    2

    3

    1Service modules1.1 Service spine Service panels and electrical modules Add-on parts Service modules Suspended service booms 1.2 Service wing

    1.3 Service distribution terminal

    Work benches2.1 Add-on tables and supporting structures 2.2 Worktops

    2.3 Mobile table 2.4 Swing 2.5 Height adjustable table 2.6 Balance table 2.7 VarioTHEK 2.8 Rack 2.9 Analysis tables Assistant Round table

    Add-on plate Oval Skate Caddy 2.10 Sink work benches Laboratory sinks

    Sink modules Drip cups 2.11 Mobile sinks Sink station AquaEl

    Fume cupboards and exhaust devices3.1 Bench mounted fume cupboards 3.2 Low ceiling bench mounted fume

    cupboards 3.3 Low level fume cupboards 3.4 Walk-in fume cupboards

    3.5 Secuflow bench mounted fume cupboards 3.6 Secuflow low volume

    bench mounted fume cupboards 3.7 Bench mounted fume cupboards with

    side installation 3.8 Low ceiling bench mounted fume cupboards with sideinstallation 3.9 Special application fume cupboards 3.10 Filter fume cupboards

    3.11 Radio-isotope fume cupboards 3.12 Mobile fume cupboards 3.13 Local

    exhaust devices 3.14 Monitoring and control

    4Storage space4.1 Underbench units Underbench unit on plinth Underbench unit on castors Suspended underbench unit Underbench unit for fume cupboards Underbench

    unit for fume cupboards for the storage of solvents Underbench unit for fume

    cupboards for the storage of acids and alkalis 4.2 Disposal Underbench unit for waste

    disposal Waste disposal system for acids and alkalis Waste disposal system for flammable

    liquids 4.3 Wall mounted cabinets 4.4 Instrument cupboards 4.5 Top-mounted

    cabinets 4.6 Exhaust cabinets Pull-out cabinet Emergency cabinet Acid and alkalicabinet Gas cylinder cabinet Solvents cabinet Supply system for flammable liquids

    5Teaching laboratories5.1 Teacher desk 5.2 Students desk with service pillar/service module 5.3 Students desk with service module 5.4 Mobile table

    6General

    6.1 Patents, brands, design patents and registered designs 6.2 Certifiedsafety and quality 6.3 Design and colours 6.4 From planning to the finished

    laboratory 6.5 Standards and technical rules

    Page 3 21

    Page 23 45

    Page 47 87

    Page 89 115

    Page 117 127

    Page 129 136

  • 8/13/2019 Catalog Lab

    4/140

  • 8/13/2019 Catalog Lab

    5/140

    3

    Service modules1.1 Service spine

    Service panels and electrical modules

    Add-on partsService modules

    Suspended service booms

    1.2 Service wing

    1.3 Service distribution terminal

    The philosophy of our mc6fur-

    niture system is to enable

    changes to be made easily andtherefore to provide more

    mobility in the laboratory.

    Energy and service systems have

    a central role.

    Pipes and cables for supplies

    and waste are run to the service

    spine, service module, sus-

    pended service boom and the

    service wing. These units then

    provide the services at fittings

    and integrated electrical units.

    Due to the modular layout andflexible installation technology,

    services and energy are always

    available where they are needed.

    1

  • 8/13/2019 Catalog Lab

    6/140

    4

    Service modules

    General

    The special feature of the serv-

    ice spine is its modular layout

    and its logical separation from

    the furniture in the laboratory.

    The service spine gives the

    designer freedom of choice

    during the design of the labo-

    ratory environment and pro-

    vides the best possible options

    for different designs and rapid

    changes. The backbone of thissystem is formed by the strong

    pillar profile without grooves,

    which can be combined

    together with the freely selec-

    table bench frames to form a

    work bench against a wall or a

    double-sided work bench.

    The strong profile frame is per-

    fectly designed for fitting dif-

    ferent services and electrical

    modules, reagent shelves,

    instrument shelves and muchmore.

    The surfaces of the supporting

    structure and the panels are

    protected with high-grade

    powder coatings. As the items

    fitted to the panel can be

    changed rapidly without the

    need to undo bolts, laboratory

    operations are hardly inter-

    rupted at all.

    The supply pipes, for example

    for water and compressed air,

    can be rapidly extended and

    fitted using a quick release

    coupling system. Overbench

    cabinets can also be fastened

    when pillar extensions are

    used.

    For the perfect planning of the

    preparatory work on-site, weprovide planners and architects

    with installation plans contain-

    ing exact sizes for supply pipes

    and cables as well as informa-

    tion on dimensioning and flow

    rates.

    Flexibility and rapid upgrad-

    ability

    The modular design principle

    enables all modules to be rap-

    idly replaced or changed for

    expansion. This feature signi-fies low installation costs.

    Shelves can be fitted as

    required independent of modu-

    lar size. Ready to plug-in elec-

    trical modules are installed in

    no time.

    Full independence from the

    work bench

    As a result of the separation of

    the service spine and furniture,

    worktops and underbench

    units can be moved or

    changed without dismantling

    the mechanical services and

    electrical installation.

    Two service spines used fordouble-sided work benches

    Here the double spine com-

    prises two self-contained spines

    that are independent of each

    other. When making changes,

    double bench spines can be

    converted into two individual

    spines set up against the wall.

    Service spine

  • 8/13/2019 Catalog Lab

    7/140

    1

    5

    Use of electrical modules

    that can be positioned as

    required

    Our modules only need space

    where the power is required.

    There is no awkward electrical

    trunking with the service

    spines. You can use the full

    depth of the bench.

    Wear-free cold/hot water

    fittings

    Unlike upper sections with a

    grease chamber, our fittings

    with ceramic sealing in the

    upper section are wear-free

    and maintenance-free for sig-

    nificantly increased economy.

    The basic structure of the

    service spine based on the

    example of a double work

    bench with the various

    elements fitted

  • 8/13/2019 Catalog Lab

    8/140

    6

    Service modules

    General

    Service wing

    Since the introduction of our

    laboratory furniture system

    mc6 , the term freedom in

    the laboratory has been com-

    pletely re-defined: Maximum

    flexibility of all systems and

    maximum freedom of move-

    ment without interface prob-

    lems provide people working in

    laboratories freedom to be able

    to implement their ideas.

    The revolutionary service wing

    supplements the new freedom

    in the laboratory in a special

    manner: as the central furni-

    ture element in which all key

    services such as mechanical,

    electrical, lighting, EDP, extract

    air and of course the disposal

    of waste water are integrated,

    our system provides a very

    high-degree of flexibility.

    The possibility of being able toplug-in to the service wing,

    practically anywhere, using

    quick-release couplings signifies

    a very high level of freedom of

    movement in the laboratory.

    Using the service wing simpli-

    fies the entire laboratory fitting

    out process including the

    mechanical and electrical serv-

    ices. A central feed point is

    sufficient and the service wing

    securely mounted to the ceil-

    ing with adjustable steel bars

    makes further installations,

    holes in the walls and visible

    suspension systems superflu-

    ous. Installation costs are

    reduced significantly, exten-

    sions can be made without

    problems without the need to

    make structural changes.

    With the service wing we have

    created a service supply area

    outside the direct working

    area. All services can be rapidly

    fed to the work bench without

    complications.

    Extraordinary flexibility in

    the laboratory

    All areas of the laboratory are

    supplied using cross and T ele-

    ments. All services are pre-

    installed in the wing. In this

    way we achieve a large num-

    ber of possible configurations.

    It is possible to dock any-

    where at any time. Standard-

    ised units for analysis tables

    and work benches, mobile

    tables, racks, mobile sink units

    or mobile fume cupboards

    everything can be used as

    required under the wing.

    Simplification of the prelimi-nary structural planning

    Complex installations for differ-

    ent systems and their planning

    are no longer required when

    the service wing is used. The

    saving in room height due to

    the lack of need for a sus-

    pended ceiling for improved

    appearance signifies a consid-

    erable saving in planning and

    fitting out costs.

    Easy to reach connection

    quick-release couplingsWhen standing, the connec-

    tions on the service wing

    can be reached easily without

    problems.

    The service wing is also

    easy to integrate in difficult

    circumstances

    The existing room architecture

    and the characteristics of the

    ceiling, wall and floor often

    make extensive installations

    more difficult. On listed build-ings it is not even allowed to

    make major changes to the his-

    toric structure. Here the service

    wing can be easily integrated

    with minimal effort without

    complications.

    Perfectly planned, pre-

    assembled and equipped in

    the factory

    The service wing system for

    your laboratory is fully pre-assembled in our factory to suit

    your needs. You save assembly

    time on-site and your service

    wing is rapidly installed and

    ready to use.

    Uncomplicated to extend

    and maintain

    Due to its independent design

    as a self-contained unit, the

    service wing is almost mainte-

    nance-free. Extending, upgrad-

    ing and checking the system

    are possible without major

    effort.

  • 8/13/2019 Catalog Lab

    9/140

    1

    7

    The service wing with cross

    element and systems and

    furniture components that can

    be placed below it.

  • 8/13/2019 Catalog Lab

    10/140

    4

    3

    2

    1

    5

    6

    7

    8

    9

    3

    2

    1

    4

    5

    6

    8

    7

    8

    Service modules

    1.1 Service spine

    Service spines provide fittings

    and sockets, can be fitted with

    supply pipes and cables and

    waste pipes, and are used for

    mounting a wide range of shelf

    systems.

    Service panels and electrical

    modules can be individually

    equipped and upgraded. The

    powder-coated panels made ofsteel sheet are clipped into the

    lateral profiles on the service

    spine above the worktop with-

    out the need for tools.

    The uprights for the service

    spine can be extended in

    height at any time as required,

    e. g. for installing overbench

    cabinets.

