wolstenholme chpt 1
TRANSCRIPT
-
AUGER ELECTRON SPECTROSCOPY
-
AUGER ELECTRON SPECTROSCOPY
Practical aPPlication to Materials analysis and characterization of
surfaces, interfaces, and thin filMs
JOHN WOLSTENHOLME
MOMENTUM PRESS, LLC, NEW YORK
-
Auger Electron Spectroscopy: Practical Application to Materials Analysis and Characterization of Surfaces, Interfaces, and Thin Films
Copyright Momentum Press, LLC, 2015.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means electronic, mechanical, photocopy, recording, or any otherexcept for brief quotations, not to exceed 400 words, without the prior permission of the publisher.
First published by Momentum Press, LLC222 East 46th Street, New York, NY 10017www.momentumpress.net
ISBN-13: 978-1-60650-681-3 (print)ISBN-13: 978-1-60650-682-0 (e-book)
Momentum Press Materials Characterization and Analysis Collection
Cover and interior design by Exeter Premedia Services Private Ltd., Chennai, India
10 9 8 7 6 5 4 3 2 1
Printed in the United States of America
-
abstract
Auger electron spectroscopy (AES) is capable of providing elemental composition and, in some restricted cases, chemical bonding information for the elements present near the surface of solid materials. The surface specificity of this technique is such that only atoms in the top 5 to 10 nm are detected. The great strength of AES is its ability to provide this infor-mation with excellent spatial resolution (down to
-
contents
List of figures ix
List of tabLes xxi
foreword xxiii
Preface xxv
acknowLedgments xxvii
chaPter 1 introduction 1
chaPter 2 the interaction of eLectrons with soLid materiaLs 13
chaPter 3 aes methodoLogies 65
chaPter 4 instrumentation for auger anaLysis 89
chaPter 5 auger eLectron sPectroscoPy in materiaLs anaLysis 127
chaPter 6 anaLyticaL methods for the characterization of materiaLs 167
aPPendix 1 abbreviations and acronyms 193
aPPendix 2 Quantum numbers 197
aPPendix 3 comParison of surface and thin fiLm anaLysis techniQues 201
aPPendix 4 standardization in surface anaLysis 207
aPPendix 5 sources of the figures 209
further reading 211
index 215
-
list of figures
Figure 2.1. Types of electron and photon emission from a thin solid sample as a result of being bombarded by a beam of primary electrons. 14
Figure 2.2. A schematic representation of the primary excitation volume at the surface of a solid irradiated by a beam of high-energy electrons. 17
Figure 2.3. The effect of beam energy and the atomic number (Z ) of the sample on the shape and size of the primary excitation volume. 17
Figure 2.4. The kinetic energy spectrum of electrons emitted from an aluminum sample during bombardment with a 3-keV primary electron beam. 18
Figure 2.5. The proportion of electrons emitted from greater than a given depth (solid line) and the proportion of electrons emitted from less than a given depth (dashed line). The depth scale has been divided by the AL to make these curves independent of energy and material properties. This figure applies to those electrons that are emitted parallel to the surface normal. 21
Figure 2.6. The variation of the AL with electron kinetic energy according to the Seah and Dench equation (dashed curve) and according to the more recent Seah equation for carbon (upper solid curve) and tungsten (lower solid curve). 22
Figure 2.7. The origin of each type of radiation shown in relation to the primary excitation volume. 24
Figure 2.8. The mechanism by which an Auger electron or an X-ray photon is emitted. Note that this is a schematic and simplified diagram, and the atom having the electronic configuration shown would be neon. Electrons occupying at least some of the M states should be assumed. 25
-
x LiSt Of figuRES
Figure 2.9. Part of the Auger spectrum from copper showing the LMM region with the peaks annotated. 27
Figure 2.10. The kinetic energy of Auger transitions as a function of atomic number. The dots indicate the kinetic energy of the most intense peak in the group and the lines show the minimum and maximum kinetic energy of transitions in the group. 29
Figure 2.11. Auger spectra from a selection of elements showing the trends in the energy of the peaks as the atomic number increases. Spectra displayed in the direct mode. 30
Figure 2.12. Auger spectra from a selection of elements showing the trends in the energy of the peaks as the atomic number increases. Spectra displayed in the differential mode. 31
Figure 2.13. The four most common ways to display an Auger spectrum, in the case of as-received copper. 32
Figure 2.14. The ionization cross-section, a generalized curve using the overvoltage, which is the ratio of the primary beam energy and the critical excitation voltage. 33
Figure 2.15. The relative yield of Auger electrons (KLL, LMM and MNN) and X-rays (K, L and M) from an ionized, excited atom. 34
Figure 2.16. The Cu LMM Auger signal following subtraction of a background. 35
Figure 2.17. An example of a path of a primary electron that undergoes a series of inelastic collisions with the atoms in a solid. In this instance, an SE is emitted at each collision. An SE1 electron is emitted from a site close to the original point of impact, while an SE2 electron is emitted from a point some distance from the point at which the primary beam strikes the surface. The BSE is an electron from the original electron beam that has undergone one or more inelastic collisions and is emitted from the surface of the sample. 39
Figure 2.18. Secondary electron current density distribution as a function of the distance of electron emission from the point at which the primary electron beam strikes the surface. 40
-
LiSt Of figuRES xi
Figure 2.19. An example of the secondary electron yield as a function of the primary beam energy at a number of emission angles. The detailed shapes of these curves depend on the nature of the sample. 41
Figure 2.20. The effect of tilting the sample on the volume from which SE can be emitted from the sample surface. 42
Figure 2.21. The effect of the atomic number of the atoms in the sample on the yield of SE and BSE. 43
Figure 2.22. The radius of BSE emission in comparison with the radius of electron penetration. 45
