  • MEGOHMMETER 50505060

    E N G L I S H User Manual

  • Statement of Compliance

    Chauvin Arnoux, Inc. d.b.a. AEMC Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

    We guarantee that at the time of shipping your instrument has met its published specifications.

    An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

    The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.

    Serial #: ________________________________

    Catalog #: 2130.20 / 2130.21

    Model #: 5050 / 5060

    Please fill in the appropriate date as indicated:

    Date Received: _________________________________

    Date Calibration Due: _______________________

    Chauvin Arnoux, Inc.d.b.a AEMC Instrumentswww.aemc.com

  • Megohmmeter Models 5050/5060 1

    Table of Contents

    INTRODUCTION ..................................................................................... 31.1 International Electrical Symbols ...........................................................31.2 DefinitionofMeasurementCategories .................................................41.3 ReceivingYourShipment .....................................................................41.4 OrderingInformation ............................................................................4

    1.4.1 AccessoriesandReplacementParts ......................................41.5 AccessoryInformation..........................................................................5

    1.5.1 DataViewSoftware(forModel5060) ....................................5

    PRODUCT FEATURES ............................................................................. 62.1 MegohmmeterFeatures .......................................................................82.2 DigitalFeatures ....................................................................................92.3 Bargraph ............................................................................................102.4 Symbols .............................................................................................102.5 Push-buttonFunctions .......................................................................11

    2.5.1 2ndButton .............................................................................11

    2.5.2 V-TIME/ Button ..............................................................112.5.3 R-DAR-PI-DD/R(t)Button ...................................................12

    2.5.4 /ALARMButton ..............................................................162.5.5 /SmoothButton ................................................................17

    2.5.6 Button ...............................................................................172.5.7 SET-UPFunction ..................................................................182.5.8 ClearingtheMemory.............................................................202.5.9 CommunicationRate(RS-232) .............................................202.5.10 DefaultDeviceConfiguration ................................................202.5.11 VoltageDisturbanceLimit .....................................................212.5.12 AutomaticMeasuringRange.................................................212.5.13 Blocking(Disabling)TestVoltages ........................................22

    SPECIFICATIONS ................................................................................. 233.1 ReferenceConditions.........................................................................233.2 Voltage ...............................................................................................233.3 InsulationResistance .........................................................................233.4 PowerSupply .....................................................................................283.5 EnvironmentalSpecifications .............................................................29

  • 2 Megohmmeter Models 5050/5060

    3.6 MechanicalSpecifications ..................................................................293.7 SafetySpecifications ..........................................................................293.8 VariationsinOperatingRange ...........................................................30

    MEASUREMENT FUNCTIONS ............................................................... 314.1 AC/DCVoltage ...................................................................................314.2 InsulationMeasurement .....................................................................31

    OPERATION ......................................................................................... 335.1 MeasurementProcedure....................................................................335.2 InsulationMeasurement .....................................................................335.3 OperationExamples...........................................................................345.4 CapacitanceMeasurement ................................................................355.5 ResidualandLeakageCurrentMeasurement ...................................35

    MEMORY / RS-232 (5060) .................................................................. 366.1 RS-232Specifications ........................................................................366.2 Saving/RecallingValues(MEM/MRButton) .....................................366.3 PrintingMeasuredValues ..................................................................386.4 InstantaneousPrintingofMeasurements(PRINTbutton) ................396.5 PrintingDatainMemory(PRINTMEMbutton) ..................................40

    DATAVIEW SOFTWARE ....................................................................... 427.1 InstallingDataView ...........................................................................427.2 ConnectingtheModel5060toyourComputer ..................................467.3 UsingDataView ................................................................................46

    7.3.1 ConfiguringtheInstrument....................................................477.3.2 RunningtheTest ...................................................................49

    7.4 DataViewMegohmmeterTemplates .................................................527.4.1 MegohmmeterSummaryReportTemplate ...........................52

    MAINTENANCE ................................................................................... 548.1 RechargingtheBattery ......................................................................548.2 FuseReplacement .............................................................................558.3 Cleaning .............................................................................................558.4 Storage...............................................................................................55

    RepairandCalibration ........................................................................................56TechnicalandSalesAssistance .........................................................................56LimitedWarranty ................................................................................................57WarrantyRepairs ................................................................................................57

  • Megohmmeter Models 5050/5060 3

    CHAPTER 1

    INTRODUCTION

    WARNING Thesesafetywarningsareprovidedtoensurethesafetyofpersonnelandproperoperationoftheinstrument.

    Donotattempttoperformanytestswiththeseinstrumentsuntilyouhavereadtheinstructionmanual.

    Safetyistheresponsibilityoftheoperator!

    Testsaretobecarriedoutonlyonnon-energizedcircuits!Checkforlivecircuitsbeforemakingresistancemeasurements(safetycheck).

    Highvoltageispresent,asisthesampleconnectedtoit.Anyoneperformingorassistingintestingmustfollowallsafetyprecautionstopreventelectricalshocktothemselvesandtoothers.

    AEMCconsiderstheuseofrubberglovestobeanexcellentsafetypractice,eveniftheequipmentisproperlyoperatedandcorrectlygrounded.

    Whentestingsampleswithacapacitivecomponent,makesuretheyhavebeenproperlydischargedandaresafetotouch.Dielectricinsulationsamplesshouldbeshort-circuitedforatleastfivetimestheamountoftimetheywereenergized.

    Megohmmetersshouldneverbeusedinanexplosiveenvironment.

    Onlyusetheleadsthataresuppliedwiththemegohmmeter.Iftheyaredefectiveorworn,replacebeforetesting.

    Thisinstrumentcanbeusedoninstallationsratedfor1000V,CategoryIII,or2500V,CategoryI.

    1.1 International Electrical Symbols

    This symbol signifies that the instrument is protected by double or reinforced insulation.

    This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.

    Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.

    In conformity with WEEE 2002/96/EC

  • 4 Megohmmeter Models 5050/5060

    1.2 Definition of Measurement CategoriesCAT I: For measurements on circuits not directly connected to the AC supply

    wall outlet such as protected secondaries, signal level, and limited energy circuits.

    CAT II: For measurements performed on circuits directly connected to the electrical distribution system. Examples are measurements on household appliances or portable tools.

    CAT III: For measurements performed in the building installation at the distribution level such as on hardwired equipment in fixed installation and circuit breakers.

    CAT IV: For measurements performed at the primary electrical supply (

  • Megohmmeter Models 5050/5060 5

    Lead Replacement 1 ft jumper lead............................................... Cat. #2119.78

    Fuse, Set of 5, 0.1A, 380V, 5 x 20, .10kA ......................................... Cat. #2119.84

    Replacement Lead Set of 3, Color-coded 10ft (5kV) Safety with Integral Hippo Clips (JUMPER LEAD INCLUDED) ............................. Cat. #2119.85

    Lead Set of 3, Color-coded 25ft (5kV) Safety with Integral Hippo Clips (JUMPER LEAD NOT INCLUDED) ............................. Cat. #2119.86

    Lead Set of 3, Color-coded 45ft (5kV) Safety with Integral Hippo Clips (JUMPER LEAD NOT INCLUDED) ............................. Cat. #2119.87

    Extra Large Classic Tool Bag ...........................................................Cat. #2133.73

    Inverter 12VDC to 120VAC 200 Watt for Vehicle use ......................Cat. #2135.43

    Battery Rechargeable 9.6V ...........................................................Cat. #2960.21

    US 115V Power Cord .......................................................................Cat. #5000.14

    Adapter RS-232 to USB 2.0 (5060) ...............................................Cat. #5000.60

    1.5 Accessory Information

    1.5.1 DataView Software (for Model 5060)

    Dataview software makes it possible to: Retrieve data from memory and plot graphs of the changes in insulation

    as a function of the time over which the test voltage is applied, R(t). Print out protocols of personalized tests, depending on the users needs.

