6.5 oscillating coriolis densitometer (gas, liquid, … oscillating coriolis densitometer (gas,...

5
831 6.5 Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services) C. H. HOEPPNER (1982) C. APPLE (1995) J. E. JAMISON (2003) Density Range: 0.0 to 5.0 g/cm 3 (0 to 5000 kg/m 3 ); can also measure mass flow, volumetric flow, and temperature Sizes (Diameter): 1/25 to 10 in. (1 to 250 mm) Flow Range: 0 to 63,000 lb/min (0 to 28,300 kg/min) Fluids: Liquids, slurries, compressed gases, liquified gases (not suited for gas-liquid mixtures) Output Signal: Linear frequency, analog, digital serial protocols, display, Highway Addressable Remote Transducer (HART), Profibus, Foundation Fieldbus, Modbus, scaled pulse, display, alarm outputs, manufacturer specified protocols. Detector Types: Electromagnetic, optical, and capacitive Operating Pressure: Depends upon tube size and flange rating; 1400 PSIG (100 bars) is the typical standard; 5000 PSIG (345 bars) is a typical high pressure. At 650 ° F temperature, the maximum operating pressure is 1160 PSIG (80 bars). Required Pressure Drop: From under 10 PSIG (0.7 bar) to over 100 PSIG (6.9 bars) Operating Temperature: Depends upon design: 400 to 400 ° F ( 240 to 204 ° C) is the standard rating; 32 to 800 ° F (0 to 426 ° C) is the rating of high temperature units. Special low temperature designs can also be used on cryogenic applications from 40 to 140 ° F ( 40 to 60 ° C). Temperature ratings can be affected by hazardous area classification. Materials of Construction: 304/316L/316 Stainless steels, high alloy Ferritic Stainless 17-7PH, Hastelloy C, titanium, Ni-Span C as standard; tantalum, zirconium, Tefzel-lined as special Inaccuracy: ± 0.0002 to 0.05 g/cc, depending upon design; typical is ± 0.002 g/cm 3 over a 0.3- to 3.0-g/cm 3 range Repeatability: ± 0.0001 to 0.01 g/cm 3 , depending upon design; typical is ± 0.001 g/cm 3 over a 0.3- to 3.0-g/cm 3 range. Digital Communications: Most manufacturers provide electronics, which are equipped with digital communi- cations that support RS232/RS422 or RS485 physical interfaces and/or protocols such as ModBus. Other electronic transmitters support a variety of other digital commu- nications protocols such as HART as well as FOUNDATION Fieldbus, DeviceNet, and Profibus-PA/DP. Costs: 1/25th in. (1.0 mm), $5000; typical 1 in. (25 mm) meter, with full-scale flow rate of 400 to 1000 lb/min (180 to 450 kg/min), $7000; typical 3 in. (75 mm) meter, with full-scale flow rate and FOUNDATION Fieldbus transmitter, with full-scale flow of 9800 to 39,500 lb/h (4500 to 18,000 kg/h), $13,500; 6 in. (150 mm), $27,500. To Receivers Flow Sheet Symbol FIT FE DE TE DIT TIT © 2003 by Béla Lipták

Upload: dangtram

Post on 19-Mar-2018

225 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: 6.5 Oscillating Coriolis Densitometer (Gas, Liquid, … Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services) 833 even though the mass of the system has not changed

831

6.5 Oscillating Coriolis Densitometer(Gas, Liquid, and Slurry Services)

C. H. HOEPPNER

(1982)

C. APPLE

(1995)

J. E. JAMISON

(2003)

Density Range:

0.0 to 5.0 g/cm

3

(0 to 5000 kg/m

3

); can also measure mass flow, volumetric flow, andtemperature

Sizes (Diameter):

1/25 to 10 in. (1 to 250 mm)

Flow Range:

0 to 63,000 lb/min (0 to 28,300 kg/min)

Fluids:

Liquids, slurries, compressed gases, liquified gases (not suited for gas-liquid mixtures)

Output Signal:

Linear frequency, analog, digital serial protocols, display, Highway AddressableRemote Transducer (HART), Profibus, Foundation Fieldbus, Modbus, scaled pulse,display, alarm outputs, manufacturer specified protocols.

