value analysis brief - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/llnwmb8/int...
TRANSCRIPT
VALUE ANALYSIS BRIEF
Executive Summary
Unmet clinical need:
Distal cortical impingement may occur in up to 25%
of cases and is often the result of the curve of the natural femoral anatomy
being greater than the curve of the hip nail causing a “mismatch”.1-3 This
complication may lead to a fracture at the distal nail tip, called anterior
perforation, which requires revision surgery.1,4,5
Cut-out is the major cause of implant failure in the fi xation
of proximal hip fractures, and may cause severe injuries in hard
and soft tissues surrounding the hip joint.7,8 Cut-out rates for
cephalomedullary nails have been reported as high as 8%, and
frequently require reoperation.7
Nail breakage may occur in as many as 5% of hip fracture
patients treated with cephalomedullary nails and requires revision
surgery.11,13 Nail breakage is often a result of fractures that take longer
to heal or fail to heal (delayed union or nonunion, respectively).14
CLINICAL VALUE
1
The TFNA System was designed to address clinical challenges:
TFNA Helical Blade technology was
designed to compress bone during insertion, which enhances
implant anchorage and may reduce the risk of cut-out. Resistance
to cut-out in osteoporotic bone is further improved by cement
augmentation of the TFNA Helical Blade or screw.9,10
• A prospective, randomized clinical trial showed lower cut-out
rates in the cohort treated with a helical blade (1.5%)
compared to the screw (2.9%).11
• Biomechanical testing of off-center placement showed signifi cant
improvement in cut-out resistance for the TFNA Helical Blade
compared to both the TFNA Screw and Gamma3 Screw.12 These
results show the TFNA Helical Blade is more forgiving in regards to head element positioning.
• A clinical study with 15.3 months of follow-up showed no cut-out events for patients treated with
Helical Blade Technology and augmentation (n=62 patients).9
Off-center Position of Head Element – Mean Failure Loads (N)18
1021 ± 33 N
577 ± 41 N 612 ± 58 N
Gamma3 ScrewTFNA ScrewTFNA Helical Blade
Distal Nail Tip Position is less anterior with the TFNA Nail than the Gamma3 Nail
TiMo titanium alloy and BUMP CUT™ Design provides the TFNA System with improved fatigue
strength when compared with existing nails of similar size.15
Results of biomechanical testing showed greater fatigue
strength for the TFNA System compared to Gamma3 and
InterTAN nails (p<0.05).15
TFNA
Gamma3
Fatigue Limit is Greater for TFNA Nails than Gamma3 and InterTAN Nails15 ‡
The TFNA System was designed with a 1.0m radius of curvature
(ROC) to reduce the risk of distal cortical impingement and anterior perforation
compared to competitive nail systems with larger ROCs.
A multi-ethnic, 3-D computational study showed the TFNA nail (1.0 m ROC) resulted
in a better fi t than Gamma3 (1.5 m ROC).6
• Nail protrusion measurements are signifi cantly smaller for the TFNA Nail compared
to Gamma36
Mean total surface area Mean maximum distance in axial plane p
TFNA Nail 915.8 mm2 1.9 mm2 <0.05
Gamma3 Nail 1181.6 mm2 2.1 mm2 <0.05
• Distal Nail Tip Position:
– TFNA Nail had a higher number
of center positions than Gamma3
(13 vs. 7, respectively).6
– The far anterior position for the nail tip was observed
markedly less often with the TFNA nail compared
with the Gamma3 nail (16 vs. 30, respectively).6
*Differences are statistically signifi cant (p<0.05)
1900
1800
1700
1600
1500
1400
1300
1200
1100
1000TFNA System Gamma3* InterTAN*
24% difference 47% difference
Med
ian
Fatig
ue L
imit
(N)
1200
1000
800
600
400
200
0
2TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
ECONOMIC VALUE
Economic Challenge: TFNA Solution:
High Cost of Reoperation
Reduction in revision surgeries due to cut-out or other complications may result in cost
reductions for the hospital.
• A budget impact analysis was developed to show the potential economic impact to
a hospital using the TFNA System with a helical blade compared with a proximal nail
system with a screw. Results showed a hospital performing 200 proximal hip nailing
cases per year may save £23,923 annually due to the difference in cut-out rates for
helical blades versus screws.
Hospital Standardization
Aligning surgeons with hospital cost reduction initiatives, such as standardization of
physician preference items, is an important step in reducing clinical supply spending and
creating opportunities for hospital savings.16
• 86% of surgeons using the TFNA System "Strongly Agreed/Agreed" that they would
"Recommend this new proximal femoral nailing system"17
• The fl exibility of the TFNA System allows the surgeon to optimize the procedure based
on patient need and surgeon preference.
For the hospital, the TFNA System offers a single hip nail system providing surgeons with
the choices they need to treat a wide variety of fracture types while promoting hospital
standardization strategies.
Procedural Effi ciency in OR
The instruments used with the TFNA System introduce design features, such as QUICK
CLICK™ Self-Retaining Technology and radiolucent insertion handles with radiographic
indicators designed to streamline the procedure in the OR, potentially reducing OR time
and minimizing pain points within the surgical procedure for OR staff and surgeons.
• 74% of surgeons "Strongly Agreed/Agreed" that "I felt the new system improved
the overall procedural effi ciency compared to previously used nailing systems"17
• 77% of surgeons "Strongly Agreed/Agreed" that "The new instrument is easier
than what I used previously"17
• In patients with severely osteoporotic bone, the use of augmentation with the helical
blade enhances implant anchorage and further reduces the risk of cut-out.9 The use
of augmentation provides additional value to the hospital when enhanced implant
anchorage is needed in patients with severe osteoporotic bone.
£60,000
£50,000
£40,000
£30,000
£20,000
£10,000
0
Screw Helical Blade
£25,632
£49,555
Ann
ual C
ost
of In
patie
nt H
ospi
tal C
are
(GBP
)
£23,923 difference
Annual Cost of Inpatient Hospital Care
3
BACKGROUND
Pertrochanteric, simple Pertrochanteric, multifragmentary
Intertrochanteric Subtrochanteric
Most hip fractures are treated by orthopedic surgery, which involves implanting an orthopedic device. The fracture
takes approximately 4-6 months to heal.22 The surgery is a major stress on the patient, particularly in the elderly.
Revision procedures should be avoided given the increased surgical risk in these patients.
