mri physics2

80
MRI PHYSICS arun

Upload: drarunrao

Post on 26-Oct-2014

155 views

Category:

Documents


14 download

TRANSCRIPT

Page 1: Mri Physics2

MRI PHYSICS

arun

Page 2: Mri Physics2

During T1 (spin-lattice) relaxation, the longi- tudinal magnetization recovers as the spinning nuclei release energy into the environment .

During T2 (spin-spin) relaxation, the trans- verse magnetization is dephased because of interaction between the spinning nuclei and their magnetic fields .

Page 3: Mri Physics2

T1 recovery (spin-lattice relaxation) involves recovery of the longitudinal magnetization (yellow) because of the release of energy (green) into the environment. The lattice is indicated in tan

Page 4: Mri Physics2

T2 decay (spin-spin relaxation) is decay of the transverse magnetization because of the interaction of the individual magnetic fields of spinning nucle .Note that all nuclei initially spin in phase (as indicated by the similar position of the red bands at the bottom of each circle), then move out of phase (with red bands in different positions)

Page 5: Mri Physics2

T2 decay is decay of the transverse magne tization because of magnetic field inhomogeneities

Page 6: Mri Physics2

T1 RELAXATION

Tissue composition

Structure

surroundings

Inhomogeneity of magnetic field

T2 RELAXATION

Page 7: Mri Physics2

TISSUE CONTRAST

Depend on

• differences in T1, T2, and proton density (ie, the number of 1H nuclei)

• repetition time (TR) and echo time (TE)

Page 8: Mri Physics2
Page 9: Mri Physics2
Page 10: Mri Physics2

TR is the time (usually measured in millisec- onds) between the application of an RF excitation pulse and the start of the next RF pulse.

TE (also usually measured in milliseconds) is the time between the application of the RF pulse and the peak of the echo detected

Page 11: Mri Physics2

Both parameters affect contrast on MR images because they provide varying levels of sensitivity to differences in relaxation time between various tissues.

At short TRs, the difference in relaxation time between fat and water can be detected (longitudi- nal magnetization recovers more quickly in fat than in water)

at long TRs, it cannot be detected.

Page 12: Mri Physics2

TR relates to T1 and affects contrast-T1W

At short TEs, differences in the T2 signal decay in fat and water cannot be detected; At long TEs, they can be detected.

Therefore, TE relates to T2 and affects contrast on T2-weighted images

Tissues with more protons have higher signal intensity, and those with fewer protons have lower signal intensity.

Page 13: Mri Physics2

Short TE Long TE

Short TR

Long TR

T1-weighted

T2-weightedProton density

T1 and T2 weighted

Page 14: Mri Physics2
Page 15: Mri Physics2
Page 16: Mri Physics2
Page 17: Mri Physics2

T1 lenghtening means dark ( Hypointense)

T1 shortening means bright ( Hyperintense)

Page 18: Mri Physics2

T2 lenghtening means bright ( Hyperintense )

T2 shortening means dark ( Hypointense )

Page 19: Mri Physics2
Page 20: Mri Physics2

MR Signal Localization

Gradients are linear variations of the magnetic field strength in a selected region .

Along with the magnetic field strength, the precessional frequency of 1H nuclei is also changed in the specific region

Three types of gradients are applied, according to the axis of imaging (x-, y-, or z-axis)

Page 21: Mri Physics2
Page 22: Mri Physics2

The phase-encoding gradient causes a phase shift in the spinning protons so that the MR imaging system computer can detect and encode the phase of the spin.

The frequency-encoding gradient also causes a shift— one of frequency rather than phase—that helps the MR system to detect the location of the spinning nuclei.Because this shift of frequency usually occurs when the echo is read, it is also called the readout gradient

Page 23: Mri Physics2

k-Space and the Image Matrix

k-Space (k, the symbol for wave number) -matrix of voxels within which the raw imaging data are stored

The horizontal axis (x-axis) of the matrix usually corresponds to the frequency,

the vertical axis (y-axis) usually corresponds to the phase

easily interchanged

The center of k-space contains information about gross form and tissue contrast, whereas the edges (periphery) of k-space contain information about spatial resolution (details and fine structures).

The raw imaging data in k-space must be Fourier transformed to obtain the final image.

