proline promass f heating jacket operation instructions

Upload: prasanna-kumar

Post on 14-Apr-2018

225 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    1/20

    The Most Trusted Name In Measurement

    Coriolis Mass Flowmeter

    Promass F Heating Jacket

    Operating Instructions

    Issue/Rev. 0.0 (5/09) Bulletin MN0M036

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    2/20

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    3/20

    Heating jacket for Promass F Cont

    Contents

    1 Safety instructions . . . . . . . . . . . . . . . . 4

    1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    1.2 Installation, commissioning and operation . . . . . . . . 41.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    1.5 Notes on safety conventions and icons . . . . . . . . . . . 5

    2 Identification . . . . . . . . . . . . . . . . . . . . 6

    2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6

    2.1.1 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . 6

    2.2 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 6

    3 Installation . . . . . . . . . . . . . . . . . . . . . . 7

    3.1 Incoming acceptance and transport . . . . . . . . . . . . . 73.1.1 Incoming acceptance . . . . . . . . . . . . . . . . . . 7

    3.1.2 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 7

    3.2.1 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . 7

    3.2.2 Temperature ranges . . . . . . . . . . . . . . . . . . . 7

    3.2.3 Orientation . . . . . . . . . . . . . . . . . . . . . . . . . 8

    3.3 Installing the heating jacket . . . . . . . . . . . . . . . . . . . 8

    3.3.1 Mounting kit . . . . . . . . . . . . . . . . . . . . . . . . 8

    3.3.2 Mount the isolation . . . . . . . . . . . . . . . . . . . 8

    3.3.3 Connecting to in-house heating pipes . . . . . 9

    3.4 Postconnection check . . . . . . . . . . . . . . . . . . . . . 10

    4 Technical data . . . . . . . . . . . . . . . . . . 11

    4.1 Technical data at a glance . . . . . . . . . . . . . . . . . . . 11

    4.1.1 Application . . . . . . . . . . . . . . . . . . . . . . . . 11

    4.1.2 Function and system design . . . . . . . . . . . . 11

    4.1.3 Operating conditions . . . . . . . . . . . . . . . . . 11

    4.1.4 Operating conditions: environment . . . . . . 11

    4.1.5 Operating conditions: Process . . . . . . . . . . 11

    4.1.6 Mechanical construction . . . . . . . . . . . . . . 13

    4.1.7 Certificates and approvals . . . . . . . . . . . . . 13

    4.2 Dimensions of heating jacket (SI Units) . . . . . . . . . 14

    4.2.1 Promass F . . . . . . . . . . . . . . . . . . . . . . . . . 14

    4.2.2 NPT Adapter . . . . . . . . . . . . . . . . . . . . . . . 154.2.3 Flange Adapter . . . . . . . . . . . . . . . . . . . . . 15

    4.3 Dimensions of heating jacket (US Units) . . . . . . . . . 16

    4.3.1 Promass F . . . . . . . . . . . . . . . . . . . . . . . . . 16

    4.3.2 NPT Adapter . . . . . . . . . . . . . . . . . . . . . . . 17

    4.3.3 Flange Adapter . . . . . . . . . . . . . . . . . . . . . 17

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    4/20

    Safety instructions Heating jacket for Promass F

    1 Safety instructions

    1.1 Designated use

    The heating jacket serves to transfer heat to maintain temperature of fluids in the Promass sensorsof the F.

    Resulting from incorrect use or from use other than that designated, the operational safety of the

    measuring devices can be suspended. The manufacturer accepts no liability for damages being

    produced from this.

    1.2 Installation, commissioning and operation

    Installation, commissioning and maintenance of the instrument must be carried out exclusively by

    trained specialists authorised by the facilitys owneroperator. The specialist must have read and

    understood these Operating Instructions and must adhere to them.

    The manufacturer reserves the right to modify technical data without prior notice. Your FMC

    Technologies distributor will supply you with current information and updates to these OperatingInstructions.

    1.3 Operational safety

    Note the following points:

    Heating jackets can not be combined with sensor which have purge connections and/or rupture

    disks.