    Layout

    1 Installation com-partment panel

    2 Cantilever3 Sanitary panel4 Reagent shelf

    (add-on part)5 Slot panel6 Cover panel7 Electrical module8 Uprights9 Frame with

    levelling screw

    1 Underbench units2 Table connectionpanel

    3 Row of panels4 Instrument shelf

    (add-on part)5 Reagent shelf

    (add-on part)6 Worktop7 Uprights8 Frame with

    levelling screw

  • 8/13/2019 Catalog Lab

    11/140

  • 8/13/2019 Catalog Lab

    12/140

  • 8/13/2019 Catalog Lab

    13/140

  • 8/13/2019 Catalog Lab

    14/140

    12

    Service modules

    1.1 Service spine

    Service panels and electrical modules

    The panels and modules are

    clipped into a mounting frame

    and can be arranged as

    required.

    The outlet valves in the panels

    can be connected as required

    to the main pipes using

    branch pipes. The main pipes

    are laid in the installation com-

    partment underneath theworktop or the cantilever.

    It is possible to fit up to five

    angle taps to sanitary panels

    (for gas, water, vacuum, com-

    pressed air or high purity gas

    valves).

    In addition, a sanitary panel

    can be fitted with a drip cup.

    Up to eight sockets for alter-

    nating current or up to two

    sockets for three-phase cur-

    rent can be fitted to electrical

    modules. Different sockets

    can be combined.

    As an option, the sockets can

    also be supplied fitted flush

    with the service spine.

    The supply is provided using

    an electrical module with an

    interface to the electrical

    supply and, on request, anelectrical module with up to

    14 fuses under a cover.

    Cover panels cover the areas

    between sanitary panels and

    electrical modules. Individually

    mounted electrical modules

    are always fitted with a half

    height cover panel.

    Behind slot panels there are

    slot rails to which add-on parts

    of varying height can be fitted.

    Sanitary panelwith angle taps

    Sanitary panelwith drip cup

    Electrical modulewith eight sockets foralternating current

    Electrical module withvarious sockets

    Electrical module withinterface to the electrical

    supply and electrical modulewith fuses

    Cover panel with anelectrical module

    Slot panel

  • 8/13/2019 Catalog Lab

    15/140

    1

    13

    Add-on parts

    Add-on parts can be fitted at

    different heights to panels

    and runners with slots on the

    service spine.

    The following add-on parts

    are available:

    - Reagent shelf

    - Instrument shelf

    - Scaffold points

    Reagent shelf

    A glass plate is fastened to

    two brackets using screws.

    An angle behind the glass

    plate is used as a stop and for

    stiffening.

    Bracket for reagent

    shelf

    The reagent shelf can beextended using rods sup-

    ported with braces at the top

    or bottom. For this purpose

    special brackets with a fasten-

    ing feature for rods with a

    diameter of 13 mm are used.

    The rods are fixed in the holes

    in the brackets using clamping

    screws. Braces can be clamped

    to the rods, the braces are fas-

    tened to the slots in the serv-

    ice spine using hooks andincrease the load capacity of

    the rods.

    Instrument shelf

    A panel is fastened to two

    brackets from below using

    countersunk screws. An angle

    behind the panel is used as a

    stop and for stiffening. The

    panel is supported underneath

    or on top using braces.

    Scaffold points

    Scaffold points are used for

    fitting a rod with a diameter

    of 13 mm. The rod is fixed by

    turning the knob on the scaf-

    fold point by a quarter of a

    turn.

    Reagent shelf

    Bracket withhole for rod

    Reagent shelfwith brace underneath

    Reagent shelfwith brace on top

    Bracket for reagentshelf

    Instrument shelfwith brace underneath

    Instrument shelfwith brace on top

    Scaffold points

  • 8/13/2019 Catalog Lab

    16/140

    14

    Service modules

    Service modules

    1.1 Service spine

    Service modules are used for

    the provision of electrical and

    mechanical services at the

    laboratory workstation. They

    are fastened to the raw ceil-

    ing using a ceiling anchor.

    The sanitary panels and elec-

    trical modules fastened to the

    uprights can be positioned on

    both sides and can beequipped as required.

    Suspended service booms are

    used for the provision of elec-

    trical and mechanical services

    at the laboratory workstation.

    The modular spines are fas-

    tened to the raw ceiling. The

    sanitary panels and electrical

    modules fastened between

    the uprights can be posi-

    tioned on both sides and can

    be equipped as required.

    Suspendedservice booms

  • 8/13/2019 Catalog Lab

    17/140

    1

    15

    Technical data

    Service spine

    Widths (mm)

    Height (mm)

    Depths, single-sided with table (mm)

    Depths, double-sided with table (mm)

    Working heights (mm)

    Pillar extension (mm)

    Pillar extension for suspended cabinet

    with height of 540 mm (mm)

    Pillar extension for suspended cabinet

    with height of 720 mm (mm)

    Pillar extension up to

    the ceiling (mm)

    Height S2 panel

    for working height of 900 mm (mm)

    600, 900, 1200,

    1500, 1800

    1745

    750, 900

    1500, 1800

    720, 900

    960

    515

    695

    1725

    280

    Dimension of the panels

    Width, panel (mm)

    Width, slot panel (mm)

    Height, panel (mm)

    300

    300, 75

    180, 360

    Dimension of the electrical modules

    Width, electrical module (mm)

    Height, electrical module (mm)

    300

    180

    Bench mounted service module

    Widths (mm)

    Height (mm)

    Working heights (mm)

    Opening

    for working height of 720 mm (mm)

    Opening

    for working height of 900 mm (mm)

    900, 1200,

    1500, 1800

    1563

    720, 900

    563

    383

    Reagent shelf

    Widths (mm)

    Depth (mm)

    Load carrying capacity

    600, 900, 1200,

    1500, 1800

    150

    20 kg

    Instrument shelf

    Widths (mm)

    Depth (mm)

    Load carrying capacity

    600, 900, 1200,

    1500, 1800

    300

    30 kg

    Scaffold points

    Diameter of the hole for rods (mm)

    Load carrying capacity in conjunction

    with reagent shelf without brace

    Load carrying capacity in conjunction

    with reagent shelf with brace

    13

    5 kg

    20 kg

    Service module

    Width (mm)

    Length of the uprights (mm)

    300

    adjusted to suit

    customer needs

    Width, panel (mm)

    Heights, panel (mm)

    300

    180, 360

    Dimension of the electrical modules

    Width, electrical module (mm)

    Height, electrical module (mm)

    300

    180

    Suspended service boom

    Widths (mm)

    Length of the uprights (mm)

    600, 900, 1200,

    1500, 1800

    adjusted to suit

    customer needs

    Dimension of the panels

    Width, panel (mm)

    Height, panel (mm)

    300

    180

    Dimension of the electrical modules

    Width, electrical module (mm)Height, electrical module (mm)

    300180

    Dimension of the panels

  • 8/13/2019 Catalog Lab

    18/140

    1

    1

    2

    3

    4

    56

    8

    3

    2

    5

    6

    7

    4

    700 mm

    211mm

    100 mm300 mm 300 mm

    520 mm90 mm 90 mm

    1

    16

    Service modules

    1.2 Service wing

    The service wing provides

    services and waste facilities at

    the laboratory workstation.

    The service wing support is

    fastened underneath the lab-

    oratory ceiling, independent

    of modular size. To avoid

    vibrations, the support is

    braced in the longitudinal and

    transverse directions.Different fittings or electrical

    modules can be fitted to the

    Layout

    Components ofthe service wing

    1 Cover panel2 Wing element

    3 Electrical module4 Support5 Front cover6 Workplace light7 Fittings8 Alsident exhaust

    (optional)

    underside of the elements on

    both sides.

    Due to the modular layout, the

    services and energy supplies in

    the laboratory can be varied as

    required during the design of

    the laboratory space and can

    also be upgraded at any time.

    Cross sectionthrough the

    service wing

    1 Support2 Indirect room

    lighting3 Service duct4 Workplace lighting

    and possibility forextract air spigot

    5 Electrical supplyarea

    6 Mechanicalservices supply andwaste area

    Section

  • 8/13/2019 Catalog Lab

    19/140

    1

    17

    Service wing elements

    Feed element

    T-element

    Wing element

    Cross element

    Design variants

    Mechanical

    services

    Electrical

    Lighting

    Ventilation

    Dust protection

    A very wide range of supply services such as flammable gas,

    technical gases, pure gas, vacuum, compressed air, water in

    a very wide range of quality stages

    Waste pipes for waste water

    Electrical supply for alternating current and three-phase current

    Fuse modules

    Workplace lighting

    - Indirect room lighting

    Alsident exhaust

    Extract air spigot for mobile fume cupboard AeroEm

    Plate valves for room ventilation

    Cover over the spacing elements from above

  • 8/13/2019 Catalog Lab

    20/140

    18

    Service modules

    1.2 Service wing

    Service wing elements

    Feed element

    The supply pipes are fed

    through the wall at the height

    of the service wing using a feed

    element. As an alternative to

    the feed element, the supply

    pipes can be fed through a

    service duct from the ceiling to

    the service wing. In this case

    the isolating valves are outsidethe service wing.

    Easy access to the isolation

    valves and pipes in the service

    wing is provided by opening

    the feed element.

    Mechanical services supply

    and waste facilities directly

    at the service wing

    The fittings (angle taps, high

    purity gas valves or quick

    release outlets) supply the labo-

    ratory workstation with flam-

    mable gas, technical gases,

    pure gases, vacuum or com-

    pressed air and water.

    The pipes from, e. g., service

    distribution terminals (see sec-

    tion 1.3) or mobile sinks (see

    section 2.11) are connected to

    quick release outlets.

