film thickness measurement by absolute methods
TRANSCRIPT
Film thickness measurement by absolute methods received 1 July 1970; accepted 3 July 1970
C Greaves BSc, PhD, Alnst, Lanchester Polytechnic, Eastlands, Rugby, Warwicks
Absolute methods for measuring the thickness of thin films are considered and a performance survey is given for a selection of commercial instruments.
1. Introduction strate with a suitable step formed by masking during deposition. The treatment of thin film monitoring techniques previously A vertical tracking force as low as 1 mg is often used to reduce given by the author’ is extended in the present article toabsolute the possibility of damage to the surface of the deposit. In prac- methods for measuring film thickness. As pointed out in the tice, however, the surface is usually marked by the stylus which previous article many monitoring techniques require pre- can cut into the material, particularly if it is a soft material. calibration against a suitable absolute method. A selection of This can lead to inaccuracies and, therefore, it is often desirable such methods is reviewed in this present article with particular that an even coating should be deposited over the whole area of emphasis being placed on those methods where measuring the specimen after the step has been formed, so that the effect instruments are commercially available. Absolute methods are should be self cancelling across the step. usually carried out on the bench rather than in situ in a vacuum The vertical movement of the stylus, as it traverses the speci- system and often require more skill in operation than the moni- men, is magnified by a pivoted arm which carries the stylus and toring techniques used during the production of thin films. is detected and converted into an electrical signal by a solid
Some of the monitoring measurements previously described state transducer or an electromechanical system. The resulting can be performed on the bench and in this sense can be regarded electrical signal is amplified and displayed on a meter or prefer- as absolute methods provided a satisfactory correlation factor ably a pen recorder chart in terms of the vertical movements of between thickness and the actual measured physical property is the stylus. Step height measurements can be readily obtained known. Two suitable methods are the modulated beam photo- from the distance between the mean lines corresponding to the meter and microbalance techniques. The modulated beam film surface and substrate surface as shown in Figure 1. photometer can be operated just as readily on the bench to measure optical reflectance or transmittance which, if required, can be converted to thickness. However, this conversion is only possible over a rather limited range and calibration against a truly absolute method is really required. Provided that the mass of the substrate is known, microbalance techniques may be Film surface used to measure the surface density of a deposit quite accur-
-- A
ately. The range of suitable microbalances is even wider than those available for monitoring since null balance techniques are no longer essential as they are in monitoring applications. How- ever, better reliability may be obtained if null balance tech- niques are used since the weighing operation can then be carried out in a vacuum enclosure thereby reducing inter- ference from draughts, errors due to contaminants and other unwanted effects.
Step height 10 nm
1 -- - 2. Stylus Method Substrate The method is an extension of well established surface profile surface
measurement techniques and has been described by Wright?. The thickness of a film of any material is measured by traver- sing a small radius blunt diamond stylus across a test groove formed in the material or over the edge of a step from the deposit to the substrate. The thickness is recorded as the Figure 1. Typical trace for stylus method on X 1000 000 range. difference in levels between film surface and substrate surface.
It is desirable that the substrate should be smooth compared The instrument is not absolute in the strictest sense since for with the thickness of the deposit. To ensure this, measurements the best reliability it should be calibrated from time to time are often made on a sample deposited on an optically flat sub- using a set of calibrated grooves. The calibration grooves them-
Vacuum/volume 20/number 10. Pergamon Press LtdlPrinted in Great Britain 437
C Greases: Film thickness measurement by absolute methods
selves require initial calibration by, for esample, an inter- ference method.
PhotographIc plate
3. Interferometer Methods An interferometer technique was first used to measure film thickness by Wiener3 in 1887. Similar techniques’-” arc now well established in many applications other than film thickness measurement, for example surface topography measurements and hardness measurements. The advantage of such methods is that they give an absolute metrical thickness rather than an optical thickness or a thickness derived from some other property.
3.1 Multiple beam interferometry using fringes of equal thickness. The method is suitable for any film material. A film is deposited onto a glass substrate, preferably an optical flat; if half of the substrate is masked the deposited film will form a step. A fairly thick layer of opaque material, such as 80 nm or more of silver, must be deposited over the film and substrate, as shown in Figure 2, in order to increase the reflectivity of the film to pro- duce high definition reflection fringes and to ensure that any phase change occurring in reflected light from each side of the step should be the same.