    Create text files for use on spreadsheets.

    Set up and control the instrument entirely by the RS-232.

    Order Accessories and Replacement Parts Directly OnlineCheck our Storefront at www.aemc.com/store for availability

  • 6 Megohmmeter Models 5050/5060

    CHAPTER 2

    PRODUCT FEATURES

    The Megohmmeter Models 5050 and 5060 are portable instruments, housed in rugged casing.These megohmmeters are designed to check the safety of electrical installations and equipment.

    Measures: Voltage

    Insulation Resistance

    Capacitance

    Leakage Current

    Advantages: Automatic voltage detection in all functions, before or during the mea-

    surement, which inhibits or stops the measurement. Automatic detection of external AC or DC voltages on the terminals

    Threshold programming in each function, triggering audible alarms

    Timed control of measurement duration

    Fuse protection, with defective fuse detection and indication

    Digital filtering of insulation measurements

    Automatic discharge of residual high voltage on the tested device

    Automatic shut-off to save the batteries

    Battery level indicator

    Large easy-to-read back-lit LCD

    Additional Features (Model 5060): 128kB memory, real-time clock and serial interface

    Control and programming of the instrument from a PC (with DataView software) Printing with a serial printer

  • Megohmmeter Models 5050/5060 7

    MODEL 5050

    ( )

    FF 0.1A380V - 10kA

    5x20mm

    110 - 230V50/60 Hz

    20 VA max

    CALIBRATION

    2nd

    SMOOTH ALARM

    R(t) MEGOHMMETERMODEL 5050

    R-DAR-PI-DD

    V/TIME

    START/STOP1000V CAT III

    2500V

    G OFF

    M -1000V

    M - 2500V

    M - 5000V

    M 50-5000V

    SET-UP

    M - 500V1

    6

    7

    5

    32 4

    MODEL 5060

    ( )

    FF 0.1A380V - 10kA

    5x20mm

    110 - 230V50/60 Hz

    20 VA max

    RS-232

    2nd

    SMOOTH ALARM

    R(t)PRINT MEMMR MEGOHMMETERMODEL 5060

    R-DAR-PI-DDPRINTMEM

    V/TIME

    START/STOP1000V CAT III

    2500V

    G OFF

    M -1000V

    M - 2500V

    M - 5000V

    M 50-5000V

    SET-UP

    M - 500V1

    6

    7

    5

    32 4

  • 8 Megohmmeter Models 5050/5060

    2.1 Megohmmeter Features1. Safetyterminals:+,Gand-andaccesstotheprotectivefuse.

    2. ACpowerplug(directoperationonACandbatteryrecharge).

    3. Back-litliquidcrystaldisplay.

    4. RS-232serialinterfacemaleplug(9-pin)

    Forcalibrationonly(Model5050)

    ForconnectiontoaPCorprinter(Model5060)

    5. Rotaryselectorswitchwith7positions:

    OFF-Instrumentisoff M - 500V-Insulationmeasurement(to2T) M - 1000V-Insulationmeasurement(to4T) M - 2500V-Insulationmeasurement(to10T) M - 5000V-Insulationmeasurement(to10T) M - 50V-5000V-Insulationmeasurementwithselectabletestvoltage SET-UP-Set-upofthemegohmmeter

    6. START/STOPbutton

    7. 6buttons(5050)or8buttons(5060),eachwithafirstandsecondfunction.Thesecondfunctionsarehighlightedinyellowbeloweachbutton:

    2nd (yellow button)-Selectsthesecondfunctiononeachbutton. R-DAR-PI-DD - Lets you choose the type ofmeasurement to be per-

    formed. Instantaneous Resistance (R), Dielectric Absorption (DAR),PolarizationIndex(PI),orDielectricDischarge(DD).

    R(t)-Shows/hidestheintervalvaluesofinsulationresistance,testvolt-ageandtime-stamping,followingatime-controlledtest.

    V/TIME -Whenmeasuring insulation,pressing thisbuttondisplays thetimeelapsed since the beginning of themeasurement, then the exactvoltagegenerated.Whenperformingresistanceorcapacitancemeasure-ment,thisbuttonhasnoeffect.InMR(memoryrecall),itdisplaysthedateandtimeatwhichthemeasurementwasstored, theexact testvoltageandtheOBJ:TESTnumber.

    -Activatesordeactivatesthetime-controlledtestmode.

    -TurnsthedisplaybacklightONorOFF.

    ALARM-ActivatesordeactivatesthealarmsprogrammedinSET-UP.

    -Selectsaparametertobemodified. SMOOTH-Stops/startssmoothingofdisplayedvaluesduringinsulation

    testing.

  • Megohmmeter Models 5050/5060 9

    -Increasestheflashingparameterbeingdisplayed.Tomoveaboutthelistofintervalinsulationmeasurements,intheR(t)function.

    -Decreasestheflashingparameterbeingdisplayed.TomoveaboutthelistofintervalinsulationmeasurementsintheR(t)function.

    Ifthe and buttonsarehelddown,themovementbetweenparam-etersisincreasedtoafasterrate.

    Model 5060 only: MEM-Savesmeasuredvalues MR-Recallssaveddata PRINT-Printsmeasurementresults PRINT MEM-Printsmemorycontents

    2.2 Digital Features

    8.8.8.80.1M

    1M10M 100M 1G 10G 100G

    1T

    A nA F

    COM

    P

    GPIDARDDALARM

    T M k

    MEM MR

    SMOOTH REMOTE

    OBJ.

    min sec

    TEST

    V DCAC

    2nd 8.8.8.8Main DisplayIndicatesthevalueof:

    Insulationmeasurement(resistance,DARandPI,DD,capacitance).

    Small DisplayIndicates:

    Voltagemeasuredorappliedbytheinstrument.

    Elapsedtimeortheoutputvoltage,duringinsulationmeasurement.

    Afterrecordingasetofdata(5060),thesmalldisplayalsoindicatesthetimeanddateinMR(memoryrecall)mode,andthememoryaddresswiththeOBJ:TESTnumber.

  • 10 Megohmmeter Models 5050/5060

    2.3 Bargraph Activeduringinsulationmeasurement(0.1M to 1T). Indicatesthebatterychargeatstart-up.

    Indicatesfreememoryspace-onesegmentrepresentingapproximately100groupsofsavedvalues,witheachgroupholdingapproximately50recordings(1OBJand1TEST).

    2.4 Symbols

    MEM/MR:Memoryaddress(5060)-indicatesthememoryoperation.

    OBJ:TEST: Memoryaddress(5060)-thenumberisdisplayedonthesmalldigitaldisplay.

    COM:Flashesonthescreenwhendataistransmittedtotheserialinterface(5060)orremainspermanentlydisplayedifthereisaproblemduringtransfer.

    DAR/PI/DD: Indicates themode chosen before insulationmeasurement or theresultsofthesemeasurements.

    Dangerousvoltagegenerated;V>120V.

    Externalvoltagepresent,symbolisactivatedafterpressingSTART,ifV>25VAC3Vor>35VDC.

    Activationoftime-controlledtestmodeorclockadjustmentwhen inSET-UP(5060).

    (Yellowbutton)Selectsthesecond(bottom)functionofabutton.

    Flashesifthebatteryvoltageislow,indicatingthatthebatteryshouldberecharged(see8.1).Thevoltageisdisplayedonthesmalldisplayfor 2 seconds when the instrument is turned on. The main displayindicatesbatwhenthetestiscarriedon.