Figure 2.23. Comparison of an SEM image (a) with a BSE image with all four quadrants positive (b) and with a BSE image with all four quadrants negative (c). 45
Figure 2.24. Elastic scattering from an atom in the solid and the definition of the scattering angle, q. 46
Figure 2.25. The energy loss occurring to electrons that strike atoms at a kinetic energy of 2 keV and are scattered through an angle of 120. 47
Figure 2.26. (a) Synthetic electron scattering spectra showing peak positions and intensities expected from a sample of Al(OH)3. (b) Expanded version of (a) showing only the total spectrum and the hydrogen spectrum. Primary beam energy = 1.0 keV and scattering angle = 120 48
Figure 2.27. Electron energy loss spectra from Al(OH)3, AlO(OH) and Al2O3, obtained using a primary beam energy of 1 keV. 49
Figure 2.28. The energy loss from a primary electron as it causes an atom in the solid to become ionized. 51
Figure 2.29. A REELS spectrum showing the energy loss associated with plasmon excitation in aluminum. 52
Figure 2.30. REELS measurements from a silicon sample that has a layer of oxide at its surface. Spectra were acquired at a series of primary beam energies. 53
Figure 2.31. REELS measurements taken with 1 keV primary electrons from a series of polymers. 53
Figure 2.32. An illustration of how bremsstrahlung is produced. 54
-
xii LiSt Of figuRES
Figure 2.33. Bremsstrahlung at a series of beam energies as a function of wavelength. The dotted line shows the position of the maximum in the spectrum as the beam energy is changed. 55
Figure 2.34. X-ray spectra from a steel showing the characteristic peaks superimposed on the background bremsstrahlung radiation at beam energies between 5 and 25 keV. 58
Figure 2.35. An illustration of the experimental method for obtaining XPS data from proximal emission of X-rays. 59
Figure 2.36. Data produced from proximal emission of X-rays. The SEM image shows the copper-containing particle under investigation on top of an aluminum substrate. The wide range spectrum shows the presence of copper Auger lines as well as the Auger lines from the aluminum substrate. The spectrum on the right is a magnified area of the spectrum showing the Cu 2p XPS spectral lines. 60
Figure 3.1. A means for storing or transporting a sample that avoids the surface to be analyzed coming into contact with the container. 68
Figure 3.2. A spectrum of a steel acquired in its as-received state and following a brief etch using an argon ion beam. The upper pair of spectra are direct spectra and the lower spectra are differentiated spectra. Only the more major of the LMM and MNN peaks have been labeled. 71
Figure 3.3. An illustration of the methods for obtaining signal intensities from (left) a direct spectrum and from (right) a differential spectrum. 74
Figure 3.4. An illustration of the effect of resolution on the spectrum and the ability to resolve differing chemical states. This example is from Al foil with a thin oxide at the surface. (a) Scaling and offset applied to each spectrum so that the effect of analyzer resolution on peak shape can be seen clearly. (b) The same data as shown in (a) but without the application of scaling or offset to show the effect of analyzer resolution on relative signal intensity. 76
Figure 3.5. The effect of emission angle on relative peak intensities for an aluminum sample that has a thin layer of oxide at its surface. 77
-
LiSt Of figuRES xiii
Figure 3.6. An SEM image showing a small particle at the surface of a hard disk. The spectra are from the particle (upper) and the disk (lower). The spectra indicate that the particle is largely composed of silicon dioxide. Elemental images are shown to the right. 79
Figure 3.7. Analysis of a small (
-
xiv LiSt Of figuRES
Figure 3.16. Depth profile from an X-ray mirror, which consists of alternating layers of silicon and tungsten. The data were collected under conditions optimized for depth resolution. 88
Figure 4.1. An example of a scanning Auger spectrometer. This one is based on a HSA. 90
Figure 4.2. An example of a scanning Auger electron spectrometer based on a CMA and coaxial electron gun. 91
Figure 4.3. A schematic energy versus distance diagram for electronic states near the solidvacuum interface of a metal. Ew is the work function of the metal. Two energy distributions are shown representing the energy distributions of electrons above the Fermi level. The darker distribution represents the situation at room temperature and the lighter one represents the distribution when the metal is heated to a high temperature. 94
Figure 4.4. Emission current density as a function of temperature for a tungsten emitter and for a thoria-coated tungsten emitter, (a) on a logarithmic scale and (b) on a linear scale. 95
Figure 4.5. Schematic diagram of an electron emitter module based on a tungsten filament. The photograph to the right is that of a typical filament module. The white ceramic filament support in this case is 12 mm in diameter. 96
Figure 4.6 . The structure of a LaB6 emitter. 97Figure 4.7. The potentials near the tip of a field emission source. 98
Figure 4.8. The arrangement of the electrodes and the electron trajectories for a field emission source, showing the relationship between the tip and the virtual source that is formed. 98
Figure 4.9. The structure of a Schottky emitter. 99Figure 4.10. Illustration of the way in which an image potential
is generated. 100Figure 4.11. The potentials near the tip of a Schottky
emission source. 101Figure 4.12. Electron-beam spot size as a function of sample
current for each type of source shown in Table 4.3. 103
-
LiSt Of figuRES xv
Figure 4.13. The overall design of an electron optical column suitable for use in the electron source of an Auger spectrometer. In this diagram, a light-optics analogy is used, in which the lenses are represented as glass lenses. 104
Figure 4.14. A simple, two-cylinder electrostatic lens shown in cross-section. 105
Figure 4.15. (a) A simple magnetic lens shown in cross-section and (b) an illustration of the way in which the image plane is rotated with respect to the object plane. 106