Reference plote- I I Portiolly tronsporent-I\\ coating
Step
Opoque coating -
Film
Substrote
Figure 2. End view of wedge shaped air gap system.
The step film is used to form the bottom of a wedge and the top is formed by a similar substrate with a partially transparent coating of about 80 per cent reflectivity as shown in Figures 2 and 3. The reflection coating on the reference plate is provided to enhance the definition of the fringes. The substrates are fixed so that lines of equal air gap thickness run along the substrates parallel to the common edge and at right angles to the film step; effectively two wedges are formed side by side.
Reference plate Reference plate Poftlolly transparent Poftlolly transparent cootlng cootlng
1 1 Substrote Figure 3. Side view of wedge shaped air gap system.
Monochromatic light is used to illuminate the wedge arrange- ment perpendicularly and the reflected light is viewed with a low power microscope as shown in Figure 4. For reflected light an interference pattern of straight dark fringes on a mono-
433
Beam\ Splitter
Low power mlcroscope
Monochromatic
Collimator
Wedge shaped air gap
Figure 4. Optical arrangcmcnt for multiple beam interferometry using fringes of equal thickness in the reflection mode.
chromatic background is set up as shown in Figure 5. The pattern indicates points of equal air gap thickness with a dis- placement at the step in the film and by suitable analysis will reveal the film thickness at the step.
For normal incidence, from conventional interference con- siderations, it can be shown that the difference in air gap between two adjacent fringes is 1/2. The height of the step (i.e. film thickness) can be obtained from the displacement of a fringe at the step in terms of a fraction of fringe spacing and thus give the film thickness in terms of units of A/2. In practice
Substrate
.
Film
/
/*I 2
Figure 5. Part of a typical fringe pattern of the step between substrate and film using fringes of equal thickness and monochromatic light.
C Greases: Film thickness measurement by absolute methods
a common technique is to photograph the fringe pattern and carry out the analysis on the photographic plate.
This technique is limited to films of thickness greater than about 3 nm by the physical character of the films rather than by optical limitations. Tolansky” has stated that, from a resolution point of view, the displacement in fringes \\;hich arc on oppo- site sides of a dividing line, as in Figure 5. can be judged to i/5th of a fringe and for highly defined fringes this corresponds to approximately 0.5 nm. The restriction to 3 nm is mainly due to the phenomenon of “bedding-in”“*” caused by the granular nature of very thin lilms. If a film is thicker than i/Z (i.e. approximately 300 nm for sodium light), then the step will show an overlap of orders which may lead to confusion. However, in practice it is possible to overcome this difficulty. It is almost certain that any mask used to form the step \vill not be perfectly straight and sharp but will impart a certain recognizable pattern on the fringes which can, with skill, bc used to identify the relevant order of displaced fringes.
Thickness measurements can be satisfactorily achieved using general purpose interference equipment but the amount of specialist skill required can be reduced by using the specially designed multiple beam interferometers which are commercially available for the purpose.
The equal thickness fringe patterns required in this tech- nique may be set up at the step in the film by methods other than the formation of a wedge shaped air gap. For example, a modification of the classical Michelson interferometer arrange- ment may be used in which the specimen takes the place of the usual movable mirror in the standard Michelson instrument, but in which the required relative movement is obtained by moving the entire interferometer with respect to the stationary specimen. Another method uses a double quartz prism and polarized light to form lines of equal path difference by com- paring the specimen surface at two positions. A possible advan- tage of these two methods is that no separate reference plate is required and the need for a physical contact between such a plate and the specimen is avoided thus reducing the risk of accidental damage to the film surface.
3.2 Multiple beam interferometry using fringes of equal chro- matic order. This is again a method which is suitable for any film material. It is often preferred to the method using fringes of equal thickness since flatness of the substrate is no longer critical and it is possible to use direct readings from the drum of a constant deviation spectrometer when obtaining the thick- ness. It has the disadvantage that considerably more skill is required in interpreting the fringe pattern since white light is used to illuminate the film.