    When using heating jackets, it is recommended to use the remote version of the sensor and

    transmitter.

    When using the remote version, and particularly with the Promass F, at very high fluid and

    heating temperatures, we recommend the use of the "longnecked version"

    (= remote version for heating), so that the ambient temperature of the connection housing and

    the connecting cable is not exceeded.

    With the compact version, make sure you take appropriate measures (orientation, etc.) so that the

    maximum ambient temperature for the transmitter is not exceeded, otherwise electronic

    components will be driven outside their specifications. You should particularly avoid this for

    components which are used for explosion protection.

    The permitted pressure of the heating medium is specified on the nameplate Page 6.

    Heating mediums are acceptable, water, water vapour, oil and other noncorrosive liquids.

    If using oil as heating medium consultation is to be held with Endress+Hauser.

    Deposits by dirty heating media or calcareous water can limit an even distribution of the heating

    medium in the heating jacket. A faultless function of the heating jacket will not be then ensured.

    #Warning!Danger of injury/burning from hot surfaces!

    During operation, the surfaces of the heating jacket and those of the conveyance lines for the heating

    medium may become very hot and will lead to burns if touched.

    The customer is responsible for mounting insulation.

    If the heating jacket and the conveyance lines are insulated, you must wait a long enough period of

    time for the jacket to cool after the insulation is removed.

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    5/20

    Heating jacket for Promass F Safety instruct

    1.4 Return

    The following procedures must be carried out before a flowmeter requiring repair, for example

    returned to FMC Technologies:

    Always enclose a duly completed "Declaration of contamination" form. Only then can FMC

    Technologies transport, examine and repair a returned device.

    Enclose special handling instructions if necessary, for example a safety data sheet as perEN 91/155/EEC.

    Remove all residues. Pay special attention to the grooves for seals and crevices which could

    contain residues. This is particularly important if the substance is hazardous to health, e.g.

    flammable, toxic, caustic, carcinogenic, etc.

    ! Note!You will find a preprinted "Declaration of contamination" form at the back of this manual.

    # Warning! Do not return a heating jacket if you are not absolutely certain that all traces of hazardous

    substances have been removed, e.g. substances which have penetrated crevices or diffused

    through plastic.

    Costs incurred for waste disposal and injury (burns, etc.) due to inadequate cleaning will becharged to the owner operator.

    1.5 Notes on safety conventions and icons

    The heating jackets are designed to meet stateoftheart safety requirements, have been teste

    and left the factory in a condition in which they are safe to operate. They can, however, be a sou

    of danger if used incorrectly or for use other than that designated.

    Consequently, always pay particular attention to the safety instructions indicated in these Operat

    Instructions by the following icons:

    # Warning!"Warning" indicates an action or procedure which, if not performed correctly, can result in inju

    or a safety hazard. Comply strictly with the instructions and proceed with care.

    " Caution!"Caution" indicates an action or procedure which, if not performed correctly, can result in incorr

    operation or destruction of the device. Comply strictly with the instructions.

    ! Note!"Note" indicates an action or procedure which, if not performed correctly, can have an indirec

    effect on operation or trigger an unexpected response on the part of the device.

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    6/20

    Identification Heating jacket for Promass F

    2 Identification

    2.1 Device designation

    The heating jackets of the Promass F consist of two halfshells.

    2.1.1 Nameplate

    a0006934

    Fig. 1: Nameplate specifications for heating jackets (example)

    1 Ordering code / Serial number: See the specifications on the order confirmation for the meanings of the individual

    letters and digits.

    2 Heating jacket material

    3 Process connection of the heating jacket

    4 Maximum permitted pressure (heating medium)

    5 Maximum permitted temperature (heating medium)

    6 Corresponding sensor7 Reserved for information on special products

    8 Please refer to operating instructions / documentation

    2.2 Certificates and approvals

    The devices are designed to meet stateoftheart safety requirements, have been tested, and left

    the factory in a condition in which they are safe to operate. The devices correspond to article 3 (3)

    of the EEC guideline 97/23/BG (Pressure Equipment Directive) and are laid out and manufactured

    after good engineer practice.

