Keysight N7081A Microwave Transceiver

Page 2: Keysight N7081A Microwave Transceiver

Warranty Statement

The material contained in this document is provided "as is," and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement will control.

U.S. Government Rights U.S. Government Rights. The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/find/sweula. The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Contacting Keysight

Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at: http://www.keysight.com/find/assist

If you do not have access to the Internet, please contact your Keysight field engineer.

In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number. With this information, the Keysight representative can determine whether your product is still within its warranty period.

Safety and Regulatory Information

The safety and regulatory information pertaining to this product is located beginning on Safety Symbols.

Where to Find the Latest Information

Documentation is updated periodically. For the latest information, please visit: http://www.keysight.com/find/n7081asupport

Software Updates

Is your product software up-to-date? Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search for software updates for your product, go to the Keysight Technical Support website at: http://www.keysight.com/find/n7081asupport.

2

Page 3: Keysight N7081A Microwave Transceiver

Technology Licenses

The hardware and / or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.

Safety Symbols

The following safety symbols are used throughout this manual. Familiarize yourself with each of the symbols and its meaning before operating this instrument.

Denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, would result in damage to or destruction of the product. Do not proceed beyond a caution note until the indicated conditions are fully understood and met.

Denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.

3

Page 4: Keysight N7081A Microwave Transceiver

Table of Contents

Operating Precautions ..........................................................................................5

Preventing Electrostatic Discharge Damage ...............................................5

Preventing Electrical Overstress Damage ...................................................6

Preventing Test Port Connector Damage ....................................................7

Installing Your N7081A ................................................................................8

STEP 1. Check the Shipment .......................................................................8

STEP 2. Meet Requirements ........................................................................8

STEP 3. Install the Software .........................................................................9

N7081A Supported Operating Systems ......................................................9

N7081A Supported Programming Languages ............................................9

STEP 4. Connect Power and Cables ............................................................13

STEP 5. Run the N7081A GUI (Graphical User Interface) ...........................21

Getting Help with Your N7081A ...........................................................................23

N7081A on the Internet................................................................................23

N7081A Example Programs .........................................................................23

Contacting Keysight .....................................................................................23

Miscellaneous Information ...................................................................................24

Installing an N7081A Transceiver in an Original Equipment Manufacturer’s (OEM) Product ....................................................................24

Warranty .......................................................................................................27

Maintenance .................................................................................................27

Shipment for Service ....................................................................................27

General Safety Considerations ....................................................................27

Regulatory Information ................................................................................28

Environmental Information ...........................................................................30

4

Page 5: Keysight N7081A Microwave Transceiver

Operating Precautions

Some older desktop PCs (e.g. HP Z400 Workstation) will fail to boot if an N7081A device is connected via USB. This is a bug in the PC BIOS where the N7081A’s VID (0x0957) and PID (0x5F18) are not being recognized. The PC BIOS will freeze during the boot process and the PC will be unstable. To prevent this fail-to-boot experience, do one of the following: (a) remove the N7081A’s USB cable before booting these older PCs, or (b) insert a USB hub between these older PCs and the N7081A, or (c) replace the older PC with a newer model.

Most causes of instrument failure can be avoided if you will apply the information in this section. PLEASE READ!

Preventing Electrostatic Discharge Damage

Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables and assemblies (DUT) to the N7081A. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. To prevent damage to the instrument:

• always have a grounded, conductive table mat in front of your test equipment.

• always wear a grounded wrist strap, connected to a grounded conductive table mat, having a 1 MΩ resistor in series with it, when handling components and assemblies or when making connections.

• always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.

• always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.

• always ground the center conductor of a test cable and DUT before making a connection to the N7081A test port or other static-sensitive device. This can be done as follows:

1. Connect a short (from your calibration kit) to one end of the cable and DUT to short the center conductor to the outer conductor.

2. While wearing a grounded wrist strap, grasp the outer shell of the cable and DUT connector.

3. Connect the other end of the cable and DUT to the test port and remove the short from the cable.

5

Page 6: Keysight N7081A Microwave Transceiver

Figure 1 ESD Protection Setup

The following table lists Keysight part numbers for ESD protection devices. To order, refer to “Contacting Keysight” on page 23.

