chapter - metoclopramide - shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/38966/17... ·...
TRANSCRIPT
CCCHHHAAAPPPTTTEEERRR ---111000
MMMEEETTTOOOCCCLLLOOOPPPRRRAAAMMMIIIDDDEEE
Chapter 10-Metaclopramide Page 218
10.1 DRUG PROFILE
METOCLOPRAMIDE
Metoclopramide is an antiemetic and gastroprokinetic agent [127]. It is used mainly
for the treatment of nausea, vomiting and facilitates gastric emptying in patients with
gastro paresis [128-129]. It is also a drug of a choice for the treatment of migraine headaches
Metoclopramide is safe drug for usage in all stages of pregnancy with no harmful
effect observed in mother and foetus [130]. The antagonist activity at D2 receptors in the
chemoreceptor trigger zone (CTZ) in the central nervous system (CNS) is reason for anti-
emetic action of Metoclopramide. This is the action which is responsible to prevent
nausea and vomiting which are triggered by stimuli [131]. 5-HT3 antagonist activity at
higher doses may also contribute to the anti-emetic effect. The muscarinic activity, D2
receptor antagonist activity and 5-HT4 receptor agonist activity are responsible for pro
kinetic activity of Metoclopramide [132-133]. The lists of brand names of Metoclopramide
which are available in the market are listed in Table 10.1
Physicochemical Properties: Metoclopramide hydrochloride is a white crystalline,
odorless substance. It occurs as a white or almost white, crystalline powder or crystals,
very soluble in water, freely soluble in Alcohol, sparingly soluble in Methylene chloride,
insoluble in Ether.
Figure–10.A: Structure of Metoclopramide
Chapter 10-Metaclopramide Page 219
TABLE 10.1 Lists of Brand Names of Metoclopramide
S. No. Brand Name Formulation Available
Strength(mg or mg/ml)
Manufacturer
1 EMENIL Tablet 10
Astra Zeneca Injection 5
2 MAXERON
Tablet 10
Wallace Syrup 5
Injection 5
3 METAJEX Injection 5 Cadila
4 METOCONTIN CR-Tablet 15 Modi Mundi
5 PERINORM
DT-Tablet 5
IPCA (INTIMA) Tablet 5, 10
Syrup 5
Injection 5
6 PERINORM-CD Cap 15 IPCA (INTIMA)
7 PROMET Injection 5 LUPIN
10.2 LITERATURE SURVEY
Several analytical methods have been reported for the determination of
Metoclopramide Hydrochloride in pure drug, pharmaceutical dosage forms and in
biological samples using Liquid Chromatography and electrochemical detection, either in
single or in combined forms.
Supriya Shidhaye et al [134] have developed and validated an HPLC method for
estimation of Metoclopramide Hydrochloride in a novel pellet formulation. The liquid
chromatographic separation was achieved isocratically using a mobile phase of
Acetonitrile: Water (25:75), with 0.06% Triethyl amine and pH was adjusted to 4 using
Chapter 10-Metaclopramide Page 220
Orthophosphoric acid. The analysis was carried out using Hi-Q-Sil C18 column (250 mm
x 4.6 mm, 5µ) at flow rate of 1 mL/min and the UV detection was made at 274nm.
Lamparczyk H. et al [135] have developed a RP-HPLC method with
electrochemical detection for the determination of Metoclopramide in serum is used in
pharmacokinetic studies. The assay was performed after single extraction with Ethyl
ether using Methyl para hydroxyl benzoate as internal standard. The quantification limit
for Metoclopramide in serum was 2 mg mL-1. Linearity of the method was confirmed in
the range of 5-120 ng mL-1. The analytical method was successfully applied for the
determination of pharmacokinetic parameters after ingestion of 10 mg dose of
Metoclopramide.
Gloria Nygard et al [136] have developed a simple isocratic HPLC method for
the determination of Metoclopramide in plasma and urine. The mobile phase consisted of
0.03M Sodium acetate (pH 7.4) and Acetonitrile. The Glucuronide and Sulfate conjugates
of Metoclopramide were also quantitated after differential acid hydrolysis of urine
samples. These methods have been applied to the analysis of plasma and urine samples
obtained from human volunteers.
