Getting the most from your Agilent Instruments: New Advancements in OpenLAB CDS Instrument Control

OpenLAB CDS

Steve Miller October 2013

New Release: OpenLAB CDS A.01.05

7890B Series GC Support• All New 7890B Series GC Driver

[CS] [EE] • 5890 Re-instated Patch no longer required

Customer Experience • Single Sample Method Loading [CS]• Acquisition Method Viewer: Review methods

without loading a method to the instrument and without triggering method resolution [CS]

• Improved handling of ALS errors• Agilent Parts Finder• Integrate Intelligence

[CS] = ChemStation Edition[EE] = EZChrom Edition

Deployment • Better migration support for mixed systems

ChemStation rev. B/ OpenLAB CDS ChemStation Edition on one ECM [CS]

• Better migration support allows "mixed version" clients on the latest OLSS server

• Mixed topology of Networked Workstations on a Distributed System (AIC)

Page 2

INSTRUMENT CONTROL

Page 3

Different Software, same drivers

Page 4

Superior GC Customer ExperienceIntegrated Intelligence: A Smarter GC System

Integrated Intelligence Customer Benefit

Integrated GC calculators Develop methods more easily

Barcode Scanning with Auto-Input Easier configuration

Integrated early-maintenance-feedback Manage assets more efficiently

3D Interactive Consumable and Parts Finder Find parts faster & order moreeasily

Sleep-wake functionality Lower cost of ownership

Page 5

New 7890B Driver

Agilent Early Maintenance Feedback

Page 6

Agilent Early Maintenance Feedback

• Set-up the Counters within the Agilent Data Systems

• Based on the configuration of the GC

• User-defined early maintenance feedback counters

• Notification of the Warning and Service Due

Page 7

Agilent Early Maintenance Feedback

• Some EMF counters have default limits; however the maintenance required will depend on your conditions

Another way to set them up:• When you perform

maintenance look at the settings of the counters.

• Set the warning limits to the amount.

• Check the counters next time when performing maintenance and adjust the limits.

Page 8

EMF Counters appear in Instrument Actuals and Reports

Instrument actuals are displayed next to set-points in the data system.

Page 9

Sleep/Wake Mode

Page 10

Resource Conservation: Sleep/Wake Mode

• Schedule the time of the Wake and Sleep Methods either as “custom”, “7 days a week” or “Sat/Sun”

• Manually put the instrument to sleep by clicking on “Sleep” on the “Instrument” menu

• Instrument will still go to Sleep or Wake mode even if the data system is not connected for any reason.

Page 11

Integrated Method Calculators

Page 12

Problem / Solution

Problem:

Helium is way too expensive and I need to change my method to Hydrogen

Solution:

Agilent Integrated Method Calculators

Method Translation Calculators

Page 13

Barcode Scan to Method

Page 14

Barcode Scanning Input

Liners

Columns

*Also syringes

Inventory and full Agilent catalog

Periodic updates available

Barcode in the columns, liners, and syringes right into the Agilent

data system method

Page 15

Autosampler messed up last night!

Page 16

Handling of ALS Errors

Page 17

Agilent Parts Finder Tool

Page 18

Agilent Parts Finder

• Quickly locate the part number and add to the parts list or favorites

• Print the parts list or save it to a file

• Links to the Agilent.com store for quick ordering through the Agilent website.

• Update file with up-to-date parts and part numbers available via the Agilent website for download.

Page 19

Agilent Parts Finder Tool

Page 20

OpenLAB Data Analysis

Page 21

Data Analysis Preview

Page 22

Improved Data Navigation

Function Ribbon

“Flat” intuitive User InterfaceIntegrated

Intelligent Reporter

Agilent Parts Finder Tool

Page 23

Agilent Parts Finder Tool

Page 24

Intelligent Reporter

Compound classes reported by weight and weight % using custom calculations

Normalized results according to ASTM formulas using custom calculations

Sample Information

Sample chromatogram scaled to show only peaks of interest

Compounds reported by weight and weight % using custom calculations

Custom Biodiesel Report

Page 25

Agilent Parts Finder Tool

Page 26

Agilent Parts Finder Tool

Page 27

Agilent Parts Finder Tool

Page 28

The new drivers and what versions of EZChrom and ChemStation they work with

7890 Series: Feature Availability from GC SW*

Platform

GC

OpenLAB CDSA.01.05

OpenLAB CDSA.01.04 and

Earlier

ChemStation B.04.03 SP1

EzChromElite

3.3.2 SP2

EzChrom Elite 3.3.2 SP1

GC Driver5.01

GC Driver4.02 [043]

GC Driver4.02 [31]

GC Driver3.02.006

GC Driver2.12.042

7890B7890B

HW/SWFeatures

7890AHW/SWFeatures

7890AHW/SWFeatures

7890AHW/SWFeatures

7890AHW/SWFeatures

7890A with G3444A board

upgrade

7890BIntegrated

Intelligent SW Features, Ability to

Add the FPD+

7890AHW/SWFeatures

7890AHW/SWFeatures

7890AHW/SWFeatures

7890AHW/SWFeatures

7890A

7890AHW/SWFeatures

PLUS• Parts Finder• Calculators

7890AHW/SWFeatures

7890AHW/SWFeatures

7890AHW/SWFeatures

7890AHW/SWFeatures

*Functionality is limited to 7890A configurations and SW functionality only supported by the specific SW platform with 7890A.

