(kevin: omni streak logo here)...the omni control module is shipped complete and ready to run. the...

28
Omni Personal Air Sampler and BGI 400 Omni NIST Traceable – ISO 9001:2008 Operation/Calibration /Maintenance BGI Incorporated 58 Guinan Street Waltham, MA 02451 Tel: 781 891 9380 Fax: 781 891 8151 www.bgiusa.com [email protected] Version 1.1.1 May 2011

Upload: others

Post on 09-Apr-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

Omni Personal Air Sampler and BGI 400 Omni

NIST Traceable – ISO 9001:2008

Operation/Calibration /Maintenance

BGI Incorporated 58 Guinan Street

Waltham, MA 02451 Tel: 781 891 9380 Fax: 781 891 8151 www.bgiusa.com [email protected]

Version 1.1.1 May 2011

Page 2: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

2

Notice Beginning with Version 1.3.02 an important improvement to the “Run Programmed” screens was introduced. Operating instructions for these new screens are contained in Appendix A Table of Contents Section Topic Page 1.0 Quick Start 3 2.0 Specifications 4 3.0 Hardware setup 5 4.0 Calibration 7 5.0 Maintenance 8 6.0 Printed/Illustrated Operating Instructions 12 Appendix A Run Set up instructions for Firmware 1.3.02 and higher 25

Page 3: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

3

1.0 Quick Start If you have already set up the Omni to be placed in service, as described in Section 3.0, you may run the instrument immediately. Referring to Figure 1 press the button labeled I/O.

Figure 1: Omni Keypad

When the main idle display screen appears press Enter. The instrument will immediately begin running at the previous flow rate. To stop the Omni, press Enter. The complete details of the run will be displayed on the screen as shown in Figure 2. To cancel this information press Escape.

Figure 2: Image of a Completed Run

Page 4: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

4

The Omni has been designed to be a highly interactive instrument. Investigators familiar with air sampling using microprocessor controlled hardware will have little difficulty following the critical paths without further instruction. For complete instructions regarding the set up and operation of the Omni we have prepared a virtual instruction manual (VIM). This manual may be viewed on your computer screen after loading the CD provided with your instrument. The most up to date version of the VIM may be downloaded from our site: http://www.bgiusa.com/aam/Omni.htm This VIM is the LCD screen and button array of the actual instrument. Thus a virtual instrument may be run ‘on screen’. Each screen has been instructionally annotated. The second half of this manual is a printed version of the instructions. 2.0 Specifications.

Primary Flow rate 3-12 lpm (± 1%) Secondary Flow Rate 15 lpm for 20 min Temp. operational range -30° C to 50° C Temp. reading range -30° C to 50° C (±0.5° C) Barometric pressure range 400 to 800 mm of Hg (± 5mm) Pulsation Ratio Under 10% Noise Flow Rate Static Pressure Noise Level Lpm cm of H2O dB(A) 12 28.5 58 10 25 56 8 17.5 46 6 12.6 44 4-2 <12.6 <40 Dimensions: 6.00 in. high (15.24 cm) x 3.75 in. wide (9.52 cm) x 3.00 in. deep (7.62 cm) NiMh Battery Pack: 5.00 in. high (12.70 cm) x 3.5 in. wide (8.89 cm) x 1.5 in. deep (3.81 cm) Weight: 1.86 lbs (0.845 kg) NiMh Battery Pack weight: 1.36 lbs (0.616 kg) One complete instrument comprises: Quantity Description Catalog/Part No. 1 Omni module 6012 1 Battery Pack NiMH 6015 1 Power Supply 6007 1 Instruction Manual Download PDF file from

BGI website & Disk 1 Software Disk CD Catalog

Page 5: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

5

Available Accessories: 1 Adaptor for direct battery charging 6008 3.0 Hardware setup 3.1 Battery connection. The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will contain only a limited amount of charge. The battery is normally charged through the instrument proper. Connect the Battery and the charger power supply to the Omni module as shown in Figure 3. If it is desired to charge additional batteries, separately, an adaptor (P/N 6007) is available which will permit the power supply to be connected, directly to the battery.

