promass q for the process industry · 2018-08 promass q for the process industry accuracy...
TRANSCRIPT
2018-08
Products Solutions Services
Promass Q for the process industry
A unique Coriolis innovation from Endress+Hauser
Slide 1 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Promass Q – Design
• Coriolis flowmeter
• Available on Proline 300/500
• Light weight, compact and drainable
• Constructed entirely from 316/316L stainless steel
• Four line sizes: DN 25 to 100 (1" to 4")
• All industry standard process connections:
• ASME, EN, JIS
• Process temperature range:
• Standard -50 °C to 205 °C (-58 °F to 400 °F)
• Cryogenic -196 °C to 180 °C (-320 °F to 356 °F)
• Max. process pressure: 102.1 bar (1480 psi)
Slide 2 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Size & weight
Promass Q face to face dimensions (installation length) are identical to
Promass F of the same nominal line size and flange connection!
Slide 3 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Key improvements & features
8 new improvements and features!
Pressure Drop
Turndown
Size &Weight
Safety
RepeatabilityProcess
Sensitivity
Entrained Gas
Density
Designed for real benefit!
Slide 4 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Applications for Promass Q – Process
Hydrocarbon custody transfer
LNG NGL light hydrocarbons cryogenicsHeavy crudes
Well cementLiquid leg of separators
Slide 5 Product Management Endress+Hauser Flowtec
2018-08
Entrained Gas
Pressure Drop
Promass Q for the process industry
Key improvements & features
Turndown
Size &Weight
Safety
RepeatabilityProcess
Sensitivity
Density
Slide 6 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Changing viscosity and density sensitivity
Slide 7 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Density vs. density
Slide 8 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Density vs. viscosity
Slide 9 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Water across pressure & temperature
Slide 10 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Isobutanol across pressure & temperature
Slide 11 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Ambient temperature sensitivity on density
But what about when my meter is installed:
HERE!?
- OR -
HERE?!
Slide 12 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Ambient temperature sensitivity on density
Slide 13 Product Management Endress+Hauser Flowtec
2018-08
Pressure Drop
Promass Q for the process industry
Key improvements & features
Turndown
Size &Weight
Safety
RepeatabilityProcess
Sensitivity
Entrained Gas
Density
Slide 14 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Promass Q has a very wide useable turndown
Yellow line Blue line
Flow where thebase accuracy begins
Flow where 1 barpressure loss is reached
Useable turndown
1 bar pressure loss
0.05
Operating conditions
Promass QDN 100 (4”)
Slide 15 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Pressure drop specifications
Promass Q offers the highest flowrates at the lowest pressure drops
Opens the ability to utilize smaller meters with less pressure loss!
Slide 16 Product Management Endress+Hauser Flowtec
2018-08
Entrained Gas
Pressure Drop
Promass Q for the process industry
Key improvements & features
Turndown
Size &Weight
Safety
RepeatabilityProcess
Sensitivity
Density
Slide 17 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Accuracy statements & GVF
• All three of the above have 15% GVF.• Will a Coriolis meter respond in the same way to these situations? • No. It cannot due to the differing motion of the fluid relative to the tube motion.
• Over simplification of reality can be misleading. Making a hard stated accuracy specification is not responsible nor is it especially useful.
Slide 18 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Multi-Frequency Technology (MFT) – Patented!
Raw ሶ𝑚 und
1st mode
2ndmode
Corrected ሶ𝒎 and
Active compensation!
MFT - the same set of dual tubes are
oscillated at two different resonant
frequencies or modes simultaneously.
The Coriolis meter is measuring the same
fluid with the same amount of entrained
gas, however the way these two modes
react to this will be different.
Through analyzing these results and the
unique resonance properties of the two
modes, we are then able to effectively
and actively compensate for the
associated measurement error.
