agilent 1290 infi nity binary lc with iset – emulation of

8
Abstract Transfer of conventional methods to UHPLC systems typically needs a change of the method, due to the smaller system delay volume and gradient mixing behavior of these systems. In most cases, this is done by installing an additional delay volume or by adding an isocratic hold step at the beginning of the run. This Application Note shows the emulation of the Waters Alliance 2695. The transfer of the analysis of analgesics onto the Agilent 1290 Infinity Binary LC using Intelligent System Emulation Technology (ISET) is shown, based on a Waters Application Note. Using ISET to enable a method transfer through simple mouse clicks was evaluated. The agreement of retention times and resolution was determined. Agilent 1290 Infinity Binary LC with ISET – Emulation of the Waters Alliance 2695 LC System Analyzing Analgesics Application Note Small Molecule Pharmaceuticals & Generics Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany min 3 4 5 6 7 8 9 AU 0 25 50 75 100 125 150 175 Waters ACQUITY UPLC H - class Agilent 1290 Infinity Binary LC w/o ISET Alliance

Upload: others

Post on 07-Jun-2022

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

AbstractTransfer of conventional methods to UHPLC systems typically needs a change of the method, due to the smaller system delay volume and gradient mixing behavior of these systems. In most cases, this is done by installing an additional delay volume or by adding an isocratic hold step at the beginning of the run.

This Application Note shows the emulation of the Waters Alliance 2695. The transfer of the analysis of analgesics onto the Agilent 1290 Infi nity Binary LC using Intelligent System Emulation Technology (ISET) is shown, based on a Waters Application Note. Using ISET to enable a method transfer through simple mouse clicks was evaluated. The agreement of retention times and resolution was determined.

Agilent 1290 Infi nity Binary LC with ISET – Emulation of the Waters Alliance 2695 LC System Analyzing Analgesics

Application NoteSmall Molecule Pharmaceuticals & Generics

AuthorA.G.HuesgenAgilent Technologies, Inc.Waldbronn, Germany

min3 4 5 6 7 8 9

AU

0

25

50

75

100

125

150

175

Waters ACQUITY UPLC H - class Agilent 1290 Infinity Binary LC w/o ISET Alliance

Page 2: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

2

IntroductionSeamless instrument-to-instrument method transfer is often a strong demand for most industries because changing established methods is expensive and time consuming, and comparison with reference data is diffi cult. Nevertheless, old equipment has to be replaced occasionally. Seamless method transfer from the Agilent 1100 or 1200 Series LC System to the Agilent 1260 or 1220 Infi nity LC System is possible without the need to change the chromatographic conditions1,2. Method transfer to an Agilent 1290 Infi nity Binary LC system is possible by activating the ISET with a simple mouse click

The Agilent 1290 Infi nity Binary LC, in combination with ISET, offers the possibility to emulate older non-Agilent instrumentation such as the Waters Alliance LC systems. A conventional method from the Waters Alliance LC system was transferred to the 1290 Infi nity Binary LC without ISET and onto a Waters ACQUITY UPLC H-Class. In a second step, the method was transferred onto the 1290 Infi nity Binary LC enabling ISET. The same results regarding retention times and typically better resolutions were received.

ExperimentalThe following instruments were used (Table 1).

Agilent 1290 Infi nity Binary LC

Waters ACQUITY UPLC H-Class Waters Alliance

Binary Pump G4220A Quaternary Pump 2695Autosampler G4226A AutosamplerALS cooler G1330B Autosampler coolerColumn compartment G1316C Column compartmentDiode array detector G4212A VWD (TUV) VWD (Dual absorbance

Detector)

Table 1. Instrumentation used.