N. P. Dudhane et al[137] have developed simple, specific, accurate, precise and
reproducible method has been developed and validated for the simultaneous estimation of
Metoclopramide Hydrochloride and Paracetamol in combined dosage form by RP-HPLC
method.
Mahasen A. Radwan et al [138] have developed a method for determination of
Metoclopramide in serum by HPLC and is applied to a Pharmacokinetic study in rat with
UV detection at 270 nm. The separation was performed on a Novapak C18, 4 µm (3.9
150 mm) column. Acetonitrile (18%) in 0.02 M Ammonium acetate containing 0.1%
Triethyl amine was used as the mobile phase.
Chapter 10-Metaclopramide Page 221
10.2.1 MOTIVATION FOR THE METHOD DEVELOPMENT
The primary purpose of this research project was to develop and to validate a
simple, precise and accurate HPLC method for determination of Metoclopramide in the
APIs, finished product and its applicability in analysis of drug in plasma samples
Metoclopramide is commonly used to treat nausea including that due to chemotherapy
and that occurring post operatively. Metoclopramide is an antiemetic and gastro-
prokinetic agent. Clearly, it is highly important to accurately measure its concentration
alone or in combination with other compounds. A high speed method was sought to
measure the concentration of this compound within a short span of time. This is
beneficial in any pharmaceutical analysis/clinical environment where the concentration of
Metoclopramide is needed to understand any patient issues along with the pharmaceutical
industry to prepare the multiple steps that may be needed to prepare the APIs for
production. The high speed method will eliminate/reduce any waste or costs that are
required with the preparation of the APIs.
There are few analytical methods that have been reported for the determination of
Metoclopramide in formulations at the time of commencement of research work. But
they mainly deal with biological fluids. Supriya Shidhaye et al have developed HPLC
method suitable for the for novel pellet formulation. Lamparczyk H.et al have developed
a method with the electrochemical detection for the analysis of the drug in serum samples
and this method is suitable for the pharmacokinetic studies. Gloria Nygard et al have
developed method which is suitable for the estimation of drug from samples of urine and
plasma from human volunteers. N.P.Ddhane et al have developed a method suitable for
estimating the drug from combination of drugs. Mahasen A.Radwan et al have developed
a HPLC method suitable for analysis of the sample collected from rats and this method
has high run time of 7 mins. This method is suitable only for PK analysis. Such methods
have some setbacks and disadvantages of one or the other like requirement of special
sample treatment and detection which is not suitable for regular pharmaceutical analysis..
Such methods may not be suitable for regular/routine analysis for Metoclopramide in
pharmaceutical industry because of diversity and complexity. The determination of
Chapter 10-Metaclopramide Page 222
Metoclopramide in a APIs sample is yet to be found. In addition, stability-indicating
methods have been able to be found for the Metoclopramide in fixed dosage forms along
with other antiviral drugs. Such simultaneous estimations are not adopted for single
component analysis due to expensive factors. Complete validation parameters were not
able to be found in any of the methods completed in the past. Studying the stability of a
drug and being able to monitor degradation products aids in the clinical treatments/early
product development and shelf life for the drug. Hence, by considering all these factors,
attempts were made, hoping to fill this gap, and succeeded in developing analytical
methods using HPLC methods.
10.3 EXPERIMENTAL
10.3.1 Instrumentation
To develop a liquid chromatographic method for quantitative estimation of
Metoclopramide, an isocratic PEAK HPLC instrument and a Chromosil C18 column (250
mm x 4.6 mm, 5µ) was used. The instrument is equipped with a LC 20AT pump and
variable wavelength programmable UV-Visible detector, SPD-10AVP. A 20µL Hamilton
syringe was used for injecting the samples. Data was analyzed by using PEAK software.
Elico SL159 UV-Visible spectrophotometer was used for spectral studies. Degassing of
the mobile phase was done by using a Loba ultrasonic bath sonicator. A Denwar balance
was used for weighing of the materials.
10.3.2 Chemicals and Solvents
The reference sample of Metoclopramide was obtained from Lupin, Ahmedabad.
The formulation was procured from the local market. Acetonitrile, Methanol and Triethyl
amine used were of HPLC grade and purchased from Merck Specialties Private Limited,
Mumbai, India.
