752 Ophthalmic Surgery, Lasers & Imaging Retina | Healio.com/OSLIRetina INSTRUMENTS/DEVICES/TECHNOLOGY Assistive Device for Efficient Intravitreal Injections Franziska Ullrich, MSc; Stephan Michels, MD, MBA; Daniel Lehmann, MSc; Roel S. Pieters, PhD; Matthias Becker, MD, PhD, MSc; Bradley J. Nelson, PhD ABSTRACT: Intravitreal therapy is the most common treatment for many chronic ophthalmic diseases, such as age-related macular degeneration. Due to the increasing worldwide demand for intravitreal injections, there exists a need to render this medi- cal procedure more time- and cost-efficient while increasing patient safety. The authors propose a medical assistive device that injects medication intravitreally. Compared to the manual intravitreal injection procedure, an automated device has the potential to increase safety for patients, decrease procedure times, allow for integrated data storage and documentation, and reduce costs for medical staff and expensive operating rooms. This work demonstrates the development of an assistive in- jection system that is coarsely positioned over the patient’s head by the human operator, followed by automatic fine positioning and intravitreal injec- tion through the pars plana. Several safety features, such as continuous eye tracking and iris recogni- tion, have been implemented. The functioning sys- tem is demonstrated through ex vivo experiments with porcine eyes. [Ophthalmic Surg Lasers Imaging Retina. 2016;47:752-762.] INTRODUCTION Since the approval of the first intravitreally applied drugs for the treatment of age-related macular degen- eration (AMD), 1,2 intravitreal injections have become one of the most frequently performed surgical proce- dures in ophthalmology, with more than 4 million in- jections worldwide in 2014. The demand for intravit- real therapy (IVT) of various drugs, including mostly vascular endothelial growth factor (VEGF) inhibitors, evolved dramatically during the past decade. The de- mand for IVT to treat the chronic diseases AMD and diabetic maculopathy/retinopathy is especially high and will increase in the foreseeable future 3 due to increasing prevalence of diabetes mellitus 4,5 and age- ing demographics, especially in industrialized coun- tries. 6,7 To cope with the increasing demand for IVTs, we propose an assistive device capable of performing intravitreal injections into the vitreous cavity of the eye. The system allows for precise and safe injections but is still controlled and monitored by the treating physician. Intravitreal Therapy Pharmaceutical companies discovered IVTs as a new area of investment and development. For the first time, ophthalmic drugs achieve more than $1 billion in U.S. revenue. Even more intravitreal drugs for an expanding spectrum of ophthalmic diseases are likely From the Multi-Scale Robotics Laboratory, ETH Zurich, Zurich, Switzerland (FU, DL, RSP, BJN); the Department of Ophthalmology, Triemli Hospital Zurich, Zurich, Switzerland (SM, MB); the University of Zurich, Zurich, Switzerland (SM); and the Department of Ophthalmology, University of Heidelberg, Heidel- berg, Germany (MB). Originally submitted March 29, 2016. Revision received May 23, 2016. Accepted for publication June 7, 2016. Acknowledgement: The authors acknowledge Sandro Kränzlin for supplying the results for the iris recognition algorithm. Drs. Becker, Michels, Nelson, and Ullrich report a connection with Ophthorobotics AG, including a pending patent. Ophthorobotics was financially sup- ported by Novartis Switzerland AG. The remaining authors report no relevant financial disclosures. Address correspondence to Stephan Michels, MD, MBA, Department of Ophthalmology, Triemli Hospital Zurich, Birmensdorferstrasse 497, 8063 Zurich, Switzerland; email: [email protected]. doi: 10.3928/23258160-20160808-09

Upload: dokhue

Post on 04-Jun-2018

281 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

752 Ophthalmic Surgery, Lasers & Imaging Retina | Healio.com/OSLIRetina

■ I N S T R U M E N T S / D E V I C E S / T E C H N O L O G Y ■

Assistive Device for Efficient Intravitreal InjectionsFranziska Ullrich, MSc; Stephan Michels, MD, MBA; Daniel Lehmann, MSc; Roel S. Pieters, PhD; Matthias Becker, MD, PhD, MSc; Bradley J. Nelson, PhD

ABSTRACT: Intravitreal therapy is the most common treatment for many chronic ophthalmic diseases, such as age-related macular degeneration. Due to the increasing worldwide demand for intravitreal injections, there exists a need to render this medi-cal procedure more time- and cost-efficient while increasing patient safety. The authors propose a medical assistive device that injects medication intravitreally. Compared to the manual intravitreal injection procedure, an automated device has the potential to increase safety for patients, decrease procedure times, allow for integrated data storage and documentation, and reduce costs for medical staff and expensive operating rooms. This work demonstrates the development of an assistive in-jection system that is coarsely positioned over the patient’s head by the human operator, followed by automatic fine positioning and intravitreal injec-tion through the pars plana. Several safety features, such as continuous eye tracking and iris recogni-tion, have been implemented. The functioning sys-tem is demonstrated through ex vivo experiments with porcine eyes.

