What is the HAZMATCADTM?

The HAZMATCAD™ is a next generation handheld trace chemical agent detector for Nerve (G) and Blister (H) agents as well as Hydrogen Cyanide (AC) or Phosgene (CG). The HAZMATCAD™ is designed to detect for these warfare agent gases with minimal interference from other chemicals which are often present in the battlefield or urban environment.

Page 4: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved
Page 5: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

The HAZMATCAD™ Plus expands the

performance of HAZMATCAD™ by adding the ability to detect selected TICs (Toxic Industrial Chemicals). These chemicals are detected and reported by class identification (hydride and halogen gases).

What is the HAZMATCAD™ Plus?

Page 6: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCAD™

SAW (nerve and blister)

One electrochemical cell (blood or choke)

HAZMATCAD™ Plus

SAW (nerve and blister)

Four electrochemical cells (blood, choke, halogen and hydride gases)

HAZMATCAD™ Configurations

Page 7: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Fast Response Time – IMS – Typically (5 to 30 sec)Direct readout of total contamination

No sample pre-collection or clean up

Increased risk of false positives

HAZMATCAD™ Analysis Time – SAW (20 to 120 sec)

Fast Mode (20 sec) or High Sensitivity Mode (120 sec)

Sample collection on absorbant media

Sample control

Reduced risk of false positives

False Alarms vs. Analysis Time

Page 8: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Small Piezoelectric Crystal

Interdigital Array

Polymer Surface Coating

RF Electronics to Drive SAW

Data Collection - frequency changes as mass absorbs into polymer

How Does a SAW Sensor Work?

Page 9: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

INTERDIGITALELECTRODES

RF AMP

QUARTZ CRYSTAL

FREQUENCY OUT

RAYLEIGHSURFACE WAVE

SORBENT POLYMERCOATING²f= k fo

2hp

SAW OSCILLATOR FREQUENCY IS HIGHLY SENSITIVETO SMALL CHANGES IN SURFACE MASS DENSITY

ON A TYPICAL 250 Mhz SAW SENSOR, ²f = 1000 hz / nanogramTHE SORBENT POLYMER DETERMINES WHICH VAPORS ARE DETECTED

SAW Sensor Fundamentals

Page 10: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Long Term Sensor Stability

Semi Selective Chemical Detection

Partition Coefficient Equilibrium Detection Mechanism

Quick Response and Fast Recovery

Better Sample Characterization

Benefits of SAW Technology

Page 11: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Sample Pump

ElectrochemicalCell

SAWArrayConcentrator

Assembly

SampleVent

SampleInlet

HAZMATCAD™ System Diagram

Page 12: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Sample Pump

E-Chem Cells (4)

SAWArray

ConcentratorAssembly

SampleVent

Sample Inlets

Sample Pump

ScrubberTube

SampleVent

HAZMATCAD™ PlusSystem Diagram

Page 13: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM and Rubber Boot

Internal Battery and 2 Li-Ion Batteries (NP-F550)

RS-232 Communication (Optional IrDA)

Pelican Storage and Carrying Case

External Battery Charger

Operations Manual

Verification Check Source

HAZMATCAD™ Components

Page 14: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Solid State SAW (Surface Acoustic Wave) Array

LED and Audible Tone Alarms

Quick Warm Up Time (less than 30 seconds)

Dual Mode Operation

20 Second Fast Mode Analysis Cycle Time

120 Second High Sensitivity Mode Analysis CycleTime

3 Concentration Levels Reported (High, Medium, Low)

AC or CG (TICs) Analysis in Real-time

Internal Data Log

Limited System Maintenance

HAZMATCAD™ Features

Page 15: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

LED Status Display

D C

B A

Status LED

A. SYS. OK

B. ALL CLEAR

C. LO BAT.

D. ALARM

Color

Green (blinking)

Green

Yellow

Red (blinking)

Meaning

Indicates proper electrical operation of the instrument

Indicates that no threat vapors are present

Indicates that a threat vapor has been detected

Indicates that battery power is running out

HAZMATCAD™ LED Screens

Page 16: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCAD™ LED Screens

Page 17: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCAD™ Plus LED Screens