    Warningbuzzerisactivated.

    Indicatesthattheautomaticswitch-offfunctionhasbeendeactivated.

    SMOOTH:Smoothstheinsulationmeasurementsdisplayed.

    REMOTE: Remotecontrolviaan interface (5060). In thismode,all thebuttonsandtherotaryswitchontheinstrumentarenon-functional,exceptfortheOFFposition.

    FUSE -G-: FlashesiftheGinputfuseisdefective.

  • Megohmmeter Models 5050/5060 11

    2.5 Push-button Functions

    2.5.1 2nd Button Selectsthesecondfunction(highlightedinyellow)onthebuttons.

    The symbol appears. This symbol disappears upon pressing thefunctionbuttonchosen,exceptifthebuttonisactivated.Inthiscase,itonlydisappearswhenthe buttonispressedagain,orifotherfunctionbuttonsarepressed.Thisallowsyoutorapidlydecreaseparameterswiththebutton,withouthavingtopressthe buttoneverytime.

    2.5.2 V-TIME / Button

    First Function - V/TIMEDisplaysallthesecondaryinformationavailableonthesmalldisplay.

    For Insulation Measurement: Thetimeelapsedsincethestartofthemeasurement.

    Thevoltagebetweentheinstruments+andterminals.

    Date,time,testvoltageandOBJ:TESTnumberinmemoryrecallmode(MR).

    Second Function - Time-controlled TestDisplaysthemeasurementdurationprogrammedinSET-UPonthesmalldisplay.

    The symbollightsup. PressingtheSTARTbuttonbeginsthemeasurementprocess.Themea-

    surementdurationissetto15minutesbydefault.

    Assoonasthemeasurementstarts,thesmalldisplaycountsdowntheremainingdurationofthemeasurement.Onceitreacheszero,themea-surementstops.

    As the time-controlled test is carriedout, interval samples (resistance/voltagevaluesasafunctionoftime)areautomaticallysaved.Thetimebetweeneachsampleissetto30sbydefault,butitcanbechangedintheSET-UPmenu.

    ThesamplesaredisplayedwiththeR(t)function(see2.5.3)aslongasanewmeasurementhasnotbeenstarted.Witheachnewmeasurement,the previous sample value is erased from thememory.These samplevaluesaresavedwith the last valueof the resistance if usedwith theMEM (memorystorage)function.

    NOTE: If the selector switchs position is altered, or if the STOP button is pressed during measurement, the measurement is interrupted.

    This function is only active for insulation measurement.

  • 12 Megohmmeter Models 5050/5060

    2.5.3 R-DAR-PI-DD / R(t) Button

    First Function - R-DAR-PI-DD

    TheR-DAR-PI-DDbuttonallowsautomaticmeasurementof: DielectricAbsorptionRatio(DAR)

    PolarizationIndex(PI)

    DielectricDischargeTest(DD)

    ThePIandDARvaluesareusefulonlongcables,orwhenmonitoringtheagingofinsulationonelectricalmachines.

    Onthismaterial,measurementisadverselyaffectedatthebeginning,duetointer-feringcurrents (fromcapacitivecharges,dielectricabsorption)whicheventuallycanceleachotherout.Therefore,inordertogetanaccuratemeasurementoftheleakagecurrent,itisnecessarytocarryoutthemeasurementoveralongperiodoftime,toreducetheinfluenceofanyinterferingcurrentsatthebeginningofthemeasurement.

    ThenthePIortheDAR ratiosarecalculated:PI*=R10min/R1 min(2valuesrecordedduringa10minutemeasurement)

    DAR=R1 min/R30s(2valuesrecordedduringa1minutemeasurement)

    Thequalityofinsulationdependsontheresultsfound.

    DAR PI State of the insulation

    < 1.25< 1

    Inadequate, even dangerous< 2

    < 1.6 > 2 to < 4 Good

    > 1.6 > 4 Excellent

    *The 10 and 1 minute times can be modified in the SET-UP menu, to adapt to any changes or specific applications.

    Formulti-layerinsulation,ifoneofthelayersisdefective,butalltheothersshowstrongresistance,thecalculationofthePIandPARratiosisnotsufficienttoshowthistypeofproblem.

    Therefore,itisnecessarytosupplementthePIandDARindicationsbyadielectricdischargetest(DD).

    Thistestmeasuresthedielectricabsorptionofheterogeneousormulti-layerinsu-lationwhileignoringtheparallel-surfaceleakcurrents.

    Itconsistsofapplyingatestvoltageforaperiodsufficienttoelectricallychargetheinsulationtobemeasured(atypicalvalueistheapplicationofa500Vvoltagefor30minutes).

  • Megohmmeter Models 5050/5060 13

    ThetestvoltageischoseninthesamewayastheinsulationmeasurementintheSET-UPmenu.Thedevicethencausesafastdischargeduringwhichtheinsula-tioncapacity ismeasured, thenmeasured1minuteafter theresidualcurrent iscirculatingintheinsulation.

    TheDDvalueiscalculatedfromtheformulabelow:

    DD=

    currentmeasuredafter1minute(mA)

    testvoltage(V)xcapacitymeasured(F)

    The indicationof the insulationqualityasa functionof the resultingvalue isasfollows:

    DD value Insulation quality

    DD >7 Very bad

    7 >DD >4 Bad

    4 >DD >2 Doubtful

    DD 1min)

    PI(formeasurementslasting>theprogrammedPItimes)

    DDcanonlybecalculated1minuteaftertheendoftheinsulationmea-surementandcircuitdischarge,andonlyifitwasselectedbeforebegin-ningthemeasurement.

    CapacitanceinF(onlyafterthemeasurementhasstoppedandthecir-cuithasbeendischarged)

    ResidualleakagecurrentcirculatingintheinstallationinAornA

    InsulationresistanceinM,G or T

    NOTE: During the measurement, the DAR value is not available if the DD value was preselected before the measurement. The PI value is not available if DAR or DD values were preselected before the measurement.

  • 14 Megohmmeter Models 5050/5060

    Automatic Measurement of DAR or PI:

    IftheR-DAR-PI-DDbuttonispressedduringthevoltagemeasurementbeforethestartofameasurement,thefollowingisdisplayed:

    M DAR PI DD

    NOTE: The input current may be a depolarization current resulting from an earlier insulation measurement. It is recommended to start a new DAR and PI measure-ment after the current has dropped to a negligible value (on the order of 100pA) to avoid variations on these measurements.

    Dependingonthechoices(DARorPI),themeasurementgoesasfollows:

    a) DAR: PressSTART-theDARsymbolflashesandthedisplayindicates----aslongasthecalculationoftheratioisnotpossible(t

  • Megohmmeter Models 5050/5060 15

    Therefore: ifthemeasurementlasts1min.-DAR ifthemeasurementlastslongerthantheprogrammedPItime-PI if1min.aftertheendofthemeasurement-DD

    NOTE: If during DAR or PI measurement (automatic or not), a high external inter-ference voltage becomes present, or the insulation resistance goes beyond the devices measurement range, then the DAR or PI measurements are interrupted and the screen indicates:

    DAR orM PI M or DD M

    ThedisplayoftheDDvalueis:

    Unknown(----)ifC (dISt + 1.05) Vn

    Note: AdjustmentofthedIStfactoroptimizesthemeasurementbuild-uptime.Ifthere isno interferencevoltagepresent, thedISt factormaybeadjusted to theminimumvaluetoobtainabuild-uptimefortheminimummeasurement.