Description Keysight Part Number

Grounding wrist strap 9300-1367

5 ft grounding cord for wrist strap 9300-0980

2 x 4 ft conductive table mat and 15 ft ground wire

9300-0797

Preventing Electrical Overstress Damage

Excessive RF power or voltage applied to the connectors on the front panel will damage the N7081A. Damage levels are listed in the N7081A Data Sheet, available online at: http://www.keysight.com/find/n7081asupport.

6

Page 7: Keysight N7081A Microwave Transceiver

Preventing Test Port Connector Damage

Proper connector care and connection techniques are critical for accurate and repeatable measurements. The following table contains tips on connector care.

Handling and Storage Do

• Keep connectors clean • Extend sleeve or connector nut • Use plastic end-caps during storage

Do Not • Touch mating-plane surfaces • Set connectors contact-end down • Store connectors or adapters loose

Visual Inspection

Do • Inspect all connectors carefully • Look for metal particles, scratches, and

dents

Do Not • Use a damaged connector - ever

Connector Cleaning

Do • Try compressed air first • Use isopropyl alcohol (see warning below) • Clean connector threads

Do Not • Use any abrasives • Get liquid into plastic support beads

Gaging Connectors

Do • Clean and zero the gage before use • Use the correct gage type • Use correct end of calibration block • Gage all connectors before first use

Do Not • Use an out-of-specification connector

Making Connections

Do • Align connectors carefully • Make preliminary connection contact lightly • Turn only the connector nut • Use a torque wrench for final connection

Do Not • Apply bending force to connection • Over tighten preliminary connection • Twist or screw any connection • Tighten past torque wrench “break” point

Keysight strongly recommends that adapters be used as connector savers to minimize the wear on the N7081A test port connectors. Use instrument-grade quality adapters or better.

Cleaning connectors with alcohol shall only be done with the DC power source disconnected and in a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior to energizing the instrument.

For more information on the proper use and maintenance of connectors, see http://na.support.keysight.com/pna/connectorcare/Connector_Care.htm

7

Page 8: Keysight N7081A Microwave Transceiver

Installing Your N7081A

STEP 1. Check the Shipment

1. Unpack the contents of the shipping container. Keep the packaging material in case the N7081A needs to be returned to the factory.

The N7081A is shipped with ESD-safe protective end caps on the test ports. Save these items for reuse to ensure shipment safety of the N7081A.

2. Carefully inspect the N7081A to make sure that it was not physically damaged during shipment. If damaged, refer to “Contacting Keysight” on page 23.

3. Use the Contents List in the shipping container to verify the completeness of your shipment. If not complete, refer to “Contacting Keysight” on page 23.

STEP 2. Meet Requirements

Before installing the N7081A, ensure the operating environment meets the requirements found in “Environmental Information” on page 30, and the requirements for the DC power source shown below.

Some N7081A specifications and characteristics apply over the internal temperature range of 42°C to 48°C, as indicated by the internal thermal sensor. For complete specifications, refer to the Data Sheet, available online at: http://www.keysight.com/find/n7081asupport.

Requirements for the DC Power Source (Not Supplied)

Voltage: 6 VDC1

Power Consumption: 24 W

1 Can operate from 6 to 18 VDC but higher voltages may increase power consumption and module temperature.

8

Page 9: Keysight N7081A Microwave Transceiver

STEP 3. Install the Software

First time users need to install the N7081A software. In addition, they need to install Microsoft .NET Runtime and the Matlab Compiler Runtime, if not already installed. See the following sections for instructions. It can take up to 30 minutes to install and verify the software. If a previous version of the N7081A software is installed on your computer, uninstall it before installing the new version by clicking Start > All Programs >Keysight N7081A Microwave Sensor API > Uninstall.

N7081A Supported Operating Systems

• Windows 32-bit

• Windows 64-bit

• Linux

N7081A Supported Programming Languages

• C++ (all OS)

• C# (Windows 64-bit, Windows 32-bit)

• VB .NET (Windows 64-bit, Windows 32-bit)

• VB6 (Windows 64-bit, Windows 32-bit)

• MATLAB (Windows 64-bit, Windows 32-bit)

If Not Already Done, Install Microsoft .NET 3.5 Runtime

Microsoft .NET Runtime is normally installed on computers with the Windows 7 or Windows 8 operating system. To see if Microsoft .NET Runtime, version 3.5 or later, is already installed, on the PC desktop click Start > Control Panel > Programs and Features.

