Bio-Rad Laboratories B I O P L E X 2 2 0 0 S Y S T E M
BioPlex™ 2200 Syphilis IgMThe first and only fully-automated, random access, multiplex platform for Syphilis IgM
Bio-Rad Laboratories B I O P L E X 2 2 0 0 S Y S T E M
Like no otherThe Syphilis IgM kit delivers a composite result for a comprehensive examination of the causative agent. Reportable results include:
• SyphilisIgMComposite*
• T. pallidum17kDa**
• T. pallidum47kDa**
*SyphilisIgMCompositeisarollupoftheindividualresults.**IndividualresultsavailableoutsideUS.
The BioPlex™ 2200 Syphilis IgM kit simultaneously detects two of the most clinically relevant antibodies directed against individual recombinant proteins to Treponema pallidum in a multiplex format.
TheBioPlex™ 2200SyphilisIgMkitisamultiplexflowimmunoassayforthe
qualitativenumericdetectionofT. pallidumIgMantibodiesinhuman
serum.ThetestsystemcanbeusedinconjunctionwithaSyphilisIgGtest
asascreenofactiveorpastinfection.Inaddition,theSyphilisIgMkitcan
beuseinconjunctionwithtreponemaland/ornon-treponemalbasedtests
asanaidinidentificationofacuteinfectionwithT. pallidum.
Performance Characteristics
PrecisionFollowingCLSIEP%-A2guidelines,theBioPlex™ 2200SyphilisIgMkit
demonstrated%CV’sofunder11%.
Low % CV’s can be attributed to:
a)fullautomation
b)magneticbead,heterogenousassayformat
c)minimumof150measurementsperanalyte
Specificity
BioPlex™ 2200 Syphilis IgM
Specificitycanbedefinedastheabilityofatesttogiveanegativeresultinnormalserafrompopulationsthatmay
betested.ThespecificityoftheBioPlex™2200SyphilisIgMassayon1247sampleswasexcellentat98%orbetter.
Detailedresultsareshownbelow.
Sensitivity
BioPlex™ 2200 Syphilis IgMData Summary:
• IgMwaspositivein43%(39/90)ofpatientswithprimary
andsecondarysyphilisinfection,helpingdistinguishacute
(primaryandsecondary)vs.non-acutediseasestatus.
• IgMandIgGcombinedwerepositivein94%*(85/90)of
patientswithprimaryandsecondarysyphilisinfection.
Please reference the IgM and IgG results, shown in the bar graph to the right.
* Of the 5 samples nonreactive by BioPlex, 3 out of 5 were also nonreactive for all other testing methods including RPR, TPPA and Western Blot. By excluding these 3 samples, BioPlex assay sensitivity is further improved.
3
1
1
45
40
35
30
25
20
15
10
5
0Untreated Treated Untreated Treated
Primary Secondary
Clinical Status
Co
unt
61.5% (8/13)
76.9% (10/13)
n=20
95.0% (19/20)
30.0% (6/20)
n=17
94.1% (16/17)
47.1% (8/17)
100% (40/40)
42.5% (17/40)
n=40Syphilis Nonreactive
Syphilis IgG Reactive
Syphilis IgG & IgM Reactive
n=13
Near Cutoff 1 – Tp 17kDa
Near Cutoff 2 – Tp 17kDa
Near Cutoff 1 – Tp 47kDa
Near Cutoff 2 – Tp 47kDa
80
80
80
80
1.0
1.0
1.0
1.2
Syphilis IgM Panel Members
Sample
N
Grand
Mean AI SD
0.05
0.06
0.06
0.08
Within-Run Between-Day Between-Run Total
%CV
5.4%
6.4%
6.1%
6.5%
SD
0.03
0.05
0.04
0.03
%CV
3.3%
5.8%
4.4%
2.6%
SD
0.07
0.10
0.07
0.09
%CV
7.3%
10.6%
7.6%
7.5%
SD
0.03
0.05
0.00
0.03
%CV
3.3%
5.8%
0.3%
2.6%
Population N BioPlex™ 2200 Syphilis IgM 95% CI
Unselected pregnant women 497 99.4% (494*/497) 98.2 to 99.8%
Pregnant women with known RPR/TPPA nonreactive results 250 98.0% (245**/250) 95.4 to 99.1%
Patients who had a Syphilis test ordered 500 98.2% (491***/500) 96.6 to 99.1%
* Two(2)sampleswereSyphilisIgMreactive,one(1)samplewasSyphilisIgMequivocal,andallthree(3)sampleswereRPR/TPPAnonreactive.