    Wing elementwith angle taps

    Wing element withhigh purity gas valve

    Quick release outlets forgas, water, waste water

    Feed elementOpen

  • 8/13/2019 Catalog Lab

    21/140

    1

    19

    Dimension of the elements

    Length, feed element (mm)

    Lengths, wing elements (mm)

    Lengths, T elements (mm)

    Lengths, cross elements (mm)

    Width (mm)

    Height (incl. dust cover) (mm)

    600

    600, 900, 1200,

    1500

    1500

    1500

    700

    210

    Dimensions panels and modules

    Length (mm)

    Width (mm)

    300

    180

    Dimensions of the workplace lights/cover panels

    Lengths (mm)

    Width (mm)

    600, 900,

    1200, 1500

    100

    Electrical supply directly at

    the service wing

    Up to eight sockets for alter-

    nating current or up to two

    sockets for three-phase cur-

    rent can be fitted to electrical

    modules.

    Connections, e. g., for tele-

    phones, monitors, data or

    loudspeakers can also beinstalled on electrical modules.

    Extract air removal directly

    at the service wing

    Extract air spigots can be fitted

    to the underside of the service

    wing between the two halves

    of the wing.

    The spigots are connected to

    the central exhaust system

    using an extract air duct.

    It is possible to connect, e. g.the ducts from mobile fume

    cupboards AeroEm (see section

    3.12) or specific Alsident

    exhausts (see section 3.13).

    Wing element withextract air spigots

    Wing elementwith electricalmodules

    Technical data

    Alsident exhaust onthe extract air spigot

  • 8/13/2019 Catalog Lab

    22/140

    3

    2

    1

    4

    20

    Service modules

    1.3 Service distribution terminal

    In combination with the serv-

    ice wing as a supply and

    waste disposal system, the

    service distribution terminal

    enables services and electrical

    supplies to be drawn directly

    at the workplace. The service

    distribution terminal is

    clamped to worktops or steel

    support frames using aclamping system.

    Layout

    Components of aservice distributionterminal

    1 Service distributionterminal with

    sanitary panel2 Service distribution

    terminal withelectrical module

    3 Clamping system4 Service beam

    The service distribution termi-

    nal is supplied using the service

    wing. Services and the electri-

    cal supply are provided using

    flexible pipes and cables thatare connected to the service

    distribution terminal.

    The pipes and cables are con-

    nected to the service wing

    using quick release outlets and

    plugs.

    A very wide range of mechani-

    cal services such as flammable

    gas, technical gases, pure

    gases, vacuum, compressed air

    can be provided.

    Design variants

    Up to three angle taps or high

    purity gas valves are fitted to

    the sanitary panel.

    Up to eight sockets for alter-nating current or up to two

    sockets for three-phase current

    can be fitted to the electrical

    panel. Different sockets can be

    combined.

    The pipes and cables between

    the service distribution terminal

    and the service wing are fed to

    the service beam and bundled

    with two straps. In addition, a

    steel rope on the service beam

    is used to provide strain relieffor the pipes and cables.

    The service distribution termi-

    nal can be fastened horizon-

    tally, e.g. to a worktop or ver-

    tically to a steel frame, e. g. to

    the rack (see section 2.8) .

  • 8/13/2019 Catalog Lab

    23/140

    1

    21

    Examples of the design variants

    Service distribution termi-

    nal with angle taps

    Service distribution terminal

    with high purity gas valves

    Service distribution terminalwith 8 sockets for alternating

    current

    Service distribution termi-

    nal with different sockets

    Fastening two service distri-

    bution terminals one above

    the other

    Vertical fastening of the

    service distribution terminal

    Horizontal fastening of the

    service distribution terminal

    Dimension service distribution terminal

    Width (mm)

    Height of a service distribution

    terminal (mm)

    Height of two service distribution

    terminals (mm)

    Depth (mm)

    Clamping range for the fastening (mm)

    300

    205

    410

    125.5

    10 to 90

  • 8/13/2019 Catalog Lab

    24/140

  • 8/13/2019 Catalog Lab

    25/140

  • 8/13/2019 Catalog Lab

    26/140

  • 8/13/2019 Catalog Lab

    27/140

    2

    25

    C frame steel support

    frames

    are extremely steady and can

    be loaded with 200 kg/m2.

    They carry suspended under-

    bench units without problems

    and provide users a large

    amount of knee and legroom.

    The height is adjusted using

    fully integrated adjusting

    strips.

    The cantilever frame

    provides the greatest legroom

    and lightest visual impact. It is

    fitted to service spines or

    directly to walls via its can-

    tilever bracket design.

    The multi-talent, the rack

    is used primarily for fitting

    items of equipment and has

    a load carrying capacity of

    150 kg.

    The heights of shelves can be

    adjusted as required in 45 mm

    increments.

    The shelves are available made

    of either melamine coated

    particleboard or powder-

    coated perforated steel plate.The shelves can be fitted with

    a tilt of up to 15 without

    tools. The extendable shelves

    mounted on caged ball-bear-

    ings and with 60 kg maximum

    load can be pulled out to 2/3

    of their depth. The castors

    enable the rack to be moved

    quickly.

    Further refined accessories

    The skate as a castor-mountedbase unit for your PC, the

    caddy as a trolley for the

    printer, add-on plate, assistant

    and round table are only a few

    examples of useful accessories

    in the laboratory environment.

    Our modular sinks

    Available in various installa-

    tion variants, the sink module

    impresses with its height of

    950 mm, which facilitates

    fatigue-free work. Sink and

    module are manufactured as

    one unit without joints.

    Our sink and drip modules are

    manufactured from self-sup-

    porting, acid proof technical

    ceramic material. The sup-porting structure is formed by

    underbench cupboards with

    plinth.

  • 8/13/2019 Catalog Lab

    28/140

    26

    The worktops for the laboratory benches are

    mounted on height adjustable bench frames

    (4-leg, C frame or cantilever frame) or directly

    on underbench units with plinth.

    The complete WIG-welded sides and supports

    are made of precision rectangular steel tube.

    Supports are bolted to the frame such that

    underbench units can be suspended in the

    bench frame or pushed underneath on

    castors.

    Work benches

    2.1 Add-on tables and supporting structures

    Supporting structures

    Add-on tables with H frame

    Material frame and supports (mm)

    Powder coating

    Load capacity

    Working heights (mm)

    Frame heights 4-leg (mm)

    Frame depths 4-leg (mm)

    Standard lengths (mm)

    The individual H frame has adjustable levelling feet made of steel each with a plastic

    plate.

    Levelling: - 5 to +25 mm with key projection.

    ST 50/25/2

    Layer thickness 80

    200 kg/m2

    720, 900

    690, 870

    552, 702, 852

    600, 900, 1200, 1500, 1800

    Variant

    Levelling covered by tube sleeve

    Two-part installation compartment cover that can be slid to the side for

    standard length frame 600 to 1800 mm in front of a service spine

    Add-on tables with C frame

    Material frame and supports (mm)

    Powder coating

    Load capacity

    Working heights (mm)

    Frame heights C frame (mm)

    Frame depths C frame (mm)

    Standard lengths (mm)

    C frames can be individually adjusted in height and have two levelling screws in the

    bottom steel brackets with a continuous pressure pad.

    ST 50/25/2

    Layer thickness 80

    200 kg/m2

    720 or 900

    690, 870

    552, 702

    600, 900, 1200, 1500, 1800

    Variant

    Two-part installation compartment cover that can be slid to the side for

    standard length frame 600 to 1800 mm in front of a service spine

  • 8/13/2019 Catalog Lab

    29/140

    2

    27

    Add-on table with cantilever frame

    Material frame and supports (mm)

    Powder coating

    Load capacity

    Working heights (mm)

    Frame heights (mm)

    Frame depths (mm)

    for wall fastening

    Standard lengths (mm)

    Fastening to the service spine or to the wall adapter using clamp strap.

    The individual cantilever frames have adjustable levelling feet made of steel each

    with a plastic plate. Levelling: - 5 to +25 mm with key projection.

    ST 50/25/2

    Layer thickness 80

    200 kg/m2

    720 bzw. 900

    690, 870

    661, 811

    Wall adapter ST 50/25/2

    WH 720 mm length 572 mm

    WH 900 mm length 752 mm

    600, 900, 1200, 1500, 1800

    Variant

    Levelling covered by tube sleeve

    Two-part installation compartment cover that can be slid to the side for

    standard length frame 600 to 1800 mm

    Notes on underbench units:

    Moveable underbench units for the working heights 720 and 900 mm can be

    pushed in.

    Notes on fastening:

    The cantilever frame can only be used in conjunction with a service spine or

    fixed to a strong wall.

    The welded tube construction is not self-supporting and must be firmly

    joined to the service spine or the wall.

  • 8/13/2019 Catalog Lab

    30/140

    28

    Work benches

    2.2 Worktops

    Worktops for wall benches without service spine

    or for free-standing benches

    The laboratory worktops are either mounted

    on bench frames (4-leg, C frame or cantilever

    frame) or directly on underbench units with

    plinth.