    If a significant interference voltage is present, the dISt factor may be increased so that the measurement is not interrupted by the appearance of a negative alternation during the generation of the test voltage, which is the same as optimizing the measurement build-up time in the presence of the interference voltage.

    Ifthemeasurementsareunstable,theSMOOTHfunctioncanbeenabled.

    BypressingtheV/TIMEbuttonduringmeasurement,youcanalternatebetweendisplayingthedurationofthevoltagemeasurementandtheexactvoltagegener-ated,onthesmalldisplay.

    Pressing theSTOPbuttonstops themeasurement.After themeasurementhasbeenstoppedtheresultremainsdisplayed.

    It ispossibletoscroll throughall theotherresultsavailableonthemaindisplayusingtheR-DAR-PI-DDbutton.Thisbuttoncanalsobeusedbeforethemeasure-mentisbegun.

    Ifthetime-controlledtest- modehasbeenselected,theR(t)buttonmakesitpossibletoaccessallthesavedintervalmeasurementsautomatically.

    IftheALARMfunctionisactivated,abuzzeristriggeredassoonasthemeasure-mentcrossesthethresholdprogrammedintheSET-UPconfigurationmenu.

    Display of values after measurement

    R-DAR-PI-DD Button V-TIME Button

    Main Display Small display Small display if the MR button is pressed (Model 5060)

    ResistanceDARPIDD*Current

    duration (min. sec)duration (min. sec)duration (min. sec)duration (min. sec)duration (min. sec)

    date, time, test voltage, OBJ : TESTdate, time, test voltage, OBJ : TESTdate, time, test voltage, OBJ : TESTdate, time, test voltage, OBJ : TEST

    Capacitance**R(t) duration (min. sec) last test voltage

    * The value of DD is displayed only one minute after the measurement stops.

    **Capacitance (F) measurement is only displayed after the measurement is finished and the circuit has been discharged.

  • Megohmmeter Models 5050/5060 33

    CHAPTER 5

    OPERATION

    NOTE: Charge the instrument fully before use.

    5.1 Measurement Procedure Starttheinstrumentbyturningtheswitchtothecorrespondingposition.

    AllthesegmentsontheLCDscreenaredisplayed,thenthebatteryvoltage.

    Connecttheleadstothe+and-leadstothemeasurementpoints.

    Thevoltageiscontinuouslymeasuredanddisplayedonthesmalldisplay.

    PresstheSTART/STOPbuttontostartthemeasurement.

    Press START/STOP again to stop the measurement. The last resultremains on the display until the next measurement is started, or theselectorswitchisturned.

    WARNING: If an external voltage is present that is greater than the thresholds described in 4.2, the instrument indicates this voltage on the small display, the warning symbol flashes, and the measurement is stopped.

    Note:Thereareseveralspecialfunctionsthatcanbeused(see2.5).

    5.2 Insulation MeasurementThis function allows the instrument to measure insulation between 10k and 10Tatatestvoltageof500V,1000V,2500V,and5000Vorprogrammedvoltagebetween40Vand5100V.

    Usetherotaryswitchtoselecttherequiredtestvoltage(M-500V,orM-1000V,orM-2500V,M-5000V,orVar.M-50-5000V).

    Connecttheinstrumenttotheinsulatortobetested.

    Start the measurement using the START/STOP button and read theresults.

    WARNING: If an external voltage is present that is greater than the thresholds described in 4.2, then measurement is prohibited.

  • 34 Megohmmeter Models 5050/5060

    ScrollthroughtheresultsonthemaindisplaywiththeR-DAR-PI-DDbutton,oronthesmalldisplaywiththeV/TIMEbutton.

    UseR(t)whencarryingouta time-controlled test toscroll through the intervalmeasurement values saved at the sampling rate specified in SET-UP. Thesevaluesareavailableuntilanothermeasurementistakenortheselectorswitchisturned.

    When measuring high levels of insulation (>1G), it isadvised that the Gguardterminalbeusedtoeliminatetheinfluenceofsurfaceleakagecurrents.Theguardisconnectedbetweenthetwomeasurementcontactpoints,andthesurfacesusceptibletosurfacecurrents,(e.g.dusty,dampcableortransformerinsulation).Inthiscase,alligatorclipsarepreferabletotest-probesheldinthehand.

    As soonas insulationmeasurement is stopped, the test circuit is automaticallydischargedusingtheinstrumentsinternalresistance.

    5.3 Operation ExamplesConnection diagram for measurement of low insulation (e.g. motor)

    ( )

    FF 0.1A380V - 10kA

    5x20mm

    110 - 230V50/60 Hz

    20 VA max

    RS-232

    2nd

    SMOOTH ALARM

    R(t)PRINT MEMMR MEGOHMMETERMODEL 5060

    R-DAR-PI-DDPRINTMEM

    V/TIME

    START/STOP1000V CAT III

    2500V

    G OFF

    M -1000V

    M - 2500V

    M - 5000V

    M 50-5000V

    SET-UP

    M - 500V

    +

    -M

    Connection diagram for measurement of strong insulation

    a)Exampleofamotor(reductionofcapacitiveeffects)

    ( )

    FF 0.1A380V - 10kA

    5x20mm

    110 - 230V50/60 Hz

    20 VA max

    RS-232

    2nd

    SMOOTH ALARM

    R(t)PRINT MEMMR MEGOHMMETERMODEL 5060

    R-DAR-PI-DDPRINTMEM

    V/TIME

    START/STOP1000V CAT III

    2500V

    G OFF

    M -1000V

    M - 2500V

    M - 5000V

    M 50-5000V

    SET-UP

    M - 500V

    +

    G

    -M

  • Megohmmeter Models 5050/5060 35

    b) Exampleofacable(reductionofsuperficialleakeffects)

    G

    -

    +

    ( )

    FF 0.1A380V - 10kA

    5x20mm

    110 - 230V50/60 Hz

    20 VA max

    RS-232

    2nd

    SMOOTH ALARM

    R(t)PRINT MEMMR MEGOHMMETERMODEL 5060

    R-DAR-PI-DDPRINTMEM

    V/TIME

    START/STOP1000V CAT III

    2500V

    G OFF

    M -1000V

    M - 2500V

    M - 5000V

    M 50-5000V

    SET-UP

    M - 500V

    External insulator Braid

    CableGuard

    Insulator

    5.4 Capacitance MeasurementCapacitance measurement is automatically carried out during insulation mea-surement.Itisdisplayedafterthemeasurementisstoppedandthecircuitisdis-charged,usingtheR-DAR-PI-DD button.

    5.5 Residual and Leakage Current MeasurementThemeasurementofresidualcurrentcirculatingintheinstallationisautomaticallymadeduringaninsulationmeasurement.Itisdisplayedattheendofthemeasure-mentwiththeR-DAR-PI-DDbutton.

  • 36 Megohmmeter Models 5050/5060

    CHAPTER 6

    MEMORY / RS-232 (5060)

    6.1 RS-232 SpecificationsTheBaudRatecanbesetto300,600,1200,2400,4800,9600,orParallelforprintingresultsonparallelprintersviatheoptionalserial-to-paralleladapter.

    ThissettingiscarriedoutintheSET-UPmenu(see2.5.7).Dataformat:8databits,1stopbit,noparity,Xon/Xoffprotocol.

    Connection to a PC or to a Parallel PrinterCable,PCRS-232,DB9F/F6ftNullModemCable(5060)............. Cat. #2119.45

    2 2

    3 3

    5 5

    9-pin female connector

    to Model 5060

    9-pin female connector

    to serial printer

    6.2 Saving / Recalling Values (MEM/MR Button)

    NOTE: After replacing or completely discharging the battery, the message Memory OFF appears when you try to save a measurement in memory.