1. If Microsoft .NET Runtime, version 3.5 or later, is already installed on your computer, go to the Install the Matlab Compiler Runtime instructions later in this document. It is not necessary to reinstall Microsoft .NET Runtime.

2. If Microsoft .NET Runtime, version 3.5 or later, is not already installed on your computer, navigate to http://www.microsoft.com/en-us/download/details.aspx?id=21.

9

Page 10: Keysight N7081A Microwave Transceiver

3. In the Select Language box, use the drop-down menu to make a selection.

4. Click Download.

5. Click No thanks and continue.

6. Click Run.

7. Follow the prompts to complete the installation.

10

Page 11: Keysight N7081A Microwave Transceiver

If Not Already Done, Install Matlab Compiler 8.0 Runtime

To see if Matlab Compiler Runtime, version 8.0 or later, is already installed, on the PC desktop click Start > Control Panel > Programs and Features.

1. If Matlab Compiler Runtime, version 8.0 or later, is already installed on your computer, go to the Install the N7081A software instructions on the next page of this document. It is not necessary to reinstall Matlab Compiler Runtime.

2. If Matlab Compiler Runtime, version 8.0 or later, is NOT already installed on your computer, navigate to http://www.keysight.com/find/n7081asupport and click N7081A Software Download.

3. Download and install the appropriate 32-bit or 64-bit Matlab Compiler Runtime software, depending on your operating system.

4. At the prompt, click Run.

5. Follow the prompts to complete the installation. This may take 10 minutes or more.

11

Page 12: Keysight N7081A Microwave Transceiver

Install the N7081A Software

1. Navigate to http://www.keysight.com/find/n7081asupport and click N7081A Software Download.

2. Download and install the appropriate 32-bit or 64-bit N7081A Microwave Transceiver software, depending on your operating system.

The following window is from N7081ASetup64.msi. If you are running N7081ASetup.msi, some text in the window will be different.

3. At the prompt, click Run.

4. Follow the prompts to complete the installation.

12

Page 13: Keysight N7081A Microwave Transceiver

STEP 4. Connect Power and Cables

If you have difficulties, refer to Troubleshooting at the end of this step.

1. Attach the 6 VDC power source (not supplied) to the DC IN port on the N7081A rear-panel. The DC IN power jack has a 2 mm center pin and a 6.5 mm outer diameter. When the DC supply is attached, the N7081A is ON immediately; there is no power switch. When ON, the lower left of four LEDs shown in the photo above is continuously lit and the LEDs in the top row blink.

2. Make one of the following connections:

o Attach the supplied USB cable to the N7081A rear-panel and to a USB port on the remote computer. The N7081A rear panel USB port has a USB Type-B mini connector. If the Windows Plug and Play mechanism prompts you to install the USB driver, navigate to the appropriate subdirectory under C:\Program Files (x86)\Keysight\N7081A\CypressUsbDriver.

o Attach a LAN cable (not supplied) to the N7081A rear-panel and to a LAN port in your data system.

Important LAN Notes

• The N7081A does NOT support DHCP (Dynamic Host Configuration Protocol).

• The factory-assigned IP address is: 192.168.0.100. This can be changed from the API or the GUI. For more information, see Connecting to the N7081A in the Help file, available online at: http://www.keysight.com/find/n7081asupport.

• The N7081A LAN port must be connected to a 1Gbit capable LAN port provided by a PC, router, or switch. Networking hardware with a slower speed (for example, a 10/100 Mbit network) will not function with the N7081A directly. However, this situation is accommodated by connecting the N7081A to a 1Gbit switch and then connecting the 1GBit switch to the slower 10/100Mbit network.

3. Optional: connect Digital I/O, and 10 MHz REF cables (not supplied). The Digital I/O port has an HDMI Type-C mini connector and is used for triggering applications – for more information, refer to the Help system, available online at: http://www.keysight.com/find/n7081asupport. The 10 MHz REF ports have MCX connectors

4. To check USB functionality, on the PC desktop click Start > All Programs > Keysight N7081A Microwave Transceiver API, and run the N7081A Blink Application. Refer to the following graphic.