The sample may be prepared, in a similar way to that des- cribed in Section 3. I, to form a step with a highly reflective coat- ing covering the whole substrate or an alternative method, used successfully by Scott et al”, of scraping a groove out of the film down to the substrate level may be employed. In either case a localized fringe system is set up between the specimen and a partially transparent reference plate by illuminating normally with white light and viewing with a spectrometer as shown in Figure 6. The interference plates may be adjusted until a small area of nearly constant thickness becomes visible, when a pattern of dark fringes on a continuous background with a displacement at the step is obtained (Figure 7).
Perhaps the simplest way to obtain film thickness from this fringe pattern is that originally described by Scott et all” in
Beam, Splitter
l--v--l Spectrometer
White light source
Parallel sided
Adiustina screw
Figure 6. Optical arrangement for multiple beam interferometry using fringes of equal chromatic order in the reflection mode.
which an accurately calibrated constant deviation spectrometer is used to measure the wavelength and wavelength displacement of particular lines in the pattern. From conventional inter- ference considerations, for normal incidence and for reflected light, the interference equation is
16, d.=2y+n
where 1~ is the order, A is a particular wavelength, y is the air gap thickness and 8 is the phase change on reflection at the sur- face. This assumes that there is no phase change due to passage of the light through the partially transparent coating of the reference plate; an assumption which is justifiable if the reflec- tivity of this coating is greater than 80 per cent. At the step in the film there is a displacement of a particular fringe and assum- ing that the phase change is the same on both sides of the step by differentiation of (1).
and since ~nA,=(rrz+l)l~ (3) eliminating 111 gives
a, da h’=m2 . z
(4)
where the displacement dy corresponds to the film thickness, &,, is the displacement of the fringe corresponding to r3, and d, is the wavelength of the adjacent fringe, as shown in Figure7.
439
C &eaves: Film thickness measurement by absolute methods
m
k I -- di --
m+l
m+2
Order Wovelength
Substrate Film
/
Figure 7. Part of a typical fringe pattern at the step between substrate and film using fringes of equal chromatic order and white light.
Thus by reading i. ,, I2 and CL?,, from the drum of the spectro-
meter the film thickness may be obtained. Figure 7 is much
simplified and in practice a much more complex pattern occurs, with suitable skill many combinations of adjacent wavelengths may be used to determine the film thickness.
Since fringes of equal chromatic order can be made narrower than the Fizeau fringes used in the method described in Section 3.1, much greater accuracy can be obtained and the method is ideally suited for use with extremely thin films.
4. Ellipsometer Methods It has been well known for many years” that ellipsometry offers a precise method for studying and measuring the thickness and refractive index of thin reflecting films deposited on to a suit- able substrate. Ellipsometer methods have not been as exten- sively used as inteference methods but can prove to be extremely sensitive. The high degree of skill required, particularly in the cumbersome process of converting experimental readings to thickness and refractive index values, has probably been the main contributor to this lack of popularity. Ellipsometer methods utilize measurements of angles, phase differences and amplitude ratios.
When plane polarized light is reflected from a thin reflecting film the state of polarization of the light is altered, usually becoming elliptically polarized. The magnitude of these polarization changes is dependent on the thickness of the film and its refractive index. Thus a study of the state of polarization can reveal both the thickness and refractive index of the film. The quantities tan y, the ratio of the reflected amplitudes for light polarized parallel to and perpendicularly to the plane of incidence, and A, the relative phase difference between them, can be measured directly’* and used to calculate film thickness. Alternatively w and A may be obtained from measurements of the orientation of the major axis of the elliptically polarized light to the plane of incidence, and 11 the ellipticity. The full significance and relationship of these quantities has been ex- plained in a review by Bennett and Bennett13.
Experimental apparatus required for ellipsometry is fairly standard consisting essentially of a spectrometer table on which the specimen is mounted, a light source, polarizer, com- pensator, analyzer and detector. Accuracy depends to a large extent on the rate of change of light intensity with angular
440
rotation of the compensator or analyzer near to the extinction
positions. An mstrument which can be used to accurately determine s and ;I has been developed by the British Scientific
Instrument Research Association. The instrument, shown in Figure 8. incorporates a magneto-optical detection system to
* Light source
// d7 ii Collimator
,’ /
\ Quarter wove plate
Photo -multiplier
Figure 8. Optical system of photoelectric ellipsometer developed by BSIRA.