    To be able to save your measurements, you will need to do the following:SettheinstrumenttoSET-UPandpresstheMEMbutton(thedisplayshowsMEMAllclr).PresstheMEMbuttonfor2s(thedisplayshowsMEMALLFrEE).Mea-surementsarenowreadytobesaved.

    First Function - MEM (Save)Results can be recorded in the instruments memory and stored along withaddressesdefinedbyanobjectnumber(OBJ)andatestnumber(TEST).

    Anobjectrepresentsaboxintowhichwecanplace99tests.Thus,anobjectcanrepresentamachineoraninstallationuponwhichacertainnumberofmeasure-ments(insulation,resistanceandcontinuity)arecarriedout.

  • Megohmmeter Models 5050/5060 37

    1) Whenthebuttonispressed,theMEMsymbolflashesandthesmalldisplayindicatesthefirstfreeOBJ : TESTnumber(e.g.02:01).Themaindisplayindicates FrEE. TheOBJnumberisthesameasthatofthelastsavedmeasurement,and

    theTESTnumbergoesupinincrementsof1. ItisalwayspossibletomodifyOBJ : TESTwith and . Iftheuserhappenstoselectamemoryaddressthatisalreadyoccupied,

    thenOCCappearsonthemainscreen. IfanewOBJisselected,theTESTgoesbackto01.

    2) By pressing MEM once again, the present measurement results will berecordedintheselectedmemoryaddress(occupiedorotherwise).TheMEM symbolstopsflashingandremainsdisplayed.Thetimeandthedateofthisrecordingarestoredwiththeavailabledata(R,V,t).If a button other than MEM is pressed, or the selector switch is moved before pressing MEM a second time, the instrument exits from MEM mode without saving the results.

    3) Ifatime-controlledtesthasbeeninitiated,intervalvalues(samples)areavail-able.TheyareautomaticallysavedunderthesameOBJ : TESTnumberasthefinalmeasurement.

    Memory Capacity Totalmemoryspace:128kb

    Datamanagement:8kb

    Freememoryspace:120kb

    An insulation measurement requires 80 bytes. A time-controlled test samplerequires10additionalbytes.Therefore,itispossibletorecordabout1500insula-tionmeasurements.

    Free Memory Space

    Thisfunctionisautomaticallyactivatedwhenaresultissaved.

    PressMEMoncetogetthefollowingfreeOBJ : TESTnumber;thebargraphdis-playedcorrespondstotheavailablefreememoryspace.

    Iftheentirememoryisfree,allofthebargraphsegmentsappear.

    Iftheentirememoryisfull,thearrowtotheleftofthebargraphflashes.

    Assoonasthestorageiscomplete,thebargraphdisappears.

    Eachsegmentofthebargraphequalsapproximately50recordings.

  • 38 Megohmmeter Models 5050/5060

    Second Function - MRTheMRfunctionallowsarecallofanysaveddatafrommemory,regardlessofthepositionoftherotaryselectorswitch.

    Whenthebuttonispressed,theMRsymbolisdisplayed.Thesmalldis-playindicatesthelastOBJ : TESTnumbertobeassigned(e.g.02:11).

    0211flashesoppositetheTESTsymbol.Usethe and buttonstoselectthedesiredOBJ : TESTnumber.

    If a newOBJ is selected,TEST is automatically set to themaximumstorednumber.Atthisstage,it ispossibletoreviewtheentirememorywiththe and buttons,sincethemeasurementvaluescorrespondingtotheselectedOBJ : TESTnumberaredisplayedonthemaindisplay.UsetheR-DAR-PI-DDbuttontoscrollthroughthem.

    The V/TIMEbuttongivesaccesstothedate/ time/V/OBJ-TESTnumberforeachresult.

    If therecordingselectedbytheOBJ : TESTnumbercorrespondstoatime-controlled test, theR(t) values can be accessed by pressing theR(t)button.Thesmalldisplaychangesandindicatesmin:sec(timeof1stsample)andthesymbol flashesonthescreen.Youcanscrollthroughtheothersampleswiththe button.

    ToexitfromR(t) modeandreturntothenormalmemoryrecallmode(OBJ : TEST),press R(t) or R-DAR-PI-DDagain.

    To exit the MR function,pressMRbuttononceagainorturntheselectorswitch.

    6.3 Printing Measured Values (PRINT/PRINT MEM Button - Model 5060)If youusea serial printer, choose theappropriate communication speed (baudrate)intheSET-UPmenu,between300-9600baud,thenprogramtheprintertotheformatrunbytheinstrument(see6.1).

    Therearetwoprintingmodespossible:

    Instantaneousprintingofmeasurement(PRINT)

    Printingofrecordeddata(PRINTmemory)

    NOTE: If the data transmission to the printer is successful, the COM symbol flashes once on the screen. If a problem has occurred, the COM symbol remains on the LCD screen without flashing.

  • Megohmmeter Models 5050/5060 39

    6.4 Instantaneous Printing of Measurements (PRINT button)FollowingameasurementorafteraccessingvaluesinMRmode,usethePRINT functiontoprintthemeasurementresults.

    NOTE: PRINT only works when the test is stopped by pressing the START/STOP button and the value is displayed on the LCD.

    PressthePRINTbuttontoprintthefollowing: 1setofmeasurements(V/R/DAR/PI/DD/date/ time)fornormaltests.

    TheR(t)valuesifthetime-controlledtesthasbeenused.

    To stop printing,movethepositionoftheselectorswitch.

    Depending on the function used, the following models are obtained:

    Insulation MeasurementAEMCInstrumentsModel5060Instrumentnumber:000001INSULATIONRESISTANCETESTOBJECT:01 .....................................TEST:01(only printed in MR mode)Description:............................................................................................................................Date: ............................................03.31.2001Starttime: ............................................14:55Testduration: .........................15min.30secTemperature: ..............................C.........FRelativehumidity: ...................................... %Testvoltage: .......................................1000VInsulationresistance(R): ...............385GOhmDAR(R1/R30): .................................1.234PI(R10/R1):......................................2.345Comments: ............................................................................................................................Dateofnexttest: ............................../..../.......

    After a time-controlled test other results are printed (interval samples): Time Resistance Voltage00:30 35.94GOhm 1005V01:00 42.00GOhm 1005V01:30 43.50GOhm 1005Vetc...

    A space for the operators signature is available at the end of the printout.

  • 40 Megohmmeter Models 5050/5060

    6.5 Printing Data in Memory (PRINT MEM button)Printsoutthecontentsoftheinstrumentsread-writememory.

    Thesmalldisplayindicates01:01fortheOBJ : TESTnumber(startingaddressofprinting).

    Themaindisplayindicatesthelastrecordinginmemory(theendaddressofprint-ing).e.g.12:06.

    01flashesoppositethe OBJposition(usethe and buttonstodefinethestart/endaddressesoftheprintout).

    To exit without printing,alterthepositionoftheselectorswitch. To proceed with printing,pressthePRINTbuttononcemore. To stop printing,alterthepositionoftheselectorswitch.

    Onlythemainresultsareprintedout.

    Example:AEMCInstrumentsModel5060Instrumentnumber:000001

    INSULATION RESISTANCE TEST OBJECT:01 .....................................TEST:01Description:.............................................................................................................................Date: ............................................03.31.2001Startingtime: ......................................13h35Runningtime: ..........................16min27secTemperature: ................................C.........FRelativehumidity: ....................................... %Testvoltage: ......................................5000VInsulationresistance(IR): .................3.85 TDAR: .....................................................1.234PI: .........................................................2.382DD: .................................................................Capacitance:.............................................. FIresidual: ................................................... nAComments: .............................................................................................................................Dateofnexttest: ............................................