Figure 3: Omni module schematic

3.2 Operational set up. It is important, not to run the Omni without a sampling device attached to the inlet. The instrument has been provided with a feature, which makes it impossible to do so for more than a few seconds. If an inlet resistance is detected, which is less than 0.5 cm of H2O, the pump will

Page 6: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

6

stop running and an error message will be displayed which alternates between RUN INTERRUPTED/SP TOO LOW. When this occurs, press Escape and the pump will be reset. The pump can use a very wide range of filters and may be set to run from 3 to 15 lpm at resistances up to 50 cm of H20. The exceptions to this rule are illustrated in Table 1. Table 1 Omni Flow Envelope

Flow Rate LPM

Max. Resistance Cm of H2O

Primary Precision %

Secondary Precision %

Min. run time Hours Max. Resistance Cm of H2O

2 50 2 1 24 50

3-12 50 1 0.5 24 25

15 12 1 0.5 0.3 12

Primary precision is the result obtained when one calibration value is used for the entire flow rate range. Secondary precision is the result obtained when the calibration is performed at or near the desired flow rate.

New units are delivered set up and calibrated for Volumetric air flow rate (Q), Temperature (T) and Barometric Pressure (BP). All initial calibrations are performed during initial setup of a new instrument. All calibrations are performed using NIST Traceable Standards. In the absence of physical or electrical damage it is not known what time interval will transpire, if any, before a calibration is no longer valid. Therefore, the user should be guided by one or both of two philosophies. 1. It is generally held by air sampling experts throughout the world, that good practice requires a calibration audit at the beginning and end of a study. If the study is protracted in length, monthly or quarterly checks are considered adequate. If no significant change is observed, longer intervals can be considered. 2. It is very common for air sampling studies to be conducted under the aegis of a local, National or trade organization. It is very likely that a client or governing body will have specific rules or suggestions regarding audits and recalibration. In some cases, highly detailed instructions will exist. BGI manufactures three instruments for audit/calibration purposes. They are NIST traceable and will provide readings for Volumetric/Standard flow rate, as well as Barometric Pressure and Temperature. They can be found on the internet at the following locations: http://www.bgiusa.com/cal/tetracal.htm http://www.bgiusa.com/cal/the_challenger.htm http://www.bgiusa.com/cal/deltacal.htm

Page 7: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

7

4.0 Calibration Any, user provided BP or Temp. instruments which are satisfactory to the investigator may be utilized as no direct connection to the Omni is required. When performing a flow audit other, conventional volumetric flow measuring devices are suitable providing they do not impose very high static pressures on the instrument. Calibrators that introduce a pressure spike into the system are not suitable for use with the Omni. If there is any doubt regarding the device to be used contact BGI at: [email protected] . Always perform a flow calibration using a clean filter of the type to be utilized in your sampling study. Suitable filters are fiberglass, quartz and the type of PTFE, which is labeled 2μm pore size and has a plastic ring surrounding it. Note: the pore size has nothing to do with the filtration efficiency of the material. For further information on this topic see: www.bgiusa.com/cau/filter.htm A typical schematic of the setup for performing a volumetric flow rate calibration is shown below in Figure 4. While an investigator may attach BGI or other calibrators to the filter tennon stub of the Omni after removing the inlet with user provided hoses and adaptors, BGI provides two useful appliances: p/n 2738 for use with the deltaCal p/n 2739 for any other calibrator

Figure 4: Schematic Drawing of Flow Calibration Arrangement

Page 8: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

8

The exact screen/button sequences for performing calibration are detailed in the virtual instruction manual: http://www.bgiusa.com/Omni.htm and in the printed instructions in the second half of this publication. 5.0 Maintenance The Omni itself has no particular maintenance requirements. The exterior surfaces should be kept clean and free of contamination. The instrument is splash proof but not waterproof. Immersion will destroy the unit. When the instrument is not in service, the battery should be charged and run down at monthly intervals. After this cycle the battery should be recharged. Various devices of BGI or other manufacture should be cleaned and maintained according to their individual instruction manuals. In the event that wear or damage is suspect and return to the factory is inconvenient, the following instructions are intended to be of assistance. A complete, detailed illustration of the Omni is shown on the next page as BGI print number C2892. If a larger view is required it may be obtained electronically at; http://www.bgiusa.com/aam/Omni.htm Principle maintenance items, in decreasing order of importance are: no. 41, Omni pump assembly, no. 4, Pulsation dampner assembly, no. 35, orifice and the external battery.