10.3.3 The Buffer Solution
About 10.0 mL of Triethylamine was diluted to 1000 mL with water. This
solution was mixed and filtered through 0.45µm nylon filter.
Chapter 10-Metaclopramide Page 223
10.3.4 The Mobile Phase
A mixture of Acetonitrile and 1% Triethylamine in the ratio of 50:50%, v/v was
prepared and used as mobile phase. They were filtered before use through a 0.45µm
membrane filter and degassed by sonication.
10.3.5 Standard Solution of the Drug
For analysis a 100 ppm standard solution was prepared in mobile phase, and
further required concentrations were obtained from 100 ppm solution by proper dilution.
10.3.6 Sample Solution
Twenty tablets of Metoclopramide (PERINORM-10 mg) were weighed, and then
powdered. A sample of the powdered tablets, equivalent to 10mg of the active ingredient,
was mixed with 30 ml of diluent in 50 ml volumetric flask. The mixture was allowed to
stand for 1 hr with intermittent sonication for complete solubility of the drug, and then
filtered through a 0.45 µm membrane filter, followed by addition of diluent upto 50 ml
which is further diluted to obtain a stock solution of 2.0 mcg/ml and then filtered through
Ultipor membrane sample filter paper.
10.3.7 Calculations in Validation Studies
Percentage recovery and area ratio were calculated using the following equation:
% Recovery = ([Peak Area] sample / [Peak Area] standard) × 100
For a set of “n” replicate measurements, percentage relative standard deviation was
calculated as follow: % RSD=SD/Average × 100
The detector sensitivity was determined by calculating the signal to noise ratio using the
following equation: Sensitivity = S/N=Signal/ Noise
Signal = Amount of detector response to the peak from the middle of the
noise to the summit of the peak.
Chapter 10-Metaclopramide Page 224
Noise = Amount of noise resulting from the detector that is taken from a
portion of the baseline without any distortions.
10.4 METHOD DEVELOPMENT
The goal of this research was to develop a RP-HPLC method that can be used to
determine the active pharmaceutical ingredient in Metoclopramide APIs with an adequate
resolution and in a minimum analysis time. The intention was to separate the degradants
(if any) and process impurities (if present) from the active ingredient. To begin method
development, important structural information such as chemical structure, molecular
weight, UV spectrum and sample solubility were reviewed.
For developing the method [17-23], a systematic study on the effect of various
factors was carried out by varying one parameter at a time and keeping all other
conditions constant i.e, OFAT (one factor at a time) mode of study. Method development
consists of selecting the appropriate detection wave length and stationary and mobile
phases. The following studies were conducted for this purpose.
10.4.1 Wavelength Detection
Selection of proper wavelength was needed to determine maximum detector
response. The first step was to run a UV-VIS spectrum (from 190-400 nm) using an
HPLC system equipped with the Photo Diode Array Detector (PDA). From the spectrum,
it is clear that Metoclopramide absorbs maximum light between 265 nm to 275 nm. The
longer wavelength of 270 nm was selected since it produces less noise, which minimizes
problems that may exhibit around the active ingredient when attempting to quantify
Metoclopramide. The shorter wavelength have a setback of back ground noise where the
mobile phase solvents might be absorbing the light, which is indicated by rough base line
in blank chromatogram.
The UV absorption spectrum of diluted solutions of the Metoclopramide (at 0.5
and 3.5ppm) in Methanol was recorded separately on a UV spectrophotometer. Both the
Chapter 10-Metaclopramide Page 225
spectra of Metoclopramide showed maximum absorbance at 270 nm which is selected as
detection wavelength for further studies.
10.4.2 Selection of Stationary Phase
Preliminary development trials have been performed with octadecyl columns of
different types and configurations from different manufacturers. Review of the literature
found a variety of separation columns are used in the analysis of Metoclopramide.
Because the compound of interest is predominantly polar and of low molecular mass, two
reversed-phase columns of different polarity, a Chromosil C18 column (250 X 4.6 mm,
5µm) column and a Inertsil ODS (150 mm × 4.6 mm i.d., 5-µm particles) were tried.
