ultrasonic transdermal glucose monitoring and insulin

112
The Pennsylvania State University The Graduate School Department of Bioengineering ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN DELIVERY USING CYMBAL TRANSDUCER ARRAYS A Dissertation in Bioengineering by Eun-Joo Park 2010 Eun-Joo Park Submitted in Partial Fulfillment of the Requirements for the Degree of Doctor of Philosophy May 2010

Upload: others

Post on 19-Dec-2021

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

The Pennsylvania State University

The Graduate School

Department of Bioengineering

ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

DELIVERY USING CYMBAL TRANSDUCER ARRAYS

A Dissertation in

Bioengineering

by

Eun-Joo Park

2010 Eun-Joo Park

Submitted in Partial Fulfillment of the Requirements

for the Degree of

Doctor of Philosophy

May 2010

Page 2: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

ii

The dissertation of Eun-Joo Park was reviewed and approved* by the following:

Nadine Barrie Smith Associate Professor of Bioengineering Dissertation Advisor Chair of Committee

Andrew G. Webb Professor of Bioengineering

Ryan S. Clement Assistant Professor of Bioengineering

Richard J. Meyer Jr. Associate Professor of Materials Science and Engineering

Herbert H. Lipowsky Professor of Bioengineering Head of the Department of Bioengineering

*Signatures are on file in the Graduate School

Page 3: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

iii

ABSTRACT

As a practically portable ultrasound system for transdermal drug delivery, cymbal

transducer arrays have shown promise in several in vivo experiments using small animals.

The goal of this study was to evaluate the potential of the cymbal transducer array as a

pre-clinical application for transdermal insulin delivery and glucose sensing. In order to

achieve this goal, several aspects have been investigated, such as the feasibility of

ultrasonic transdermal insulin delivery and glucose sensing in large animals, and a

closed-loop feedback controller for regulating glucose levels. To evaluate further the

practicability of the cymbal transducer array as a pre-clinical application, in vivo

experiments with transdermal insulin delivery were performed in large pigs (~ 200 lbs).

The results of the ultrasound exposure group indicate that the glucose level decreased to

-74 ± 5 mg/dl at 60 minutes and continued to decrease to -91 ± 9 mg/dl at 90 minutes. In

addition to the insulin delivery, the cymbal transducer array has been used in determining

the glucose concentration of interstitial fluid. From the perspective of application in

humans, the in vivo studies using large pigs were designed to determine the feasibility of

the cymbal transducer array in ultrasonic transdermal glucose sensing. After 5 minutes of

ultrasound exposure at Isptp = 100 mW/cm2, the glucose levels determined by the

ultrasound system and glucose meter were 127 ± 16 mg/dl and 131 ± 5 mg/dl,

respectively. Based on the results of transdermal insulin delivery and glucose sensing in

large pigs, the ultrasound systems for both applications were combined using a closed-

loop feedback controller. Four in vivo experiments were designed with large pigs to

demonstrate the feasibility of the combined ultrasound system in noninvasive glucose

Page 4: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

iv

control. Based on the glucose level measured by ultrasound and biosensors at 20-minute

intervals, the ultrasound system of insulin delivery was shown to regulate the glucose

level. During the 120-minute experiments, glucose levels were controlled in the range of

104-137 mg/dl as determined by the ultrasound system. Through the in vivo experiments

of transdermal insulin delivery and glucose, the ultrasound system using the cymbal

transducer array has shown the positive potential of a practically portable device for a

clinical application of diabetes care.

Page 5: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

v

TABLE OF CONTENTS

LIST OF FIGURES .....................................................................................................vii

LIST OF TABLES....................................................................................................... ix

Chapter 1 INTRODUCTION......................................................................................1

1.1 Motivations .....................................................................................................1 1.1.1 Diabetes ................................................................................................1 1.1.2 Glucose monitoring and insulin delivery .............................................4

1.2 Specific Aims and Scope ................................................................................5 1.3 Dissertation Outlines ......................................................................................7

Chapter 2 BACKGROUND - LITERATURE REVIEW ...........................................8

2.1 Transdermal Drug Delivery............................................................................8 2.2 Glucose Monitoring ........................................................................................13 2.3 Closed-loop Control for Diabetes Care ..........................................................14

Chapter 3 BACKGROUND - ACOUSTIC THEORY ...............................................17

3.1 Design of cymbal transducer ..........................................................................17 3.2 Acoustic Characteristics of Cymbal Transducer ............................................18 3.3 Acoustic Cavitation ........................................................................................20

3.3.1 Definition..............................................................................................21 3.3.2 Cavitation Threshold ............................................................................22

Chapter 4 MATERIALS AND METHODS - APPRATUS .......................................25

4.1 Construction of Cymbal Transducer Arrays...................................................25 4.1.1 Fabrication of cymbal arrays ................................................................25 4.1.2 Construction of matching circuits ........................................................29

4.2 Exposimetry: Acoustic Characteristics of Cymbal Arrays.............................30 4.3 Fabrication and Calibration of Electrochemical Glucose Sensors .................32

4.3.1 Fabrication of glucose sensors..............................................................33 4.3.2 Calibration of glucose sensors..............................................................34

4.4 Closed-Loop Feedback Controlled Ultrasound System for Insulin Delivery and Glucose Sensing.......................................................................35

Chapter 5 MATERIALS AND METODS – IN VIVO EXPERIMENTS ...................38

5.1 Transdermal Insulin Delivery .........................................................................39 5.2 Ultrasonic Transdermal Glucose Sensing.......................................................43 5.3 Closed-Loop Feedback Controlled Ultrasound System for Glucose

Control ...........................................................................................................46

Page 6: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

vi

Chapter 6 RESULTS...................................................................................................50

6.1 Acoustic Pressure Distribution of Cymbal Transducer Arrays ......................50 6.2 Transdermal Insulin Delivery ........................................................................51

6.2.1 In Vivo Experiments in Large Pigs .......................................................51 6.2.2 Analysis of Insulin Concentration in Transdermal Insulin Delivery....53 6.2.3 Ultrasonic Glucose Sensing..................................................................56 6.2.4 Glucose control in large pigs using a closed-loop feedback

controlled ultrasound system..................................................................58

Chapter 7 CONCLUSION ..........................................................................................62

7.1 Transdermal Insulin Delivery .........................................................................62 7.2 Ultrasonic Glucose Sensing............................................................................64 7.3 Closed-Loop Feedback Controlled Ultrasound System for Glucose

Regulation......................................................................................................66 7.4 Future Work....................................................................................................68

BIBLIOGRAPHY........................................................................................................71

Appendix A FUNDAMENTAL RESONANCE FREQUENCY OF CYMBAL TRANSDUCER – FEA ........................................................................................80

A.1 In Air..............................................................................................................80 A.2 In Water .........................................................................................................81

Appendix B DETAIL INSTRUCTION OF CYMBAL ARRAY CONSTRUCTION ...............................................................................................83

B.1 Metal Endcaps................................................................................................83 B.2 Cymbals .........................................................................................................85 B.3 Polyurethane Base..........................................................................................85 B.4 Connections....................................................................................................86 B.5 Casting ...........................................................................................................87

Appendix C A COMPARISON BETWEEN INTRAPERITONEAL INJECTION AND TRANSDERMAL DELIVERY..................................................................91

Appendix D MATLAB CODES.................................................................................98

D.1 Low pass π-circuit for impedance matching..................................................98 D.2 Feedback controller........................................................................................100

Page 7: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

vii

LIST OF FIGURES

Figure 3.1: Diagram of cymbal transducer in cross sectional view............................18

Figure 4.1: PZT-4 disk and a brass endcap.................................................................26

Figure 4.2: Dimensions of endcaps and PZT-4 disk...................................................27

Figure 4.3: Lightweight cymbal arrays were constructed using four transducers in a two-by-two pattern, (left) or nine cymbal transducers in a three-by-three pattern (right). The two-by-two array was used for ultrasonic glucose sensing, while the three-by-three array was used in the transdermal insulin delivery. ......29

Figure 4.4: Electric impedance matching network using a π-type circuit. After constructing the circuit based on the theoretical values, actual value of each component was adjusted to have maximum acoustic output................................30

Figure 4.5: Equipment setup of the exposimetry to characterize the cymbal arrays. Acoustic pressure field of each cymbal array was measured in the water tank filled with degassed water. ...................................................................................31

Figure 4.6: Photograph of an enzyme based electrochemical biosensor which consists of three electrodes: working/sensing, counter, and reference electrodes ..............................................................................................................34

Figure 5.1: Photograph of a transdermal insulin delivery experiment with pig placed in a lateral recumbent position with the array attached.............................40

Figure 5.2: (a) Layout of the glucose sensor incorporated with the cymbal transducer array placed on the skin surface. (b) Photograph of the ultrasonic glucose sensing on a pig with the array attached to the skin and the reservoir filled with PBS solution........................................................................................44

Figure 5.3: (a) The layout of the glucose sensing system incorporated with the insulin delivery system as placed on the skin surface. After five minutes of ultrasound exposure, the biosensor determined the glucose concentration of the interstitial fluid. Based on the glucose level, the insulin delivery system was automatically operated. (b) Photograph of the ultrasound systems placed on the pig for insulin delivery and glucose sensing. The ultrasound system was controlled by a closed-loop controller. ..........................................................47

Figure 5.4: Schematic diagram shows the experimental setup with the closed-loop feedback controller for noninvasive glucose regulation. The ultrasound operation for both the glucose sensing and insulin delivery systems was automatically operated..........................................................................................49

Page 8: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

viii

Figure 6.1: Acoustic intensity profiles of the two-by-two (left) and three-by-three (right) arrays were determined by measuring acoustic pressure and plotted in dB. The spatial-peak-temporal-peak intensities were 100.8 ± 0.9 mW/cm2 for the two-by-two array and 100.0 ± 0.5 mW/cm2 for the three-by-three array. ............................................................................................................51

Figure 6-2: Change of blood glucose level in the ultrasound mediated transdermal insulin delivery on pigs.........................................................................................52

Figure 6.2: (a) The concentrations of human (, µU/l) and porcine (◊, ng/l) insulin were determined for 90-minutes periods. While porcine insulin concentrations stayed in the rage of 12.23 - 12.38 ng/l, human insulin concentration increased by 0.5 and 23.7 µU/l after 15 30 minutes ultrasound exposure, respectively. (b) Human insulin concentrations were determined in ng/l by unit conversion and compensated for the cross-reaction between porcine and human insulin for ELISA tests. The cross-reaction ratio was obtained at 13.3%. ................................................................................................55

Figure 6.3: The total insulin concentrations in ng/L were presented with the change of blood glucose level in mg/dl over 90-minute experiments. As insulin concentrations increased faster, the glucose level decreased at higher rate. .......................................................................................................................56

Figure 6.4: The glucose concentrations of the interstitial fluid determined using the ultrasound system and a conventional glucose meter. Asterisks above the data bar were used if there was no statistical difference between the ultrasonic glucose sensing method and the conventional method based on the ANOVA with the 0.05 or 0.01 level of significance. ...........................................57

Figure 6.5: (a) – (d) Glucose concentrations of four pigs, which were regulated by the ultrasonic transdermal glucose sensing and insulin delivery systems combined by a feedback controller. ......................................................................60

Figure C.1: Photograph of a transdermal insulin delivery experiment on a rat place in a dorsal decubitus position with the array attached. ...............................92

Figure C.2: Changes of blood glucose by the ultrasonic transdermal insulin delivery and subcutaneous injections of insulin. ..................................................94

Page 9: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

ix

LIST OF TABLES

Table 1.1: Comparison of Type 1 and Type 2 diabetes. .............................................3

Table 2.1: Enhanced skin permeability by ultrasound.................................................10

Table 4.1: To control the ultrasound system using cymbal arrays, a simple proportional controller was used. .........................................................................37

Table 5.1: The detail information of ultrasound exposure conditions. .......................45

Table A.1: Comparison of the in-air resonance frequency of standard size brass-capped cymbals for various PZT types.................................................................80

Table A.2: Comparison of the in-air resonance frequency of standard size cymbals for various endcap materials...................................................................80

Table A.3: Comparison of the in-air resonance frequency of standard size brass-capped cymbals for different endcap thickness. ...................................................81

Table A.4: Comparison of the in-air resonance frequency of standard size brass-capped cymbals for different cavity depths. .........................................................81

Table A.5: Comparison of the in-water resonance frequency of standard size brass-capped cymbals for various PZT types. ......................................................81

Table A.6: Comparison of the in-water resonance frequency of standard size cymbals for various endcap materials...................................................................82

Table A.7: Comparison of the in-water resonance frequency of standard size brass-capped cymbals for different endcap thickness. .........................................82

Table A.8: Comparison of the in-water resonance frequency of standard size brass-capped cymbals for different cavity depths. ...............................................82

Page 10: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

x

ACKNOWLEDGEMENTS

I would like to thank my family for their countless love and support. They have

been with me all the time and made me get through this long journey. Without them, I

wouldn’t be able to overcome all the obstacles. Also, I would like to thank Dr. Nadine

Barrie Smith. She was my role model, mentor, big sister, and a good friend although I

did not have a chance to tell her how much I was happy to have her. She was, is, and will

be my advisor.

Page 11: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Chapter 1

INTRODUCTION

1.1 Motivations

More than 23.6 million Americans suffer from diabetes – the seventh leading

cause of death in the United States (American Diabetes Association 2007). Diabetes is a

disease in which the body does not properly produce or utilize insulin, a hormone that

converts food (such as sugar and starches) into energy needed for daily life. Diabetes

costs the nation in excess of $170 billion per year and accounts for approximately 25% of

all Medicare expenditures. In addition, it is a risk factor for heart disease, stroke, and

birth defects and shortens average life expectancy by up to 15 years. Although much

research has been dedicated to finding the cause and a cure, the cause of diabetes remains

unknown.

1.1.1 Diabetes

Diabetes is a metabolic disorder characterized by chronic hyperglycemia with

disturbances of carbohydrate, fat, and protein metabolism resulting from defects in

insulin secretion, insulin action or both. Although the symptoms are often not severe or

may even be absent, diabetes may present characteristic symptoms such as thirst, polyuria,

weight loss, and blurred vision. Complications stemming from diabetes include long-

term damage, dysfunction and failure of several organs, especially the eyes, kidneys,

Page 12: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

2

heart, blood vessels and teeth (Beaser & Staff of Joslin Diabetes Center 2007). In the

absence of effective treatment, the complications can ultimately lead to death.

When an individual has symptoms such as thirst, polyuria, unexpected weight loss,

drowsiness or coma, and marked glucosuria, diagnostic screening using a fasting plasma

glucose test or an oral glucose tolerance test (OGTT) is required. Because of its

simplicity, the fasting plasma glucose test is recommended by the American Diabetes

Association (ADA). Under normal conditions, the plasma glucose concentration after an

eight-hour fast is less than 100 mg/dl while the 2-hour postload plasma glucose level is

lower than 140 mg/dl. If the fasting plasma glucose level is higher than 126 mg/dl or the

2-hour postload plasma glucose level is higher than 200 mg/dl, the patient is diagnosed as

having diabetes.

In 1997, the ADA proposed the current clinical classification, which was adopted

by the World Health Organization (WHO) in 1999, based on the fact that the majority of

diabetes cases falls into two broad etiopathologic categories: Type 1 and Type 2 diabetes.

Given the increasing number of forms of diabetes recognized by a specific etiology,

diabetes is currently classified using both etiologic types and clinical stages etiologic

types. The terms Type 1 and Type 2 of the current classification system refer to the most

common forms of diabetes.

Type 1 diabetes is caused by the destruction of the insulin producing β-cell of the

islets of Langerhans in the pancreas and usually leads to the clinical stage that requires

insulin treatment for survival. Type 1 diabetes is characterized by the presence of

antibodies that reflect the autoimmune process leading the β-cell destruction. In

individuals who have Type 1 diabetes, their immune system recognizes the β-cell as an

Page 13: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

3

antigen or pathogen and attacks the cells with antibodies (Beaser & Staff of Joslin

Diabetes Center 2007). The destruction of the β-cell results in the loss of insulin

secretion and absolute insulin deficiency (i.e., insulinopenia).

Type 2 diabetes is the most common form of diabetes, comprising approximately

90 percent of all diabetes cases. It is characterized by disorders of insulin action and

insulin secretion due to insulin resistance and reduced insulin sensitivity. Compared to

Type 1 diabetes, symptoms are not as severe in individuals having Type 2 diabetes since

the body produces enough or even too much insulin. Because Type 2 diabetes is

asymptomatic, it is often unrecognized and untreated during the early stages of

development. The differences between Type 1 and Type 2 diabetes are summarized in

Table 1.1.

Table 1.1: Comparison of Type 1 and Type 2 diabetes.

Type 1 diabetes Type 2 diabetes

Percentage of people with diabetes

10% 90%

Onset Usually below 30 years Usually over 30 years

Condition when discovered

Usually moderately to severely ill

Often not ill at all, or having mild symptoms

Cause of diabetes Reduced or absent insulin production

Insulin resistance and relative or absolute insulin secretory deficiency

Insulin level None to small amounts Markedly elevated early

Acute complications Ketoacidosis

Non-ketotic hyperosmolar hyperglycemic coma. Usually not prone to ketoacidosis

Usual treatment Insulin, meal plan, exercise Diet, exercise, antidiabetes medications and insulin if needed

Page 14: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

4

1.1.2 Glucose monitoring and insulin delivery

In the management of diabetes, a strict control of blood glucose concentration can

reduce the risk of short- and long-term complications. Multiple measurements of glucose

concentration are necessary to maintain the blood glucose level within the normal range.

In particular, self-blood glucose monitoring (SBGM) is very important in diabetes

treatment as it provides information related to glucose patterns to help the healthcare

team make appropriate treatment decisions, provides data with which patients can

independently make daily treatment-related decisions, and provides feedback on the

efficacy of the treatment, especially in regard to physical activities and nutritional

therapy. Conventional SBGM techniques use a portable glucose monitor, which requires

a blood sample obtained by pricking fingers. Although this method is widely used, it

limits the number of measurements because it is painful and invasive.