Hip fractures are common in the elderly; The number and costs of hip fractures are expected
to rise as the population ages.18 Reducing the reoperation rate, which has been estimated
at 6.3%,19 provides an opportunity for hospitals and health care systems to save costs.20
HIP FRACTURESA hip fracture is a femoral fracture that occurs in the proximal end of the femur (the long bone running through
the thigh), near the hip. The term "hip fracture" is commonly used to refer to the fracture patterns as shown
in Figure 1. In the vast majority of cases, a hip fracture is a fragility fracture due to a fall or minor trauma in
someone with weakened osteoporotic bone.21 Hip fractures in people with normal bone are often the result
of high-energy trauma such as car accidents, falling from heights (>10 ft.), or sports injuries.21
FIGURE 1: Types of Hip Fracture Patterns:
4TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
EpidemiologyHip fractures are a major public health problem in terms of patient morbidity, mortality, and costs to health and
social care.23 The incidence of hip fracture increases steeply with age due to higher rates of osteoporosis and falls
in the elderly population. Hip fractures account for the majority of osteoporotic fragility fractures and for over
40% of the estimated burden of osteoporosis worldwide.23 In 2010, there were an estimated 600,000 incident
hip fractures in the European Union, and this number is expected to rise to 972,000 by 2050.23,24
Trochanteric fractures constitute up to 55% of proximal femoral fractures and occur predominantly in elderly
patients.25 Most commonly, trochanteric fractures are caused by low-energy trauma events, such as falls from a
standing position, usually in combination with osteoporosis.26,27
Due to the patients’ high age and comorbidities, fractures of the proximal femur are often life threatening: in
the fi rst postoperative year, mortality rates may be as high as 30%.26,28 In young patients, trochanteric fractures
are typically associated with high-energy trauma events, such as motor vehicle, bicycle, and skiing accidents.25
Economic BurdenIn 2010, there were an estimated 600,000 incident hip fractures in the European Union, costing €20 billion and
accounting for 54% of the total costs of osteoporosis.23 These costs are expected to increase exponentially
following the growing incidence rate.24
The economic burden of treating hip fractures is substantial:
• UK – Total annual hospital costs associated with incident hip fractures were estimated at £1.1 billion in 2013.23
• Germany – 109,341 patients experienced a hip fracture in 2002, and the annual cost of care was
€2,998,000,000.29
• Portugal – Mean individual fracture-related costs were estimated at €13,434 for the fi rst year and €5985 for
the second year following the fracture. In 2011 the economic burden attributable to osteoporotic hip fractures
in Portugal could be estimated at €216 million.30
• Netherlands – The healthcare costs due to osteoporosis-related hip fractures (approximately €11,000–€13,000
per person) substantially exceeded those of other osteoporosis related fractures, such as spine, upper
extremity, lower extremity, and wrist/distal forearm.31
• Italy – A total of 85,762 hospitalizations due to hip fractures (17,597 men and 66,674 women) were recorded
in 2005 for people >65 years old.32 Hospital costs were €467 million in 2005, with rehabilitation costs reaching
€532 million in the same year.32
5
Clinical BurdenHip fractures result in pain, loss of mobility, and high rates of mortality.33 Nearly all patients are hospitalized,
and most undergo surgical repair of unstable fractures using cephalomedullary nails. Fractures of the hip are
associated with significant loss of function; one year after the fracture, fewer than 50% of patients have the
same walking ability they had prior to the fracture.34 Many patients lose their independence and need long-term
care. Comorbidity is an important contributory factor to hip fractures and is often a determinant of outcome.33
The reoperation rate of cephalomedullary hip nailing has been estimated at approximately 6.3%.19 The most
common types of complications resulting in revision include distal cortex penetration (≤3% revision rate4,5),
proximal cut-out (≤8% revision rate7,19), and implant breakage (≤5% revision rate13). Revision surgeries are
associated with a poor prognosis, an increase in mortality, and a decrease in the number of patients able
to return to their original residence.35
6TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
METHODS
This value analysis brief presents information on the potential clinical, and economic benefi ts of using
the TFNA System. The referenced data were obtained through a systematic literature review of Ovid
Medline, Ovid Embase, and PubMed for clinical and economic studies published from 2003-2016.
This literature search resulted in a total of 97 publications that met the inclusion and exclusion
criteria. Papers were selected for use in this value analysis brief based on the highest level of clinical,
biomechanical, and economic evidence. Recently completed biomechanical studies were also
included to support the value propositions for the TFNA System and referenced as “Data on File”.
Published results included studies reporting outcomes for proximal hip nails currently on the market
with similar features as the TFNA System. The TFNA System builds upon the clinical heritage of
existing DePuy Synthes Trauma technology:
• The TFNA Helical Blade technology is similar to the existing Helical Blade technology used in the
Trochanteric Fixation Nail (TFN) and Proximal Femoral Nail Antirotation (PFNA) Systems.
• The TFNA LATERAL RELIEF CUT™ is comparable to the lateral relief cut of the Proximal Femoral Nail
Antirotation-II (PFNA-II) System.
• The augmentation option available with the TFNA System is analogous to the cement
augmentation option available with the PFNA System.
7
TFNA SYSTEM CLINICAL VALUE
ROC = radius of curvature
Note: The gold nail represents the TFNA Nail with 1.0 ROC. The blue nail represents a nail with 1.5 m ROC.
Penetration of the anterior cortex of the distal femur is a complication associated with treating proximal
femoral fractures with intramedullary devices.1 Use of long cephalomedullary nails may result in the distal tip
of the nail abutting the anterior cortex of the femur, which is called “nail impingement”.1 Distal cortical
impingement is often the result of the curve of the femoral anatomy being greater than the curve of the hip
nail (nail-canal mismatch, see Figure 2).1-3 Published clinical studies have reported rates of distal cortical
impingement of up to 25%,1 and this complication may lead to fracture at the distal nail tip in the early
post-operative period.1 This fracture event, called anterior cortex perforation, occurs in up to 3% of cases and
requires revision surgery.4,5
Cephalomedullary nail designs include both short and long nails. The long nails extend to the end of the distal
femoral metaphysis (i.e. wide portion of the bone above the femoral condyles).36 The distal nail design,
specifi cally the radius of curvature (ROC) of a long nail, as well as the nail entry point, and proximal nail
geometry are important factors for clinical success.3
Prior to the launch of the TFNA System, the ROCs for commercially available cephalomedullary nails ranged
from 1.3 m to 3.0 m,37-39 and clinical experiences from recent studies have shown that these ROCs may lead
to complications resulting from nail-canal mismatch.3,39,40 Collinge and Beltran (2013) reported that femoral
nails with a ROC of 1.5 m more closely approximate the femoral bow of geriatric patients with hip fracture
than nails with a ROC of 2.0 m, and may be less likely to cause complications, such as anterior cortical
abutment, perforation, or fracture.3 This study reported rates of distal cortical impingement of 12% with a 2.0
m ROC nail (InterTAN Nail System with 2.0m AP bow, Smith and Nephew) but only 3% with a 1.5 m ROC nail
(InterTAN Nail System with 1.5m AP bow, Smith and Nephew,).3 A nail resting against the anterior femur also
may be associated with thigh or knee pain.3 Therefore, continuing to decrease the nail ROC to more closely
match anatomical measurements may further improve treatment outcomes.
DISTAL CORTICAL IMPINGEMENT AND ANTERIOR CORTEX PERFORATION
FIGURE 2: Curvature of the TFNA Hip Nail System (1.0 m ROC)
and Nail with 1.5 m ROC‡
The TFNA System was designed to enhance anatomic fi t, which may improve patient outcomes
and reduce the risk of postoperative complications.
8TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
The TFNA Nail was designed with a ROC of 1.0 m to more closely match the femoral anatomy than hip nails
with larger ROCs (i.e. straighter nails). A 3D computational study showed the 1.0m ROC of TFNA resulted in a
better fi t than the 1.5 m ROC nail (Gamma3 Long Nail R1.5, Stryker Trauma).6 This study included samples
derived from Caucasian (n=31), Japanese (n=28), and Thai (n=4) subjects with a mean age of 77 years (range
65 to 103 years). The results showed:
Nail Protrusion• Nail protrusion measurements are signifi cantly smaller for the TFNA Nail compared to Gamma3:6
ROCMean total surface
areaMean maximum distance
in axial plane p
TFNA Nail 1.0m 915.8 mm2 1.9 mm2 <0.05
Gamma3 Nail 1.5m 1181.6 mm2 2.1 mm2 <0.05
• Subgroup analyses showed mean total surface area of nail protrusion was signifi cantly smaller with the
TFNA nail compared to Gamma3 in Caucasian (p=0.0009) and Asian samples (p=0.000002).6
Distal Nail Tip Position• TFNA Nail had a higher number of center positions than Gamma3 (13 vs. 7, respectively).6
• The far anterior position for the nail tip was observed markedly less often with the TFNA nail compared
with the Gamma3 nail (16 vs. 30, respectively).6
These biomechanical results showing the improved fi t of the TFNA Nail of this study showed the TFNA Nail,
with a 1.0 ROC design, resulted in a better fi t compared with the Gamma3 nail with a 1.5m ROC design.