    The devices described in these Operating Instructions thus complies with the statutory

    requirements of the EC Directives. FMC Technologies confirms successful testing of the device.

    DK8HJ-XXXX

    Heating Jacket

    F DN50

    G3/4" / RP3/4"

    25bar / 375psi

    +200C / 392F

    1.4571 / 316Ti

    XXXXXXXXXXXXSer.No.:

    Order Code:

    For Promass Sensor Type:

    Material:

    Process Connection:

    p max.:

    TM max.:

    3.1B

    321566-0001A

    1

    2

    3

    4

    5

    7

    i8

    6

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    7/20

    Heating jacket for Promass F Installa

    3 Installation

    3.1 Incoming acceptance and transport

    3.1.1 Incoming acceptance

    On receipt of the goods, check the following points:

    Check the packaging and the contents for damage.

    Check the shipment, make sure nothing is missing and that the scope of supply matches you

    order.

    3.1.2 Transport

    Transport the devices in the containers in which they are delivered.

    3.2 Installation conditions

    3.2.1 Dimensions

    The dimensions are to be found on Page 14.

    3.2.2 Temperature ranges

    Temperature ranges of the heating medium find you on Page 11.

    Heating medium Operating temperature

    "Caution! Always select the orientation of the sensor and the heating jacket so that the maximum ambi

    temperatures for the measuring electronics are not exceeded Page 11.

    With additional heating jacket insulation, always keep to the minimum safety distances to th

    device Page 14 ff.

    Keep to the following ambient temperatures depending on the design of the measuring syste

    The material load curves (pressuretemperature diagrams) for the heating jacket are to be found

    Page 13.

    Device version Max. ambient temperatures

    Compact version Transmitter: maximal +60 C (+140F)

    Remote version (standard

    version)

    Connection housing: maximum +60 C (+140F)

    Operating temperature of connecting cable: maximum +105 C (+221F)

    Remote version for heating(longnecked version)

    Always use the longnecked version when the max. ambient temperature of +60 C(+140F) for using the standard remote version cannot be kept to. The longnecked vers

    has a housing support for the thermal separation of sensor and transmitter.

    This version is used, for example, for applications in which very high fluid and heating

    medium temperatures are encountered.

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    8/20

    Installation Heating jacket for Promass F

    3.2.3 Orientation

    3.3 Installing the heating jacket

    3.3.1 Mounting kit

    Note the following points:

    Screws, nuts, etc. are included in the scope of supply.

    Dimensions and required space Page 14 ff.

    Promass F

    Join the two halfshells of the heating jacket by nuts and bolts.

    #Warning!With large nominal diameters, a support may be required due to the heavy weight of the sensor and

    the heating jacket. Never use the feed and return pipes of the heating system to support the heating

    jacket!

    The weight data for the sensors are listed in the "Technical data" of the corresponding Operating

    Instructions.

    3.3.2 Mount the isolation

    With additional heating jacket insulation, always keep to the minimum safety distances to the device

    Page 14 ff.

    Vertical Horizontal Horizontal Inclined Inclined

    Transmitterat the side at the top

    at the

    bottomat the bottom at the top

    Sketch

    a0004572

    a0004576 a0004580

    A0008136A0008137

    Promass FRecommended

    Recommended

    (m)

    Recommended

    (n, o)

    Recommended

    (m,p)

    Recommended

    (n, o, q)

    Promass F HT

    Campact

    version

    Recommended Not suitableRecommended

    (n, o)

    Not suitable/

    Possible (m,p)

    Recommended

    (n, o, q)

    Promass F HT

    Remote versionRecommended Possible (m)

    Recommended

    (n, o)

    Recommended

    (m,p)

    Recommended

    (n, o, q)

    T = Process connection heating jacket

    HT = High temperature version for medium temperatures (TM) > 200 C (> 392 F)

    m This orientation is not suitable for fluids with entrained solids.

    n This orientation is not suitable for outgassing fluids.

    o This orientation is not suitable for low fluid temperatures.

    p Depending on medium temperatures.

    q Depending on heating medium.