Insulin treatment is essential for Type 1 diabetes; however, it is not required in the

early stages of other types of diabetes—although may be required, along with a

combination of other treatments (e.g., diet, oral agents, and exercise) when the glucose

level is not well controlled. The goals of insulin treatment are to eliminate symptoms of

hyperglycemia, prevent diabetic ketoacidosis, avoid developing severe catabolic state,

reduce frequent infections, decrease fetal and maternal morbidity in pregnancy, and

prevent and delay microvascular and macrovascular complications. The traditional

insulin delivery method involves using a syringe and needle for injection; this needle

injection system is flexible and allows dosages to be adjusted readily. However, the

number of injections and injection sites on a body are limited, and this method involves

Page 15: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

5

pain that can affect patient compliance. To provide advantages over traditional injection

drug delivery methods, several noninvasive methods have been studied.

The challenge with the management of diabetes is to avoid the continuous process

of finger pricking for blood glucose samples and daily insulin injections, which can often

lead to noncompliance. Noncompliance typically occurs in the form of refusing

treatment, reducing or increasing insulin doses, and skipping medications. Such actions

are common problems that can prevent the efficacy of the treatment and negatively

impact the patient’s health. Ultimately, noncompliance can lead to complications such as

nerve damage, retinal damage, microvascular damage, chronic renal failure, kidney

failure, dental disease, and cardiovascular disease. In order to overcome noncompliance,

a great deal of effort has focused on finding alternative methods that can deliver insulin

noninvasively.

1.2 Specific Aims and Scope

Previous studies using ultrasound to facilitate noninvasive insulin delivery or

extraction of interstitial skin fluid for glucose monitoring have shown positive results.

However, these studies used a bulky commercial sonicator to generate the ultrasound

energy. Interest in a compact and lightweight ultrasound device indicates the need to

develop a portable ultrasound device for transdermal drug delivery.

Several possible low frequency transducer designs can be used in a drug delivery

application, such as the low frequency flextensional resonators, tonpilz transducers, or

“thickness”-type resonators. Another inexpensive candidate for a portable device is the

Page 16: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

6

low-profile, lightweight cymbal transducer. The cymbal transducer is a flextensional

transducer that is less than 2 mm thick, weighs less than 3 grams, and resonates between

1 and 100 kHz depending on the geometry. Multiple cymbals can be patterned for

increased spatial intensity fields for drug transport. Indeed, the use of the low-profile,

lightweight cymbal ultrasound array has previously demonstrated transport enhancement

of insulin across in vitro human skin as well as in vivo rats and rabbits. The cymbal

design has also been used for noninvasive glucose sensing in rat models.

The goal of the current research is to develop a practical ultrasound system for

noninvasively monitoring blood glucose and transporting insulin across the skin, which is

operated by a closed-loop feedback controller. To overcome the drawbacks of ultrasound

system that used in previous studies (El-Kamel, Al-Fagih, & Alsarra 2008; Kost 2002;

Mitragotri, Blankschtein, & Langer 1995c; Tachibana & Tachibana 1991; Zhang, Shung,

& Edwards 1996) and to improve the mobility of ultrasound devices, a small-sized low-

frequency (1 – 100 kHz) cymbal transducer has been developed. Using the cymbal

transducer, the efficacy of transdermal insulin delivery has been evaluated within in vivo

experiments on rats, rabbits, and large pigs (Lee et al. 2004; Lee, Newnham, & Smith

2004; Park, Werner, & Smith 2007). This transducer has a compact design with a

thickness of less than 2 mm and a weight of less than 3 grams. Multiple cymbal

transducers can be patterned side by side into 2 × 2 and 3 × 3 arrays. Arranging the

transducer patterns in such a way increases the spatial acoustic intensity fields for drug

transport. As the specific aims of this study, in vivo experiments were designed to

demonstrate the feasibility of noninvasive transdermal insulin delivery and glucose

sensing in large animals as a milestone to a future clinical application, and feasibility of a

Page 17: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

7

closed-loop feedback-controlled ultrasound system using cymbal transducers. In addition,

the study to answer the essential questions of transdermal insulin delivery was included.

1.3 Dissertation Outlines

In chapter 2, background information about alternative methods of glucose

monitoring, insulin delivery, and closed-loop between the monitoring and delivery are

presented based on literature reviews. Acoustic theories of the cymbal transducer and

potential mechanism of ultrasonic transdermal method are described in chapter 3.

Chapter 4 presents detailed descriptions of the construction procedures of cymbal

transducers, glucose sensors, and feedback controller for the ultrasound system.

Details of in vivo experiments on the transdermal insulin delivery, ultrasonic glucose

monitoring, and the regulation of glucose level by the closed-loop feedback

controlled ultrasound system are explained in chapter 5. In chapter 6, the results of

each in vivo experiment as well as the measured acoustic properties of the cymbal

array are presented. Chapter 7 provides the conclusions drawn from each task with in

vivo experiments. In addition, fundamental resonance frequencies of the cymbal

transducer, the step-by-step construction procedures of the cymbal arrays, comparison

of the efficacy of transdermal insulin delivery to intraperitonial injection, and a

program for the feedback controller used in this study are provided in the appendix.

Page 18: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Chapter 2

BACKGROUND - LITERATURE REVIEW

As an alternative to overcoming the compliance negatively impacted by the

current diabetes treatment (i.e., finger pricking and multiple injections of insulin), several

studies have presented positive results of ultrasonic transdermal insulin delivery and

extraction of glucose in the interstitial fluid (ISF). In this section, researches on

transdermal drug delivery and glucose sensing were revised. In addition, several studies

of a feedback controlled system for diabetes care were revised.

2.1 Transdermal Drug Delivery

One alternative to noninvasive drug administration is transdermal drug delivery,

which provides advantages over traditional injection drug delivery methods. Several

noninvasive methods exist for transdermal drug delivery, including chemical mediation

using liposomes and chemical enhancers or physical methods such as microneedles,

iontophoresis, electroporation, and ultrasound (Montorsi et al. 2000; Prausnitz 1997;

Sershen & West 2002; Tezel, Sens, & Mitragotri 2002; Whelan 2002). However,

currently, few drugs, proteins, or peptides have been successfully administered

transdermally for clinical applications due to the low skin permeability of these relatively

Page 19: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

9

large molecules, a problem attributed to the stratum corneum (e.g., the outermost skin

layer). One hypothesis related to transdermal drug delivery is that, once the drug has

traversed the stratum corneum, the next layer is easier to cross meaning the drug can thus

reach the capillary vessels to be absorbed(Mitragotri et al. 1995; Tezel, Sens, &

Mitragotri 2002). However, at this time, even with these enhancing mechanisms,

transdermal administration of drugs for clinical applications remains limited.

Recent reviews have demonstrated that ultrasound-mediated transdermal drug

delivery offers promising potential for noninvasive drug administration(Mitragotri 2005;

Pitt, Husseini, & Staples 2004; Smith 2007; Tachibana & Tachibana 2001). Researchers

have suggested that the working principle of ultrasonic transdermal drug delivery stems

from cavitation (Guzman et al. 2003b; Mitragotri, Edwards, Blankschtein, & Langer

1995; Mitragotri, Blankschtein, & Langer 1997b; Schlicher et al. 2006). Low frequency

ultrasound is capable of generating microbubbles in water and tissue, which allow water

channels to be produced within lipid bilayers. The resulting disorder created in the

stratum corneum facilitates the crossing of a hydrophilic drug or molecule. To this end,

the number of drugs and compounds shown to transdermally cross skin via ultrasound

continues to increase (Smith 2007). In vivo studies using ultrasound in transdermal drug

delivery showed that the skin permeability was enhanced for up to 6 hours. The

enhanced skin permeability is listed in Table 2.1 for each ultrasound operating condition.

Page 20: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

10

Table 2.1: Enhanced skin permeability by ultrasound.

Ultrasound Drug /

Animals fre-quency

mode intensity exposure

time Permeability

(a) Insulin / rabbit

105 kHz

50% / 10 sec

repetition time

5 kPa 90 min

Glucose level was recovered after 6 hours from the beginning of experiments.

(b) D-[3H]Mannitol / pigs (17-20 kg)

20 kHz

50% / 10 sec

repetition time

6.5 W/cm2 5 min Mannitol was delivered up to 4.5 hours from the beginning of experiments.

(c) Insulin / hairless

rats 20 kHz

20% / 8 sec

reputation time

5 W/cm2 15 min

Glucose level decreased and kept in hypoglycemic condition up to 3 hours of experiments.

(d) Insulin / hairless

rats 20 kHz

20% / 8 sec

reputation time

5 W/cm2 15 min

Glucose level decreased and kept in hypoglycemic condition up to 3 hours of experiments.

(e) Insulin / hairless

rats 20 kHz

40% / 8 sec

repetition time

2.5 W/cm2 15 min

Glucose level decreased and kept in hypoglycemic condition up to 2.5 hours of experiments.

(f) Insulin / hairless

rats 20 kHz

10% / 1 sec

repetition time

125 and 225

mW/cm2 60 min

Glucose level decreased and kept in hypoglycemic condition up to 4 hours from the beginning of ultrasound exposure.

(g) Salicylic

acid / hairless

rats

20 kHz

10% / 1 sec

repetition time

125 mW/cm2

60 min

The amount of salicylic acid recovered in urine was increased until the end of the 2.5 hours experiments.

(h) Insulin / hairless

rats 20 kHz 10%

225 mW/cm2

30 min

Glucose level decreased and kept in hypoglycemic condition up to 1.5 hours from the beginning of ultrasound exposure.

Page 21: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

11

Noninvasive ultrasonic transdermal delivery of insulin has generated significant

public interest due to the increasing number of people affected by diabetes. In the 20 to

105 kHz frequency range, enhanced transport in the presence of ultrasound has been

shown in both in vitro and in vivo experiments (Boucaud et al. 2000; Boucaud et al. 2002;

Kost 2002; Mitragotri, Blankschtein, & Langer 1995c; Smith et al. 2003; Smith 2007;

Tachibana 1992b; Zhang, Shung, & Edwards 1996). Many such early experiments were

performed using an ultrasound sonicator, ultrasonic bath, or commercial transducer. For

example, investigators demonstrated effective in vivo transport of insulin at 48 kHz using

an ultrasonic bath and at 105 kHz using a commercially obtained transducer (Tachibana

1992a). However, the major drawback thus far in exploiting ultrasound for noninvasive

drug delivery is the large size and poor mobility of the ultrasound device. Commercial

sonicators are large, heavy, table-top devices specifically designed for lysis of cells,

catalyzation of reactions, creation of emulsions, or cleaning purposes.

In transdermal insulin delivery using mechanical (Mitragotri, Edwards,

Blankschtein, & Langer 1995; Nordquist et al. 2007), electrical (Murthy et al. 2006),

chemical enhancers (Sintov & Wormser 2007), or the combination of these enhancers

(Chen et al. 2009), researchers monitored insulin concentrations to determine how

effectively insulin was delivered through the skin by their methods. Researchers also

measured glucose level as an indicator of insulin delivery.

Using ultrasound at 20 kHz, in vitro experiments were performed to investigate

transdermal insulin delivery and insulin concentrations were measured either by

radioimmunoassay (RIA) or enzyme-linked immunosorbent assay (ELISA) (Mitragotri,

Blankschtein, & Langer 1995b). In the in vitro study of iontophoresis-driven insulin

Page 22: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

12

transport through microneedle-induced skin microchannels, insulin concentrations were

measured by high performance liquid chromatography (HPLC) (Chen, Zhu, Zheng, Mou,

Wan, Zhang, Shi, Zhao, Xu, & Yang 2009). Insulin concentrations were also measured

by HPLC during the in vitro experiments of enhanced permeation flux of insulin using

trypsin as a biochemical enhancer (Li et al. 2008). In vitro experiments of

electroporation used fluorescein-isothiocyanate (FITC) insulin to investigate the

enhancement of transdermal transport by measuring fluorescence (Connolly, Cotton, &

Morin 2002). During the in vivo study of enhanced insulin delivery by co-administrating

short synthetic peptide, insulin concentrations were measured by immunoradioassay

(Chen et al. 2006). Using immunoradiometric assay (IRMA), insulin concentrations were

measured in transdermal insulin delivery using microneedles in in vivo condition

(Martanto et al. 2004).

While insulin concentrations were measured in both in vitro and in vivo studies,

glucose concentrations were widely used as the indicator of transdermal transport in in

vivo studies. For example, during the in vivo experiments using ultrasound, blood

glucose levels were monitored every 30 minutes (Boucaud, Garrigue, Machet, Vaillant, &

Patat 2002; Mitragotri, Blankschtein, & Langer 1995c). In transdermal insulin delivery

studies using needle-free liquid jet injectors (Arora et al. 2007), microneedles (Prausnitz

2004), and biochemical enhancers (Li, Quan, Zang, Jin, Kamiyama, Katsumi, Yamamoto,

& Tsutsumi 2008), blood glucose levels were measured periodically.

Monitoring insulin concentrations in serum directly answers the questions of

whether insulin is delivered through the skin and how much insulin is transported by

enhancers including ultrasound. It also provides insulin delivery rates at the given

Page 23: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

13

conditions of enhancing methods. Ultimately, measuring insulin concentrations in serum

will supply the information for the insulin dose guideline of each method, and this

guideline will provide essential information for developing an insulin delivery system.

2.2 Glucose Monitoring

Measuring glucose in interstitial fluid has shown to be an alternative to blood

glucose measurements. One device used to measure the glucose concentration of

interstitial fluid is a subcutaneous sensor, such as the MiniMed Continuous Glucose

Monitoring System (Medtronic, Minneapolis, MN) (Mastrototaro 1999). Depending on

the model, the subcutaneous sensor can measure the glucose level every one to five

minutes. However, it has some drawbacks, such as the need to replace sensors every

three to seven days and the interaction of biological components, causing fibrosis (Garg

et al. 2006; Weinstein et al. 2007; Wong et al. 2006). Devices using microdialysis also

measure the glucose concentration of interstitial fluid (Heinemann 2003). Despite the

advantage of providing sufficiently precise measurements, such devices are either only

available to physicians or too bulky to wear during daily activities. Additional possible

methods of glucose sensing such as impedance spectroscopy (Caduff et al. 2003), reverse

iontophoresis (Sekkat et al. 2002), optical technique using near-infrared fluorescence

(Ballerstadt et al. 2004), and tear glucose monitoring have been underdeveloped.

Page 24: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

14

2.3 Closed-loop Control for Diabetes Care

The importance of tightly controlled blood glucose concentrations has been well

recognized in the management of diabetes. The publication on the landmark Diabetes

Complications and Control Trial (DCCT) describes improved glycemic control as having

the potential to reduce the risk of long term complications such as chronic renal failure,

retinal damage, nerve damage, cardiovascular disease, and microvascular damage(The

Diabetes Control and Complications Trial Research Group (DCCT) 1993). As a

longstanding goal in diabetes care, a system that can automatically restore near-

physiologic glycemic control, a so-called ‘artificial pancreas’, has been studied for over

30 years(Bessman & Schultz 1972; Felts 1973; Kumareswaran, Evans, & Hovorka 2009).

Such a system requires three main elements: continuous glucose monitoring, insulin

delivery, and a controller that regulates the insulin delivery.

Continuous insulin delivery using an infusion device called an insulin pump was

introduced in the 1970s, and it is commercially available for Type 1 diabetes(Jeitler et al.

2008; Volpe et al. 2009; Weissberg-Benchell, ntisdel-Lomaglio, & Seshadri 2003). In a

recent comprehensive review involving a meta-analysis, it was concluded that severe

hypoglycemic rates were markedly lower with continuous subcutaneous insulin infusion

versus multiple daily injection therapy(Pickup & Renard 2008). Continuous

subcutaneous insulin infusion (CSII) is effective in managing brittle or unstable Type 1

diabetes. However, it has drawbacks such as infections that can occur at the site of

needle insertion. Also, after changing infusion sets every three days, flow problems tend

to occur(Cohen & Shaw 2007).

Page 25: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

15

Research in glucose monitoring has resulted in the development of the

continuous glucose monitoring (CGM) system(Cox 2009; Hanaire et al. 2009; Ryan &

Germsheid 2009; Skyler 2009). The CGM system measures glucose concentrations in

the interstitial fluid every one to five minutes using sensors inserted subcutaneously.

Although the CGM system has shown a benefit in glucose control, it has some limitations

such as the need to replace sensors every three to seven days and the interaction of

biological components causing physiological clots(Garg, Zisser, Schwartz, Bailey,

Kaplan, Ellis, & Jovanovic 2006; Weinstein, Bugler, Schwartz, Peyser, Brazg, &

McGarraugh 2007; Wong, Buckingham, Kunselman, Istoc, Leach, & Purvis 2006). Also,

it is still an invasive technique, requiring the subcutaneous insertion of sensors.

Currently, CSII and CGM techniques are used as an open-loop system that

requires patients or caretakers to interpret the glucose information(Kowalski 2009). To

close the loop between the two systems so as to have the systems provide the

interpretation, several algorithms have been studied and evaluated either in in silico

conditions(Marchetti et al. 2008; Patek et al. 2007) or in hospital conditions(Renard

2008; Weinzimer et al. 2008). The evaluation results of clinical trials have demonstrated

the feasibility of glucose control in 24 – 48 hours. While these evaluations show positive

results in the development of a closed-loop system for diabetes care, the technique relies

on subcutaneous injection of needles and, though minimal, is still invasive. To address

the limitations of subcutaneous insertion of CSII and CGM systems, ultrasound has been

used to transdermally extract glucose in the interstitial fluid(Cantrell, McArthur, &

Pishko 2000; Lee et al. 2005; Mitragotri et al. 2000; Park et al. 2009) as well as to

Page 26: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

16

transport insulin(Lee, Snyder, Newnham, & Smith 2004; Mitragotri 2005; Park, Werner,

& Smith 2007; Pitt, Husseini, & Staples 2004).

Page 27: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Chapter 3

BACKGROUND - ACOUSTIC THEORY

3.1 Design of cymbal transducer

A cymbal transducer, defined as a piezoelectric electroacoustic transducer, is

classified as a type V flextentional transducer. A cymbal transducer capable of producing

a very low frequency has a compact, lightweight structure. Based on the dimensions and

materials of the transducer, the cymbal transducer has an adjustable frequency ranging

between 1 and 100 kHz (Maione et al. 2002; Newnham, Xu, & Yoshikawa 1991;

Newnham, Xu, & Yoshikawa 1994).