obtain a higher sensitivity and accuracy. A Nicol prism ensures that the incident light is plane polarized before it is reflected from the specimen when it becomes elliptically polarized. A quarter-wave plate compensator in a particular orientation with reference to the axis of the elliptically polarized light restores the light to its plane polarized form. A Faraday cell modulates the plane of polarization of the light about a mean position and the light then passes to the analyzer, a second Nicol prism. At the extinction position balanced amounts of light pass through the analyzer to be detected by the photo- multiplier. The output from the photomultiplier is combined vectorially with the modulation signal on an oscilloscope pro- ducing Lissajous figures which provides an accurate detection method. This procedure allows x and y to be determined from the orientations of the compensator and analyzer at extinction
Table 1. Com
parison of
film
thickness m
easurement
methods
Method
Classification Principle
Comm
ents
Comm
ercial instrum
ent
Manufacturer
Type Approxim
ate operating range
Comm
ents
Stylus m
ethod Direct
measurem
ent of
height Suitable
for m
etallic, Rank
Precision Industries
Ltd, Talystep
I I nm
to
10jrm
of step
from
lilm
to substrate
scmiconducting
or Rank
Taylor Hobson,
(8 ranges)
by traversing
a stylus
across diclcctric
lilms.
For Lcicester,
England it.
Vertical m
ovcmcnt
is cxtrcm
cly thin
lilms
converted by
a transducer deposit
should bc
on an
G V
Planar Ltd.
Surface Profile
5 nm
to 5 jirn
system
to give
electrical optically
flat substrate
Windmill
Road. M
onitor SPM
IO (3
ranges) signal
proportional to
the Sunbury-on-Tham
cs, thtckncss
Middlescu,
England
Multiple
beam
An equal
thickness fringe
Suitable for
metallic.
Varian Vacuum
Division.
A-scope 3nm
to2jtm
Simple
to operate
since tntcrfcrom
ctry ustng
system
is set
up using
mono-
scmiconducting
or 61 I
tlansen W
ay, Palo
Alto, interferom
eter specially
designed for
thin frinacs
of equal
chromatic
light. The
diclcctric lilm
s. if
totally California
94303, USA
film
measurem
ent. Fringe
thickness dtsplaccm
cnt bctwccn
the two
covcrcd by
a thick
pattern set
up using
wedge fringe
patterns al
the step
rcllcctivc him
. Prcfcrablc
shaped air
gap bctwcen
botwccn the
tilm
and to
“SC optic;llly
Rat spccim
cn and
similar
plate substm
tc gives
the step
subsrrntc acting
as a
reference height
in term
s of
the wavclcngth
of the
light Gacrtner
Scicntilic Corp.
M307 3
nm
to 3 jrm
M
odifcd Michelson
Inter- I201
Wrightwood
Ave. M
icro fcrom
cter. Specim
en replaces
Chicago, III
60614. USA
intcrfcromctcr
movable
mirror,
it rem
ains stationary
while entire
j” intcrferom
ctcr is
moved
0 AZ
Fringe pattern
set up
by xi
C Rcichcrt
Optischc W
crke AG,
McF
Microscope
IO
nm
to a
few ,,rn
Wicn
XVII, E
and polarization
creating lines
of equal
path Hcrnalscr
Haupstr 219,
Interferometer
difference using
a double
22 Austria
attachment
quartz prism
and
polarized 3
5 light
to exam
ine the
specimen.
2 Rofcrcnce
plate not
rcquircd
Ill
2 Intcrferom
etry using
An equal
chromatic
order Suitable
for m
etallic. Rank
Precision Industries
Ltd, Thin
film
measuring
3 nm
IO
a few
I’m
Manufacture
of N 130
now frinacs
of equal
frtngc system
is
set up
using scm
iconducting and
Analytical Division.
Interfcrencc discontinued.
Fringe pattern
5 chrom
atic order
\\ hitc light.
The displaccm
cnt diclcctric
lilms
if totally
Hilgcr 61 W
atts, M
icroscope N
130 set
up using
parallel sided
air
E of
the fringes
at the
s~cp covcrcd
by a
thick 31
Camden Road,
and W
avclcngth gap
bctwccn specim
en and
?i bcf\\ccn
the lilm
and
the rcllccting
lilm.
