8 8 40% 47.5
Human IL-33 ND-383.7 1 5% 383.7
ND=Non-detectable*The mean is calculated based on detectable samples only.
Cell Culture Supernatant
Human PBMC (1 x 106cells/mL)wereculturedundervariousconditions(LPS,1µg/mL;PolyIC,50µg/mL;CPG,5µg/mL;R488,2µg/mL;IFN-γ,100ng/mL).Supernatantswerecollectedafter42hoursandassayedwiththe LEGENDplexTMHumanCytokinePanel2kit.Theresults(allinpg/mL)are summarized below.
Analyte Control LPS Poly IC CPG R488 LPS+IFN-γ*
Human TSLP ND ND ND ND ND 2.7
HumanIL-1α 1.5 480.6 1.6 1.5 159.1 2825.9
HumanIL-1β 0.8 1515.4 9.5 0.8 859.2 4571.7
Human GM-CSF
1.3 20.9 4.0 1.3 5.5 23.1
HumanIFN-α2 ND ND 51.7 ND 334.2 12.8
Human IL-23 ND 5.3 4.2 ND 5.7 2410.4
Human IL-12p40
ND 44.0 16.5 1.5 196.6 7889.8
Human IL-12p70
ND ND 43.0 ND 3.2 >10,000
Human IL-15 ND 2.4 2.3 2.0 2.6 6.5
Human IL-18 ND 15.2 3.4 1.3 14.1 38.9
Human IL-11 0.7 4.7 1.6 1.4 5.5 10.5
Human IL-27 ND ND 1.2 ND 1.4 54.3
Human IL-33 ND ND ND ND 1.1 2.9 ND=Non-detectable*CellswereprimedwithIFN-γfor2hoursbeforeLPSstimulation.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
48
TROUBLESHOOTING
Problem Possible Cause Solution
Bead popula-tionshiftingupward or downward dur-ingacquisition
The strong PE signal from high concentra-tionsamplesorstan-dards may spill over to classificationChannel(e.g.,FL3/FL4/APC)and mess up the bead separation.
OptimizeinstrumentsettingsusingKitSetup Beads, and make appropriate com-pensationbetweenchannels.
Filter plate will not vacuum or some wells clogged
Vacuum pressure is insufficientorvacuummanifold does not seal properly.
Increase vacuum pressure such that 0.2 mLbuffercanbesuctionedin3-5seconds.Clean the vacuum manifold and make sure no debris on the manifold. Press down the plate on the manifold to make a good seal.
Samples have insoluble particlesorsampleistoo viscous (e.g., serum and plasma samples)
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Ifsomewellsarestillcloggedduringwash-ing, try the following:
1).Addbuffertoallthewells,pipetteupand down the clogged wells and vacuum again.
2). Use a piece of clean wipe, wipe the un-der side of the clogged wells and vacuum again.
3). Take a thin needle (e.g., insulin needle), while holding the plate upward, poke the littleholeundereachofthecloggedwellsand vacuum again. Do not poke too hard ortoodeepasitmaydamagethefilterand cause leaking.
Filter plate was used without pre-wet.
Pre-wetplatewithwashbufferbeforerun-ning the assay.
biolegend.com49
LEGENDplex™ Human Cytokine Panel 2
Insufficientbead count or slow reading
Beads inappropriately prepared
Sonicate bead vials and vortex just prior toaddition.Agitatemixedbeadsintermit-tentlyinreservoirwhilepipettingthisintothe plate.
Samples cause beads aggregationduetoparticulatematterorviscosity.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Beads were lost during washing for in-tube assay
Make sure beads are spun down by visu-ally check the pellet (beads are in light blue or blue color). Be very careful when removing supernatant during washing.
Probe might be par-tiallyclogged.
Sample probe may need to be cleaned, or if needed, probe should be removed and sonicated.
Plate leaked
Vacuum pressure set too high
Adjust vacuum pressure such that 0.2 mL buffercanbesuctionedin3-5seconds.Donot exceed 10” Hg of vacuum.