[Ophthalmic Surg Lasers Imaging Retina. 2016;47:752-762.]

INTRODUCTION

Since the approval of the first intravitreally applied drugs for the treatment of age-related macular degen-eration (AMD),1,2 intravitreal injections have become one of the most frequently performed surgical proce-dures in ophthalmology, with more than 4 million in-jections worldwide in 2014. The demand for intravit-real therapy (IVT) of various drugs, including mostly vascular endothelial growth factor (VEGF) inhibitors, evolved dramatically during the past decade. The de-mand for IVT to treat the chronic diseases AMD and diabetic maculopathy/retinopathy is especially high and will increase in the foreseeable future3 due to increasing prevalence of diabetes mellitus4,5 and age-ing demographics, especially in industrialized coun-tries.6,7 To cope with the increasing demand for IVTs, we propose an assistive device capable of performing intravitreal injections into the vitreous cavity of the eye. The system allows for precise and safe injections but is still controlled and monitored by the treating physician.

Intravitreal TherapyPharmaceutical companies discovered IVTs as a

new area of investment and development. For the first time, ophthalmic drugs achieve more than $1 billion in U.S. revenue. Even more intravitreal drugs for an expanding spectrum of ophthalmic diseases are likely

From the Multi-Scale Robotics Laboratory, ETH Zurich, Zurich, Switzerland (FU, DL, RSP, BJN); the Department of Ophthalmology, Triemli Hospital Zurich, Zurich, Switzerland (SM, MB); the University of Zurich, Zurich, Switzerland (SM); and the Department of Ophthalmology, University of Heidelberg, Heidel-berg, Germany (MB).

Originally submitted March 29, 2016. Revision received May 23, 2016. Accepted for publication June 7, 2016.Acknowledgement: The authors acknowledge Sandro Kränzlin for supplying the results for the iris recognition algorithm.Drs. Becker, Michels, Nelson, and Ullrich report a connection with Ophthorobotics AG, including a pending patent. Ophthorobotics was financially sup-ported by Novartis Switzerland AG. The remaining authors report no relevant financial disclosures.Address correspondence to Stephan Michels, MD, MBA, Department of Ophthalmology, Triemli Hospital Zurich, Birmensdorferstrasse 497, 8063 Zurich, Switzerland; email: [email protected]: 10.3928/23258160-20160808-09

Page 2: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

August 2016 · Vol. 47, No. 8 753

to be introduced to the market in the near future, thus severely raising the number of IVTs and the related health care costs. The access to IVT will further in-crease with the beginning of the expiration of the pat-ent of the popular approved anti-VEGF drug ranibi-zumab (Lucentis; Genentech, South San Francisco, CA) in 2017. This may open the market for generics. In addition, the off-label use of other VEGF inhibitors, such as bevacizumab (Avastin; Genentech, South San Francisco, CA) and ziv-aflibercept (Zaltrap; Sanofi Genzyme, Bridgewater, NJ) — both approved for can-cer treatment — is common in some countries (eg, the United States), despite serious issues of compound-ing. However, the cost of off-label drugs is up to 10-times lower than that of approved agents.8 With decreasing drug cost, a larger population will gain ac-cess to treatment.

There is increasing pressure on reducing the com-pensation for intravitreal injection, mostly performed

by doctors.9 Physicians and hospitals are struggling to manage an increasing workload while obtaining re-duced reimbursement. This situation calls for a radi-cal change in the way ophthalmologists deliver IVTs. A challenge arises as the IVT needs to remain under the control of physicians, but costs are to be mini-mized. Thus, the time the physician spends perform-ing IVTs has to be significantly reduced while main-taining or even increasing safety and precision of the IVT for patient safety. In this context, the system we propose allows for automated, remotely controlled intravitreal injection and fulfills a large number of re-quirements.

Related RoboticsPerhaps the most well-known example of a surgi-

cal robot is the da Vinci Surgical System (Intuitive Surgical, Sunnyvale, CA) that was designed for gen-eral laparoscopic surgery and gained U.S. Food and

Figure 1. The system overview summarizes the system’s functions, such as the general automated injection, the local sterile environ-ment, a communication system, and the user interface.

Page 3: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

754 Ophthalmic Surgery, Lasers & Imaging Retina | Healio.com/OSLIRetina

Drug Administration approval in 2000. This master-slave surgical system consists of three arms for surgi-cal tools and one arm with a camera that provides ste-reoscopic 3-D vision for the surgeon. Currently, it is being used for many medical procedures in the fields of urology and gynecology.10,11 Other medical devices that do not comprise of the more traditional robotic manipulation systems are based on eg, magnetic ac-tuation, or radiotherapy. Respective examples are the EPOCH system (Stereotaxis, St. Louis, MO) and the Aeon Phocus (Aeon Scientific, Zurich, Switzerland) that provide catheter steering by magnetic guidance and the CyberKnife (Accuray, Sunnyvale, CA) that delivers radiotherapy.

