2
GA32 Model Operating Temperature Range Liquid Sample Flux 40 200 up to 28 /min. Spraying Nozzle for Liquid and Air Water Evaporation Max.1300 / h Dryer, Spray "Pulvis Mini-Spray" A small sized dryer for laboratory use which produces uni- formalized fine particles. The GA32 is a compact spray dryer using a spray dry method which enables sim- ple use for laboratory experiments. It can be used for a wide variety of drying operations, from preliminary examinations for pilot plants to drying operation in general laboratories. No danger of damaging foodstuffs, medical products and products that are sensitive to biochemical-heat. No oxidation takes place and the water content of the formed powders is low, so that there is no contamination. Even a solid body of ca.0.5g can be dried. Easy operation and maintenance GA32 Control Panel Inlet temperature, outlet tem- perature and drying air volume are displayed digitally. Condition setting dials and stage level control switch are located on the front panel for easy use. In this Spray Dryer, the sample are dried and transformed into a uniform fine powder of amorphous form, so both its solubility and reactivity are extremely high in comparison to crystal. Drying chamber, cyclone, product vessel can be easily detached and cleaned. Inlet temperature, outlet tempera- ture and drying air volume are dis- played digitally. Condition setting dials and stage level control switch are located on the front panel for easy use. Obtained powder (example)

Upload: hathien

Post on 04-May-2019

221 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dryer, Spray Pulvis Mini-Spray - dongindis.co.krdongindis.co.kr/web_data/down/GA32.pdf · Model GA32 Operating Temperature Range Liquid Sample Flux 40℃~200℃ up to 28握/min

GA32Model

Operating Temperature Range

Liquid Sample Flux

40℃~200℃

up to 28握/min.

Spraying Nozzle

for Liquid and Air

Water Evaporation

Max.1300握/h

Dryer, Spray "Pulvis Mini-Spray"

A small sized dryer for laboratory use which produces uni-formalized fine particles.The GA32 is a compact spray dryer using a spray dry method which enables sim-ple use for laboratory experiments. It can be used for a wide variety of dryingoperations, from preliminary examinations for pilot plants to drying operation ingeneral laboratories.

No danger of damaging foodstuffs, medical products and products that aresensitive to biochemical-heat.

No oxidation takes place and the water content of the formed powders is low,so that there is no contamination.

Even a solid body of ca.0.5g can be dried.

Easy operation and maintenance

GA32

Control Panel

●Inlet temperature, outlet tem- perature and drying air volume are displayed digitally. Condition setting dials and stage level control switch are located on the front panel for easy use.

●In this Spray Dryer, the sample are dried and transformed into a uniform fine powder of amorphous form, so both its solubility and reactivity are extremely high in comparison to crystal.

●Drying chamber, cyclone, product vessel can be easily detached and cleaned.

●Inlet temperature, outlet tempera- ture and drying air volume are dis- played digitally. Condition setting dials and stage level control switch are located on the front panel for easy use.

●Obtained powder (example)

Page 2: Dryer, Spray Pulvis Mini-Spray - dongindis.co.krdongindis.co.kr/web_data/down/GA32.pdf · Model GA32 Operating Temperature Range Liquid Sample Flux 40℃~200℃ up to 28握/min

SINCE 1889

YAMATO SCIENTIFIC CO.,LTD.2-1-6, Nihonbashi Honcho, Chuo-ku, Tokyo 103-8432 Japan

Telephone : 81-3-3231-1124Facsimile : 81-3-3231-1144Website : http : // www.yamato-net.co.jp/englishEmail : [email protected]

Structure

Standard

GB22 (Basic Unit)Model

HeaterAspiratorSample liquid feed pumpStirring systemCompressed air blowing systemTemperature control rangeTemperature control accuracyTemperature setting and display methodDry air flow meterSpray pressure gauge

2 kWBypass-type blowerQuantitative peristaltic pump, flow rate variable up to 28握/min.Using an induction motorPulse jet system, an electromagnetic valve and electronic timer40℃~200℃±1℃DigitalDigital displayBurdon tube pressure gauge, measurement range : 0~294kPa (0~3kg/cm2)

Power source (50/60Hz)

External dimensions(W×D×Hmm)Weight

AC 200/230V, single phase 14A/13A760×420×1,350Approx. 110kg

Structure and StandardGF32 (Pulvis Mini Bed Attachment)Model

Water evaporation rateSpray nozzleDrying chamberCycloneProduct vesselNozzle orifice cleaningWeight

Max. approx. 1,300握/HTwo-liquid nozzle, 1ASuper-hard glassSuper-hard glassSuper-hard glassPulse jet system (using a compressed air blowing system of Model GB22)Approx. 11kg

Specifications

Dryer, Spray "Pulvis Mini-Spray"

① ②

③④

Exhaustair

Air supply

Pressure air

Sample

System diagram

① Fluid Pump② Spray nozzle③ Needle valve④ Electromagnetic valve⑤ Heater

⑥ Distributor⑦ Drying chamber⑧ Cyclone⑨ Product vessel⑩ Aspirator

⑪ Cap⑫ Inlet temperature sensor⑬ Outlet temperature sensor

A

Liquid nozzle

Air nozzle

B

C

Spraying nozzle

Nozzle No.

Nozzle size (μm)

Model 1A : Standard nozzle

Model

Nozzle No.

1A

4061270

1626

5081270

1626

508 1270

1778

711 1270

1778

7111650 2050 2050 2850 2850

1270

162664 64 70 70 64

1

A侃 B

C粥

2A 2 3