Page 24: Mri Physics2
Page 25: Mri Physics2

EPI imaging and k-space

Source: Traveler’s Guide to K-space (C.A. Mistretta)

x = frequency and y = phase or angle

Page 26: Mri Physics2

Fourier transformation.

k-space is magically transformed into our image via a Fourier transformation.

Source: Buxton book Ch 5

Page 27: Mri Physics2

MR Sequences

There are only two fundamental types of MR pulse sequences: SE and GRE.

All other MR sequences are variations of these, with different parameters added on

MR pulse sequences can be either two-dimensional (2D), with one section acquired at a time, or three-dimensional (3D), with a volume of multiple sections obtained in a single acquisition.

Page 28: Mri Physics2

SE SequencesIn SE sequences, a 90° pulse flips the net magnetization vector into the transverse plane

As the spinning nuclei go through T1, T2, and T2* relaxation, the transverse magnetization is gradually dephased.

A 180° pulse is applied at a time equal to one-half of TE to rephase the spinning nuclei. When the nuclei are again spinning in phase (at total TE), an echo is produced and read

Page 29: Mri Physics2

After the RF pulse is switched off, the protons dephase (a-c). The 180° pulse causes them to precess in the opposite direction and so they rephase again (d-f)

Page 30: Mri Physics2
Page 31: Mri Physics2
Page 32: Mri Physics2

Conventional SE

Fast SE sequences

Page 33: Mri Physics2
Page 34: Mri Physics2

34

CSE is used in 2 ways:

Single spin echo-

90 ˚ excitation pulse - single 180˚ RF.

Produces T1 WT images

Dual echo sequence-

Two 180 ˚ pulses are applied to produce two spin echoes.

Provides two images per slice:

proton density WT & T2 WT

Page 35: Mri Physics2

35

Uses:

Spin echo sequences are the gold standard in that the contrast they produce is understandable & is predictable & can be used in any part of the body for any indication.

ADVANTAGES

Good image quality

Versatile

True T2 wt images

Available on all systems

Gold standard for image contrast &

weighting.

DISADVANTAGE:

Long scan time

Page 36: Mri Physics2

36

FAST/ turbo spin echo :

Mechanism : A train of 180˚rephasing pulses are

employed, each producing a spin echo- called an echo train.

ETL( echo train length)/turbo factor :number of 180˚ pulse & their resultant echoes.

After each rephasing, a phase encoding step is performed & the data from the resultant echo is stored in the K space.

As several lines of K space are filled every TR (instead of one line as in conventional spin echo ), K space is filled more rapidly decreased scan time.

2,4, 8,or 16 180˚ RF pulse applied in every TR, reducing the scan time by ½, ¼, 1/8, 1/16 time resp.

Use : brain ,spine

joints

extremity

pelvis

Page 37: Mri Physics2

Fast SE VariantsIn a fast or turbo SE sequence, a single 90° pulse is applied to flip the net magnetization vector, after which multiple 180° rephasing pulses are applied each of which creates a Hahn echo .

All the echoes together are called an echo train, and the total number of 180° RF pulses and echoes is referred to as the echo train length.

Page 38: Mri Physics2
Page 39: Mri Physics2

Axial T2-weighted fast SE image (b) and conventional SE image (c) provide comparable depiction of a brain tumor. The acquisition time for conventional SE imaging was 7 minutes 17 seconds, whereas that for fast SE imaging with an echo train length of 16 was 34 seconds.

Page 40: Mri Physics2

Conventional Inversion Recovery

180° preparatory pulse is applied to flip the net magnetization vector 180° and null the signal from a particular entity

When the RF pulse ceases, the spinning nuclei begin to relax. When the net magnetization vector for water passes the transverse plane (the null point for that tissue), the conventional 90° pulse is applied, and the SE sequence then continues as before .

The interval between the 180° pulse and the 90° pulse is the TI. [inversion time]

Page 41: Mri Physics2

Two important clinical implementations of the inversion- recovery concept are

1. short tau inversion-recovery (STIR) sequence

2. the fluid-attenuated inversion-recovery (FLAIR) sequence

Page 42: Mri Physics2
Page 43: Mri Physics2

Coronal STIR image shows an insufficiency fracture of the distal tibia, with an extensive area of high signal intensity in bone marrow near the site of the fracture

Page 44: Mri Physics2

Nulling of the water signal is seen at TI, when there is virtually no net magnetization vector (NMV) in water. When the 90° pulse flips the net magnetization vector into the transverse plane, little or no transverse magnetization (Tm) is present, and, therefore, no signal is detected in water.