This could result in clinical improvements in implant fi t and potentially fewer post-operative complications.6
FIGURE 3: Distal Nail Tip Position for TFNA Compared to Gamma36‡
35
30
25
20
15
10
5
0
TFNA System Gamma3
Num
ber
of b
ones
Anterior
20
13
Center
13
7
Posterior
2 1
Far posterior
1 0
Far anterior
16
30
9
Case StudyApplying 3D computer modelling to an average Caucasian sample with an ROC of 1.015m resulted in a slightly
smaller misfi t in the subtrochanteric region for the TFNA Nail compared with the Gamma3 nail (Figure 4).6
Distally, the TFNA achieved a center position whereas the Gamma3 nail showed an anterior position (Figure 4).6
FIGURE 4: Case Study of Caucasian Model with an ROC of 1.015m Shows
Better Fit with TFNA Nail (1.0m ROC) vs. Gamma3 (1.5m ROC)
This 3D computational study showed the TFNA Nail with a 1.0m ROC design resulted in a better fi t compared
with the Gamma3 nail with a 1.5m ROC design as evaluated by protrusion area, protrusion distance, and far
anterior nail tip positions.6
In addition to nail-canal mismatch and anterior perforation of the cortex, lateral extrusion of the nail and
impingement are also potential complications associated with the fi t of the nail. The small proximal diameter
and the LATERAL RELIEF CUT™ Design (Figure 5) of the TFNA Nail avoid impingement on the lateral cortex while
preserving bone in the insertion area, potentially reducing the risk of fracture displacement. Additionally, the
oblique cut on the lateral end of the TFNA Blade and Screw reduces lateral protrusion on the soft tissues when
compared with that of a standard cut head element.41
FIGURE 5: TFNA System: Designed with a Small
Proximal Diameter, Oblique Cut, and LATERAL
RELIEF CUT™
Oblique cut on the lateral end of head element
Small proximal diameter
LATERAL RELIEF CUT
TFNA: 1.0m ROC
Gamma3: 1.5m ROC
10TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
TFNA Helical Blade technology is designed to compress bone during insertion, which enhances implant anchorage
and may reduce the risk of cut-out, a serious post-operative complication often resulting in reoperation.
Cut-out is the major cause of implant failure in dynamic hip screws, accounting for more than 80% of failures.7
PROXIMAL CUT-OUT
Defi nition of Cut-OutImplant cut-out is a loss of implant anchorage in the bone that causes the femoral
neck-shaft angle to collapse, leading to the extrusion, or cutting-out, of the screw or
blade element from the femoral head (Figure 6). Revision surgery is often necessary
when cut-out occurs.42
Cut-out is the most common43 cause of implant failure in the fi xation of trochanteric
fractures, accounting for more than 80% of failures in cases using dynamic hip screws.7
Cut-out rates for cephalomedullary nail devices were reported as 3.2% in a Cochrane
review of the literature,19 are often reported as high as 8%,7,43 and have also been
reported as high as 33.3%.43
A recent study published by Mingo-Robinet and colleagues (2015) evaluated the cut-out
rates of 218 fractures treated with either the Gamma Nail (Stryker) or Gamma3 Nail
(Stryker) System with lag screws.43 Cut-out rates in patients treated with Gamma3 were
4.87% for stable fractures and 33.3% for unstable fractures.43 These values for
Gamma3, a second-generation nail system, were greater than cut-out rates reported for
Gamma, a fi rst-generation nail system (p<0.05).43 The authors concluded the Gamma3
lag screw design may be the cause of the worse clinical results compared with the
Gamma Nail in unstable fractures.43
Cut-out continues to be a major complication for intramedullary hip nailing devices11
and may cause severe injuries in hard tissues as well as in soft tissues surrounding
the hip joint.8
Advantages of Helical BladesHelical blade devices offer an advantage in fracture repair because they allow for bone
compaction around the head element and avoid bone loss that occurs with the drilling
and insertion of the standard hip screw. Figure 7 shows an image of the helical blade
and screw from the TFNA System. The helical blade is designed to be implanted without
pre-drilling, which compresses the surrounding bone7,44 resulting in an increase in
trabelcular bone density in the surrounding area (Figure 8).7 This enhances implant
anchorage45 and provides additional purchase (i.e. grip) in osteoporotic bones,7 which
may result in a decrease in the risk of cut-out. Additionally, this increase in bone
compaction minimizes the potential for rotation of the blade.7
Helical blades have shown a higher potential of rotational stability
compared with the screw-based nails.8 Furthermore, the helical
blade is associated with a statistically signifi cant 2- to 4-fold higher
torque resistance reducing rotational forces during insertion
compared with a screw system.8 Anti-rotation wires are often
required with screws to counterbalance these rotational forces.44
Use of the blade lessens the need for the anti-rotation wires, which
may add to the procedural effi ciency of the surgical technique.
FIGURE 6: Example of Cut-Out
Source: DePuy Synthes Trauma.
FIGURE 7: TFNA Screw and
TFNA Helical Blade
TFNA Helical Blade
TFNA Screw
11
The cut-out resistance of the TFNA Helical Blade was compared to the TFNA Screw and
the Gamma3 Screw in a biomechanical study conducted by the AO Foundation using
a foam model with properties that mimic osteoporotic bone.10 Head elements were
tested for fatigue strength in the foam model based on position of the head element
as either center or off-center.10 These measurements show a range of placement
options that may occur during a hip nailing procedure. Center position is the optimal
placement of the head element;46 however, placement may vary from surgeon to
surgeon resulting in off-center positioning of the head element.10
The mean failure load was calculated for each study group (TFNA Helical Blade, TFNA
Screw, and Gamma3) to determine the resistance to cut-out. Results showed failure
loads in a similar range for all head elements tested in the center position (range of
1489N to 1613N).12 For the off-center study group, the TFNA Helical Blade showed
statistically signifi cant higher resistance to cut-out compared to the TFNA Screw and
Gamma3 Screw (Figure 9).10,12 These results show that the TFNA Helical Blade is
more forgiving than a screw in terms of positioning and also shows greater resistance
to cut-out.10
In addition to the AO study evaluating the TFNA System,10 several
biomechanical studies have been published showing the improved
cut-out resistance of helical blades compared to lag screws.7,47,48
• A computational study was published by Goffi n and colleagues
(2013) evaluating the role of bone compaction with respect to
implant cut-out using a CT scan-based fi nite element model of
trochanteric fractures.7 The effect of bone compaction resulting
from the insertion of a proximal femoral nail anti-rotation (PFNA)
with a helical blade provided additional bone purchase in an
osteoporotic femoral head bone model characterized as 75% of
the initial bone density. The authors concluded that the helical
blade has the potential to decrease the number of cut-out cases
in severely osteoporotic patients.7
• A biomechanical study by Sommers and colleagues (2004)
simulated implant cut-out in an unstable pertrochanteric
fracture model.48 This model accounted for dynamic loading,
osteoporotic bone, and a defi ned implant offset. Foam models
were used to determine differences in cut-out resistance
between two lag screws (dynamic hip screw, Gamma3) and two
blade-type implant designs (dynamic helical hip system,
trochanteric fi xation nail). Results showed trochanteric fi xation
nail implants with a helical blade demonstrated the highest
cutout resistance of all implants included in the analysis.47
• A biomechanical study published by Lenich and colleagues
(2011) compared previously published experimental data of two
fi xation screws (DHS/Synthes®, Gamma3 nail/Stryker®) and
helical blades (TFN/Synthes®, PFNA/Synthes®) and compared
them to a theoretical model using loads acting in the hip during
daily activities.46 The helical blades showed greater stability
compared to the screws with regards to implant rotation in the
de-central position. This study concluded that center positioning
of the head element is the optimal position for both helical
blades and screws; however, the helical blade provides greater
stability in a de-central position compared to screws.46
FIGURE 9: Cut-Out Resistance In Vitro Modeling‡
1200
1000
800
600
400
200
0
1200
1000
800
600
400
200
0
Load
at f
ailu
re in
N
TFNA Helical Blade Shows Greater Resistance in the Off-Center Position Compared to TFNA Screw and Gamma3 Screw10
Gamma3 ScrewTFNA ScrewTFNA Blade
FIGURE 8: TFNA Helical Blade
Technology: Designed to
Compress Bone During Insertion
Screw
Helical Blade
12TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
The improved cut-out resistance of helical blades compared to lag screws has also been studied clinically.