    The following dimensions are available

    Worktop widths (mm)

    Worktop depths (mm)

    600, 900, 1200, 1500, 1800

    600, 750, 900

    Material worktop

    Melamine resin coating on base

    material

    Particleboard, self-supporting*

    Polypropylene coating on

    particleboard base*

    Solid grade laminate without

    increased edge*

    Stainless steel on particleboard base*

    Tiles on coated plywood board*

    Technical ceramic material

    self-supporting

    Technical ceramic material -

    compound worktop on structural

    element

    Epoxy resin self-supporting

    Glass (safety glass ESG)

    * Work surfaces also available by the

    metre. Maximum length that can be

    supplied 3600 mm

    Grooved upstand edging strip made

    of plastic

    Wall sealing strip

    Installation ventilation grid

    750x100 mm

    Flush edge profile

    Marine edging

    Installation ventilation grid

    750x100 mm

    Flush stainless steel edge

    Marine edging

    Marine edging on two

    or three sides

    Profiled plastic end cap

    Marine edging on two

    or three sides

    Marine edging

    Profiled plastic end cap

    Without marine edge

    Marine edging on two or three sides

    Marine edging

    Material increased edge: Epoxy resin

    Marine edging on two or three sides

    Marine edging

    Variants

  • 8/13/2019 Catalog Lab

    31/140

    2

    29

    Worktops for wall benches and

    double work benches in front of a service spine

    The laboratory worktops are either mounted

    on bench frames (4-leg, C frame or cantilever

    frame) or directly on underbench units with

    plinth.

    The following dimensions are available

    Worktop widths (mm) 600, 900, 1200, 1500, 1800

    Wall bench

    Module depth 750 mm

    Module depth 900 mm

    Worktop depth (mm)

    675

    825

    Double work bench

    Module depth 1500 mm

    Module depth 1800 mm

    Worktop depth per side (mm)

    675

    825

    Material worktop

    Melamine resin coating on base

    material

    Particleboard, self-supporting*

    Polypropylene coating on

    particleboard base*

    Solid grade laminate without

    increased edge*

    Stainless steel on particleboard base*

    Tiles on coated plywood board*

    Technical ceramic material

    self-supporting

    Technical ceramic material -

    compound worktop on structural

    element

    Epoxy resin self-supporting

    Glass (safety glass ESG)

    * Work surfaces also available by the metre. Maximum length that can be

    supplied 3600 mm

    Flush edge profile

    Marine edging

    Flush stainless steel edge

    Marine edging

    Marine edging on two

    or three sides

    Profiled plastic end cap

    Marine edging on two

    or three sides

    Marine edging

    Profiled plastic end cap

    Without marine edge

    Marine edging on two or three sides

    Marine edging

    Material increased edge: Epoxy resin

    Marine edging on two or three sides

    Marine edging

    Variants

  • 8/13/2019 Catalog Lab

    32/140

    30

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - Acids with concentration > 10%

    - Concentrated hydrochloric acids

    - Nitric acids

    - Heated sulphuric acid

    - Resistance to moisture

    - Flat

    - Easy to dispose of

    - Low coating thickness

    - Wet room

    - Benches with medium stress

    26.4 kg/m2

    19 mm

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - Acids with concentration > 10%

    - Concentrated hydrochloric acids

    - Nitric acid

    - Heated sulphuric acid

    - Flat

    - Joints sensitive to moisture

    - Medium chemical resistance

    - Mobile tables, add-on tables, window

    benches

    - Instrument benches and laboratory

    benches in the dry area

    - Cannot be used in the moist or wet

    area

    19.6 kg/m2

    30 mm

    Melamine resin/postformed Polypropylene

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - Hydrocarbons

    - Citric acid

    - Oxalic acid

    - Carbon tetrachloride

    - Diesel oil

    - Ozone

    - Concentrated nitric acid

    - Chloroform

    - Petrol

    - Benzol

    - No joints

    - Flat

    - Light

    - High chemical resistance to acids and

    many solvents

    - Easy to dispose of

    - Less glass breakage

    - Soft surface sensitive to scratches

    - Sensitive to heat

    - Areas with high chemical resistance

    - Working with hydrofluoric acid

    - Area in which the lack of joints is

    important

    - Radio-isotope area

    20.3 kg/m2

    30 mm and 7 mm increased edge

    Tiles

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - None

    - Hydrofluoric acid

    - Flat

    - High resistance to chemicals

    - Tops have gaps with low resistance

    to chemicals

    - Gaps difficult to clean and difficult

    to decontaminate

    - Laboratory benches of all types that

    are subjected to chemicals

    30 kg/m2

    30 mm and 7 mm increased edge

    Work benches

    2.2 Worktops

    Properties of the worktops

    Solid grade laminate

  • 8/13/2019 Catalog Lab

    33/140

    2

    31

    Epoxy resin

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - Various solvents

    - Diluted acids

    - Hydrofluoric acid

    - Concentrated, warm mineral acids

    - No joints

    - Flat

    - Solid panel

    - High mechanical loading possible

    - Easy to dispose of

    - Surface sensitive to scratches

    - Sensitive to concentrated acids

    - Laboratory workstation of all type

    32 kg/m2

    19 mm and 7 mm increased edge

    Stainless steel

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - Cadmium

    - Lactic acid

    - Oxalic acid

    - Compounds containing chlorine

    and bromine

    - Formic acid

    - Sulphuric acid

    - No joints

    - High resistance to solvents

    - High temperature resistance

    - Sensitive to halogens and their

    compounds

    - For maximum loads in the area of

    decontamination and moisture

    resistance as well as solvent resistance

    - Biology

    - Microbiology

    - Pharmacy

    - Radio-isotope area

    - Pathology

    27.5 kg/m2

    30 mm and 7 mm increased edge

    Technical ceramic material

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - None

    - Hydrofluoric acid

    - Best chemical resistance

    - Mechanically stable

    - Easy to dispose of

    - Evenness tolerances due to firing process

    - Thermodynamic stress limited

    - Areas with very high chemical and

    mechanical stress

    56 kg/m2

    26 mm and 7 mm increased edge

    Glass

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - None

    - Hydrofluoric acid

    - Flat

    - High resistance to chemicals

    - Sensitive to knocks at corners and

    edges

    - Laboratory benches of all types subject

    to large amounts of chemicals

    38 kg/m2

    30 mm

    Stoneware compound worktop

    Critical substances

    Damaging

    substances

    Advantage

    Disadvantage

    Use

    Weight

    Total thickness

    - None

    - Hydrofluoric acid

    - Flat

    - Lighter than technical ceramic material

    - Best chemical resistance

    - Easy to dispose of

    - Thermodynamic stress limited

    - Areas with very high chemical stress

    40 kg/m2

    30 mm and 7 mm increased edge

    made of epoxy resin

  • 8/13/2019 Catalog Lab

    34/140

    32

    Work benches

    2.3 Mobile table

    Mobile table

    The rectangular worktop on our mobile tables

    is fitted to a seamless WIG-welded precision

    rectangular steel tube frame with four heavy

    duty castors. The two front castors are

    equipped with a full fixing device that simulta-

    neously locks castors and steering axle.

    Material worktop

    Melamine resin coating on structural

    element, self-supporting

    Polypropylene coating on

    particleboard base

    Solid grade laminate

    Stainless steel on particleboard base

    Tiles on coated plywood board

    Technical ceramic material

    self-supporting

    Technical ceramic material - compound

    worktop on structural element

    Epoxy resin self-supporting

    Glass (safety glass ESG)

    Marine edging

    Marine edging

    Marine edging

    Marine edging

    Marine edging;

    Material increased edge: Epoxy resin

    Marine edging

    Mobile table

    Working heights (mm)

    Widths (mm)

    Depths (mm)

    Load capacity

    720, 900

    1200, 1500

    750, 900

    150 kg/m2

    Variants

    Shelf fitted to bottom of frame

    Suspended underbench units

    Underbench units fitted onto bottom shelf

    Edges and increased edge

  • 8/13/2019 Catalog Lab

    35/140

    2

    33

    Swing

    2.4 Swing

    The curved worktop on the Swing sits on a

    4-leg frame. In combination with the Swing

    two tables of different depth can be joined

    together.

    Material worktop

    Melamine

    Melamine with beech finish

    Glass (safety glass ESG)

    3mm polypropylene edge, 4-sides

    3mm ABS edge, 4-sides

    Edges

    Swing

    Working heights (mm)

    Widths (mm)

    Depths (mm)

    Load capacity

    720, 900

    1200, 1500

    600 / 750

    675 / 825

    750 / 900

    200 kg/m2

  • 8/13/2019 Catalog Lab

    36/140

    1

    2

    3

    1

    2

    34

    Work benches

    2.5 Height adjustable table

    Height adjustable table

    The working height of the height adjustable

    C-frame tables can be infinitely varied

    between sitting working height (720 mm) and

    standing working height (900 mm) using an

    electric motor or a crank.

    Material worktop

    Melamine

    Postformed

    Melamine with beech finish

    3mm polypropylene edge, 4-sides

    3mm polypropylene edge on side and

    rear

    3mm ABS edge, 4-sides

    Edges

    Height adjustable table

    Working height (mm)

    Width (mm)

    Depth (mm)

    Load capacity

    720 to 900

    1500

    750

    70 kg/m2

    Height adjustabletable with electricmotor

    1 Switch for electricmotor

    2 Electric motor3 Bench frame with

    height adjustment

    Height adjustabletable with crank

    1 Crank2 Bench frame with

    height adjustment

  • 8/13/2019 Catalog Lab

    37/140

    2

    35

    Balance table

    The balance table is used for analytical balances

    and other sensitive measuring equipment.

    A heavy slab of concrete is specially mounted

    on a supporting construction.

    The supporting construction for the slab is

    vibration isolated from the outer worktop.

    Material worktop

    Melamine

    Postformed

    3mm polypropylene edge, 4-sides

    3mm polypropylene edge on side and

    rear

    Edges

    Balance table

    Working height (mm)

    Width (mm)

    Depths (mm)

    Width of the balance slab (mm)

    Depth of the balance slab (mm)

    Weight of the balance slab

    Total weight of the balance table

    720

    900

    750, 900

    400

    450

    65 kg

    120 kg

    2.6 Balance table

  • 8/13/2019 Catalog Lab

    38/140

    36

    Work benches

    2.7 VarioTHEK

    VarioTHEK

    The VarioTHEK provides the ideal workplace

    for laboratories as well as preparation rooms

    for teaching and research.