Plate set directly on table or absorbent tow-elsduringincubationsorreagentadditions
Set plate on plate holder or raised edge sobottomoffilterisnottouchinganysurface.
Liquid present on the under side of the plate aftervacuum
Afterwashing,pressdownplatefirmlyona stack of clean paper towels to dry the underside of the plate.
Pipettetouchinganddamagedplatefilterduringadditions.
Pipettetothesideofwells.
High Back-ground
Background wells were contaminated
Avoidcross-wellcontaminationbychang-ingtipsbetweenpipettingwhenperform-ingtheassayusingamultichannelpipette.
InsufficientwashesThe background may be due to non-specificbindingofSA-PE.Increasenumberof washes.
Debris(FSC/SSC) during sample acquisi-tion
Debris or platelet may exist in sample solu-tion.
Centrifuge samples before analyzing samples. Remove platelet as much as possible.
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
50
Variationbe-tweenDuplicate samples
Beadsaggregation Sonicate and vortex the Beads prior to use.
Multichannelpipettemay not be calibrated or inconsistent Pipet-ting
CalibratePipette.Ensuregoodpipettingpractice.Primepipettebeforeusemayhelp.
Plate washing was not uniform
Make sure all reagents are vacuumed out completely in all wash steps.
Samples may contain particulatematters.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Low or poor standard curve signal
The standard was in-correctlyreconstituted,stored or diluted
Followtheprotocoltoreconstitute,storeand dilute standard. Double check your calculation.
Wrongorshortincuba-tiontime
Ensurethetimeofallincubationswasappropriate.
Signals too high, standard curves satu-rated
PMTvalueforFL2/PEset too high
MakesurethePMTsettingforthere-porter channel is appropriate
Plateincubationtimewas too long Useshorterincubationtime.
Sample read-ings are out of range
Samples contain no or below detectable levels of analyte
Make sure the experiment to generate thesamplesworked.Useproperpositivecontrols.
Samplesconcentrationshigher than highest standard point.
Dilute samples and analyze again.
Standard curve was saturated at higher end of curve.
MakesurethePMTsettingforthere-porter channel is appropriate. Use shorter incubationtimeifincubationtimewastoolong
Missed beads populationsduring reading, ordistributionis unequal
Sample may cause some beads to ag-gregate.
Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.
Beadspopulationsarenot mixed properly
Makesureallbeadpopulationsaremixed.and in similar numbers.
biolegend.com51
LEGENDplex™ Human Cytokine Panel 2