The following clinical studies report the cut-out rates of helical blades compared to screws:
• Stern and colleagues (2011) published a prospective, randomized clinical trial of 335 pertrochanteric and
intertrochanteric fractures and compared cut-out rates of screw vs. helical blade constructs. Results showed
cut-out rates were lower in the blade group (1.5%) compared with the screw group (2.9%). All cases of
cut-out resulted in reoperation.11
• Lenich and colleagues (2010) published a single-center, case-series examining 322 patients with trochanteric
fractures treated with either third generation nails with helical blades or second generation nails with lag
screws. Results showed cut-out rates were lower in the cohort treated with helical blades (cut-out rate range
of 2.5%-7%) compared with lag screws (14% cut-out rate).48
• A multicenter case-series of 315 fractures published by Simmermacher and colleagues (2008) concluded
that nails with a helical blade limit the effects of early rotation of the head/neck-fragment in unstable
trochanteric fractures, likely preventing or at least delaying rotation-induced cut-out.49
The TFNA Helical Blade Technology provides the enhanced stability that is critical for reducing the risk of
cut-out compared to a lag screw.
13
AUGMENTATION
The TFNA System offers cement augmentation of the head element. Augmentation provides fi xation stability,
enhanced cut-out resistance, and reduced cut-through and medial migration, especially in osteoporotic bone.50
Failure of fi xation and cut-out are common problems in the treatment of osteoporotic hip fractures.51,52 Low
bone mineral density and thin cortices not only are major risk factors for hip fractures but also contribute to the
failure of fi xation postfracture.51 Achieving stable fi xation contributes to early patient mobilization and good
fracture healing.52
Augmentation of the weak bone with polymethylmethacrylate (PMMA) or calcium phosphate bone cement may
increase the stability of nail osteosynthesis, especially in unstable fractures and osteoporotic bone.9
Augmentation involved injecting the cement at the level of the neck of the femur;53 the process takes
approximately 10 to 15 minutes.50 The decision to augment may be made during surgery, allowing for full
intra-operative fl exibility for the surgeon.
Biomechanical StudiesSeveral biomechanical studies have been conducted to evaluate the performance of cement augmentation.
Sermon and colleagues (2012) conducted a study comparing PFNA blades implanted in paired femurs either
with or without cement augmentation.54 Cut-out resistance and rotational stability under cyclic loading were
evaluated. Results indicated that:
• Bone mineral density was signifi cantly related to the number of cycles to failure in non-augmented group
(P < 0.001), but not in the augmented group (P = 0.91).54
• Augmented samples showed a signifi cantly greater number of cycles to failure (P = 0.012) than
non-augmented.54
• In the groups with center position of the PFNA blade, cement augmentation led to an increase in loading
cycles of 225%. In the groups with off-center positioning of the blade, this difference was even more
substantial (933%) (see Figure 10).54
The authors concluded that augmentation with small amounts of PMMA signifi cantly enhances cut-out
resistance in proximal femoral fractures. The procedure may improve patient care, especially in
osteoporotic bone.54
FIGURE 10: Cement Augmentation Increases Cut-Out Resistance54 ‡
Augmented center
Augmented off-center
Non-Augmented center
Non-Augmented off-center
0 10,000
Cycles to 5º varus collapse
20,000 30,000PFNA = proximal femoral nail antirotation.
14TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
In a cadaveric-bone study, Fensky and colleagues (2013) compared augmented with non-augmented PFNA
fi xation.55 Cyclic testing was performed with axial loads up to 1,400 N at 10,000 cycles to simulate full
postoperative weight bearing. Results showed that stiffness after instrumentation was signifi cantly greater for
the cement-augmented group than for the non-augmented group (300.6 ± 46.7 N /mm vs. 250.3 ± 51.6 N/
mm; P = 0.001). Results indicated that the cement augmentation of PFNA may increase implant stability,
especially in osteoporotic pertrochanteric fractures.55
Clinical Studies Kammerlander and colleagues (2011) reported the results of a prospective, multi-center study to evaluate the
technical performance and early clinical results of augmentation of the PFNA blade with PMMA bone cement
(mean volume=4.2 mL).50 A total of 59 patients with osteoporosis were included in the study (mean age=84.5
years); mean follow-up was 4 months. Results showed 55.3% of the patients reached the same or better
mobility than before the fracture. No events of cut-out, cut-through, unexpected blade migration, implant
loosening, or implant breakage were observed. The overall surgical complication rate was 3.4%; however, no
complications were related to the cement augmentation. Results indicated that augmentation of the PFNA blade
prevents blade migration within the head-neck fragment, enhances implant anchorage, and leads to good
functional outcomes.50
Kammerlander and colleagues (2014) reported long-term results (mean follow-up=15.3 months) from an
enlarged population of the same patient group from the study published in 2011.9,50 In the 62 patients included
in the analysis, 59.6% of patients reached their pre-fracture mobility level within the follow-up time frame. The
overall surgical complication rate was 3.2%, with no complications related to the cement augmentation. The
mean hip joint space did not change signifi cantly at follow-up, and there were no signs of osteonecrosis in the
follow-up x-rays. In addition, no blade migration was observed. Augmentation with the PFNA blade leads to
good functional results and is not associated with cartilage or bone necrosis.9 Table 1 presents a side by side
comparison of the results from the two analyses of this patient group.
TABLE 1: Side-by-Side Comparison of Short-term and Long-term Results of Cement
Augmentation of the PFNA9,50
PFNA = proximal femoral nail antirotation.