Page 45: Mri Physics2
Page 46: Mri Physics2

STIRIn STIR sequences, an inversion-recovery pulse is used to null the signal from fat .

When the net magnetization vector of fat passes its null point (at approximately 140 msec), the conventional 90° RF pulse is applied.

Little or no longitudinal magnetization is present in fat, and the transverse magnetization in fat is insignificant.

Page 47: Mri Physics2

STIR sequences provide excellent depiction of bone marrow edema , which may be the only indication of an occult fracture.

Unlike conventional fat-saturation sequences , STIR sequences are not affected by magnetic field inhomogeneities, so they are more efficient for nulling the signal from fat

Page 48: Mri Physics2

FLAIRIn FLAIR sequences, an inversion- recovery pulse is used to null the signal from cerebrospinal fluid

useful for detecting lesions that otherwise are not easily distinguishable or for delineating hyperintense lesions that border fluid-containing spaces such as sulci or ventricles in the brain

Page 49: Mri Physics2
Page 50: Mri Physics2
Page 51: Mri Physics2

GRE SequencesIn a GRE sequence, an RF pulse is applied that partly flips the net magnetization vector into the transverse plane (variable flip angle)

Gradients, as opposed to RF pulses, are used to dephase (negative gradient) and rephase (positive gradients) transverse magnetization

Page 52: Mri Physics2

Because gradients do not refocus field inhomogeneities, GRE sequences with long TEs are T2* weighted (because of magnetic susceptibility) rather than T2 weighted like SE sequences.

Page 53: Mri Physics2
Page 54: Mri Physics2

Diffusion Weighted Imaging

Characterizes difference in the brownian motion of water molecules , depending on the environment

Biologic system impede the free motion of water

So cellullar structure ,permeability barriers and various macromolecules within the brain restrict the free diffusion

Page 55: Mri Physics2

DWI is an Echo Planar Imaging Sequence

EPI – 2 separate and equal magnetic gradients are applied opposite side of radiofrequency pulse during image acquisition

Water molecule that do not travel significantly between excitation pulse and read pulse will not dephase and retain much of its signal.

So increased signal in DWI suggest loss of normal extracellullar space,presence of highly viscous , proteinaceous material, oe cellullar environment

Page 56: Mri Physics2

Diffusion-weighted imaging sequence diagram Axial diffusion-weighted brain image shows areas of restricted diffusion with high signal intensity.

Page 57: Mri Physics2
Page 58: Mri Physics2
Page 59: Mri Physics2

DW sequences are usually applied in conjunction with apparent diffusion coefficient (ADC) mapping techniques .

For the calculation of ADC maps, two sets of images are required: one set obtained with- out application of a diffusion gradient (which have an appearance similar to that of T2-weightedimages), and one obtained with a diffusion gradi- ent. The ADC calculation is based on the negative logarithm of the ratio of those two image sets

T2 shine-through.

Page 60: Mri Physics2

Normal Brain ADC value – 740 to 840 x 1/10 mm2/sec

ADC value less than 500 x 10 mm2/sec – suggest irreversible infarction

Page 61: Mri Physics2

Perfusion weighted Imaging

The Hemodynamic values generated by the PWI are

cerebral blood flow

cerebral blood volume

mean transit time

Page 62: Mri Physics2

Brain volume is scanned repeatedly before , during and after the bolus

Perfusion and time intensity curve is generated in each voxel

Area under curve – CBV

Maximum slope of the curve – CBF

Width of Susceptibiliy curve – MTT

CBF = CBV / MTT

Page 63: Mri Physics2

Functional MRI

Shortly after presentation of stimuli there is increase in cortical activity ,increase in local blood flow

The oxygen delievered by increase in blood flow is greater then the demand of tissuue for oxygen

So there is net increase in oxyhaemaoglobinwith relative decreased deoxyhaemoglobin

This reduction in deoxy hb cause net increase in local signal --- So callled as BOLD technique

Page 64: Mri Physics2

Magnetic Resonance Spectroscopy

Based on chemical shift

MRS data are plotted as a linear graph with chemical shift ( in ppm) on the X axis and relative signal amplitude on Y axis