Page 9: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

9

Fig. 5: Exploded View of Omni Components

Page 10: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

10

5.1 Pump assembly It is highly recommended that repairs and service are carried out at BGI, by trained personnel. Nevertheless, it is recognized that there may be reasons or need for user performed servicing. Replacement intervals for pump parts are difficult to predict due to variations in severity of service. Nevertheless it is possible to offer a general guide: Part Replacement Interval in Hours Valves 2000 Pump diaphragm 2500 Dampner 2500 Ball Bearing on eccentric 4000 Motor 4500 Utilizing exploded diagram Figure 6 as a guide:

1. Remove the double valve chest. Note position and method of securing valves and discard, unless they are in reusable condition. 2. Examine valve seat Detail 10. Use non-metallic probe, for example, a wooden toothpick, to remove dirt. Use care at valve edge.

3. Fit new valves, or replace old ones. 4. Replace valve chest. Do not use too much pressure when tightening screws. 5. If, when the valve chest is removed, the slightest damage to the diaphragms

was observed, then the following operations are to be performed. 6. Withdraw the screw, now visible in the center of the support disk. Withdraw

the diaphragm and both support disks. Note direction of convolution in diaphragm and replace new diaphragm in the same direction. This is to be done on both sides of the assembly.

7. If for any reason the motor or eccentric bearings are suspect due to age, reduced performance or increasing noise then they must be serviced/replaced.

It is not recommended, that the replacement of the bearings on the bearings on the eccentric shaft be attempted. The bearings and eccentric are available as a complete assembly.

8. Remove valve chests, clamp rings and diaphragm as above and draw away the plastic yoke assembly.

9. Loosen set (Grub) screw and withdraw eccentric and bearings as a complete assembly.

10. Remove three screws holding motor in place and withdraw motor from chassis.

11. Carefully unsolder electrical leads from old motor and transfer to new motor. 12. Reassemble in reverse order.

Page 11: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

11

Fig 6: Exploded Diagram of Pump Assembly

Page 12: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

12

5.2 Pulsation Dampner

The Dampeners, detail 4, figure 5 are a glued up assembly and are not user reparable. The complete dampner assembly may be returned to the factory for rebuilding or a complete new assembly acquired. 5.3 Orifice and inlet fittings The orifice and fittings leading to it from the inlet should not require any maintenance, provided an inlet filter has always been utilized. If upon inspection dirt has found to have accumulated in the internal passages, then careful disassembly and cleaning is required. The only recommended cleaning materials are lint free cloths and clean water. If necessary a pipe cleaner may be utilized on the orifice. Be careful not to loose part no. 29. This is a fine screen to prevent the ingress of foreign objects. While this segment of the pump is apart be certain to check for cracked tubing, gaskets or “O” rings. Replace damaged components as required. It is utterly vital to reassemble all components in a manner completely free of leaks. 5.4 Battery The battery used in the Omni is a detachable Nickel metal hydride (NiMH) cell. It should provide years of maintenance free service. The end of the batteries useful life will reveal itself by a consistent inability to provide at least 24 hours of service under moderate sampling conditions in temperate climates. However, one service item must be observed if the unit has not been used for more that two months. Full performance may not be achieved until the battery has been charged and discharged 2-4 times. 6.0 Printed/Illustrated Operating Instructions The main operating instructions for the Omni are in virtual format. They are included with each new instrument in the form of a CD disc or they may be downloaded at: http://www.bgiusa.com/aam/Omni.htm For those investigators who do not or cannot utilize a virtual format the instructions have been printed and are presented herein.

Page 13: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

BGI 58 Guinan St. Waltham, MA

02451 V: 781 891 9380 F: 781 891 8151

www.bgiusa.com August 2009 INSTRUCTIONS: Omni Introduction Operating the Omni 0

2

>Turning ON the Omni: 1) Press the button. An initiation screen will briefly appear, then advance to the “Main Menu”. This screen will have the serial number and the code version number. >At the Main Menu:

1) Scroll using the and buttons to make a selection. Position the

in front of the selection and press the button. Run Now Initiates a sampling event. Run Programmed Setup and initiate a programmed sampling

event. Set Fcns/Cal Setup Time, Units and calibration Functions. Select Language Change language Note: Prior to using the Omni, it is wise to set up the Date, Time and

Preferences. Advance to the “Set Preferences” section of this manual.

Page 14: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

14

3

4

9

>Run Now: From the “Main Menu”:

1) If necessary, scroll to the “Run Now” selection, using the and

buttons, press the button. 2) The pump begins running and the conditions and run parameters are

displayed on the LCD. While the pump is running the conditions and run parameters are continuously displayed on the LCD. These are the parameters from the previous set run. >Stopping the Run: 1) While the pump is running, press the “Enter” button, to stop the run. 2) The final run data will be displayed on the LCD. Note: After the “Run”, pressing “ESC” will cause the elapsed run data to disappear. Pressing “ESC” will cause it to reappear. Run information is not lost until overwritten by a new run. >Power Off: 1) Press the button to turn “OFF” the Omni and save the data for

downloading later.

