Refining Corneal Trephination and Donor Preparation for Artificial Cornea
Surgery
World Cornea Congress VIIApril 15 - 17 2015, San Diego 2015
Behrooz Azizi, MD; Ronan Conlon, MD; Salina Teja. MD; Joshua C Teichman, MD, MPH; Season Yeung, MBBS, FRANZCO;
Setareh Ziai, MD, FRCSC; Kashif Baig, MD, MBA, FRCSCUniversity of Ottawa Eye Institute, Ottawa, Ontario, Canada
Financial Disclosures
Drs. Azizi, Conlon, Teja, Yeung, Teichman, & Ziai have no financial relationships to disclose
Dr. Baig has the following relationships:
Financial Interest/ Affiliation Name of Company(s)
Grant/Research Support Allergan, Bausch & Lomb, Merck, Moria
Employment/Honoraria/Consulting Fees/Travel Expenses Alcon, Allergan, Bausch & Lomb, Labtician
Major Stock Shareholder None
Member: Advisory Panel, Standing Committee, Board of Directors Allergan, Bausch & Lomb
Other Financial or Material Interest None
Introduction
Corneal transplantation is a standard surgical treatment for many corneal diseases
However, there are situations in which replacement of the diseased cornea is best managed with an “artificial cornea”
The Boston keratoprosthesis (Kpro) is the most commonly utilized artificial cornea
Difficulty in achieving concentric punches
Khalifa YM, Moshirfar M. Improved centration of the type 1 Boston Keratoprosthesis in donor carrier tissue. Clin Ophthalmol. 2010 Aug 19;4:931-3
Difficulty in achieving concentric punches
Khalifa YM, Moshirfar M. Improved centration of the type 1 Boston Keratoprosthesis in donor carrier tissue. Clin Ophthalmol. 2010 Aug 19;4:931-3
Difficulty in achieving concentric punches
John T. Use of trypan blue corneal staining and centration ring in Boston keratoprosthesis surgery. Techniques in Ophthalmology. 2009 Dec. 7(4):127-130
Purpose
To devise a simple and cost-effective device for the preparation of the donor corneal tissue for Kpro surgery.
This device will perform the outer 8mm and inner 3mm cuts with increased accuracy and precision to improve centration.
Methods 30 donor corneas that were not suitable for transplantation
3 arms (1:1:1 design) 3mm followed by 8mm punch 8mm followed by 3mm punch Paired 3mm & 8mm punches utilizing the prototype
Computer software (Adobe Photoshop, San Jose, CA) was used to measure the centration of the punches by a masked examiner
Descriptive & inferential statistics were performed
Results – Decentration
Error bars = ± SEM
* P<0.0001
Results – Decentration
Range (mm) Avg (mm) SD (mm)
8mm 3mm 0.22 - 0.57 0.39 0.11 3mm 8mm 0 - 0.20 0.13 0.06 Paired 0 - 0.22 0.11 0.09
Conclusions
Our study demonstrates that pairing the two trephinations on the donor cornea with the use of a single device leads to increased predictability & ease of trephination