comparison of central corneal thickness measurements with a new optical device and a standard...
TRANSCRIPT
J CATARACT REFRACT SURG - VOL 32, MARCH 2006
Comparison of central corneal thickness
measurements with a new optical device
and a standard ultrasonic pachymeter
Gabor Nemeth, MD, Alexis Tsorbatzoglou, MD, Katalin Kertesz, MD, Attila Vajas, MD,
Andras Berta, MD, PhD, DSci, Laszlo Modis Jr, MD, PhD
PURPOSE: To compare central corneal thickness (CCT) values obtained with ultrasonic pachymetryand a new optical method using partial coherence interferometry (PCI).
SETTING: Department of Ophthalmology, Medical Health and Science Center, University of Debrecen,Debrecen, Hungary.
METHODS: The study comprised 136 eyes of 70 patients whose spherical refractive error was notgreater than G6.0 diopters (D) and whose keratometric astigmatism was not greater than 2.0 D.Central corneal thickness was measured 5 times with a new optical device (ACMaster, Zeiss) andwith an ultrasonic pachymeter (AL-2000, Tomey). All measurements were obtained by the sameinvestigator.
RESULTS: Mean CCT was 531.2 mm G 3.9 (SD) with PCI and 547.8 G 36.0 mm with the ultrasonicdevice. The difference between groups was significant (P Z .001). There was no difference betweenCCT values measured in right and left eyes (P Z .55) with ultrasonography and PCI (P Z .67). Thecoefficient variation was 0.73% for PCI and 6.5% for ultrasonography. Correlation between the CCTmeasurements with both devices was strong and statistically significant (Spearman correlation Z .91,P Z .001).
CONCLUSIONS: Mean CCT values measured by the PCI method were significantly smaller than thosemeasured by the ultrasonic device. Central corneal thickness measured with PCI is more reproducibleand seems to be more reliable than that measured by ultrasonography.
J Cataract Refract Surg 2006; 32:460–463 Q 2006 ASCRS and ESCRS
Corneal thickness measurement has become an important
factor in most types of anterior segment surgery in the eval-
uation of corneal dehydrating function and for determining
intraocular pressure. To avoid complications in refractive
surgery such as keratectasia after laser in situ keratomileu-
sis or perforation in photorefractive keratotomy, accurate
corneal thickness measurements are also important.1–5
Accepted for publication August 5, 2005.
From the Department of Ophthalmology, Medical and HealthScience Center, University of Debrecen, Debrecen, Hungary.
No author has a financial or proprietary interest in any material ormethod mentioned.
Reprint requests to Gabor Nemeth, MD, Department of Ophthal-mology, University of Debrecen, Nagyerdei Blvd. 98, H-4012 De-brecen, Hungary. E-mail: [email protected].
Q 2006 ASCRS and ESCRS
Published by Elsevier Inc.
460
For decades, the standard technique of central corneal
thickness (CCT) measurement was the contact ultrasono-
graphic method; however, differences in corneal thickness
measurements have been reported between ultrasonic pa-
chymeter devices.6,7 The aim of this study was to evaluate
CCT by 2 different methods: a new device using noncontact
partial coherence interferometry (PCI) and the standardcontact ultrasound (US) method.
PATIENTS AND METHODS
The study comprised 136 eyes of 70 patients (70 right and 66left eyes). The mean age of the patients was 66.2 years G 11.3(SD) (range 36 to 86 years) with a male-to-female ratio of36:34. Central corneal thickness was measured with a new devicethat used the PCI method (ACMaster, Zeiss) and with an ultra-sonic pachymeter (AL-2000, Tomey). Patients with a history ofwearing contact lenses, a spherical refractive error greater thanG6.0 diopters (D), keratometric astigmatism greater than 2.0 D,
0886-3350/06/$-see front matterdoi:10.1016/j.jcrs.2005.12.138
CCT MEASUREMENTS BY PCI AND US
or any anterior segment abnormalities were excluded from thestudy.
Five separate, sequential measurements were performed by 1examiner, first with the noncontact PCI device and then with theUS device. The mechanism of the PCI method has been de-scribed.8–10 For the PCI method, the patient assumed a sittingposition, placing the chin on a chin rest and the forehead intoa headband. The patient was instructed to keep both eyes openand to focus the eye being tested on the built-in measurement tar-get of the device.