Detector Types:

Electromagnetic, optical, and capacitive

Operating Pressure:

Depends upon tube size and flange rating; 1400 PSIG (100 bars) is the typicalstandard; 5000 PSIG (345 bars) is a typical high pressure. At 650

°

F temperature, themaximum operating pressure is 1160 PSIG (80 bars).

Required Pressure Drop:

From under 10 PSIG (0.7 bar) to over 100 PSIG (6.9 bars)

Operating Temperature:

Depends upon design:

400 to 400

°

F (

240 to 204

°

C) is the standard rating; 32 to800

°

F (0 to 426

°

C) is the rating of high temperature units. Special low temperaturedesigns can also be used on cryogenic applications from

40 to 140

°

F (

40 to 60

°

C).Temperature ratings can be affected by hazardous area classification.

Materials of Construction:

304/316L/316 Stainless steels, high alloy Ferritic Stainless 17-7PH, Hastelloy C,titanium, Ni-Span C as standard; tantalum, zirconium, Tefzel-lined as special

Inaccuracy:

±

0.0002 to 0.05 g/cc, depending upon design; typical is

±

0.002 g/cm

3

over a 0.3- to3.0-g/cm

3

range

Repeatability:

±

0.0001 to 0.01 g/cm

3

, depending upon design; typical is

±

0.001 g/cm

3

over a 0.3-to 3.0-g/cm

3

range.

Digital Communications:

Most manufacturers provide electronics, which are equipped with digital communi-cations that support RS232/RS422 or RS485 physical interfaces and/or protocols suchas ModBus. Other electronic transmitters support a variety of other digital commu-nications protocols such as HART as well as FOUNDATION Fieldbus, DeviceNet,and Profibus-PA/DP.

Costs:

1/25th in. (1.0 mm), $5000; typical 1 in. (25 mm) meter, with full-scale flow rate of400 to 1000 lb/min (180 to 450 kg/min), $7000; typical 3 in. (75 mm) meter, withfull-scale flow rate and FOUNDATION Fieldbus transmitter, with full-scale flow of9800 to 39,500 lb/h (4500 to 18,000 kg/h), $13,500; 6 in. (150 mm), $27,500.

To Receivers

Flow Sheet Symbol

FIT

FEDE TE

DIT TIT

© 2003 by Béla Lipták

Page 2: 6.5 Oscillating Coriolis Densitometer (Gas, Liquid, … Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services) 833 even though the mass of the system has not changed

832

Density Measurement

Partial List of Suppliers:

ABB Inc. (www.abb.com)

Brooks Instrument, a Unit of Emerson Process Management(www.emersonprocess.com/brooks)

Danfoss A/S (Denmark) (www.danfoss.com)Endress & Hauser Inc. (www.us.endress.com)Fischer and Porter, a Unit of ABB (www.us.abb.com)Foxboro-Invensys (www.foxboro.com)Heinrichs GmbH & Co. (www.heinrichs-mt.com)Krohne (www.krohne.com)Micro Motion Inc., a Unit of Emerson Process Management(www.emersonprocess.com/micromotion)

Oval (www.oval.co.jp)Rheonik (www.rheonik.de)Schlumberger Industries (www.slb.com)Smith Meter Inc. (www.fmcenergysystems.com)Yokogawa (www.yokogawa.com)

INTRODUCTION

It requires the least amount of energy to cause an object to gointo natural frequency vibration. This frequency is a functionof the mass of the object. In case of a Coriolis densitometer,the mass of the vibrating object is the sum of the pipe, the fluidwithin the pipe, and the mass of the driver and sensors(Figure 6.5a). Therefore, since the masses of the pipe, driver,and sensors are constant, a change in natural frequency is adirect indication of a change in the density of the process fluid.