In the design of a cymbal transducer, the lead zirconate titanate (PZT) ceramic

disk is sandwiched between metal endcaps, maintaining a shallow cavity beneath the

surface. The radial motion of the ceramic is converted into the axial displacement of the

endcaps, which is the fundamental vibration mode of the transducer. A cross-sectional

diagram of the cymbal transducer is provided in Figure 3.1. The metal endcap has a

cavity radius ( cR ), a cavity depth ( cd ), dimple diameter ( dR ), and thickness ( et ). As a

standard design with a 1 mm thick PZT disk, two brass endcaps of cR = 3 mm, cd = 0.32

mm, dR = 9 mm, and et =0.25 mm were used to construct a circular cymbal transducer.

Page 28: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

18

3.2 Acoustic Characteristics of Cymbal Transducer

In two-dimensional finite element models of the cymbal transducers, the metal

caps and the piezoelectric element were considered. In addition, an acoustic fluid as a

medium was considered in the models. Based on the finite element models, the

resonance frequencies of the cymbal transducer in air and water were determined.

The fundamental vibration mode of the endcap shapes corresponding to the (0, 1)

vibration mode of the endcap and the equation of motion of the endcap is very similar to

that of a thin plate:

Figure 3.1: Diagram of cymbal transducer in cross sectional view.

)()( 00 KrBIKrAJr , (Equation 3.1)

Cavity

Cavity

PZTBondinglayers

Endcaps

Cavity

Cavity

PZTBondinglayers

Endcaps

Page 29: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

19

where A and B are unknown constants, is the displacement of the endcap across its

radius ( r ), K is elastic constant of the plate, 0J is the zero order Bessel function and 0I

is the modified zero order Bessel function (Kinsler et al. 1982). The solution of the

equation of motion is given by Equation 3.2:

where rw is the endcap displacement across the radius, the C ’s are unknown constants

and is defined by the endcap material density ( e ), the endcap thickness ( et ), flexural

modulus of elasticity ( D ), the Young’s modulus ( E ) and the endcap radius ( eR ) as

Equation 3.2 is re-written by applying a rigid clamping boundary condition around the

periphery of the endcap ( ar ):

Since 1C is the unknown constant, proportionality can be derived from this equation

instead of exact solutions. Based on the proportionality and some algebra, the

fundamental resonance frequency ( arf , ) in the air is derived from Equation 3.3 and

described by Equation 3.5:

where, d is the dimple diameter ( dR2 ) and is Poisson’s ratio (Tressler 1997).

30201 )()( CrICrJCrw , (Equation 3.2)

2

24

e

eee

DR

Et

D

t

. (Equation 3.3)

)(

)()()()()(

1

100001 aI

aJaIrIaJrJCrw

. (Equation 3.4)

222,

1

1

1~

ede

arR

Ef

, (Equation 3.5)

Page 30: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

20

If the transducer is operated in water or another medium, the resonance frequency

changes due to the extra mass, which dampens the amplitude of resonance and decreases

the frequency at resonance. In water or another medium, the resonance frequency is

given as Eq. 3.6:

where w,rf is the in-water resonance frequency, a is the radius of the cymbal transducer,

and w is the water density (kg/m3).

The fundamental resonance frequency is affected by the type and thickness of

PZT disk, the material and thickness of the endcap, the depth and radius of cavity, and

the bottom radius of the cavity. For different PZT types and endcap dimensions, the

resonance frequency in air or water is presented in the Appendix A.

3.3 Acoustic Cavitation

Ultrasound-mediated transdermal method is based on the hypothesis that, once the

drug has been transported through the outermost layer of skin (i.e., stratum corneum), the

next layer is easier to cross, enabling the drug to reach the capillary vessel. Such an

assumption is based on the facts that ultrasound can permeate the skin and that the

glucose level of interstitial fluid is highly correlated to the blood glucose level. Although

ultrasound is known to enhance transdermal drug or protein delivery, the working

mechanisms have yet to be completely understood. The cavitation effect, which is one of

ee

w

arwr

t

a

ff

7885.0

3

81

,,

, 3.6

Page 31: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

21

the bioeffects of ultrasound, has been suggested as the primary mechanism for enhancing

the efficacy of ultrasonic transdermal drug delivery. (Guzman, McNamara, Nguyen, &

Prausnitz 2003b; Mitragotri, Edwards, Blankschtein, & Langer 1995; Mitragotri,

Blankschtein, & Langer 1997a; Schlicher, Radhakrishna, Tolentino, Apkarian, Zarnitsyn,

& Prausnitz 2006).

3.3.1 Definition

Acoustic cavitation is defined as the generation and activities of gas- or vapor-

filled bubbles in response to large rarefractional pressure in an aqueous solution

(Leighton 1994). In addition to the excitation of gaseous bubbles already existing in the

solution, acoustic waves with large rarefractional pressure can create bubbles when the

pressure is great enough to overcome the tensile strength of the solution. Common

terminology distinguishes two types of cavitations: stable and transient cavitation

(Carstensen & Flynn 1982).

In stable cavitation, which represents the oscillation of bubbles, the bubble radius

varies about an equilibrium value and violent collapse does not occur as the acoustic

intensity and frequency are not high enough to cause such a collapse. However,

microstreaming can occur near the boundaries as a result of the conversion of absorbed

acoustic energy into the flow (Leighton 1994). Such microstreaming can produce sheer

forces in tissue that have a rigid boundary. If the sheer forces are great enough to disrupt

cells, they can enhance skin permeability. Meanwhile, transient cavitation, which

oscillates in an unstable manner of the equilibrium radius, occurs at greater acoustic

Page 32: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

22

pressure. Transient cavitation includes the violent collapsing of bubbles after a few

cycles of oscillation. As the result of the violent collapsing of transient cavitation,

significant energy is transformed to create a high energy jet stream, which can disrupt the

cells (Mitragotri & Kost 2004).

3.3.2 Cavitation Threshold

Although several thresholds may be involved in acoustic cavitations, Neppiras

defined two thresholds, the stable cavitation threshold and transient cavitation threshold.

Thresholds are described in terms of acoustic pressure amplitude ( AP ), ambient pressure

( 0P ), initial bubble radius ( 0R ), and angular frequency of applied ultrasound ( ).

During refraction, the gas flow rate into a bubble ( dtdm / ) that includes surface

tension is given by

where D is the diffusion constant, C is the gas concentration in the liquid in the

absence of the bubble, and is surface tension (Neppiras 1980a). Similarly, the gas

flow rate out of the bubble during compression is given by

A

A

PRP

PRDC

dt

dm

000

21

3

8 , (Equation 3.7)

000

0

214

C

C

PRRDC

dt

dm , (Equation 3.8)

Page 33: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

23

where 0C is the saturation gas constant in the liquid. When the gas flow during both

phases of pressure cycles is equal, a threshold condition is reached and it is described by

Equation 3.9

The threshold is more exactly described by considering damping factor ( ), imposed

angular frequency ( ) and resonant angular frequency ( 0 ) (Neppiras 1980a; Neppiras

1980b):

where 0

220

2

0

2

3 P

R

. While the resonance of a stable bubble for frequencies up

to 100 kHz occurs at (Minnaert 1933)

where is the specific heat ratio of the gas inside the bubble, the resonance above 100

kHz is defined by taking into account the surface tension:

Known as the ‘Blake’ threshold (Blake 1949), the threshold of transient cavitation

for small bubbles (well below resonance size) is described by

1

000

2

0

211

2

3

PRC

C

P

PA . (Equation 3.9)

2222

1

000

2

0

)1(2

112

3

PRC

C

P

PA (Equation 3.10)

2/1

0

00

31

P

R, (Equation 3.11)

2/1

00

0

00 3

)13(21

31

PR

P

R

(Equation 3.12)

Page 34: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

24

For bubbles have greater size than resonance size, this threshold becomes

where rR is the resonant radius and 0/ PPp A .

1

00

30

30

232

3

4

RPRPPT

. (Equation 3.13)

,)1(3

21)1(46.0

3/12/10

ppp

R

R

r

(Equation 3.14)

Page 35: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Chapter 4

MATERIALS AND METHODS - APPRATUS

4.1 Construction of Cymbal Transducer Arrays

Based on the design and basic theory of the cymbal transducers described in

sections 3.1 and 3.2, three types of transducer arrays were constructed using the circular

and rectangular cymbal transducers. The design and fabrication procedures of the

cymbal transducer are protected by US Patents 5729007 and 6232702 (Newnham &

Dogan 1995; Newnham & Zhang 1999). This section gives the detail descriptions of the

assembly and acoustical evaluation of the cymbal transducers and arrays and presents an

acoustical evaluation of them.

4.1.1 Fabrication of cymbal arrays

The piezoelectric disk having a diameter of 12.7 mm and a thickness of 1 mm was

made from PZT-4 (Piezokinetics, Inc., Bellefonte, PA). To allow electricity to be

conducted uniformly through the piezoelectric disk, both axial surface of the disk were

covered with silver electrodes (Figure 4.1). Before bonding the disk to metal endcaps,

the electrodes were slightly scraped with 500 grit sandpaper (3M, St. Paul, MN) to

remove an accumulated oxidation layer and expose a clean silver surface. After scraping

the oxidation layer, residual dust on the surface of the disk was removed using acetone.

Page 36: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

26

The metal endcaps with the shape of a musical cymbal were made of 0.25 mm

thick brass. To obtain flat and circular disk with a diameter of 12.7 mm, purchased brass

sheet (0.25 mm thick) was punched using a manual punching set (Precision Brand,

Downners Grove, IL). The punched brass disk was molded to have a cymbal shape by

pressing it at 60 kpsi in a hydraulic press. The diagram in Figure 4.2 shows the

dimensions of the endcaps having the standard cavity depth, 0.32 mm (Tressler 1997).

For symmetry, the cavity depth of each pressed endcap was measured at the center of the

cap by using a Starrett dial gauge. Then, caps with the closest cavity depth were paired

together.

Figure 4.1: PZT-4 disk and a brass endcap

Page 37: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

27

Two cymbal shape endcaps were bonded to the piezoelectric disk using

Eccobond® 45 LV epoxy base and 15 LV (Emerson & Cuming, Billerica, MA) catalyst

with a 3:1 ratio. To increase the electrical conductivity between the caps and the

piezoelectric disk, silver powder (<250 μm, Sigma-Aldrich Corp., St. Louis, MO) was

added to the epoxy mixture in a 1:4 ratio. The epoxy mixture was carefully spread on the

edge of the disk, and then the caps were placed on the disk. To ensure a better seal,

binder clips were placed along the edges of the transducer. After binding caps on the

disk, the cymbal transducer was placed at room temperature for 24 hours.

While the cymbal transducer was cured, a 1 mm thin polyurethane sheet was

constructed for a cymbal array template. A two part polyurethane (Por-A-Mold 2160

A/B, PathwayPolymers, Chattanooga, TN) was thoroughly mixed in 1:1 volume ratio.

Figure 4.2: Dimensions of endcaps and PZT-4 disk

Page 38: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

28

To minimize air bubbles, the polyurethane mixture was degassed and carefully poured

onto a glass plate coated with a mold release (ease release 200, Mann Formulated

Products, Easton, PA). The mixture was pressed by another glass plate with 1 mm thick

spacers at each corner and left for 24 hours for curing. To construct the cymbal

transducer array in 2 × 2 and 3 × 3 patterns, the polyurethane sheet was cut in a size of 30

× 30 mm2 and 50 ×50 mm2. Four holes of 12.7 mm diameter were punched on the sheet

for the 2 × 2 array and nine holes were punched for the 3 × 3 array. The polyurethane

sheet was also used as a spacer when potting transducer arrays.

After placing the cymbal transducers into the template, a conductive adhesive

(silver epoxy, E-Solder No. 3021, VonRoll ISOLA, New Haven, CT) and flexible,

nonmagnetic, stranded wires (26 AWG 66/44, Pegasus Cable, Inc., Wilmington, DE)

were used to connect the circular cymbal transducers. To ensure a proper electrical

connection, the transducers were left to dry overnight. Once the transducers and wires

were connected, they were encased in the same polyurethane used for the cymbal

template. After mixing and degassing to avoid capturing air bubbles, the polyurethane

mixture was carefully poured in a silicon compound mold for each array shape. Once the

transducers are submerged in the polyurethane mixture, the array was left for 24 hours to

cure. The 2 × 2 and 3 × 3 cymbal arrays are shown in Figure 4.3. The step-by-step

instruction of the cymbal array construction is described in appendix B.

Page 39: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

29

4.1.2 Construction of matching circuits

In order to obtain maximum power transfer from the operating system to the

transducer arrays that have input electrical impedance, it is necessary to match the

impedance of the transducer to the output impedance of the operating system. In this

study, a π-type network circuit was used for the impedance matching to a 50 Ω output

impedance of amplifiers. Theoretical values of C1, C2, and L1 were obtained using a

MATLAB code for measured impedance value of each array. After getting theoretical

values, actual circuits shown in Figure 4.4 were constructed to obtain the maximum

acoustic output by adjusting the value of each component (C1, C2, and L1).

Figure 4.3: Lightweight cymbal arrays were constructed using four transducers in a two-by-two pattern, (left) or nine cymbal transducers in a three-by-three pattern (right). The two-by-two array was used for ultrasonic glucose sensing, while the three-by-three array was used in the transdermal insulin delivery.

Page 40: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

30

4.2 Exposimetry: Acoustic Characteristics of Cymbal Arrays

According to the exposimetry guidelines established by the American Institute of

Ultrasound in Medicine (AIUM 1993), the acoustic pressure field of the array was

determined with a calibrated miniature omnidirectional reference hydrophone (Model

TC4013, S/N: 5199093, RESON, Inc., Goleta, CA). The cymbal array was submerged in

a water tank (51 x 54 x 122 cm3) which was made almost anechoic by placing 1.27 cm

thick rubber sound absorbing material around its wall (Figure 4.5). To avoid cavitation

effects, the tank was filled with distilled water degassed by a custom made degasser to

reduce the dissolved oxygen content to 1-2 ppm. The position of the hydrophone was

precisely controlled by a Velmex Positioning System (Velmex Inc., East Bloomfield,

NY). In order to obtain the acoustic pressure profile over a plane 2 mm away from the

array surface, automatic scannings with 1 mm step size were performed by the computer

Figure 4.4: Electric impedance matching network using a π-type circuit. After constructing the circuit based on the theoretical values, actual value of each componentwas adjusted to have maximum acoustic output.

Page 41: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

31

controlled exposimetry positioning system. The automatic scanning was performed over

the area of 50 × 50 mm2. Based on three scannings, spatial peak-temporal peak (Isptp)

intensities of each array in a mean and standard deviation were determined by the

Equation 4.1 .

Maximum voltage amplitude of signal ( opV ) received by hydrophone were converted

into maximum pressure amplitude ( opP ) by the sensitivity of the hydrophone ( A ). With

the density ( ) of 998 kg/m3 and the speed of sound ( c ) of 1482 m/s, Isptp was

determined for each array.

cA

V

c

PI opop

sptp 2

22

22 . (Equation 4.1)

Figure 4.5: Equipment setup of the exposimetry to characterize the cymbal arrays. Acoustic pressure field of each cymbal array was measured in the water tank filled withdegassed water.

Cymbal Array Agilent

33250A

AR

25A250

Ch1 Ch2

Oscilloscope Agilent 54622A

Matching circuit

Hydrophone Reson TC4013

Preamp Reson VP1000

Water tank

Page 42: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

32

To drive the array, a radio frequency (RF) signal was generated by a

function/arbitrary waveform generator (33250A, Agilent, Santa Clara, CA) and amplified

by an RF amplifier (Model 25A250, Amplifier Research, Souderton, PA). Using an

external tuning network, the electrical impedance of the array was tuned to the output

impedance of the amplifier. The pulse duration and repetition period of the RF signal

were monitored using a digital oscilloscope (54622A, Agilent, Santa Clara, CA). In

order to determine electrical input condition to the cymbal array with the matching circuit,

input voltage to the array was measured for each array by using an electrical test probe

(100×, Tek P5100, Tektronix, Inc., Beaverton, OR) with the exposimetry setup.

4.3 Fabrication and Calibration of Electrochemical Glucose Sensors

As part of the noninvasive glucose monitoring system, disposable biosensors are

used in the current study. Biosensors using electrochemical reaction have been used to

monitor changes in physiological parameters. Based on the electrochemical properties,

enzyme-coated biosensors can be used as glucose sensors.

In the presence of a glucose oxidase enzyme (GOx), glucose reacts with oxygen

(O2) and water (H2O) to produce gluconic acid and hydrogen peroxide (H2O2) (Wilkins &

Atanasov 1996):

Glucose + O2 + H2O (GOx) Gluconic acid + H2O2.

When a positive electric potential is applied in the presence of platinum as a catalyst, two

hydrogen peroxide molecules are oxidized into oxygen, water, and four electrons:

2H2O2 (Pt) 2O2 + 2H2O + 4 e-.

Page 43: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

33

The current produced in this reaction is proportional to the concentration of hydrogen

peroxide, which is proportional to the glucose concentration.

4.3.1 Fabrication of glucose sensors

In this study, the biosensors for the transdermal glucose sensing were fabricated

based on the electrochemical reaction of enzyme coated electrodes. For the biosensors,

polymer thick film electrodes were acquired from Conductive Technologies Inc. (York,

PA). The biosensor consists of three electrodes: working, counter and reference

electrodes (Figure 4.6). In order to get the current from the electrochemical reaction, the

working electrode should be deposited with platinum which worked as a catalyst. Once

platinic acid of 25 mM (Sigma-Aldrich Corp., St. Louis, MO) in 0.1 M NaCl solution

was placed over three electrodes, a voltage of -0.7 V between working and counter

electrodes was applied for 60 minutes using a potentiostat (Model 283, Princeton Applied

Research, Oak Ridge, TN). After the washing the platinum deposited electrodes with

deionized distilled water, glucose oxidase (1,000,000 unit/g, Sigma-Aldrich Corp., St.

Louis, MO) was immobilized on the three electrodes by polyethylene glycol (PEG)

hydrogel. After PEG diacrylate (MW = 575, Aldrich Corp., St. Louis, MO) was diluted

with distilled water to produce a 20% solution, glucose oxidase (GOx) was mixed. After

adding liquid photoinitiator (2-hydroxy–2-methyl-1-phenyl-1-propanone, Ciba Specialty

Chemicals, Tarrytown, NY) the PEG-GOx solution, three electrodes were covered with

the solution and exposed to 6 W, 365 nm UV light for 20 seconds in air.