    T

    T

    TD

    T

    T

    E T

    T

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    9/20

    Heating jacket for Promass F Installa

    3.3.3 Connecting to in-house heating pipes

    Designing and making calculations for a tempered heating system is generally based on the high

    heating needs ever to be expected. Pipe network sizing is based on the permitted flow velocitie

    the heated media (steam, water, oil) or pipe friction loss. The pump capacity, and thus the sizin

    parameters, result from the total of the overall resistances in the heating system together with

    required amount of heated media.Partial load operation always means an increase in the differential pressure in the overall system

    the case of normal, non-regulated pumps, this effect is intensified by the increasing pump

    characteristic. In the case of regulated pumps (p = constant), the pump pressure is maintaine

    even if the total flow is reduced but is passed on almost completely to existing, still open valves

    pipe restrictions due to the greatly reduced pipe loss.

    " Caution!Heating pipes with thermal oils

    When thermal oils are used, a high differential pressure can occur during system startup. This is

    to the high viscosity of thermal oils at low temperatures.

    Differential pressure regulating values or overflow valves can be used to reduce an increase in h

    differential pressures. The overflow valve should be installed between the feed and return sectof the heating jacket. The valve limits the feed pressure of the circulation pump and ensures a

    minimum volume flow in the boiler system by opening as soon as the feed pressure exceeds th

    value set at the valve. This value should not be above the permitted nominal pressure of the heat

    jacket.

    Once the heating jacket, and the oil it contains, are heated up, an differential pressure according

    Fig. 3 (Fig. 4) on Page 12 is achieved.

    A

    Fig. 2:

    1 Promass heating jacket

    2 Shutoff valve

    3 Dirt trap

    4 Overflow valve

    1

    22 3

    4

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    10/20

    Installation Heating jacket for Promass F

    3.4 Postconnection check

    Perform the following checks after completing the installation:

    Device condition and specifications Notes

    Is the device damaged (visual inspection)?

    Does the device correspond to specifications at the measuring point ? Page 11 ff.

    Installation Notes

    Are measuring point number and inscription correct (visual inspection)?

    Was the correct installation position for the heating jacket selected? Page 14 ff.

    Is safety margin kept by 20 mm (0.79") between user isolation and Sensor? Page 14 ff.

    Process environment / process conditions Notes

    Does the run of the line to the connections at the heating jacket accordingly shifted?

    (seal the connections)?

    Is the device sufficiently secured against vibrations (attachment, support)? Acceleration up to 2 g by analogy

    with IEC 6826

    The maximum pressure rating of heating system is not exceeded? Page 11 ff.

    Flow of heated medium ensured? Pump characteristic checked? Page 9 ff.

    Customer documents on

    heating system

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    11/20

    Heating jacket for Promass F Technical

    4 Technical data

    4.1 Technical data at a glance

    4.1.1 Application

    The heating jacket serves to transfer heat to maintain temperature of fluids in the Promass senso

    Chocolate, condensed milk, liquid sugar

    Oils, fats

    Lacquers

    Pharmaca, catalysts, inhibitors

    Suspensions

    4.1.2 Function and system design

    Design The heating jacket of the Promass F consists of two halfshells.

    4.1.3 Operating conditions

    Installation instructions Installation position arbitrary (horizontal or vertically) restrictions and further references

    Page 7 ff.

    4.1.4 Operating conditions: environment

    Ambient temperatureTransmitter or connection

    housing

    All versions: max. +60C (+140 F)Detailed description Page 7

    4.1.5 Operating conditions: Process

    Heating medium temperature

    range

    Heating jacket/ Promass F: max. +200 C (+392 F)

    Heating jacket / Promass F (high temperature): max. +350 C (+662 F)

    Pressure rating of heating

    system

    Heating jackets for all sensors: 25 bar (363 PSI)

    Heat transfer coefficient DN 8 15 25 40 50 80 100 15

    Heat transfer

    coefficient* in C/h130 130 130 120 110 90 70 40

    * uninsulated

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    12/20

    Technical data Heating jacket for Promass F

    Pressure loss Typical pressure loss of the heating jacket for Promass F independent of the nominal diameter, for

    a half-shell (parallel/series installation).