Clinical Outcome Kammerlander et al., 2011 (N = 59) Kammerlander et al., 2014 (N = 62)
Mean follow-up 4 months 15.3 months
Mean volume of cement injected 4.2 mL 3.8 mL
Percentage of patients reaching their pre-fracture mobility level
55.3% 59.6%
Overall surgical complication rate 3.4% 3.2%
Complications related to cement augmentation
None None
15
As with PFNA, the TFNA head elements can be augmented with TRAUMACEM™ V+ bone cement. This cement
is applied through the head element around the tip of the implant with syringes and a specifi c needle kit
compatible with the TFNA HELICAL BLADE™ Technology and the TFNA Screw.41,56
The failure load of the TFNA Helical Blade and the screw were evaluated with and without augmentation in a
foam bone model that mimics human osteoporotic bone in a study conducted by the AO.10 The analysis
included samples with the head elements in the center position as well as the off-center position.10 Figure 11
shows the use of cement augmentation with the helical blade signifi cantly improves the resistance to cut-out in
both the center and off-center positions (p<0.001).10
FIGURE 11: TFNA Helical Blade with Augmentation Shows Greater Resistance
to Cut-Out in the Center and Off-Center Position in Osteoporotic Bone10
Similar fi ndings were observed for the TFNA Screw. The use of augmentation with the screw signifi cantly
improves the resistance to cut-out in both the center and off-center positions (p<0.001).10 This study shows
augmentation of the TFNA Helical Blade and Screw provides increased cut-out resistance in an osteoporotic
bone model.10
5000
4000
3000
2000
1000
0
Non-Augmented Augmented
Load
at
failu
re in
N
Off-center position
P<0.001
Center position
P<0.001
16TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
IMPLANT STRENGTHDelayed unions or nonunions are defi ned as broken bones that take longer than usual to heal or fail to heal,
respectively.14 Instances of delayed unions or nonunions create excess stress on the nail57 and may cause nail
failure, which frequently occurs at the proximal hole of the nail. Nail breakage may occur in as many as 5% of
hip fracture patients treated with cephalomedullary nails.13 Nail breakage requires revision surgery to replace
the broken nail with a total hip arthroplasty, a hemiarthroplasty, or another hip nail.58
Resistance to Nail Breakage: Nail StrengthThe TFNA Nail was designed with specifi c features and materials to increase its fatigue strength while allowing
it to have a reduced proximal diameter without compromising strength.
The TFNA System is constructed of T-15Mo (TiMo) titanium alloy. Other commercially available hip nails are
made of either Ti-6Al-4v (TAV) ELI alloy [Gamma3 (Stryker) and InterTAN (Smith and Nephew)] or Ti-6Al-7Nb
(TAN). TiMo was chosen as the alloy for the TFNA System because of its combination of high strength and
fatigue resistance. TiMo is a biocompatible titanium alloy that meets the requirements of ASTM F 2066 testing
protocol. When heat treated, the minimum mechanical strength of TiMo is 33% higher than TAV and 28%
higher than TAN (see Figure 12).59,60 Additional testing showed TiMo is not only stronger than TAV and TAN,
but also maintains a similar level of fl exibility.61,62
The increased strength in combination with the BUMP CUT™ Design of the proximal hole (Figure 13), and
other design improvements in the TFNA System, provides improved fatigue strength when compared with
existing nails of similar size.15
FIGURE 13: TFNA System BUMP CUT Design FIGURE 12: TFNA Material Strength‡
1400
1200
1000
800
600
400
200
0
1400
1200
1000
800
600
400
200
0
Mpa
min
TiMo is Stronger than TAV and TAN61,62
TANTAVTIMo
The TFNA Nail offers improved fi t without compromising nail strength.
17
The strength of the TFNA System was compared to Gamma3 and InterTAN nails in a biomechanical study.15 The
median fatigue limit for the TFNA Nails was 24% greater than the Gamma3 nail (p<0.05) and 47% greater than
the InterTAN nail (p<0.05).15 These results show the TFNA Nail is stronger than other nail systems with similar
proximal diameters (Figure 14).
The increase in fatigue strength of the TFNA system compared to Gamma3 was also observed using fi nite
element analysis (FEA).63 These results are shown in Figure 15.
Static LockingMost cephalomedullary nails systems provide surgeons the ability
to rotationally lock the head element in position allowing for
translation of the head element but not rotation; this is
commonly referred to as guided collapse. The TFNA System
provides surgeons with the option to statically lock the head
element thus preventing both rotation and translation. Static
locking involves fully tightening the set screw onto the head
element so that linear movement of the lag screw is prevented by
friction.64 Post-operative complications, such as excessive fracture
collapse from lag screw sliding, and prominent hardware pain,
may be prevented with static locking.65
Mean slippage load of the statically locked TFNA System was
compared to statically locked Gamma3 and InterTAN constructs.65
The TFNA System showed a 47% improvement in slip load
compared to InterTAN (p<0.001) and 48% improvement
compared to Gamma3 (p<0.001).65 These results are shown in
Figure 16. The TFNA System was the only implant system able to
complete the dynamic portion of the testing without slippage of
the head element.65,66
Note: *differences compared to TFNA are statistically signifi cant (p < 0.001)
TFNA System Gamma3* InterTAN*
48% difference 47% difference
FIGURE 16: Static Locking: Mean Slippage Load of
TFNA is Greater than Gamma3 and InterTAN65
1000
900
800
700
600
500
400
300
200
100
0
Load
(N)
FIGURE 15: FEA Results: TFNA
Nail Shows Greater Fatigue
Strength than the Gamma3 Nail63 ‡
1750
1500
1250
1000
750
500
TFNAGamma3M
edia
n Fa
tigue
Lim
it (N
)
FIGURE 14: Fatigue Limit is Greater for TFNA
Nails than Gamma3 and InterTAN Nails15 ‡
1900
1800
1700
1600
1500
1400
1300
1200
1100
1000
TFNA System Gamma3* InterTAN*
24% difference 47% difference
proximal diameters of the nails
Med
ian
Fatig
ue L
imit
(N)
*Differences are statistically signifi cant (p<0.05)
18TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
ECONOMIC VALUE OF THE HELICAL BLADE TECHNOLOGY Although the overall revision rate for hip fractures is low (6.3%19), revision procedures increase the risk of
postoperative complications. The costs of hospitalization for patients who experience a complication after
hip fracture surgery are substantially higher than those for patients without complications.35 Reduction in
complications and reoperations are direct ways to reduce healthcare costs.
The TFNA System offers Helical Blade Technology to enhance stability and reduce the risk of cut-out.
Complications, such as cut-out, cut-through, and medial migration are expensive to treat and frequently
require a reoperation.35 Costs of complications and revision surgeries place a substantial economic
burden on the hospital and healthcare system.
Quantifi cation of the economic impact of complications and
revisions may be assessed using a budget impact analysis. The
analysis in Figure 17 shows the potential cost savings for a
hospital that switches from using a proximal hip nailing system
with a screw to a system with a helical blade. The following
input parameters were used:
• Stern and colleagues (2011) reported cut-out rates of 2.9%
with a screw and 1.5% with the helical blade.11
• Mean hospital costs of revision were £8,544 (converted to
€10,944) as published by Leal and colleagues (2016)23
• Annual volume was estimated at 200 cases per year.
Under these assumptions, the annual economic impact to the
hospital of revision surgeries is estimated to be of £49,555 for
cephalomedullary nail systems using a screw and £25,632 for
systems using a helical blade. Based on this model, a hospital
performing 200 hip fracture cases per year may save £23,923
annually by using a helical blade hip nailing system compared to
a hip nailing system with a screw. This potential cost reduction is
due to the difference in revision rates for helical blades versus
screws. These data show how small differences in revision rates
have the potential to make a large economic impact to the
healthcare system.
ECONOMIC VALUE OF THE TFNA SYSTEM
The TFNA System includes Helical Blade Technology and the option for cement augmentation,
both features reduce the risk of costly reoperations due to cut-out.11,50 Reduction in reoperations
may result in substantial economic savings to the hospital system.