Under these chromatographic conditions, the analyte of interest has exhibited poor peak
efficiencies (N) and peak symmetries (As) and a partial resolution (Rs) between drug and
mobile phase components. With the same mobile phase, peak symmetries and peak
efficiencies improved on an X-Terra RP18 (250 mm x 4.6 mm, 3.5 µm) column but
selectivity for the component has lost. Finally Chromosil C18 column (250 mm x 4.6
mm, 5µm) was selected based on the peak shape and the baseline separation from the
other interfering peaks in the formulation sample.
10.4.3 Selection of the Mobile Phase
Different mobile phases were tested to optimize analytical performance. In order
to get sharp peak and base line separation of the components, number of experiments by
varying the composition of various solvents and its flow rate was carried. To effect ideal
separation of the drug under isocratic conditions, mixtures of solvents like Methanol,
Water and Acetonitrile with or without different buffers in different combinations were
tested as mobile phase. A mixture of Acetonitrile and 1% Triethylamine in the ratio of
50:50%, v/v was proved to be the most suitable out of all combinations, as the
chromatographic peak obtained was better defined and resolved and almost free from
tailing.
Chapter 10-Metaclopramide Page 226
10.4.4 Flow Rate
Flow rate of the mobile phase was changed from 0.5 – 1.5 mL/min for optimum
separation. A minimum flow rate as well as minimum run time gives the maximum
saving on the usage of solvents. It was found from the experiments that 1.0 mL/min flow
rate was ideal for the successful elution of the analyte.
10.4.5 Optimized Chromatographic Conditions
Chromatographic conditions as optimized in the experiment are shown in Table
10.2. These optimized conditions were followed for determination of Metoclopramide in
APIs samples and in its tablet formulations. The chromatograms of standard, blank, tablet
sample, plasma blank and plasma sample are shown in Graph 10.A, 10.B, 10.C, 10.D and
10.E respectively.
Table 10.2: Optimized Chromatographic Conditions for Estimation of
Metoclopramide
Mobile phase Acetonitrile : 1% Triethylamine
50:50%, v/v
Pump mode Isocratic
Mobile phase pH 6.8
Diluent Mobile phase
Column Chromosil C18 column (250 mm x 4.6
mm, 5µ)
Column Temp Ambient
Wavelength 270 nm
Injection Volume 20 µL
Flow rate 1.0 mL/min
Run time 6 min
Retention Time 2.56 min
Chapter 10-Metaclopramide Page 227
Graph - 10.A: Chromatogram of Standard Solution
Graph - 10.B: Chromatogram of Blank solution
Graph - 10.C: Chromatogram of Tablet Formulation
Chapter 10-Metaclopramide Page 228
Graph - 10.D: Chromatogram of Plasma Sample Blank
Graph - 10.E: Chromatogram of Plasma Sample Solution
10.5 VALIDATION OF THE PROPOSED METHOD
The proposed method was validated [24-32] as per ICH guidelines. The parameters
studied for validation were specificity, linearity, precision, accuracy, robustness, system
suitability, limit of detection, limit of quantification, and solution stability.
10.5.1 Specificity
The specificity of the method was assessed by comparing the chromatograms
obtained from drug standards and from placebo solution prepared from the excipients
most commonly used in pharmaceutical formulations, including starch, lactose
monohydrate, aerosil, hydroxypropylmethyl cellulose, titanium dioxide, and magnesium
Chapter 10-Metaclopramide Page 229
stearate. No excipient peaks coeluted with the analytes and internal standard, indicating
the method is selective and specific in relation to the excipients used in this study. The
specificity results are shown in Table 10.3.
Table 10.3: Specificity study
Name of the solution Retention Time in Min
Blank No peak
Metoclopramide Standard 2.56
Metoclopramide Sample 2.58
10.5.2 Linearity
Linearity was performed by preparing standard solutions of Metoclopramide at
different concentration levels including working concentration mentioned in experimental
condition i.e. 2.0 ppm. Twenty micro liters of each concentration was injected in
duplicate into the HPLC system. The peak response was read at 270 nm and the
corresponding chromatograms were recorded. From these chromatograms, the mean peak
areas were calculated and linearity plots of concentration over the mean peak areas were
constructed individually. The regressions of the plots were computed by least square
regression method. Linearity results were shown in Table10.4 and calibration plot shown
in Graph 10.F
Chapter 10-Metaclopramide Page 230
Table 10.4: Linearity Results
Level Concentration of
Metoclopramide in ppm Mean peak area
Level -1 0.5 78000
Level -2 1.0 137155
Level -3 1.5 203690
Level -4 2.0 261471
Level -5 2.5 312874
Level -6 3.0 377123
Level-7 3.5 435145
Range: 0.5 ppm to 3.5 ppm
Slope
Intercept
Correlation coefficient
118611
20700
0.9993
Graph - 10.F: Calibration Plot for Metoclopramide
Chapter 10-Metaclopramide Page 231
10.5.3 Method Precision
Precision of the assay method was determined by repeatability (intraday) and
intermediate precision (interday) using the triplicate analysis of the samples.