    " Caution!With serial installation the pressure losses of each half-shell add themselves.

    SI Units

    a0008130en

    Fig. 3: Pressure loss diagram for 1 cP and 30 cP (SI units)

    US Units

    a0008130ae

    Fig. 4: Pressure loss diagram for 1 cP and 30 cP (US units)

    30 cP

    1 cP

    10000

    1000

    100

    1010 100 1000 10000

    [mbar]

    [kg/h]

    100

    10

    1

    0.1 1 10 100

    [psi]

    [lb/min]

    30 cP

    1 cP

    0.1

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    13/20

    Heating jacket for Promass F Technical

    4.1.6 Mechanical construction

    Design, dimensions Page 14 ff.

    Weight Page 14 ff.

    Materials Heating jackets: 1.4571 / 316 Ti

    Material load diagrams Heating jackets

    Materials: 1.4571 / 316 Ti

    " Caution!The following material load diagrams (reference curves) are referring to a temperature range of

    50 to +350 C (58 to +662 F). But the maximal permissible fluid temperature always depe

    on sensor.

    A00

    Fig. 5: Material load through the heating medium (pressuretemperature diagram)

    4.1.7 Certificates and approvals

    Pressure device approval The devices correspond to article 3 (3) of the EEC guideline 97/23/BG (Pressure Equipment

    Directive) and are laid out and manufactured after good engineer practice.

    18

    19

    20

    21

    22

    24

    23

    25

    26

    17

    [bar]

    50 0 50 100 150 200 250 300 350 [C]

    [psi]

    -40 120 200 440 520 [F]40 600280 360

    280

    270

    300

    290

    310

    320

    330

    340

    350

    260

    360

    370

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    14/20

    Technical data Heating jacket for Promass F

    4.2 Dimensions of heating jacket (SI Units)

    US Units Page 16 ff.

    4.2.1 Promass F

    A0008131

    Fig. 6: Heating jacket Promass F

    R Insulation, customer supply (e.g. mineral wool)

    S Safety distance: min. 20 mm

    T Process connection of the heating jacket acc. to tabele

    C

    T

    A

    BT

    R

    D

    S

    G1

    H

    G2

    K

    FE

    DN 25HT40(1"HT1")

    FE

    DN 5080

    (2"3")

    FE

    DN 825(3/8"1")

    FE

    DN 150(5")

    FE

    DN 100(4")

    DN A B C D E F G1 G2 H K

    8, 15, 25 292 167.7 52 154 40 106 121 81 27 100.5

    25HT 335 182.2 66 140 50 117.5 142.5 102.5 27 114

    40 444 182.9 66 140 50 172 197 157 27 110.6

    50, 50HT 580 214.4 86 154 50 160 257.5 202.5 32.5 130

    80, 80HT 710 293.9 125 195 50 195 322 262.5 37.5 193.2

    100 840 367 146 214 50 170 387 325 47.5 247.7

    150 1040 431 213 287 415 105 487.5 418 50 291

    DN 825 25HT 40 50, 50HT 80, 80HT 100 150

    weight* [kg] 1.25 1.9 2.5 3.9 6.8 10.3 15

    volume* [Liter] 0.15 0.15 0.2 0.3 0.5 0.6 1.0

    All dimensions in [mm]; HT = high temperature; *per Jacket

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    15/20

    Heating jacket for Promass F Technical

    4.2.2 NPT Adapter

    A

    Abb. 7: Dimensions of Adapter NPT inner thread on conical ISO external thread

    4.2.3 Flange Adapter

    Abb. 8: Dimensions of Adapter for Flanges

    A

    B

    C ISO

    SNP

    T

    NPT ISO2) S A1) B1) C1)