Optically London
NWI,
England Spcctrom
cter D9003
reference plate
: substrate
gives the
step height
flat substrate
not ncccssary
3 in
terms
of line
wavclcngths
E in
the illum
rnuting light
ul 2 Ellipsom
etcr m
ethods The
state of
polarization of
Suitable for
metallic,
Measurem
ents are
generally
5
polarized light
rellcctcd from
sem
iconducting and
more
difficult to
perform
than the
film
gives inform
ation diclcctric
tilms.
The m
ethods above
r which
can bc
used to
obtain tcchniquc
is m
ade casicr
Gacrtncr Scientific
Corp. Lll8
I nm
to a few
jtrn Rotation
readings to
0.1’. thickness
and refractive
if the
lilm
is non-
I201 W
rightwood Ave.
Ellipsometer
Uses Nicol
prisms
E indcs
of the
film
I= absorbing.
If the
him
is Chicago.
Ill 60614,
USA LL
absorbing a m
ore Ll
l8G-f 1 nm
to
a few I’m
Greater
accuracy achieved
by . .
Ellipsomctcr
using Glan
Thompson
2 com
plex m
ethod is
rcquircd prism
s
5 G
Ll19 1 nm
to
a few
,*rn Rotation
readings to
0.01”. Ellipsom
eter Uses
Glan Thom
pson prism
s c,
C &eaves: Film thickness measurement by absolute methods
thus w and A can be calculated using the following expressions from the electromagnetic theory of light: cos 2 v=cos 2y cos 2s (3 and tan A=tan Zy/tan 2s. (6)
Drude” laid the theoretical foundations of ellipsometry in 1889 but used simplifying approximations which are only true if the film thickness is very much smaller than the wavelength of the incident light. A more exact treatment has been developed by Lucy” which gives an explicit formula when the film thick- ness is less than the light wavelength. For films of greater thickness the only approach open is to find, by graphical or arithmetical methods, the values of thickness and refractive index which fit the experimental observations. There are various analytical techniques available1s~17 but all are extremely com- plex and often require the use of a computer. A comprehensive coverage of experimental and theoretical techniques of ellipso- meter methods for studying surfaces and thin films has been given at a Symposium in 1964l*.
Va&kls has shown that the sensitivity of ellipsometer methods varies with film thickness and has indicated that an acceptable accuracy can still be achieved down to film thick- nesses of the order of 1 nm.
5. Comparison of methods The basic principles and ranges of the various methods are compared in Table I. Where applicable brief details of a selec- tion of commercial instruments are given.
References
3O Wiener. ll’ied.4r~~1. 31, ISS7. 629. ,’ S Tolansky, Alultipl~~ Beonr Il,r~,t:f~,~u,,rc/~~, cf Sut:frrc~~~,~ u~rd Films 194S, (o.Yford: Clorcv/dlJn Pwss). 5 S Tolansky. Sw:fhe t\li~~~otop~~~~rrrph~~, (I OhO), (Lo~~dwr: Lo~rgt~rowrr). o 0 S Heavens. Proc Phys SW. 64, 1951, 419. ’ A F Gunn and R A Scott, Ntrrrw, 158, 1946, 671. ” D G Avery, Nor,rw. 163, 1949, 916. !’ S Tolansky, ~\lrrlriple BCUIII Irrrerfirem~e Microscopy qf klerols, 1970, p 95. (Lo/rdo/r: .Actrd1wric PImY). *‘I G D Scott. T A McLauchlin and R S Scnnctt, J Appl P/t.vs. 21, 1950. 843. ” P Drude, H7ed Am. 36, 1589. 432. ” R E Hartman, J Opf Sot .Avr. 41. 1951. 244. “I H E Bennett and J M Bcnnclt. Ph.rsics o/Thin Filers, 4, (Ed G Hess mrd R E T/m, 1967. p I. (New York: AC&W;C Prc~s,s). I’ F Lucy, J C/w,,, Ph.w 16, 1948. 167. Ii A V&i&k, J Opr SW Am, 37. 1947, 979. I6 F L McCracklin. E Passoglia. R R Stromberg and H L Steinberg, J RPS Nnrl Bur Std. 67c\, 1963. 363. Ii D A Holmes, Appl Oprics. 6, 1967, 168. Iy E Passoglia, R R Stromberg and J Kruger (Editors), E//ipsontrfr.v itr /he ~~T~SIII’E~I~~II oJ’.wqk~s crud tbi~r~films. Nnr Brrr SIP Mis Pub1 No 256 1964.
442