13

Page 14: Keysight N7081A Microwave Transceiver

o Click Query Number Sensors. The number returned should be 1 or greater, depending on the number of transceivers you are using.

o Click Connect and Blink.

If the green LEDs next to the N7081A front panel test ports blink for several seconds, this verifies that USB communications are working properly. Proceed to Step 5. If the LEDs do not blink, refer to the following section on Troubleshooting.

14

Page 15: Keysight N7081A Microwave Transceiver

Troubleshooting: Scenario 1: When you connect the USB cable between the N7081A and the remote computer, your computer displays an error message such as “unknown device” or “the device driver software was not successfully installed”, or you are unable to blink the lights in the Blink Application. Try any or all of the following suggestions.

• Suggestion 1: Open a command window and run the command set path. A correct response is shown in the following graphic:

If the response to your command does not match that shown in the graphic, try editing the environmental variable for Path. To do so, click Computer > System Properties > Advanced System Settings > Environmental Variables. In the lower pane, scroll to Path, click Edit, and enter the following variable value: %N7081A_SENSOR_PATH%bin;%SystemRoot%\system32;%SystemRoot%;%SystemRoot%\System32\Wbem;%SYSTEMROOT%\System32\WindowsPowerShell\v1.0\;C:\Program Files\MATLAB\MATLAB Compiler Runtime\v80\runtime\win64

• Suggestion 2: Reinstall the Cypress USB driver.

o Click Start, right-click Computer, and click Manage.

o In the left pane of the “Computer Management” window, click Device Manager.

o If you see IO Interface, update its driver to the one provided.

o Otherwise, in the right pane, click the arrow next to Universal Serial Bus controllers.

o In the right pane, look to see if N7081A Microwave Transceiver or Unknown Device is listed. If you see N7081A Microwave Transceiver, the Cypress USB driver is properly installed, so you can stop following the instructions in this suggestion. If you see Unknown Device, click it, because the Cypress USB driver is not properly installed.

15

Page 16: Keysight N7081A Microwave Transceiver

o In the “Unknown Device Properties” window, click the Driver tab, then click Update Driver….

16

Page 17: Keysight N7081A Microwave Transceiver

o In the “Update Driver Software – Unknown Device” window, click Browse my computer for driver software.

o In the “Update Driver Software – Unknown Device” window, click Browse, and navigate to the appropriate subdirectory under C:\Program Files (x86)\Keysight\N7081A\CypressUsbDriver. (If the CypressUsbDriver directory is missing, repeat the step Install the N7081A Software.)

o Click Next.

17

Page 18: Keysight N7081A Microwave Transceiver

o Click Install the driver software anyway and follow the prompts.

18

Page 19: Keysight N7081A Microwave Transceiver

• Suggestion 3: Run the CyControl.exe tool to verify the Cypress USB driver installed properly.

o Navigate to C:\Program Files (x86)\Keysight\N7081A\bin.

o Run CyControl.exe

o If N7081A Microwave Transceiver appears at the top of the left pane in the “USB Control Center” window (as shown in the following graphic), the driver is working properly. (If missing, repeat the step Install the N7081A Software or Reinstall the Cypress USB Driver.)

19

Page 20: Keysight N7081A Microwave Transceiver

Scenario 2: When you try to run the N7081A software, you get the following error message.

Suggestion: Load the appropriate version of Matlab Compiler Runtime. For example, if you have a 64-bit system, but want to run the N7081A 32-bit program WatsonGUI.exe, you must load the 32-bit version of Matlab. Refer to Install the Matlab Compiler Runtime.

20

Page 21: Keysight N7081A Microwave Transceiver

STEP 5. Run the N7081A GUI (Graphical User Interface)

• On the PC desktop, click Start > All Programs > Keysight N7081A Microwave Transceiver API.

• Click either N7081A MATLAB GUI (32-bit version) or N7081A x64 MATLAB GUI (64-bit version), depending on your OS.

• The following text-free window will appear for up to a minute. Please wait for a dialog box.

• If using only one N7081A transceiver, Click Connect

• If using multiple N7081A transceivers,

o Click Query.

o Click the transceiver of your choice in the list.

o Click Connect.

