electro – surgical unit diathermy

21
ELECTRO – SURGICAL UNIT ELECTRO – SURGICAL UNIT DIATHERMY DIATHERMY What is about the unit? What is about the unit? It is a high frequency electric It is a high frequency electric current producing machine current producing machine generates enough heat to cut, generates enough heat to cut, coagulates and destroys body coagulates and destroys body tissue tissue

Upload: salinder-kaur-dhaliwal

Post on 20-Jan-2016

76 views

Category:

Documents


2 download

DESCRIPTION

what is an electro surgical unit? An introduction to the electro surgical unit used in the operation theatres

TRANSCRIPT

Page 1: ELECTRO – SURGICAL UNIT DIATHERMY

ELECTRO – SURGICAL UNIT ELECTRO – SURGICAL UNIT DIATHERMYDIATHERMY

ELECTRO – SURGICAL UNIT ELECTRO – SURGICAL UNIT DIATHERMYDIATHERMY

What is about the unit?What is about the unit?

It is a high frequency electric current It is a high frequency electric current producing machine generates producing machine generates

enough heat to cut, coagulates and enough heat to cut, coagulates and destroys body tissuedestroys body tissue

Page 2: ELECTRO – SURGICAL UNIT DIATHERMY

ELECTRO – SURGICAL UNIT DIATHERMY

OBJECTIVE: Describe the safe use of a monopolar

and bipolar Electro-surgical unit.Describe the effects of electro-

surgical unit.Describe the dangers of Electro-

surgical unit.

Page 3: ELECTRO – SURGICAL UNIT DIATHERMY

ELECTRO – SURGICAL UNIT DIATHERMY

The Machine consist of 4 important units:

Electrosurgical Generator

Active Electrode

Dispersive electrodes

Page 4: ELECTRO – SURGICAL UNIT DIATHERMY

ELECTRO – SURGICAL UNIT DIATHERMY

1. Electrosurgical Generator.

Different model have its own manufacturing instructions but basically, All generators will have the following:

Cable for power sources On/Off switch to flow power through active

electrode. A dial for selecting coagulation, cutting or blend

dials and to select the amount of current for cutting or coagulation mode.

A receptacle for the dispersive electrode cable or indifferent plate.

A receptacle for active electrode cable A receptacle for the foot control

Page 5: ELECTRO – SURGICAL UNIT DIATHERMY

ELECTRO – SURGICAL UNIT DIATHERMY

2. Active Electrode

Active electrodes are used by the surgeon to accomplish the desired function. They comes in various sizes and styles, may be useable or disposable and may be hand control or foot control.

3. Dispersive ElectrodesThese may be useable metal plates, lead plates, disposable aluminum foil or flexible rubber type of plates which can be moulded to the thigh or calf and secure with bandages and lastly the pregelled type dispersive plates

Page 6: ELECTRO – SURGICAL UNIT DIATHERMY

TYPE OF ELECTRO SURGICAL UNIT

1. Monopolar Unit

The inactive electrode is in the form of a patient grounding pad or plate which is placed in direct contact with the skin.

Current will flow from the active electrode through the body to the inactive electrode. The current then returns to the diathermy via the conductor cord to complete the circuit.

Page 7: ELECTRO – SURGICAL UNIT DIATHERMY
Page 8: ELECTRO – SURGICAL UNIT DIATHERMY

2.Bipolar diathermy• The active electrode is in the inner side of the

forcep and the inactive is in the others side.

• The current flows from the unit through the active side,arcs through the tissue to the inactive side and returns to the unit via the

same cord.• The current does not disperse through the

patient.