Repeatability shows the applicability of the analytical procedure within a laboratory over
a short period of time that is evaluated by assaying the samples during the same day.
Intermediate precision is assessed by comparing the assays on different days. Precision is
the degree of repeatability of an analytical method under normal operational conditions.
Precision of the method was performed as intraday precision and interday precision
10.5.3.1. Intraday Precision: To study the intraday precision, six replicates of standard
solution of Metoclopramide (at 2ppm) were prepared and injected using the proposed
method conditions. The percent relative standard deviation (% RSD) for peak responses
was calculated and was found to be 0.7%, which is well within the acceptance criteria of
not more than 2.0%. Results of intraday system precision studies are shown in Table
10.5.
Table10.5: Intraday Precision Results for Metoclopramide
Sample Concentration(ppm) Injection no Peak Area RSD(Acceptance
criteria ≤ 2.0%)
Metoclopramide 2
1 262549
0.7
2 266384
3 260994
4 262856
5 264189
6 263005
10.5.3.2. Interday Precision: To study the interday precision, six replicate standard
solutions of Metoclopramide at 2ppm concentration was injected on third day of sample
preparation. The percent relative standard deviation (% RSD) was calculated and it was
Chapter 10-Metaclopramide Page 232
found to be 0.9%, which is well within the acceptance criteria of not more than 2.0%.
Results of interday system precision studies are shown in Table 10.6.
Table 10.6: Interday Precision Results for Metoclopramide
Sample Concentration
(ppm) Injection no Peak Area RSD(Acceptance criteria ≤ 2.0%)
Metoclopramide
2
1 263618
0.9
2 265472
3 269085
4 267634
5 266057
6 262517
The %RSD of Metoclopramide during intraday precision study was within 0.6%
and the %RSD values of interday precision study were within 0.8%. This data indicates
that the repeated preparations of the Metoclopramide samples are highly precise.
10.5.4 Accuracy
The accuracy of the developed method was tested to determine closeness.
Accuracy of the developed method was studied by evaluating the recovery of
Metoclopramide from spiked solutions
The accuracy of the method was determined by standard addition method. A
known amount of standard drug was added to the fixed amount of preanalyzed tablet
solution. Percent recovery was calculated by comparing the area before and after the
addition of the standard drug. The standard addition method was performed at 50%,
100% and 150% levels of 2ppm. The percent recovery and % RSD was calculated and
results are presented in Table 10.7. Satisfactory recoveries ranging from 98.0 to 101.0
were obtained for the proposed method with mean RSD of 1.0% for %recoveries, across
the three levels. This indicates that the proposed method was accurate.
Chapter 10-Metaclopramide Page 233
Table 10.7: Accuracy Results
Level Amount of Metoclopramide (ppm)
% Recovery %RSD Spiked Recovered
50 %
1 0.99 99.0
1.5 1 0.98 98.0
1 1.01 101.0
100%
2 2.02 101.0
1.0 2 1.99 99.5
2 1.98 99.0
150%
3 2.99 99.7
0.5 3 3.02 100.7
3 3.01 100.3
Mean 99.8 1.0
10.5.5 Robustness
The robustness of the method was evaluated by analyzing standards and test
solutions at the method nominal concentration of Metoclopramide (2µg mL-1) in the
presence of all mobile phase and excipients. The parameters altered were, the flow rate
(±10%), 0.8, 1.2 mL min-1 instead of 1.0 mL min-1, pH (± 0.1 pH unit), 7.0, 6.6 instead of
4.8 and the Acetonitrile content in mobile phase composition. The results were
summarized in Table10.8. Evaluation of the results was based mainly on the most
important parameter of the separation, e.g. resolution. The method proved to be robust
since resolution (Rs≥ 1.8) met the acceptance limit (Rs ≥ 1.5) for Metoclopramide.