    G 1 1/16 AF (SW 27) 49.3 19.1 11.9

    G 1 5/16 AF (SW 33) 51.3 19.1 15.7

    1) Dimensions in [mm]2) DIN EN ISO 228

    S

    IS

    O

    D

    L

    EF

    G

    Flange according to ASME B16.5: 1.4404/316L/316

    Pressure stage ISO2) D1) E1) F1) G1) L1) S

    Cl 150G 15.7 60.5 88.9 4 15.7 87.8

    SW 2

    (7/8 A

    G 15.7 60.5 88.9 4 15.7 87.8

    Cl 300G 15.7 66.5 95.2 4 15.7 92.3

    G 15.7 66.5 95.2 4 15.7 92.3

    Flange EN 1092-1 (DIN 2501) / PN 40: 1.4404/316L/316

    Pressure stage ISO2) D1) E1) F1) G1) L1) S

    PN 40G 17.3 65.0 95 4 14 78.0 SW 2

    (7/8 AG 17.3 65.0 95 4 14 78.0

    1) Dimensions in [mm]2) DIN EN ISO 228

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    16/20

    Technical data Heating jacket for Promass F

    4.3 Dimensions of heating jacket (US Units)

    SI Units Page 14 ff.

    4.3.1 Promass F

    A0008131

    Fig. 9: Heating jacket Promass F

    R Insulation, customer supply (e.g. mineral wool)

    S Safety distance: min. 0.79"

    T Process connection of the heating jacket acc. to tabele

    C

    T

    A

    BT

    R

    D

    S

    G1

    H

    G2

    K

    FE

    DN 25HT40(1"HT1")

    FE

    DN 5080(2"3")

    FE

    DN 825(3/8"1")

    FE

    DN 150(5")

    FE

    DN 100(4")

    DN A B C D E F G1 G2 H K

    3/8"1" 11.50 6.60 2.05 6.06 1.57 4.17 4.76 3.19 1.06 3.961" HT 13.19 7.17 2.60 5.51 1.97 4.63 5.61 4.04 1.06 4.49

    1" 17.48 7.20 2.60 5.51 1.97 6.77 7.76 6.18 1.06 4.35

    2", 2"HT 22.83 8.44 3.39 6.06 1.97 6.30 10.14 7.97 1.28 5.12

    3", 3"HT 27.95 11.57 4.92 7.68 1.97 7.68 12.68 10.33 1.48 7.61

    4" 33.07 14.45 5.75 8.43 1.97 6.69 15.24 12.80 1.87 9.75

    6" 40.94 16.97 8.39 11.30 16.34 4.13 19.19 16.46 1.97 11.46

    DN 3/8"1" 1"HT 1" 2", 2"HT 3", 3"HT 4" 6"

    weight* [lbs] 2.8 4.2 5.5 8.6 15.0 22.7 33.1

    volume* [gallon] 0.04 0.04 0.05 0.08 0.13 0.16 0.26

    All dimensions in [inch]; HT = high temperature; *per Jacket

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    17/20

    Heating jacket for Promass F Technical

    4.3.2 NPT Adapter

    A

    Abb. 10: Dimensions of Adapter NPT inner thread on conical ISO external thread

    4.3.3 Flange Adapter

    Abb. 11: Dimensions of Adapter for Flanges

    A

    B

    C ISO

    SNP

    T

    NPT ISO2) S A1) B1) C1)

    G 1 1/16 AF 1.94 0.75 0.47

    G 1 5/16 AF 2.02 0.75 0.62

    1) Dimensions in [inch]2) DIN EN ISO 228

    S

    IS

    O

    D

    L

    EF

    G

    Flange according to ASME B16.5: 1.4404/316L/316

    Pressure stage ISO2) D1) E1) F1) G1) L1) S

    Cl 150G 0.62 2.38 3.5 4 0.62 3.46

    7/8 AG 0.62 2.38 3.5 4 0.62 3.46

    Cl 300G 0.62 2.62 3.75 4 0.62 3.63

    G 0.62 2.62 3.75 4 0.62 3.63

    Flange EN 1092-1 (DIN 2501) / PN 40: 1.4404/316L/316

    Pressure stage ISO2) D1) E1) F1) G1) L1) S

    PN 40G 0.68 2.56 3.74 4 0.55 3.07

    7/8 AG 0.68 2.56 3.74 4 0.55 3.07

    1) Dimensions in [inch]2) DIN EN ISO 228

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    18/20

    Technical data Heating jacket for Promass F

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    19/20

    taXIV

    Because of legal regulations and for the safety of our employees and operating equipment, we need the "Declaration of Hazardous Materialand De-Contamination", with your signature, before your order can be handled. Please make absolutely sure to attach it to the outside of thepackaging.Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, bentigen wir die unterschriebene"Erklrung zur Kontamination und Reinigung", bevor Ihr Auftrag bearbeitet werden kann. Bringen Sie diese unbedingt auen an derVerpackung an.