    OBJECT:01 .....................................TEST:02Date: ............................................03.31.2001Startingtime: ......................................15h10Runningtime: ..........................15min30sec

  • Megohmmeter Models 5050/5060 41

    Temperature: ................................C.........FRelativehumidity: ....................................... %Testvoltage: ......................................1000VInsulationresistance(IR): ................3.85 GDAR: .....................................................1.234PI: .........................................................2.345DD: .................................................................Capacitance:.............................................. FIresidual: ................................................... nAComments: .............................................................................................................................Dateofnexttest: ............................................

    A space for the operators signature is available at the end of the printout.

  • 42 Megohmmeter Models 5050/5060

    CHAPTER 7

    DATAVIEW SOFTWARE

    7.1 Installing DataView

    DO NOT CONNECT THE INSTRUMENT TO THE PC BEFORE INSTALLING THE SOFTWARE AND DRIVERS.

    Minimum Computer Requirements: WindowsXP/WindowsVista&Windows7/8(32/64bit)

    256MBofRAMforWindowsXP 1GBofRAMforWindowsVista&Windows7/8(32bit) 2GBorRAMforWindowsVista&Windows7/8(64bit)

    80MBofharddiskspace(200MBrecommended)

    USBport

    Windows is a registered trademark of Microsoft Corporation in the United States and other countries.

    NOTE:When installing, the user must haveAdministrative access rightsduring the installation.Theusers access rights canbe changedafter theinstallationiscomplete.

    DataViewmustbereinstalledforeachuserinamulti-usersystem.

    USB Flash Drive Install

    1. InserttheUSBstickintoanavailableUSBport(waitfordrivertobeinstalled).2. IfAutorunisenabledthenanAutoPlaywindowshouldappearasshown.

  • Megohmmeter Models 5050/5060 43

    NOTE: If Autorun is disabled, it will be necessary to open Windows Explorer, then locate and open the USB stick drive labeled DataView to view the files on the drive.

    3. IntheAutoPlaywindow,selectOpen Folder to view Files.

    4. Double-clickonSetup.exefromtheopenedfolderviewtolaunchtheData-Viewsetupprogram.

    NOTE: If installing onto a Vista based computer the User Account Control dialog box will be displayed. Select the Allow option to proceed.

    5. A Set-up window,similartotheonebelow,willappear.

    Figure 7-1

    There are several different options to choose from.Someoptions(*) require aninternet connection.

    DataView, Version x.xx.xxxx-InstallsDataViewontothePC.

    *Adobe Reader-LinkstotheAdobewebsitetodownloadthemostrecentversionofAdobeReadertothecomputer.Adobe Reader is required for viewing PDF documents supplied with DataView.

    *DataView Updates -Links to theonlineDataViewsoftwareupdates tocheckfornewsoftwareversionreleases.

    *Firmware Upgrades -Linkstotheonlinefirmwareupdatestocheckfornewfirmwareversionreleases.

  • 44 Megohmmeter Models 5050/5060

    Documents -Showsa listof instrument relateddocuments thatyoucanview. AdobeReaderisrequiredforviewingPDFdocumentssuppliedwithDataView.

    6. DataView, Version x.xx.xxxxoptionshouldbeselectedbydefault.SelectthedesiredlanguageandthenclickonInstall.

    7. TheInstallation Wizardwindowwillappear.ClickNext.

    8. Toproceed,acceptthetermsofthelicenseagreementandclickNext.

    9. IntheCustomer Informationwindow,enteraNameandCompany,thenclickNext.

    10. IntheSetup Typewindowthatappears,selecttheCompleteradiobuttonoption,thenclickNext.

    11. IntheSelect Featureswindowthatappears,selecttheinstrumentscontrolpanelthatyouwanttoinstall,thenclickNext.

    NOTE: The PDF-XChange option must be selected to be able to generate PDF reports from within DataView.

    Figure 7-2

    12. IntheReady to Install the Programwindow,clickonInstall.

    13. If the instrumentselected for installation requires theuseofaUSBport,a warningboxwillappear,similartoFigure7-3.ClickOK.

  • Megohmmeter Models 5050/5060 45

    Figure 7-3

    NOTE: The installation of the drivers may take a few moments. Windows may even indicate that it is not responding, however it is running. Please wait for it to finish.

    14. Whenthedriversarefinishedinstalling, the Installation Successfuldialogboxwillappear.ClickonOK.

    15. Next,theInstallation Wizard Completewindowwillappear.ClickonFinish.

    16. A Questiondialogboxappearsnext.ClickYestoreadtheprocedureforcon-nectingtheinstrumenttotheUSBportonthecomputer.

    NOTE: The Set-up window remains open. You may now select another option to download (e.g. Adobe Reader), or close the window.

    17. Restartyourcomputer,thenconnecttheinstrumenttotheUSBportonthecomputer.

    18. Onceconnected,theFound New Hardwaredialogboxwillappear.Windowswillcompletethedriverinstallationprocessautomatically.

    Shortcuts forDataView andeach instrument controlpanel selectedduring theinstallationprocesshavebeenaddedtoyourdesktop.

    NOTE: If you connected your instrument to the computer before installing the software and drivers, you may need to use the Add/Remove Hardware utility to remove the instrument driver before repeating the process.

  • 46 Megohmmeter Models 5050/5060

    7.2 Connecting the Model 5060 to your ComputerTheModel5060issuppliedwithaserialinterfacecablenecessaryforconnectingtheinstrumenttothecomputer.Thiscable(Cat.#2119.45)isequippedwitha9-pinfemaleconnectoroneachend,thusprovidingtheabilitytointerfacewithcomput-ersthathave9-pinconnectors(seediagramin6.1).

    Toconnectthemegohmmetertoyourcomputer:

    Connectthe9-pinconnectorfromoneendofthecabletotheserialportonthefrontpaneloftheModel5060megohmmeter.

    Connectthe9-pinconnectorfromtheotherendofthecabletoanavail-ableserialportonthebackofyourcomputer.

    YouarenowreadytocommunicateusingDataViewwiththemegohmmeter.

    7.3 Using DataView

    OncetheserialconnectionbetweenthecomputerandtheModel5060hasbeencompleted,asdescribedabove,startDataView.

    TherearetwowaystoopenandusetheDataViewsoftware:

    Using the Megohmmeter Icon

    Double-clickthe Megohmmeter Icon thatwascreatedduringinstallation,locatedonthedesktop.

    DataViewwillopenandtheConnectionwindowwillappear (seeFigure7-6).

    Using the DataView Icon

    Double-click the DataView Icon that was created during installation,locatedonthedesktop.

    DataViewwillopenanddisplaytheDataView Quick Start window (seeFigure7-4).

    ClicktheConfigure Instrument buttonandtheConnection window will appear(seeFigure7-5).

    TheQuickStartDialogBoxcontainsthefollowingoptions:

    Configure Instrument: Createsaconfigurationforthemegohmmeter.

    Create View from Template: Creates a printable report fromexistingdataalreadystoredonthecomputer.

  • Megohmmeter Models 5050/5060 47

    Figure 7-4

    7.3.1 Configuring the Instrument

    Toconfiguretheinstrument,performthefollowingsteps.

    1. ClicktheConfigure Instrumentbutton.Adialogboxwillopen,allowingyoutoselecttheCommunicationRateandtheSerialPortforcommunicatingwiththeModel5060.

    Figure 7-5

    2. Makesurethattheserialportdisplayedinthedialogboxmatchestheportyoupluggedtheserialcableinto.Ifthecorrectserialportisnotselected,clickonthedropdownarrowandselecttheproperone.