The robustness study was performed by slight modification in mobile phase flow
rate, pH and its composition. Metoclopramide at 2 ppm concentration was analyzed under
these changed experimental conditions. Three replicate injections were performed with
each of the altered chromatographic condition and the mean peak area was compared
Chapter 10-Metaclopramide Page 234
against the mean peak area obtained with the unaltered conditions. It was observed that
there were no marked changes in chromatography and the %assay when compared with
unaltered conditions was within ± 2%, demonstrating that the developed method was
robust in nature. The results of robustness study are shown in Table 10.8.
Table 10.8: Robustness of Metoclopramide
Condition Mean area % assay % Difference
Unaltered 260514 100.0 -
Flow rate at 0.8 mL/min
Flow rate at 1.2mL/min
263687
259143
101.2
99.5
1.2
-0.5
Mobile phase: ACN : 1% TEA
48% 52%
52% 48%
259036
258317
99.4
99.2
-0.6
-0.8
pH of mobile phase at 7.0 259680 99.7 -0.3
pH of mobile phase at 6.6 261327 99.2 0.3
10.5.6 System Suitability
The system suitability test is very important because it ensures the validity of the
analytical procedure. In the present work, a mixed standard consisted of Metoclopramide
and the sample solution prepared from tablet dosage form was the system suitability
solution used daily for this purpose. Thus, the consistency of the retention time of
Metoclopramide and the area of the peak along with their asymmetry factor (As, the USP
tailing factor), the relative retention times (tR), and the resolution (Rs), were the critical
parameters examined every day. During method validation the critical parameters of Rs
and As met the acceptance criteria (Rs≥ 1.5, As ≤ 2.0) and assured for a good separation
analyte and its reliable quantification.
System suitability was studied under each validation parameter by injecting six
replicates of the standard solutions at 2ppm concentration. For the method following
Chapter 10-Metaclopramide Page 235
limits were considered as acceptance criteria, tailing factor ≤ 2, theoretical plates > 2000
and %RSD for peak area ≤ 2%. The system suitability results are given in Table 10.9.
Table10.9: System Suitability Results for Metoclopramide
Parameter Tailing factor Theoretical plates % RSD for peak response
Specificity study 1.61 14025.6 1.2
Linearity study 1.81 13751.6 0.7
Precision study 1.88 13406.2 0.7
10.5.7 Solution Stability Studies
To perform the solution stability, three replicates of standard and sample solutions
at 2ppm (stability samples) were prepared and stored separately at ambient temperature
(25±10°C) for two days. After the intended storage period, both the standard and sample
stability solutions were compared against a freshly prepared standard solution
(comparison sample) using the proposed method. It is noticed that the % stability of
Metoclopramide was more than 98%, demonstrating insignificant degradation in both
standard and formulation samples. The results of solution stability are shown in Table
10.10.
