© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
AV Synchrony Restored
RYTHMIQ operates in AAI(R) pacing mode with VVI backup during times of normal conduction, switching to
DDD(R) mode when a conduction block is detected
Loss of AV
Synchrony
OperationRYTHMIQ™
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
• The device provides AAI(R) at the LRL and/or sensor indicated rate with backup VVI pacing at a rate of 15 bpm slower than the LRL
• When there is good conduction, ventricular pacing does not occur as the VVI backup mode runs in the background
During normal conduction, RYTHMIQ operates like two pacemakers running concurrently – one in AAI(R)
mode and one in VVI mode.
Normal Conduction
OperationRYTHMIQ™
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
The device monitors for loss of AV synchrony to determine when to switch to DDD(R) mode
Loss of AV Synchrony
• If 3 slow ventricular beats are detected in a window of 11 beats, then the device automatically switches to DDD(R) mode.
• A slow beat for RYTHMIQ is defined as a ventricular pace or ventricular sensed event that is at least 150 ms slower than the AAI(R) pacing rate.
Loss of AV Synchrony
OperationRYTHMIQ™
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
#1 #2 #3
INGENIO™ records a RYTHMIQ Episode along with 20 seconds of electrogram data when RYTHMIQ
detects loss of AV synchrony.
This can help the clinician to diagnose patient rhythm conduction disorders and determine appropriate algorithms and pacemaker settings
OperationRYTHMIQ™
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
If AV Search+ remains in AV hysteresis for at least 25 cardiac cycles, and less than 2 of the last 10 cycles are ventricular paced, then the device automatically switches the pacing mode back to AAI (R) with VVI backup.
The device uses AV Search+ to periodically check for a return of intrinsic conduction and determine when to mode switch back to AAI(R) with VVI backup.
Normal Conduction Restored
AV Synchrony Restored
OperationRYTHMIQ™
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
AV Synchrony Restored
RYTHMIQ is designed to adapt to a patient’s conduction disorder, by mode switching when AV synchrony is lost or
restored, without clinically significant pauses
OperationRYTHMIQ™
Loss of AV
Synchrony
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
If RYTHMIQ went directly from AAI(R) into ATR Mode Switch, when the mode switch ends, it goes directly back
into AAI(R) mode
Exception During ATR Mode SwitchRYTHMIQ™
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
RYTHMIQ is designed to minimize unnecessary pacing without clinically significant pauses
RYTHMIQ operates like two pacemakers running concurrently—one in AAI(R) and one in VVI
RYTHMIQ does not require dropped ventricular beats to switch to DDD(R) pacing
Reference Guide, INGENIO Pacemaker, Page 2-53, 358577-001 EN US 11/10
OperationRYTHMIQ™
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
Patients may require different algorithms at various stages of their conduction disease. RYTHMIQ adds another RV pacing algorithm
to Boston Scientific devices, because no one algorithm is appropriate for every patient.
AV Search+ is designed to reduce unnecessary RV pacing by providing backup ventricular pacing for every beat while searching for intrinsic activity
Managing RV Pacing – Two AlgorithmsRYTHMIQ™
BSC also offers AV Search+
© 2012 by Boston Scientific Corporation or its affiliates. All rights reserved. CRM-59709-AA
Patient’s AV Conduction
Status
Boston Scientific1 Medtronic2 St. Jude3
Normal RYTHMIQ™
MVP®Nominally ON for all patients
Search AV™
VIP™
Only available algorithm
1st Degree Block RYTHMIQ™
2nd Degree Block AV SEARCH+
3rd Degree Block5
Fixed AV Delay
1) Therapy recommendations made by Indications-Based Programming2) Source: Medtronic Adapta/Versa/Sensia Pacemaker Programming Guide (M929948A001C 2008-01-18)3) Source: St. Jude Medical Unity Platform Reference Guide (60020476A, 07/09)Boston Scientific Indications-Based Programming would recommend paced/sensed AV delays. While not an algorithm, an equivalent setting would be available in competitor devices
With both RYTHMIQ™ & AV Search+, clinicians now have options to appropriately manage RV pacing in patients with varying degrees of conduction block
INGENIO™ pacemakers have two algorithms to help minimize unnecessary RV pacing
Managing RV Pacing – Two AlgorithmsRYTHMIQ™