Figure 4.16. Designs for two types of magnetic lens: (a) is a suitable shape for a condenser lens and (b) is suitable for an objective lens. 107
Figure 4.17. An illustration of the origin of spherical aberration and how this can be controlled using the lens aperture. 109
Figure 4.18. (a) Astigmatism in a lens causing the position of the focal point to depend on the plane of the electron trajectories and (b) a common form of octopole stigmator. 110
Figure 4.19. (a) The arrangement of deflector electrodes and the beam trajectory in an electron column that uses electrostatic deflection. (b) The beamscan or raster pattern used when acquiring an Auger image or an SEM image. (c) The arrangement used for scanning and deflection when a single set of deflectors is used ahead of the objective lens. (d) The arrangement used for scanning and deflection when two sets of deflectors are used ahead of the objective lens. Note that in (c) and (d) only the xz plane is shown, but there would be similar arrangements in the yz plane. 111
Figure 4.20. Trajectories of electrons passing through a CMA. 113Figure 4.21. (a) Trajectories of electrons having kinetic energies
of 550 eV, 500 eV and 450 eV passing through the CMA tuned for 500 eV electrons and (b) the effect of sample position on the energy of the detected electrons. 114
Figure 4.22. CMA with a coaxial electron gun. 115Figure 4.23. A schematic diagram illustrating an arrangement
that allows the CMA to be operated with improved energy resolution. 116
-
xvi LiSt Of figuRES
Figure 4.24. Trajectories of electrons passing through an HSA in the dispersive plane. 117
Figure 4.25. Trajectories of electrons passing through an HSA in the nondispersive plane and viewed from the detector side of the analyzer. 118
Figure 4.26. HSA with multiple detectors, each detecting electrons of a different energy. 118
Figure 4.27. Potentials applied to the hemispheres as a function of electron kinetic energy when using the CRR mode of operation. 119
Figure 4.28. Arrangement of the components of a typical AES instrument based on an HSA. 120
Figure 4.29. Auger instrument in which the vacuum system is surrounded by an acoustic enclosure. 123
Figure 4.30. An ion source of the type commonly used on an AES instrument. 124
Figure 4.31. The plan view of a CMA-based instrument showing the components required for Auger analysis as well as some of the optional equipment that can be fitted. 125
Figure 5.1. (a) SEM image from a bond pad structure and (b) a silicon map of the same bond pad on which the three areas from which spectra were extracted are marked. Electron beam energy: 10 keV and current: 10 nA. 129
Figure 5.2. (a) The average Si KLL spectrum from the whole area of the image shown in Figure 5.1b. (b) The spectra extracted from the three areas marked in Figure 5.1b. Area 1 is silicon oxynitride, Area 2 is elemental silicon, and Area 3 is silicon in the form of a silicide. Data are acquired using the analyzer set to an energy resolution of 0.1%. 130
Figure 5.3. Chemical state images extracted from the spectra acquired at each pixel of the image. The images are: (a) silicon oxynitride, (b) elemental silicon, and (c) a silicide. 131
Figure 5.4. Secondary electron images of a semiconductor structure following the deposition of poly-silicon. Images are shown
-
LiSt Of figuRES xvii
at both low (left) and high magnification. A cross-section through the device and a contaminating particle at its surface has been cut using a FIB. 132
Figure 5.5. Scanning Auger maps of the cross-section shown in Figure 5.4. 132
Figure 5.6. Depth profiles from a superlattice structure obtained using two different ion energies for sputtering (500 eV for the profile on the left and 100 eV for the profile on the right). 133
Figure 5.7. Images from a cross-section of a GaAs/AlAs superlattice structure. The image on the left shows the secondary electron image, where layers of three different thicknesses are clearly visible, the white rectangle shows the area imaged in the middle secondary electron image, and the gallium Auger image is on the right. 134
Figure 5.8. A line scan that plots the intensity of the gallium signal as a function of distance across the 10-nm layers in the superlattice structure. 134
Figure 5.9. A depth profile through a nickel layer on a silicon substrate (left) and spectra of silicon (middle) and nickel (right) acquired over a large area in the nickel layer and in the substrate. 135
Figure 5.10. A secondary electron image of the nickel on a silicon sample following the depth profile acquisition and a pair of spectra taken from the particle visible in the image (point 1) and the surrounding area (point 2). 136
Figure 5.11. Spectra from TiN and Ti (left), and the depth profiles constructed using the energy regions indicated on the spectra. 137
Figure 5.12. Left: the basis spectra acquired from the sample during the profile. Right: the depth profile of the titanium chemical states constructed from the data using LLS fitting. 138
Figure 5.13. Full depth profile constructed using the LLS fitting procedure. 139
Figure 5.14. Secondary electron image showing the grain structure in a duplex steel and the positions of two points from which the spectra on the right were acquired. 140
-
xviii LiSt Of figuRES
Figure 5.15. Elemental Auger maps from chromium and iron taken from the duplex steel. 141
Figure 5.16. SEM and elemental Auger images acquired from the fractured surface of the iron. 142
Figure 5.17. Spectrum taken from the tin-coated crater visible in Figure 5.16. 143
Figure 5.18. SEM and elemental Auger images acquired at high magnification from the fractured surface of the iron. 144
Figure 5.19. Auger maps of Fe, Cr, and Sb from the fractured surface of the failed rotor blade. 146
Figure 5.20. Auger spectra from the four labeled points shown in Figure 5.19. 146
Figure 5.21. An Auger electron spectrometer fitted with a fracture stage mounted on an UHV preparation chamber (left) and an EDS spectrometer mounted on the analysis chamber (right). 147
Figure 5.22. Auger (left) and EDS (right) spectra from a stainless steel weldment. The upper spectra are from the hot-cracked region of the sample and the lower spectra are from the brittle fractured surface. 148