2) At the Power “OFF” screen, select “yes” and press the button to turn “OFF” the unit. When the Omni is turned “ON” the next time, the data will be available for download or it can be recalled from the main idle display screen (screen 2) by pressing “ESC.”

Page 15: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

15

Download the Data to a computer: 1) Open BGI Link. 2) Select the screen shown below.

If you have selected the com port already, you will be switched to the next screen. If you have not, go to the instructions for BGI Link.

Click on the Save icon and the data will be given a code number and saved for you in a JOB file:

Page 16: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

16

If your Version Number is 1.3.02 or higher follow the instructions in Appendix A

2

>Programming the Run:

1) Scroll the to “Run Programmed” using the and buttons then

press the button.

Page 17: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

17

5

>Program Menu: Select Delay The amount of time before the sample event

begins. Select Duration Sample Run Time. Start Run Start the program. >Select Delay:

1) Scroll to “Select Delay”, then press the button.

> Setup via computer. A single or multiple number of frmOmni Instruments may be programmed to run at any future time through a computer. 1) Connect a computer to the frmOmni via serial cable and RS232 port and turn it on. 2) Open BGI Link and select the Setup Run screen:

Double clicking on this screen will open the program section dealing with this function

Device type should be selected first, immediately followed by Flow Rate. If the frmOmni is selected the flow rate is locked at 5 lpm. It is not possible to set the PQ100 or Omni above or

Page 18: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

18

below their built in limits. The flow rate may be highlighted or selected with the up/dn button:

. Midnight is a popular start point for many applications, as is 24-hour duration. They may be selected by clicking on the icons and the display will instantly display your selections as shown below:

In order to insert other dates and times you may use the arrow selection buttons , or by highlighting the values and typing in new ones. >Setting Preferences:

2

> Set Flow Rate:

1) Scroll to “Set Flow Rate”, then press button.

Page 19: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

19

13

2

14

>Setting the target flow rate:

1) The LCD shows the target flow rate. Press the and buttons to change the flow rate.

Note: Each press equals 10 ml/min. Press and hold to accelerate

the speed of the numeric change.

2) Press the button to accept the target flow rate and return to the “Main Menu”.

> Set Fcns/Cal:

2) Scroll to “Set Fcns/Cal,” then press button. >Setup and Calibration Menu: Date and Time Set current date and time. Cal. Temp. Adjust temperature to match a TetraCal

(formally triCal) temp. reading. Cal. BP Adjust Barometric Pressure to match the

TetraCal (formally triCal) BP Reading. Cal. Flow rate Adjust flow rate to match a TetraCal

(formally triCal) Flowmeter. Select BP unit Set the units preference for Barometric

Pressure, (mm of Hg is the default value). >Making a Selection:

1) To select “Date and Time”, scroll using the and buttons,

then press the button to accept.

Page 20: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

20

15

>Setting the Date and Time:

1) Move the () using the and buttons.

2) Press the button to select the item. The item will then flash.

3) Use the and buttons to correct the numeric value. Press and hold to accelerate the speed of the numeric change.

4) Press the button to accept the value and the () will automatically advance to the next item.

5) Select “Done” to return to the “Setup and Cal Menu”.

14

16

> Calibrate Temperature:

1) To select “Cal. Temp.”, scroll using the and buttons,

then press the button to accept. >Adjusting the Temperature 1) Compare the temperature reading from the Omni to a TetraCal

(formally triCal) or other standard. 2) If they differ, change the numeric value on the Omni using the

and buttons.

3) Press the button to accept the value and return to the “Setup and Cal Menu.”

Page 21: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

21

14

> Calibrate Barometric Pressure:

1) To select “Cal. BP.”, scroll using the and buttons, then

press the button to accept.

20

14

>Adjusting the Barometric Pressure 1) Compare the Barometric Pressure reading from the Omni to a

TetraCal (formally triCal) or other standard. 2) If they differ, change the numeric value on the Omni using the

and buttons.

3) Press the button to accept the value and return to the “Setup and Cal Menu”.