Budget Impact Calculations: hospital cost = annual volume x hospital cost of reoperation x reoperation rate
FIGURE 17: Potential Annual Hospital Costs
of Reoperation are Less for Helical Blades Compared
to Screws Due to Differences in Cut-Out Rate
£60,000
£50,000
£40,000
£30,000
£20,000
£10,000
0
Screw Helical Blade
£25,632
£49,555
Ann
ual C
ost
of In
patie
nt H
ospi
tal C
are
(GBP
)
£23,923 difference
Annual Cost of Inpatient Hospital Care
19
AUGMENTATION OPTION PROVIDES ADDITIONAL VALUE IN PATIENTS WITH SEVERE OSTEOPOROTIC BONEFailures due to cut-out mainly occur in patients with severe osteoporotic bone; therefore, the cut-out rates may
be even higher than the 2.9% published by Stern et al.11 Using cement augmentation with Helical Blade
Technology has been shown to enhance implant anchorage and reduce the rate of cut-out in severely
osteoporotic patients.9,50 Cement augmentation of the blade gives the fi xation construct much more stability
due to a larger bone-implant interface.9 In a clinical study published by Kammerlander and colleagues (2014),
no complications or revisions due to cut-out were reported for patients treated with augmentation and Helical
Blade Technology (n=62 patients, average follow-up of 15.3 months).9 The use of augmentation provides
additional value to the hospital when enhanced implant anchorage is needed in patients with severe
osteoporotic bone (Figure 18).
FIGURE 18: Potential for Additional Value to the Hospital by Reducing Costs
of Reoperation Due to Reduced Rates of Cut-Out with Augmentation
Budget Impact Calculations: hospital cost = annual volume x hospital cost of reoperation x reoperation rate
This economic value analysis only focused on one post-operative complication, cut-out. The economic impact
to the hospital may be even greater when the reductions in other post-operative complications are factored into
the analysis. Reducing the risk of complications and revisions may result in opportunities for additional cost
savings and reductions in the overall economic burden to the healthcare system.
Budget Impact Calculations: hospital cost = annual volume x hospital cost of reoperation Budget Impact Calculations: hospital cost = annual volume x hospital cost of reoperation
£60,000
£50,000
£40,000
£30,000
£20,000
£10,000
0
Ann
ual C
ost
of In
patie
nt H
ospi
tal C
are
(GBP
)
Screw
£49,555
2.9% cut-out rate11
Helical Blade
£25,632
1.5% cut-out rate11
With Augmentation
0% cut-out rate9,50
Annual Cost of Inpatient Hospital Care
20TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
FACILITATING STANDARDIZATION AND OPERATING ROOM EFFICIENCYHospitals that choose to use the TFNA System may realize indirect cost-savings opportunities from the
standardization of surgeon preference items and improving operating room effi ciency. These indirect cost
savings combined with direct cost savings from potential reduction in complications and revisions contribute
to the overall value of the TFNA System to the hospital.
Facilitating StandardizationStandardization of physician preference items is one method for enhancing a hospital’s supply chain and
driving profi tability.67 In addition to cost reduction, standardizing implants can improve effi ciency and quality
of care.68
Aligning surgeons with hospital cost-reduction initiatives, such as standardization of physician preference
items, is an important step in reducing clinical supply spending and creating opportunities for cost savings.16
However, surgeons often develop a strong preference for a specifi c device or manufacturer creating a
challenge for the hospital to incentivize alignment with standardization strategies that require surgeons to
change devices.69
A survey of 77 surgeons included three questions related to their clinical experience with the TFNA System:17
• 86% of surgeons stated they “Strongly Agreed/Agreed” they “would recommend this new proximal
femoral nailing system”.
• 74% of surgeons “Strongly Agreed/Agreed” they “felt the new system improved the overall procedural
effi ciency compared to previously used Nailing systems”.
• 77% of surgeons “Strongly Agreed/Agreed” that “the new instrumentation is easier than what
I used previously”.
These survey results indicate a high level of surgeon satisfaction with the TFNA System. The strong
willingness to recommend the TFNA System is a good indicator of potential surgeon alignment in support of
hospital standardization strategies.
Procedural Effi ciency in the Operating RoomOrthopedic instruments should be intuitive to use allowing the surgeon and operating room (OR) team to focus
completely on the patient and the procedure. The instruments used with the TFNA System introduce features,
such as QUICK CLICK™ Self-Retaining Technology and radiolucent insertion handles with radiographic
indicators. These instruments were designed to streamline the procedure in the OR, potentially reduce OR time,
and minimize pain points within the surgical procedure for surgeons and OR staff.
The TFNA System is optimized for hospital standardization, designed for procedural effi ciency,
and offers a wide array of options to address surgeon preference when treating a full range
of fracture types.
21
QUICK CLICK Self-Retaining Technology (Figure 19) was designed to ensure a fast and
effective link between the insertion handle and intramedullary nail, potentially improving
surgical effi ciency and reducing OR time. A mishandled instrument or implant may result
in the need for immediate-use steam sterilization or traditional steam sterilization,
respectively. Unexpected sterilization may delay the surgical procedure as much as 30
min.70 Re-sterilizations and longer surgical times lead to a greater risk of infection and
blood loss for the patient.70,71 For hospitals, longer procedure times and re-sterilizations
result in increased costs in addition to the risk of infection and readmissions.
Radiolucent insertion handles with radiographic indicators allow x-ray visualization and
assist with guide wire placement (Figure 20). Placement of the guide wire in the femoral
head is a critical step in a hip-nailing procedure. Guide wire position dictates fi nal
placement of the femoral head element. Studies have shown that proper positioning is
correlated with clinical success of the implant.11
FIGURE 19: TFNA System QUICK
CLICK Technology Instrumentation
FIGURE 20: TFNA System Radiolucent Insertion Handle
Flexibility of the TFNA SystemThe TFNA System offers the surgeon a wide portfolio of intramedullary nailing options
for the proximal femur. The fl exibility of the TFNA System allows the surgeon to optimize
the procedure based on patient need and surgeon preference. For the hospital, the TFNA
System offers a single hip nail system providing surgeons with the choices they need to
treat a wide variety of fracture types while promoting hospital standardization strategies.
SUMMARYThe TFNA System was designed to solve a wide range of unmet needs for surgeons,
OR staff, and hospital administrators. This system offers advancement in hip fracture
treatment, including outcome-based design, reduced procedural complexity, and
comprehensive surgical options. The TFNA System was developed to deliver clinical and
economic value to patients, surgeons, and hospitals through improved outcomes and
cost savings opportunities.
22TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
PRODUCT DESCRIPTION
To suit a wide variety of clinical needs and surgeon preferences,
the system includes an array of options, including both short
and long nails and the option for augmentation. Further, it is the
only system to offer both helical blade and screw options using
one nail. In addition, the long nail provides three distal locking
options, including a unique oblique distal hole offset 10°,
designed to better target bone in condyles.
The TFNA System also offers a preassembled locking mechanism
in the nail with the ability to both rotationally and statically lock
the helical blade or screw. All nails in The TFNA System are
made from a high-strength titanium alloy (TiMo alloy), and the
instrumentation is designed for procedural effi ciency and
improved x-ray visualization.
‡ Based on biomechanical bench testing. May not be indicative of clinical performance.
For a complete list of indications for use, warnings, and precautions, please see the package insert or surgical technique.
FIGURE 21: TFN-ADVANCED Proximal Femoral
Nailing System, Showing Both Blade and
Screw Options
The DePuy Synthes Trauma TFN-ADVANCED Proximal Femoral Nailing System is a cephalomedullary proximal
femoral nailing system (Figure 21) designed to match patient femoral anatomy, to help improve patient
outcomes, and address a wide variety of patient needs. Specifi cally, the TFNA Nail includes:
• A radius of curvature of 1.0 m
• LATERAL RELIEF CUT Design
• BUMP CUT Design
• Helical Blade Technology
• Cannulated head elements to allow for augmentation with PMMA bone cement
23
DEPUY SYNTHES TRAUMA: FOCUSED ON PATIENTS AND HOSPITALS
Trusted Quality and Innovation
• A century of breakthroughs that create value
Delivering Solutions That Help Improve Clinical Outcomes
• Industry leader in trauma
• Provide a broad, high-quality product portfolio that addresses all your trauma needs
Advanced Technical Support and Training
• Highly-trained, trauma-focused team
• Commitment to education and training
• Industry-leading, customisable education and training programs for entire OR staff
24TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
1. Roberts JW, Libet LA, Wolinsky PR. Who is in danger? Impingement and penetration of the anterior cortex of the distal femur during intramedullary nailing of proximal femur fractures: preoperatively measurable risk factors. J Trauma Acute Care Surg. 2012;73(1):249-254.