    This item can be variably equipped and is

    supplied ready to plug into the service wing.

    A waste water pump is integrated in the

    underbench unit.

    VarioTHEK

    Working height (mm)

    Width (mm)

    Length (mm)

    Height incl. castors (mm)

    Worktop

    Lighting

    Electrical supply

    Mechanical services supply

    Weight

    Fold down analysis table

    Working height 720 mm

    Suspended underbench units

    900

    825

    1900

    1920

    Melamine resin with or without raised

    drip cup arranged in the middle of the

    bench

    Variant

    Stoneware compound worktop.

    Integrated in the services transition

    above the working area

    Integrated in service modules on

    request

    Integrated in service panels for various

    water, gases, pure and high purity

    gases on request

    Approx. 250 kg

    Work surface beech finish

    Various variants on request

  • 8/13/2019 Catalog Lab

    39/140

    2

    37

    Rack

    Ideally suited for fitting items of equipment,

    our rack is available with or without castors.

    For analysis workplaces the rack is equipped

    with the secretaire worktop.

    Rack

    Widths (mm)

    Height (mm)

    Depth (mm)

    Widths, shelves (mm)

    Depths, fixed shelves (mm)

    Depths, extendable shelves (mm)

    Load carrying capacity per shelf

    Lad carrying capacity, total

    Shelves

    850, 1150

    1746

    677

    800, 1100

    440, 730

    440

    50 kg

    150 kg

    4 pieces

    Material: coated OSB panels

    Height adjustable in increments

    of 45 mm

    Optional

    Powder-coated perforated plate

    2.8 Rack

    Analysis workplace secretaire

    Widths (mm)

    Depth (mm)

    Load carrying capacity

    800, 1100

    900

    50 kg/m2

    Material worktop secretaire

    Melamine, colour grey

    Melamine with beech finish

    3 mm polypropylene edge, 4-sides

    3 mm ABS edge, 4-sides

    Edges

  • 8/13/2019 Catalog Lab

    40/140

    38

    Work benches

    2.9 Analysis tables

    Assistant

    The analysis workplace assistant is an addi-

    tional swivelling table for 900 mm high and

    750 mm deep add-on tables with 4-leg frame.

    In the extended position the swivelling table

    provides an additional, quarter circle shaped

    work surface with a working height of

    720 mm.

    Assistant

    Working height (mm)

    Width (mm)

    Depth (mm)

    Load capacity

    Worktop

    720

    695

    672

    30 kg/m2

    Melamine, colour grey or beech finish

    The round table can be fitted to 4-leg add-on

    tables.

    It comprises a round worktop with a cut-out

    and table leg.

    Round table

    Round table

    Working heights (mm)

    Diameter (mm)

    Load capacity

    Worktop

    720, 900

    1200

    50 kg/m2

    Melamine, colour grey or beech finish

  • 8/13/2019 Catalog Lab

    41/140

  • 8/13/2019 Catalog Lab

    42/140

    40

    Skate

    Work benches

    2.9 Analysis tables

    As a mobile unit the skate is used for the PC

    tower at the analysis workplace. It comprises a

    beech finish panel and a perforated plate side

    wall.

    The PC tower is fastened using a holding

    strap.

    Caddy

    As a moveable printer table, the caddy is used

    at the PC analysis workplace.

    Skate

    Width (mm)

    Height (mm)

    Depth (mm)

    Base

    215

    467

    450

    Melamine resin with beech finish

    Thickness 19 mm

    Caddy

    Width (mm)

    Height (mm)

    Depth (mm)

    Top work surface

    450

    400

    500

    Melamine resin with beech finish

    Thickness 30 mm

  • 8/13/2019 Catalog Lab

    43/140

    2

    41

    Laboratory sinks

    Laboratory sinks are permanently installed

    parts of the laboratory furniture and are

    placed either against a wall or a service spine.

    The sinks can be fitted to the service spine

    along the laboratory bench or as end sink units

    on double work benches.

    The sinks can be manufactured from

    polypropylene, stainless steel, epoxy resin or

    technical ceramic material. On worktops made

    of technical ceramic material and polypropy-lene, the sinks are fitted from below. On work-

    tops with tiles, or made of epoxy resin or

    melamine, the sinks are inserted from above.

    As standard underbench units with hinged

    doors and plinths are used.

    2.10 Sink work benches

    Material worktop

    Technical ceramic material

    Tiles*

    Melamine resin*

    Melamine resin*

    Polypropylene*

    Stainless steel*

    Technical ceramic material

    Compound worktop

    Epoxy resin

    * Work surfaces also available by the metre.

    Technical ceramic material

    Technical ceramic material

    Polypropylene

    Stainless steel

    Polypropylene

    Stainless steel

    Technical ceramic material

    Epoxy resin

    Options

    Wide range of different versions

    for water bench mounted service

    outlets

    Eye shower

    Draining rack

    The following dimensions are available for worktops

    Widths (mm)

    Depths (mm)

    900, 1200, 1500, 1800

    600, 675, 750, 825, 900

    500 x 400 x 250 (W x D x H) and

    400 x 400 x 250 (W x D x H)

    460 x 360 x 330 (W x D x H) and

    360 x 360 x 330 (W x D x H)

    500 x 400 x 250 (W x D x H) and

    400 x 400 x 250 (W x D x H)

    500 x 400 x 250 (W x D x H) and

    340 x 370 x 160 (W x D x H)

    500 x 400 x 235 (W x D x H) and

    400 x 400 x 235 (W x D x H)

    500 x 400 x 250 (W x D x H)

    510 x 360 x 265 (W x D x H) and

    360 x 360 x 265 (W x D x H)

    457 x 381 x 279 (W x D x H) and

    406 x 406 x 190 (W x D x H)

    On request

    On request

    On request or see Internet

    accessories catalogue

    Sinks bonded to the worktop

    from below using silicon and also

    secured with steel straps

    Sinks fitted flush with worktop

    and also secured with steel straps

    Sinks with beading fitted in the

    worktop from above

    Sinks with beading fitted in the

    worktop from above

    Sinks fitted in the worktop from

    below and welded

    Sinks welded flush with worktop

    Sinks fitted flush with worktop

    Sinks fitted flush with worktop

    Material sinks Sink dimensions i. L. (mm) Type of installation

  • 8/13/2019 Catalog Lab

    44/140

    42

    Sink modules

    Work benches

    2.10 Sink work benches

    Sink modules are primarily used as end sink

    units on double work benches but can also be

    fitted along the laboratory bench.

    The sink module and the adjacent work

    surfaces on the end sink unit are made of

    glazed, acid-proof technical ceramic material.

    A polypropylene splash bench runs along the

    entire width behind the sink module and the

    adjacent work surfaces.

    As standard underbench units with hinged

    doors and plinths are used.

    Sink modules for head sink unit

    Head sink unit standard width 1500 mm

    Width (mm)

    Depth (mm)

    600

    710

    Head sink unit standard width 1800 mm

    Width (mm)

    Depth (mm)

    900

    710

    Adjacent work surface left and right for standard width 1500 and 1800 mm

    Width (mm)

    Depth (mm)

    450

    670

    Sink height (mm)

    Thickness of the splash protection

    panel (mm)

    made of grey polypropylene

    300

    10

    Sink modules for service spine and as sink to stand against walls

    Sink module in front of service spine with depth 750 mm

    Widths (mm)

    Depth (mm)

    Sink module in front of service spine with depth 900 mm

    Widths (mm)

    Depth (mm)

    Sink module as sinks to stand against walls

    Widths (mm)

    Depth (mm)

    Sink height (mm)

    600, 900

    710

    600, 900

    860

    600, 900

    785, 935

    300

  • 8/13/2019 Catalog Lab

    45/140

    2

    43

    Drip cups

    Drip cups are integrated in the service spine

    or fitted in the worktop.

    Material drip cupsPolypropylene

    Technical ceramic material

    270 x 82 x 113 (W x D x H)

    260 x 82 x 108 (W x D x H)

    Sink dimensions i. L. (mm)

    Drip cups for service spines

    Drip cups for fitting in worktop

    Material worktop

    Melamine resin;

    Polypropylene

    Polypropylene

    Technical ceramic material;

    Tiles

    Melamine resin

    Stainless steel

    Epoxy resin

    Technical ceramic material

    Compound worktop

    Variants:

    various different versions for water fittings on request

    Polypropylene

    Polypropylene

    Technical ceramic material

    Stainless steel

    Stainless steel

    Epoxy resin

    Technical ceramic material

    Material drip wastes Sink dimensions i. L. (mm)

    255 x 105 x 150 (W x D x H)

    255 x 105 x 140 (W x D x H)

    250 x 95 x 112 (W x D x H)

    240 x 80 x 120 (W x D x H)

    240 x 80 x 120 (W x D x H)

    246 x 95 x 180 (W x D x H)

    245 x 95 x 150 (W x D x H)

    Type of installation

    Sinks with beading fitted in the

    worktop from above

    Sinks fitted to the worktop from

    below and welded

    Sinks with beading fitted in the

    worktop from above

    Sinks with beading fitted

    in the worktop from above

    Sinks welded flush with worktop

    Sinks fitted flush with worktop

    Sinks with beading fitted flush in

    the worktop from above

  • 8/13/2019 Catalog Lab

    46/140

  • 8/13/2019 Catalog Lab

    47/140

    2

    45

    The ready to plug-in compact system for the

    supply and disposal of water can be used on

    mobile or fixed laboratory benches and

    connected to the Waldner service wing using

    flexible pipes.

    A pump removes the waste water via the

    service wing.