>Calibrate Flow rate: The preferred way to calibrate the Omni is to use the TetraCal (formally triCal) Auto Adjust Mode. The TetraCal (formally triCal) Auto Adjust works as follows: The TetraCal (formally triCal) puts out a continuous stream of flow rate information in ascii format. When the “TetraCal (formally triCal)” auto mode is selected on the pump menu, the pump is instructed to look for the stream of flow rate data. It then compares the TetraCal (formally triCal) flow rate data to it’s own flow rate information and calculates an offset and then automatically adjusts the pump motor speed to match the data coming from the TetraCal (formally triCal). At the Setup and Calibration Menu:

1) Scroll using the and buttons to the “Cal. Flow Rate”

position. Press the button to accept.

Page 22: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

22

20

At the TetraCal (formally triCal) Auto Adjust Menu Connect the Pump and calibrator using tubing and filter.

TetraCal (formally triCal) and pump connected 1) Turn the TetraCal (formally triCal) “ON” and allow it to zero,

itself.

2) Using the and buttons, scroll to the “TetraCal (formally

triCal)” position and press the button. The pump will automatically begin to run.

22

At this point the pump instantaneously compares it’s data to the TetraCal (formally triCal) data and calculates an offset.

1) When the flow readings on the Omni is stable, press the

button.

Page 23: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

23

20

13

>Manual Calibration

Manual

1) Using the and buttons, scroll to the “Manual”

position and press the button. The pump will automatically begin to run. > Adjusting Flow:

1) Using the and buttons, adjust the flow reading on the Omni, to match the reading on the TetraCal (formally triCal) or other flow meter.

2) Press the button to accept the Calibration.

Page 24: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

24

14

25

> Selecting Barometric Pressure Units:

1) Scroll using the and buttons, to the “Select BP unit”

position, then press the button to accept. >Setting the Barometric Pressure Units

1) Move the () using the and buttons to select the barometric pressure units that are correct for you.

2) Press the button to accept the units.

Page 25: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

25

Appendix A Run Set up instructions for Firmware 1.3.02 and higher With the introduction of firmware Version 1.4.02 in August of 2009, setting up a programmed run became simpler. The need to calculate the number of days, hours and minutes prior to the start of a run was eliminated. A run will now commence upon the exact date and time selected. From the main menu screen select Run Programmed by using the Up/Dn buttons and press Enter.

The screen will change to the one shown below with Select Start as the default choice, press Enter.

The screen will change to the one shown below with Select Start as the default choice, press Enter unless you wish to start at Midnight, in which case use the Up/Dn buttons to select Start @ Midnight and then press Enter.

If you selected Start at Midnight, you have completed the set up choice. If you selected Select Start, you will be taken to the following screen.

Page 26: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

26

Press enter in order to select the year. If it is correct, press enter again to select the month. If it is incorrect, use the Up/Dn buttons to correct the year and press enter to select the month. If the month is correct, press enter again to select the day. If it is incorrect, use the Up/Dn buttons to correct the day and press enter to select the day. If the day is correct, press enter again to select the hour. If it is incorrect, use the Up/Dn buttons to correct the day and press enter to select the hour. If the hour is correct, press enter again to select the minute. If it is incorrect, use the Up/Dn buttons to correct the hour and press enter to select the minute. If the minute is correct, press enter again to select the minute. If it is incorrect, use the Up/Dn buttons to correct the minute and press enter to select Done. Press Enter. The screen will change to the one shown below with Select Duration as the default choice. If you desire to run for 24 hours, use the Up/Dn buttons to select it. If you wish a custom run time choose Select Duration and press Enter. The is will take you to the following screen.

Press Enter to select Hrs. and use the Up/dn keys to select the number of hours required. Press Enter. Use the Up/dn keys to select the number of hours required. Press Enter. This leads to the Done selection, press Enter to complete the set up procedure. This action leads to the Start screen.

Page 27: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

27

When the Start Run Key is selected, the countdown to the initiation of the run begins, resulting in a screen like the one shown below.

Once the countdown screen appears the run may still be canceled, by simply pressing the Esc Key. After the run commences it may only be canceled by pressing, in sequence:

Page 28: (Kevin: omni streak logo here)...The Omni control module is shipped complete and ready to run. The Standard NiMH battery is shipped in an as received and tested condition and will

28

Revision List Rev. Change Date 1 Creation Sept. 2005 1.0.1 Updated ISO Registration March 2009 1.0.2 Changed Run Time Set Up Screen August 2009 1.1.0 Updated to 400 Model April 2010 1.1.1 Added to Title May 2011