2. Egol KA, Chang EY, Cvitkovic J, Kummer FJ, Koval KJ. Mismatch of current intramedullary nails with the anterior bow of the femur. Journal of orthopaedic trauma. 2004;18(7):410-415.
3. Collinge CA, Beltran CP. Does modern nail geometry affect positioning in the distal femur of elderly patients with hip fractures? A comparison of otherwise identical intramedullary nails with a 200 versus 150 cm radius of curvature. Journal of orthopaedic trauma. 2013;27(6):299-302.
4. Parker MJ, Bowers TR, Pryor GA. Sliding hip screw versus the Targon PF nail in the treatment of trochanteric fractures of the hip: a randomised trial of 600 fractures. The Journal of bone and joint surgery. British volume. 2012;94(3):391-397.
5. Miedel R, Ponzer S, Tornkvist H, Soderqvist A, Tidermark J. The standard Gamma nail or the Medoff sliding plate for unstable trochanteric and subtrochanteric fractures. A randomised, controlled trial. The Journal of bone and joint surgery. British volume. 2005;87(1):68-75.
6. Schmutz B, Amarathunga J, Kmiec S, Jr., Yarlagadda P, Schuetz M. Quantifi cation of cephalomedullary nail fi t in the femur using 3D computer modelling: a comparison between 1.0 and 1.5m bow designs. Journal of orthopaedic surgery and research. 2016;11(1):53.
7. Goffi n JM, Pankaj P, Simpson AH, Seil R, Gerich TG. Does bone compaction around the helical blade of a proximal femoral nail anti-rotation (PFNA) decrease the risk of cut-out?: A subject-specifi c computational study. Bone & joint research. 2013;2(5):79-83.
8. Al-Munajjed AA, Hammer J, Mayr E, Nerlich M, Lenich A. Biomechanical characterisation of osteosyntheses for proximal femur fractures: helical blade versus screw. Studies in health technology and informatics. 2008;133:1-10.
9. Kammerlander C, Doshi H, Gebhard F, et al. Long-term results of the augmented PFNA: a prospective multicenter trial. Archives of orthopaedic and trauma surgery. 2014;134(3):343-349.
10. Hofmann L, Hagen J, Agarwal Y, et al. Impact of bone cement augmentation on fi xation strength of TFNA blades & screws. European Soceity for Trama and Emergency Surgery (ESTES); 24-26 April, 2016; Vienna, Austria.
11. Stern R, Lubbeke A, Suva D, Miozzari H, Hoffmeyer P. Prospective randomised study comparing screw versus helical blade in the treatment of low-energy trochanteric fractures. International orthopaedics. 2011;35(12):1855-1861.
12. Hofmann L. AO Foundation: Final Report for biomechanical evaluation of non-augmented nail head elements in surrogate femoral heads [Synthes GmbH: USTRA09022 Trochanteric Fixation Nail – Advanced (TFNA)]. 2015.
13. Brammar TJ, Kendrew J, Khan RJ, Parker MJ. Reverse obliquity and transverse fractures of the trochanteric region of the femur; a review of 101 cases. Injury. 2005;36(7):851-857.
14. American Academy of Orthopaedic Surgeons. Nonunions. March 2014.
15. DePuy Synthes Trauma. Data on File. Fatigue strength testing of cephalomedullary nails. 2014.
16. Moran C. Four steps to engage physicians in clinical supply cost reduction. The Advisory Board: At the Margins. 2015:1-6.
17. Depuy Synthes Trauma. Data on File. Market Preference Evaluation. 2014.
18. Gu Q, Koenig L, Mather RC, 3rd, Tongue J. Surgery for hip fracture yields societal benefi ts that exceed the direct medical costs. Clinical orthopaedics and related research. 2014;472(11):3536-3546.
19. Parker MJ, Handoll HH. Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures in adults. The Cochrane database of systematic reviews. 2010(9):1-240.
20. American Academy of Orthopaedic Surgeons. Management of hip fractures in the elderly: evidence-based clinical practice guideline. September 5, 2014.
21. Hahnhaussen J, Hak DJ, Weckbach S, Ertel W, Stahel PF. High-Energy Proximal Femur Fractures in Geriatric Patients: A Retrospective Analysis of Short-Term Complications and In-Hospital Mortality in 32 Consecutive Patients. Geriatric Orthopaedic Surgery & Rehabilitation. 2011;2(5-6):195-202.
22. Barwick JF, Nowotarski PJ. Femur Shaft Fractures (Broken Thighbone). 2011. http://orthoinfo.aaos.org/topic.cfm?topic=A00521. Accessed 1 August 2015.
23. Leal J, Gray AM, Prieto-Alhambra D, et al. Impact of hip fracture on hospital care costs: a population-based study. Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 2016;27(2):549-558.
24. European Commission: Report on Osteoporosis in the European Community - Action for Prevention. 1998.
25. Rüedi TP, Buckley RE, Moran CG. AO principles of fracture management. Vol 2. 2 ed. Davos: AO Publishing; 2007.
26. Bonnaire F, Strassberger C, Kieb M, Bula P. [Osteoporotic fractures of the proximal femur. What's new?]. Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen. 2012;83(10):882-891.
27. Watts JJ, Abimanyi-Ochom J, Sanders KM. Osteoporosis costing all Australians: A new burden of disease analysis - 2012 to 2022. Melbourne, Australia: Deakin Worldly Univeristy and University of Melbourne. 2013.
28. Kraus M, Krischak G, Wiedmann K, et al. [Clinical evaluation of PFNA(R) and relationship between the tip-apex distance and mechanical failure]. Der Unfallchirurg. 2011;114(6):470-478.
29. Konnopka A, Jerusel N, Konig HH. The health and economic consequences of osteopenia- and osteoporosis-attributable hip fractures in Germany: estimation for 2002 and projection until 2050. Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 2009;20(7):1117-1129.
30. Marques A, Lourenco O, da Silva JA. The burden of osteoporotic hip fractures in Portugal: costs, health related quality of life and mortality. Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 2015;26(11):2623-2630.
31. Lotters FJ, van den Bergh JP, de Vries F, Rutten-van Molken MP. Current and Future Incidence and Costs of Osteoporosis-Related Fractures in The Netherlands: Combining Claims Data with BMD Measurements. Calcifi ed tissue international. 2016;98(3):235-243.
32. Piscitelli P, Iolascon G, Argentiero A, et al. Incidence and costs of hip fractures vs strokes and acute myocardial infarction in Italy: comparative analysis based on national hospitalization records. Clinical interventions in aging. 2012;7:575-583.
33. Woolf AD, Pfl eger B. Burden of major musculoskeletal conditions. Bulletin of the World Health Organization. 2003;81(9):646-656.
34. Sernbo I, Johnell O. Consequences of a hip fracture: a prospective study over 1 year. Osteoporosis international : a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 1993;3(3):148-153.
REFERENCES
25
35. Palmer SJ, Parker MJ, Hollingworth W. The cost and implications of reoperation after surgery for fracture of the hip. The Journal of bone and joint surgery. British volume. 2000;82(6):864-866.