Additional information
*Animation exaggerated for clarity
Slide 19 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Outgassing field trial results
Slide 20 Product Management Endress+Hauser Flowtec
2018-08
Promass Q for the process industry
Entrained gas examples – Process
Ou
tgas
sin
g:
Cry
oge
nic
N2
:
Wel
l cem
ent:
2% GVF | <0.5% | mass flow 3% GVF | 0.3% | mass flow
10% GVF | <1% | density
Em
uls
ion
:
5% GVF | <1% | mass flow
60
cP o
il:
8% GVF | <2% | mass flow
Slide 21 Product Management Endress+Hauser Flowtec
2018-08
Products Solutions Services
Thank you very much for your attention
Slide 22 Product Management Endress+Hauser Flowtec
2018-06
Products Solutions Services
Proline Promass A
For process applications
Slide 1 Coriolis Product Management
2018-06
Smallest footprint for integration – compact and lightweight
Proline Promass A – for process applications
DN 04 (1/8") 5.15 kg (11 lbs)310 mm (12.2 in)
DN 02 (1/12") 3.3 kg (7 lbs)264 mm (10.39 in)
DN 01 (1/24") 1.75 kg (4 lbs)187 mm (7.36 in)
(sensor only, fitted with VCO)
Slide 2 Coriolis Product Management
2018-06
New Promass A – key characteristics at a glance
Proline Promass A – for process applications
Topic New Promass A design
Measuring tube materials 1.4435 (316/316L) stainless steel & Alloy C22
Sensor housing material Complete external design in 316L stainless steel (suitable for offshore)
Max pressure rating 430.9 bar / 6250 psi (corresponds to ASME B16.5 Material Group 3.8)
Medium temperature range -50 to +205 °C (-58 to +401 °F)
Mass and volume flow accuracy liquids ±0.10%
Mass and volume flow repeatability liquids ±0.05%
Mass flow accuracy gas ±0.35%
Mass flow repeatability gas ±0.15%
Density accuracy ±0.0005 g/cm3 field density & ±0.002 g/cm3 wide range density
Temperature accuracy (ref. conditions) ±0.5 °C (±0.9 °F)
Transmitter and process connection offer See next slides
Slide 3 Coriolis Product Management
2018-06
Transmitter overview new Promass A sensor design
Proline Promass A – for process applications
Transmitter
Material
100 200 300 500500
digital
Standard
aluminum
Heavy duty, offshore
stainless, cast CF3M “316L”- - -
Hygienic compact
stainless, sheet-
Hygienic ultra-compact
stainless, sheet- -
Slide 4 Coriolis Product Management
2018-06
Process connection offer overview
Proline Promass A – for process applications
Flanges
1.4404 (316/316L) Alloy C22
Tri-Clamp
1.4435 (316L)
4-VCO-4 G ¼" NPT ¼"
1/2"≤ 40 bar (580 psi)
EN 1092-1
≤ PN 400
JIS B2220
≤ 63K
ASME B16.5
≤ Cl. 2500 (430.9 bar (6250 psi))
Couplings and threads
1.4404 (316/316L)Alloy C22
All of the above ≤ 430.9 bar (6250 psi)
Note: All process connections are directly welded on the sensor (no adapters needed)
All flanges: DN 15 (½")
Slide 5 Coriolis Product Management
2018-06
Process connection offer overview
Proline Promass A – for process applications
Tri-Clamp and all flanges: DN 15 (1/2’’) (*) In addition also several Alloy C22 lap joint flange versions are available
Standard –All wetted parts in st. steel
Standard –All wetted parts in Alloy C22
High pressure –Measuring tube always in C22
Tri-Clamp
≤ PN 100 ≤ PN 100 (*) ≤ PN 400 (St. steel or C22)
≤ Cl. 600 ≤ Cl. 600 (*) ≤ Cl. 2500 (St. steel or C22)
≤ JIS 63K ≤ JIS 63K (*)
4-VCO-4 4-VCO-4 4-VCO-4 (C22 only)
G 1/4" internal thread G 1/4“ internal thread G 1/4" int. thread (C22 only)
NPT 1/4" internal thread NPT 1/4" internal thread NPT 1/4" int. thread (C22 only)
Slide 6 Coriolis Product Management
2018-06
• Superior corrosion resistance – complete exterior
sensor design in 316L as standard
• Directly welded process connections (no gaskets or
adapters)
• International hazardous area approvals
• Developed according to IEC 61508 (SIL)
• Sensor housing burst pressure >100 bar (1450 psi)
• Additional safety layer – rupture disk
• always included for high pressure versions
• optional for standard versions
• Significant reduced risk of particle clogging due to
single tube design (cp. to dual tube systems with
similar flow rate capacity, see next slide)
Safety by design
Proline Promass A – for process applications
Predefined breaking point – ruptures at a pressure of 10 to 15 bar (145 to 217.5 psi)
Rupture disk
Slide 7 Coriolis Product Management
2018-06
Significant reduced risk of particle clogging due to single tube design
