What are the dose quantities in CT ?

IAEA/RCA Kampala

Kampala

Page 2: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

12345678

12

3

45

6

7

8

CT Principles

Kampala

Page 3: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Kampala

Page 4: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Reminder : CT terminology • Axial scanning

– 1 rotation of beam followed by stepwise couch movement

– Increment – distance moved by couch• Helical/spiral scanning

– Continuous table movement & beam rotation– Pitch – table movement relative to nominal

slice width

Kampala

Page 5: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CT Slice

Kampala

Page 6: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CT Slice

Kampala

Page 7: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CT Slice

Kampala

Page 8: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Dose Profile

Kampala

Page 9: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Dose profile for a series of slices

Kampala

Page 10: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Dose Profiles in Phantoms / PatientsEffects of scatter

Broadened profilePeak kerma dependent on beam widthLong scatter tails for wide beams

Kampala

Page 11: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Which Dosemeter ?

Kampala

Page 12: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Integrated Profile

Kampala

Page 13: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Integrated Profile

Kampala

Page 14: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDose Index

• Measure of dose for a single axial slice (mGy)– no account of Helical/MDCT gaps or overlaps

• CTDI = area under dose profile (mGy.mm) nominal slice width, T (mm)

Kampala

kerma

T

CTDI(mGy

)

Z-axis (mm)

CTDI = height of rectangle if area

under profile was all contained

within nominal slice width

Page 15: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDI100 – What is it ?• Analagous to

Multiple Scan Average Dose.

• Originally defined for single slice scanners.

Kampala

Page 16: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

• In air : CTDI100,Air

• In air : CTDI100,Air

• Sometimes normalise to 1 mAs:

• nCTDI100,Air

• Just like in radiography!

Kampala

Page 17: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

In Perspex / PMMA

Kampala

Page 18: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CT dosimetry in PerspexThe computed tomography air kerma index is also measured inside PMMA head and body phantoms

The quantity CTDI100,c is measured at the centre of the standard CT dosimetry phantom and CTDI100,p is the average of values measured at four positions around the periphery of the same phantom.

Kampala

Page 19: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDI100

• CTDI can be measured in air or in phantoms

• CTDI in air can be used as a measure of tube output

• For dosimetry purposes CTDI is measured in perspex phantoms• 16cm diameter ‘head’• 32cm diameter ‘body’

Kampala

Page 20: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Patient Size.• CTDI is a measure of scanner output.• CTDI does take patient size into account• CTDI is NOT patient dose• Important for paediatrics, small adults

and possibly bariatrics.• You can correct to patient size

– To be mentioned later.

Kampala

Page 21: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Phantoms for CT dosimetry

• Cylindrical PMMA phantoms with holes for pencil chamber

• 32 cm body phantom• 16 cm head phantom

Kampala

Page 22: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Centre Periphery – 1cm depth (mean of 4 positions)

C

P1

P2

P3

P4

32 cm

‘Head’ ‘Body’

16 cm

140 mm

CTDI Phantoms

Kampala

Page 23: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

40

‘Head’

‘Body’

40

40 40

40

20

20 20

20

10

Periphery : centre 1:1 Periphery : centre 2:1

Typical CTDI100 ValuesCTDI100,p values quoted in mGy

Kampala

Page 24: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDIW• Weighted average CTDI: represents the

average dose for contiguous irradiation • CTDIw = 1/3 CTDIcentre + 2/3 CTDIperiphery • nCTDIw = CTDIw/mAs

CTDIC CTDIP

Kampala

Page 25: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDIvol

Pitch = 1 CTDIvol = CTDIw

Pitch = 2 CTDIvol = CTDIw

2

Pitch = 0.5 CTDIvol = 2 x CTDIw

Kampala

Page 26: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDIvol

• CTDIW is the average dose in the scan plane

• Volume CTDI considers non-contiguous scanning

• The CTDIvol can be considered to be the average CTDIW in the scan volume

• CTDIvol = CTDIW Pitch

Kampala

Page 27: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Dose Length Product

• CTDIvol is a measure of absorbed dose, energy absorbed per unit mass

• Radiation risk is based on total energy absorbed

• Estimate using dose length product (DLP)• DLP = CTDIvol x scan length• Units are mGy cm

Kampala

Page 28: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Dose Quantities in CT

Kampala

Page 29: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Dose Summary Sheet - Commercial CT Scanner

Huda W , Mettler F A Radiology 2011;258:236-242

Kampala

Page 30: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDI on the Console

• The CTDI value displayed will depend on the phantom that it is based on.

• Body : 32 cm; Head 16 cm.• CTDI for 16cm reference twice that of

CTDI for 32cm reference.• Paediatric protocol MAY reference to 16

cm phantom• Implementation manufacturer

dependent.Kampala

Page 31: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

CTDI MEASUREMENT IN PRACTICE

Kampala

Page 32: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Equipment for CT measurements

• CT ionization chamber and electrometer calibrated for CT beam qualities

• Chamber support stand• CT head & body phantoms

Kampala

Page 33: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Methodology for CT measurements in air

• Position chamber in stand so beyond couch

• Adjust couch position so chamber along axis of scanner and beam centred at central point of chamber

• Record dose measurements for single axial scan at selected parameters (no couch movement)

• Record all data

Kampala

Page 34: What are the dose quantities in CT ? IAEA/RCA Kampala Kampala

Methodology for CT measurements in phantoms

• Position phantom on couch (use head rest if appropriate)

• Position chamber in central position in phantom• Adjust couch position so chamber centred as

before• Record dose measurements for single axial

scan at selected parameters (no couch movement)

• Repeat for peripheral positions (north, south, east & west)

Kampala