A unique feature of the Coriolis flowmeter is its ability todetermine, in addition to its flow rate, the density of the pro-cess fluid. This ability can be used even when the measurementof the fluid flow rate is of no interest. The density measurementrequires the main components illustrated in Figure 6.5a.

THE SENSOR

One design of a Coriolis densitometer is shown inFigure 6.5b. This density sensor is comprised of either oneor two flow tubes (usually two), a driving mechanism for

oscillating the flow tubes, detectors for the measurement ofnatural frequency, and a temperature-measuring device.

The process fluid stream is split into two and flowsthrough the flow tubes. A coil-and-magnet driver is used tooscillate the flow tubes in opposition at their natural frequency.The oscillation of the tubes causes the flow detectors tooutput a sinusoidal voltage signal that reflects this motion.The frequency of the sinusoidal voltage from the flowdetectors represents the natural frequency of the tube vibra-tion. Typically, the flow detectors are electromagnetic sen-sors, although capacity detectors and optical sensors arealso used.

Changes in the density of the process fluid will cause themass of the fluid-filled flow tube system to change. This inturn will change the natural frequency of the sensor. Thischange will cause the frequency of the sinusoidal voltage fromthe flow detectors to change. By measuring the frequency ofthe sinusoidal voltage developed by the Coriolis detector, thedensity of the process fluid is determined.

Temperature Correction

The temperature of the tube affects its natural frequency ofvibration. If the tube temperature increases, the tube becomesmore elastic, which causes the natural frequency to decrease

FIG. 6.5a

The components and the tube assembly of a Coriolis meter.

Flow Splitters

Driver

Sensors

FIG. 6.5b

Design of a typical Coriolis density meter.

RTD Driver

FlowDetector

Flow Tube

© 2003 by Béla Lipták

Bopp & Reuther (www.burhm.com)

Page 3: 6.5 Oscillating Coriolis Densitometer (Gas, Liquid, … Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services) 833 even though the mass of the system has not changed

6.5 Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services)

833

even though the mass of the system has not changed. Themeasurement of the tube’s temperature is used to correct thenatural frequency for changes in the elastic modulus of the flowtube material. A resistance temperature detector is typicallyused to measure the tube temperature.

THEORY OF OPERATION

The natural frequency of vibration is described by the equation:

6.5(1)

where

ω

n

=

natural frequency, rad/sk

=

spring constant, pound-force/in. or lb/s

2

(N/m or kg/s

2

)m

=

mass of the system, lb (kg)

For the Coriolis meter, the mass of the system is thecombined mass of the fluid and the flow tube assembly andcan be expressed by the following relationship:

6.5(2)

where

ρ

f

=

fluid density, lb/in.

3

(g/cc)

ρ

t

=

tube material density, lb/in.

3

(g/cc)A

f

=

tube internal area, in.

2

(cm

2

)A

t

=

tube cross-sectional area, in.

2

(cm

2

)l

t

=

tube length, in. (cm)

The spring constant (k), from Equation 6.5(1), is a func-tion of geometry and material properties and is determinedfrom the following equation:

6.5(3)

whereM

=

modal constantE

=

modulus of elasticity, pound-force/in.

2

(kPa)I

=

moment of inertia, in.

4

(cm

4

)

The natural frequency (

ω

n

), from Equation 6.5(1), canalso be expressed by the following equivalent relationships:

6.5(4)

wheref

=

oscillation frequency, cycles/sT

=

tube period, s (time for one cycle of oscillation)

Substituting Equations 6.5(2), 6.5(3), and 6.5(4) intoEquation 6.5(1) and solving for

ρ

f

gives the followingrelationships:

6.5(5)

The Calibration Constant

The variables that depend upon the tube geometry and mate-rials properties can be combined to obtain the calibrationconstants, such as: K

1

=

(MI)

/

and

/

A

f

.A correction factor C

T

is used to compensate for changes inthe material modulus of elasticity (E) with temperature. Equa-tion 6.5(5) can then be rewritten as:

6.5(6)

Values for the calibration constants are determined bymeasuring the period of oscillation of the tube at two knownfluid densities. With the two fluid densities and their respec-tive tube periods, two simultaneous equations with twounknowns can be solved to obtain K

1

and K

2

. Any two fluidsof known density can be used to determine the calibrationconstant.