Page 44: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

34

4.3.2 Calibration of glucose sensors

To determine the unknown glucose concentration of interstitial fluid, each

biosensor was calibrated using a concentration curve created by a series of glucose

solutions, 0, 50, 100, 150, 200 and 300 mg/dl. The glucose solutions were prepared by

dissolving glucose (Dextrose Anhydrous, VWR, West Chester, PA) in 0.9 % phosphate

buffered (PBS) solution. Once the three electrode part of the biosensor was submerged in

the calibration solution and connected to the potentiostat, a voltage of 0.7 V was applied

to the working electrode against the counter electrode. The voltage applied to the

working electrode was increased from 0 V to 0.7 V with 10 mV/sec increasing ratio. To

minimize the residual effect of the glucose solution, the biosensors was gently washed

Figure 4.6: Photograph of an enzyme based electrochemical biosensor which consists of three electrodes: working/sensing, counter, and reference electrodes

Counter electrode (Carbon)

Working electrode

(Platinized Carbon)

Reference electrode (Silver/ Chloride)

Page 45: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

35

with deionized distilled water and the calibration was performed from the lower glucose

concentration to the higher concentration.

4.4 Closed-Loop Feedback Controlled Ultrasound System for Insulin Delivery and Glucose Sensing

A simple proportional-plus-integrator-plus-derivative (PID) controller has been

used in this study. There are three parts to the PID controller: the proportional gain PK ,

the integral gain IK , and the derivative gain DK . These three values determine how the

controller responds to error in the current state of the system ( G ) with respect to the

desired state ( refG ). In a closed-loop system, the tracking error ( E ) is the difference

between the current output and the desired output. After the tracking error is sent to the

PID controller as its input (Figure 4.7), the output signal of the controller (U ) will be

obtained as the summation of the proportional gain multiplied by the error, the integral

gain multiplied by the integral of the error, and the derivative gain multiplied by the

derivative of the error:

This signal will be sent to the plant and the new output will be obtained. The new output

will be sent back again to find the new error. For a PID controlled system, increasing the

proportional gain factor PK , decreases the rise time but oscillations increase. By

increasing the integral gain factor IK , the noise tolerance improves but the settling time

dt

dEKdt EKKU DIP . (Equation 4.2)

Page 46: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

36

lengthens. The damping is increased and any overshoot is reduced by increasing the

derivative gain factor DK .

The ultrasound systems for the transdermal glucose sensing and insulin delivery

were combined through a proportional closed-loop feedback controller. In the controller,

the reference glucose concentration, reference glucose concentration was set as 115

mg/dl. The difference between the reference and measured values was defined as an

error and used as the input of the controller. Using the constant of proportionality of the

system, PK set as 1, the input on/off signal U was sent to the ultrasound systems. Using

a control program written in MATLAB® (Mathwork, Inc., Natick, MA), the ultrasound

systems were operated automatically. After the reference glucose concentration and the

exposure time of the ultrasonic glucose sensing system were set, the repetition times of

the measurements and the total number of measurements were set. Once the control

program was started, the glucose sensing system was operated and the first glucose level

measurement was taken by converting measured current value based on the calibration

curve of each biosensor. Depending on the glucose levels, the insulin delivery system

Figure 4.7: Block diagram of a proportional-plus-integral-plus-derivative (PID) controller. The variables are defined in table 1. The plant is a system controlled by thecontroller. In this study, it is the ultrasound system for insulin delivery and glucosemonitoring.

Controller

KP, KI, KD

Plant U

+ E

G

Gref

Page 47: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

37

either operated or stopped. In this study, the ultrasound exposure time for glucose

sensing was set at 5 minutes, and the measurements were performed every 20 minutes for

120 minutes. The definitions and values of the variables in Figure 4.7 are listed in Table

4.1.

Table 4.1: To control the ultrasound system using cymbal arrays, a simple proportionalcontroller was used.

Symbol Definition

refG Desired blood glucose concentration, 15 mg/dl

E Error between the measured and reference values,

refGGE

PK Proportional gain factor, 1

IK Integral gain factor, 0

DK Derivative gain factor, 0

U Input to the system, EKU P

G Measured blood glucose concentration

Page 48: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Chapter 5

MATERIALS AND METODS – IN VIVO EXPERIMENTS

To evaluate the cymbal transducers in transdermal insulin delivery and glucose

sensing, several in vivo experiments were performed in large pigs. This section decribes

each in vivo experiments for (1) the feasibility of transdermal insulin delivery, (2) the

feasibility of using the cymbal array in ultrasonic glucose sensing , and (3) the feasibility

of glucose control by a feedback controlled ultrasound system using the cymbal arrays.

As a milestone to a preclinical application, in vivo experiments were performed in large

pigs possessing a weight and blood volume similar to humans.

All procedures described in this report involving live animals were approved by

the Institutional Animal Care and Use Committee (IACUC) at the Pennsylvania State

University. Each animal was pre-anesthetized for intubation with a combination of

ketamine hydrochloride (10 - 12 mg/kg intramuscularly, Ketaject®, Phoenix, St. Joseph,

MO) and sodium xylazine (1 - 2 mg/kg intramuscularly, Xyla-Ject®, Phoenix, St. Joseph,

MO). Pigs were fitted with an intravenous (IV) catheter in the auricular vein and an

endotracial tube (size 6-7) was inserted into the airway. Anesthesia throughout the

remaining experiment was maintained to surgical depth via inhalant isoflurane

(IsothesiaTM, Abbott Laboratories, North Chicago, IL) using an inhalational anesthesia

unit (Narkovet Deluxe, North American Drager, Telford, PA).

Page 49: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

39

5.1 Transdermal Insulin Delivery

Six Yorkshire pigs (100 – 140 lbs) obtained from the Penn State Swine Center

were divided into two experimental groups. As the control (n = 3), the first group did not

receive any ultrasound exposure with the insulin while second group (n = 3) was treated

with ultrasound and insulin. Under the anesthesia, the axillary area of the pigs were

shaved using an electric shaver, and a depilatory agent was applied to the skin of both

groups, control and exposure, to eliminate any remaining hair. With the pig in the lateral

recumbency (Figure 5.1), a 1 mm thick, water-tight standoff was attached between the

skin and the 3 × 3 array using tissue glue (Vetbond®, 3M, St. Paul, MN). The reservoir

within standoff was filled with insulin (Humulin® R, rDNA U-100, Eli Lilly and Co.,

Indianapolis, IN) through a small hole in the back of the array for both the control and

exposure experiments. To prevent disruption of ultrasound transmission, care was taken

to remove any bubbles from the solution in the reservoir (standoff) between the axillary

area and the array.

Page 50: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

40

At the beginning of the experiment, blood sample (0.3 ml) was collected from the

ear vein of each pig for a baseline glucose level analysis. The glucose level (mg/dl) in

the blood was determined using ACCU-CHEKTM blood glucose monitoring system

(Roche Diagnostics Co., Indianapolis, IN, USA). Multiple blood samples (2-4 each time)

were taken every 15 minutes for 90 minutes. Elapsed time from the initial injection of

the ketamine-xylazine until the first glucose measurement was no greater than 15

minutes. For comparison between the pigs, the change in the blood glucose level was

normalized to a baseline with respect to each animal's initial blood glucose recording at 0

minutes.

For comparison between the results of each pig, the change of blood glucose level

was normalized with respect to the initial blood glucose level of each pig. The control

group (three pigs) used insulin inside the reservoir without ultrasound exposure while the

Figure 5.1: Photograph of a transdermal insulin delivery experiment with pig placed in alateral recumbent position with the array attached.

ArrayArrayArray

Page 51: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

41

second group (three pigs) was treated with ultrasound and insulin at 30 kHz with an Isptp

= 100 mW/cm2 for 60 minutes. For both groups, the standoff reservoir with the insulin or

saline was removed at 60 minutes although glucose determination continued until 90

minutes from the start. At the end of the experiments, the pig was euthanized

(Pentobarbitol, Fatal Plus, 130 mg/kg IV, Vortech Pharmaceuticals, Ltd., Dearborn MI)

under anesthesia.

Statistical analysis was performed using Microsoft Excel® (Microsoft Corp.,

Redmond, WA). The blood glucose verses time data were pooled for each group and

analyzed as its mean and standard error (x ± s.e.). An ANOVA was used to analyze the

statistical significance of the differences among the means of groups. The p-value was

used to determine if the between-group differences are significantly greater than chance.

For all the data, a single or double asterisk was used if the p-value is less than the 0.05 or

0.01 level of significance, respectively.

To answer how much insulin is transported through the skin, additional

transdermal insulin delivery experiments were performed at 20 kHz. Six pigs were

divided into a control and ultrasound exposure groups. Blood samples of 5 ml were

collected every 15 minutes for 90 minutes to analyze insulin concentrations as well as

glucose levels. At the completion of the experiments, each pig was euthanized

(Pentobarbitol, Fatal Plus, 130 mg/kg IV, Vortech Pharmaceuticals, Ltd., Dearborn MI)

under anesthesia.

The concentrations of insulin in the blood samples were analyzed by enzyme

linked immunoassay (ELISA) tests. In the blood samples, both human insulin and

porcine insulin were found and there is 99% similarity between the two types of insulin.

Page 52: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

42

Therefore, the blood samples were analyzed for both human insulin and porcine insulin

using Mercodia ELISA kits (Insulin ELISA, 10-1113-01 and Porcine Insulin ELISA, 10-

1200-01, Mercodia, Uppsala, Sweden). The insulin ELISA is a solid-phase two-site

enzyme immunoassay and is based on the direct sandwich technique in which two

monoclonal antibodies are directed against separate antigenic determinants on the insulin

molecule. During incubation, insulin in the sample reacts with peroxidase-conjugated

anti-insulin antibodies and anti-insulin antibodies bound to the microtitration well. A

simple washing step removes the unbound enzyme-labeled antibody. The bound

conjugate is detected by reaction with 3, 3', 5, 5'-tetramethylbenzidine (TMB). The

reaction is halted by adding acid to give a colorimetric endpoint that is read

spectrophorometrically. After preparing the enzyme conjugate solution, wash buffer,

and sufficient microplate wells that can accommodate calibrators and samples in

duplicate, 25 µl of calibrators and samples were pipetted into the microplate wells. Then,

100 µl of enzyme conjugate solution was added to each well and the plate was incubated

on a plate shaker (700 - 900 rpm) for one hour at room temperature (18 - 25 °C). The

incubated plate was washed six times with an automatic plate washer. After the final

wash, the plate was inverted and firmly tapped against absorbent paper to remove the

remaining washing solution. During the next step, 200 µl of substrate TMB was added

into each well and the plate was incubated for 15 minutes at room temperature. After the

incubation, 50 µl of stop solution was added into the well and the plate was placed on a

shaker for 5 minutes to ensure thorough mixing. Within 30 minutes after the final step,

the plate was placed in an ELISA reader and the optical density of each well was read at

450 nm and the insulin concentration of each well was calculated. Since there was a

Page 53: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

43

maximum of 28% cross-reaction between human and porcine insulin for the ELISA tests,

additional calculation of the insulin concentration was performed to compensate for the

cross-reaction.

5.2 Ultrasonic Transdermal Glucose Sensing

For perspective ultrasonic noninvasive glucose monitoring, eight Yorkshire pigs

(~ 200 lbs) obtained from The Pennsylvania State University Swine Center were used in

five groups according to the ultrasound exposure conditions. After anesthesia, hair on the

axillary area of pigs in lateral recumbency was removed and the 2 × 2 cymbal array with

double layers of a 1 mm thick and water-tight standoff made of plastic was place on the

skin. Figure 5.2 shows the placement of array and the biosensors on the pig. Four

biosensors were placed between two layers of the standoff and the reservoir within the

standoff was filled with 0.9% PBS solution. The volume of PBS solution was 2.25 ml.

Since it was assumed that the concentration of glucose in the collection chamber depends

on the volume of PBS in the reservoir, the volume was kept as 2.25 ml for all

experiments. The experiments were performed on the same location of each pig.

Page 54: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

44

To permeabilize the skin for extraction of glucose in interstitial fluid, the cymbal

array was operated at a frequency of 20 kHz with an intensity of Isptp = 100 or 50

mW/cm2 for 5, 10 or 20 minutes (Table 5.1). After connecting the biosensors to the

potentiostat at the end of the ultrasound exposure, the current from the electrochemical

reaction on the biosensors was measured by applying a voltage of 0.7 V to the working

electrode against the counter electrode. Using the potentiostat (Model 283, Princeton

Figure 5.2: (a) Layout of the glucose sensor incorporated with the cymbal transducerarray placed on the skin surface. (b) Photograph of the ultrasonic glucose sensing on a pig with the array attached to the skin and the reservoir filled with PBS solution.

Pig

Cymbal Array

Saline

Biosensors

2 mm

Pig

Cymbal Array

Saline

Biosensors

2 mm

Cymbal Array

Biosensors

Cymbal Array

Biosensors

Cymbal Array

Biosensors

(b)

(a)

Page 55: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

45

Applied Research, Oak Ridge, TN), the voltage applied to the working electrode was

increased from 0 V to 0.7 V with 10 mV/sec increasing ratio. Any false currents caused

by interference were observed. Using the potentiostat control program (Electrochemistry

PowerSuite™ v. 2.1.1, Princeton Applied Research, Oak Ridge, TN), the current was

recorded on a computer. Based on the biosensor calibration results and the linear

relationship between the current and the glucose concentration, the glucose levels in

interstitial fluid were determined. For the comparison to the glucose levels determined

using biosensors, blood samples were collected at the same time when the currents from

the biosensors were measured. The glucose level of blood sample was measured by

ACCU-CHEKTM blood glucose meter (Roche Diagnostics Co., Indianapolis, IN) and i-

STAT portable clinical analyzer (MODEL 200, Abbott Laboratories, Abbott Park, IL).

Table 5.1: The detail information of ultrasound exposure conditions.

Acoustic intensity, Isptp Exposure time (minutes)

20

10 100 mW/cm2

5

10 50 mW/cm2

5

Page 56: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

46

5.3 Closed-Loop Feedback Controlled Ultrasound System for Glucose Control

Four Yorkshire pigs (~ 200 lbs) were obtained from the evaluation of the closed-

loop feedback controlled ultrasound system. After anesthesia, hairs on the axillary area

and medial area of the front leg of the pig in lateral recumbency were removed.

Figure 5.3 shows the layout of the glucose control system and the experimental setup.

For extracting glucose in interstitial fluid, the 2 × 2 array with double layers of a 1 mm

thick and water-tight standoff was attached to the skin on the medial area of the front leg

using tissue glue (Vetbond®, 3M, St. Paul, MN). Two biosensors were placed between

two layers of the standoff and connected to the potentiostat to obtain current

measurements. For extracting glucose in interstitial fluid, the 2 × 2 array with double

layers of a 1 mm thick and water-tight standoff was attached to the skin on the medial

area of the front leg using tissue glue (Vetbond®, 3M, St. Paul, MN). Two biosensors

were placed between two layers of the standoff and connected to the potentiostat to

obtain current measurements. The interstitial fluid reservoir within the standoff was filled

with 0.9 % PBS solution through a small hole in the back of the array. To deliver insulin

transdermally, the 3 × 3 array with a single layer of the standoff was attached to the skin

on the axillary area, and the reservoir was filled with insulin (Humulin® R, rDNA U-100,

Eli Lilly and Co., Indianapolis, IN).

Page 57: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

47

Figure 5.3: (a) The layout of the glucose sensing system incorporated with the insulin delivery system as placed on the skin surface. After five minutes of ultrasound exposure, the biosensor determined the glucose concentration of the interstitial fluid. Based on theglucose level, the insulin delivery system was automatically operated. (b) Photograph of the ultrasound systems placed on the pig for insulin delivery and glucose sensing. Theultrasound system was controlled by a closed-loop controller.

(a)

(b)

Page 58: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

48

Figure 5.4 shows the experimental setup with a closed-loop feedback controlled

ultrasound system using cymbal arrays. To permeabilize the skin for glucose extraction,

a RF signal of 20 kHz was generated, amplified, and sent to the 2 × 2 cymbal array in the

glucose sensing system (Figure 5.4 solid line). The array was operated at a 20% duty

cycle (a pulse duration of 200 ms and a pulse repetition period of 1 second) with an

intensity (Isptp) of 100 mW/cm2 for 5 minutes. At the end of the ultrasound exposure, the

current from the electrochemical reaction on the biosensors was measured by applying a

voltage of 0.7 V to the working electrode against the counter electrode. Using the

potentiostat control program (Electrochemistry PowerSuite™ v. 2.1.1, Princeton Applied

Research, Oak Ridge, TN), the current was recorded in a computer. According to the

biosensor calibration results and the linear relationship between the current and glucose

concentration, the glucose levels in the interstitial fluid were determined. Based on the

glucose level determined by the biosensor, the error between the reference and measured

values was calculated. According to the calculated error, the insulin delivery system

using the 3 × 3 cymbal array (Figure 5.4 dashed line) was either turned on or off. If the

glucose level was determined to be higher than the reference value, 115 mg/dl, the

control program turned on the arbitrary waveform generator to generate a RF signal at 30

kHz. The amplified signal was sent to the 3 × 3 cymbal array through an impedance

matching network. For the insulin delivery, the array was operated at a 20% duty cycle

with Isptp = 100 mW/cm2. After 15 minutes of ultrasound exposure for the insulin

delivery, the glucose sensing with the biosensor was performed again. Using the same

procedure, the glucose level was determined, and the ultrasound system for the insulin

delivery was operated. The closed-loop control of the glucose level was performed for

Page 59: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

49

120 minutes. Every 20 minutes for the 120 minutes period, the biosensors determined the

glucose level in the interstitial fluid. Additionally, for comparison with the glucose levels

that were determined using the biosensors, the glucose levels of blood samples were also

measured by the ACCU-CHEKTM blood glucose meter (Roche Diagnostics Co.,

Indianapolis, IN) and the i-STAT portable clinical analyzer (MODEL 200, Abbott

Laboratories, Abbott Park, IL). The results of the glucose meter tests were presented in a

mean and standard error.

Figure 5.4: Schematic diagram shows the experimental setup with the closed-loop feedback controller for noninvasive glucose regulation. The ultrasound operation forboth the glucose sensing and insulin delivery systems was automatically operated.

