rigid gas permeable contact lens and corneal topography
TRANSCRIPT
Rigid gas permeable contact lens and cornealtopography
E. Yebra-Pimentel1, M. J. Gira ldez1, F. L. Arias4, J. Gonza lez1,J. M. Gonza lez1, M. A. Para®ta2 and M. Febrero3
1Department of Applied Physics (Optometry Group), School of Optics and Optometry, University ofSantiago de Compostela, 15706 Galicia, Spain, 2Department of Ophthalmology, School of Opticsand Optometry, University of Santiago de Compostela, Galicia, Spain, 3Department of Statistic andO. R., Faculty of Mathematics, University of Santiago de Compostela, Galicia, Spain and4Department of Optometry, University AutoÂnoma of Aguascalientes, Mexico, Mexico
Summary
We investigated the effect of high Dk daily wear rigid gas permeable contact lenses on cornealtopography. Eight young myopic subjects wore hard contact lenses for 21 days. Cornealtopography was monitored using the EyeSys system. It was measured every day during thenext 21 days after the ®tting. We recorded the corneal radius of curvature at 16 peripheral pointsapproximately 1, 2, 3 and 4 mm along the four principal meridians (nasal, superior, inferior andtemporal). Our study showed that daily wear RGP Forum 210 does not produce signi®cantalterations of the corneal curvature as a function of time. q 2001 The College of Optometrists.Published by Elsevier Science Ltd. All rights reserved.
Introduction
The shape of the human cornea is best represented by an
ellipse, which progressively ¯attens in curvature from
centre to edge. As the cornea is the major refracting compo-
nent of the eye, accurate measurement of the corneal shape
is important to understand the optical characteristics of an
individual eye and in the ®tting and optics of contact lenses.
Variations in the shape of the cornea have important impli-
cations for the optical quality of the eye.
Contact lens wear may alter corneal shape and induce
corneal distortion (Wilson and Klyce, 1994). Rengstorff
showed that rigid contact lenses can induce transient
changes in corneal curvature (Rengstorff, 1969a). Poly-
methylmethacrylate lenses (PMMA) produce initial corneal
steepening in association with central corneal edema, parti-
cularly during the adaptation period, followed by a slowly
progressive ¯attening of the central corneal curvature, as
measured by keratometry (Rengstorff, 1969b). Corneal ¯at-
tening has also been reported with long-term wear of rigid
gas-permeable lenses (RGP) on daily and extended wear
schedules (Briceno-Garbi, 1984; Polse et al., 1987; Iskeleli
et al., 1996). Gradual reductions in corneal toricity
(Briceno-Garbi, 1984; Polse et al., 1987; Iskeleli et al.,
1996) and corneal asphericity (Kame and Kennedy, 1987)
have also been reported.
The effect of the base curve ®tting relationship on corneal
topography has been evaluated by Carney (1974) and
Carney (1975), who found that during ®tting, ¯at, steep
and alignment ®tting lenses all reduced the asphericity of
the cornea. After ®tting, the steeply ®tted group continued to
show a decrease in asphericity; the ¯at-®tting group,
however, actually showed an increase in asphericity, with
the alignment-®tting group between the other groups. In
some cases the inferior quadrant was more affected, espe-
cially in the case of ¯at-®tting lenses. These changes are a
result of the moulding force of the lens against the eye or are
caused by contact lens interference with corneal metabolism
or, perhaps, by a combination of both factors (Calossi et al.,
1996).
Videokeratoscopes permit the measurement of the
shape of a large proportion of the corneal surface (Horner
et al., 1998). This measurement is achieved in most
236
Ophthal. Physiol. Opt. Vol. 21, No. 3, pp. 236±242, 2001q 2001 The College of Optometrists. Published by Elsevier Science Ltd
All rights reserved. Printed in Great Britain0275-5408/01/$20.00
www.elsevier.com/locate/ophopt
PII: S0275-5408(00)00018-1
Received: 23 July 1999
Revised form: 22 December 1999
Accepted: 19 March 2000
Correspondence and reprint requests to: E. Yebra-Pimentel. Tel.: 134-
981-56-31-00; fax: 134-81-59-04-85.
E-mail address: [email protected] (E. Yebra-Pimentel).
videokeratoscopes by re¯ecting light from a series of
concentric rings from the cornea/tear surface, similar in
principle to the Placido disk. A CCD is used to capture
the re¯ected image, which is then digitized and analyzed
by computer to provide a detailed reconstruction of the
central and peripheral corneal shape.