Promass A vs competition – reduced risk of particle clogging
Proline Promass A – for process applications
Other brand – dual tube designPromass A – single tube design
Slide 8 Coriolis Product Management
2018-06
Proline Promass A – for process applications
Services, non destructive tests (NDTs), approvals and certificates
A full range of services, NDTs, approvals and certificates are available to provide documented evidence
that Promass A fulfills its stated high quality standards. How to order:
• Group 570, option
• HA: Cleaned from oil + grease (wetted parts)
• Group 580, option
• JB: Pressure test
• JC: 3.1 material certificate wetted parts + sensor housing
• JK: Positive material identification (PMI) test
• KE, KI, KN, KS: X-ray test (film) acc. to ISO 10675-1, ASME B31.3, ASME VIII & NORSOK M-601
• K5, K6, K7, K8: X-ray test (digital) acc. to ISO 10675-1, ASME B31.3, ASME VIII & NORSOK M-601
• Group 590, option
• LA: SIL
• LD: CRN
• LN: Sensor housing burst pressure, type test
• LR: NACE MR0175 / ISO 15156
• LS: NACE MR0103 / ISO 17945
Slide 9 Coriolis Product Management
2018-06
New dimension of low flow measurement – innovative robust sensor design
Proline Promass A – for process applications
Slide 10 Coriolis Product Management
2018-06
New dimension of low flow measurement – innovative robust sensor design
Proline Promass A – for process applications
• Higher frequency vibrational symmetry very stable zero point
• Sets the bar for small DN measuring performance
PRELIMINARY VIDEO – WORK IN PROGRESS
Slide 11 Coriolis Product Management
2018-06
Useable turndown – definition
Proline Promass A – for process applications
Green line Blue line
Flow where thebase accuracy begins
Flow where 1 barpressure loss is reached
Useable turndown
1 barpressure loss
Slide 12 Coriolis Product Management
2018-06
New Promass A design with significantly improved useable turndown
Proline Promass A – for process applications
Improvedzero point stability
Largermeasuring
tubes
HP: High pressure
Slide 13 Coriolis Product Management
2018-06
New Promass A DN 01 – new dimension of low flow measurement
Proline Promass A – for process applications
1.0%0.05 kg/h
=0.8 g/min
0.1%0.48 kg/h
= 8.0 g/min
0.1%1.2 kg/h
= 20.0 g/min
Other Brand DN 06
Other Brand DN 04
Other Brand DN 02
Other Brand DN 01
Slide 14 Coriolis Product Management
2018-06
…including high pressure applications
Proline Promass A – for process applications
0.1%0.8 kg/h
= 13.3 g/min
0.1%4.4 kg/h
= 73.3 g/min
1.0%0.08 kg/h
=1.3 g/min
Other Brand DN 04-HP
Other Brand DN 02-HP
Slide 15 Coriolis Product Management
HP: High pressure
2018-06
New design – stable performance over wide a viscosity range
Proline Promass A – for process applications
Slide 16 Coriolis Product Management
2018-06
Recommended when there are concerns about the mechanical strength of the
piping or the rotational stability of the installation (Tri-Clamp, VCO, G, NPT)
Universal sensor holder
Proline Promass A – for process applications
One design for all DN/purposes
Sensorneck
Wall, table or pipe
Promass A 300DN 04 VCO pipe mount
Sensorholder
How to order
• Group 620, option PR: Sensor holder (for wall, table and pipe mount)
• As an accessory: material number 71392563
Slide 17 Coriolis Product Management
2018-06
Heating jacket for the new Promass A sensor design
For applications where keeping a constant process temperature is important
Proline Promass A – for process applications
DN 01(¹⁄₂₄")
DN 02 (¹⁄₁₂")
DN 04(¹⁄₈")
How to order
• Group 620, option RB: Heating jacket, G ½" internal thread
• Group 620, option RD: Heating jacket, NPT ½" internal thread (served with a "G NPT" adapter)
• Via separate structure DK8003 (material number 71386499)
Slide 18 Coriolis Product Management
2018-06
Proline Added Values
Proline Promass A – for process applications
Data storage concept
Seamless system integration Web server
Heartbeat Technology
Operation concept (HMI)
W@M Life Cycle Management
Slide 19 Coriolis Product Management
2018-06
Heartbeat Technology – reliable and safe processes
Proline Promass A – for process applications
FunctionalSafety (SIL)
Developed acc. to
IEC 61508
Slide 20 Coriolis Product Management
2018-06
Products Solutions Services
Promass A – The lightweight champion for lowest flow rates and highest performance
Thank you
Slide 21 Coriolis Product Management