The first automated laser-assisted surgery was reported in the 1980s to reshape the eye’s cornea in order to improve visual acuity.12 In general, robotic assistance is of particular interest for ophthalmic interventions as the required manipulation dexter-ity and force perception lies below human capabili-ties.13,14 To overcome these limitations, various ro-botic platforms have been proposed as assistive tools for ophthalmic surgery. Examples are active support systems that filter human hand-tremor (Micron15) or systems that provide cooperative manipulation of surgical tools (Steady-Hand16). Teleoperated systems, such as the da Vinci surgical system (Intuitive Surgi-cal, Sunnyvale, CA),17 the Preceyes Surgical System (Preceyes, Eindhoven, the Netherlands),18 or a paral-lel robot for vitreoretinal surgery19 translate the sur-geon’s hand motion to the scaled motion of tethered surgical instruments. Although such tethered systems

leave an incision with protruding surgical tools dur-ing intravitreal operation, untethered devices exist that can be inserted with minimal invasiveness in the eye and guided without any physical connec-tion to the outside world. The OctoMag (ETH Zurich, Zurich, Switzerland) is such a system20 and can guide an untethered magnetic microrobot in the vitreous humor of the eye.21,22

Although all mentioned systems provide capa-bilities for direct medical interventions (eg, surgery), more recently, an increasing interest has been ob-served in the system development for more specific procedures. One example is a medical device that enables automated intravenous injections for draw-ing blood or delivering fluids. Currently, two devices have been designed for such automated venipunc-ture, namely Veebot (Veebot LLC, Mountain View, CA) and VenousPro (VascuLogic LLC, Piscataway, NJ). Both systems use real-time imaging and robotic techniques for improved needle-based procedures and aim for complete automation. One reason for the development of such system is the frequency these procedures are carried out. Venipuncture is one of the most routinely performed invasive procedures and, depending on regulations, performed by physicians or other nursing staff.

Robot-Assisted Injection ProcedureHere we propose an automated system that assists

ophthalmologists during the administration of in-travitreal injections with high positioning accuracy, repeatability, and several integrated safety features.

Figure 2. (A) The assistive device for intravitreal injection is coarsely positioned over the patient’s head with a passive arm and locked into position. (B) The positioning module fine positions the system and the injection module allows for orientation of a syringe and injec-tion of medication into the eye.

Page 4: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

August 2016 · Vol. 47, No. 8 755

The proposed system consists of two modules and is suspended from a passive arm with lockable joints. The positioning module allows for computer assist-ed placement of the injection module. The injection module is suspended from the positioning module and is responsible for orienting a prefilled syringe with an attached injection needle to a preselected location on the sclera and inject medication into the vitreous cavity of the eye.

System RequirementsThe assistive system for automated intravitreal

injections was designed according to several require-ments concerning patient and physician safety, cost and time efficiency, and usability of the system as summarized in the Table. To ensure safety of the sys-tem, it must be sterilizable and the precision of the injection must be high, with position accuracy of the needle of less than 100 µm and orientation accuracy less than 1°. The patient and eye must be correctly identified and the pupil of the treated eye must be

tracked at all times, such that the injection procedure can be stopped in case of patient motion. To prevent eye motion, the system should include a stable fixa-tion of the patient’s direction of gaze. The position control of the system should be based on anatomic landmarks of the patient to account for anatomical differences among patients. To prevent multiple in-jections into the same location, which might cause thinning of the sclera, a history of previous injections and their locations should be saved for each patient. In order to render the assistive device attractive to the market, time and cost efficiency of the injection pro-cedure must be provided by reducing the time spent by the physician to a maximum of 1 minute per in-jection. Thus, the system should allow for automated documentation of the procedure and the possibility to inject different drugs into the patient’s eye. Also, the system should be remotely controlled, such that one physician can supervise several patients at the same time. To ensure easy handling and usability of the system, it must allow for quick positioning over

TABLE

Requirements for an Automated Assistive Device for Intravitreal InjectionsRequirements

Safety Reduction of bacterial contamination and sterilizability; better position accuracy < 100 µm; orientation accuracy < 1°; safety automated recognition of patient and eye; eye tracking at > 15 Hz; stable fixation of the patient’s direction of gaze; control based on anatomic land-marks; possibility to save injection history

Cost/Time Efficiency Remotely controlled system; possibility to inject several different drugs; automated docu-mentation of procedure

Usability Lightweight (< 5 kg); audio-visual communication between patient and doctor; located in an operating room or defined injection room

Figure 3. A computer vision algorithm allows for continuous pupil tracking during the injection process. The pupil and its center are tracked from different sides to determine the eye’s orientation in 3-D: (A) top view, (B) side view.

