10{ nmr hardware - university of british columbia · 2013. 4. 10. · 10{ nmr hardware in this...
TRANSCRIPT
'
&
$
%
10– NMR Hardware
In this lecture, we will begin to familiarise ourselves
with the more practical aspects of NMR - i.e. major
components of a spectrometer, with probeheads, and
with the definition of signal-to-noise.
10.1 The spectrometer
A spectrometer consists of the following components:
superconducting coil
shim coil
probehead
filter
shim power supply
receiver
computer
1H amplifier
13C amplifier
1H frequency
13C frequency
HX
pulse programmer
'
&
$
%
i) The magnet
The magnet, which provides the ~B0 field needed for
precession to take place, is a superconducting
magnet. This means it is made out of
superconducting material (e.g. Bi-Sr-Ca-Cu-O based
superconductors) that has a minimal resistance at
0K (-273.15 oC). These temperatures can be
achieved by immersing the superconducting material
in liquid helium.
The superconducting coil is made out of wire which
is several miles long and wound into a multi-turn
solenoid. The wire itself is made out of different
materials, arranged as illustrated below. The reason
for this arrangement is to minimize the chances of a
quench if there should be material failure in the coil.
Superconducting Material
Insulating Material
'
&
$
%
To keep the coil cold, it is immersed in a dewar
containing liquid helium and kept at 4.2 K. Some
magnets have a pump built-in so that the helium can
be supercooled. This is used to reduce the resistance
in the coil further, allowing higher B0 fields to be
achieved.
In order to slow down the evaporation of helium, the
dewar is surrounded by dewars of nitrogen, as shown
in the diagram below:
'
&
$
%
Used with permission, JEOL USA, Inc.
Used with permission, JEOL USA, Inc.
ref: http://www.cis.rit.edu/htbooks/nmr/chap-7/chap-7.htm
'
&
$
%
The outer vacuum region is filled with many layers of
reflextive mylar film, which serves to insulate the
magnet by diminishing the amount of heat which can
enter the helium region.
ii) The shim coils
The shim coils, which are situated inside the bore of
the magnet, provide compensatory magnetic fields,
such that the net ~B0 field is spatially homogeneous.
Inhomogeneities in the ~B0 field arise from the way in
which the magnet is designed, from materials in the
probe (e.g. metallic objects in proximity to the RF
coil in the probe), from the sample tube, from
sample permeability, and from ferromagnetic
materials around the magnet.
As a result of the inhomogeneities, a field gradient
may exist across the sample, e.g.
'
&
$
%
The shim coils consist of wire through which current
is passed so that small magnetic fields are generated.
These coils have particular shapes which correspond
to particular functional forms. Some typical shim
coils and their corresponding functional forms are:
'
&
$
%
Z0
Z1
Z2
'
&
$
%
X
XZ
'
&
$
%
Y
Y Z
'
&
$
%
XZ3
ref: http://www.cis.rit.edu/htbooks/nmr/chap-7/chap-7.htm
Shimming consists thus in finding the optimal
current settings such that the ~B0 field is
homogeneous. This can be determined by either
maximizing the lock signal (see below) or by
maximizing the size of the free induction decay
(FID) of, for example, water.
Because the linewidths observed in the solid state are
typically broader than in solution state, it is not
necessary to shim very precisely. This is often why
most people only optimize the lower order shims (Z0,
Z1, Z2, X, XZ, Y , and Y Z) and do not modify the
'
&
$
%
higher order shims (e.g. Z4,Z5,XZ2,XZ3,...).
iii) The lock
The lock is needed to compensate for external drifts
in the magnetic field (e.g. from a tram, metallic
trolleys). In solution state, it is very common and
works by using the signals arising from deuterated
solvents (in the sample) which are then monitored by
the computer. The signal is “locked” on to the
frequency of something like D2O and monitored
constantly during the course of the experiment.
Small changes are compensated for by a coil in the
shim stack.
'
&
$
%
In solid state, the lock isn’t as commonly used, but
can be useful, as illustrated here for a 13C line of
adamantane, spinning at 4 kHz. For some solids
probe assemblies, the lock is situated in the duct
used to eject the MAS rotors. The deuterium
sample, in this case, sits close to the RF coil but is
not surrounded by it.
iv) The probehead
The probehead contains an RF coil which produces
the ~B1(t) magnetic field needed to perturb the spins
'
&
$
%
away from equilibrium. The major component
within the probe is a resonance circuit, built from
inductors and capacitors which allow the probe to be
tuned and matched to a particular frequency
(Larmor frequency of the nuclei to be observed
and/or excited).
We saw in Chapter 2, that a probe consists of a
slightly more complex version of an LC circuit. In
our circuit, we said last time that the inductor L is
the coil where the sample is placed. There are a
number of different coil geometries which are used.
In solution, standard coils are the saddle coil:
and Helmholtz coils
'
&
$
%
whereas, in solids, a solenoid or flat coil is used. The
choice on geometry depends on power applied to the
coil, filling factor, and homogeneity.