    3. TheCommunicationRatewilldefault to9600andshouldbeacceptable toanycomputer.ThisisalsothedefaultratefortheModel5060.YoumayselectadifferentratebyclickingontheCommunicationRatedrop-downmenuandselectingfromtheavailablevalues.

    4. OncethepropercommunicationparametershavebeenspecifiedclickontheOKbutton.

  • 48 Megohmmeter Models 5050/5060

    NOTE: The instrument must be configured for the same communications rate specified in Figure 7-5. If it is not, then you must either select the corresponding rate or change the instruments rate.

    When a serial communication link is established, DataView will automaticallyidentifytheinstrumentthatitisconnectedto.Aninstrumentselectionscreenwillappearforafewsecondswhilethistakesplace.

    Onceidentificationiscomplete,theSetupdialogboxwillappearonthescreen.

    Figure 7-6

    ThisdialogboxletsyouconfigureeveryaspectoftheModel5060megohmmeter.Eachfield in thedialogbox is identical to theprogrammable featuresavailablefromtheinstrumentfrontpanelitself.Referto2.5foracompletedescriptionofeachconfigurablefunction.

    Severalofthefunctionsareconfiguredbytypingtheappropriatevalueinthefieldprovided,suchastestruntime,whichmaybeprogrammedfromavalueofoneminuteto59minutes59seconds.

    Othersareconfiguredbyclickingontheappropriateradiobutton,suchas,activat-ingordeactivatingthebuzzerbyeitherselectingONorOFF.

    Therighthandsideofthissetupdialogboxcontainsninecommandbuttons:

    Close:ClosestheConfigurationdialogboxandwillbringuptheControlPanel.

    Write to Inst:Programsthemegohmmeterusingthecurrentsettings.

    Read from Inst:Reads thecurrentconfigurationof themegohmmeterattachedviatheserialcable.

  • Megohmmeter Models 5050/5060 49

    Save to Disk:Savesthecurrentconfiguration.Thisfilewillresideonthecomputersdiskdrive.Savingdifferentconfigurationsetupscanbeusefulforfuturefunctionsandtests.

    Load from Disk:Retrievesasavedfilefromthecomputersdiskdrivetobeusedinprogrammingthemegohmmeter.

    Defaults: Resetsallfields in theSetupdialogboxback to theoriginalfactorysettings.

    Set Clock:Programsthecomputerstimeanddateintotheconfigurationofthemegohmmeter.

    Clear Memory:Erasesallcontentsinallmemorylocationsinthemegohmmeter.

    Download:Retrievesthestoreddatafromallmemorylocationswithinthemegohmmeterforuseonthecomputersdiskdrivememory.

    OnceallfieldsintheSetupdialogboxhavebeenconfigured,itisrecommendedthatyouperformthefollowingsteps.

    1. ClickontheSave button

    2. Nameandsavethefile.

    3. ClickontheWritebuttontoconfigurethemegohmmeter.

    4. ClickontheClose buttontoexitSetupandopentheControlPanel.

    NOTE: While connected, both the DataView and Megohmmeter windows are open. They have similar main menu options, so you need to be aware of which window you are working in.

    You can tell which window you are working in by looking at the top left corner of the active window. It will either display DataView or Megohmmeter.

    You can toggle between both windows via the Windows taskbar at the bottom of your screen.

    7.3.2 Running the Test

    Afterconfiguringtheinstrument(see7.3.1),theControlPanelwillautomaticallyopen.TheControlPanelisusedtoinitiateatimedtestfromthecomputer.

    TheControlPaneldialogboxconsistsoffivecommandbuttons:

    Start:Thisbuttonbeginsatimedtest.

    Clear Display:Clearsthestatisticaldata,onthelowerleftsideofthescreen,andthegraph,ontherightsideofthescreen.

    Save to Disk:Savesthecurrentdatatoafileinthecomputer.

  • 50 Megohmmeter Models 5050/5060

    Configure:ReturnstothemegohmmeterSetupdialogboxforconfigur-ingtheinstrument.

    Download: Retrievesdatafromthememoryofthemegohmmeterforstorageinthecomputer.

    Figure 7-7

    Torunatimedtest,performthefollowingsteps:

    1. Selecttheappropriatevoltageforthetestbyclickingontheradiobuttonnexttothedesiredselection.

    2. PresstheStartbutton.Thiswillrunthetestattheselectedvoltage. Asthetestisrunning,allotherbuttonsaregrayedoutandtheStart button

    becomes a Stopbutton.

    Oncethetesthasstarted,agraphwillbegintoappearandbuildontherightsideoftheControlPanel(seeFigure7-7).Thestatisticaldataforthetimeofthetest,DAR,PI,resistancevalueandotherinformationwillappearastheyarecompletedonthelowerleftsideoftheControlPanel.

    Thetestwillstopautomaticallyattheendoftheprogrammedruntime.YoumayalsostopthetestatanytimebypressingtheStopbutton.Whenthetestiscom-plete,iftheCalculateDDcheckboxwasenabled,theModel5060willautomati-callycalculatedielectricdischargeratio.Thistesttakes60seconds.

    NOTE: On small motors and machinery it is possible that the capacitance values are too low for this test to be completed accurately. In these cases, simply press the Stop button to stop the dielectric discharge test.

  • Megohmmeter Models 5050/5060 51

    3. OncethetesthasbeencompletedpresstheSave to Diskbuttontosavetheresultsofthetestjustcompletedtoadiskdriveinyourcomputer.

    4. Namethefile,thenpresstheSavebutton.

    Thiswillopenthesessionpropertiesdialogbox:

    Figure 7-8

    5. Onceyouhavecompletedfilling inall the informationon thescreen,pressOK.Thiswillopenadialogboxallowingyoutoaddadescriptivenamefortheobjectandtestjustperformed.

    6. PresstheOK button.Theinformationisnowsavedtothecomputersmemory.

    7. Closethedialogboxbyclickingonthesmall atthetoprighthandcornerofthedialogbox.

    8. Youwillthenbeaskedifyouwanttoopenthedatabasejustsaved.ClicktheYes button,whichwillopenadialogboxlabeledCreateViewfromTemplate.

    9. IntheGroupswindowofthisdialogbox,clickonmegohmmeterthenselectthedesiredtemplate,ifotherthanthedefaultisdesired.

    10. InStep2:SpecifyDatabase(s)window, thefileyou justsavedandnamedshouldbevisible.Ifthisisthefileyouwishtogenerateareportfrom,clicktheOKbutton.Thereportwillnowbeavailableon thescreen includinggraph,individualdatapointsandallstatisticaldata.

    11. Toprintthisreport,eitherclickonthePrinticonorclickontheFilecommandandthenPrint.

  • 52 Megohmmeter Models 5050/5060

    7.4 DataView Megohmmeter Templates

    7.4.1 Megohmmeter Summary Report Template

    TheMegohmmeterSummaryReport templategeneratesaviewwith twowork-sheets.AnexampleofthefirstpageisshowninFigure7-9below.

    Figure 7-9

    Thisworksheetcontainsthefollowingframes:

    Trend Graph: Thisframeprovidesagraphofthemeasurementpointsintheselectedtest.

    Channel Summary: This frame displays a textual summary of theselected test.

    Channel List: Thisframedisplaysa listof tests. Ifmorethanonetestislistedyoucandisplaytheresultsofeachtestbyselectingitwiththemousepointer.

    Trend Summary List: Thisframedisplaysatextuallistsummaryofthesamplepointsforaselectedtest.Eachsummarylinecontainsuser-spec-ifiedparametersofagroupofsamplepoints.

