diathermy (1)

Upload: mindbook

Post on 09-Apr-2018

234 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/8/2019 Diathermy (1)

    1/43

  • 8/8/2019 Diathermy (1)

    2/43

    2

    DiathermyApplication of High-FrequencyElectromagnetic Energy

    Used To Generate Heat In Body TissuesHeat Produced By Resistance of TissuesAlso Used For Non-Thermal Effects

  • 8/8/2019 Diathermy (1)

    3/43

    3

    Physiologic Responses ToDiathermy

    Not Capable of Producing Depolarization and

    Contraction of Muscles Wavelengths Too Short

  • 8/8/2019 Diathermy (1)

    4/43

  • 8/8/2019 Diathermy (1)

    5/43

    5

    Diathermy HeatingDoses Are Not Precisely Controlled ThusThe Amount of Heating Cannot BeAccurately Measured Basically means amount of heating patient

    receives cannot be directly measured

    Heating= Current 2 X Resistance

  • 8/8/2019 Diathermy (1)

    6/43

    6

    Non-Thermal EffectsPulsed SWD Used To Treat Soft Tissue Injuries andWoundsRelated To Depolarization of Damaged Cells Loss of Cell Division Loss of Proliferation Loss of Regenerative capabilities

    Repolarization Corrects Cell DysfunctionGenerates A Magnetic Field To Increase Na PumpActivity

  • 8/8/2019 Diathermy (1)

    7/43

    7

    SWD Physiological Events SummaryElectromagnetic Radiation

    CSWD PSWD

    Electromag. Energy absorption

    Incr. molecular KE cell ion-binding properties, proteinsynthesis, ATP production

    Thermal Effect Athermal Effect

    Increased cell metabolism and function

    Enhanced soft-tissue healing

  • 8/8/2019 Diathermy (1)

    8/43

    8

    Shortwave DiathermyRadio Transmitter With FCC AssignedFrequencies 2 7. 12 MHz at 11 M 13 .5 6 MHz at 22 M 40 .68 MHz at 7. 5 M

  • 8/8/2019 Diathermy (1)

    9/43

    9

    Shortwave Diathermy UnitPower Supply Powers Radio FrequencyOscillator (RFO)RFO Provides Stable Drift-Free Oscillationsat Given FrequencyPower Amplifier Generate Power To DriveElectrodesOutput Resonant Tank Tunes In The Patientfor Maximum Power Transfer

  • 8/8/2019 Diathermy (1)

    10/43

    10

    Possible Shortwave Diathermy Unit

    A=Power SwitchB=Timer

    C=Power Meter(monitors current from power supply notcurrent entering patient-volume control)D=Output Intensity(%max power to patient)E=Tuning Control(tunes output from RFO)

  • 8/8/2019 Diathermy (1)

    11/43

    11

    Shortwave Diathermy UnitPower Output Should Provide Energy ToRaise Tissue Temp To Therapeutic Range

    (40 -45 deg C) ( 8 0 -120 watts)Should Exceed SAR-Specific AbsorptionRate (rate of energy absorbed /unit area of tissue mass)

  • 8/8/2019 Diathermy (1)

    12/43

    12

    Adjusting Resonance of SWD Unit(Tuning)

    Manual vs Automatic Tuning When patients circuit (biologic tissue) oscillates at same

    frequency as device frequency Only when tuned will is the electromagnetic energy fully delivered Most automatic Can lose tuning due to movement at skin/electrode interface

    Manual Tuning (adjusts patient circuit) Set Output Intensity at 30 -40 % Adjust Tuning Control Until Power Output Meter Reaches Max Then Adjust Down to Patient Tolerance Which Is About 50 % If More Than 50 % Patient Is Out of Resonance

  • 8/8/2019 Diathermy (1)

    13/43

    13

    Shortwave Diathermy UnitGenerates Both an Electrical and aMagnetic FieldRatio Depends on Characteristics of BothThe Generator and the Electrodes SWD Units at 1 3 .5 6 MHz= Stronger Magnetic

    Field SWD Units at 2 7. 12 MHz = Stronger Electrical

    Field

  • 8/8/2019 Diathermy (1)

    14/43

    14

    SWD ElectrodesCapacitor ElectrodesInductor ElectrodesSelection of Appropriate Electrodes CanInfluence The Treatment

  • 8/8/2019 Diathermy (1)

    15/43

    15

    Capacitor (Condenser) ElectrodesCreate Stronger Electrical Field Than MagneticField