21

Page 22: Keysight N7081A Microwave Transceiver

• Use the GUI (shown below) to make a measurement. For information on making measurements, refer to the N7081A Help system at: http://www.keysight.com/find/n7081asupport.

The Graphical User Interface (GUI) supplied with the N7081A provides some of the basic features and capabilities required to measure and display S-parameters with the N7081A. While not intended to be a full featured network analyzer application, this GUI can still be a useful tool to help understand the operation or to evaluate the performance of the N7081A. The GUI can also be useful in developing or troubleshooting solutions based on the N7081A.

22

Page 23: Keysight N7081A Microwave Transceiver

Getting Help with Your N7081A

N7081A on the Internet

The N7081A Support website is a resource for help. You can find the N7081A documentation (listed below), the outline drawing, and other resources at: http://www.keysight.com/find/n7081asupport.

The N7081A Measurement Transceiver Support website includes the following online documentation, available in English only:

• N7081A Microwave Transceiver Installation Guide (this document)

• Help file

• N7081A Data Sheet

• Outline Drawing

N7081A Example Programs

• C++ measurement example programs can be found at C:\Program Files (x86)\Keysight\N7081A\examples\MeasurementExample.

• C# measurement example programs can be found at C:\Program Files (x86)\Keysight\N7081A\examples\MeasurementExampleCs.

• Matlab measurement example programs can be found at C:\Program Files (x86)\Keysight\N7081A\matlab\MatlabExample

• VB.NET measurement example programs can be found at C:\Program Files (x86)\Keysight\N7081A\examples\N7081AVB

• VB6 measurement example programs can be found at C:\Program Files (x86)\Keysight\N7081A\examples\N7081AVB6

Contacting Keysight

Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at: www.keysight.com/find/assist

If you do not have access to the Internet, please contact your Keysight field engineer.

In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number. With this information, the Keysight representative can determine whether your product is still within its warranty period.

23

Page 24: Keysight N7081A Microwave Transceiver

Miscellaneous Information

Installing an N7081A Transceiver in an Original Equipment Manufacturer’s (OEM) Product

When intended for installation as a component within an OEM device for use in Hazardous Locations contact CTD Support ([email protected]) for additional information. The most important consideration when installing the N7081A module in an OEM product is temperature level and stability. Here are some general guidelines for installation of a module:

• The N7081A is warranted to operate between 0° C and 70° C as indicated by the internal thermal sensor. The module will automatically turn off if the module temperature exceeds 70° C.

• The module’s temperature should be verified to be within operating limits by monitoring its internal thermal sensor.

• Up to 24 watts can be dissipated within the N7081A (with all receivers ON), resulting in approximately a 32° C rise in module temperature over the ambient temperature. This assumes there is no air flow around the module and the module is insulated from any thermal conducting material.

• It is recommended the N7081A be cooled with forced air in the direction indicated in the outline drawing, viewable at: http://www.keysight.com/find/n7081asupport. The drawing also shows the locations and sizes of the N7081A mounting holes.

• Mounting the N7081A to a thermally conductive surface can also lower the temperature rise above ambient, but it is the responsibility of the OEM to determine if this reduces the temperature sufficiently for their application and environment. It is recommended to put a thermally conductive material between the N7081A and the mounting surface for better heat transfer.

• The module requires a stable temperature in order to retain its calibration. Some N7081A performance parameters are specified within a temperature range between 42° C and 48° C as indicated by the internal thermal sensor. For more temperature information, refer to the N7081A Data Sheet, viewable at: http://www.keysight.com/find/n7081asupport.

24

Page 25: Keysight N7081A Microwave Transceiver

25

Page 26: Keysight N7081A Microwave Transceiver

?

26

Page 27: Keysight N7081A Microwave Transceiver

Warranty

The actual warranty on your instrument depends on the date it was ordered as well as whether or not any warranty options were purchased at that time. To determine the exact warranty on your instrument, contact Keysight Technologies with the model and serial number of your instrument. Refer to “Contacting Keysight” on page 23.

The N7081A warranty is voided if the cover is removed.

Maintenance

To remove dirt or dust from the external case of the N7081A, clean the case using a dry or slightly-dampened cloth only.

