edax on the xl-30 tutorial #3 introduction to using the x-ray analysis program

29
EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Upload: kathryn-turner

Post on 17-Jan-2016

223 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

EDAX on the XL-30

Tutorial #3Introduction to Using the X-ray

Analysis Program

Page 2: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Remember Purpose• Purpose of these tutorials is to enable the

user to understand the basics of OPERATING the XL-30 to obtain Secondary Electron and Back Scattered Electron Images, and to obtain an Energy Dispersive X-Ray Spectrum.

• Very little theory is given.• If the user’s requirements demand further

instruction, he or she should seek further formal instruction.

Page 3: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Assumptions

• This tutorial assumes that you have reviewed the previous tutorials;

• That you are familiar with and competent to use the XL-30 SEM;

• You have a sample loaded in the SEM chamber;

• You have the SEM operational.

Page 4: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Working Distance for EDS

• The optimum analytical working distance is 10 mm. This means that you sample should be in focus at a 10 mm working distance.

Page 5: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Start the EDS (DX) program

• Always start the XL-30 microscope control program before the EDS program

Page 6: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

eDX Opening Screen

Page 7: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Collect a spectrum

• Place the beam where you want to collect of spectrum. – If it is from a spot you will need to use spot mode

in the Microscope Control.• I find it easier to control the beam using the XL-30

microscope control program than to try to control the beam from the eDX program.

– If you want to scan an area to collect a spectrum, scan on the area you wish to collect from.

Page 8: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Start the Collection

Page 9: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Stop the Collection

Page 10: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Identifying the Peaks

Page 11: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Quantify Results

Page 12: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Menus and Buttons

• The next slides will give a quick overview of the menus and buttons on the graphic interface.

• REMEMBER – IF YOU NEED MORE INFORMATION START WITH THE HELP FILES.

Page 13: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

File Menu

Page 14: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Edit Menu

Page 15: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

View Menu No. 1

Page 16: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

View Menu No. 2

Page 17: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Qual Menu

Page 18: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Quant Menu

Page 19: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Proc Menu

Page 20: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Auto Menu

Page 21: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Setup Menu

Page 22: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Options, Window, and Help Menus

• There is only one option – SEM• Windows menu is self explanetory.• Help Menu – see this for more information.

Page 23: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Window Menu

Page 24: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Help Menu

Page 25: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Icons

Page 26: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Peak ID Page

Page 27: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Quantify Page

Page 28: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

Microscope Control Page

Page 29: EDAX on the XL-30 Tutorial #3 Introduction to Using the X-ray Analysis Program

That’s all

• You are less likely to use the other pages, but feel free to explore.

• To start, I suggest that you analyze some samples that you know the composition. That way you can tell if you have a problemor not.