O

HN

CH3

HO

CH3

CH3

H3C

N

NN

N

O

O

HN CH

3

OH

O

NH

H3C

OCH

3

OH

O

O

O

NH

OO

Paracetamol/acetaminophen Caffeine 2-Acetamidophenol

Acetanilide Acetylsalicyclic acid Phenacetin

Analyzed compounds

Chromatographic conditionsColumn Atlantis T3, 4.6 × 150, 5 µm Mobile phases A= 0.1 % formic acid pH = 2.3, B = ACNGradient at 0 minutes 15 % B,

at 10 minutes 60 % B, at 11 minutes 15 % B, at 15 minutes 15 % B

Flow rate 1mL/minInjection volume 10 µLColumn temperature 30 °CDetection 260 nm, 5Hz

Page 3: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

3

Figure 2. Overlay of chromatograms of the Waters Alliance 2695, an Agilent 1290 Infi nity Binary LC without ISET, and the Waters ACQUITY UPLC H-Class.

Selection of manufacturer

Selection of instrument

Selection of mobile phases

Selection of autosampler configuration

Figure 1. Selection of the instrumentation to be emulated.

Acquisition and evaluation software OpenLAB CDS ChemStation version C.01.04 and ISET

Figure 1 shows the parameter that must be entered in the ISET parameter screen of the 1290 Infi nity Binary LC to select the instrument to be emulated.

Results and DiscussionThe following experiments were done to prove the seamless method transfer from the Waters Alliance 2695 LC system to the Agilent 1290 Infi nity Binary LC in combination with ISET.

Experiments• Analysis of analgesics on the

Alliance 2695, based on a Waters Application Note3

• Transfer of the developed method onto the 1290 Infi nity Binary LC without applying ISET and onto a Waters ACQUITY UPLC H-Class

• Analysis of analgesics on the 1290 Infi nity Binary LC enabling ISET

Evaluated parameters• Determination of the deviation of

retention times, specifi ed deviation is < ± 5 %

• Determination of the resolution, typically better on the 1290 Infi nity Binary LC. Specifi ed maximum deviation < –5 %

Transfer of method to the Agilent 1290 Infi nity Binary LC without ISET and onto the Waters ACQUITY UPLC H-ClassFigure 2 shows the obtained chromatograms for the Water Alliance 2695 and both UHPLC systems overlaid. As expected, both UHPLC systems showed less retention for all peaks due to its signifi cant lower delay volumes, (Figure 2).

min3 4 5 6 7 8 9

AU

0

25

50

75

100

125

150

175

Waters ACQUITY UPLC H - class Agilent 1290 Infinity Binary LC w/o ISET Alliance

Page 4: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

4

Figures 3 and 4 show the comparison of the deviation of retention times and resolution. The retention times deviated from the original value signifi cantly more than –5 % for both instruments.

The deviation of resolution was within the –5 % limit for both systems.

-15-13-11-9-7-5-3-1135

Devia

tion

(%)

Deviation of retention times

Agilent 1290 Infinity Binary LC w/o ISETWaters ACQUITY UPLC H-Class

Acetam

inoph

en

Caffein

e

2-Acet

amido

pheno

l

Acetani

lide

Acetyls

alicylic

acid

Phenace

tin

Figure 3. Deviation of retention times of the Agilent 1290 Infi nity Binary LC without ISET and the Waters ACQUITY UPLC H-Class.

Figure 4. Deviation of resolution of the Agilent 1290 Infi nity Binary LC without ISET and the Waters ACQUITY UPLC H-Class.

Caffein

e

2-Acet

amido

pheno

l

Acetani

lide

Acetyls

alicylic

acid

Phenace

tin-10

-5

0

5

10

15

20

25

30

35

Devia

tion

(%)

Deviation of resolution

Agilent 1290 Infinity Binary LC w/o ISETWaters ACQUITY UPLC H-Class

Page 5: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

5

Transfer of conventional method onto the Agilent 1290 Infi nity Binary LC enabling ISETThe best correlation was obtained using the 1290 Infi nity Binary LC with ISET and fi ne tuning, see the violet trace (Figure 6). Fine tuning means that an additional delay volume of 200 µL was added in the ISET parameter screen by enabling the manual fi ne tuning parameters (Figure 5). Without the fi ne tuning step, the agreement of retention times using the 1290 Infi nity Binary LC was already close, see the red trace (Figure 6).