Page 5: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

756 Ophthalmic Surgery, Lasers & Imaging Retina | Healio.com/OSLIRetina

the patient’s face by a nurse, constraining size and weight. In order to account for guidelines concerning location of the intravitreal injection procedure,9 the injection system must fit into an operating room or a designated injection room. Audiovisual communica-tion between patient and doctor should be included to increase patient acceptance.

System OutlineThe proposed system for intravitreal injections

and its multiple functions are outlined in Figure 1. According to the requirements, the system consists of a wall- or ceiling-mounted dome that can easily be positioned over the patient’s face by a human opera-tor (ie, a nurse), using only one hand. The shape of the dome allows for the establishment of a local ster-ile environment utilizing laminar airflow. The hemi-spherically shaped casing covers all essential parts for the injection procedure, such as the mechatronic system, which moves a needle with high precision to the selected injection site on the sclera. For patient convenience, an audiovisual communications system is added, which consists of a camera system, a screen, a speaker, and a microphone. To allow for several medications to be injected into the eye, external drug reservoirs are attached. Alternatively, a prefilled sy-ringe23 can be directly clipped into a fitted holder for injection. For physician convenience, the system fea-tures an intuitive user interface that allows for remote control of the injection procedure. Prior to injection, the camera identifies the patient through an iris scan.

In order to increase time efficiency, the system auto-matically documents and saves the patient’s injection history in a database.

Preparation of the ProcedurePrior to the intravitreal injection procedure, the

patient is prepared by a nurse according to the stan-dard of care. The nurse gives local topical anesthetics, spreads the eyelids using a speculum, and disinfects the eye. The nurse then manually positions the sys-tem coarsely over the patient’s head by moving the passive arm into position and locking the joints. The system is illustrated hanging over a patient’s head in Figure 2. Visual information provided by a vision sys-tem during the injection process allows for automated fine adjustment of the system over the treated eye.

Iris RecognitionThe human iris offers a unique, stable pattern,

which has been shown to be suitable for person and eye identification.24 Thus, automated iris recognition of the patient’s iris has been implemented into the system for safety. In daily practice, human error may lead to injections into the wrong eye (eg, left instead of right), or the wrong patient (eg, two patients with the same or a similar name). If the wrong medication or a different dose of the drug is injected, treatment of the eye’s disease may not be optimal. Thus, we implemented an iris scan that is executed after ini-tialization of the injection procedure. The algorithm compares the iris to previously stored patient images

Figure 4. The safe region for intravit-real injection spans between ± 80° from the horizontal on the temporal side of the eye with a radial dis-tance of 3.5 mm to 4 mm from the limbus.

Page 6: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

August 2016 · Vol. 47, No. 8 757

in a database. The injection process is started only if the iris is recognized as the right patient and eye.

The iris recognition algorithm initially acquires an image of the eye under near infrared illumination, thereby reducing specular reflections and reveal-ing the complex features of the iris pattern. Subse-quently, the pupil and iris are detected utilizing the intensity gradient at the boundaries of the latter. The iris is isolated from the image and converted into a rectangular form by applying a coordinate transfor-mation. Specular reflections in the iris are filtered by thresholding the image and can be ignored during fur-ther analysis. Convoluting the unwrapped iris with a Gabor filter results in a binary iris code25 that can be compared to the saved image codes in the database. In a next step, the degree of similarity of two codes is determined by deriving the Hamming distance as a

ratio between disagreeing bits to the total number of bits in the binary codes.

Pupil TrackingA tracking algorithm has been implemented to

track the position of the pupil of the eye. During the injection process, several cameras continuously take images of the targeted eye. An algorithm analyzes the images and outputs the relative position and orientation of the eye. Pupil tracking, as illustrated in Figure 3, has been implemented to supply visual feedback to the motor controllers. The pupil and its center are tracked from different angles, such that the eye’s orientation can be determined in 3-D. Ad-ditionally, pupil tracking offers increased safety to the patient. As soon as the eye moves more than the allowed distance or if the eye moves fast, the system

Figure 5. Four examples of successful iris segmentation under demanding conditions, such as eyelash and eyelid occlusions and chang-ing light conditions.

Page 7: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

758 Ophthalmic Surgery, Lasers & Imaging Retina | Healio.com/OSLIRetina

detects the eye motion and automatically stops the injection procedure.

Eye and Head FixationThe standard of care to fixate the patient’s direc-

tion of gaze is to ask to look at a specific point in the room during manual injection. To fixate the eye dur-ing an automated injection process, a screen is inte-grated into the system. The patient can choose to view a personal image. Alternatively, the patient is given a task, such as counting the red stars on the screen. This small activity helps to focus the patient’s atten-tion to the screen rather than to the injection needle and fixates the eye in a predetermined orientation.