Figure 5.23. (a) Cross-section of a ball-milled crater in an oxide layer on a metal. (b) Cross-section through a similar sample following angle lapping. 151
Figure 5.24. Secondary electron image of a steel following immersion in an electrolyte for sufficient time for its corrosion to become visible. Auger spectra are taken from the areas marked as anodic and cathodic on the image. 153
Figure 5.25. Auger depth profile through DLC coating on a silicon substrate clearly showing the composition of the intermediate layers. 155
Figure 5.26. Secondary electron images from a 40 nm chromium oxide layer on a cobalt substrate, which has been scratched at increasing loads. The thin horizontal lines on these images and those in Figure 5.27 are from a layer of Al2O3 insulating material, which acts as a dielectric gap between the poles. Each image is collected from an area measuring 3 m 4 m. 156
Figure 5.27. Secondary electron image and Auger images of a scratched coating. 156
-
LiSt Of figuRES xix
Figure 5.28. A secondary electron image of a carbon nanotube on which an Auger line scan of the C KLL peak has been superimposed. 160
Figure 5.29. A secondary electron image of a SNW (left). A spectrum taken from the point labeled 1 on the image (middle) and a sputter depth profile of the phosphorus contained in the nanowire (right). 161
Figure 5.30. The concentration of phosphorus in the nanowire as a function of position along the wire and depth within the wire. 162
Figure 5.31. Images from a nanocone. Left: a secondary electron image. Middle: a nitrogen image. Right: an iron image. 163
Figure 5.32. Si KLL spectrum from the cone wall compared with standard spectra of silicon oxide and silicon nitride. These spectra were collected using an analyzer resolution of 0.1%. 163
Figure 5.33. Images from the smaller nanocones. Left: a secondary electron image. Middle: a nitrogen image. Right: an oxidized silicon image. 164
Figure 6.1. Photoelectron emission caused by an X-ray photon. 172Figure 6.2. An atomic force microscope. 189
-
list of tables
Table 1.1. Components that are fitted to an Auger electron spectrometer 8
Table 2.1. Some of the analytical techniques that are based on the interaction of electrons with a solid surface 14
Table 2.2. The relationship between the quantum numbers and the notation used in XPS and AES 26
Table 2.3. The temperature rise at the analysis position that can be expected during Auger analysis of a series of samples under a range of analysis conditions 61
Table 3.1. A quantification table for the direct and differential spectra of the etched sample shown in Figure 3.2 75
Table 4.1. Manufacturers of specialist Auger spectrometers. The model names, websites, and analyzer types are correct at the time of writing 89
Table 4.2. Manufacturers of XPS spectrometers to which an AES capability may be added as an option. The model names, websites, and analyzer types are correct at the time of writing 90
Table 4.3. Operating parameters and characteristics of electron sources 102
Table 4.4. Advantages of electrostatic and magnetic lenses 108Table 5.1. Some of the general types of analysis that
are commonly undertaken by AES 128Table 5.2. Labeling of Auger peaks in this example 137Table 5.3. The concentration of each of the elements observed
in the spectrum (Figure 5.17) 143Table 5.4. Bulk composition of the material recovered from
the failed rotor 145
-
xxii LiSt Of tabLES
Table 5.5. Changes in the Auger spectrum of a Fe20Cr steel as it is heated in the Auger spectrometer 150
Table 6.1. Analytical techniques having some degree of surface or thin-film specificity shown in relation to the input and output radiation associated with the technique 168
Table 6.2. The types of ion-scattering measurements 180Table A.1. The shaded cells show some of the atomic orbitals
that exist. The darker shading shows those orbitals that are occupied in the ground-state atom of at least one element up to lawrencium (atomic number: 102) 200
-
foreword
The goal of the Collection is to provide a set of definitive texts of moderate length on the techniques available for the analysis and characterization of materials, particularly technological and industrial materials. The empha-sis, at least for the initial volumes, will be on techniques used extensively for thin films, interfaces, and surfaces, or just very small amounts of material, or very low concentrations of a species in a host matrix. Many of the techniques will be spectroscopic in nature. There will be a wide variety of abilities in terms of spatial resolution, ranging from the sub-nanometer to millimeters, and in the depth probed, ranging from the top atomic layer to essentially bulk material. There are literally hundreds of techniques that could be included in such a materials analysis collection, but the aim here is primarily to cover those that currently have significant practical usage. Suitable techniques are those that tend to be available in commercial analytical laboratories, national facilities, university laboratories serving the materials user community, and in the research, development, and quality control laboratories of technological industry companies.
The texts will include a summary of the capabilities and main uses of the technique; an explanation of the physical basis of the technique in terms a non-specialist science or engineering major can easily follow; a section on instrumentation; a discussion of the range of information available and the range of materials to which it is applicable, and, equally important, the limitations of the technique; some typical examples in various technology areas; likely future developments; and finally, a com-parison with the capabilities of other techniques, some of which will also be in the Collection. The style of the volumes will inevitably vary some-what, ranging from the more academic textbook style at one end through to a how to hands on manual at the other, but the emphasis should always be on practicality, that is how is the technique actually used today.