Page 60: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Chapter 6

RESULTS

6.1 Acoustic Pressure Distribution of Cymbal Transducer Arrays

From multiple exposimetry scans, the intensity of the 2 × 2 array was obtained as

Isptp = 100.8 ± 0.9 mW/cm2 while the intensity of the 3×3 array was Isptp = 100.0 ± 0.5

mW/cm2. Spatial distributions of temporal peak intensities of the arrays were plotted in

dB with respect to the highest intensity (Isptp) to show spatial distribution of the intensity

field from the maximum radiation point (Figure 6.1). Using similar Isptp driving

conditions, the intensity field was determined in a plane 1 mm from the array face for

both the rectangular and the circular designs. Figure 6.1 shows typical two dimensional

scanning plots of the temporal peak intensity for the 2 × 2 and 3 × 3 cymbal arrays over a

50 × 50 mm2 area with 1 mm steps. The color bar on the side of the graph illustrates the

normalized temporal peak intensity change in dB. In order to obtain the intensity of Isptp

= 100.0 mW/cm2, the 2 × 2 cymbal array was operated at the input voltage of 613 ± 12.5

V (peak-to-peak) while the 3 × 3 cymbal array was operated at 1.03 ± 0.05 kV. Both

arrays had the spatial peak intensity near the center of each array. In order to estimate the

power consumption of each array, small resister (1 Ω) was added between the amplifier

and the matching circuit. Currents to the 2 × 2 and 3 × 3 cymbal arrays, which were

determined by measuring the voltages across the small resister, were 525 ± 11 mA and

1.25 ± 0.01 A, respectively.

Page 61: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

51

6.2 Transdermal Insulin Delivery

6.2.1 In Vivo Experiments in Large Pigs

Results of the ultrasonic transdermal insulin delivery in large pigs for the two

groups are graphed in Figure 6.2 as change in the blood glucose level during the 90

minute experiment in terms of the mean and standard error. After the pigs were

anesthetized, the average initial glucose level of the six pigs was 146 ± 13 mg/dl.

Generally for pigs, blood glucose level is approximately 100-110 mg/dl (Danfaer 1999;

Pond & Houpt 1978). As mentioned, for a comparison between the pigs, the change in

Figure 6.1: Acoustic intensity profiles of the two-by-two (left) and three-by-three (right) arrays were determined by measuring acoustic pressure and plotted in dB. The spatial-peak-temporal-peak intensities were 100.8 ± 0.9 mW/cm2 for the two-by-two array and 100.0 ± 0.5 mW/cm2 for the three-by-three array.

(a) (b)

Page 62: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

52

the blood glucose level was normalized to a baseline which was the initial glucose level

for each pig.

For the control group (insulin without ultrasound), the glucose level increased to

31 ± 21 mg/dl compared to the initial baseline over the 90 minute experiment. The slope

of this increase was +20 mg/dl/hour (r2 =0.9). In contrast with the ultrasound exposure

group (insulin with ultrasound), the glucose level decreased to –74 ± 5 mg/dl at 60

minutes and continued to decrease to –91 ± 9 mg/dl at 90 minutes after the standoff was

removed. Additionally, a gross examination of the pig's skin was performed after

exposure to look for visible lesions on the skin surface. Visual examination of the post

Figure 6-2: Change of blood glucose level in the ultrasound mediated transdermal insulin delivery on pigs.

146 ±13

mg/dl

***

* * *

Control, insulin-no ultrasound (n = 3)Ultrasound with insulin (n = 3)

Ch

an

ge

of

blo

od

glu

cose

lev

el (

mg

/dl)

50

0

-50

-100

-150150 45 7530 60 90

Time (minutes) Standoff removed

146 ±13

mg/dl

***

* * *

Control, insulin-no ultrasound (n = 3)Ultrasound with insulin (n = 3)

Ch

an

ge

of

blo

od

glu

cose

lev

el (

mg

/dl)

50

0

-50

-100

-150

146 ±13

mg/dl

146 ±13

mg/dl

***

* * *

***

* * *

Control, insulin-no ultrasound (n = 3)Ultrasound with insulin (n = 3)Control, insulin-no ultrasound (n = 3)Ultrasound with insulin (n = 3)Ultrasound with insulin (n = 3)

Ch

an

ge

of

blo

od

glu

cose

lev

el (

mg

/dl)

50

0

-50

-100

-150

Ch

an

ge

of

blo

od

glu

cose

lev

el (

mg

/dl)

50

0

-50

-100

-150

50

0

-50

-100

-150150 45 7530 60 90

Time (minutes)

150 45 7530 60 90150 45 7530 60 90

Time (minutes) Standoff removed

Page 63: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

53

ultrasound exposed skin did not indicate any noticeable damage or significant change to

the skin.

To determine the statistical significance between the results in Figure 6-2 of the

two groups at the 15 minute increment time points of the experiment, an ANOVA was

used to analyze the data. Asterisks above a glucose-time data point were used if a

statistical difference existed between the control and exposure. At 15 minutes the

analysis showed no statistical difference between the groups. However, a comparison of

the control to the ultrasound exposure groups at 30 minutes and greater indicated that the

results were statistically significant at a p-value of 0.05 or better.

6.2.2 Analysis of Insulin Concentration in Transdermal Insulin Delivery

Uncompensated human insulin concentrations (µU/l) of serum samples in a mean

and standard deviation are shown in Figure 6.2a with porcine insulin concentrations

(ng/l). After 15 minutes of ultrasound exposure, the human insulin concentration

increased only by 0.5 µU/l from the initial value, 40.3 µU/l. However, at 30 minutes, the

concentration increased by 23.7 µU/l for 15 minutes. After 30 minutes, human insulin

concentrations were obtained in the range of 41.8 - 48.6 µU/l. For 90-minute periods,

porcine insulin concentrations stayed in the range of 12.23-12.38 ng/l. In order to

compare the change of porcine concentrations, the human insulin concentrations were

determined in ng/l by unit conversion (MW = 5808 Da, 1 IU = 6 nmol) (Newnham, Xu,

& Yoshikawa 1991). Due to the small value, the standard deviation of human insulin

concentration is not clearly presented in the plot. Only one different amino acid is found

Page 64: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

54

between human and porcine insulin (Berson & Yalow 1963). Therefore, there is a

maximum of 28% cross-reaction for ELISA tests between the two types of insulin.

Based on this finding that initial insulin concentration of serum samples is solely due to

the porcine insulin, the cross-reaction ratio of 13.3% was obtained. With the cross-

reaction ratio, the compensated human insulin concentrations were determined. Figure

6.2b shows the changes in porcine insulin concentrations with the changes of human

insulin concentrations for both with and without compensating for the cross-reaction.

The results of comparison between two insulin concentrations indicate that human insulin

was transported through the skin due to the ultrasonically enhanced skin permeability.

To compare the change of glucose levels to the insulin concentrations, the total

insulin concentrations (human and porcine insulin) in ng/l were presented in Figure 6.3

with the change of blood glucose level in mg/dl over 90-minute experiments. For the

rapid increase of insulin from 15 minutes to 30 minutes, the glucose level decreased at

higher rate. As the insulin concentration increased and stayed higher than the initial

value, the glucose level continuously decreased.

Page 65: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

55

Figure 6.2: (a) The concentrations of human (, µU/l) and porcine (◊, ng/l) insulin were determined for 90-minutes periods. While porcine insulin concentrations stayed in therage of 12.23 - 12.38 ng/l, human insulin concentration increased by 0.5 and 23.7 µU/l after 15 30 minutes ultrasound exposure, respectively. (b) Human insulin concentrationswere determined in ng/l by unit conversion and compensated for the cross-reaction between porcine and human insulin for ELISA tests. The cross-reaction ratio was obtained at 13.3%.

0

10

20

30

40

50

60

70

0 15 30 45 60 75 90

Time (minutes)

12.0

12.2

12.4

12.6

12.8

13.0

Human insulinPorcine insulin

Hu

man

in

sulin

co

nce

ntr

atio

n (

µU

/l) P

orcin

e insu

lin co

ncen

tration

(ng

/l)

-2.0

0.0

2.0

4.0

6.0

8.0

10.0

12.0

14.0

16.0

1 2 3 4 5 6 7

Time (minutes)

Porcine

Human (uncompensated)

Human (compensated)

Insu

lin c

on

cen

trat

ion

(n

g/l)

(a)

(b)

Page 66: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

56

6.2.3 Ultrasonic Glucose Sensing

Based on the concentration curve previously created for each biosensor, the

glucose concentrations of the interstitial fluid were determined using the current

measured by the biosensors. For each exposure condition, the results of ultrasonic

glucose sensing using the cymbal array and biosensor system were graphed and

compared to the results of a conventional glucose meter in Figure 6.4.

Figure 6.3: The total insulin concentrations in ng/L were presented with the change of blood glucose level in mg/dl over 90-minute experiments. As insulin concentrations increased faster, the glucose level decreased at higher rate.

-120

-100

-80

-60

-40

-20

0

0 15 30 45 60 75 90

Time (minutes)

0

2

4

6

8

10

12

14

16

18

GlucoseInsulinC

han

ge

of

glu

cose

(m

g/d

l) In

sulin

con

centratio

n (µ

U/l)

Page 67: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

57

With an acoustic intensity of Isptp = 100 mW/cm2, the glucose level determined by

the biosensors after 20 minutes of ultrasound exposure was 175 ± 7 mg/dl while the level

measured by the conventional glucose meter was 166 ± 5 mg/dl. After 10 minutes of

ultrasound exposure, the glucose levels determined by the biosensor and glucose meter

were 152 ± 9 mg/dl and 130 ± 3 mg/dl, respectively. In a comparison to the glucose

level, 131 ± 5 mg/dl, measured by the conventional glucose meter, the glucose

concentration of the interstitial fluid was 127 ± 16 mg/dl which was determined by the

biosensors after 5 minutes of ultrasound exposure. The glucose levels of the interstitial

fluid after the ultrasound exposure with the intensity of Isptp = 50 mW/cm2 for 10 and 5

minutes were determined by the biosensors as 140 ± 10 mg/dl and 82 ± 15 mg/dl,

Figure 6.4: The glucose concentrations of the interstitial fluid determined using theultrasound system and a conventional glucose meter. Asterisks above the data bar wereused if there was no statistical difference between the ultrasonic glucose sensing method and the conventional method based on the ANOVA with the 0.05 or 0.01 level ofsignificance.

0

20

40

60

80

100

120

140

160

180

200

20 kHz, 20 min,100 mW/cm2

20 kHz, 10 min,100 mW/cm2

20 kHz, 5 min,100 mW/cm2

20 kHz,10 min,50mW/cm2

20 kHz, 5 min, 50mW/cm2

Glu

co

se

le

ve

l [m

g/d

l]Biosensor

Glucose meter

frequency: exposure time:

Isptp:

20 kHz10 min 100 mW/cm2

20 kHz5 min 100 mW/cm2

20 kHz 20 min 100 mW/cm2

20 kHz5 min 50 mW/cm2

20 kHz 10 min 50 mW/cm2

** ** ** **

*

Page 68: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

58

respectively. Compared to the results of the biosensors, the glucose levels measured by

the conventional glucose meter after 10 and 5 minutes exposure were 124 ± 4 mg/dl and

125 ± 4 mg/dl, respectively.

In Figure 6.4, asterisks above the data bar were used if there was no statistical

difference between the ultrasonic glucose sensing method and the conventional method

based on the ANOVA with the 0.05 or 0.01 level of significance. For the intensity of Isptp

= 100 mW/cm2, the glucose concentrations determined by the ultrasound system were

statistically the same as the value determined by the glucose meter. In addition, the

results of the 10 minute ultrasound exposure at the intensity of Isptp = 50 mW/cm2 showed

no difference between two measurements. At the end of each experiment, the skin area

that was exposed to the ultrasound was examined for the skin reaction and no skin

irritation was found from all experiments.

6.2.4 Glucose control in large pigs using a closed-loop feedback controlled ultrasound system

For all four pigs designated as pig #1 – #4, the currents from the biosensors were

measured every 20 minutes for a 120-minute period. Based on the concentration curve

previously created for each biosensor, the glucose concentration of interstitial fluid was

determined. From the comparison of the glucose levels, the glucose concentrations

determined by the biosensors were up to 10 mg/dl higher than those measured by the

glucose meter (Figure 6.5).

Using the ultrasound and the biosensors, the initial glucose level of pig #1,

determined as 117 mg/dl, was lower than the reference level, 115 mg/dl (Figure 6.5a).

Page 69: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

59

Therefore, the ultrasonic insulin delivery system was initially turned on for insulin

delivery. After 20 minutes of experiments, the glucose level was 106 mg/dl, and the

ultrasonic transdermal delivery system was turned off. Additional 20 minutes after the

experiments, the insulin delivery system was turned on again since the glucose level was

116 mg/dl and turned off again 80 minutes after the experiments because of lower

glucose concentration than the reference value. For pig #2, the insulin delivery system

was turned on at the second glucose measurement; 40 minutes after the experiment had

started (Figure 6.5b). By operating the insulin delivery system for 40 minutes, the

glucose level decreased to 104 mg/dl. During the period from 60 to 120 minutes

following the experiments, the glucose level was controlled within 104 – 116 mg/dl by

the automatically operated insulin delivery system. The initial glucose level of pig #3

was 137 mg/dl, and the insulin delivery system was automatically operated (Figure 6.5c).

After 80 minutes of transdermal insulin delivery from the beginning of the experiments,

the glucose level decreased to 107 mg/dl and, after the delivery system was turned off,

increased to 127 mg/dl. The glucose level of pig #4 was regulated in the range of 109 –

130 mg/dl by the ultrasonic glucose sensing and the insulin delivery system that was

automatically controlled by the closed-loop controller (Figure 6.5d). Without the

operation of the insulin delivery system, the glucose level increased by 9-10 mg/dl for 20

minutes(Park, Werner, & Smith 2007).

Page 70: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

60

Figure 6.5: (a) – (d) Glucose concentrations of four pigs, which were regulated by the ultrasonic transdermal glucose sensing and insulin delivery systems combined by a feedback controller.

Case #1

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

on

Case #1

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

onon

Case #2

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

on

offoff

on

Case #2

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

onon

offoffoffoff

onon

(b)

(a)

Page 71: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

61

Figure 6.5: (continued)

Case #3

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

on

offon

Case #3

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

onon

offoffon

Case #4

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

onon

off

off

on

Case #4

80

90

100

110

120

130

140

0 20 40 60 80 100 120

Time (min)

Glu

cose

Lev

el (

mg

/dl)

BiosensorGlucose meter

onononon

offoff

offoff

onon

(d)

(c)

Page 72: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Chapter 7

CONCLUSION

7.1 Transdermal Insulin Delivery

The goal of this task was to determine if a light-weight, low-profile ultrasound

device based cymbal transducer could be used for in vivo transdermal insulin in animals

which approximate the size and weight of a human. Although commercial sonicators

have been admirable devices for demonstrating drug delivery, the ultrasonic probe or

converter from a sonicator can weigh a kilogram or more; the cymbal array weighs less

than 38 grams. Individual cymbals can be arranged into multi-element array designs

since this can increase the effective aperture of ultrasound area with respect to skin area

and some research indicates that the delivery dose increases with ultrasound exposure

area (Smith, Lee, Maione, Roy, McElligott, & Shung 2003). Interestingly, the cymbal

design originates from underwater research for naval applications and current research is

underway to incorporate existing battery technology in the miniaturization of portable

power.

Generally, xylazine causes hyperglycemia in rats and rabbits (Harkness &

Wagner 1995; Hillyer & Quesenberry 1997; Kawai, Keep, & Betz 1997; Pavlovic et al.

1996). In pigs, xylazine suppresses insulin release which results in the higher initial

average blood glucose level of 146 mg/dl pigs (Heim et al. 2002). The mean value for

normal blood glucose levels for pigs is approximately 108.5 mg/dl yet (Danfaer 1999) the

Page 73: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

63

use of the xylazine appears to have the continual effect of increasing the glucose level at

a rate of 20 mg/dl/hour even after the experiment started. This may indicate either the

lack of passive permeability of the pig's skin to the insulin or the sustained consequence

of the xylazine. Yet the useful effect of using these insulin-suppressed pigs was to

demonstrate the feasibility of reducing a high glucose level (≈150 mg/dl) to a normal

glucose level (below 100) albeit for a pig model. In contrast to the control group, the

ultrasound with insulin group showed a blood glucose level decrease of -91 mg/dl at 90

minutes compared to the baseline. Moreover, compared to the control at 90 minutes, the

difference between the two groups was approximately 120 mg/dl and the ANOVA

analysis indicated a statistical difference (p > 0.05) between the two groups at the time

points of 30 minutes and greater.

When fasting, glucose of 110-126 mg/dl is classified as impaired fasting glucose,

140-200 mg/dl is impaired glucose tolerance and greater than 200 mg/dl is considered

diabetic (Carnevale Schianca et al. 2003). In this case a diabetic person would need to

inject enough insulin to reduce his or her blood glucose by about 100 mg/dl.

Summarizing our previous ultrasound exposure results using hyperglycemic animals at

90 minutes, the glucose level continued to decrease to -296.7 ± 52.8 mg/dl in rats

and -208.1 ± 29 mg/dl in rabbits (Lee, Snyder, Newnham, & Smith 2004; Smith, Lee,

Maione, Roy, McElligott, & Shung 2003). Although the decrease in blood glucose in this

pig research was not as large as it was in the rats and rabbits, this may be due to the fact

that pigs have a far greater blood volume and body mass. Nevertheless the results

indicate that the array was capable of safely bringing the diabetic glucose level within the

normal range. In conclusion, the results herein demonstrate a promising pre-clinical

Page 74: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

64

outcome for the low profile cymbal array to be used for ultrasound enhanced in vivo

insulin transport using an animal model which mimics human use.