Page 18: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCAD™ Operation- Power Up

Press On Button

LED will display (HCAD and Segment Test)

System Diagnostics

Battery Charge Indication and SYS. OK will illuminate

Mode of operation (FAST MODE DEFAULT)

Pump Activates – Concentrator Clean Up (30 Seconds)

First Analysis Begins and Displays Findings: All Clear – Green LED On Alarm Status – Red LED On

Page 19: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Alphanumeric LED Display

Power ButtonsAlarm Silence

Button

Mode Select Button

LED Status Display

HAZMATCAD™ Operation

Page 20: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Controls On / Off Switch Battery Door External Communication / RS-232 Connection

− DIN 9-Pin or IrDA (Optional Accessory)

Sample Inlet Port Three Position Open, Closed, and Dust Filter

Information / Status Panel Display LED 4-Character Alphanumeric

HAZMATCAD™ User Interface

Page 21: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Analysis Cycle is 20 Seconds Sample Collection

− Pump On− 1-11 Seconds

Sample Preheat− Pump Off / Concentrator On− 12-14 Seconds

Sample Injection− Pump On− 15-20 Seconds

Repeats Cycle Sequence Until Turned Off

HAZMATCAD™Analysis Cycle – Fast Mode

Page 22: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Analysis Cycle is 120 Seconds Sample Collection

− Pump On− 1-111 Seconds

Sample Preheat− Pump Off / Concentrator On− 112-114 Seconds

Sample Injection− Pump On− 115-120 Seconds

Repeats Cycle Sequence Until Turned Off

HAZMATCAD™ Analysis Cycle- High Sensitivity Mode

Page 23: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Press “MODE” Button

Current Mode will display on LED “FAST” or “HI S” (High Sensitivity)

While “Mode” is displayed, press “MODE” again

New Mode will flash on LED

A purge cycle will begin (i.e. Start Over)

HAZMATCAD™Change Mode Cycle

Page 24: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Press the ON Button

HAZMATCAD will perform a self test

When /////// is displayed on the LED, press and hold the “Mode Select” button

Audible alarm will beep to confirm training mode has started

After warm up, HAZMATCAD will chirp every three seconds

Turn off HAZMATCAD to deactivate

HAZMATCAD™Training Mode

Page 25: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM

Performance Nerve “G” Agent Detection

HAZMATCADTM Low Alarm Level

Type

GA (Tabun)

GB (Sarin)

GD (Soman)

GF

VX

Fast Mode

0.3 mg / m3

1.0 mg / m3

0.3 mg / m3

0.1 mg / m3

0.1 mg / m3

High Sensitivity

0.05 mg / m3

0.25 mg / m3

0.05 mg / m3

0.02 mg / m3

0.002 mg / m3

Page 26: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM

Performance Blister/Mustard “H” Agent Detection

HAZMATCADTM Low Alarm Level

Type

HD (Mustard)

HN3

(Nitrogen Mustard)

Fast Mode

0.4 mg / m3

High Sensitivity

0.09 mg / m3

0.4 mg / m3 0.09 mg / m3

Page 27: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM

Performance Blood and Choke Agent Detection

HAZMATCADTM Low Alarm Level

Type

Blood AC (Hydrogen Cyanide)

Choke CG (Phosgene)

Alarm Level

5.0 ppm

5.0 ppm

Page 28: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

-200

-100

0

100

200

300

400

500

600

DATA POINT

Series1

Series2

Series3

SAW Sensor Output

Page 29: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

F-76

-1000

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

10000

DATA POINT

Series1

Series2

Series3

G Agent

-150

-100

-50

0

50

100

150

200

250

300

350

DATA POINT

Series1

Series2

Series3

Mustard

-200

-100

0

100

200

300

400

500

600

700

800

DATA POINT

Series1

Series2

Series3

SAW Sensor Outputs

Page 30: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM

Plus PerformanceToxic Industrial Chemical Detection

“HYdR” Arsine AsH3 0.4 30 5 3.0 0.05

“HYdR” Diborane B2H6 3.0 30 5 15 0.1

“HYdR” Hydrogen Sulfide H2S 3.0 20 5 100 20

“HYdR” Hydrogen Cyanide HCN 10.0 25 5 50 10

“HALO” Chlorine Cl2 2.5 60 15 10 1

“HALO” Bromine Br2 4.0 45 10 3

“HALO” Nitrogen Dioxide NO2 5.0 15 5 20

ChemicalClass

Alarm TypeClassification

ChemicalName

Chem.Symbol

LowAlarm(ppm)