** Five(5)sampleswereSyphilisIgMreactive,allfive(5)sampleswereRPR/TPPAnonreactive.
***Eight(8)sampleswereSyphilisIgMreactive.Ofthoseeight(8)samples,five(5)sampleswereRPR/TPPAnonreactive,two(2)sampleswereRPR/TPPAreactive, andone(1)samplewasRPRnonreactive/TPPAreactive.One(1)samplewasSyphilisIgMequivocalandRPR/TPPAnonreactive.
Syphilis IgM Assay Procedure
Ordering InformationCatalog No. Description
BioPlex™ 2200 Syphilis IgM
665-1550 SyphilisIgMReagentPack...................................................................................................................................1pack663-1500 SyphilisIgMCalibratorSet...................................................................................................................................... 1set663-1520 SyphilisIgMCalibratorLotDataCD........................................................................................................................ 1CD663-1530 SyphilisIgMControlSet ......................................................................................................................................... 1set663-1540 SyphilisIgMControlLotDataCD........................................................................................................................... 1CD665-1560A SyphilisIgMInstructionsforUsePackage......................................................................................................1package
BioPlex™ 2200 System and Bulk Reagents
660-0000 BioPlex™ 2200....................................................................................................................................................1system660-0800 SampleRacks(30perbox) ................................................................................................................................... 1box660-0801 SampleTray.......................................................................................................................................................... 1tray660-0817 InstrumentSheathFluid(2perbox)....................................................................................................................... 1box660-0818 InstrumentWashBuffer......................................................................................................................................... 1box660-2003 ReactionVessels(1000perbag)........................................................................................................................... 1bag666-0001 InstrumentDetectorCalibrationPack...................................................................................................................1pack666-0002 InstrumentDetectorCleanPack...........................................................................................................................1pack666-0003 InstrumentProbeCleaningSolution(4perbox)..................................................................................................... 1box
B I O P L E X 2 2 0 0 S Y S T E M
1.Patientsample(5μL)automaticallyaddedtoReactionVessel
2.Samplediluent&beads(dilution1:60addedandincubatedat37°C
3.Washstep4.Conjugateaddedand
incubatedat37°C
5.Washstep6.Beadresuspended
inwashbuffer
7.Flow-based,duallaserdetect
Syphilis IgM Reagent Pack
Containsallthenecessaryreagents(samplediluent,
beadsandconjugate)toprocess100patientsper
packandupto100resultsperpatient.
Syphilis IgM Calibrator Set
Calibratorsareready-to-useand
prebar-codedforeaseofuse.
Syphilis IgM Control Set
QualityControlsareready-to-use
andprebar-codedforeaseofuse.
For further information, please contact the Bio-Rad office nearest you or visit our website at www.bio-rad.com/diagnostics
© 2009 Bio-Rad Laboratories, Inc. Printed in USA 12/09 DG09-0960 T-145
Clinical Diagnostics Group
Website www.bio-rad.com/diagnostics U.S. 1-800-2BIO-RAD Australia 61-2-9914-2800 Austria 43-1-877-8901 Belgium 32-9-385-5511 Brazil 5521-3237-9400 Canada 1-514-334-4372 China 86-21-64260808 Czech Republic 420-241-430-532 Denmark +45-4452-1000 Finland 358-9-804-22-00 France 33-1-47-95-60-00 Germany +49-(0)89-318-840 Greece 30-210-7774396 Hong Kong 852-2789-3300 Hungary +36-1-459-6100 India 91-124-4029300 Israel 972-3-9636050 Italy +39-02-216091 Japan 81-3-6361-7070 Korea 82-2-3473-4460 Mexico 52(55)5200-0520 The Netherlands +31-318-540666 New Zealand 64-9-415-2280 Norway 47-23-38-41-30 Poland 48-22-3319999 Portugal 351-21-472-7700 Russia 7-495-721-14-04 Singapore 65-6415-3188 South Africa 27-11-442-85-08 Spain 34-91-590-5200 Sweden 46-8-555-127-00 Switzerland 41-61-717-95-55 Thailand 662-651-8311 United Kingdom +44-(0)20-8328-2000