Calibration fluids should be so selected as to have suffi-ciently different densities to minimize the amount of errorassociated with the calibration constants K

1

and K

2

. Air andwater are commonly used as the calibration fluids. The valueof C

T

depends upon the material of the flow tube and isdetermined by the manufacturer from experimental data.

ELECTRONICS

A typical block diagram of the density measurement elec-tronics is shown in Figure 6.5c. The electronics is comprisedof interfaces to the sensor, signal-processing components,and outputs to external devices. The interfaces between thesensor and the electronics include the flow detectors, the drivemechanism, and the temperature detector.

The electronics vibrates the tubes at their natural fre-quency by minimizing the amount of energy input into thedriver while maintaining sufficient tube displacement toproduce the forces required to measure the fluid flow rate.The electronics applies an alternating current to the drivecoil that is mounted on one of the flow tubes. This alternatingcurrent generates an alternating magnetic field in the coil.The alternating magnetic field causes the fixed magnetmounted on the other tube to be repelled and attracted, forc-ing the tubes alternately away from and toward one anotherat their natural frequency. The flow detectors generate asinusoidal voltage signal that reflects the oscillation of theflow tubes.

ω n

km

=

m A l A lf f t t t t= +ρ ρ

kMEI

It

= 3

ω π πn f

T= =2

2

ρπ

ρf

t f

t t

f

MEIl A

TA

A=

4 2 42

4 2 4π I At f K At t2 = ( )ρ

ρf TK C T K= −12

2

© 2003 by Béla Lipták

Page 4: 6.5 Oscillating Coriolis Densitometer (Gas, Liquid, … Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services) 833 even though the mass of the system has not changed

834

Density Measurement

The sinusoidal voltage signal from one of the flow detec-tors is input into a counter that detects the start and end ofeach tube cycle. The tube cycles are gated by the counter,which triggers the counter to make a time measurement overthe duration of the tube cycle. The time or period over whichthe tube cycle occurred is obtained from a precision crystaloscillator. The microprocessor reads the counter time mea-surement and uses this value along with the tube temperatureand calibration constants, K

1

and K

2

, to calculate the fluiddensity using Equation 6.5(6).

Output Options

Electronics are available to provide several different outputoptions. The most common output is an analog (4 to 20 mA)output, which can be scaled to represent the desired densityrange. Some manufacturers provide a scalable frequency out-put that can also be used to represent the fluid density. Manyof the meters are equipped with built-in displays that can beused to directly display the density of the process fluid (orany other process variable acquired by the meter includingvolumetric flow, mass flow, temperature, and total flow). Theuser can select a variety of units for the density display, suchas g/cc, lb/ft

3

,

°

API,

°

Baumé, etc.Most manufacturers provide electronics that are equipped

with digital communications that support RS232/RS422 orRS485 physical interfaces and/or protocols such as ModBus.Other electronic transmitters support a variety of other digitalcommunications protocols, such as Highway AddressableRemote Transducer (HART), as well as FOUNDATION

Fieldbus, DeviceNet, and Profibus-PA/DP. The latter two canprovide high-speed transmission of all sensor process vari-ables back to the host distributed control system (DCS) fordisplay and/or control purposes.

A density output is occasionally optional on somedevices. A typical Coriolis flowmeter comes standard withone pulse or frequency output that usually represents flowrate and another output that can represent density. Two analogoutputs can be configurable for a choice of flow rate, density,or temperature; a local display can be purchased which isindependent of the other outputs. Digital discrete outputs arealso normally standard issue including alarm outputs. Severaldata communications options are available including basicRS-232/422/485 physical links, ModBus, and HART. Otheroptional multivariable transmitters can provide fieldbus serialdigital data communication outputs such as FOUNDATIONFieldbus, DeviceNet, and Profibus-PA/DP, and can sendmany process variables (flow rate [mass or volumetric], den-sity, temperature, and flow total) back to a host DCS. Thenumber and type of outputs, both standard and optional, varyfrom one manufacturer to another.