To prevent electrical shock, disconnect the N7081A from the DC power source before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.

Shipment for Service

Contact Keysight Technologies for instructions on where to ship the N7081A for service. Refer to Contacting Keysight on the previous page.

Ship the N7081A using the original packaging materials. Shipping the N7081A in anything other than the original packaging may result in non-warranted damage.

Safety Symbols

The following safety symbols are used throughout this manual. Familiarize yourself with each of the symbols and its meaning before operating this instrument.

Denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, would result in damage to or destruction of the product. Do not proceed beyond a caution note until the indicated conditions are fully understood and met.

Denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.

General Safety Considerations

Safety Earth Ground

The product chassis should be properly connected to earth ground. Failure to ensure adequate earth grounding may cause product damage and serious injury.

27

Page 28: Keysight N7081A Microwave Transceiver

Before Applying Power

If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.

The measuring terminals on this instrument are designed to be used with external signals described in Measurement Category I, but NOT with external signals described in Categories II, III, and IV.

Servicing

These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.

No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers.

Regulatory Information

This section contains information that is required by various government regulatory agencies.

Instructions for Use

This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.

Instrument Markings

The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.

This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005. All electrical and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE Directive, 2002/96/EC.

The CE mark is a registered trademark of the European Community. (If accompanied by a year, it is when the design was proven.)

This is a symbol of an Industrial Scientific and Medical Group 1 Class A product.

This is a marking to indicate product compliance with the Canadian Interference-Causing Equipment Standard (ICES-001).

Cet appareil ISM est conforme a la norme NMB du Canada.

28

Page 29: Keysight N7081A Microwave Transceiver

Direct current.

The RCM mark is a registered trademark of the Australian Communications and Media Authority.

China RoHS regulations include requirements related to packaging, and require compliance to China standard GB18455-2001.

South Korean Certification (KC) mark; includes the marking's identifier code which follows the format: KCC-REM-YYY-ZZZZZZZZZZZZZZ or MSIP-REM-YYY-ZZZZZZZZZZZZZZ.

This symbol indicates compliance with China RoHS regulations for paper/fiberboard packaging.

UL Recognized Component Mark for U.S.A. and Canada.

Declaration of Conformity

A declaration of conformity is available upon request, or a copy is available on the Keysight Technologies Web site at http://regulations.about.keysight.com/DoC/search.htm or by contacting Keysight. See “Contacting Keysight” on page 23.

Acoustic Statement

European Machinery Directive 2006/42/EC Acoustic noise emission LpA <70 dB Operator position Normal operation mode

EMC Information

Complies with European EMC Directive 2014/30/EU.

• IEC/EN 61326-1

• CISPR Pub 11 Group 1, class A

• AS/NZS CISPR 11

• ICES/NMB-001 This ISM device complies with ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.

29

Page 30: Keysight N7081A Microwave Transceiver

South Korean Class A EMC Declaration

This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.

Environmental Information

Samples of this product have been type-tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation and end-use; those stresses include but are not limited to temperature, humidity, shock, vibration, and altitude. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.

Parameter Required Values/Ranges

Temperature

Operating 0 °C to +70 °C (as indicated by internal thermal sensor)

Storage −40 °C to +70 °C (ambient)

Operating Altitude 0 to 4,600 meters (≈15,000 feet)

Relative humidity Type tested, 0% to 95% at 40 °C, non-condensing

This product has been certified for use as a component within process control equipment for use in Class I, Division 2, Groups A, B, C, and D Hazardous Locations. For further environmental information, see the N7081A Data Sheet (Keysight part number N7081-90002). To view this document online, use the following steps:

1. Got to http://www.keysight.com.

2. In the Search boxes:

a. Use the drop-down menu to select Electronic Test and Measurement.

b. Enter N7081-90002, the part number of the Data Sheet and click Search.

3. Click the hyperlink title of the N7081A Data Sheet to view the PDF.

30

Page 31: Keysight N7081A Microwave Transceiver

This information is subject to change without notice.

© Keysight Technologies 2014-2015 Manufacturer: Keysight Technologies 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1783, USA Print Date: November 6, 2015 Supersedes: September 30, 2015

www.keysight.com