2 basic methods are used – single voxel and multivoxel

Page 65: Mri Physics2

• N- Acetyl Aspartate( NAA) - Highest peak in normal spectra

Marker of neuronal integrity

Decreased in malignancy / tumour death

• Choline –

Marker of membrane density

Increased in malignancy/inflammatory process

Page 66: Mri Physics2

• Creatine –

Marker of energy metabolism

Creatine values are relatively stable

• Glutamate / Glutamine

• Lactate –

Marker of anaerobic glycolysis

seen in brain ischemia and seizures

Page 67: Mri Physics2

Myoinositol –Marker of degree of glial proliferation

Alanine Role in citric acid cycle , increased in meningioma

Lipid Seen in areas of increased cell turn over

Page 68: Mri Physics2

Quick Assessment

3 most prominent metabolitesNAA , creatine , choline

Imaginery line which connects these three peaks from left to right ----- produces an angle of 45 to 50 degree ( Hunter angle)

Reduction in angle seen in neuronal destruction

Reversal of this angle s/o tumour , ischemic process, inflammatory condition

Page 69: Mri Physics2

Tumours – Reduction in NAA( further decrease as grade increases)

Elvation of choline

Choline is said to be abnormally high when its peak is twice the height of NAA and creatine

Areas of necrosis in Tumours – show lactate peak

Hippocampal sclerosis –Reduction in NAA

Elevated Glx and myoinositol

Alzeihmers Disease Reduction in NAA

Elevated Myoinositol

Page 70: Mri Physics2

The techniques used for MRS are

Point resolved Spectroscopy

Stimulated ECHO Acquistion Mode

Page 71: Mri Physics2
Page 72: Mri Physics2
Page 73: Mri Physics2

MR Angiography

The most common MR angiographic techniques are

time-of-flight imaging, [TOF]multiple overlapping thin-slab acquisitions (MOTSA), phase-contrast imaging, contrast-enhanced MR angiography.

Page 74: Mri Physics2

Time-of-Flight Imaging and MOTSA

Multiple RF pulses applied with short TRs saturate the spins in stationary tissues .

This results in suppression of the signal from stationary tissues in the imaging slab

In- flowing blood is unaffected by the repetitive RF pulses; therefore, as it enters the imaging slab, its signal is not suppressed and appears hyperintense compared with that of stationary tissue.

Time-of- flight imaging -2D, with section-by-section acquisition or 3D, with acquisition of a larger volume.

MOTSA is the hybrid result of 2D and 3D time-of-flight imaging.

Page 75: Mri Physics2

Phase-Contrast Imaging.Phase (or direction) of flow and the velocity (or magnitude) of flow .

It requires two measurements that are sensitized to flow in different directions; typically, one measurement is sensitized to flow in one direction, and the second is sensitized equally to flow in the opposite direction.

Both measurements are subtracted to eliminate any contribution to image phase that does not arise from flow or motion As a result, no signal is detected from stationary tissue; signal is received only from flowing blood

2D or 3D.

Page 76: Mri Physics2

One convention is that high signal is seen from flow that moves from a superior location to an inferior one, from right to left, and from an anterior to a posterior location. No signal (a signal void) is seen from flow that moves from an inferior to a superior location, from left to right, or from a posterior to anterior location

Page 77: Mri Physics2

TOF vs PCA

Page 78: Mri Physics2

Contrast-enhanced MR Angiography

an intravenous contrast material (usually gadopentetate dimeglumine ) is used .

Such agents typically shorten the T1 of blood (ie, hasten the recovery of longitudinal magnetization),

so a higher-magnitude net magnetization vector is available to be flipped with the next TR,-- high signal intensity on T1-weighted images

2D or 3D methods

Page 79: Mri Physics2

Maximum intensity projection from 2D time-of-flight imaging of the cerebral veins. (b) Axial image obtained with MOTSA. (c) Maximum intensity projection from MOTSA data

Page 80: Mri Physics2

. (d) Phase- contrast image shows left subclavian steal syndrome. In this example, caudal-to-cranial flow appears dark, and cranial-to-caudal flow appears bright. The left vertebral artery appears bright (arrow), a feature that indicates subclavian steal. (e) Maximum intensity projection from contrast-enhanced MR angiography at the level of the heart, aorta, and vessels of the aortic arch.