Ultrasound has been used as an enhancer for transdermal drug delivery, especially

for transdermal insulin transport (Boucaud, Garrigue, Machet, Vaillant, & Patat 2002;

Pitt, Husseini, & Staples 2004; Prausnitz & Langer 2008; Smith 2007). However,

concerns have been expressed that the insulin secreted by the subject animals might

solely control the change of glucose. In order to address this essential issue in ultrasonic

transdermal insulin delivery, an other task was designed to investigate how much insulin

was transported through the skin and the effects of transdermally delivered insulin on the

glucose control especially using the cymbal transducer array as a practically portable

ultrasound device. The cross-reaction compensated results of insulin concentrations from

90-minute experiments demonstrated that insulin is transdermally delivered due to the

ultrasound exposure and indicates that the glucose levels are controlled by the effects of

both delivered (human insulin) and secreted insulin (porcine insulin). In addition, these

results confirm transdermal insulin delivery by ultrasound and supply essential

information for developing a wearable ultrasound device for transdermal drug transport.

7.2 Ultrasonic Glucose Sensing

In several studies, researchers have demonstrated that ultrasound can permeabilize

skin to enhance the transdermal delivery of drugs (Mitragotri 2005; Park, Werner, &

Smith 2007; Pitt, Husseini, & Staples 2004; Tachibana & Tachibana 2001). The results of

cavitation have been suggested as the main mechanism of ultrasound mediated

Page 75: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

65

transdermal drug delivery although it has not been completely understood (Guzman,

McNamara, Nguyen, & Prausnitz 2003b; Guzman et al. 2003a; Mitragotri, Edwards,

Blankschtein, & Langer 1995; Schlicher, Radhakrishna, Tolentino, Apkarian, Zarnitsyn,

& Prausnitz 2006). In addition to the drug delivery, transdermal extraction of glucose in

the interstitial fluid by ultrasound has shown potential in becoming an alternative

technique of glucose measurement (Cantrell, McArthur, & Pishko 2000; Lee, Nayak,

Dodds, Pishko, & Smith 2005).

Based on the effects of ultrasound that enhances the skin permeability and the

mechanism of glucose diffusion due to the concentration differences, this research was

designed. Previously, in vivo experiments of transdermal glucose sensing on rats were

conducted using the light-weight cymbal transducer array and electrochemical biosensors

and the results have demonstrated the feasibility of using the cymbal array in ultrasonic

glucose measurement. Additionally, the results have shown a reliable accuracy of this

technique compared to the conventional method (Lee, Nayak, Dodds, Pishko, & Smith

2005).

In perspective to human application, the purpose of this research was to

demonstrate the feasibility of the cymbal array in ultrasonic glucose monitoring on large

pigs which have a similar size and weight to humans. As the initial study of in vivo pig

experiment, the results of the ultrasound exposure at the same exposure condition (Isptp =

100 mW/cm2, 20 minutes) used in a previous study, have shown that the ultrasound

system consisting of the cymbal array and biosensors was able to noninvasively

determine the glucose concentration on large animals having a similar size to humans. To

explore the reliability of the ultrasound system for glucose sensing, further experiments

Page 76: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

66

have been performed with different ultrasound exposure conditions. From the statistical

analysis, the results of ultrasound exposure with Isptp = 100 mW/cm2 for 10 and 5 minutes

have shown a reliable accuracy compared with the results of conventional glucose meter

measurements.

The results presented herein have demonstrated the feasibility of using the cymbal

transducer array for ultrasonic glucose sensing on large animals for preclinical

applications. Furthermore, the results of ultrasonic glucose measurements with 100

mW/cm2 (Isptp) have indicated that the ultrasound system using the cymbal array has been

able to determine the glucose concentration in short exposure time.

7.3 Closed-Loop Feedback Controlled Ultrasound System for Glucose Regulation

The challenge with diabetes care is frequent patient non-compliance due to the

unpleasant continuous process of finger pricking for blood glucose samples and daily

insulin injections. Noncompliance typically involves refusing treatment, reducing or

increasing insulin doses, or skipping medications. These actions are common problems

that can prevent the efficacy of treatment and reduce patient health. Eventually, non-

compliance leads to complications such as nerve damage, retinal damage, microvascular

damage, chronic renal failure, kidney failure, dental disease, and cardiovascular

disease(The Diabetes Control and Complications Trial Research Group (DCCT) 1993;

Wickramasinghe, Yang, & Spencer 2004). Overcoming non-compliance requires a

feedback-controlled active insulin delivery system that is noninvasive and easy to use.

Page 77: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

67

Current drug delivery technologies allow controlled dosing of a drug, but are

limited as they do not respond to actual biological status, meaning they lack a feedback

loop. To address this need, a paradigm shift is happening in diabetes care from the

current passive (one drug at a single dose over time) controlled- release to an active

delivery system that includes sensing and biofeedback. With it known that ultrasound

enhances transdermal drug delivery by permeabilizing skin to drugs(Mitragotri 2005;

Smith 2007; Smith 2008), the research is presented using ultrasound within an initial

study of a feedback-controlled active insulin delivery system with cymbal arrays that can

continuously monitor biological status and control insulin delivery in response to

changing glucose levels.

For the glucose sensing to occur, size was a consideration; a smaller sized array

was needed to minimize the dilution of glucose into the PBS solution, and a large enough

sized one was needed for ultrasound exposure to permeabilize the skin. To meet these

two requirements, the 2 × 2 cymbal array was used for the glucose sensing. Glucose

levels in previous in vivo experiments with rats and rabbits were determined by using 2 ×

2 arrays. The errors between the measurements by conventional glucose meters and the

ultrasonic method were -16% and -3% for rats and rabbits, respectively. In the studies of

using the cymbal array for the transdermal insulin delivery, a broader acoustic intensity

field without increasing the spatial-peak temporal-peak intensity (Isptp) was demonstrated

that performed glucose reduction better(Luis et al. 2007). Based on this result, the 3 × 3

cymbal array was used in the transdermal insulin delivery to expose a larger skin area to

ultrasound. Using the closed-loop feedback algorithm, the systems were combined and

Page 78: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

68

evaluated in in vivo experiments on four pigs. The glucose levels of each pig were

controlled for 120 minutes with a maximum error of 9%.

To develop an active ultrasound system that tightly controls glucose levels and

one that is practically portable so that patients can wear it on a daily basis, tone must

evaluate the portable design of transducers and the closed-loop feedback control system.

In addition to the design of cymbal transducer arrays that have previously shown

potential as portable ultrasound systems, the results presented herein demonstrated the

feasibility of the ultrasound systems using cymbal arrays for controlling glucose levels.

7.4 Future Work

The future work of this research includes (1) modifying the feedback controller

with a complex control algorithm that counts the changing rate of glucose concentrations,

(2) integrating the relationship between the ultrasound exposure time and the duration of

enhanced skin permeability into the controller, and (3) improving the biosensor design to

extend its lifetime. A simple proportional feedback controller was used in this study to

evaluate the feasibility of the cymbal array system to control glucose concentrations.

While the simple controller showed potential to close the ultrasonic glucose sensing and

insulin delivery systems, the controller responded only to the difference of glucose levels

from a reference value. In order to control rapid increment of glucose concentrations

such as post-meal glucose levels, a modification of the simple controller that counts the

changing ratio of glucose concentrations is necessary.

Page 79: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

69

Another consideration to improve the controller is integrating the relationship

between the insulin delivery dose and the ultrasound exposure time. In previous studies

of transdermal drug delivery, 15 minutes ultrasound exposure at 20 kHz with an intensity

of 2.5 - 5 W/cm2 showed enhanced skin permeability for 2.5 – 3 hours. Also, 5 minutes

ultrasound exposure at 6.5 W/cm2 resulted the skin permeable to insulin for 4.5

hours(Mitragotri, Blankschtein, & Langer 1995a). At lower intensities, 30 – 60 minutes

ultrasound exposure with an intensity of 125 – 225 mW/cm2 permeabilized the skin for

1.5 – 4 hours(Mitragotri, Blankschtein, & Langer 1995b; Prausnitz 1997). While these

studies demonstrated that ultrasound enhanced the skin permeability for several hours,

comprehensive studies of the relationship between ultrasound exposure time, the duration

of enhanced permeability, and the dose of delivered insulin have not been explored.

Therefore, quantitative studies of the insulin delivery dose as a function of ultrasound

exposure time is necessary for the modification of the controller to automatically

determine the operating time of the arrays in the system.

For more accurate and convenient glucose sensing, the design of biosensors

should be modified to stabilize the hydrogel as well as to extend the sensors lifetime of

sensors. The biosensors used in this study have several advantages that they are easy to

prepare, economical, and disposable. Three days after preparing the biosensors with

hydrogel, the biosensor could sense glucose concentration with a maximum error of 3%.

Although the biosensors had stable sensitivities, the lifetime of the biosensors was limited

due to the lack of hydrogel adhesion. The hydrogel capturing GOx can be easily

detached from the screen printed electrodes by bending motions. Therefore, technical

improvements on the sensor design are required to over come the limited lifetime. As the

Page 80: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

70

next step to an active ultrasound system, integration of the modifications in the

ultrasound system will promote the capability of a closed-loop feedback controlled

ultrasound system in favor of an active glucose control in diabetes care.

Page 81: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

BIBLIOGRAPHY

AIUM 1993, Bioeffects and safety of diagnostic ultrasound American Institute of Ultrasound in Medicine, Rockville, MD.

American Diabetes Association 2007, 2007 Annual Report: Individual in Impact, Broad in Reach.

Arora, A., Hakim, I., Baxter, J., Rathnasingham, R., Srinivasan, R., Fletcher, D. A., & Mitragotri, S. 2007, "Needle-free delivery of macromolecules across the skin by nanoliter-volume pulsed microjets", Proc.Natl.Acad.Sci.U.S.A, vol. 104, no. 11, pp. 4255-4260.

Ballerstadt, R., Polak, A., Beuhler, A., & Frye, J. 2004, "In vitro long-term performance study of a near-infrared fluorescence affinity sensor for glucose monitoring", Biosensors & Bioelectronics, vol. 19, no. 8, pp. 905-914.

Beaser, R. S. & Staff of Joslin Diabetes Center 2007, Joslin's Diabetes Deskbook, 2nd edn, Joslin Diabetes Center, Boston, MA.

Berson, S. A. & Yalow, R. S. 1963, "Antigens in Insulin Determinants of Specificity of Porcine Insulin in Man", Science, vol. 139, no. 3557, pp. 844-845.

Bessman, S. P. & Schultz, R. D. 1972, "Sugar electrode sensor for the "artificial pancreas"", Horm.Metab Res., vol. 4, no. 6, pp. 413-417.

Blake, F. G. 1949, The onset of cavitation in liquids. I. Cavitation threshold sound pressures in water as a function of temperature and hydrostatic pressure, Dept. of Engineering, Sciences and Applied Physics. Harvard University, Cambridge, MA, Tech. Mem. 12.

Boucaud, A., Garrigue, M. A., Machet, L., Vaillant, L., & Patat, F. 2002, "Effect of sonication parameters on transdermal delivery of insulin to hairless rats", J.Control Release, vol. 81, no. 1-2, pp. 113-119.

Boucaud, A., Tessier, L., Machet, L., Vaillant, L., & Patat, F. "Transdermal Delivery of Insulin using Low Frequency Ultrasound", in IEEE Ultrasonics Symposium, pp. 1453-1456.

Caduff, A., Hirt, E., Feldman, Y., Ali, Z., & Heinemann, L. 2003, "First human experiments with a novel non-invasive, non-optical continuous glucose monitoring system", Biosensors & Bioelectronics, vol. 19, no. 3, pp. 209-217.

Cantrell, J. T., McArthur, M. J., & Pishko, M. V. 2000, "Transdermal Extraction of Interstitial Fluid by Low-Frequency Ultrasound Quantified with 3H2O as a Tracer Molecule", Journal of Pharmaceutical Sciences, vol. 89, no. 9, pp. 1170-1179.

Page 82: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

72

Carnevale Schianca, G. P., Rossi, A., Sainaghi, P. P., Maduli, E., & Bartoli, E. 2003, "The significance of impaired fasting glucose versus impaired glucose tolerance: importance of insulin secretion and resistance", Diabetes Care, vol. 26, no. 5, pp. 1333-1337.

Carstensen, E. L. & Flynn, H. G. 1982, "The potential for transient cavitation with microsecond pulses of ultrasound", Ultrasound Med.Biol., vol. 8, no. 6, p. L720-L724.

Chen, H., Zhu, H., Zheng, J., Mou, D., Wan, J., Zhang, J., Shi, T., Zhao, Y., Xu, H., & Yang, X. 2009, "Iontophoresis-driven penetration of nanovesicles through microneedle-induced skin microchannels for enhancing transdermal delivery of insulin", J.Control Release, vol. 139, no. 1, pp. 63-72.

Chen, Y., Shen, Y., Guo, X., Zhang, C., Yang, W., Ma, M., Liu, S., Zhang, M., & Wen, L. P. 2006, "Transdermal protein delivery by a coadministered peptide identified via phage display", Nat.Biotechnol., vol. 24, no. 4, pp. 455-460.

Cohen, N. D. & Shaw, J. E. 2007, "Diabetes: advances in treatment", Intern.Med.J., vol. 37, no. 6, pp. 383-388.

Connolly, P., Cotton, C., & Morin, F. 2002, "Opportunities at the skin interface for continuous patient monitoring: a reverse iontophoresis model tested on lactate and glucose", IEEE Trans.Nanobioscience., vol. 1, no. 1, pp. 37-41.

Cox, M. 2009, "An overview of continuous glucose monitoring systems", J.Pediatr.Health Care, vol. 23, no. 5, pp. 344-347.

Danfaer, A. 1999, "Carbohydrate and Lipid Metabolism," in A Quantitative Biology of the Pig, I. Kyriazakis, ed., CABI Publishing, New York, NY, pp. 333-362.

El-Kamel, A. H., Al-Fagih, I. M., & Alsarra, I. A. 2008, "Effect of sonophoresis and chemical enhancers on testosterone transdermal delivery from solid lipid microparticles: an in vitro study", Curr.Drug Deliv., vol. 5, no. 1, pp. 20-26.

Felts, P. W. 1973, "Pancreas transplantation and the artificial pancreas: status of current experimental approaches", South.Med.J., vol. 66, no. 1, pp. 66-73.

Garg, S., Zisser, H., Schwartz, S., Bailey, T., Kaplan, R., Ellis, S., & Jovanovic, L. 2006, "Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor", Diabetes Care, vol. 29, no. 1, pp. 44-50.

Guzman, H. R., McNamara, A. J., Nguyen, D. X., & Prausnitz, M. R. 2003b, "Bioeffects caused by changes in acoustic cavitation bubble density and cell concentration: a unified explanation based on cell-to-bubble ratio and blast radius", Ultrasound Med.Biol., vol. 29, no. 8, pp. 1211-1222.

Page 83: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

73

Guzman, H. R., McNamara, A. J., Nguyen, D. X., & Prausnitz, M. R. 2003a, "Bioeffects caused by changes in acoustic cavitation bubble density and cell concentration: a unified explanation based on cell-to-bubble ratio and blast radius", Ultrasound Med.Biol., vol. 29, no. 8, pp. 1211-1222.

Hanaire, H., Dubet, A., Chauveau, M. E., Anduze, Y., Fernandes, M., Melki, V., & Ritz, P. 2009, "Usefulness of Continuous Glucose Monitoring for the Diagnosis of Hypoglycemia after a Gastric Bypass in a Patient Previously Treated for Type 2 Diabetes", Obes.Surg.

Harkness, J. E. & Wagner, D. J. 1995, The Biology and Medicine of Rabbits and Rodents Williams and Willkins, Baltimore MD.

Heim, K. E., Morrell, J. S., Ronan, A. M., & Tagliaferro, A. R. 2002, "Effects of ketamine-xylazine and isoflurane on insulin sensitivity in dehydroepiandrosterone sulfate-treated minipigs (Sus scrofa domestica)", Comp Med., vol. 52, no. 3, pp. 233-237.

Heinemann, L. 2003, "Continuous glucose monitoring by means of the microdialysis technique: underlying fundamental aspects", Diabetes Technol.Ther., vol. 5, no. 4, pp. 545-561.

Hillyer, E. & Quesenberry, K. E. 1997, Ferrets, Rabbits, and Rodents: Clinical Medicine and Surgery W.B. Saunders Co, Philadelphia PA.

Hodgson, B. B. & Robert J. 2006, Saunders Nursing Drug Handbook 2006 ELSEVIER SAUNDERS, St. Louis, Missouri.

Jeitler, K., Horvath, K., Berghold, A., Gratzer, T. W., Neeser, K., Pieber, T. R., & Siebenhofer, A. 2008, "Continuous subcutaneous insulin infusion versus multiple daily insulin injections in patients with diabetes mellitus: systematic review and meta-analysis", Diabetologia, vol. 51, no. 6, pp. 941-951.

Kawai, N., Keep, R. F., & Betz, A. L. 1997, "Hyperglycemia and the Vascular Effects of Cerebral Ischemia", Stroke, vol. 28, pp. 149-154.

Kinsler, L. E., Frey, A. R., Coppens, A. B., & Sanders, J. V. 1982, Fundamentals of Acoustics, 3rd edn, John Wiley & Sons, New York.

Kost, J. 2002, "Ultrasound-Assisted Insulin Delivery and Noninvasive Glucose Sensing", Diabetes Technology & Therapeutics, vol. 4, no. 4, pp. 489-497.

Kowalski, A. J. 2009, "Can we really close the loop and how soon? Accelerating the availability of an artificial pancreas: a roadmap to better diabetes outcomes", Diabetes Technol.Ther., vol. 11 Suppl 1, p. S113-S119.

Page 84: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

74

Kumareswaran, K., Evans, M. L., & Hovorka, R. 2009, "Artificial pancreas: an emerging approach to treat Type 1 diabetes", Expert.Rev.Med.Devices, vol. 6, no. 4, pp. 401-410.

Lance, L. L., Lacy, C. F., Goldman, M. P., & Armstrong, L. L. 2002, Quick Look Drug Book Lippincott Williams & Wilkins, Philadelphia, PA.

Lee, S., Nayak, V., Dodds, J., Pishko, M., & Smith, N. B. 2005, "Glucose measurements with sensors and ultrasound", Ultrasound in Medicine and Biology, vol. 31, no. 7, pp. 971-977.

Lee, S., Newnham, R. E., & Smith, N. B. 2004, "Short ultrasound exposure times for noninvasive insulin delivery in rats using the lightweight cymbal array", IEEE Trans.Ultrason.Ferroelectr.Freq.Control, vol. 51, no. 2, pp. 176-180.