Table 10.10: Stability Results for Metoclopramide
S. No Concentration
(ppm) Solution Mean Peak Area % Stability
1 2 Fresh standard Solution (Comparison sample)
26554 -
2 2 Stored Standard Solution
(Stability sample) 264284 99.5
3 2 Stored Sample Solution
(Stability sample) 264437 99.6
Chapter 10-Metaclopramide Page 236
10.5.8 Limit of Detection and Limit of Quantification
Limit of detection (LOD) and Limit of quantification (LOQ) were estimated from
the signal- to-noise ratio. The detection limit was determined as the lowest concentration
level resulting in a peak area of three times the baseline noise. The quantitation limit was
determined as the lowest concentration level that provided a peak area with signal-to-
noise ratio 10. To determine the Limit of Detection (LOD) sample was dissolved in
mobile phase and injected until peak was disappeared. After 0.01ppm dilution, peak was
not clearly observed. So it confirms that 0.01ppm is Limit of Detection and Limit of
Quantification was found to be 0.05ppm. For establishing LOQ, six replicates of
standard at 0.05ppm were prepared and quantified with a relative standard deviation of
1.9%. The LOD and Limit of Quantification of Metoclopramide are given in Table 10.11
Table 10.11: Limit of Detection and Limit of Quantification for Metoclopramide
Parameter Measured volume
Limit of Quantification 0.05ppm
Limit of Detection 0.01ppm
10.5.9 Analysis of Pharmaceutical formulations
For the assay of Metoclopramide in tablet formulation (PERINORM - 10 mg), 20
tablets were weighed and their average weight was calculated. The tablets were powdered
and accurately weighed, quantity of powder equivalent to 10mg of Metoclopramide
transferred into a 10mL volumetric flask. 5mL of mobile phase was added and sonicated
to completely dissolve the drug and final volume was made with the same diluent. The
solution was mixed well and filtered the solution through 0.45µm filter. Further pipetted
out 0.1mL of the above solution into a 50mL volumetric flask and diluted upto the mark
with mobile phase to get a final concentration of 2ppm. An aliquot of the prepared
solution was injected into HPLC system and the peak area of Metoclopramide was
measured. The %assay was calculated by applying the molecular weight correction for
Chapter 10-Metaclopramide Page 237
hydrochloride salt. The proposed method was able to estimate Metoclopramide in the
tablet formulation with an accuracy of 98.9%.
10.5.10 Plasma sample analysis
To evaluate the general applicability of the HPLC method, Metoclopramide was
analyzed in different sample matrices. The proposed HPLC procedure was further tested
using spiked human plasma samples. In all cases, no interferences from the indigenous
plasma components were observed. The simplicity of the proposed extraction procedure
and the high extraction efficiency are among the essential features of the proposed HPLC
method, making this method suitable for routine, efficient and fast extraction of
Metoclopramide from complex matrices. The method is accurate, precise and specific.
The described HPLC assay can be easily applied for the quantification of the degradation
products.
Human fresh frozen plasma containing K2 EDTA as anticoagulant was procured
from M/S. Lakshmi sai clinical labs and required concentration of Metoclopramide
spiked samples (in plasma) were prepared using working solutions.
Preparation of Plasma Spiked Samples: Fresh frozen plasma stored at -20°C was
thawed and used for the preparation of Metoclopramide spiked samples. 0.2mL of
100ppm working solution was pipetted out into a 10mL volumetric flask and further
diluted up to the mark with plasma, to get 2ppm concentration sample. During the spiking
of sample volume of diluents was kept ≤ 5% of the total plasma volume, to avoid
unnecessary changes in matrix components.
Sample Extraction / Preparation: 0.5mL of plasma spiked sample was taken into a test
tube and added 100µL of 0.1M sodium carbonate solution and vortexed. Further to the
sample, 5mL of Eethylacetate was added and subjected to extraction on a plasma
extractor, using 50rpm for 20 min. Later centrifuged the sample for 10min at 3000 rpm,
collected 4mL of the separated organic layer and evaporated to dryness at 50°C. The
residue was reconstituted with 200µL of mobile phase and injected into HPLC system
with the developed chromatographic conditions.
Chapter 10-Metaclopramide Page 238
10.6 RESULTS AND DISCUSSIONS
A RP-HPLC method in isocratic mode was developed and validated for the
determination of Metoclopramide in APIs and pharmaceutical dosage forms. The present
study is the first report on assay of Metoclopramide in pharmaceutical dosage forms by
HPLC. In this method isocratic elution method is selected for the analysis of
Metoclopramide API because it gave better base line separation and peak width, which is
suitable for the routine analysis of Metoclopramide. The developed method was validated
as per ICH guidelines (ICH, 1996) and its updated international convention (ICH, 2002).
The presently described method appears to be alternative to existing methods in
terms of good linearity, and broad range of concentration in the lower concentration
levels. It also provides the good system suitability parameters and recovery studies. The
retention was short and hence the present method is suitable of rapid analysis. During
method development, it was observed that it has high selectivity and no excipient/solvent
interferences.
To optimize the RP-HPLC parameters, several mobile phase compositions were
tried. A satisfactory separation and good peak symmetry was found in a mixture of
Acetonitrile: 1% Triethylamine in the ratio of 50:50%, v/v at a flow rate of 1.0 mL/min.