The audience the Collection should appeal to includes college instruc-tors for whom the collection, or volumes in it, might serve as material
-
xxiv fOREWORD
for graduate courses; graduate students in the areas of materials science, chemical engineering, catalysis, nanoscience and nanotechnology, surface physics or chemistry; and, in industry, technologists, process engineers and managers in these areas. It should also be of interest to scientists and engineers in National Laboratories and in the materials analysis laborato-ries of the technology industries.
C. Richard BrundleCollection Editor
July 2015
-
Preface
This book was written as part of the series on Materials Characterization and Analysis. Such a series, which includes dynamic and static SIMS, TEM and XPS, would be incomplete if it did not include a volume on Auger electron spectroscopy (AES). AES is a major contributor to the field of surface characterization and analysis and is complementary to the techniques that are the basis for the other books in the field.
This book provides an introduction to the Auger technique for ana-lysts or those who may need to use or commission surface analysis. It attempts to answer a number of key questions that someone new to AES will ask. Such people will need to know the capabilities of this tech-nique. Will Auger provide the information I seek? They will also need to understand what the results of an Auger analysis are telling them. How are Auger data interpreted? To answer these questions and to understand the answers, it is essential to have a good understanding of the physical basis of the technique. What is the mechanism by which Auger electrons are emitted? A good analyst will have an understanding of the instrumen-tation. How does my instrument work? That way the analytical conditions can be set to provide optimum results. The analyst should also know the range of analytical options available and understand when to use each option. Would a spectrum be sufficient for this analysis or would an image be more appropriate? Some AES instruments provide the opportunity for additional types of analysis. Should these techniques be used to enhance the data coming from AES? It may be that AES will not provide all of the answers to a given analytical problem. Which additional analytical method should I use? The final chapter outlines the capabilities and physi-cal basis of other methods that might provide additional or complementary analytical data.
A largely qualitative approach to the physics underlying the Auger emission process and the theoretical aspects associated with Auger instru-mentation has been taken. The practical side of the subject is covered from
-
xxvi PREfacE
sample handling and preparation to quantification of the Auger spectra. There is a major chapter providing examples of AES analysis from a wide range of technologies (e.g. semiconductor, metals, nanotechnology etc.).
Input to this book was provided by a number of experts in the field of AES (see the Acknowledgments). When experts allowed the use of their data, these have been attributed in Appendix 5.
John WolstenholmeJuly 2015
-
acknowledgMents
Without the help I received from a number of people, it would not have been possible for me to produce this book; these people deserve my sincere thanks.
First, I should thank Dr. C.R. Brundle. As commissioning editor for this series of books, he offered me the opportunity to write this book. He has also provided a great deal of help and many useful suggestions.
A book of this sort cannot be written without the inclusion of a large amount of illustrative data. I would like to thank the following people for allowing me to use their data and, in some cases, for useful discussions regarding data interpretation: Prof. J.E. Castle, Dr. C.F. Mallinson, and Prof. J.F Watts of Department of Mechanical Engineering Sciences, University of Surrey, UK; Dr. C. Crawford of Kimball Physics Inc., USA; Prof. J. Sullivan of Midlands Surface Analysis Ltd., Aston University, UK; Dr. M.P. Seah of The National Physical Laboratory, UK; Dr. J. Hammond of Physical Electronics Inc, USA; Dr. T. Nunney, Dr. K.S. Robinson, and Dr. A. Wright of Thermo Fisher Scientific, UK.
Finally, I would like to thank my wife, Pat, for her patience and under-standing during the preparation of this volume and for checking that the draft manuscript made sense.
-
cHaPtER 1
introduction
1.1 SuRfacE aNaLYSiS
Auger electron spectroscopy (AES)1 is used to analyze the surfaces of solids. More precisely, it is used to analyze the solidvacuum or possibly the solidgas interface. Therefore, in this book, reference to surfaces will usually mean solid surfaces.
Solid materials interact with their surroundings via their surfaces. A surface is an abrupt boundary between a material and its environment and so the atoms at the surface of a solid will have a smaller number of nearest neighbors compared with those in the bulk of the material. This difference leads to a different electronic structure which, in turn, leads to higher chemical reactivity. For example, elements that do not naturally oxidize in air often have a layer of oxide or suboxide at the surface, the oxygen atoms replacing the missing nearest neighbor metal atoms. Similarly, supposedly clean surfaces very frequently have a thin layer of hydrocarbon contaminant (often called adventitious carbon).
The nature of the chemical interactions with surfaces will depend on the chemical structure, composition, and topography of the surface. To get a reasonably comprehensive understanding of a given surface the following properties will need to be determined, measured, and understood:
The chemical elements at the surface. The relative amount of each element at the surface. The chemical states of those elements. The uniformity of the elements and their chemical states to a very
fine scale. The variation with depth of the composition in the near-surface region.
1The abbreviation AES can be used to mean either Auger electron spectroscopy or Auger electron spectrometer. It is normally clear from the context which of these is meant. If not, it will be made explicit in the text.
-
2 augER ELEctRON SPEctROScOPY
The physical topography of the surface. The presence, nature, and chemical composition of impurities or
contaminants.
There is no single analytical technique that can provide a complete understanding of the surface but AES can be used to determine many of these properties with varying degrees of proficiency.
Some of the areas in which AES can provide a valuable contribution include:
Microelectronics Catalysis Corrosion Adhesion Lubrication Grain boundary segregation Particulate contamination Delamination
From this short list alone it can be seen that AES is an important tool in many areas of both industry and research. Examples of the application of the technique in most of the above areas are provided in Chapter 5.