Lee, S., Snyder, B., Newnham, R. E., & Smith, N. B. 2004, "Noninvasive ultrasonic transdermal insulin delivery in rabbits using the light-weight cymbal array", Diabetes Technol.Ther., vol. 6, no. 6, pp. 808-815.

Leighton, T. G. 1994, The Acoustic Bubble Academic Press, San Diego, CA.

Li, Y. Z., Quan, Y. S., Zang, L., Jin, M. N., Kamiyama, F., Katsumi, H., Yamamoto, A., & Tsutsumi, S. 2008, "Transdermal delivery of insulin using trypsin as a biochemical enhancer", Biol.Pharm.Bull., vol. 31, no. 8, pp. 1574-1579.

Luis, J., Park, E. J., Meyer, R. J., & Smith, N. B. 2007, "Rectangular cymbal arrays for improved ultrasonic transdermal insulin delivery", J.Acoust.Soc.Am., vol. 122, no. 4, pp. 2022-2030.

Maione, E., Shung, K. K., Meyer, R. J., Hughes, J. W., Newnham, R. E., & Smith, N. B. 2002, "Transducer design for a portable ultrasound enhanced transdermal drug delivery system", IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control, vol. 49, pp. 1430-1436.

Marchetti, G., Barolo, M., Jovanovic, L., Zisser, H., & Seborg, D. E. 2008, "An improved PID switching control strategy for type 1 diabetes", IEEE Trans.Biomed.Eng, vol. 55, no. 3, pp. 857-865.

Martanto, W., Davis, S. P., Holiday, N. R., Wang, J., Gill, H. S., & Prausnitz, M. R. 2004, "Transdermal delivery of insulin using microneedles in vivo", Pharm.Res., vol. 21, no. 6, pp. 947-952.

Mastrototaro, J. 1999, "The MiniMed Continuous Glucose Monitoring System (CGMS)", J.Pediatr.Endocrinol.Metab, vol. 12 Suppl 3, pp. 751-758.

Page 85: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

75

Meeuwisse-Pasterkamp, S. H., van der Klauw, M. M., & Wolffenbuttel, B. H. 2008, "Type 2 diabetes mellitus: prevention of macrovascular complications", Expert Rev.Cardiovasc.Ther., vol. 6, no. 3, pp. 323-341.

Minnaert, M. 1933, "On musical air-bubbles and the sounds of running water", Phil.Mag., vol. 16, pp. 235-248.

Mitragotri, S. 2005, "Healing sound: the use of ultrasound in drug delivery and other therapeutic applications", Nat.Rev.Drug Discov., vol. 4, no. 3, pp. 255-260.

Mitragotri, S., Blankschtein, D., & Langer, R. 1995a, "Ultrasound-Mediated Transdermal Protein Delivery", Science, vol. 269, no. 5225, pp. 850-853.

Mitragotri, S., Blankschtein, D., & Langer, R. 1995b, "Ultrasound-Mediated Transdermal Protein Delivery", Science, vol. 269, no. 5225, pp. 850-853.

Mitragotri, S., Blankschtein, D., & Langer, R. 1995c, "Ultrasound-Mediated Transdermal Protein Delivery", Science, vol. 269, no. 5225, pp. 850-853.

Mitragotri, S., Blankschtein, D., & Langer, R. 1997a, "An explanation for the variation of the sonophoretic transdermal transport enhancement from drug to drug", J.Pharm.Sci., vol. 86, no. 10, pp. 1190-1192.

Mitragotri, S., Blankschtein, D., & Langer, R. 1997b, "An explanation for the variation of the sonophoretic transdermal transport enhancement from drug to drug", J.Pharm.Sci., vol. 86, no. 10, pp. 1190-1192.

Mitragotri, S., Coleman, M., Kost, J., & Langer, R. 2000, "Analysis of ultrasonically extracted interstitial fluid as a predictor of blood glucose levels", Journal of Applied Physiology, vol. 89, pp. 961-966.

Mitragotri, S., Edwards, D. A., Blankschtein, D., & Langer, R. 1995, "A mechanistic study of ultrasonically-enhanced transdermal drug delivery", J.Pharm.Sci., vol. 84, no. 6, pp. 697-706.

Mitragotri, S. & Kost, J. 2004, "Low-frequency sonophoresis: a review", Adv.Drug Deliv.Rev., vol. 56, no. 5, pp. 589-601.

Montorsi, F., Salonia, A., Guazzoni, G., Barbieri, L., Colombo, R., Brausi, M., Scattoni, V., Rigatti, P., & Pizzini, G. 2000, "Transdermal electromotive multi-drug administration for Peyronie's disease: preliminary results", J.Androl, vol. 21, no. 1, pp. 85-90.

Murthy, S. N., Zhao, Y. L., Marlan, K., Hui, S. W., Kazim, A. L., & Sen, A. 2006, "Lipid and electroosmosis enhanced transdermal delivery of insulin by electroporation", J.Pharm.Sci., vol. 95, no. 9, pp. 2041-2050.

Page 86: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

76

Neppiras, E. A. 1980a, "Acoustic Cavitation", Physics Reports-Review Section of Physics Letters, vol. 61, no. 3, pp. 159-251.

Neppiras, E. A. 1980b, "Acoustic Cavitation Thresholds and Cyclic Processes", Ultrasonics, vol. 18, no. 5, pp. 201-209.

Newnham, R. E. & Dogan, A. 1995, Metal-electroactive ceramic composite transducer, US Patent 5729077 (patent).

Newnham, R. E., Xu, Q. C., & Yoshikawa, S. 1991, Transformed stress direction acoustic transducer, 4,999,819 (patent).

Newnham, R. E., Xu, Q. C., & Yoshikawa, S. 1994, Metal-electroactive ceramic composite actuators, 5,276,657 (patent).

Newnham, R. E. & Zhang, J. 1999, Flextensitional metal-ceramic composite transducer, US Patent 6232702 (patent).

Nordquist, L., Roxhed, N., Griss, P., & Stemme, G. 2007, "Novel microneedle patches for active insulin delivery are efficient in maintaining glycaemic control: an initial comparison with subcutaneous administration", Pharm.Res., vol. 24, no. 7, pp. 1381-1388.

Park, E. J., Werner, J., Beebe, J., Chan, S., & Smith, N. B. 2009, "Noninvasive Ultrasonic Glucose Sensing with Large Pigs (~200 Pounds) Using a Lightweight Cymbal Transducer Array and Biosensors", Journal of Diabetes Science and Technology, vol. 3, no. 3, pp. 517-523.

Park, E. J., Werner, J., & Smith, N. B. 2007, "Ultrasound mediated transdermal insulin delivery in pigs using a lightweight transducer", Pharmaceutical Research, vol. 24, no. 7, pp. 1396-1401.

Patek, S. D., Breton, M. D., Chen, Y., Solomon, C., & Kovatchev, B. 2007, "Linear Quadratic Gaussian-Based Closed-Loop Control of Type 1 Diabetes", J.Diabetes Sci.Technol., vol. 1, no. 6, pp. 834-841.

Pavlovic, M., Wroblewski, K., Manevich, Y., Kim, S., & Biaglow, J. E. 1996, "The importance of choice of anaesthetics in studying radiation effects in the 9L rat glioma", Br.J.Cancer Suppl, vol. 27, p. S222-S225.

Pickup, J. C. & Renard, E. 2008, "Long-acting insulin analogs versus insulin pump therapy for the treatment of type 1 and type 2 diabetes", Diabetes Care, vol. 31 Suppl 2, p. S140-S145.

Pitt, W. G., Husseini, G. A., & Staples, B. J. 2004, "Ultrasonic drug delivery--a general review", Expert.Opin.Drug Deliv., vol. 1, no. 1, pp. 37-56.

Page 87: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

77

Pond, W. G. & Houpt, K. A. 1978, The Biology of the Pig Cornell University Press, London, UK.

Prausnitz, M. R. 1997, "Reversible skin permeabilization for transdermal delivery of macromolecules", Critical Reviews in Therapeutic Drug Carrier Systems, vol. 14, no. 4, pp. 455-483.

Prausnitz, M. R. 2004, "Microneedles for transdermal drug delivery", Adv.Drug Deliv.Rev., vol. 56, no. 5, pp. 581-587.

Prausnitz, M. R. & Langer, R. 2008, "Transdermal drug delivery", Nat.Biotechnol., vol. 26, no. 11, pp. 1261-1268.

Renard, E. 2008, "Clinical experience with an implanted closed-loop insulin delivery system", Arq Bras.Endocrinol.Metabol., vol. 52, no. 2, pp. 349-354.

Rifkin, H. & Porte, D. 1990, Ellenberg and Rifkin's Diabetes, 4th edn, Elsevier Science, New York, NY.

Ryan, E. A. & Germsheid, J. 2009, "Use of continuous glucose monitoring system in the management of severe hypoglycemia", Diabetes Technol.Ther., vol. 11, no. 10, pp. 635-639.

Schlicher, R. K., Radhakrishna, H., Tolentino, T. P., Apkarian, R. P., Zarnitsyn, V., & Prausnitz, M. R. 2006, "Mechanism of intracellular delivery by acoustic cavitation", Ultrasound Med.Biol., vol. 32, no. 6, pp. 915-924.

Sekkat, N., Naik, A., Kalia, Y. N., Glikfeld, P., & Guy, R. H. 2002, "Reverse iontophoretic monitoring in premature neonates: feasibility and potential", Journal of Controlled Release, vol. 81, no. 1-2, pp. 83-89.

Sershen, S. & West, J. 2002, "Implantable, polymeric systems for modulated drug delivery", Adv.Drug Deliv.Rev., vol. 54, no. 9, pp. 1225-1235.

Shaw, J. E., Zimmet, P. Z., de Court, M., Dowse, G. K., Chitson, P., Gareeboo, H., Hemraj, F., Fareed, D., Tuomilehto, J., & Alberti, K. G. 1999, "Impaired fasting glucose or impaired glucose tolerance. What best predicts future diabetes in Mauritius?", Diabetes Care, vol. 22, no. 3, pp. 399-402.

Sintov, A. C. & Wormser, U. 2007, "Topical iodine facilitates transdermal delivery of insulin", J.Control Release, vol. 118, no. 2, pp. 185-188.

Skyler, J. S. 2009, "Continuous glucose monitoring: an overview of its development", Diabetes Technol.Ther., vol. 11 Suppl 1, pp. S5-10.

Smith, N. B. 2007, "Perspectives on transdermal ultrasound mediated drug delivery", Int.J.Nanomedicine, vol. 2, no. 4, pp. 585-594.

Page 88: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

78

Smith, N. B. 2008, "Applications of ultrasonic skin permeation in transdermal drug delivery", Expert Opin.Drug Deliv., vol. 5, no. 10, pp. 1107-1120.

Smith, N. B., Lee, S., Maione, E., Roy, R. B., McElligott, S., & Shung, K. K. 2003, "Ultrasound mediated transdermal transport of insulin through in vitro human skin using novel transducer designs", Ultrasound in Medicine and Biology, vol. 29, no. 2, pp. 311-317.

Suetsugu, M., Takebayashi, K., & Aso, Y. 2007, "Association between diabetic microangiopathy and vascular endothelial function evaluated by flow-mediated vasodilatation in patients with type 2 diabetes", Int.J.Clin.Pract., vol. 61, no. 6, pp. 920-926.

Tachibana, K. 1992b, "Transdermal delivery of insulin to alloxan-diabetic rabbits by ultrasound exposure", Pharm.Res., vol. 9, no. 7, pp. 952-954.

Tachibana, K. 1992a, "Transdermal delivery of insulin to alloxan-diabetic rabbits by ultrasound exposure", Pharm.Res., vol. 9, no. 7, pp. 952-954.

Tachibana, K. & Tachibana, S. 1991, "Transdermal delivery of insulin by ultrasonic vibration", J.Pharm.Pharmacol., vol. 43, no. 4, pp. 270-271.

Tachibana, K. & Tachibana, S. 2001, "The use of ultrasound for drug delivery", Echocardiography., vol. 18, no. 4, pp. 323-328.

Tezel, A., Sens, A., & Mitragotri, S. 2002, "Incorporation of lipophilic pathways into the porous pathway model for describing skin permeabilization during low-frequency sonophoresis", Journal of Controlled Release, vol. 83, no. 1, pp. 183-188.

The Diabetes Control and Complications Trial Research Group (DCCT) 1993, "The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes-Mellitus", N Engl J Med, vol. 329, no. 14, pp. 977-986.

Tressler, J. F. 1997, Capped Ceramic Underwater Sound Projector The "Cymbal", Ph. D., The Pennsylvania State University.

Volpe, L., Pancani, F., Aragona, M., Lencioni, C., Battini, L., Ghio, A., Resi, V., Bertolotto, A., Del, P. S., & Di, C. G. 2009, "Continuous subcutaneous insulin infusion and multiple dose insulin injections in Type 1 diabetic pregnant women: a case-control study", Gynecol.Endocrinol. pp. 1-4.

Weinstein, R. L., Bugler, J. R., Schwartz, S. L., Peyser, T. A., Brazg, R. L., & McGarraugh, G. V. 2007, "Accuracy of the 5-day freestyle navigator continuous glucose monitoring system - Comparison with frequent laboratory reference measurements", Diabetes Care, vol. 30, no. 5, pp. 1125-1130.

Page 89: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

79

Weinzimer, S. A., Steil, G. M., Swan, K. L., Dziura, J., Kurtz, N., & Tamborlane, W. V. 2008, "Fully automated closed-loop insulin delivery versus semiautomated hybrid control in pediatric patients with type 1 diabetes using an artificial pancreas", Diabetes Care, vol. 31, no. 5, pp. 934-939.

Weissberg-Benchell, J., ntisdel-Lomaglio, J., & Seshadri, R. 2003, "Insulin pump therapy: a meta-analysis", Diabetes Care, vol. 26, no. 4, pp. 1079-1087.

Whelan, J. 2002, "Electroporation and ultrasound for gene and drug delivery", Drug Discovery Today, vol. 7, no. 11, pp. 585-586.

Wickramasinghe, Y., Yang, Y., & Spencer, S. A. 2004, "Current Problems and Potential Techniques in In Vivo Glucose Monitoring", Journal of Fluorescence, vol. 14, no. 5, pp. 513-520.

Wilkins, E. & Atanasov, P. 1996, "Glucose monitoring: State of the art and future possibilities", Medical Engineering & Physics, vol. 18, no. 4, pp. 273-288.

Wong, L. J., Buckingham, B. A., Kunselman, B., Istoc, E., Leach, J., & Purvis, R. 2006, "Extended use of a new continuous glucose monitoring system with wireless data transmission in children with type 1 diabetes mellitus", Diabetes Technol.Ther., vol. 8, no. 2, pp. 139-145.

Zhang, I., Shung, K. K., & Edwards, D. A. 1996, "Hydrogels with enhanced mass transfer for transdermal drug delivery", J.Pharm.Sci., vol. 85, no. 12, pp. 1312-1316.

Page 90: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Appendix A

FUNDAMENTAL RESONANCE FREQUENCY OF CYMBAL TRANSDUCER – FEA

A.1 In Air

Table A.1: Comparison of the in-air resonance frequency of standard size brass-capped cymbals for various PZT types.

PZT type rf (kHz)

PZT-552 23.5 PZT-5A 23.6 PZT-4 24.0 PZT-8 24.1

Table A.2: Comparison of the in-air resonance frequency of standard size cymbals forvarious endcap materials.

Cap material rf (kHz)

Brass 23.5 Titanium 34.8

Kovar 26.7 Steel 32.3

Molybdenum 34.0

Page 91: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

81

A.2 In Water

Table A.3: Comparison of the in-air resonance frequency of standard size brass-capped cymbals for different endcap thickness.

Cap thickness (mm) rf (kHz)

0.12 19.7 0.20 22.1 0.25 23.5 0.3 25.0 0.38 27.7 0.5 31.9

Table A.4: Comparison of the in-air resonance frequency of standard size brass-capped cymbals for different cavity depths.

Cavity depth (mm) rf (kHz)

0.12 16.7 0.18 18.4 0.25 20.8 0.32 23.4 0.47 29.2

Table A.5: Comparison of the in-water resonance frequency of standard size brass-capped cymbals for various PZT types.

PZT type rf (kHz)

PZT-552 16.5 PZT-5A 16.1 PZT-4 16.3 PZT-8 16.5

Page 92: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

82

Table A.6: Comparison of the in-water resonance frequency of standard size cymbals forvarious endcap materials.

Cap material rf (kHz)

Brass 16.5 Titanium 20.0

Kovar 18.6 Steel 22.0

Molybdenum 24.7

Table A.7: Comparison of the in-water resonance frequency of standard size brass-capped cymbals for different endcap thickness.

Cap thickness (mm) rf (kHz)

0.12 10.9 0.20 14.5 0.25 16.5 0.3 18.3 0.38 21.3 0.5 25.8

Table A.8: Comparison of the in-water resonance frequency of standard size brass-capped cymbals for different cavity depths.

Cavity depth (mm) rf (kHz)

0.12 11.5 0.18 12.8 0.25 14.5 0.32 16.5 0.47 20.6

Page 93: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Appendix B

DETAIL INSTRUCTION OF CYMBAL ARRAY CONSTRUCTION

B.1 Metal Endcaps

1. Punch 0.5" diameter caps from brass sheet (see table below for thickness; generally 0.2-0.7mm) with Precision Brand Punch and Die Set.

2. Set cavity depth with shaping die. Use specific pressure (60kpsi) because excess pressure will flatten the die.

Array dimensions

Thickness (mm) Depth (mm)

1x1 0.12 0.25 2x2 0.25 0.25 3x3 0.25 0.32

Page 94: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

84

Page 95: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

85

B.2 Cymbals

1. Rub off the white powder from the 1mm thick, 0.5" diameter piezoelectric material (PZT) with ≥ 400 grain sand paper. The dot marks positive polarity and must be redrawn after powder is removed. Gently rub in a circular motion; do not “shine” the PZT, it will damage the electrode.

2. Weigh 3:1:1 2741LV epoxy to 15LV epoxy to silver powder. Mix thoroughly. Apply a thin strip to the edges of the PZT disc with a toothpick.

3. Carefully place the endcap onto the PZT disk and align by touch. Redraw the dot to mark polarity. Repeat with other side and align again. Wipe off excess epoxy with silver powder from the sides with a Kim-wipe. Clip sides and allow curing for 24hrs at room temperature. Leave excess epoxy as indicator. (Clipping the center will change the cavity depth.)