    Serial number

    Seriennummer ________________________Type of instrument / sensor

    Gerte-/Sensortyp ____________________________________________

    Process data/Prozessdaten Temperature _____ [F] [C]

    Conductivity / ________

    _____

    Leitfhigkeit

    /

    [S/cm]

    Temperatur Pressure _____ [psi] [ Pa ]

    Viscosity _____ [cp] [mm

    _____

    _____

    /

    /

    Druck

    Viskositt 2

    corrosivetzend

    harmlesunbedenk

    other *sonstiges*

    toxicgiftig

    Processmedium

    IdentificationCAS No.

    flammable

    entzndlich

    harmful/irritant

    gesundheits-schdlich/

    reizend

    Medium /concentrationMedium /Konzentration

    Returned partcleaned with

    Medium forprocess cleaning

    Medium and warnings

    Warnhinweise zum Medium

    * explosive; oxidising; dangerous for the environment; biological risk; radioactive* explosiv; brandfrdernd; umweltgefhrlich; biogefhrlich; radioaktiv

    Please tick should one of the above be applicable, include safety data sheet and, if necessary, special handling instructions.Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen.

    Description of failure / Fehlerbeschreibung __________________________________________________________________________

    ______________________________________________________________________________________________________________

    ______________________________________________________________________________________________________________

    We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge.We further certify that the returneparts have been carefully cleaned. To the best of our knowledge they are free of any residues in dangerous quantities.Wir besttig

    w

    besttigen, die vorliegende Erklrung nach unserem besten Wissen wahrheitsgetreu und vollstndig ausgefllt zu haben. Wir

    eiter, dass die zurckgesandten Teile sorgfltig gereinigt wurden und nach unserem besten Wissen frei von Rckstnden in gefahrbringender Menge sind.

    Company data /Angaben zum Absender

    Company / ________________________________

    _________________________________________________

    Address /

    _________________________________________________

    _________________________________________________

    Firma ___

    Adresse

    Phone number of contact person /

    ____________________________________________

    Fax / E-Mail ____________________________________________

    Your order No. / ____________________________

    Telefon-Nr. Ansprechpartner:

    Ihre Auftragsnr.

    Medium zurEndreinigung

    Medium zurProzessreinigung

    Medium imProzess

    Used as SIL device in a Safety Instrumented System / Einsatz als SIL Gert in Schutzeinrichtungen

    RA No.

    Erklrung zur Kontamination und ReinigungDeclaration of Hazardous Material and De-Contamination

    Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.

    Bitte geben Sie die von E+H mitgeteilte Rcklieferungsnummer (RA#) auf allen Lieferpapieren an und vermerken Sie die

    auch auen auf der Verpackung. Nichtbeachtung dieser Anweisung fhrt zur Ablehnung ihrer Lieferung.

  • 7/29/2019 Proline Promass F Heating Jacket Operation Instructions

    20/20

    The specifcations contained herein are subject to change without notice and any user o said specifcations should veriy rom the manuacturer that the specifcations are currentlyin eect. Otherwise, the manuacturer assumes no responsibility or the use o specifcations which may have been changed and are no longer in eect.

    Headquarters:500 North Sam Houston Parkway West, Suite 100, Houston, TX 77067 USA, Phone: +1 (281) 260-2190, Fax: +1 (281) 260-2191

    Gas Measurement Products:Erie, PA USA +1 (814) 898-5000Ellerbek, Germany +49 (4101) 304-0Thetford, England +44 (1842) 82-2900Kongsberg, Norway +47 (32) 286-700Buenos Aires, Argentina +54 (11) 4312-4736