    AquaEl

    AquaEl

    Material housing and sink Epoxy resin

    Width (mm)

    Depth (mm)

    Height (mm)

    Weight (incl. pipes)

    304

    500

    480

    11 kg

    Lengths, feed and waste pipes (mm)

    Lengths, power connecting cables (mm)

    2500

    2500

  • 8/13/2019 Catalog Lab

    48/140

  • 8/13/2019 Catalog Lab

    49/140

    47

    Fume cupboards and exhaust devices3.1 Bench mounted fume cupboards

    3.2 Low ceiling bench mounted fume cupboards

    3.3 Low level fume cupboards3.4 Walk-in fume cupboards

    3.5 Secuflow low volume bench mounted fume cupboards

    3.6 Secuflow low ceiling bench mounted fume cupboard

    3.7 Bench mounted fume cupboards with side panels

    3.8 Low ceiling bench mounted fume cupboards

    with side panels

    3.9 Special application fume cupboards

    3.10 Filter fume cupboards

    3.11 Radio-isotope fume cupboards

    3.12 Mobile fume cupboards

    3.13 Local exhaust ventilation

    3.14 Monitoring and control

    The requirements on laboratoryfume cupboards in relation to

    work safety, ease of use andeconomy have significantlyincreased, not least due to thenew EN 14175.Waldner fume cupboards, irre-spective of the type, providehighly innovative technologyfor every type of laboratory,and that way beyond the levelsrequired in the standards.

    3

  • 8/13/2019 Catalog Lab

    50/140

    48

    Fume cupboards and exhaust devices

    General

    Fume cupboards are a key ele-ment in laboratories. All workduring which gases, fumes,particles or liquids are handledin dangerous quantities andconcentrations must be per-formed in fume cupboards.

    Our fume cupboards, irrespec-tive of the type and version,ensure maximum safety, excel-

    lent ergonomics and maximumeconomy.

    During the development of ourfume cupboards, many years ofdesign know-how are accom-panied by scientific findings our designs are not based onthe example of others. Numer-ous patents demonstrate ourtechnical lead. All Waldnerfume cupboards of coursecomply with the requirements

    of EN 14175.As an important part of theoverall laboratory ventilationscheme, our fume cupboardscan be integrated into thebuilding ventilation concept.

    The fact that our Secuflowfume cupboard technologyalso reduces the overall costsfor the ventilation system is acommercial advantage thatcannot be ignored and that ismade possible by the inte-grated supportive flow technol-ogy. You will find further infor-

    mation on this topic atwww.waldner-lab.com.

    Our in-house certified test

    laboratory for fume cup-

    board measurements

    Previously the effectiveness offume cupboards was definedby the extract volume flowrates. With the re-publicationof DIN 12924 Part 1 in 1991, alimit for the leakage of test gaswas defined instead. In this

    way it was intended to facili-tate innovations, particularly inrelation to energy saving via areduction in the extract volumeflow rates.

    Our test facility, which hadalready been in existence formany years and that had beensetup for studies on fluidmechanics and the optimisa-tion of fume cupboards, was asa result expanded to form atest laboratory for ventilationengineering tests according toDIN 12924 Part 1 in 1991. Overthe last somewhat more than10 years, over 150 differenttypes of fume cupboard havebeen type tested in this testfacility.

    On the publication of EN14175 we setup our new testlaboratory for fume cup-boards. The latest technicalequipment and the GS certifi-cation via TV Product ServiceGmbH guarantee the bestmeasurement results in rela-tion to accuracy and repro-ducibility.

    In two independent test cham-bers it is possible to independ-ently test two fume cupboardsin parallel due to the duplica-tion of equipment.

  • 8/13/2019 Catalog Lab

    51/140

    3

    49

    We test fume cupboards tothe old standards DIN, BS(UK), and AFNOR (France), aswell as to the new EuropeanEN 14175 that will replace theprevious national standards.Furthermore, we can also per-form measurements accordingto ASHRAE 110/1995 as areoften required by customersbased in the USA.

    Anybody who designs, buildsand tests fume cupboards hasa major responsibility. Notleast to the staff in the labora-tory whose health depends onthe safe and reliable operationof the fume cupboards.Beyond our ISO 9001:2000certification and the GS mark-ing for our entire productrange, we have closed theloop in relation to fume cup-board tests and had our newtest laboratory tested and cer-tified by TV Product ServiceGmbH according to the Ger-man law on equipment safety(Gertesicherheitsgesetz). As aresult the first certified testlaboratory for fume cupboardsin Europe is in Wangen im All-gu. Visitors are welcome atany time.

    Safety via the patented

    intake airflow profile on

    the front edge of the fume

    cupboard worktop

    It prevents turbulence thatcould carry pollutant emissionsin the direction of the user inthe laboratory.

    Air flowing into the fume cup-board is guided via the airfoil-

    like profile geometry (with lowturbulence over the worktop)to the rear panel low levelextraction which ensures thesafe removal of heavy gases,e.g. solvent fumes.

    The technology in the new Waldner test laboratory:

    Two independent ventilation systems incl. cooling andheating for supply air and extract air with a flow rate of2,500 m3/h each

    Input air either via ceiling ventilation or supply air outletsof varying design

    Pressure and air quantity conditions in the test chambervariably adjustable using high resolution ventilation controltechnology

    Temperature conditions variably adjustable, from isother-mal to the simulation of various temperature gradients inthe chamber

    Reference air exchange rate determination according to

    DIN 1952 and VDI 2080 Calibrated test gas control Transducers with electronic data acquisition Data analysis and documentation using PC software

    Clear view of all processes in

    the workspace

    The high level glazed panelenables the experimentalequipment and processes tobe clearly seen.

    We offer the best possible

    utilisation of the working

    width The slender, patentedfume cupboard side posts,

    only 25 mm wide, provide themaximum workspace widthand due to their special shapeensure that there is little tur-bulence in the intake air.

  • 8/13/2019 Catalog Lab

    52/140

    50

    The removable panels with

    perimeter extraction in the

    fume cupboard rear panel

    provide maximum grouping

    of services

    In this way we ensure thatmechanical and electrical serv-ices can be quickly and eco-nomically changed orupgraded.

    For more safety andeconomy

    Maximum user safety is pro-vided by our toothed belt sashmounting along with signifi-cantly reduced maintenanceeffort. The stainless steel rein-forced toothed belts also pro-vide the supply of electricalpower and signals to thephoto-electric barrier in themoving sash.

    Endurance tests with morethan 200,000 load cycles ver-ify that this mounting is atleast ten-times more resistantto load changes compared toa conventional steel ropemounting.

    The track mechanism also

    provides additional safety

    In the unlikely case that bothsash mountings fail, the sashis stopped in fractions of asecond.

    All functions at a glance

    The control switch and dis-plays integrated into the fumecupboard side posts provideinformation on the operationalstate of the fume cupboard atthe ideal eye height.

    Sash handle with aero-

    dynamic design the air frowsinto the workspace when the

    sash is opened and pollutantemissions due to the openingsash are prevented. The bal-anced and free-moving sashmechanism including therelease for the sash stop canbe operated with one hand.

    The motion detector

    The sash is closed automati-cally if there is nobody work-ing at the fume cupboard.The photo-electric barrierstops the closing process ifthere are objects protrudingfrom inside the workspace.

    Cost saving due to exten-

    sive pre-assembly

    The factory pre-assembly andtesting of our fume cupboardsenable them to be very quicklyinstalled on-site.Due to the easy to maintaindesign, all components arequick and easy to access whenrequired.During installation, servicingand maintenance, our trainedinstallation and service person-nel ensure that everything

    goes smoothly and quickly.

    Fume cupboards and exhaust devices

    General

  • 8/13/2019 Catalog Lab

    53/140

  • 8/13/2019 Catalog Lab

    54/140

    5

    4

    3

    2

    1

    6

    7

    8

    9

    10

    11

    52

    Bench mounted fume cup-boards according to EN 14175

    prevent fumes, aerosols and

    dusts escaping into the labora-

    tory in dangerous quantities or

    concentrations.

    The following substances arenot approved for use in bench-mounted fume cupboards:

    Radioactive substances Microorganisms Genotype modified substances

    Bench mounted fume cup-boards are not suitable for pulp-ing works. This instruction alsoapplies to bench mounted fumecupboards with ceramic liner.

    Fume cupboards and exhaust devices

    3.1 Bench mounted fume cupboards

    Fume cupboard EN 14175

    1 Sash handle2 Worktop

    3 Display andoperating device

    4 Upper sashwindow

    5 Fascia panel6 Extract-air

    manifold7 Baffle with

    integrated servicepanels

    8 Vertical sash

    9 Glazed side panel10 Cable through-put

    11 Service module

    1200 1500 1800 mm 900 mm

    WH900mm

    Extractairconnectionforairflowcontroller2

    890mm

    2710mm

    Extractairconnectionforfunctiondisplay2610mm

    250

    190

  • 8/13/2019 Catalog Lab

    55/140

    3

    53

    * Control system for the regulation of air exchange rates mustbe at a distance of at least 250 mm from the soffit for thecollection duct to avoid interference with the FAZ measure-ment.

    All air quantity figures relate to a sash opening heightof 500 mm and the recommended trace gas maximumvalues from the German industry body BG Chemie.

    On the use of on-site extract air control systems and otherproduct rules, the air quantities may vary.For this reason the operating limits must be agreed withWaldner in advance.

    A maximum extract pressure of 600 Pa should not be exceeded.

    The minimum extract volume must not drop below 400 m3/hper running metre fume cupboard width according to EN14175 and should therefore not be used as the basis for thedesign of the ventilation system.

    We are of course always available at any time to answer anyquestions during the planning or implementation of yourproject.