B.3 Polyurethane Base

1. Make a sheet of polyurethane after degassing. Spread out and remove air bubbles. Place spacers and top plate. For larger arrays, prepare thicker sheets. Cut squares (depends on molding; distances between are arbitrary but symmetry is preferred; ~37mm

Page 96: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

86

x 37mm for 2x2) for the arrays. (Leaving too large of a gap between Cymbal transducers will cause the 3d Gaussian output to be distorted. After using polyurethane, flush with nitrogen gas before sealing.) 2. Punch the appropriate amount of 0.5" diameter holes in the squares. (The polyurethane bases are necessary for alignment and spacing, and to ensure that cymbals will be parallel to transducer surface.)

B.4 Connections

1. For 1 × 1 pattern, bare portion of red wires (arbitrary, but ~4.5ft) are connected directly to positive side of the cymbal and black, to the negative side. For 2 × 2 arrays, single-strand wires are used to connect the cymbals together, and the red/black wires are soldered onto the intersection.

2. Mix the silver epoxy, a conducting adhesive. The mix ratio is 1:0.08 solid to liquid. The mixture may seem thick but is acceptable.

3. Use a toothpick and the least amount possible to connect wires to cymbals. Use as little tape as possible. Again, keep excess epoxy as an indicator. Allow 8hrs curing time for silver epoxy.

Page 97: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

87

B.5 Casting

1. Before casting, test connections with a multimeter.

2. Mix the two-part polyurethane in a 10:1 Part A to Part B ratio. Mix and degas in the vacuum. Polyurethane will rise when bubbles are drawn out, therefore, use a large container.

Page 98: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

88

3. Spray the mold with Ease Release. Slowly pour polyurethane into mold to avoid capturing air bubbles, filling only up to 1mm. Remove air bubbles with a toothpick. 4. Slowly place the base with the cymbals into the mold (red/positive side down), one side first to prevent air bubble formation.

5. Use a toothpick to ensure that the base with the cymbals is parallel to the bottom surface.

Page 99: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

89

6. Fill the rest of the mold with polyurethane.

(For 1 × 1 pattern, fill the entire mold with polyurethane before adding the base with the cymbal transducer. Then, slowly submerge the base. Also, when connecting the wires to the cymbals, make sure there is symmetry so that the weight is not all on one side. This will cause the base to rotate in the polyurethane. Shaping the base to fit tightly to the mold will also help prevent rotation.)

7. Tape the top and side of the groove where the wires come out for support. Allow polyurethane to cure for 24hrs. Leave a small amount of polyurethane as an indicator. (If the atmosphere is especially dry, the polyurethane has a tendency to dry with curve. To avoid this, cover with a towel.)

Page 100: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

90

8. To remove transducer from molding, gently bend the mold. Excessive pressure may damage transducer or disrupt connections.

Page 101: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Appendix C

A COMPARISON BETWEEN INTRAPERITONEAL INJECTION AND TRANSDERMAL DELIVERY

One of the many questions with ultrasound mediated transdermal insulin delivery

is the relationship between the levels of glucose decrease from ultrasonic method versus a

direct subcutaneous injection of insulin. Therefore the blood glucose response from

direct injections of insulin against the transdermal delivery using the cymbal array was

examined.

A total of 16 experiments using eight Sprague-Dawley rats (350 - 550 g) were

performed in four experimental groups with four rats in each group: one ultrasonic

transdermal delivery and three subcutaneous injection groups. Each rat was used for

multiple experiments. Rats were anesthetized with a combination of ketamine

hydrochloride (60 mg/kg intramuscularly, Ketaject®, Phoenix, St. Joseph, MO) and

xylazine hydrochloride (10 mg/kg intramuscularly, Xyla-Ject®, Phoenix, St. Joseph,

MO). In addition to its role in general anesthesia, xylazine was used to induce a

temporary, but sustained (up to 12 hrs), hyperglycemia in rats.

For the ultrasonic transdermal delivery, the abdominal area of the rat was shaved

using an electric shaver and a depilatory agent was applied to the skin to eliminate any

remaining hair. After shaving, a 1 mm thick, water-tight standoff was attached between

the skin and the array (Figure C.1). With the rat in the dorsal decubitus position, a

reservoir within the standoff was filled with insulin (Humulin® R, rDNA U-100, Eli Lilly

and Co., Indianapolis, IN) through a small hole in the array. Care was taken to remove

Page 102: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

92

all bubbles from the solution in the reservoir to prevent disruption of ultrasound

transmission. The elapsed time from the initial injection of the anesthetic until the start

of ultrasound exposure was no longer than 25 minutes. For the ultrasound exposure, the

cymbal array was operated at 20 kHz with an Isptp = 102.2 ± 2.3 mW/cm2 for 60 minutes.

After the 60 minutes ultrasound exposure, the array was removed and the skin examined

for visible lesions.

The dose of insulin selected for the injection groups was based on published

insulin doses used to control diabetes mellitus in other species and on pilot experiments

(not reported). Humulin® R Insulin (100 U/ml) was diluted with a 0.9% saline solution

(Phoenix Pharmaceutical, Inc., St. Joseph, MO) to 0.5 U/ml and delivered subcutaneously

(under the skin of the ventral abdomen) at 0.15, 0.20, or 0.25 U/kg.

Figure C.1: Photograph of a transdermal insulin delivery experiment on a rat place in a dorsal decubitus position with the array attached.

Page 103: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

93

Blood was collected from the tail vein of each rat to obtain a baseline glucose

level and, following insulin administration, additional samples were collected every 15

minutes for 90 minutes. For the ultrasonic transdermal delivery, the base line glucose

level was measured at the beginning of the ultrasound exposure. The blood glucose level

(mg/dl) for each sample was determined using the ACCU-CHEKTM blood glucose

monitoring system (Roche Diagnostics Co., Indianapolis, IN). Each sample was tested at

least two times to confirm the accuracy of the reading

The data was corrected by subtracting the baseline glucose for each animal from

each data point such that only changes in blood glucose were compared. Statistical

analysis was performed using Microsoft Excel® (Microsoft Corp., Redmond, WA) and

the data of blood glucose versus time were pooled for each group and analyzed as the

mean and standard deviation. A t-test was used to analyze the statistical significance of

the differences among the means of groups. The p-value was used to determine if the

between-group differences are significantly greater than chance.

Results of the ultrasound delivery compared to injections doses for the four

groups are graphed as a decrease in the blood glucose level during the 90- minute

experiment (Figure C.2). The data were graphed and reported as the mean ± standard

deviation (x ± sd) of each group. Due to the anesthesia, the average initial glucose level

at the beginning of the experiment was 340 ± 69 mg/dl for the 16 experiments. Rats not

anesthetized with xylazine would have a blood glucose concentration closer to a normal

level of ≈100 mg/dl.

Page 104: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

94

For direct subcutaneous injections with a dose of 0.15 U/kg and 0.20 U/kg, the

blood glucose level deviated little from the baseline value. Overall the glucose level

varied no greater than 32 mg/dl from the initial value over a 90 minute experiment period

for both doses. Yet for a subcutaneous insulin injection of 0.25 U/kg, the blood glucose

decreased by 190 ± 96 mg/dl after 90 minutes. In comparison, the ultrasound produced a

blood glucose decrease of 263 ± 40 mg/dl at 90 minutes. An t-test analysis at 90 minutes

indicated that all the groups were statistically different from each other at a p-level less

than 0.01. Visual examination of the skin exposed to ultrasound did not indicate any

damage or significant changes.

Figure C.2: Changes of blood glucose by the ultrasonic transdermal insulin delivery andsubcutaneous injections of insulin.

-350

-300

-250

-200

-150

-100

-50

0

50

100

0 15 30 45 60 75 90

Time (min)

Ch

ang

e o

f b

loo

d g

luco

se l

evel

(m

g/d

l)

340 ± 69 mg/dl

0.15 U/kg injection (n=4)

0.20 U/kg injection (n=4)

0.25 U/kg injection (n=4)

60 minutes ultrasound exposure (n=4)

-350

-300

-250

-200

-150

-100

-50

0

50

100

-350

-300

-250

-200

-150

-100

-50

0

50

100

0 15 30 45 60 75 900 15 30 45 60 75 90

Time (min)

Ch

ang

e o

f b

loo

d g

luco

se l

evel

(m

g/d

l)

340 ± 69 mg/dl

0.15 U/kg injection (n=4)

0.20 U/kg injection (n=4)

0.25 U/kg injection (n=4)

60 minutes ultrasound exposure (n=4)

0.15 U/kg injection (n=4)

0.20 U/kg injection (n=4)

0.25 U/kg injection (n=4)

60 minutes ultrasound exposure (n=4)

Page 105: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

95

For humans to regulate their blood glucose level, the required insulin injection

dose is 0.5 - 1 U/kg/day for adults and children and 0.8 - 1.2 U/kg for adolescents

experiencing growth spurts (Hodgson & Robert J. 2006; Lance et al. 2002). Direct

subcutaneous injection doses for animals range from 0.1 - 0.4 U/kg subcutaneous (SC)

for dogs and 0.1 - 0.5 U/kg (SC) for ferrets (Plumb 2005). Ranges are given since

physiological variables have a direct effect on the specific blood glucose decrease from

injections. As many clinicians and diabetes patients know, the body's glucose response to

direct injections varies according to a host of variables such as body weight, fat

percentage, exercise level, and composition of the most recent meal. Consistency of diet

and exercise along with routine dose-glucose recording assists in the control of blood

glucose. Without proper glucose control, diabetic complications may result, including

renal failure, peripheral vascular disease, and limb amputation (Meeuwisse-Pasterkamp,

van der Klauw, & Wolffenbuttel 2008; Suetsugu, Takebayashi, & Aso 2007).

To facilitate the ability of a diabetic patient to avoid repeated and painful daily

injections of insulin, a safe, light-weight, low-profile, inexpensive and potentially

portable ultrasonic device is proposed. The goal of this task was to develop an

approximate relationship between dose levels from direct subcutaneous injections and

noninvasive ultrasound at a Isptp ≈100 mW/cm2 for 60 minutes. While an exact

mathematical relationship was not determined, the results in Figure C.2 indicate that the

ultrasound dose appears to be greater than an injection dose of 0.25 U/kg for rats. Not

included in the results was a single rat experiment which used a direct injection dose of

0.44 U/kg which resulted in a rapid decrease in blood glucose of 290.5 ± 8 mg/dl after

only 60 minutes. Given this rapid decrease, the animal was removed from the experiment

Page 106: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

96

and this dose was determined to be too high for the rats. Yet the single point result can

indicate that the ultrasound dose would be somewhere between injection dose levels of

0.25 - 0.44 U/kg. Nevertheless the results are the first steps in determining a relationship

between ultrasound intensity levels and insulin dose responses. Further experiments

should explore the use of larger animals with a similar size and weight as humans or

animals which are truly diabetic such as pancreatectomized pig.

A person is considered diabetic if his or her blood sugar level is above 126 mg/dl

after eight hours of fasting. People without diabetes have fasting sugar levels that

generally run between 70 - 110 mg/dl. A glucose concentration of 110 - 126 mg/dl is

classified as impaired fasting glucose. In the oral glucose tolerance test, 140 - 200 mg/dl

is impaired glucose tolerance and greater than 200 mg/dl is considered diabetic

(Carnevale Schianca, Rossi, Sainaghi, Maduli, & Bartoli 2003; Rifkin & Porte 1990;

Shaw et al. 1999). In this last situation, a diabetic person would need to inject enough

insulin to reduce his or her blood glucose by about 100 mg/dl. Both the ultrasound and

direct injection of 0.25 U/kg achieve blood glucose level decreases of 190 mg/dl or

greater.

Use of transdermal drug delivery techniques has practical clinical application to

medications which need to be injected multiple times either daily or weekly. A recent

review on ultrasound drug delivery states that "small-sized low-frequency transducers

need to be developed so that patients can wear them" (Pitt, Husseini, & Staples 2004).

As with diagnostic ultrasound imaging, drug delivery using therapeutic ultrasound

requires a delicate balance between safety and efficacy and requires careful scientific

study. For a transdermal device to replace conventional needles, the bioeffects and safety

Page 107: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

97

of each device needs to be carefully evaluated, since it will not matter how much of any

drug can be transported if the skin is damaged or the procedure is painful.

Page 108: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

Appendix D

MATLAB CODES

D.1 Low pass π-circuit for impedance matching

% ============================================================= % Calculates a low pass pi matching circuit for any array. % Required values of array: % Resonance Frequency (fr), measured complex impedance at fr, % desired Q of the circuit. % Source impedance = 50 ohms % Required condition: % Q>0 % % Eun-Joo Park, modified on July 3, 2007, Tuesday % ============================================================= function [C1,C2,L]=ejpimc(f,RL,XL) %fr=Resonance frequency of array %RL=Transducer resistance %XL=Transducer reactance %Transducer array parameters w=2*pi*f; ZL=RL+j*XL; Q=sqrt(RL/50)+0.01 %Desired Q %Source output impedance parameters RS=50; %Source resistance XS=0; %Source reactance ZS=RS+j*XS; Rc=max(RS,RL)/(Q^2+1); QL=-XL/RL; %Node load Q RpL=RL*(1+QL^2); %Equivalent parallel load resistance CpL=QL/(RpL*w); %Equivalent parallel load capacitance q1=sqrt(RpL/Rc-1); %Match to source C2=(q1/(w*RpL))-CpL %C2 value

Page 109: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

99

LL=(q1*Rc)/w; %Inductance due to load QS=-XS/RS; %Node source Q RpS=RS*(1+QS^2); %Equivalent parallel source resistance CpS=QS/(RpS*w); %Equivalent parallel source capacitance q2=sqrt(RpS/Rc-1); %Match to source C1=(q2/(w*RpS))-CpS %C1 value L_S=(q2*Rc)/w; %Inductance due to source L=L_S+LL %L value %Calculate the matched and unmatched impedance over the bandwidth BW=f/Q; %Bandwidth given by specified Q f_axis=linspace(round(f-BW/2),round(f+BW/2),1000); w_axis=2*pi*f_axis; ZC1=1./(j*w_axis*C1); ZC2=1./(j*w_axis*C2); Z_L=j*w_axis*L; Z1=(ZL.*ZC2)./(ZL+ZC2); Z2=Z1+Z_L; Zin=(Z2.*ZC1)./(Z2+ZC1); Yin=1./Zin; %Calculate the unmatched impedance over the bandwidth Ct=1/(2*pi*f*abs(XL)); Zt=RL+1./(j*w_axis*Ct); %Plots figure(1) f_axis=f_axis./1000; subplot(2,1,1) plot(f_axis,abs(Zin)) grid on xlabel('f (Hz)') ylabel('|Zin|') title('Magnitude of Impedance'); subplot(2,1,2) plot(f_axis,(180/pi)*angle(Zin)) grid on xlabel('f (Hz)') ylabel('Phase (deg)') title('Phase of Impedance');

Page 110: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

100

D.2 Feedback controller

% ============================================================= % operating program with steps of % 1. Run ultrasound system for glucose sensing (10 minutes) % 2. Wait for glucose sensing with potentiostat (8 minutes) % 3. Run ultrasound system for insulin delivery (2 minutes + 18 % minutes) % 4. Repeat step 1-3 % % Eun-Joo Park, modified on January 4th, 2009 Sunday % ============================================================= function controller1(current) for ii=1:3 turn=ii % Turn on US system of GS fg33120a; a=1 % Pause process for manual measurement of current from potentiostat t_GS=7; % time for glucose sensing by potentiostat with two biosensors pause(t_GS*60); % pause (seconds) b=2 % Read current at 700 mV load GS_FC_1.txt; load GS_FC_2.txt; load GS_FC_3.txt; load GS_FC_9.txt; i_gs1=min(GS_FC_1(:,2)) i_gs2=min(GS_FC_2(:,2)) i_gs3=min(GS_FC_3(:,2)) i_gs9=min(GS_FC_9(:,2)) % i_gs=mean([i_gs1, i_gs9]) i_gs=mean([i_gs1, i_gs2, i_gs3, i_gs9]) c=3

Page 111: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

101

if i_gs>=current; fg33250a_on; else fg33250a_off; end d=4 t_TD=3; % time to wait for the next GS (minutes) pause(t_TD*60); % pause (seconds) e=5 end % Turn on US system of GS fg33120a; aa=1 % Pause process for manual measurement of current from potentiostat t_GS=7; % time for glucose sensing by potentiostat with two biosensors pause(t_GS*60); % pause (seconds) bb=2 % Read current at 700 mV load GS_FC_1.txt; load GS_FC_2.txt; load GS_FC_3.txt; load GS_FC_9.txt; i_gs1=abs(min(GS_FC_1(:,2))) i_gs2=abs(min(GS_FC_2(:,2))) i_gs3=min(GS_FC_3(:,2)) i_gs9=min(GS_FC_9(:,2)) % i_gs=mean([i_gs1, i_gs9]) i_gs=mean([i_gs1, i_gs2, i_gs3, i_gs9]) cc=3

Page 112: ULTRASONIC TRANSDERMAL GLUCOSE MONITORING AND INSULIN

VITA

EUN-JOO PARK

Eun-Joo Park was born in Choong-Nam in Korea (Republic of). She received her

BS and MS degrees in Physics (Physical Acoustics) from Sung Kyun Kwan University,

Seoul, Korea in 1996 and 1998 respectively. The thesis of her MS degree was about

“The thermal effect of ultrasound in transdermal insulin delivery.” She received another

MS degree in Mechanical Engineering (Underwater Acoustics) from Boston University,

Boston, MA in 2003. The thesis was about “Acoustic characterization of saturated

sediments using a thick-walled, water-filled impedance tube.” After her second MS

degree, she worked at the Focused Ultrasound Lab in Brigham and Women’s Hospital,

Boston, MA as a researcher. In order to study biomedical ultrasound for her PhD degree,

she joined the interdisciplinary program in bioengineering at The Pennsylvania State

University in 2004 and has been working as a research assistant in the Therapeutic

Ultrasound Laboratory. Her research deals with biomedical applications of the cymbal

transducer arrays on noninvasive transdermal drug deliveries. Eun-Joo’s research

interests are the diagnostic and therapeutic ultrasound in medical applications especially

for cancer treatment and diabetes care. She is also interested in the research of

developing ultrasound transducers for specific medical applications.