1.2 WHat iS MEaNt bY SURFACE ANALYSIS?
AES is used for surface analysis but it is necessary to understand what that term means because the definition of the word surface can depend on the context in which it is used. In the context of AES the term surface refers to the top few atomic layers of a solid, up to about 10 nm.
Consider a silicon sphere having a radius of 1 cm. The proportion of atoms forming the top layer is approximately 1 in 3 107 and an elemental impurity at the surface at a concentration of 1 percent (a concentration that can normally be detected with AES) is therefore only present at a concentra-tion of about 0.3 ppb (parts per billion) of the sphere as a whole. It is there-fore essential that there is some form of filter that efficiently removes signal from the bulk material so that it does not swamp the signal coming from the near-surface region. It turns out that, for AES, this filter is the material itself.
As is shown in more detail in Chapter 2, when an energetic electron beam interacts with a solid surface it will penetrate to a depth of several micrometers,2 depending on its energy. It will interact with the solid over the
21 m is equal to 1,000 nm.
-
iNtRODuctiON 3
whole length of its path exciting the atoms with which it comes into contact. However, the Auger electrons which will be detected will have a range of energies which is much lower than the energy of the primary beam and so the distance they travel without experiencing a collision which changes their kinetic energy (their inelastic mean free path, IMFP) is small. It depends on the energy of the electron and the material through which it is traveling but, typically, it will be in the range 1 to 3 nm. More than 95 percent of the measured Auger signal will originate from a depth less than three times the IMFP. It is the small IMFP which provides the filter allowing only the signal from the near surface of the material to be detected. Signal is detected from more than a single monolayer but the proportion of the signal coming from the top monolayer is now much larger (~10 to ~30 percent).
1.3 WHat iS aES?
AES is a technique which is widely used for the chemical analysis of solid surfaces. In simple terms, an Auger electron spectrometer consists of a source of electrons whose energy is usually within the range 3 to 30 keV (although it is not limited to this range) and an electron energy analyzer. The analyzer measures the kinetic energy of electrons emitted from the surface under investigation as a consequence of the bombardment of the high-energy electrons. Measuring the kinetic energy of the emitted elec-trons allows the identity of the elements in the sample to be determined. The area of the surface analyzed using this technique is dependent pri-marily on the diameter of the electron beam as it strikes the surface. Since electron beams can be focused to spot sizes down to
-
4 augER ELEctRON SPEctROScOPY
referred to as a scanning Auger microscope by analogy with the scanning electron microscope.
Stand-alone Auger spectrometers are commonplace, but many have one or more additional analytical techniques added such as X-ray spectroscopy or even, by adding an X-ray source, X-ray photoelectron spectroscopy (XPS). Similarly, an electron source can be added to an X-ray photoelectron spectrometer to allow it to make AES measurements.
Emission of Auger electrons can result from irradiation of a surface with electrons, X-rays or high-energy ions. Typically, AES is performed using high-energy electrons. Auger peaks appear in XPS and observing these has value in the analysis but X-rays cannot be focused to spot sizes less than a few micrometers and cannot therefore provide the lateral resolution which is available using electrons. High-energy ions are seldom if ever used as a primary source in AES.
When an electron from an inner electron shell, close to the nucleus, is removed from an atom using high-energy radiation, an ion in an excited state is produced. It will then relax to a lower energy state by one of a number of processes (described in Chapter 2). The Auger process is one of those. According to the International Standard (ISO 18115) the Auger process is the relaxation, by electron emission, of an atom (or ion) with a vacancy in an inner electron shell. This process is referred to as a radia-tionless process since no photons are emitted. The energy of the emitted electron is determined by the element from which it is emitted and is inde-pendent of the energy of the ionizing radiation. In this respect AES differs from XPS in which the kinetic energy of the emitted electron is dependent on the energy of the incident X-ray photon.
It follows, therefore, that the elemental composition of a surface can be determined by measuring the kinetic energy of the electrons emitted by the Auger process.
AES is a process which requires that the atom has a minimum of three electrons and so this technique cannot be used to analyze for hydrogen and helium but it is a method that can be used for all other elements. Details of the Auger emission process will be provided in a later chapter.
1.4 cHEMicaL StatE iNfORMatiON
The position of a peak in an Auger spectrum is dependent on the ele-ment from which the electron is emitted and the electron energy levels involved in the Auger emission. In general, AES is not used for distin-guishing between the chemical states of the elements present at the surface because the chemical shifts in the peaks are small compared with the peak
-
iNtRODuctiON 5
widths as they appear in the spectrum. An exception to this is the ability of the technique to distinguish between the elemental state and the oxidized state of certain materials (e.g., elemental silicon and SiO2). AES would, of course, be able to detect the presence of both Si and O as elements in the spectrum. XPS is far better at providing much more subtle chemical state information than AES.
1.5 RaNgE Of REQuiRED iNfORMatiON
As has been mentioned previously, the depth from which AES provides chemical information is typically less than 3 nm, it is a highly surface- specific form of analysis. However, if the spectrometer is equipped with a source of ions (usually having an energy in the range of a few hundred electron volts to about 5 keV) it is possible to construct depth profiles of the near surface region of the solid. This is accomplished by analyzing the surface, using the ion beam to erode the sample by sputtering, repeating the analysis and continuing with this alternation of analysis and erosion until the required depth is reached. In principle, this depth profiling tech-nique can be used to analyze the solid upto any depth but, in practice, analysts do not use it for depths greater than a few microns because of the time required to measure a profile to greater depths plus complicating artifacts that get worse as a function of depth (see later).
