high performance maldi ms, ms/ms, and multiplexed ms/ms tissue imaging

24
High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging Boone M. Prentice 1 , Kevin M. Hayden 2 , Marvin L. Vestal 2 , Richard M. Caprioli 1 1 Mass Spectrometry Research Center, Department of Biochemistry, Vanderbilt University, Nashville, TN 37235 2 SimulTOF Systems, Sudbury, MA 01776 MSACL 2014 US March 1 - 5, 2014, San Diego, CA

Upload: keena

Post on 22-Feb-2016

49 views

Category:

Documents


3 download

DESCRIPTION

High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging . MSACL 2014 US March 1 - 5, 2014, San Diego, CA. Boone M. Prentice 1 , Kevin M. Hayden 2 , Marvin L. Vestal 2 , Richard M. Caprioli 1 - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

Boone M. Prentice1, Kevin M. Hayden2, Marvin L. Vestal2, Richard M. Caprioli1

1Mass Spectrometry Research Center, Department of Biochemistry, Vanderbilt University, Nashville, TN 37235

2SimulTOF Systems, Sudbury, MA 01776

MSACL 2014 USMarch 1 - 5, 2014, San Diego, CA

Page 2: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

Why Imaging Mass Spectrometry?IMS combines molecular specificity with location

The multiplexed nature of MS analysis allows for the parallel acquisition of many different molecular signals, each which can be reconstructed to give a molecular picture.

2

medullaRat Kidneycortex

Page 3: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

3

Imaging Mass Spectrometry

Lipid & Protein Ion Images

…………………………………………

…………………………………………Matrix

application

Matrix-Assisted Laser Desorption/Ionization

Laser

Images for individual m/z values integrated over all pixels

MS spectrum for each x, y coordinate (pixel)

Sample Preparation

MatrixSA – proteins/peptidesDHB –proteins/peptidesCHCA- lipids/peptidesDAN– lipids

Application MethodSpraying (manual, robotic)SublimationSpotting (robotic)

Data Processing

Page 4: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

4

IMS Performance ConsiderationsTraditional MS Figures of Merit

• Mass resolving power: m/Δm

• Mass accuracy: difference between measured mass and exact mass

• Sensitivity: overall response of the instrument for a given analyte

• Dynamic range: range of detectable signals

• MS/MS capabilities: ability to perform fragmentation experiments

Special IMS Considerations

• Spatial resolution: distance between two neighboring pixels on the sample surface

• Throughput (acquisition rate): number of spectra acquired per unit time

• Data/file sizes: storage costs and processing abilities

Page 5: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

5

MALDI Imaging Platforms

J.M Spraggins and R.M. Caprioli, J. Am. Soc. Mass Spectrom. 2011, 22, 1022-1031.

• High mass resolution (m/Δm = 20,000)• High repetition rate laser (5 kHz Nd:YLF)• High digitizer acquisition rate (50-100 pixels/sec)• Continuous laser raster sampling• MS/MS

Next Generation MALDI TOF SimulTOF 200 Combo & 300 Tandem

MALDI TOF Bruker Daltonics Autoflex II

LIT Thermo LTQ XL

FT-ICR Bruker 9.4 T FTMS

IM Q-TOF Waters Synapt G-2

𝐿𝑎𝑡𝑒𝑟𝑎𝑙 𝑆𝑝𝑎𝑡𝑖𝑎𝑙 𝑅𝑒𝑠𝑜𝑙𝑢𝑡𝑖𝑜𝑛=𝐻 . 𝐴(𝑣𝑠𝑡𝑎𝑔𝑒

𝑓 𝑟𝑒𝑝 )

Page 6: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

6

Performance TradeoffsLow Spatial Resolution (>100 μm)

High Spatial Resolution (<10 μm)

Minutes Hours

1 x 108 1 x 105

<5 GB > 10GB

Acquisition Time

Signal Intensity

File Size

Page 7: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

7

#Pixels = f(Spatial Resolution, Area)

25 pixels 2,500 pixels

High-spatial resolution and/or large area IMS experiments can require prohibitive

analysis times.

Page 8: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

8

Pixel Size: The Importance of Speed

~21,000 pixels/3hr MALDI TOF

~9,800 pixels/3hr LIT

~7,700 pixels/3hr FT-ICR

~7,200 pixels/3hr IM Q-TOF

Page 9: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

9

High Speed MALDI TOF~345,000 pixels/3hr Next Generation MALDI TOF

~21,000 pixels/3hr MALDI TOF

~9,800 pixels/3hr LIT

~7,700 pixels/3hr FT-ICR

~7,200 pixels/3hr IM Q-TOF

Page 10: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

10

MS Protein Imaging

Red: S100A8 protein

Green: Hemoglobin alpha chain

Sample: +Mouse Kidney (infected w/ S. aureus)Lateral Spatial Resolution: 25 µmVertical Step: 50 µmWash, Matrix: Carnoy’s, CHCA/DHB mix

Laser Rep. Rate: 1000 HzAcquisition Rate: 20 pixels/secondPixels: 15,940 pixelsAnalysis Time: ~20 minutes

Abscesses

~15,000 m/z

~10,160 m/z1 mm

Page 11: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

11

MS Lipid Imaging

Sample: +Rat BrainLateral Spatial Resolution: 50 µmVertical Step: 50 µmMatrix: DANLaser Rep. Rate: 1000 HzAcquisition Rate: 20 pixels/secondPixels: 39,073 pixelsAnalysis Time: ~40 minutes