    AnexampleofthelastworksheetisshowninFigure7-10.

  • Megohmmeter Models 5050/5060 53

    Figure 7-10

    hispagecontainsthefollowingframes:

    Title Box-containingthetextTrendSummaryReport Operator Summary-containingoperatorinformation Site Summary -containingtestsiteinformation Comments Summary-containingtypedinoperatorcomments Database -containingthenameof thedatabaseusedtogeneratethe

    report

    Eachoftheseframesiscustomizedbyusingtheassociatedpropertiesdialogbox.Thepropertiesdialogboxforaframeisdisplayedbyselectingtheframe,right-clickingandselectingPropertiesfromthelistdisplayed.

  • 54 Megohmmeter Models 5050/5060

    CHAPTER 8

    MAINTENANCE

    Useonlyfactoryspecifiedreplacementparts.AEMCwillnotbeheldresponsibleforanyaccident,incident,ormalfunctionfollowingarepairdoneotherthanbyitsservicecenterorbyanapprovedrepaircenter.

    8.1 Recharging the Battery

    Ifthe symbolisflashing,thebatteryneedstoberecharged.

    Connect the instrument to the120VACpower cord via theconnector (chargingstartsautomatically).

    bAtonthesmalldisplayandCHrG or chAronthemaindisplay,signi-fiesfastcharginginprogress.

    bAtonthesmalldisplayandCHrGflashingonthemaindisplay,signi-fiesslowcharging(startsoffwithfastcharge,temperatureconditionspermitting).

    bAtonthesmalldisplayandFULLonthemaindisplay,signifiesthatchargingisover.

    Iftheinstrumentisstartedupandthebatteryvoltageis>8V,thenthenormaluseofthedeviceispermitted.

    NOTE: The battery should only be changed by an authorized repair facility recog-nized by AEMC Instruments.

    Changing the battery causes data to be lost from the memory.

    PresstheMEM/MRbutton(OFFisdisplayed).

    ProceedwithatotalclearingofmemoryinSET-UP(see2.5.7)sotheMEM/MRfunctionscanbeusedagain.

  • Megohmmeter Models 5050/5060 55

    8.2 Fuse ReplacementIfFUSE -G-flashesonthedisplay,thefusemustbereplaced,takingalltheneces-saryprecautionswhenopeninguptheinstrument.

    Make sure that none of the terminals are connected and that the selector switch is set to OFF before opening up the case.

    Onlyusethetypeoffuseshownonlabelinsidetheunitscover:

    0.1A-FastActing-380V,5x20mm,10kA

    8.3 Cleaning

    Disconnect the instrument from any source of electricity.

    Useasoftclothlightlydampenedwithsoapywater.Rinsewithawetclothandthendrywithadrycloth.Donotusealcohol,solventsorhydrocarbons.

    8.4 StorageIftheinstrumentisnotusedforanextendedperiodoftime,itisrecommendedtochargetheinstrumenteverytwoorthreemonths.

  • 56 Megohmmeter Models 5050/5060

    Repair and CalibrationTo ensure that your instrument meets factory specifications, we recommendthat itbescheduledbacktoourfactoryServiceCenteratone-year intervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.

    For instrument repair and calibration:YoumustcontactourServiceCenterforaCustomerServiceAuthorizationNumber(CSA#). This will ensure that when your instrument arrives, it will be trackedandprocessedpromptly.Pleasewrite theCSA#on theoutsideof theshippingcontainer.Iftheinstrumentisreturnedforcalibration,weneedtoknowifyouwantastandardcalibration,oracalibration traceable toN.I.S.T. (Includescalibrationcertificateplusrecordedcalibrationdata).

    Ship To: ChauvinArnoux,Inc.d.b.a.AEMCInstruments 15FaradayDrive Dover,NH03820USA

    Phone: (800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail: [email protected]

    (Orcontactyourauthorizeddistributor)

    Costs for repair, standard calibration, and calibration traceable to N.I.S.T. areavailable.

    NOTE: You must obtain a CSA# before returning any instrument.

    Technical and Sales Assistance

    Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

    ChauvinArnoux,Inc.d.b.a.AEMCInstruments 200FoxboroughBoulevard Foxborough,MA02035USA

    Phone: (800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail: [email protected] www.aemc.com

    NOTE: Do not ship Instruments to our Foxborough, MA address.

  • Megohmmeter Models 5050/5060 57

    Limited WarrantyTheMegohmmeterModel5050/5060iswarrantedtotheownerforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMCInstruments,notbythedistributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMCInstruments.

    Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html.

    Please print the online Warranty Coverage Information for your records.

    What AEMC Instruments will do:Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC Instrumentswill,at itsoption,repairorreplacethefaultymaterial.

    REGISTER ONLINE AT: www.aemc.com

    Warranty Repairs

    What you must do to return an Instrument for Warranty Repair: First,requestaCustomerServiceAuthorizationNumber(CSA#)byphoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideof theshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:

    Ship To: ChauvinArnoux,Inc.d.b.a.AEMCInstruments 15FaradayDriveDover,NH03820USA

    Phone: (800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail: [email protected]

    Caution: To protect yourself against in-transit loss,we recommend you insureyourreturnedmaterial.

    NOTE: You must obtain a CSA# before returning any instrument.

  • 08/13

    99-MAN 100258 v22

    Chauvin Arnoux, Inc. d.b.a. AEMC Instruments15 Faraday Drive Dover, NH 03820 USA Phone: (603) 749-6434 Fax: (603) 742-2346

    www.aemc.com

    INTRODUCTION1.1International Electrical Symbols1.2Definition of Measurement Categories1.3Receiving Your Shipment1.4Ordering Information1.4.1Accessories and Replacement Parts

    1.5Accessory Information1.5.1DataView Software (for Model 5060)

    PRODUCT FEATURES2.1Megohmmeter Features2.2Digital Features2.3Bargraph2.4Symbols2.5Push-button Functions2.5.12nd Button2.5.2V-TIME / Button2.5.3R-DAR-PI-DD / R(t) Button2.5.4 / ALARM Button2.5.5 / Smooth Button2.5.6 Button2.5.7SET-UP Function2.5.8Clearing the Memory2.5.9Communication Rate (RS-232)2.5.10Default Device Configuration2.5.11Voltage Disturbance Limit2.5.12Automatic Measuring Range2.5.13Blocking (Disabling) Test Voltages

    SPECIFICATIONS3.1Reference Conditions3.2Voltage3.3Insulation Resistance3.4Power Supply3.5Environmental Specifications3.6Mechanical Specifications3.7Safety Specifications3.8Variations in Operating Range

    MEASUREMENT FUNCTIONS4.1AC/DC Voltage4.2Insulation Measurement

    OPERATION5.1Measurement Procedure5.2Insulation Measurement5.3Operation Examples5.4Capacitance Measurement5.5Residual and Leakage Current Measurement

    MEMORY / RS-232 (5060)6.1RS-232 Specifications6.2Saving / Recalling Values (MEM/MR Button)6.3Printing Measured Values6.4Instantaneous Printing of Measurements (PRINT button)6.5Printing Data in Memory (PRINT MEM button)

    DATAVIEW SOFTWARE7.1Installing DataView7.2Connecting the Model 5060 to your Computer7.3Using DataView7.3.1Configuring the Instrument7.3.2Running the Test

    7.4DataView Megohmmeter Templates7.4.1Megohmmeter Summary Report Template

    MAINTENANCE8.1Recharging the Battery8.2Fuse Replacement8.3Cleaning8.4Storage

    Repair and CalibrationTechnical and Sales AssistanceLimited WarrantyWarranty Repairs