For the analysis of deeper features, techniques such as in situ fractur-ing, surface lapping or ball cratering are often used in combination with AES analysis.
1.6 HiStORY Of aES
AES has its origins in the 1920s. Although the effect is named after Pierre Auger, it was first observed and reported by Elise (Lise) Meitner in 1922.
The Austrian, Lise Meitner, was a gifted, atomic physicist (Sime 1996). She was a leading figure in the early days of nuclear fission research. Her
AES is named after Pierre Auger, who was a French physicist. For people who are not French speakers it may not be obvious how his name should be pronounced. According to the International Phonetic Alphabet, the pronunciation is oe which approximates to o as in hotel, as in measure and e as in stay. It is not pronounced in the same way as the tool used for making holes.
-
6 augER ELEctRON SPEctROScOPY
work formed a major contribution to the development of both peaceful and not-so-peaceful applications of nuclear fission. One of her major achievements, working with Otto Hahn, was her discovery of a relatively long-lived isotope of protactinium (231Pa). She received several scientific awards and many people hold the view that she should have received a Nobel Prize for her work but she was overlooked. She observed and reported what we now call Auger emission during the course of her work, so it may be argued that the topic which is the subject of this book should be called Meitner electron spectroscopy. Fortunately, she has received some degree of permanent recognition because the synthetic element 109 (Meitnerium, Mt), which was discovered in 1982, has been named after her. It was named in 1997 nearly 30 years after her death. So far, this is the only element to be named after a nonmythological woman.
Pierre Victor Auger, a French physicist, reported the phenomenon of what is now known as Auger electron emission in 1925. He had been studying the X-ray photoelectron emission from gaseous atoms using a cloud chamber. He noted that some of the electrons emitted, following ionization by X-rays, originated from the ionized atom, not from the neu-tral species. This he interpreted as being due to a radiationless relaxation of the ion. The observation was reported in the French scientific journal Comptes Rendus (Auger and Perrin 1925) as a note entitled On second-ary beta-rays produced in a gas by X-rays. Later, Auger became inter-ested in cosmic rays and, with his co-workers, observed the effect of the earths magnetic field on these rays. He was later instrumental in the estab-lishment of the European Organization for Nuclear Research (CERN) in Geneva. Professor Auger was responsible for the establishment or promo-tion of nine national or international organizations.
He died in 1993 but his name lives on, not only with AES but also in the naming of what is currently the worlds largest cosmic ray observatory located in Argentina. He was nominated for a Nobel Prize but never received one. A short biography of Pierre Auger has been published (Persson 1996).
Although the phenomenon was observed independently by both Auger and Meitner in the 1920s, neither suggested its use as an analytical technique in surface analysis. This had to wait until 1953 when J.J. Lander, a scientist working for Bell Telephone Laboratories in New Jersey, recog-nized it as an interesting technique for surface analysis.
The technique began to be used for surface analysis in 1968 when L.A. Harris, working for General Electric in New York, used low-energy electron diffraction (LEED) optics to obtain Auger spectra from metal sur-faces. His breakthrough was in realizing the importance of differentiating the kinetic energy distribution in order to achieve sufficient sensitivity when using LEED optics for a viable analytical technique.
-
iNtRODuctiON 7
In 1969, Palmberg, Bohn, and Tracy invented the cylindrical mirror analyzer (CMA) which led to the development and eventual commercial-ization of dedicated Auger electron spectrometers.
Today commercial, stand-alone, Auger electron spectrometers are available and generally use either a CMA or a hemispherical sector ana-lyzer (HSA). LEED instruments are often capable of AES although their performance in AES is more limited than the other types of analyzers.
1.7 iNStRuMENtatiON fOR augER aNaLYSiS
A later chapter will deal with the details of the instrumentation but an outline of the required components of an Auger spectrometer will help in the understanding of what follows. AES can be performed either in an instrument primarily designed for the purpose or as an add-on technique in an instrument designed primarily for another technique, XPS, for example. As will be discussed later, it is essential that the Auger spectrometer be housed in an ultra-high vacuum (UHV) enclosure, this is true whether Auger is a primary or a secondary technique.
Loosely, the components of an AES instrument fall into three cate-gories: essential, desirable, and optional (as shown in Table 1.1). Essen-tial items are those without which it would not be possible to produce an
AUGER ELECTRON EMISSION FOR CANCER THERAPY?
Therapies which rely on Auger electron emission are being developed for the treatment of a number of cancers, including prostate cancer. For this type of therapy, a radiolabeled targeting reagent would be injected into the patient. This reagent then binds to a specific receptor in the targeted cell. If, for example, the radionuclide attached to the targeting reagent is 125I (half-life = 59.4 days) then it will decay by electron capture to an excited state of 125Te which further decays with the emission of 35 keV gamma rays. The emitted gamma rays cause the ionization of the molecule which then relaxes by the emission of a total of 21 Auger electrons in the energy range 50 to 500 eV. As with other forms of radiation therapy, Auger electrons damage the targeted cancer cells, including the DNA, in order to stop cell division and tumor growth. The mean free path of electrons having an energy in this range is so short that there is little or no damage to the cells surrounding the targeted cell. The gamma radiation is also of sufficiently low energy to ensure that cell damage remains very localized.
-
8 augER ELEctRON SPEctROScOPYTa
ble
1.1.