The purpose of this study was to investigate the effect of
daily wear high Dk rigid gas permeable contact lenses on
corneal topography for a period of 21 days after ®tting.
Materials and methods
Eight young myopic subjects were enrolled in this study.
Of these patients, 4 were females and 4 were males. Their
ages ranged from 20 to 30 years and their refractive errors
ranged from 22.00 and 28.00 D. The maximum cylinder
power was 22.00 D. Two patients had previously worn
hydrogel contact lenses, but they stopped using them one
month before the study. No patient had any ocular or
systemic disease and none had undergone eye surgery.
The study participants did not use any local eye medications
during contact lens wear.
All subjects were ®tted with rigid gas permeable contact
lenses (RGP) and wore the lenses on a daily wear basis. The
responses of 8 right eyes to RGP lens wear were investi-
gated. The contact lenses used are manufactured by Bausch
& Lomb (Spain). The trade name is Forum 210w and the
generic name is Si¯uorfocon-A. They are made of a copo-
lymer in which the major constituent is a monomer with
silicone and ¯uorine. The Dk value of this material is 210 £10211 cm2
=s (ml O2/ml mm Hg) at 358. We used a trial set of
lenses in which base curves ranged from 7.30 to 8.70 mm in
0.10 mm steps. The power of these lenses was 22.00 D, the
diameter was 9.80 mm and the optic zone diameter was
8.40 mm. The optimal ®t was one that achieved a central
alignment ¯uorescein pattern with moderate-to-minimum
edge lift and adequate movement. Lenses were cleaned
and stored according to the manufacturer's recommended
procedures when not being worn.
Corneal topography was measured every day during a 21-
day period after the ®tting. Sixteen corneal points were
measured on the right eye every day after 10 hours of lens
wear (20:00 pm). Measurements were taken 1 min after the
removal of the lens. During the measurements, the patient
blinked naturally, and readings were taken immediately
after completion of a blink. The time of day at which the
topography measurements were taken was standardized to
avoid the effect of diurnal variations in corneal topography
(Kiele et al., 1982).
The EyeSys Corneal Analysis Systemw videokeratoscope
was used to record the local radius of curvature at 16 periph-
eral corneal points approximately at 1, 2, 3 and 4 mm along
the superior, inferior, nasal and temporal main meridians
(with maximum and minimum power). To decide where
the 1, 2, 3 and 4 mm points were located and the radii values
at these points, the data was presented in tabular form.
Location of the peripheral points 1, 2, 3 and 4 mm along
each meridian was usually accurate within 0.05 mm (maxi-
mum error 0.14 mm), because the position of the re¯ected
mire rings varies slightly in spacing depending upon the
shape of the individual cornea along that meridian.
The calibration of the videokeratoscope was checked by
taking repeated measurements of four calibrated steel balls
both before, during and after the study. There was no signif-
icant variation in the videokeratoscope calibration over the
course of the study.
In this study, we analyzed results using the SPSS Profes-
sional Statistics 6.1 program (Statistical Package for the
Social Sciences). Two statistical analyses were used to
examine changes in corneal curvature over time: analysis
of autocorrelation and analysis of regression. Both analyses
were applied for a total of 21 corneal radii measurements
from 16 corneal points per eye. However, less than 21
measurements were obtained for the most peripheral corneal
points in some cases. In the vertical meridian this was possi-
bly due to interference from eyelids, being most noticeable
in the superior meridian. The 4 mm nasal point was prob-
ably obscured by the nose shadow.
Results
The statistical analysis of the data in this study was
performed using data corresponding with 16 corneal points
of 8 right eyes. We analyzed results from each eye sepa-
rately. We based this decision on the assumption that each
eye ®tted with Forum 210 lens would respond indepen-
dently to the particular pressures generated as a result of
the individual lens-cornea ®tting relationship achieved.
Following 21 days of RGP lens wear, no signi®cant
differences were found in corneal curvature at any of the
temporal, nasal, superior and inferior locations studied in
each of the right eyes of the 8 subjects. Several outliers data
points were observed only in one eye, but they have no
in¯uence in the estimation of the slope. Data from this
eye, the only one that shows outliers, are presented as an
example in this paper.