    Variants

    Worktops

    Internal lining

    Glazing of the side panel

    Cable through-put in the solid side

    panel

    Underbench units

    Electrical supply

    Water supply

    Drip cup on the service panel

    Lighting

    Electrical/RCCB

    Sash controller SC

    Technical ceramic material, polypropylene, stainless steel, epoxy resin, tiles, stoneware compound worktop

    Melamine resin, technical ceramic material, polypropylene, solid grade laminate

    Left and/or right side possible, not on version with internal lining in technical ceramic material

    Left and/or right side, possible with all internal lining variants

    (see section 4: Storage space - underbench unit for fume cupboards)

    Internal and/or external sockets

    Take-off valves for water, gases, pure gases and high purity gases

    Polypropylene, technical ceramic material

    In ex-proof version

    Optional

    Optional

    Air data mc6 EN 14175 for bench mounted fume cupboards

    Dimensions

    (mm)

    1200x900x2710

    1500x900x2710

    1800x900x2710

    Minimumvalue according to EN 14175

    (m3/h)

    480

    600

    720

    Technical data

    Data

    1200

    2710

    900

    1440

    1150

    250

    7

    2

    Standard

    1500

    2710

    900

    1440

    1450

    300

    10

    3

    Standard

    1800

    2710

    900

    1440

    1750

    350

    10

    3

    Standard

    Widths (mm)

    Height (mm)

    Depth (mm)

    Clear internal height in the

    workspace (mm)

    Clear internal width in the

    workspace (mm)

    Weight in kg (without installations)

    Scaffold points for 13mm

    Service panels

    Light switch, lighting

    Ventilation technics variants (see section 3.14)

    Function display

    Constant airflow damper

    Variable airflow damper

    Cross slide acquisition

    FAZ*

    Air flow controller AC

    Air flow controller AC

    Only with variable air flow controller AC

  • 8/13/2019 Catalog Lab

    56/140

    4

    3

    2

    1

    5

    7

    6

    8

    54

    Fume cupboards and exhaust devices

    3.2 Low ceiling bench mounted fume cupboards

    Low ceiling bench mounted fume cupboard EN 14175

    Low ceiling bench mountedfume cupboards are used inrooms that have low internalheight. The sash is two-partand does not project beyondthe total height of the fumecupboard.General purpose fume cup-boards according to EN 14175

    prevent fumes, aerosols anddusts escaping into the labora-tory in dangerous quantities orconcentrations.

    The following materials are notpermitted in low ceiling benchmounted fume cupboards: Radioactive substances Microorganisms Genotype modified substances

    Low ceiling bench mountedfume cupboards are not suit-

    able for pulping works. Thisinstruction also applies to lowceiling bench mounted fumecupboards with ceramic liner.

    1 Sash handle2 Worktop

    3 Display andoperating device

    4 Fascia panel5 Extract-air

    manifold6 Two-part verticalsash

    7 Baffle withintegrated service

    panels8 Service module

    1200 1500 1800 mm 900 mm

    250

    190

    WH900mmE

    xtractairconnectionforairflowcontroller2530mm

    2350mm

    Extractairconnectionforfunctiondispla

    y2250mm

  • 8/13/2019 Catalog Lab

    57/140

    3

    55

    Technical data

    * Control system for the regulation of air exchange rates mustbe at a distance of at least 250 mm from the soffit for thecollection duct to avoid interference with the FAZ measure-ment.

    All air quantity figures relate to a sash opening heightof 500 mm and the recommended trace gas maximumvalues from the German industry body BG Chemie.

    On the use of on-site extract air control systems and otherproduct rules, the air quantities may vary.For this reason the operating limits must be agreed withWaldner in advance.

    A maximum ectract pressure of 600 Pa should not be exceeded.

    The minimum extract volume must not drop below 400 m3/hper running metre fume cupboard width according to EN14175 and should therefore not be used as the basis for thedesign of the ventilation system.

    We are of course always available at any time to answer anyquestions during the planning or implementation of yourproject.

    Air data mc6 EN 14175 for low ceiling bench mounted fume cupboards

    Dimensions

    (mm)

    1200x900x2350

    1500x900x2350

    1800x900x2350

    Minimumvalue according to EN 14175

    (m3/h)

    480

    600

    720

    Variants

    Worktops

    Internal lining

    Glazing of the side panel

    Cable through-put in the solid side

    panel

    Underbench units

    Electrical supply

    Water supply

    Drip cup on the service panel

    Lighting

    Electrical/RCCB

    Sash controller SC

    Technical ceramic material, polypropylene, stainless steel, epoxy resin, tiles, stoneware compound worktop

    Melamine resin, technical ceramic material, polypropylene, solid grade laminate

    Left and/or right side possible, not on version with internal lining in technical ceramic material

    Left and/or right side, possible with all internal lining variants

    (see section 4: Storage space - underbench unit for fume cupboards)

    Internal and/or external sockets

    Take-off valves for water, gases, pure gases and high purity gases

    Polypropylene, technical ceramic material

    In ex-proof version

    Optional

    Optional

    Ventilation technics variants (see section 3.14)

    Function display

    Constant airflow damper

    Variable airflow damper

    Cross slide acquisition

    FAZ*

    Air flow controller AC

    Air flow controller AC

    Only with variable air flow controller AC

    Data

    1200

    2350

    900

    1080

    1150

    220

    4

    2

    Standard

    1500

    2350

    900

    1080

    1450

    260

    6

    3

    Standard

    1800

    2350

    900

    1080

    1750

    300

    6

    3

    Standard

    Widths (mm)

    Height (mm)

    Depth (mm)

    Clear internal height in the

    workspace (mm)

    Clear internal width in the

    workspace (mm)

    Weight in kg (without installations)

    Scaffold points for 13mm

    Service panels

    Light switch, lighting

  • 8/13/2019 Catalog Lab

    58/140

    5

    4

    3

    2

    1

    6

    7

    9

    8

    56

    Fume cupboards and exhaust devices

    3.3 Low level fume cupboards

    Low level fume cupboard EN 14175

    Low level fume cupboards areused for working with tall itemsof experimental equipment. Thesash is two-part and does notproject beyond the total heightof the fume cupboard.General purpose fume cup-boards according to EN 14175prevent fumes, aerosols and

    dusts escaping into the labora-tory in dangerous quantities orconcentrations.

    The following substances arenot approved for use in lowlevel fume cupboards: Radioactive substances Microorganisms Genotype modified substances

    Low level fume cupboards arenot suitable for pulping works.This instruction also applies to

    low level fume cupboards withceramic liner.

    1 Griffleiste2 Tischplatte

    3 Anzeige- undBedieneinheit

    4 Oberlicht-verglasung

    5 Frontblende6 Abluftsammel-

    kanal7 Luftleitwand mit

    integriertenMedien-Panels

    8 Zweigeteiltes

    vertikalesSchiebefenster9 Modultraverse

    1200 1500 1800 mm 900 mm

    250

    190

    WH540

    mm

    Extractairconnectionforairflowcontroller2890mm

    2710mm

    Extractairconnectionforfunctiondisplay2610

    mm

    180

  • 8/13/2019 Catalog Lab

    59/140

    3

    57

    Technical data

    * Control system for the regulation of air exchange rates mustbe at a distance of at least 250 mm from the soffit for thecollection duct to avoid interference with the FAZ measure-ment.

    All air quantity figures relate to a sash opening heightof 500 mm and the recommended trace gas maximumvalues from the German industry body BG Chemie.

    On the use of on-site extract air control systems and otherproduct rules, the air quantities may vary.For this reason the operating limits must be agreed withWaldner in advance.

    A maximum extract pressure of 600 Pa should not be exceeded.

    The minimum extract volume must not drop below 400 m3/hper running metre fume cupboard width according to EN14175 and should therefore not be used as the basis for thedesign of the ventilation system.

    We are of course always available at any time to answer anyquestions during the planning or implementation of yourproject.

    Variants

    Worktops

    Internal lining

    Glazing of the side panel

    Cable through-put in the solid side

    panel

    Underbench units

    Electrical supply

    Water supply

    Drip cup on the service panel

    Lighting

    Electrical/RCCB

    Side service pillar 180mm

    Sash controller SC

    Technical ceramic material, polypropylene, stainless steel, epoxy resin, tiles, stoneware compound worktop

    Melamine resin, technical ceramic material, polypropylene, solid grade laminate

    Left and/or right side possible, not on version with internal lining in technical ceramic material

    or side service pillar

    Left and/or right side, possible with all internal lining variants, not possible with side service pillar

    (see section 4: Storage space - underbench unit for fume cupboards)

    Internal and/or external sockets

    Take-off valves for water, gases, pure gases and high purity gases

    Polypropylene, technical ceramic material

    In ex-proof version

    Optional

    Optional, instead of service module below the worktop

    Optional

    Dimensions

    (mm)

    1200x900x2710

    1500x900x2710

    1800x900x2710

    Minimumvalue according to EN 14175

    (m3/h)

    480

    600

    720

    Ventilation technics variants (see section 3.14)

    Function display

    Constant airflow damper

    Variable airflow damper

    Cross slide acquisition

    FAZ*

    Air flow controller AC

    Air flow controller AC

    Only with variable air flow controller AC

    Data

    1200

    2710

    900

    1800

    1150

    270

    10

    2

    Standard

    1500

    2710

    900

    1800

    1450

    320

    12

    3

    Standard

    1800

    2710

    900

    1800

    1750

    370

    14

    3

    Standard

    Widths (mm)

    Height (mm)

    Depth (mm)

    Clear internal height in the

    workspace (mm)

    Clear internal width in the

    workspace (mm)

    Weight in kg (without installations)

    Scaffold points for 13mm

    Service panels

    Light switch, lighting

    Air data mc6 EN 14175 for low level fume cupboards

  • 8/13/2019 Catalog Lab

    60/140

    4

    3

    2

    1

    5

    6

    8

    7

    58

    Fume cupboards and exhaust devices

    3.4 Walk-in fume cupboards

    Walk-in fume cupboard EN 14175

    Walk-in fume cupboards areused for working with tall itemsof experimental equipment andwith items of experimentalequipment that are placed inthe fume cupboard using amobile table. The sash is two-part and does not projectbeyond the total height of the

    fume cupboard.General purpose fume cup-boards according to EN 14175prevent fumes, aerosols anddusts escaping into the labora-tory in dangerous quantities orconcentrations.