For the US method, the corneas were anesthetized with top-ical tetracaine hydrochloride. Patients were again asked to sit andlook straight at a target on the wall. The probe tip of the ultrasonicpachymeter was applied to the central cornea perpendicularly andthe surface of the cornea was slightly touched. The probe tip wassterilized with alcohol after use on each patient.
Statistical analysis was performed and CCT data were de-scribed in terms of means and standard deviations. The intraob-server variability was described with a coefficient of variation (CV).Differences in CCT values between the devices were recordedwith the paired test of Wilcoxon, and a P value of .01 was consid-ered the level of significance. The association between groups wasdescribed with the Spearman correlation.
RESULTS
The mean CCT was 531.2 G 3.9 mm when measured
by the PCI device and 547.8 G 36.0 mm when measured
by US pachymetry. The difference between the CCT values
was strongly significant (P Z.001). There was no significant
difference between the right and left eyes (PCI, P Z.67; US,
P Z.55) (Table 1). Figure 1 shows the measured CCTs.
Intraobserver variability was lower with the PCI
method than with the US technique (CV 0.73% and 6.5%,respectively). There were no differences in intraobserver
variability between right and left eyes with PCI (CV 0.8%
versus 0.72%) or US (CV 6.5% versus 6.58%). The correla-
tion between pachymetric devices was strong, as shown in
Figure 2 (r Z .91; P Z .001).
DISCUSSION
Ophthalmic biometry should provide rapid, objective,
and accurate measurements of different ocular parameters.
Change in corneal thickness is an important sign of wors-ening function of the corneal endothelium. In corneal re-
fractive surgery, a high degree of accuracy in corneal
thickness measurement is required.3–5
Table 1. Mean central corneal thickness values with PCI and with
ultrasonic pachymeter.
Eye PCI Ultrasonography
Right 533.6 G 4.3 mm 545.9 G 35.9 mmLeft 528.9 G 3.8 mm 549.5 G 36.2 mm
J CATARACT REFRACT SURG
Ultrasound pachymetry has been the standard methodof evaluating corneal thickness for the past few decades.
Numerous new pachymetric techniques are now available,
although ultrasonic pachymetry is still the common stan-
dard. Many studies have compared CCT measurement
using different methods and devices (eg, specular micro-
scope, optical coherence tomography, ultrasound biomicro-
scope, Orbscan (Bausch & Lomb), and devices based on
PCI).7,11–16
In this study, CCT was measured with a new noncon-
tact ophthalmic device developed for the anterior segment
of the eye using the PCI method, and the results were com-
pared with measurements obtained using the standard
ultrasonic pachymetry technique. The mean CCT values
obtained with the noncontact optical method were smaller
than those derived from the ultrasonic method. The differ-
ences in CCT values were statistically significant (P Z.001).The unquestionable advantage of the PCI technique is
the noncontact method. The noncontact (ie, noninvasive)
pachymetry method has important advantages besides its
short measuring time. There is no need for anesthesia
and no risk for corneal infection, which ensure a high de-
gree of patient comfort.
Giasson and Forthomme17 report significant left–right
differences in ultrasonic CCT measurements due to thehandedness of the examiner; ie, left corneal thickness was
measured significantly larger by a right-handed examiner.
This phenomenon has been reported in several studies
and has been attributed to measurements not being per-
formed perpendicularly to the corneal surface.11 In our
Figure 1. The CCT values measured with the ACMaster and AL-2000 ultra-
sonic pachymeters. The mean CCTs are presented with continuous (US)
and dotted lines (PCI).
- VOL 32, MARCH 2006 461
CCT MEASUREMENTS BY PCI AND US
Figure 2. Mean CCT values obtained with PCI and
ultrasonic pachymetry.
study, the investigator sat on the left side of the patient
while measuring CCT with the ultrasonic device. There-
fore, measuring CCTof the right eye was somewhat compli-
cated, but this factor did not generate large statistical
differences between right and left CCT values (P Z .55).One possible explanation for this finding is that our ultra-
sonic device measures the CCT in an automated mode only
when the probe tip is perpendicular to the corneal surface.