The optimum wavelength for detection was set at 270nm at which much better detector
response for drug was obtained. As it was shown in Graph 10.A, the retention time was
2.56 min for Metoclopramide and no interferences was observed in formulation samples.
The number of theoretical plates was found to be > 13000 which indicates the efficient
performance of column. System suitability test was performed with each of the validation
parameter and has comfortably met the set acceptance criteria. The results obtained were
represented in Table 10.9.
The calibration curve was obtained with seven standard points in the range of 0.5-
3.5ppm was found to be linear. The representative graph of calibration curve was shown
in Graph 10.F and the regression data was presented in Table 10.4. Calibration curve was
found to be linear with correlation coefficient of r=0.9993, and the intercept and slope
Chapter 10-Metaclopramide Page 239
values were found to be 20700 and 118611 respectively. In all cases, the relative standard
deviation (%RSD) for the analytic peak area for two consecutive injections was < 2.0%.
At each level of analyte, the relationship between the peak area and concentration of the
Metoclopramide was highly linear over the entire examined concentration range. The
correlation coefficients of the calibration curves being greater than r = 0.995. The relative
standard deviations (%RSD) and the standard errors of the slope, as well as of the
intercept of Metoclopramide confirm the excellent linearity.
Precision was evaluated by carrying out six independent sample preparations of
standard at 2ppm concentration. Low values of standard deviation denoted very good
repeatability of the measurement. Repeatability or intraday precision was investigated by
injecting nine replicate samples of each of the samples of three different concentrations.
Interday precision were assessed by injecting the same three samples over three
consecutive days. Thus it was showing that the equipment used for the study was correct
and hence the developed analytical method is highly reproducible. %RSD of intraday
precision was found to 0.7%. For the interday precision a study carried out on
consecutive days resulted in a %RSD of 0.9%. The degree of reproducibility of the
results obtained as a result of small deliberate variations in the method parameters and by
changing analytical operators has proven that the method is robust. This indicates good
method precision and the results are shown in Table 10.5 and 10.6.
Recovery from tablet dosage form, using the extraction procedures, was assessed
by comparing the peak area obtained from multiple analyses (n=3) of spiked samples
with the peak area from standard solution of the analyte in a mobile phase at the same
concentrations, as described by international guidelines. Standard addition method at
50%, 100% and 150% of 2ppm, showed good recoveries ranging from 98.0 to 101.0%.
The mean recovery data obtained at each level as well as on overall three levels (Table
10.7) was within 2.0%, which satisfied the acceptance criteria set for the study.
The stability studies of Metoclopramide was evaluated by comparing the results
of freshly prepared standard solution (2ppm) against the standard and sample solutions
(stability samples) after two successive days of storage at ambient temperature
Chapter 10-Metaclopramide Page 240
(25±10°C). Metoclopramide was found stable under the tested conditions with
insignificant degradation.
The proposed method has been applied to the assay of commercial tablets
(PERINORM - 10 mg) containing Metoclopramide monohydrochloride. Three
independent samples were analyzed after extracting the drug, as mentioned in assay
sample preparation of the experimental section. The results presented good agreement
with the labeled content. The proposed method was able to estimate Metoclopramide
with an accuracy of 98.9% in tablet formulation.
The statistical evaluation of the proposed method revealed its good linearity,
reproducibility and its validation for different parameters made us to conclude that it
could be used for rapid and reliable determination of Metoclopramide in tablet
formulation as well as in APIs samples. The current method can be successfully applied
in the quality control of tablet formulations, with good accuracy and precise results.
The developed method was applied to analysis of plasma spiked samples of
Metoclopramide, which showed relatively good recovery of analyte from plasma
components, however in the chromatography merging peaks were observed before and
after the analyte peak which can contribute significant variations at low concentration
levels. Extraction of plasma spiked samples was carried out using liquid-liquid extraction
with ethylacetate under basic condition as the molecule is non-polar (log P- 1.8) and the
applied extraction procedure was considered efficient based on the peak response at
analyte tR.
With the data obtained from analysis of plasma samples, it is concluded that the
current method can be applied to the analysis of biological samples for therapeutic drug
monitoring of Metoclopramide which involves high concentrations, however the method
chromatographic conditions need to be reoptimized for application to Pharmacokinetic
studies.