APPLICATIONS

A Coriolis meter can measure the density and the flow rateof a wide variety of process fluids. In addition to indicatingdensity, the electronics can be configured to provide addi-tional information such as the concentration of one or morecomponents in a mixture.

FIG. 6.5c

Signal processing block diagram.

Sensor Electronics

Drive Coil

FlowDetector

Flow TubeTemperature

Interfaces Signal Processing

DriveControl

PrecisionCrystal

Oscillator

SignalAmplifier

Microprocessor

Counter

DensityComputation

Outputs

Analog(4-20 mA)

Frequency

TemperatureAmplifier

Analog/DigitalConverter Alarm

Contacts

RS232RS422RS485

ρf = K1CTT2-K2

© 2003 by Béla Lipták

Page 5: 6.5 Oscillating Coriolis Densitometer (Gas, Liquid, … Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services) 833 even though the mass of the system has not changed

6.5 Oscillating Coriolis Densitometer (Gas, Liquid, and Slurry Services)

835

Microprocessors can be programmed to determine thecomposition of mixtures. Common options that are providedallow for the determination and monitoring of % solids bymass or volume,

°

Brix, % fructose, % alcohol, and % solidsin black liquor. Many of these relationships are temperature-dependent, and the temperature reading from the flow tubeis used to correct for the variations in fluid temperature.

Since the flow rate is measured by the Coriolis meter, thenet flow rate of one or more components in a multicomponentmixture can also be determined. An example of this is the abilityto determine the flow rate of oil in an oil-water emulsion.

DEVELOPMENTS AND TRENDS

In addition to monitoring several variables, the newer systemsallow the transfer of calibration and diagnostic related infor-mation to and from the host DCS. Some Coriolis meters notonly support the use of many different density units, such as

°

Baume,

°

Balling,

°

Plato, but are also able to convert theirreadings into specialized units as required by the particularindustry. For example, in the hydraulic well fracturing pro-cess of the oil and gas industry, the sand added per cubicmeter of liquid can be monitored.

Two- and three-dimensional polynomial calculations canyield relationships between concentration and reference densityor can correlate relationships between concentration, density,and temperature.

The Coriolis electronics can support up to four detectorassemblies and can provide high precision under widely vari-able operating conditions. For example, it can detect the den-sity within an error of

±

0.0005 g/cm

3

over a range of 0.001 to1.8 g/cm

3

, while the temperature varies from 5 to 80

°

C. The Coriolis meter is a multivariable sensor. It is hoped

that in the future it might become capable of measuring evenmore variables, possibly including viscosity.

Bibliography

Babb, M., “New Coriolis Meter Cuts Pressure Drop in Half,” Control Engi-neering, October 1991.

Babb, M., “New Mass Flowmeter Design Claims Improved Reliability,”Control Engineering, May 1990.

Benabdelkarim, M. and Galiana, C., “Non-Radioactive Densitometer forContinuous Monitoring of Cement Mixing Process,” in Proceedingsof the First International Conference on Health, Safety, and Environ-ment in Oil and Gas Exploration & Production, Part 1of 2, Richardson,TX: Society of Petroleum Engineers, November 1991.

Berkowitz, P.N. and Hardin, M.B., “Mass Flow and Density Sensor TightensControl of Alumina Extraction,” Instruments and Control Systems,December 1991.

Blickley, G.J., “Mass Flow Measurement Aided by Coriolis Methods,” Con-trol Engineering, April 1991.

Blickley, G.J., “Tank Gauging Transmitter Performs More Functions,” Con-trol Engineering, August 1991.

Blumenthal, J., “Direct Mass Flow Rate and/or Density Monitoring usingCoriolis/Gyroscopic Sensor Base,” in TAPPI Proceedings: 1985Papermakers Conference, 1985.