As illustrated in Figure 2, the patient lays on his or her back during the procedure. The patient’s head is placed on a surgical pillow with an indentation for the head to prevent backward or sideward head movement. Additionally, by tracking the pupil, the automated system has knowledge of the relative posi-tion of the patient’s eye, despite any small head or body motion, and can adjust. In the case that the pa-tient’s head or eye movement exceeds a velocity or position threshold, the system terminates the proce-dure and retreats to its initial position.

Intravitreal InjectionThe assistive device automatically suggests a point

for safe injection on the pars plana that lies within a range of 3.5 mm to 4 mm from the limbus and within an angle of ± 80° on the temporal side of the eye, as

illustrated in Figure 4. This region is chosen to avoid injecting into the crystalline lens of the eye on the one side and the retina on the other (Figure 4A) and allows for an additional safety margin. The prese-lected injection position can be altered manually by the treating physician on a user interface (UI). When all changes and data are submitted, the ophthalmolo-gist initiates the injection procedure on the UI. To ac-count for patient comfort, the system allows for direct visual and auditive interaction between physician and patient through the use of a screen, microphone, speaker, and two cameras.

After initialization of the injection process, the system first centers itself over the treated eye. Then, the injection needle is aligned with the intended in-jection point on the sclera with a predetermined ori-entation. Finally, the needle is moved linearly into the eye and medication is delivered intravitreally. The positioning accuracy during fine adjustment of the system is 0.05 µm with a bidirectional repeatabil-ity of 1.6 µm. The theoretical positioning resolution of orienting the syringe over the injection point is 6.8 × 10-5 degrees. Presumably, the precision is lowered during actuation due to backlash in the mechanical system or uncertainties during visual tracking.

Patient Data DocumentationA Swiss survey shows that the average time a

physician spends on documentation and administra-tive tasks increased from 15.8% in 2011 to 17.1% in 2013,26 leaving less time for patient interaction. In

Figure 6. Hamming distance distribu-tion of as calculated from 100 images of 20 eyes (five images of each eye). Defin-ing the threshold at 0.4, correct recogni-tion can be achieved for all images.

Page 8: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

August 2016 · Vol. 47, No. 8 759

practice, the average time for one intravitreal injection process, including preparation and documentation time, is approximately 10 minutes, of which the ac-tual injection procedure only takes 1 minute to 2 min-utes. To decrease the time needed for documentation, a system is implemented that allows to directly save a patient’s personal and medical data into a database. These data may include the patient’s name, date of birth, and patient number, as well as his/her medical condition, suggested treatment, and type and amount of the injected drug. After each intravitreal injection process, the physician can submit all changes made to the patient database, where information is saved as bi-nary code to account for storage space. Storing patient data in a database has the advantage that information is made available and searchable throughout the injec-tion process efficiently.

EXPERIMENTAL VALIDATION

To evaluate the injection system, the iris recogni-tion algorithm and the injection process are demon-strated through experiments.

Iris RecognitionThe iris recognition was tested by evaluating the

segmentation and recognition performance. Segmen-tation performance is evaluated by applying the al-gorithm to 2,739 images from an eye database, which

offers demanding conditions as images feature eye-lash and eyelid occlusions combined with changing illumination conditions. The algorithm achieved a 93.39% correct segmentation rate. However, in prac-tice, the iris recognition algorithm will be used on images of patients’ eyes held open by a speculum, reducing occlusions and increasing the success rate. Examples of iris segmentation featuring eyelash and eyelid occlusions under changing light conditions are shown in Figure 5. To analyze the recognition perfor-mance, 100 images of 20 different eyes (ie, five images per eye) were taken and the Hamming distance was derived for all possible intraclass (same eyes) and in-terclass (different eyes) combinations. The resultant distributions are normalized and shown in Figure 6. The two distributions are completely separated, such that a Hamming distance threshold of 0.4 would re-sult in 100% correct recognition rate.

Injection ProcedureTo demonstrate the functions of the proposed as-

sistive device for intravitreal injections, porcine eyes are injected during ex vivo experiments. The experi-mental setup is illustrated in Figure 7. Cadaver eyes were obtained from the local abattoir and experiments took place less than 10 hours postmortem. A porcine eye was fixed in a rubber holder and placed within the reach of the passive arm. First, a human opera-

Figure 7. Experimental setup of injecting into a porcine cadaver eye. The robot structure supports and controls the position and orienta-tion of the end-effector that allows for injecting a needle into the eye.

Page 9: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

760 Ophthalmic Surgery, Lasers & Imaging Retina | Healio.com/OSLIRetina

tor moved the passive arm coarsely over the pig eye and fixed it in position. Then the positioning module centered the injection module over the eye. Prior to injection, a region for safe injection was defined on the eye. Then the injection process was started, ignor-ing eye recognition.