Before a regression analysis can be performed, we must
ensure that the data are independent and there are no
outliers. If there is any type of dependence it is well
known that the estimation of the regression coef®cients
will not be consistent. In our case the measurements were
taken over 21 consecutive days and the suspicion of depen-
dence is strong. To analyze dependence over time the most
suitable tool is the autocorrelation function, usual tool in the
time series methodology (see, for example, Box and
Jenkins, 1976). Let xi be the measure at the day i. The
autocorrelation coef®cient at lag k (hereafter r k) is the
usual correlation between pairs of values {xi, xi1k}. The
plot of r k vs the lag k, is called the autocorrelation function.
When the data are independent r k is equal to zero for every
Rigid gas permeable contact lens and corneal topography: E. Yebra-Pimentel et al. 237
238 Ophthal. Physiol. Opt. 2001 21: No 3
Figure 1. Autocorrelation function for lags k � 1;¼; 5 on x-axis (bars), the con®dence intervals for everycoef®cient (solid line) and the probability obtained by the Ljung-Box test (at the bottom of each ®gure) on thechange in corneal radius at 1, 2, 3 and 4 mm from the center along (a) temporal, (b) nasal, (c) inferior, and(d) superior main meridians as a function of time. Data obtained from one eye are shown as an example.
Rigid gas permeable contact lens and corneal topography: E. Yebra-Pimentel et al. 239
Figure 1. (continued)
k $ 1: We have tested the hypothesis that the ®rst ®ve auto-
correlation coef®cients are simultaneously equal to zero
using the Ljung-Box test. Indeed, we have calculated con®-
dence intervals at 95% for every autocorrelation coef®cient.
Figure 1 shows the autocorrelation function for lags k �1¼5 (bars), the con®dence intervals for every coef®cient
(solid line) and the probability obtained by the Ljung-Box
test (at the bottom of each ®gure) for the eye taken as an
example. The autocorrelation function also shows that there
is no trend in the data (when the data are affected by a trend
the autocorrelation function decreases slowly and there are
signi®cant coef®cients). Also the probability obtained by the
Ljung-Box test is always greater than 0.05. Consequently
the results obtained in the regression analysis, later in this
paper, are not surprising.
Another thing we must check before doing a regression
analysis is the absence of outliers. In Figure 2 the box-
plots of the residuals from regression analysis are shown.
In this case there are several outliers in the data corre-
sponding to the measurements taken on Monday of the
third week. These may be caused by interference from
the eyelids or irregularities in the tear ®lm (Hough and
Edwards, 1999) or due to a mechanical effect of the contact
lens. But the central position of this outlier means that this
has no in¯uence in the estimation of the slope. To measure
the in¯uence of a datum point, we could use the Mahala-
nobis Distance �Mi � �xi 2 �x�2=s2x� which takes into
account the distance of a datum from its average. The
Mahalanobis Distance of this datum is 0.00502 and the
scale of this distance goes from 0.002 to 2.709. There
were no signi®cant changes in corneal topography at 1,
2, 3 and 4 mm from the center of the cornea as a function
of time in the eight patients from the study.
Analyses of regression were also carried out at each
corneal point independently to investigate trends in corneal
curvature at 1, 2, 3 and 4 mm from the center along superior,
inferior, nasal and temporal main meridians as a function of
time. The statistics of this analysis are presented in Table 1.
The determination coef®cient indicates that the variation of
the curvature is not dependent on time. The slope in all
regression lines is close to zero, with a p value much greater
than 0.05.
240 Ophthal. Physiol. Opt. 2001 21: No 3
Fig. 2. Box-plots of the residuals from regression analysis corresponding to 1, 2, 3 and 4 mm from the center alongtemporal, nasal, superior and inferior main meridians. Data obtained from one eye are shown as an example. Oneor more peripheral data points were not available possibly due to interference from eyelids and nose shadow.
Discussion
Several reports have indicated that PMMA contact lens
wear often leads to a signi®cant increase in central corneal
curvature (Miller, 1968; Rengstorff, 1971; Montenegro et
al., 1993). After the initial adaptation period, the curvature
returns to the baseline value (Rengstorff, 1969b). After
about one year, the cornea shows ¯attening beyond the
original baseline values (Saks, 1966). The ®rst of these
changes is attributed largely to localized corneal edema
caused by non-permeable PMMA material (Montenegro et
al., 1993; Hovding, 1983). Mechanical molding of cornea
may also be partly responsible for the altered corneal curva-
ture with PMMA lenses (Montenegro et al., 1993; Hovding,
1983; Carney, 1975).