Page 29

7890A Enhancements Kit, Other

7890A Logic Board Enhancements Kit (G3444A)- Makes 7890A GC ready for most (but not all) 7890B GC enhancements. For example:• “Integrated Intelligence” features:- Early Maintenance Feedback (EMF), Sleep/Wake modes - Method Translation Calculator (via CDS)- Enable 2-way communication – 7890A/5977A

• Serial Port (not “RS-232”):- 2nd communication port for Remote Advisor- Optional Bar Code Reader connects directly to GC

[also a version to connect to USB/PC] • FPD Plus

7890B Only (Cannot add this to 7890A via G3444A, or otherwise)• 2nd Multimode Inlet (MMI) mounted on 7890B GC• 3rd detector as uECD - developing IDO countries for enviro/pesticides (non-GC/MS)• Factory installed Backflush (mid-column/PUU), Option #204 (est. CPL Mar’13)• Removable GC Oven Door Accessory (primarily for GC/MS users)

Page 30

RC.Net Drivers for additional LC modules

This release of the RC.Net LC & CE Driver allows configuring Agilent 1260 Prep LC Systems• G1361A - 1260 Isocratic Prep Pump

– Prep Pump Cluster of up to 4x G1361A 1260 Isocratic Prep Pumps• G1364A Fraction Collector

Fraction Collector Cluster of up to 3x G1364A/B/C or G5664A Fraction Collectors AND additionally 1x G1364A/B/C or G5664A as Recovery

Fraction Collector• G2258A - 1260 Dual Loop Autosampler• G2260A - 1260 Preparative Autosampler

Check supported modules and firmware document to see what is supported.

Page 31

CHEMSTATION EDITION

Page 32

Acquisition Method Viewer ChemStation Edition

ProblemI need to review the Acquisition Method parameters before downloading to the instrument. I do not want to trigger Method Resolution.

SolutionAcquisition Method Viewer-provides a way to view the acquisition method parameters.

Page 33

Single Sample Workflow (1/2)ChemStation Edition

• The Navigation Table now shows the acquisition method as well as the data analysis method.

• A fly-over indicates the path to the data analysis method.

Page 34

Single Sample Workflow (2/2)ChemStation Edition

• In the Last Result Mode it is now possible to save the data analysis parameters last used (from DA.M) to a master method.

• Update any master method with the data analysis parameters last used (from DA.M).• Update the corresponding master method with the data analysis parameters last used (from

DA.M).• Combine the data analysis parameters last used (from DA.M) with acquisition parameters from

any master method and save as a new master method.

Page 35

EZCHROM EDITION

Page 36

OpenLAB CDS EZChrom Edition A.04.05

• Online instrument will always remain in Monitor mode. When “Result Set” is opened at the end of Single/Sequence run the Instrument will remain in Monitor mode.

• In “Result Review Mode” when a user opens a method, sample prep, sequence, or data file, the last opened method and sequence file, if available, will be kept open/opened rather than being set to un-assigned while closing the “Result Set”.

• Display and positioning of windows for instrument setup values and instrument curves is now persisted.

• Option to select Local Time or UTC in exported reports.• Display instrument curves on IR report.• Update signal displayed in the online plot automatically whenever a method is

downloaded to the instrument.• API support for automation of LC driver: Set instrument parameters from a 3rd party

application (for e.g., flow rate, solvent composition) for Agilent LC driver.

Page 37

OpenLAB CDS EZChrom Edition A.04.06

•The Automation Tool Kit will now work with old applications like EZLink!

• Agilent will be wanting to test this new release in the January/February time frame with companies that can’t upgrade because of the Tool Kit.

• Contact myself and I will get you on the list!• THIRD PARTY DRIVER UPDATE• Shimadzu Drivers Coming very soon! Classical C++ Drivers for 2010 and series 10

instruments.• New Shimadzu drivers already released.• Thermo available soon for Trace Ultra• Focus GC, Trace 1300, TriPlus, TriPlus RSH, AI/AS 3000, AI/AS 1310• PE Drivers available from VWR

Page 38

AIC Virtualization: EZChrom Edition

• Test and document topologies where parts of the system are deployed under VMware (either VMware workstation or VMware ESX).

Topology • EZChrom Edition Distributed

System with ECM• Clients on a VMware ESX server,

AIC on physical machines• Support statement: Supported

and recommended

Page 39

ECM or Data Store

DEPLOYMENT

Page 40

Better migration support:ChemStation rev. B/ rev. C on one ECM

ChemStation rev. B/ OpenLAB CDS ChemStation Edition on one ECM • Support of ChemStation rev. B and OpenLAB CDS ChemStation Edition on

one ECM system: With ChemStation rev. B and OpenLAB CDS rev. C on one ECM system, document limitations, e.g., the ability to reprocess ChemStation rev. B data in ChemStation rev. C, but not vice versa.

Mixed version clients on the latest OpenLAB Shared Services server• EZChrom: Software upgrades in mixed environments: In a Distributed or

Networked environment, allow a "mixed version" of clients running with the latest OpenLAB Shared Services server to minimize downtime during an upgrade.

Page 41

Mixed topology Networked Workstation / Distributed System [ChemStation]

• Test and document an OpenLAB CDS ChemStation Edition system mixing Networked Workstation and AIC on one OpenLAB Shared Services server.

• One use case for such a mixed system is to do the data acquisition on the networked workstation, e.g. with instruments that are not supported on the AIC (e.g., GPC, ELSD, GPIB instruments) and then use the AIC for data reprocessing.

• Another use case is data acquisition and reprocessing for some instruments (e.g. GPC, ELSD, GPIB instruments, …) on the networked workstation, and data acquisition and reprocessing for other instruments on the AIC.