Figure 8 depicts the steps of the injection proce-dure. Figure 8A shows the end-effector with the at-tached injection needle in its initial position. In a next step, the injection needle is aligned with the targeted point of injection on the pars plana by controlling the injection module (Figure 8B). By actuating the linear actuator in the end-effector, the injection is accom-plished. Finally, the needle is retracted linearly and the assistive device moves back to its initial position. The experimental process was repeated 10 times and the true injection point was compared to the targeted point for safe injection during post-processing. It was found that all 10 injections were within the safe re-gion for injection on the cadaver eye.

DISCUSSION

Since initial developments in the mid-1980s,27 the use of robotic technology in surgery has increased dra-

matically. Specialty areas that adopted robotics for as-sistance were first explored in neurology, urology, and gynecology,10 and more recently, in ophthalmology.28 A historical review of these developments can be found in Ballantyne et al. and Lanfranco et al.,29,30 and a more present overview is given in Bergeles et al.31 Advantages that robotic technologies have over traditional, human-operated procedures include an improved access to the surgical site, a higher technical precision of instrumen-tation positioning, and the possibility of tactile and vi-sual feedback beyond human capabilities.30 Moreover, robotic devices enable teleoperation whereby the sur-geon does not need to be present at the patient site,29 eliminating physiological tremor and fatigue while im-proving dexterity of the surgery.30 Furthermore, a surgi-cal robot may be sterilized whereas a human surgeon provides limited sterility.

Despite these clear advantages for patient and sur-geon, few automated medical assistive devices have made it past clinical trials and are routinely being used in hospitals. The disadvantages of robot-assisted surgery include limited adaptability of a system to a situation; thus, only simple surgical tasks are current-ly possible utilizing robots. Additionally, acquisition

Figure 8. Intravitreal injection procedure: (A) the end-effector in its initial position, (B) the needle is aligned with the point of injection and (C) is moved linearly into the sclera of the eye before (D) retracting it.

Page 10: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

August 2016 · Vol. 47, No. 8 761

cost of a robot-assisted system for surgery is relatively high, adding cost to procedures with low throughput. However, surgical procedures with a high patient throughput, such as intravitreal injections, may be-come more efficient in terms of time and cost.

As demonstrated in an experimental setting, intra-vitreal injections can be performed by automated, re-motely controlled systems in a most safe and precise manner. The requirements as listed in the Table can be technically fulfilled. Compared to today’s prac-tice, the proposed system has multiple advantages. First, the precision of the intravitreal injection can be increased with regard to location, depth, and vol-ume of the injection. Second, safety with regard to human errors, specifically in highly repetitive tasks, can be increased. The system recognizes the correct patient and eye using an iris scan. In clinics, treating large numbers of patients per day produces, despite all safety measures, imminent risk for human error such as selecting the wrong drug, treating the wrong patient (eg, poor hearing of elderly patients presum-ing their name to be called), or confusing left and right eye. An automated system with iris recognition avoids these risks. By having a closed drug-delivery system, mistakes in handling the drug (volume, air-bubbles, wrong drug) and bacterial contamination can be avoided. The most serious complication of an IVT is a bacterial infection in the eye, which can lead to rapid loss of not only vision but also of the whole eye. Off-label drugs used for intravitreal injec-tions (eg, bevacizumab and ziv-aflibercept) are fre-quently compounded by external pharmacies. In the past, multiple series of severe infections in the eye occurred due to misconduct in compounding.32

In addition to the features mentioned above, fur-ther development of the assistive system includes a closed laminar airflow system. Laminar airflow is ca-pable of reducing air contamination twentyfold com-pared to conventional operation room ventilation and removes 99.9% of particles larger than 3 µm.33 More-over, since disinfection is needed only at the point of injection a disinfecting patch, through which the injection is given, is currently being developed in ad-dition to a specific needle that is only exposed once it enters the eye. These features should minimize the risk of contamination and endophthalmitis.

The key aspect that makes an automated system attractive to clinics and physicians is the dramatic re-duction in the physician’s time performing and docu-menting IVTs, and thus, a reduction in cost. In the setting of a large clinic in Zurich, Switzerland, (De-partment of Ophthalmology, City Hospital Triemli), performing more than 7,000 IVTs per year, the doc-tor’s time associated with the injection can be reduced

from 10 minutes to a maximum of 1 minute. This frees up physicians’ time, allowing them to replace a tiring and very repetitive task by more valuable and demanding tasks, as well as leaving them more time for essential patient care. The cost of an automated device for intravitreal injections is similar to other technological systems used for ophthalmic surgery. Considering the saved physician time, the system am-ortizes within a time period of 2 years, without tak-ing into account other high-value procedures, which may be conducted in the freed time. Moreover, by creating a local sterile environment (ie, closed lami-nar airflow) around the patient’s head, an expensive operating room becomes obsolete. Furthermore, the system allows the ophthalmologist to control and su-pervise the IVT from outside the injection/operating room. It is therefore possible to integrate performing IVTs in a regular outpatients clinic or even to perform them from remote locations. In addition, a very pre-cise documentation of the procedure is amendable to electronic records, as well as automated billing. Both aspects can be easily integrated in the controlled con-duction and surveillance of the procedure by the phy-sician. In future, fingerprint recognition of the doctor performing the injections may help avoid situations where an untrained personnel member conducts an IVT with the system.