Resp.TimeLow

Alarm

Resp.Time

MediumAlarm

DLH(ppm)

PEL(ppm)

HydrideGas

HalogenGas

Page 31: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Check Source Verification

Chemical Simulant for “G” Agent−Dimethyl Methlyphosphonate (DMMP)

Verify Check Source Expiration Date−6 months from date on tube−MSDS - Appendix A in Operation Manual

A Successful Test indicates the HAZMATCADTM is fully operational for Nerve and Blister Agent Detection

No need to test separately for Blister

Page 32: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

How to use Simulant Check Source

Power Up HAZMATCADTM

Allow HAZMATCADTM to report “ALL CLEAR”

Prepare Check Source - remove vinyl end caps

Position Sample Inlet to “OPEN” position

Listen for Pump Operation−When pump pauses, wait 10 seconds−Hold Simulant Check Source to Sample Inlet for

3 seconds then remove

“G” Alarm should be indicated on next analysis

Allow HAZMATCADTM to return to “ALL CLEAR”

Page 33: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM Li-Ion Battery

2 SONY NP-F550 Rechargeable Lithium Ion Batteries

Can operate on one battery

8 to 12 hours of operation (dependent on Mode)• Fast Mode = 8 hours• High Sensitivity Mode = 12 hours

Can remove batteries while operating (HOT SWAP)

NP-F550 batteries require periodic charging

Typical recharge time is approximately 1 hour

Battery charger will indicate charge level

Page 34: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM Decontamination

Turn HAZMATCAD OFF

Place the three position sample port in the CLOSED position

Wipe down completely with moist, soapy cloth

Shake HAZMATCAD briskly to remove residual liquid

Place three position sample port in OPEN position and invert HAZMATCAD for two minutes

Wipe dry with clean towel, paper towel or clean rag

Place the HAZMATCAD in a sealed bag for one hour

Sample the headspace of the bag using another HAZMATCAD, M256 kit or other appropriate device

If residual contamination is detected, repeat steps 1 thru 8

Page 35: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM Check forInternal Contamination

Place an operating HAZMATCAD in a large plastic bag and seal it

Let the HAZMATCAD run for 60 minutes

Sample the headspace of the bag with another HAZMATCAD, M256 kit or other appropriate device for trace contamination

If contamination is detected, run the HAZMATCAD in clean air for several hours and repeat steps 1 thru 3

If HAZMATCAD still shows traces of contamination, contact Microsensor Systems, Inc. or the appropriate government agency

Page 36: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Data Log / Communication

External Modem Software Program (ProComm) ► Null Modem Cable ► 9600 Baud N-8-1 Serial Port Communication

Set up software to save file

When on, HAZMATCADTM will automatically transmit the data log

System parameters and data log will transmit to PC

Data log can be saved as a text file

Page 37: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

HAZMATCADTM Operation- Error Messages

Fault conditions displayed on LED ►Sensor 1 (S1), Sensor 2 (S2), Sensor 3 (S3)►CONC (Concentrator)►PUMP (Sample Pump)►XXXX Electrochemical Cell Failure►All system problems are recorded in Data Log

Low battery warning►Lo Bat. (Yellow LED) indication at 6.5V►System will not power up if battery voltage is less

than 6.0V

Page 38: Proudly Introduces HAZMATCAD™ HAZMATCAD™ PLUS ©MSA 2004 All Rights Reserved

Warnings / Precautions

Do Not Allow HAZMATCAD to Sample Water ► Avoid Rain Storms, Don’t Sample Close to Liquids

Do Not drop or abuse instrument

Do Not disassemble case - no serviceable parts

Do Not expose the HAZMATCAD to extreme high or low temperatures (Operational Range 0° to 40°C)

Charge Li-Ion batteries on a regular basis