In the present study we examined the corneal curvature
changes that occurred in eight RGP lens wearers and discov-
ered that there are no statistically signi®cant �p . 0:05�corneal curvature changes over a 21-day period after ®tting.
The high oxygen transmissibility of Forum lenses can
prevent corneal hypoxia, oedema, and the corneal curvature
changes seen with PMMA lenses. Moreover, our results
show that the mechanical effects of lid pressure do not
provoke corneal curvature variation.
Sevigny (1986) compared the corneal curvature and
refractive changes produced by Boston II lenses on daily
wear on one eye and Boston IV lenses on extended wear on
the opposite eye. He observed that corneal curvature
showed only a 0.07 D average ¯attening with daily wear.
DeRubeis and Shily (1985) found that patients ®tted with
Boston II lenses did not show any change in the radius of
curvature of the horizontal meridian, and the vertical meri-
dian ¯attened by 0.31 D. Obviously these lenses are low Dk
materials. However recent studies with higher Dk materials
have demonstrated a much more moderate keratometric and
refractive changes, often not exceeding 0.25 D. It appears
that the reduction in oedema may make the cornea less
susceptible to the mechanical force of the lens, and less
corneal change results.
Our study corroborates previous studies concerning the
minimal effect of high Dk RGP lenses on corneal curvature.
Previous studies have shown that during the ®rst three
months of lens wear, various types of silicone±acrylate
lenses, including extended wear lenses, did not produce
signi®cant changes in corneal curvature (Lydon and
Guillon, 1984; Rengstorff and Odby, 1986). Minimal
corneal variations were also found for longer periods of
lens wear (Odby and Rengstorff, 1985). Sanaty and Temel
(1996) found that the effect of ¯uorosilicone acrylate lenses
(Quantum, Bausch & Lomb) on central curvature was not
statistically signi®cant. Similar results were found by Pole
and Lowther (1987) when studying the effect of Paraperm
EW and Equalens lenses on corneal radii.
In summary we have demonstrated that Forum 210 lenses
do not induce signi®cant changes in corneal topography
over a 21-day period of lens wear. The implications of
this ®nding in terms of corneal health and optical quality
of the eye warrant further investigation over longer periods
of Forum 210 lenses wear.
The authors have no ®nancial interest in the manufacture,
supply or distribution of the contact lens used in this study.
Acknowledgements
This work was supported by CICYT research project ref.
PM 98-0225.
References
Box, G. E. P. and Jenkins, G. M. (1976). Time Serie Analysis:Forecasting and Control, Prentice Hall, Englewood Cliffs, NJ.
Briceno-Garbi, E. A. (1984). Variations in corneal curvature andrefractive error in CAB gas-permeable contact lens contact lenswearers. J. Am. Optom. Assoc. 55, 217±219.
Calossi, A., Verzella, F. and Zanella, S. G. (1996). Cornealwarpage resolution after re®tting an RPG contact lens wearerinto hydrophilic high water content material. CLAO J. 22,242±244.
Carney, L. G. (1974). Corneal topography changes during contactlens wear. Contact Lens J. 3, 5±16.
Carney, L. G. (1975). The basis for corneal shape change duringcontact lens wear. Am. J. Optom. 52, 445±454.
DeRubeis, M. J. and Shily, B. G. (1985). The effects of wearing theBoston II gas permeable contact lens on central cornealcurvature. Am. J. Optom. 62, 497±500.
Horner, D. G., Salmon, T. O. and Soni, P. S. (1998). Cornealtopography. In: Borish's Clinical Refraction (ed W. J.Benjamin), Saunders, London.
Hough, T. and Edwards, K. (1999). The reproducibility ofvideokeratoscope measurements as applied to the human cornea.Contact Lens and Anterior Eye 22, 91±99.