One of the key challenges will remain the accep-tance of automated systems in medicine. The center-piece of the patient-doctor relationship is the trust in the expertise and skills of the treating physician. With an automated system, the misperception of an anonymous robot taking over the medical care of a doctor could arise. It is therefore important that com-munication with the patient is part of the proposed system, allowing the patient to see his or her treating physician in a high-resolution screen and hear his or her voice. This will help to strengthen the fact that the doctor is treating an individual patient. Today, most patients understand that automated systems can be much more precise than a physician’s hand. Some automated systems, such as the da Vinci sys-tem for prostate surgery, have become the standard of care and the doctor’s use of an automated system is a sign for excellence and state of the art care.11 In ophthalmology, automated laser systems were first introduced many years ago. The first automated cor-neal laser surgeries were reported in the 1980s and have continuously been refined.12 No ophthalmolo-gist would consider today removing corneal tissue manually for achieving ideal optical properties in re-fractive surgery. Relevant steps in cataract surgery are also now conducted by automated laser systems (ie, femto-phaco surgery).34 The trend for automated sys-

Page 11: Assistive Device for Efficient Intravitreal Injections · Assistive Device for Efficient Intravitreal Injections. ... drugs for the treatment of age-related ... intravitreal injections

762 Ophthalmic Surgery, Lasers & Imaging Retina | Healio.com/OSLIRetina

tems that allow better precision and safety is evident throughout medicine, but especially in the field of ophthalmology. An automated intravitreal injection system is, in our eyes, the next logical step.

CONCLUSION

This work outlines the potential of an automated assistive system for intravitreal therapy for the treat-ment of AMD and other chronic ocular diseases. The development of an injection system is demonstrated that consists of a positioning and injection module. Experiments with ex vivo porcine eyes show that the system allows for precise injection into the previously defined safe region for injection, such that neither the crystalline lens nor the retina is damaged. In conclu-sion, technical feasibility of an automated system for IVT is shown that offers a wide range of advantages in today’s care of the growing number of patients with medically treatable ocular diseases.

REFERENCES

1. Gragoudas ES, Adamis AP, Cunningham ET Jr., Feinsod M, Guyer DR. Pegaptanib for neovascular age-related macular degeneration. N Engl J Med. 2004; 351(27):2805-2816.

2. Rosenfeld PJ, Brown DM, Heier JS, et al. Ranibizumab for neovascular age-related macular degeneration. N Engl J Med. 2006;355(14):1419-1431.

3. Day S, Acquah K, Lee PP, Mruthyunjaya P, Sloan FA. Medicare costs for neovascular age-related macular degeneration, 1994-2007. Am J Ophthalmol. 2011;152(6):1014-1020.

4. Badran M, Laher I. Type II diabetes mellitus in Arabic-speaking coun-tries. Int J Endocrinol. 2012;2012:902873.

5. McKay R, McCarty CA, Taylor HR. Diabetic retinopathy in Vic-toria, Australia: the Visual Impairment Project. Br J Ophthalmol. 2000;84(8):865-870.

6. World Health Organization. Priority Eye Diseases. http://www.who.int/ blindness/causes/priority/en/index7.html. Accessed: January 12, 2016.

7. Rudnicka AR, Kapetanakis VV, Jarrar Z, et al. Incidence of late-stage age-related macular degeneration in American whites: systematic re-view and meta-analysis. Am J Ophthalmol. 2015;160(1):85-93.e3.

8. Gudeman J, Jozwiakowski M, Chollet J, Randell M. Potential risks of pharmacy compounding. Drugs R D. 2013;13(1):1-8.

9. Michels S, Becker M, Wachtlin J, Binder S. The intravitreal injection: variations in regulations, cost and reimbursement in Europe. Spekt Augenheilkd. 2012;26(1):2-6.

10. Hussain A, Malik A, Halim MU, Ali AM. The use of robotics in sur-gery: a review. Int J Clinical Practice. 2014;68(11):1376-1382.

11. Yates DR, Vaessen C, Roupret M. From Leonardo to da Vinci: the his-tory of robot-assisted surgery in urology. BJU Int. 2011;108(11):1708-1713; discussion 1714.

12. Munnerlyn CR, Koons SJ, Marshall J. Photorefractive keratecto-my: a technique for laser refractive surgery. J Cataract Refract Surg. 1988;14(1):46-52.