    Ions Will Be Attracted Or Repelled Depending onthe Charge of the Pole

  • 8/8/2019 Diathermy (1)

    16/43

    16

    Capacitor Electrodes

    Electrical Field Is The Linesof Force Exerted onCharged Ions That Cause

    Movement From One PoleTo Another Center Has Higher CurrentDensity Than Periphery

  • 8/8/2019 Diathermy (1)

    17/43

    17

    Capacitor Electrodes

    Patient Is BetweenElectrodes and BecomesPart of CircuitTissue Is BetweenElectrodes in a SeriesCircuit Arrangement

  • 8/8/2019 Diathermy (1)

    18/43

    18

    Electrical FieldThe Tissue That Offers The GreatestResistance To Current Flow Develops TheMost Heat Fat Tissue Resists Current Flow Thus Fat Is Heated In An Electrical Field

    Precaution: electrical field may overheat areawith large fat content

  • 8/8/2019 Diathermy (1)

    19/43

    19

    Capacitor Electrodes(Air Space Plates)

    Two Metal Plates SurroundedBy Plastic GuardCan Be Moved 3 cm WithinGuardProduce High-Frequency

    Oscillating Current

  • 8/8/2019 Diathermy (1)

    20/43

    20

    Air Space Plate Electrodes

    Area To Be Treated Is

    Placed BetweenElectrodes BecomingPart of Circuit

  • 8/8/2019 Diathermy (1)

    21/43

    21

    Air Space Place Electrodes

    Sensation Of Heat In Direct Proportion ToDistance Of Electrode From SkinCloser Plate Generates More Surface HeatParts Of Body Low In Subcutaneous Fat BestTreated

  • 8/8/2019 Diathermy (1)

    22/43

    22

    Capacitor Electrodes(Pad Electrodes)

    Greater Electrical Field

    Patient Part of CircuitMust Have UniformContact (toweling)Spacing Equal To Cross-sectional Diameter of PadsPart To Be TreatedShould Be Centered

  • 8/8/2019 Diathermy (1)

    23/43

    23

    Pad ElectrodesIncreasing TheSpacing Will Increase

    The Depth Of Penetration But WillDecrease The CurrentDensityCapacitive MethodGood for TreatingSuperficial SoftTissues

  • 8/8/2019 Diathermy (1)

    24/43

    24

    Induction MethodCreates A Stronger Magnetic Field ThanElectrical FieldA Cable Or Coil Is WrappedCircumferentially Around An Extremity Or Coiled Within n Electrode

  • 8/8/2019 Diathermy (1)

    25/43

    25

    Induction Electrodes

    Passing Current Through A Coiled Cable Creates

    A Magnetic Field By Inducing Eddy Currents(small circular electrical fields) That GenerateHeat

  • 8/8/2019 Diathermy (1)

    26/43

    2 6

    Induction ElectrodesPatient In A Magnetic Field Not Part Of A CircuitGreatest Current Flow Through Tissue With Least

    Resistance Tissues react like a parallel circuit Fat does not provide as much resistance to

    electromagnetic energy

    Tissue High In Electrolytic Content Respond BestTo A Magnetic Field Muscle and blood

    Precaution: Not as much superficial sensation of heat with electromagnetic

  • 8/8/2019 Diathermy (1)

    27/43

    2 7

    Induction Electrodes(Cable Electrode)

    Two Arrangements: Pancake Coils

    Wraparound CoilsToweling Is EssentialPancake Coil MustHave 6 in Center Then5 -10 cm SpacingBetween TurnsBest Frequency

  • 8/8/2019 Diathermy (1)

    28/43

    2 8

    Induction Electrodes(Drum Electrode)

    One Or MoreMonopolar Coils

    Rigidly Fixed In AHousing UnitMay Use More ThanOne Drum DependingOn Area TreatedPenetration Deeper Soft Tissues

    Toweling Important

  • 8/8/2019 Diathermy (1)

    29/43

    2 9

    Heating With Continuous SWDPatient Sensation Provides Basis For Recommendations Of Continuous SWD

    Dose I (Lowest) (< 3 8 W) - No Sensation of HeatDose II (Low) (~ 8 0 W)- Mild Heating SensationDose III (Medium) ( 8 0 -300 W) - Moderate or Pleasant Heating SensationDose IV (Heavy) (> 300 W) -Vigorous HeatingWithin Pain Threshold

  • 8/8/2019 Diathermy (1)