Blumenthal, J., “Simultaneous Direct Mass Flow and Density Monitoring,”Annual Meeting: Technical Section, Canadian Pulp and Paper Asso-ciation, 72nd Annual Meeting CPPA, Montreal, 1986.

Bugher, G., “Coriolis Flowmeters,” Measurements and Control, September1990.

Capano, D., “The Ways and Means of Density,” InTech, November 2000.Corser, G.A. and Hammond, G.C., “A Combined Effects Meter,” InTech,

April, 1993.Cheesewright, R., Clark, C., and Bisset, D., The Identification of External

Factors Which Influence the Calibration of Coriolis Mass Flowmeters,”Flow Measurement Instruments, Vol. 11, 2000, pp. 1–10.

Eibl, C., “Is Mass Flow Measurement Increasing Too Rapidly?” Measure-ments and Control, September 1990.

Eide, J.M. and Gwaspari, S.C., Comparison test and calibration of Coriolismeters, North Sea Flow Measurement Workshop, Peebles, Scotland,1996.

“Flowmeter Measures Mass of Fluid Motion,” Chemical Processing,November 1982.

Ginesi, D. and Annarummo, C. Application and installation guidelines forvolumetric and mass flowmeters, ISA Transactions, Vol. 33, No. 1,1994, pp. 61–72.

Ginesi, D., “Flow: A Performance Review,” Chemical Engineering, June 22,1987.

Gordon, I., “Vibrating Element Technology for Measuring Liquid Densityin Process Applications,” ACHEMA, Frankfurt 2000.

Gupta, S.V., “Practical Density Measurement and Hydrometry,” Bristol,U.K.: Institute of Physics, 2002.

Hall, J., “Coriolis Mass Flowmeter Reduces Pressure Loss,” Instrumentationand Control Systems, October 1989.

Hall, R., “Measuring Mass Flow and Density with Coriolis Meters,” InTech,Vol. 37, No. 4, April 1990.

Heinonen, E.W. and Tapscott, R.E., “Secondary Properties of Aqueous Lith-ium Bromide Solutions,” ASHRAE Transactions, Vol. 105 (Part 1)1999.

Kalotay, P., “Density and Viscosity Monitoring Systems Using Coriolis FlowMeters,” ISA Transactions, Vol. 38 No. 4, 1999.

Keita, N.M., Behaviour of Straight Type Coriolis Mass Flowmeters in theMetering of Gas: Theoretical Predictions with Experimental Verifica-tions, Flow Measurement Instruments, Vol. 5, 1994, pp. 289–294.

Kolahi, K., Gast, T.H., and Rock, H., Coriolis Mass Flow Measurements ofGas Under Normal Conditions, Flow Measurement Instruments,Vol. 5, 1994, pp. 275–283.

Kotnik, P., et al., “Accurate Fluid Density Measurement Using OscillatingType Density Meters Under Rapid Temperature Variations,” ACHEMA,Frankfurt 2000.

“Mass Flow Control Ups Polyurethane Production,” Instruments and ControlSystems, December 1981.

Menke, D., Use of Coriolis Mass Flowmeters in Custody Transfer, FLOM-EKO ’96 8th Int. Conf. On Flow Measurement, Beijing, China, 1996,pp. 232–237.

Plache, K., “Coriolis/Gyroscopic Flow Meter,” Mechanical Engineering,March 1979.

Robinson, C., “Obstructionless Flowmeters,” InTech, December 1986.Schietinger, M., “Mass Flow vs. Volumetric Flow,” Measurements and Con-

trol, September 1990.“Standard Practice for Calibration of Transmission Densitometers,” ASTM

Standard, 1998.Valentine, J. and Duffill, R., “Coriolis Meters Used To Measure Density-

Related Physical Properties,” in Annual ISA Analysis Division Sympo-sium-Proceedings, Vol. 33, Research Triangle Park, NC: Instrumenta-tion, Systems, and Automation Society, 2000.

© 2003 by Béla Lipták