13. Gupta PK, Jensen PS, de Juan EJ. Surgical forces and tactile per-ception during retinal microsurgery. Medical Image Computing and

Computer-Assisted Intervention (MICCAI’ 99). Morrisville, NC; 1999:1218-1225.

14. Jagtap AS, Riviere CN. Applied force during vitreoretinal microsur-gery with handheld instruments. Conf Proc IEEE Eng Med Biol Soc. 2004;4:2771-2773.

15. Maclachlan R, Becker BC, Tabarés JC, Podnar GW, Lobes LA, Riviere CN. Micron: an actively stabilized handheld tool for microsurgery. IEEE Trans Robotics. 2012;28(1):195-212.

16. Uneri A, Balicki MA, Handa J, Gehlbach P, Taylor RH, Iordachita I. New steady-hand eye robot with micro-force sensing for vitreoretinal surgery. Proc IEEE RAS EMBS Int Conf Biomed Robot Biomechatron. 2010;2010(26-29):814-819.

17. Bourla DH, Hubschman JP, Culjat M, Tsirbas A, Gupta A, Schwartz SD. Feasibility study of intraocular robotic surgery with the da Vinci surgical system. Retina. 2008;28(1):154-158.

18. Meenink H, Hendrix R, Naus G, et al. Robot-assisted vitreoretinal surgery. In: Gomes P, ed. Medical Robotics: Minimally Invasive Surgery. Woodhead Publishing, Cambridge, UK; 2012:185-209.

19. Nakano T, Sugita N, Ueta T, Tamaki Y, Mitsuishi M. A parallel ro-bot to assist vitreoretinal surgery. Int J Comput Assist Radiol Surg. 2009;4(6):517-526.

20. Kummer M, Abbott JJ, Kratochvil BE, Borer R, Sengul A, Nelson BJ. Octomag: An electromagnetic system for 5-DOF wireless microma-nipulation. IEEE Trans Robotics. 2010;26(6):1006-1017.

21. Bergeles C, Kratochvil BE, Nelson BJ. Visually servoing magnetic in-traocular microdevices. IEEE Trans Robotics. 2012;28(4):798-809.

22. Ullrich F, Bergeles C, Pokki J, et al. Mobility experiments with micro-robots for minimally invasive intraocular surgery. Invest Ophthalmol Vis Sci. 2013;54(4):2853-2863.

23. Souied E, Nghiem-Buffet S, Leteneux C, et al. Ranibizumab prefilled syringes: benefits of reduced syringe preparation times and less com-plex preparation procedures. Eur J Ophthalmol. 2015;25(6):529–534.

24. Daugman JG. High confidence visual recognition of persons by a test of statistical independence. IEEE Transactions on Pattern Analysis and Machine Intelligence. 1993;15(11):1149-1161.

25. Daugman JG, Uncertainty relation for resolution in space, spatial fre-quency, and orientation optimized by two-dimensional visual cortical filters. J Opt Soc Am A. 1985;2(7):1160-1169.

26. Golder L, Longchamp C, Gauch C, Tschöpe S, Rochat P, Schwab JL. Begleitstudie anlässlich der Einfuhrung von SwissDRG im Auftrag der FMH. http://www.gfsbern.ch/de-ch/Detail/begleitstudie-anlaesslich-der-einfuehrung-von-swissdrg. Accessed: June 1, 2016.

27. Kwoh YS, Hou J, Jonckheere EA, Hayati S A. Robot with improved absolute positioning accuracy for CT guided stereotactic brain sur-gery. IEEE Trans Biomed Eng. 1988;35(2):153-160.

28. Pitcher JD, Wilson JT, Tsao TC, Schwartz SD, Hubschman JP. Ro-botic Eye Surgery: Past, Present, and Future. J Comp Sci Syst Bio. 2012;S3:001.

29. Ballantyne GH. Robotic surgery, telerobotic surgery, telepresence, and telementoring. Surg Endosc. 2002;16(10):1389-1402.

30. Lanfranco AR, Castellanos AE, Desai JP, Meyers WC. Robotic sur-gery: A current perspective. Ann Surg. 2004;239(1):14-21.

31. Bergeles C, Yang GZ. From passive tool holders to microsurgeons: Safer, smaller, smarter surgical robots. IEEE Trans Biomed Eng. 2014;61(5):1565-1576.

32. Matthews JL, Dubovy SR, Goldberg RA, Flynn HW Jr. Histopathol-ogy of streptococcus mitis/oralis endophthalmitis after intravitreal injection with bevacizumab: a report of 7 patients. Ophthalmology. 2014;121(3):702-708.

33. Dharan S, Pittet D. Environmental controls in operating theatres. J Hosp Infect. 2002;51(2):79-84.

34. Nagy Z, Takacs A, Filkorn T, Sarayba M. Initial clinical evaluation of an intraocular femtosecond laser in cataract surgery. J Refract Surg. 2009;25(12):1053-1060.