36. Kleweno C, Morgan J, Redshaw J, et al. Short versus long cephalomedullary nails for the treatment of intertrochanteric hip fractures in patients older than 65 years. Journal of orthopaedic trauma. 2014;28(7):391-397.
37. Georgiannos D, Lampridis V, Bisbinas I. Complications following Treatment of Trochanteric Fractures with the Gamma3 Nail: Is the Latest Version of Gamma Nail Superior to Its Predecessor? Surgery Research and Practice. 2014;2014:6.
38. Zimmer Natural Nail Brochure. 2010.
39. Ostrum RF, Levy MS. Penetration of the distal femoral anterior cortex during intramedullary nailing for subtrochanteric fractures: a report of three cases. Journal of orthopaedic trauma. 2005;19(9):656-660.
40. Bazylewicz DB, Egol KA, Koval KJ. Cortical encroachment after cephalomedullary nailing of the proximal femur: evaluation of a more anatomic radius of curvature. Journal of orthopaedic trauma. 2013;27(6):303-307.
41. DePuy Synthes Trauma. Data on File. Design Review Files: 0000095553 Head Element and 0000098188 Fenestrated Head Element. 2014.
42. Butler M, Forte M, Kane RL, et al. Treatment of common hip fractures. Evidence report/technology assessment. 2009(184):1-85, v.
43. Mingo-Robinet J, Torres-Torres M, Martinez-Cervell C, et al. Comparative study of the second and third generation of gamma nail for trochanteric fractures: review of 218 cases. Journal of orthopaedic trauma. 2015;29(3):e85-90.
44. Abdulkareem IH. A review of tip apex distance in dynamic hip screw fixation of osteoporotic hip fractures. Nigerian Medical Journal : Journal of the Nigeria Medical Association. 2012;53(4):184-191.
45. Windolf M, Muths R, Braunstein V, Gueorguiev B, Hanni M, Schwieger K. Quantification of cancellous bone-compaction due to DHS Blade insertion and influence upon cut-out resistance. Clinical biomechanics. 2009;24(1):53-58.
46. Lenich A, Bachmeier S, Prantl L, et al. Is the rotation of the femoral head a potential initiation for cutting out? A theoretical and experimental approach. BMC musculoskeletal disorders. 2011;12:79.
47. Sommers MB, Roth C, Hall H, et al. A laboratory model to evaluate cutout resistance of implants for pertrochanteric fracture fixation. Journal of orthopaedic trauma. 2004;18(6):361-368.
48. Lenich A, Vester H, Nerlich M, Mayr E, Stockle U, Fuchtmeier B. Clinical comparison of the second and third generation of intramedullary devices for trochanteric fractures of the hip--Blade vs screw. Injury. 2010;41(12):1292-1296.
49. Simmermacher RK, Ljungqvist J, Bail H, et al. The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008;39(8):932-939.
50. Kammerlander C, Gebhard F, Meier C, et al. Standardised cement augmentation of the PFNA using a perforated blade: A new technique and preliminary clinical results. A prospective multicentre trial. Injury. 2011;42(12):1484-1490.
51. Augat P, Rapp S, Claes L. A modified hip screw incorporating injected cement for the fixation of osteoporotic trochanteric fractures. Journal of orthopaedic trauma. 2002;16(5):311-316.
52. Lindner T, Kanakaris NK, Marx B, Cockbain A, Kontakis G, Giannoudis PV. Fractures of the hip and osteoporosis: the role of bone substitutes. The Journal of bone and joint surgery. British volume. 2009;91(3):294-303.
53. Dall'Oca C, Maluta T, Moscolo A, Lavini F, Bartolozzi P. Cement augmentation of intertrochanteric fractures stabilised with intramedullary nailing. Injury. 2010;41(11):1150-1155.
54. Sermon A, Boner V, Schwieger K, et al. Biomechanical evaluation of bone-cement augmented Proximal Femoral Nail Antirotation blades in a polyurethane foam model with low density. Clinical biomechanics. 2012;27(1):71-76.
55. Fensky F, Nuchtern JV, Kolb JP, et al. Cement augmentation of the proximal femoral nail antirotation for the treatment of osteoporotic pertrochanteric fractures--a biomechanical cadaver study. Injury. 2013;44(6):802-807.
56. DePuy Synthes Trauma. Data on File. Clinical evaluation of trochanteric fixation nail advanced. Document number: 0000087418; Version A.34. 2014.
57. Chen CE, Weng LH, Ko JY, Wang CJ. Management of nonunion associated with broken intramedullary nail of the femur. Orthopedics. 2008;31(1):78.
58. D'Arrigo C, Perugia D, Carcangiu A, Monaco E, Speranza A, Ferretti A. Hip arthroplasty for failed treatment of proximal femoral fractures. International orthopaedics. 2010;34(7):939-942.
59. ASTM. Designation: F1295-11Standard Specification for Wrought Titanium-6Aluminum-7Niobium Alloy for Surgical Implant Applications (UNS R56700). 2011.
60. ASTM. Designation: F136 − 13 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). 2013.
61. DePuy Synthes Trauma. Data on File. Implant Materials: Wrought Titanium - 15% Molybdenum, 2nd ed. . 2003.
62. DePuy Synthes Trauma Data on File. Implant Materials: Titanium-6% Aluminum-7% Niobium, 2nd ed. 1993.
63. Bushelow M, Kmiec S, Shultzabarger B, McMillan R, Coombs D, Blauth M. Use of Finite Element Analysis to Predict the Fatigue Strength of Cephalomedullary Nail Systems. Paper presented at: Orthopedic Research Society; March 28-April 1, 2015; Las Vegas, NV.
64. Kuzyk P, Shah S, Zdero R, Olsen M, Waddell J, Schemitsch E. A biomechanical comparison of static versus dynamic lag screw modes for cephalomedullary nails used to fix unstable peritrochanteric fractures. Journal of Trauma and Acute Care Surgery. 2012;72(2):E65-E70.
65. Shultzabarger B, Pappalardo D, Bushelow M, Harris R. Comparison of Statically Locked Proximal Lag Screws in Cephalomedullary Nail Systems. Paper presented at: Orthopedic Research Society; March 28-April 1, 2015; Las Vegas, NV.
66. DePuy Synthes Trauma. Data on File. MT15-273 TFNA Pre/Post Fatigue Sliding Data Analysis 2015.
67. Herman B. 11 ways hospitals and health systems can increase profitability in 2013. Becker’s Hospital Review; November 26, 2012.
68. Rodak S. How bundled payments in orthopedics can help build the foundation for a center of excellence. Becker Hospital Review; February 6, 2013.
69. Lee J. Losing Preferrential Treatment. Modern Healthcare. 2013. http://www.modernhealthcare.com/article/20130215/MAGAZINE/302169953. Accessed 21 March 2015.
70. Muscarella L. Flash or Immediate Use Steam Sterilization: A Position Statement. 2014. http://endoscopereprocessing.com/2014/08/flash-sterilization-a-position-statement/. Accessed 6 July 2015.
71. Association for Professionals in Infection Control and Epidemiology. Guide to the elimination of orthopedic surgical site infections. 2010.
26TFN-ADVANCED™ Proximal Femoral Nailing System Value Analysis Brief |
This publication is not intended for distribution in the USA.
All surgical techniques are available as PDF files at www.depuysynthes.com/ifu
Synthes GmbH Eimattstrasse 3 4436 Oberdorf Switzerland Tel: +41 61 965 61 11 Fax: +41 61 965 66 00 www.depuysynthes.com 0123 ©
DeP
uy S
ynth
es 2
016
All
right
s re
serv
ed.
DSE
M/TRM/0515/0375(1)
6/16
JC