Some authors postulate additional reasons for the
larger corneal thickness values measured with ultrasonic
pachymeters, including ultrasonic probe tip tilting, epithe-
lial edema from local anesthesia, repeated pressure of the
cornea,18,19 and inadequate calibration of the devices.Others suspect that the speed of sound varies between
different layers of the cornea and between normal and
pathologic corneal tissues.6,20,21 Some authors recommend
using a modified refractive index for optical coherence
J CATARACT REFRACT SURG462
tomography (OCT) measurements (n Z 1.365) and a mod-
ified velocity of sound (1664 m/s) for ultrasonic CCT mea-
surements.11 We measured the CCT with ultrasonography
at the sound velocity of 1640 m/s.
Mean corneal thickness values obtained with opticalpachymeters (PCI method), OCT, and Orbscan were signif-
icantly less than values obtained with US pachymetry, sug-
gesting that US-based devices slightly overestimate corneal
thickness.11–14,16,22 These differences might be clinically
relevant. However, the reason for the significantly smaller
CCT values measured with the PCI method, OCT, and
Orbscan remains unknown. Other studies mention larger
CCTs obtained with specular microscopy, US biomicros-copy, and findings opposite the above-mentioned Orbscan
findings.13 Moreover, it is impossible to state whether the
optical or the US measurements are closer to the actual
corneal thickness. The measured CCTs depend on the
Table 2. Differences between CCT values measured with different methods and devices and recorded as extracted value of the 2. method by the 1. method.
Author* 1. Method 2. Method Micrometer Difference (2.-1. method)
Bechmann22 OCT Ultrasonography 50Modis15 Noncontact specular Ultrasonography 28Rainer18 PCI Ultrasonography 21.5Rainer18 Orbscan Ultrasonography 20Wheeler7 Specular Ultrasonography 20Fishman12 OCT Ultrasonography 12Wirbelauer24 OCT Ultrasonography 8Modis15 Ultrasonography Specular 68Fishman12 Ultrasonography Orbscan 40Tam14 Ultrasonography Specular 22Tam14 Ultrasonography Ultrasound biomicroscopy 5
Present study PCI Ultrasonography 16.6
OCT Z optical coherence tomography; PCI Z partial coherence interferometry
*First author
- VOL 32, MARCH 2006
CCT MEASUREMENTS BY PCI AND US
refractive index of the cornea and the US velocity in the cor-
nea. In our study, we compared the CCT values obtained
using a new optical device and an ultrasonic device. The
mean CCTwas 16.6 mm greater when measured with ultra-
sonography than when measured with the PCI method.
This is consistent with data derived from previous studies(Table 2).11–16,22
The intraobserver reproducibility represents the vari-
ability of values in a series of measurements (all of our mea-
surements were taken by the same investigator). The results
of our study show that the intraobserver variability with the
PCI method is almost one-tenth smaller than with the US
method (CV 0.73% versus 6.5%). This contradicts the liter-
ature in which these data are comparable. However, somedata obtained with the Orbscan system show even greater
variability. This fact confirms the accuracy and reproduc-
ibility of CCT values observed with the PCI method.13,23
A high positive correlation coefficient was obtained be-
tween the 2 methods of central corneal pachymetry (r Z0.91; P Z .001), suggesting that the CCT values of these
methods are count over to each other, but are not
interchangeable.In conclusion, we have shown that the CCT values
measured by the ultrasonic pachymeter are higher than
those obtained by the PCI method. The differences between
optical and ultrasonic pachymetry measurements were sta-
tistically significant. The accuracy of these pachymetric de-
vices is acceptable in measuring CCT. Further studies are
required to confirm these findings and to evaluate the accu-
racy, reproducibility, and independence of corneal pachy-metry research using the PCI method.
REFERENCES
1. McLeod SD, Kisla TA, Caro NC, McMahon TT. Iatrogenic keratoconus:
corneal ectasia following laser in situ keratomileusis for myopia.