Page 9: ELECTRO – SURGICAL UNIT DIATHERMY
Page 10: ELECTRO – SURGICAL UNIT DIATHERMY

SUGGESTED SITES FOR PLATE PLACEMENT

• CALF • UPPER ARM• ABDOMEN• MID BACK• BUTTOCKS• ANTERIOR & POSTERIOR THIGH

Page 11: ELECTRO – SURGICAL UNIT DIATHERMY

Pad Site Burn

AORN Journal, March 1991

Page 12: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESU

Responsibilities before operating:

Generator:

1. Know which electro-surgical to be used and how to use it. Read the instruction manual

2. Only use equipment that is designed and approved for the unit.

3. Do not use an ESU with damaged and missing parts.4. Report lost items or items not working properly.5. Ensure alarm system is working.6. Do not use top of the units as a storage area.7. Do not use ESU if explosive anesthetic agent is in use.

Page 13: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESU

Responsibilities before operating:

Dispersive Electrode or Indifferent Plates

1. Do not use metal plates that is cracked, bent or corroded.

2. Do not cut, crease, or sharply bent a disposable plate.

3. Apply enough gel evenly and smoothly over the entire indifferent plate for the good conductive. Use the recommended gel.

Page 14: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESU

Responsibilities before operating:

Positioning of Dispersive Electrode:

1. Avoid body protuberances, skin folds, scar tissue, excessively fatty or hairy area, sites to be X-rayed or skin leison of any kind.

2. Avoid sites where fluids might pool3. Ensure the dispersive electrode is in good contact with

the patient and properly positioned.4. Ensure no skin surface in touching any metal parts of the

table and its accessories5. The ESU should be far away from the ESU electrodes.

Page 15: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESU

Responsibilities before operating:

Positioning of Dispersive Electrode (continue)

6. Ensure the grounding cable is firmly connected to the dispersive electrode.

7. Ensure flammable preparation solution have dried thoroughly.

8. Place foot pedal / hand control in position for use by the surgeon.

9. Turn the unit on and set the dials at the desired setting. (setting differ with each separate ESU, you must follow the manufacturer's instructions)

Page 16: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESUResponsibilities during operating:

The following precaution should be observed when using the ESU.

1. Position all cords so that they do not present a tripping hazard. Do not wheel equipment over the cords.

2. If the patient is repositioned, recheck the dispersive electrode for proper patient contact. Recheck the grounding cable connections.

3. If a second electrode is applied due to some fault, the first one must be removed.

4. Ensure the active electrode clean at all times during surgery. Escher build-up increases resistant, reduce performance and require higher settings.

5. Keep the active electrode separate from all conductors on the sterile field.

Page 17: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESUResponsibilities during operating:

The following precaution should be observed when using the ESU.

6. An unused active electrode should be placed in a protective sheath of pocket. It should never be placed where is can be activated accidentally and so burns the patient or medical personnel.

7. Do not coil the active or patient cable as this causes cracks in the cable, leading to leakages of current and so constitute a potential danger to the patient.

8. Do not activate the electrode-surgical unit for a long periods of time.

9. If the plate or foil dispersive electrode is used for a long procedure, check for gel dry-out. Re-gel is necessary.

Page 18: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESUResponsibilities after operating:

The following must be done:1. Disconnect the unit in the following orders:

- Turn the dials to 0- Turn the power switch off- Disconnect the cable from active electrode by grasping the plug not the cable.- Disconnect the cable from the dispersive/ground electrode by

grasping the plug.2. Remove the dispersive electrode.

- When removing metal plate lift the patient lightly and do not pull

on the cable to remove the plate.

Page 19: ELECTRO – SURGICAL UNIT DIATHERMY

The safe operation of An ESUResponsibilities after operating:

The following must be done:- When removing metal plate lift the patient lightly and do not pull

on the cable to remove the plate.- When using self-adhering dispersive electrode, gently support

the skin surface and peel the adhesive back slowly to remove it.- Check the dispersive electrode site for signs of the skin damage.

Page 20: ELECTRO – SURGICAL UNIT DIATHERMY

Cont….

- Document skin condition.- Wipe all reusable parts with a damp cloth.- Remove all gel thoroughly from the metal electrode and

wipe it dry.- Do not bend or kink electrical cords, coil them loosely.- Sterile the accessories with sterrad (hydrogen proxide),

if possible to prolong life. The active electrode can also be

steam sterile.