colorimetry introduction.docx

Upload: nikky-namdeo

Post on 04-Jun-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/14/2019 Colorimetry Introduction.docx

    1/7

    Introduction to Colorimetry

    A spectrophotometeris an instrument that passes a selected wavelength of

    electromagnetic radiation through a solution and measures the amount of that

    electromagnetic radiation that the solution absorbs (does not allow to pass) or

    transmits (does allow to pass). Refer for a moment to theelectromagnetic

    spectrumshown on the linked page, (from Life, The Science of Biology, by Purves

    et al., 5thed., 1998). One type of spectrophotometer uses ultraviolet (UV)

    wavelengths; those are the wavelengths that cause sunburn, for example. Another

    type uses infrared (IR) wavelengths, which are most familiar as heat. You mayhave an opportunity to use one of these in another course. The third type of

    spectrophotometer, which we will use in this course, uses visible wavelengths,

    which we think of as the different colors of light, ranging from violet to red. Since

    these are the familiar colors, this type of spectrophotometer is called a

    "colorimeter." The particular model we will use is the Spectronic 20 colorimeter.

    Observation of the visible spectrum inside the

    Spectronic 20 colorimeter

    1. Turn the power switchclockwise to turn on the instrument; you'll hear/feel the

    on-click. The pilot lampshould glow, and you may hear the gentle sound of the

    cooling fan inside. If the pilot lamp doesn't glow on one of these instruments, don't

    be concerned. For the moment you may ignore the other uses of the two knobs on

    the front of the instrument.

    2. Refer to the illustration of the instrument below. Each solution that you will test

    is placed into a small test tube, called a colorimeter tube, which is inserted into

    http://pirate.shu.edu/~rawncarr/colorimetry/spectrum.gifhttp://pirate.shu.edu/~rawncarr/colorimetry/spectrum.gifhttp://pirate.shu.edu/~rawncarr/colorimetry/spectrum.gifhttp://pirate.shu.edu/~rawncarr/colorimetry/spectrum.gifhttp://pirate.shu.edu/~rawncarr/colorimetry/spectrum.gifhttp://pirate.shu.edu/~rawncarr/colorimetry/spectrum.gif
  • 8/14/2019 Colorimetry Introduction.docx

    2/7

    thesample holder.The beam of light, of whatever wavelength is selected, passes

    through the tube in the holder from right to left.

    3. In your test tube rack is a colorimeter tube containing a rectangular strip of

    white paper that is a bit shorter than the length of the tube and just wide enough to

    fit snugly into the colorimeter tube. Insert this tube into the holder (pushing it all

    the way down); rotate the tube so that the faint light beam entering the tube from

    the right will strike the paper's flat surface.

    4. Cup your hand around the top of the open sample holder and look closely down

    into the tube to see the faint light reflected off the paper. You will need to haveyour eye close to your cupped hand (near the top of the tube) because the light is

    faint. If the room light is too bright the instructor will turn off the room lights for a

    moment.

    5. As you turn the wavelength control knob, you will see the light reflected by the

    paper change from violet/blue (about 400 nm) to green to yellow to orange and

    finally to red (about 700 nm). The lamp inside the instrument produces white light

    (a mixture of all wavelengths), but the wavelength control knob screens all but the

    one you select. In normal operation (i.e. without the paper in the colorimeter tube)

    the light that you see reflected would enter the tube containing a test solution and

    some portion of it would pass out the other (left) side of the tube to strike a

    detector, which would convert that light into an electrical signal that registers on

    thetransmittance and absorbance scales.

    6. Remove the tube from the sample chamber and replace it in the test tube rack.

    7. In normal operation the lid of the sample holder is always closed after inserting

    the sample tube, before recording measurements, to shut out stray light.

  • 8/14/2019 Colorimetry Introduction.docx

    3/7

    Standardization of the colorimeter for determining the

    absorption spectrum of riboflavin, the vitamin

    As light passes through a solution in a colorimeter tube, some small fraction of that

    light may be absorbed by the glass of the tube and by the solvent in which the

    molecule of interest is dissolved. Further, as you will see in your next lab exercise

    (the biuret test for measuring protein in solutions), the solution in the tube may

    contain other substances that are necessary to produce the color that you measure.

    Since many types of molecules are not colored, they must be reacted with

    something that will produce a color before you can use the colorimeter to measure

    them. Anything that the light passes through may absorb some of the light and

  • 8/14/2019 Colorimetry Introduction.docx

    4/7

    influence measurements. Since the point is to measure absorption of one, particular

    type of molecule in a solution, it is necessary to correct for (to "subtract") light

    absorption by everything else (the glass of the test tube, the solvent, any color-

    producing reagents present). To do that you standardize the instrumentbefore

    doing your measurements.

    Riboflavin is a water soluble vitamin. The yellow solution in one of the tubes in

    your rack contains riboflavin dissolved in water, 0.017 mg/mL; nothing else is

    present. In order to get the most accurate possible data for determining the

    absorption spectrum of riboflavin, you must standardize the colorimeter to correct

    for light absorption by the glass and the solvent (water in this case). Therefore, you

    have another tube containing only water, that is, everything except the molecule of

    interest, riboflavin; this second tube is called the blank. You use the blank to

    standardize the instrument. In effect you will be "telling the instrument to ignore"

    light absorption by everything except the molecule of interest, riboflavin.