Rigid gas permeable contact lens and corneal topography: E. Yebra-Pimentel et al. 241
Table 1. Statistics for the analysis of regression for thecorneal radius at 1, 2, 3 and 4 mm from the center alongthe temporal, nasal, superior and inferior main meridians asa function of time. Data obtained from one eye are shown asan example
Location Distance Slope Constant R2 Sig. F
Temporal 1 20.0025 8.2729 0.0612 0.293Temporal 2 20.0019 8.2436 0.0395 0.4008Temporal 3 20.0012 8.2533 0.0205 0.5473Temporal 4 20.0011 8.3327 0.0167 0.5869Nasal 1 20.0025 8.3295 0.0476 0.3847Nasal 2 20.0025 8.3610 0.0498 0.3734Nasal 3 20.0016 8.4950 0.0325 0.4738Nasal 4 0.0019 8.6128 0.0180 0.5952Superior 1 20.0016 8.1443 0.0397 0.4945Superior 2 20.0017 8.1010 0.0375 0.5069Superior 3 20.0019 8.2103 0.0051 0.8088Superior 4 20.0012 8.4043 0.0213 0.6187Inferior 1 20.0005 8.1320 0.0025 0.8379Inferior 2 20.0005 8.1790 0.0031 0.8215Inferior 3 20.0005 8.1976 0.0036 0.8068Inferior 4 20.0033 8.3500 0.0687 0.2783
Hovding, G. (1983). Variations of central corneal curvature duringthe ®rst year of contact lens wear. Acta Ophthalmol. (Copenh.)61, 117±128.
Iskeleli, G., Oral, A. Y. and Celikkol, L. (1996). Changes incorneal radius and thickness in response to extended wear ofrigid gas permeable contact lenses. CLAO J. 22, 133±135.
Kame, R. T. and Kennedy, J. R. (1987). Corneal contour changewith 3M ¯uoropolymer lenses. Contact lens Spect. 2 (5), 53±59.
Kiele, P. M., Carney, L. G. and Smith, G. (1982). Diurnalvariations of corneal topography and thickness. Am. J. Optom.Physiol. Opt. 59, 976±982.
Lydon, D. and Guillon, M. (1984). Effect of center thickness on theperformance of rigid gas permeable lenses. Am. J. Optom. Phys.Opt. 61, 23.
Miller, D. (1968). Contact lens-induced corneal curvature andthickness changes. Arch. Ophthalmol. 80, 430±432.
Montenegro, J. R., Marfa, J. H. and Wilson, S. E. (1993). Cornealtopography alterations in normal contact lens wearers.Ophthalmology 100, 128±134.
Odby, A. and Rengstorff, R. H. (1985). The Boston lens: clinicalstudy in Sweden. Int. Contact Lens Clin. 12, 104.
Pole, J. J. and Lowther, G. E. (1987). Clinical comparison of low,moderate and high rigid gas permeable lenses. CL Forum 12 (7),47±51.
Polse, K. A., Sarver, M. D., Kenyon, E. and Bonanno, J. (1987).Gas permeable hard contact lens extended wear: ocular and
visual responses to a 6-month period of wear. CLAO J. 13,31±38.
Rengstorff, R. H. (1969a). Variations in corneal curvaturemeasurements: an after-effect observed with habitual wearers ofcontact lenses. Am. J. Optom. Arch. Am. Acad. Optom. 46,45±51.
Rengstorff, R. H. (1969b). Studies of corneal curvature changesalter wearing contact lenses. J. Am. Optom. Assoc. 40, 298±299.
Rengstorff, R. H. (1971). Diurnal variations in corneal curvatureafter the wearing of contact lenses. Am. J. Optom. Am. Acad.Optom. 51, 115±117.
Rengstorff, R. H. and Odby, A. (1986). Adaptation to Parapermextended wear lenses: a clinical study in Sweden. J. Am. Optom.Assoc. 57, 600.
Saks, S. J. (1966). Fluctuation in refractive state in adapting andlong-term contact lens wearers. J. Am. Optom. Assoc. 37,229±238.
Sanaty, M. and Temel, A. (1996). Corneal Curvature in soft andrigid gas permeable contact lens wearers after two years of lenswear. CLAO J. 22, 186±188.
Sevigny, J. (1986). Clinical comparison of the Boston IV contactlens under extended wear vs. the Boston II lens under daily wear.Int. Eye Care 2 (5), 260±264.
Wilson, S. E. and Klyce, S. D. (1994). Screening for cornealtopographic abnormalities before refractive surgery.Ophthalmology 101, 147±152.
242 Ophthal. Physiol. Opt. 2001 21: No 3