    30/43

    30

    Pulsed SWDReferred To By Different Names Pulsed Electromagnetic Energy (PEME) Pulsed Electromagnetic Field (PEMF) Pulsed Electromagnetic Energy Treatment

    (PEMET)

  • 8/8/2019 Diathermy (1)

    31/43

    3 1

    Pulsed SWD

  • 8/8/2019 Diathermy (1)

    32/43

    32

    PSWD Mean Power CalculationsPulse Period (Pulse on + off time) = Peak Pulse Power (W) / Pulse Rep Freq (Hz)

    Percentage On Time = Pulse Duration (msec) / Pulse Period (msec)

    Mean Power = Peak Pulse Power (W) / Percentage on Time

  • 8/8/2019 Diathermy (1)

    33/43

    33

    ExamplePeak Pulse Power = 8 00 WPulse Duration = .4 msPulse Frequency = 200 HzPulse Period = 8 00 W / 200 Hz = 4 ms% on time = .4 / 4 = .10 or 1 0 %

    Mean Power = 1 0 % of 8 00 = 8 0 W End Result = thermal effect < 3 8 W = no heating

  • 8/8/2019 Diathermy (1)

    34/43

    34

    Pulsed SWDInterrupted Output Delivered In Series Of High-Frequency Bursts ( 40 -400 usec)Pulse Rate Selected With Pulse FrequencyControl (11 000 Hz)Off-Time Longer Than On-TimeLow Mean Power OutputUses Drum Electrode

  • 8/8/2019 Diathermy (1)

    35/43

    35

    Shortwave Diathermy vs .Ultrasound

    Pulsed SWD Produces The Same Magnitude And DepthOf Muscle Heating as 1MHz Ultrasound (Draper, JAT199 7 )

  • 8/8/2019 Diathermy (1)

    36/43

    3 6

    Treatment TimeMost Typically SWD Treatments Last For

    20 -30 MinutesRemember As Skin Temperature RisesResistance Falls

  • 8/8/2019 Diathermy (1)

    37/43

    3 7

    When Should Diathermy

    Be Used?If The Skin Or Some Underlying Soft TissueIs Tender And Will Not Tolerate PressureIn Areas Where Subcutaneous Fat Is Thick And Deep Heating Is Required

    Induction methodWhen The Treatment Goal Is To IncreaseTissue Temperatures Over A Large Area

    If The Skin Or Some Underlying Soft TissueIs Tender And Will Not Tolerate PressureIn Areas Where Subcutaneous Fat Is Thick And Deep Heating Is Required

    Induction methodWhen The Treatment Goal Is To IncreaseTissue Temperatures Over A Large Area

  • 8/8/2019 Diathermy (1)

    38/43

    3 8

    SWD GuidelinesIndications Continuous and Pulsed

    PrecautionsContraindications

  • 8/8/2019 Diathermy (1)

    39/43

    3 9

    Microwave DiathermyTwo FCC Assigned Frequencies- 245 6 MHz and

    9 15 MHzMWD Has Higher Frequency and Shorter Wavelength Than SWDGenerates Strong Electrical Field and RelativelyLittle Magnetic Field

    Advantage: better focus wave on body, therebymore local heating affectsDisadvantage: Depth Of Penetration Is Minimal In

    Areas With Subcutaneous Fat > 1 cm

  • 8/8/2019 Diathermy (1)

    40/43

    40

    Microwave Diathermy Unit

    A= Power SwitchB=Timer

    C=Output Meter (indicates relative output in wattsD= Power Output Level KnobE= Amber Light-Warming up / Red Light- Ready

  • 8/8/2019 Diathermy (1)

    41/43

    4 1

    MWD Applicators (Electrodes)

    Circular Shaped Applicators 4 or 6

    Maximum Temperature At PeripheryRectangular Shaped Applicators 4 .5 x 5 or 5 x 2 1

    Maximum Temperature At Center

  • 8/8/2019 Diathermy (1)

    42/43

    42

    Microwave Applicator Set-UpMicrowave Applicator Beams Energy ToPatient

    Must Pay Attention To Cosine LawIn 9 15 MHz Units Applicators Placed 1 cmFrom Skin

    245 6 MHz Units Have Manufacturer Recommended Distances and Power Outputs(Uses Antenna)

  • 8/8/2019 Diathermy (1)

    43/43

    43

    Best Treated areas for Microwave

    Tendons of foot, hand and wristAC and SC jointsPatellar tendonDistal tendons of hamstringsAchilles tendonOther areas of low subcutaneous fat