'
&
$
%
v) The frequency generator
The frequency generator, also known as the
synthesizer, generates a sinusoidal modulation, which
has a frequency close to the Larmor frequency of the
nucleus to be excited/perturbed. The base frequency
generated is called the carrier frequency or reference
frequency, such that the synthesizer output is defined
as
s ≈ cos(ωref t + φ(t)) (10.1)
where φ(t) is the phase of the radiofrequency
modulation.
vi) The pulse programmer
The pulse programmer “interprets” the pulse
program that the user writes. It controls how long a
pulse is on for (i.e. gating), the phase of the pulse
and the shape of the pulse (e.g. rectangular,
Gaussian,...).
'
&
$
%
M.H. Levitt’s book, p. 74-75
vii) The amplifiers
The amplifiers increase the amplitude of the output
of the frequency generator so that the final
amplitude of the RF pulses ranges from milliwatts
(no amplification) to kilowatts.
The output of an amplifier is characterised by its
peak-to-peak voltage (Vpp) or power in Watts (P), i.e.
P =Vpp
2
400∗ 10
dB
10 (10.2)
where Vpp is the voltage measured on the oscilloscope
and dB is the attenuation used.
'
&
$
%
Just as a probehead can be tuned to a specific
frequency, so can an amplifier. To tune an amplifier,
the following procedure can be used:
1. Connect the output of the amplifier into an
attenuator (e.g. 40dB). The output of that can
now be connected to an oscilloscope for
observation.
2. Start pulsing using a single pulse, with an easy
to observe pulse width (e.g. 1-2ms). Use a power
level setting which is in the middle of the
available range.
WARNING!: Beware of connecting high
power outputs directly to the oscilloscope!
3. Most amplifiers have a button for ”tuning” and
one for the ”amplifier load”. Change the tuning
and load so that the maximum amplitude
(largest Vpp) value is achieved.
vii) Filters
Although the probehead behaves as a filter for a
particular frequency that is to be excited and/or
observed, there are often spurious frequencies still
'
&
$
%
present in the lines leading up to the receiver. To
eliminate these unwanted frequencies a number of
filters can be used, namely low-pass filters, high-pass
filters, band-pass filters, or band-stop filters. A
low-pass filter allows low-frequency signals to be
transmitted, while attenuating higher frequencies. A
high-pass filter behaves in an opposite manner in
that it lets through high frequencies, while
attenuating low frequencies. A band-pass or
band-stop do not attenuate or attenuate a range of
frequencies, respectively. Schematically,
'
&
$
%
ref: R. Ludwig and P. Bretchko, RF Circuit Design: Theory and
Applications, Prentice Hall, NJ, 2000.
A low-pass filter consists of a circuit containing a
resistor R and a capacitor C. A high-pass filter
consists of a circuit containing a resistor R and an
inductor L. And a bandpass filter consists of an RLC
circuit.
For NMR purposes, the best is always to use a
bandpass filter since only the frequencies to be
excited/observed are let through. These filters are
more expensive though and so are often replaced by
a low-frequency filter or a high-frequency filter.
viii) The receiver
As with a radio, a NMR spectrometer is equipped
with a receiver to detect the RF (or magnetization)
coming from the sample in the probehead. Typically
it consists of a quadrature detector, which is a device
which separates the Mx component of the
magnetization from the My component.
Electronically, this is done with a doubly balanced
mixer. For more details cf. R. Ludwig and P.
Bretchko, RF Circuit Design: Theory and
'
&
$
%
Applications, Prentice Hall, NJ, 2000.
ix) Co-axial cables
Connecting all the different components of a
spectrometer are co-axial cables. These cables
consist of:
1. an inner cylindrical conductor of radius a,
2. surrounded by a dielectric material, such as
polystyrene, polyethylene or teflon,
3. an outer conductor, which is grounded to
minimize radiation loss and field interference
4. and an outer insulator.
To avoid losses between the different components of
'
&
$
%
the spectrometer, it is important to use good quality
co-axial cable (usually the thicker kind).
x) The computer
The computer controls most of the components listed
above and acts as an interface between the user and
the hardware.
10.2 Signal-to-Noise Ratio
When we discussed probes in Chapter 2, we
introduced the concept of the sensitivity of the
probehead. A measure of this sensitivity in terms of
the electronics is given by the quality factor, Q.
There is another measure of sensitivity, known as the
signal-to-noise ratio, given in terms of the observed
magnetization.
In a NMR experiment, the signal is given by
S ∝ ω0,xMxtot (10.3)
where ω0,x is the Larmor frequency and Mxtot is the
magnetization detected in the receiver.
'
&
$
%
The noise, known as thermal or Johnson noise,
coming from the electronics, is
N ∝ keff,x√
ω0,x (10.4)
with keff,x is a proportionality constant which takes
into account the Q of the circuit and other electronic
parameters of the detection devices.
Therefore the signal-to-noise ratio is given by
S
N= keff,x
√ω0,xMxT2,x (10.5)
where T2,x is the transverse relaxation time.
Using Curie’s law for the magnetization for a spin
1/2,
Mx =Nxγx
2h2B0
4(2π)2kT, (10.6)
the signal-to-noise ratio can be rewritten as
S
N=
keff,xNxγx5/2h2B0T2,x
4(2π)2kT. (10.7)
'
&
$
%
This latter equation shows that one can expect
improved sensitivity if:
1. the probe has a high Q (keff,x is high);
2. there is a lot of sample (Nx large);
3. the spins observed have a high γ value;
4. the magnetic field is large (e.g. 800 MHz vs 400
MHz);
5. the temperature at which the experiment is
carried out is low.