Arch Ophthalmol 2000; 118:282–284
2. Joo C-K, Kim-G. Corneal ectasia detected after laser in situ keratomi-
leusis for correction of less than 12 diopters of myopia. J Cataract Re-
fract Surg 2000; 26:292–295
3. Vinciguerra P, Camesasca FI. Prevention of corneal ectasia in laser in
situ keratomileusis. J Refract Surg 2001; 17:S187–S189
4. Argento C, Cosentino MJ, Tytiun A. Corneal ectasia after laser in situ
keratomileusis. J Cataract Refract Surg 2001; 27:1440–1448
5. Rowsey JJ, Balyeat HD. Radial keratotomy: preliminary report of com-
plications. Ophthalmic Surg 1982; 13:27–35
J CATARACT REFRACT SURG
6. Reader AL III, Salz JJ. Differences among ultrasonic pachymeters in
measuring corneal thickness. J Refract Surg 1987; 3:7–11
7. Wheeler NC, Morantes CM, Kristensen RM, et al. Reliability coefficients
of three corneal pachymeters. Am J Ophthalmol 1992; 113:645–651
8. Fercher AF, Mengedoht, Werner W. Eye-length measurement by inter-
ferometer with partially coherent light. Opt Lett 1988; 13:186–188
9. Hitzenberger CK, Baumgartner A, Drexler W, Fercher AF. Interferomet-
ric measurement of corneal thickness with micrometer precision. Am J
Ophthalmol 1994; 118:468–476
10. Drexler W, Baumgartner A, Findl O, et al. Submicrometer precision bi-
ometry of the anterior segment of the human eye. Invest Ophthalmol
Vis Sci 1997; 38:1304–1313
11. Marsich MM, Bullimore MA. The repeatability of corneal thickness
measures. Cornea 2000; 19:792–795
12. Fishman GR, Pons ME, Seedor JA, et al. Assessment of central corneal
thickness using optical coherence tomography. J Cataract Refract
Surg 2005; 31:707–711
13. Salz JJ, Azen SP, Bernstein J, et al. Evaluation and comparison of
sources of variability in the measurement of corneal thickness
with ultrasonic and optical pachymeters. Ophthalmic Surg 1983;
14:750–754
14. Tam S, Rootman DS. Comparison of central corneal thickness
measurements by specular microscopy, ultrasound pachymetry,
and ultrasound biomicroscopy. J Cataract Refract Surg 2003; 29:
1179–1184
15. Modis L Jr, Langenbucher A, Seitz B. Corneal thickness measurements
with contact and noncontact specular microscopic and ultrasonic pa-
chymetry. Am J Ophthalmol 2001; 132:517–521
16. Rainer G, Findl O, Petternel V, et al. Central corneal thickness measure-
ments with partial coherence interferometry, ultrasound, and the
Orbscan system. Ophthalmology 2004; 111:875–879
17. Giasson C, Forthomme D. Comparison of central corneal thickness
measurements between optical and ultrasound pachometers. Optom
Vis Sci 1992; 69:236–241
18. Rainer G, Petternel V, Findl O, et al. Comparison of ultrasound pa-
chymetry and partial coherence interferometry in the measurement
of central corneal thickness. J Cataract Refract Surg 2002; 28:2142–
2145
19. Herse P, Siu A. Short-term effects of proparacaine on human corneal
thickness. Acta Ophthalmol (Copenh) 1992; 70:740–744
20. Coleman DJ, Lizzi FL, Jack RL. Ultrasonography of the Eye and Orbit.
Philadelphia, Lea & Febiger, 1977; 113–114
21. Chivers RC, Round WH, Zieniuk JK. Investigation of ultrasound axially
traversing the human eye. Ultrasound Med Biol 1984; 10:173–188
22. Bechmann M, Thiel MJ, Neubauer AS, et al. Central corneal thickness
measurement with a retinal optical coherence tomography device
versus standard ultrasonic pachymetry. Cornea 2001; 20:50–54
23. Doughty MJ, Zaman ML. Human corneal thickness and its impact on
intraocular pressure measures: a review and meta-analysis approach.
Surv Ophthalmol 2000; 44:367–408
24. Wirbelauer C, Scholz C, Hoerauf H, et al. Noncontact corneal pachyme-
try with slit lamp-adapted optical coherence tomography. Am J Oph-
thalmol 2002; 133:444–450
- VOL 32, MARCH 2006 463