    The following substances arenot approved for use in walk-infume cupboards: Radioactive substances Microorganisms

    Genotype modified substancesWalk-in fume cupboards are notsuitable for pulping works.

    1 Sash handle2 Display and

    operating device3 Upper sash

    window4 Fascia panel5 Extract-air

    manifold6 Baffle with

    integrated servicepanels

    7 Two-part verticalsash

    8 Side service pillar

    1200 1500 1800 mm 900 mm

    250

    190

    Extractairconnectionforairflowcontroller2890mm

    2755mm

    Extractairconnectionforfunctiondisplay2610

    mm

    180

  • 8/13/2019 Catalog Lab

    61/140

  • 8/13/2019 Catalog Lab

    62/140

  • 8/13/2019 Catalog Lab

    63/140

  • 8/13/2019 Catalog Lab

    64/140

    4

    3

    2

    1

    5

    6

    8

    7

    62

    Fume cupboards and exhaust devices

    3.6 Secuflow low ceiling bench mounted fume cupboards

    Secuflow low ceiling bench mounted fume cupboard

    EN 14175

    Secuflow low volume benchmounted fume cupboards areused in rooms that have lowinternal height. The sash isdivided in two and does not

    project beyond the total height

    of the fume cupboard.

    Over and above the requirements

    of EN 14175, our Secuflow fume

    cupboards with supportive flowtechnology provide maximum

    economy and safety with lower

    extract air volumes and therefore

    lower power consumption.

    General purpose fume cupboards

    according to EN 14175 prevent

    fumes, aerosols and dusts escap-

    ing into the laboratory in danger-

    ous quantities or concentrations.

    The following substances are notapproved for use in Secuflow

    low volume bench mountedfume cupboards: Radioactive substances Microorganisms Genotype modified

    substances

    Secuflow low volume

    bench mounted fume

    cupboards are not suitable

    for pulping works. This

    instruction also applies to

    Secuflow low volume

    bench mounted fumecupboards with ceramic

    liner.

    1 Sash handle2 Worktop

    3 Display andoperating device

    4 Fascia panel5 Extract-air

    manifold6 Baffle with

    integrated servicepanels

    7 Two-part verticalsash

    8 Service module

    5

    1200 1500 1800 mm 900 mm

    250

    190

    WH900mmE

    xtractairconnectionforairflowco

    ntroller2530mm

    2350mm

    Extractairconnectionforfunctiondispla

    y2250mm

  • 8/13/2019 Catalog Lab

    65/140

    3

    63

    Technical data

    * Control system for the regulation of air exchange rates mustbe at a distance of at least 250 mm from the soffit for thecollection duct to avoid interference with the FAZ measure-ment.

    All air quantity figures relate to a sash opening heightof 500 mm and the recommended trace gas maximumvalues from the German industry body BG Chemie.

    On the use of on-site extract air control systems and otherproduct rules, the air quantities may vary.For this reason the operating limits must be agreed withWaldner in advance.

    A maximum extract pressure of 600 Pa should not be exceeded.

    The minimum extract volume must not drop below 300 m3/hper running metre fume cupboard width according to EN14175 and should therefore not be used as the basis for thedesign of the ventilation system.

    We are of course always available at any time to answer anyquestions during the planning or implementation of yourproject.

    You will find further information on the Secuflow benchmounted fume cupboard with supportive flow technol-ogy in printed form in our Secuflow brochure and fordownload in the Internet at www.waldner-lab.com

    Variants

    Worktops

    Internal lining

    Glazing of the side panel

    Cable through-put in the solid side

    panel

    Underbench units

    Electrical supply

    Water supply

    Drip cup on the service panel

    Lighting

    Electrical/RCCB

    Sash controller SC

    Technical ceramic material, polypropylene, stainless steel, epoxy resin, tiles, stoneware compound worktop

    Melamine resin, technical ceramic material, polypropylene, solid grade laminate

    Left and/or right side possible, not on version with internal lining in technical ceramic material

    Left and/or right side, possible with all internal lining variants

    (see section 4: Storage space - underbench unit for fume cupboards)

    Internal and/or external sockets

    Take-off valves for water, gases, pure gases and high purity gases

    Polypropylene, technical ceramic material

    In ex-proof version

    Optional

    Optional

    Ventilation technics variants (see section 3.14)

    Function display

    Constant airflow damper

    Variable airflow damper

    Cross slide acquisition

    FAZ*

    Air flow controller AC

    Air flow controller AC

    Only with variable air flow controller AC

    Data

    1200

    2350

    900

    1080

    1150

    220

    4

    2

    Standard

    1500

    2350

    900

    1080

    1450

    260

    6

    3

    Standard

    1800

    2350

    900

    1080

    1750

    300

    6

    3

    Standard

    Widths (mm)

    Height (mm)

    Depth (mm)

    Clear internal height in the

    workspace (mm)

    Clear internal width in the

    workspace (mm)

    Weight in kg (without installations)

    Scaffold points for 13mm

    Service panels

    Light switch, lighting

    Air data mc6 EN 14175 for Secuflow low volume bench mounted

    fume cupboards

    Dimensions

    (mm)

    1200x900x2350

    1500x900x2350

    1800x900x2350

    Minimumvalue according to EN 14175

    (m3/h)

    360

    450

    540

  • 8/13/2019 Catalog Lab

    66/140

    5

    4

    3

    2

    1

    7

    9

    8

    10

    6

    64

    Fume cupboards and exhaust devices

    3.7 Bench mounted fume cupboards with side panels

    Bench mounted fume cupboard with side panels

    EN 14175

    General purpose fume cup-boards according to EN 14175prevent fumes, aerosols anddusts escaping into the labora-tory in dangerous quantities orconcentrations.

    The following substances arenot approved for use in benchmounted fume cupboards with

    side installation: Radioactive substances Microorganisms Genotype modified substances

    Bench mounted fume cup-boards with side installation arenot suitable for pulping works.This instruction also applies tobench mounted fume cup-boards with side installationwith ceramic liner.

    1 Sash handle2 Worktop

    3 Service panels onthe inside in thearea of the posts

    4 Display andoperating device

    5 Upper sashwindow

    6 Fascia panel7 Extract-air

    manifold8 Baffle with

    scaffold points9 Vertical sash10 Installed side

    posts

    1200 1500 1800 2100 mm900 mm

    WH900mm

    2710mm

    Extractairconnectionforfunctiondisplayandairflowcontroller2810mm

    150 150

  • 8/13/2019 Catalog Lab

    67/140

    3

    65

    Technical data

    Data

    1200

    300

    2710

    900

    1440

    850

    320

    6

    6

    Standard

    1500

    300

    2710

    900

    1440

    1150

    390

    6

    6

    Standard

    1800

    300

    2710

    900

    1440

    1450

    450

    6

    6

    Standard

    Widths (mm) incl. service pillar

    Width installed side panels

    (left and right 150 mm each)

    Height (mm)

    Depth (mm)

    Clear internal height in the

    workspace (mm)

    Clear internal width in the

    workspace (mm)

    Weight in kg (without installations)

    Scaffold points for 13mm

    Service panels on the inside in the

    area of the posts

    Light switch, lighting

    2100

    300

    2710

    900

    1440

    1750

    510

    6

    6

    Standard

    Worktops

    Internal lining

    Glazing of the side panel

    Cable bushing

    Underbench units

    Electrical supply

    Water supply

    Drip cup fastened to the inside of

    the fume cupboard

    Lighting

    Electrical/RCCB

    Sash controller SC

    Extract-air manifold

    Technical ceramic material, polypropylene, stainless steel, epoxy resin, tiles, stoneware compound worktop

    Solid grade laminate, technical ceramic material, polypropylene, epoxy resin, powdered steel sheet,

    stainless steel

    Not possible

    Via front panel and side panel

    (see section 4: Storage space - underbench unit for fume cupboards)

    Internal and/or external sockets

    Take-off valves for water, gases, pure gases and high purity gases

    Polypropylene, technical ceramic material

    In ex-proof version

    Optional

    Optional

    For fume cupboard width (mm) Size (mm) Material

    1200 250 Polypropylene

    1500/1800/2100 315 Polypropylene

    or 600 x 250 Stainless steel

    Ventilation technics variants (see section 3.14)

    Function display

    Constant airflow damper

    Variable airflow damper

    Cross slide acquisition

    FAZ*

    Air flow controller AC

    Air flow controller AC

    Only with variable air flow controller AC

    Air data mc6

    EN 14175 for bench mounted fume cupboards with side installation

    Dimensions

    (mm)

    1200x900x2710

    1500x900x27101800x900x2710

    2100x900x2710

    Minimumvalue according to EN 14175

    (m3/h)

    360

    480600

    720

    * Control system for the regulation of air exchange rates mustbe at a distance of at least 250 mm from the soffit for thecollection duct to avoid interference with the FAZ measure-ment.

    All air quantity figures relate to a sash opening heightof 500 mm and the recommended trace gas maximumvalues from the German industry body BG Chemie.

    On the use of on-site extract air control systems and otherproduct rules, the air quantities may vary.For this reason the operating limits must be agreed withWaldner in advance.

    A maximum extract pressure of 600