    The steps in standardization are few and simple, but theymust be done correctly and carefully to obtain reliable data.

    1. If the instrument were not already turned on, you would use the power switchto

    turn on the instrument and let it warm up for about 15 minutes.

    2. Use the wavelength controlknob to select the desired wavelength; the

    wavelength values shown in the wavelength dialdisplay window are graduated in

    nanometers (nm). NOTE WELL that in next week's work you will use only one

    wavelength; today, though, you will measure light absorption at many wavelengths

    in order to determine the absorption spectrum of riboflavin. So, we'll begin with

    340 nm for the wavelength. Set that now.

  • 8/14/2019 Colorimetry Introduction.docx

    5/7

    3. With the sample holder empty and its lid closed, use the "0" controlknob

    (same as the power control knob) to set the indicator needle to zero percent on the

    transmittance scale, 0 %T. Without a tube in the chamber, no light enters the

    chamber (it's pitch black in there), so no light would strike the detector. In effect,

    you're telling the instrument that if there were a tube with a solution in the

    chamber, this is what you'd see if absolutely none of the light entering the tube

    passed through to the other side, i.e. 0 % of the light got through.

    4. Insert the blank tube into the holder and close the lid. The needle will sweep

    upward on the %T scale. Wherever the needle stops, use the light controlknob

    now to move the needle to 100 %T on the scale. In effect, you're telling the

    instrument that with a sample in the chamber, but no riboflavin present, this is

    what you'd see if all of the light (100 % of it) entering the tube passed through to

    the other side. Since the glass and the solvent will also be present with your

    riboflavin sample, this step tells the instrument to ignore light absorption by those

    factors. Now, if you see any %T values less than 100%, that must be due to light

    absorption by the riboflavin; with riboflavin present, less than 100% of the light

    entering the tube will pass through.

    5. Remove the blank from the sample holder and close the lid. The needle should

    fall back to zero %T.

    6. The instrument is now standardized for 340 nm wavelength. If you were going

    to do measurements on many samples at this single wavelength, you would nottouch the light controlknob or the "0" controlknob hereafter.

  • 8/14/2019 Colorimetry Introduction.docx

    6/7

    Determining riboflavin's absorption spectrum

    An absorption spectrum plots absorbance(A), not % transmittance, on the y-axis,

    as a function of wavelength on the x-axis. Absorbance is the other scale on your

    colorimeter. Note that absorbance and %T scales run in opposite directions; than

    means that they are reciprocally related; as one rises, the other falls. But note also

    the spacings between divisions on the scales; %T has a linear scale whereas

    absorbance has a logarithmic scale (common log, i.e. base 10 log). Specifically, A

    = log (1 / T), where T is the decimal equivalent of %T. Example: if %T = 50%,

    then T = 0.5 and (1 / T) = 2.0, and A = 0.301. Make sure you understand this

    algebraic relationship.

    To collect data for riboflavin's absorption spectrum now, you must record

    absorbance values at many wavelengths, not just one, across the visible spectrum.

    It will be sufficient to take measurements at 5 nm increments. Each time the

    wavelength is changed, however, the 100 %T setting must be redone because the

    glass and solvent may not absorb light to the same degree at different wavelengths.

    But this extra step takes little time. So, proceed as follows.

    1. With the colorimeter already standardized for 340 nm wavelength, insert your

    riboflavin tube, close the chamber lid, and record the absorbance reading. Don't

    change any knob settings yet. Then remove the riboflavin tube and close the lid.

    The needle will fall back to 0 %T, which is A = "infinity" (the maximum

    absorbance possible). You set the 0%T only once.

    2. Reset the wavelength to 5 nm higher than the last setting. Insert the blank tube,

    close the lid, and note that the needle doesn't go all the way back to 100 %T. That's

    because the tube's glass and the water absorb different amounts of light at different

  • 8/14/2019 Colorimetry Introduction.docx

    7/7

    wavelengths. So, reset the needle to 100 %T with the light controlknob now. The

    machine is now standardized for the new wavelength.

    3. Remove the blank tube, and see that the needle falls back to 0 %T (= maximum

    absorbance). Insert the riboflavin tube into the sample holder, close the lid, and

    record absorbance at the new wavelength. Don't adjust any knobs with the

    riboflavin sample in the chamber. Then remove the riboflavin tube and close the

    lid.

    4. Now repeat steps #2 and #3: increase the wavelength in 5 nm steps, reset the 100

    %T with the blank each time, and record a new absorbance value for theriboflavin solution. Your last wavelength setting will be 530 nm. This process

    goes quickly, but it must be done carefully.

    5. Return both the blank tube and the riboflavin tube to the test tube rack. Never

    leave a sample in the instrument. Turn off the instrument by turning the power

    switchcounter clockwise; you'll hear/feel the click.

    6. Plot your data on graph paper that has both axes with linear scales: absorbance

    on the y-axis and wavelength on the x-axis. What is riboflavin's absorption

    maximum wavelength? How many "peaks" does the absorption spectrum have?

    Check your results with your lab instructor.