200

80

Adding or substracting delay volume

Figure 5. Fine tuning parameters within ISET.

min3 4 5 6 7 8

mAU

0

50

100

150

200

Agilent 1290 Infinity Binary LC with ISET + fine tuning (+ 200 µL) Alliance Agilent 1290 Infinity Binary LC with ISET

Figure 6. Transfer onto the Agilent 1290 Infi nity Binary LC with ISET and with ISET and Fine Tuning.

Page 6: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

6

Figure 7 summarizes the retention time deviations obtained on the different 1290 Infi nity Binary LC confi gurations. The allowed deviation of retention times for the 1290 Infi nity Binary LC with ISET is ± 5 %. Optimum results were achieved using the ISET tool with fi ne tuning. With ISET and no fi ne tuning, only the fi rst peak was slightly out of the specifi cation. The resolution had improved for all peaks.

Figure 8 compares resolution data. Using ISET and ISET with fi ne tuning, resolution had improved for all peaks.

Agilent 1290 Infinity Binary LC with ISET

Agilent 1290 Infinity Binary LC with ISET and Fine Tuning

Acetam

inoph

en

Caffein

e

2-Acet

amido

phen

ol

Acetani

lide

Acetyls

alicylic

acid

Phenace

tin-6

-4

-2

0

2

4

Devia

tion

(%)

Deviation of retention times

Figure 7. Deviation of retention times as percentage for all used Agilent 1290 Infi nity Binary LC confi gurations.

Caffein

e

2-Acet

amido

pheno

l

Acetani

lide

Acetyls

alicylic

acid

Phenace

tin

Agilent 1290 Infinity Binary LC with ISET

Agilent 1290 Infinity Binary LC withISET and Fine Tuning

0

5

10

15

20

25

30

35

40

45

Devia

tion

(%)

Deviation of resolution

Figure 8. Deviation of resolution as percentage for all used Agilent 1290 Infi nity Binary LC confi gurations.

Page 7: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

7

ConclusionMethod transfer using the Agilent 1290 Infi nity Binary LC in combination with the ISET enables the emulation of older non-Agilent LCs, such as the Waters Alliance LC system, in a time-saving and convenient way. There is no need to change the original method by adding the additional delay volume of an isocratic hold step. With a few mouse clicks, the ISET is enabled and allows seamless method transfer. In this example, the deviation of retention times was < 4 % using the 1290 Infi nity Binary LC with ISET and fi ne tuning. The allowed deviation of retention times is < ± 5 %. The resolution was improved for all peaks. The allowed maximum deviation for the resolution is < –5 %.

References1. A.G. Huesgen, ”Seamless method

transfer from the Agilent 1200 Series LC System to the Agilent 1260 Infi nity LC System, Comparison of system suitability results according EP and USP regulations for the analysis of paracetamol (acetaminophen) and its impurities”, Agilent Application Note, Publication Number 5990-6207EN, September 1, 2010.

2. A.G. Huesgen, “Seamless method transfer from the Agilent 1200 Series to the Agilent 1220 and 1260 Infi nity LC systems, Comparison of system suitability results according to USP method USP/NF 23 for the analysis of Paracetamol (Acetaminophen) and Aspirin in pain relievers”, Agilent Technical Overview, Publication Number 5990-6206EN, July 1, 2010.

3. Waters Application note, “Analysis of analgesics using Atlantis T3“, WA60217 KK PDF, 2009.

Page 8: Agilent 1290 Infi nity Binary LC with ISET – Emulation of

www.agilent.com/chem/ISET

This information is subject to change without notice.

© Agilent Technologies, Inc., 2013Published in the USA, September 1, 20135991-2792EN