Average spectrum across tissue

ventriclecorpus callosumfornix

anterior commissure

striatum

Page 12: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

12

Average spectrum across tissue

The Need for High Mass Resolution

Sample: +Rat Brain (lipids)Lateral Spatial Resolution: 25 µmVertical Step: 25 µmMatrix: CHCA

Laser Rep. Rate: 5000 HzAcquisition Rate: 100 pixels/secondPixels: 178,154 pixelsAnalysis Time: ~90 minutes

Page 13: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

13

The Need for High Mass Resolution

~12,000 FWHM resolution at 869 m/z

869.25 ± 0.25 m/z

Sample: +Rat Brain (lipids)Lateral Spatial Resolution: 25 µmVertical Step: 25 µmMatrix: CHCA

Laser Rep. Rate: 5000 HzAcquisition Rate: 100 pixels/secondPixels: 178,154 pixelsAnalysis Time: ~90 minutes

Page 14: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

14

869.32

Δ = 0.19 m/z

869.13

The Need for High Mass Resolution

corpus callosumthalamus hippocampus

Page 15: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

15

MALDI TOF/TOF

• High resolution timed ion selector (<5 Da)• Collision cell for efficient fragmentation• High repetition rate laser (5 kHz Nd:YLF)• High digitizer acquisition rate (50-100

pixels/sec)• Continuous laser raster sampling

MS/MS can provide a way to distinguish isobaric species and give sequence information on an analyte of interest.

Vg

VVelocity and Space Focus

Vp

Detector

Ion Accelerator

Collision Cell

Timed Ion Selector

Page 16: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

16

MS Lipid Imaging

Sample: +Rat Brain (lipids)Lateral Spatial Resolution: 50 µmVertical Step: 50 µmMatrix: DAN

Laser Rep. Rate: 1000 HzAcquisition Rate: 20 pixels/secondPixels: 51,120 pixelsAnalysis Time: ~45 minutes

869.2 ± 0.4 m/z

Page 17: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

17

MS/MS Lipid ImagingMS/MS Precursor: 869.2 Sample: +Rat Brain (lipids)Lateral Spatial Resolution: 50 µmVertical Step: 50 µmMatrix: DANLaser Rep. Rate: 1000 HzAcquisition Rate: 20 pixels/secondPixels: 48,441 pixelsAnalysis Time: ~45 minutes

hippocampus

Page 18: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

18

MS/MS Drug ImagingSample: +Rat Liver (synthetically does with

Rifampicin, 400 mg liver immersed in 200 µM RIF solution for 100 hrs)

Lateral Spatial Resolution: 50 µmVertical Step: 50 µm

Matrix: THAPLaser Rep. Rate: 1000 HzAcquisition Rate: 20 pixels/secPixels: 14,192 pixelsAnalysis Time: ~15 minutes

[M-H]- [M-H-424]-CID

821 m/z 397 m/z1 mm

100%

0%

821 → 397 m/z

Page 19: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

19

“Multiplexed” MS/MS AnalysisIon AcceleratorCollision Cell

Timed Ion Selector

• High resolution timed ion selector (<5 Da)• Selection of multiple precursor ions allows

for multiple MS/MS transitions to be performed in a single laser shot.

Monitoring MS/MS transitions is useful, but throughput is limited when only examining a single precursor ion.

Page 20: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

20

“Multiplexing”Sample: +Rat Brain (lipids)Lateral Spatial Resolution: 50 µmVertical Step: 50 µmMatrix: DAN

Laser Rep. Rate: 1000 HzAcquisition Rate: 20 pixels/secondPixels: 47,204 pixelsAnalysis Time: ~40 minutes

MS Imagecorpus callosumthalamus hippocampus

cortex

Page 21: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

21

Multiplexed Imaging

MS/MS Images

MS/MS of 735 m/z AND 827 m/z

corpus callosum

hippocampus

Page 22: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

22

Conclusions• Going from hours to minutes: High speed IMS

produces high quality molecular images.• Making the unseen, seen: Isobaric or near isobaric

species require special consideration.– High resolution MS– MS/MS

• Informing biology: Coupled with the multiplexed nature of MS-based acquisition, the high throughput methodologies described herein offer viable means for studying complex biological systems in situ.

Page 23: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging

AcknowledgementsRichard CaprioliKevin ScheyJeremy NorrisMichelle ReyzerDavid HacheyKristie RoseAndre ZavalinJeff SpragginsRaf Van de PlasErin Seeley

Junhai YangPeggi AngelJere ComptonJamie AllenBrian HacheyAudra JuddLisa ManierDavid AndersonMegan GesselDhananjay Sakrikar

Domenico TavernaKerri GroveJessica MooreDavid RizzoChad ChumbleyJamie WenkeFaizan ZubairKyle FloydMonika Murphy

Mass Spectrometry Research Center

SimulTOF SystemsChristina Hsieh VestelGeorge MillsKenneth Parker

FundingNIH T32 ES007028NIH/NIGMS 5P41 GM103391-03MSACL Young Investigator Travel Award

Page 24: High Performance MALDI MS, MS/MS, and Multiplexed MS/MS Tissue Imaging