PLA
TE M
AP
(for
in-p
late
ass
ay)
1
2 3
4 5
6 7
8 9
10
11
12
A
C0
C4
Sa
mpl
e1
Sa
mpl
e5
Sa
mpl
e9
Sa
mpl
e 13
Sa
mpl
e17
Sa
mpl
e21
Sa
mpl
e25
Sa
mpl
e29
Sa
mpl
e33
Sa
mpl
e 37
B
C0
C4
Sa
mpl
e1
Sa
mpl
e5
Sa
mpl
e9
Sa
mpl
e 13
Sa
mpl
e17
Sa
mpl
e21
Sa
mpl
e25
Sa
mpl
e29
Sa
mpl
e33
Sa
mpl
e37
C
C1
C5
Sa
mpl
e2
Sa
mpl
e6
Sa
mpl
e10
Sa
mpl
e 14
Sa
mpl
e18
Sa
mpl
e22
Sa
mpl
e26
Sa
mpl
e30
Sa
mpl
e34
Sa
mpl
e38
D
C1
C5
Sa
mpl
e2
Sa
mpl
e6
Sa
mpl
e10
Sa
mpl
e 14
Sa
mpl
e18
Sa
mpl
e22
Sa
mpl
e26
Sa
mpl
e30
Sa
mpl
e34
Sa
mpl
e38
E
C2
C6
Sa
mpl
e3
Sa
mpl
e7
Sa
mpl
e11
Sa
mpl
e 15
Sa
mpl
e19
Sa
mpl
e23
Sa
mpl
e27
Sa
mpl
e31
Sa
mpl
e35
Sa
mpl
e39
F
C2
C6
Sa
mpl
e3
Sa
mpl
e7
Sa
mpl
e11
Sa
mpl
e 15
Sa
mpl
e19
Sa
mpl
e23
Sa
mpl
e27
Sa
mpl
e31
Sa
mpl
e35
Sa
mpl
e39
G
C3
C7
Sa
mpl
e4
Sa
mpl
e8
Sa
mpl
e12
Sa
mpl
e 16
Sa
mpl
e20
Sa
mpl
e24
Sa
mpl
e28
Sa
mpl
e32
Sa
mpl
e36
Sa
mpl
e40
H
C3
C7
Sa
mpl
e4
Sa
mpl
e8
Sa
mpl
e12
Sa
mpl
e 16
Sa
mpl
e20
Sa
mpl
e24
Sa
mpl
e28
Sa
mpl
e32
Sa
mpl
e36
Sa
mpl
e40
Tel: 858-768-5800
LEGENDplex™ Human Cytokine Panel 2
52
RA
CK
MA
P (
for
in-t
ub
e as
say)
1
2 3
4 5
6 7
8 9
10
11
12
A
C0
C0
C1
C1
C2
C2
C3
C3
C4
C4
C5
C5
B
C6
C6
C7
C7
Sam
ple
1 Sa
mp
le
1 Sa
mp
le
2 Sa
mp
le
2 Sa
mp
le
3 Sa
mp
le
3 Sa
mp
le
4 Sa
mp
le4
C
Sam
ple
5
Sam
ple
5
Sam
ple
6
Sam
ple
6
Sam
ple
7
Sam
ple
7
Sam
ple
8
Sam
ple
8
Sam
ple
9
Sam
ple
9
Sam
ple
10
Sa
mp
le
10
D
Sam
ple
11
Sa
mp
le
11
Sam
ple
12
Sa
mp
le
12
Sam
ple
13
Sa
mp
le
13
Sam
ple
14
Sa
mp
le
14
Sam
ple
15
Sa
mp
le
15
Sam
ple
16
Sa
mp
le
16
E Sa
mp
le
17
Sam
ple
17
Sa
mp
le
18
Sam
ple
18
Sam
ple
19
Sa
mp
le
19
Sam
ple
20
Sa
mp
le
20
Sam
ple
21
Sa
mp
le
21
Sam
ple
22
Sa
mp
le
22
F Sa
mp
le
23
Sam
ple
23
Sa
mp
le
24
Sam
ple
24
Sa
mp
le
25
Sam
ple
25
Sa
mp
le
26
Sam
ple
26
Sa
mp
le
27
Sam
ple
27
Sa
mp
le
28
Sam
ple
28
G
Sam
ple
29
Sa
mp
le
29
Sam
ple
30
Sa
mp
le
30
Sam
ple
31
Sa
mp
le
31
Sam
ple
32
Sa
mp
le
32
Sam
ple
33
Sa
mp
le
33
Sam
ple
34
Sa
mp
le
34
H
Sam
ple
35
Sa
mp
le
35
Sam
ple
36
Sa
mp
le
36
Sam
ple
37
Sa
mp
le
37
Sam
ple
38
Sa
mp
le
38
Sam
ple
39
Sa
mp
le
39
Sam
ple
40
Sa
mp
le
40
biolegend.com53
LEGENDplex™ Human Cytokine Panel 2
LEGENDplex™ Kits are manufactured by BioLegend 9727 Pacific Heights Blvd.San Diego, CA 92121Tel: 1.858.768.5800Tel: US & Canada Toll-Free: 1.877.Bio-Legend (1.877.246.5343)Fax: 1.877.455.9587Email: [email protected]
For a complete list of world-wide BioLegend offices and distributors, please visit our website at: biolegend.com
Enabling Legendary Discovery™
76119_V03