Com
pone
nts
that
are
fitt
ed to
an
Aug
er e
lect
ron
spec
trom
eter
Cat
egor
yC
ompo
nent
Com
men
t
Esse
ntia
lEl
ectro
n so
urce
Prov
ides
ele
ctro
ns h
avin
g th
e ne
cess
ary
ener
gy so
that
Aug
er e
lect
rons
ca
n be
em
itted
Elec
tron
ener
gy a
naly
zer a
nd
elec
tron
dete
ctor
Mea
sure
s the
ene
rgy
spec
trum
of t
he e
mitt
ed e
lect
rons
to a
llow
the
iden
tifi
cati
on o
f el
emen
ts in
the
sam
ple
Sam
ple
stag
eSu
ppor
ts a
nd m
anip
ulat
es th
e sa
mpl
e un
der i
nves
tigat
ion.
Idea
lly, t
his
wil
l hav
e fi
ve a
xes
of m
ovem
ent;
x, y
, z,
(ti
lt),
and
rot
atio
n. T
he
prec
isio
n an
d st
abili
ty o
f the
stag
e m
ust b
e co
nsis
tent
with
the
spot
size
of
the
elec
tron
beam
del
iver
ed b
y th
e el
ectro
n so
urce
.D
esira
ble
Seco
ndar
y el
ectro
n de
tect
or (S
ED)
If th
e pr
imar
y be
am c
an b
e sc
anne
d ov
er th
e su
rfac
e of
the
sam
ple
this
ty
pe o
f det
ecto
r allo
ws t
he in
stru
men
t to
acqu
ire se
cond
ary
elec
tron
(SE)
imag
es w
hich
can
be
extre
mel
y he
lpfu
l in
loca
ting
smal
l fea
ture
s on
the
sam
ple
for a
naly
sis.
In a
scan
ning
Aug
er sp
ectro
met
er, t
his c
ould
be
cla
ssifi
ed a
s an
ess
enti
al p
iece
of
equi
pmen
t.
Ion
gun
An
ion
gun
serv
es th
ree
usef
ul p
urpo
ses o
n an
AES
inst
rum
ent.
1.
It ca
n be
use
d to
rem
ove
cont
amin
atio
n fr
om th
e sa
mpl
e to
exp
ose
the
surf
ace
requ
iring
ana
lysi
s2.
It
can
be
used
to a
cqui
re c
once
ntra
tion
dep
th p
rofi
les
3.
It ca
n he
lp e
limin
ate
surf
ace
char
ging
if it
is c
apab
le o
f del
iver
ing
a be
am o
f io
ns w
ith
suffi
cien
tly
low
ene
rgy
-
iNtRODuctiON 9
Opt
iona
lX
-ray
sour
ceTh
is c
an b
e ad
ded
to in
stru
men
ts h
avin
g a
sphe
rical
sect
or ty
pe o
f ene
rgy
anal
yzer
. Its
use
allo
ws X
PS sp
ectra
to b
e co
llect
ed.
X-r
ay sp
ectro
met
erTh
e pr
imar
y el
ectro
n be
am c
ause
s X-r
ays t
o be
em
itted
as w
ell a
s Aug
er
elec
trons
. An
X-r
ay sp
ectro
met
er p
rovi
des t
he a
naly
st w
ith a
noth
er
anal
ytic
al te
chni
que.
Bac
ksca
ttere
d el
ectro
n de
tect
orA
s a c
ompl
emen
tary
met
hod
to th
e SE
, the
BSE
det
ecto
r pro
vide
s im
ages
w
hich
hav
e at
omic
num
ber c
ontra
st a
nd c
an p
rovi
de im
ages
whi
ch a
re
help
ful i
n di
rect
ing
subs
eque
nt A
uger
ana
lysi
s.Pr
epar
atio
n ch
ambe
r and
pr
epar
atio
n de
vice
sM
any
inst
rum
ents
hav
e a
UH
V p
repa
rati
on c
ham
ber.
Dev
ices
fitt
ed to
th
ese
incl
ude
a fr
actu
re st
age,
gas
dos
ing,
hea
ting
and
cool
ing,
pee
ling
stag
e, a
nd so
on.
Thi
s allo
ws f
resh
surf
aces
to b
e pr
epar
ed u
nder
UH
V
cond
ition
s and
ther
efor
e av
oids
con
tam
inat
ion
betw
een
prep
arat
ion
and
anal
ysis
.
-
10 augER ELEctRON SPEctROScOPY
Auger spectrum. Desirable items are those which extend the capabilities of the spectrometer but are usually supplied with an instrument that is pri-marily designed for AES. Optional items are those which can extend the capabilities further but are not so commonly included with the instrument.
If the primary technique of the instrument is XPS, it is possible to col-lect Auger spectra by adding an electron gun to the instrument. This can be a simple gun which cannot be scanned and so spectra can be obtained, probably from a relatively large area, but imaging is not possible and depth profiling may not be possible with any precision. If the electron beam in such a gun can be scanned these problems can be overcome and, by add-ing a SED as well, SE images can be collected along with SAM images. SE images can be the eyes of an Auger spectrometer because they can identify the area to be analyzed and direct the analysis to the right place. The lateral resolution obtainable with Auger using this type of instrument is likely to be worse than that obtainable from a purpose-built Auger spec-trometer by about a factor of 10 at best.
1.8 SuMMaRY Of tHE caPabiLitiES Of aES
Elements: All elements heavier than helium.Sample Types: Solid, able to survive in and not con-
taminate, a UHV. Conducting and semi-conducting samples routinely analyzed, insulating samples are more challenging and there may be limitations.
Sensitivity: 0.1 to 1 atomic percent, not a trace analy-sis technique.
Elemental Analysis: Yes, semiquantitative without standards, quantitative with standards.
Analysis Types: Spectroscopy with element identifica-tion, element mapping (imaging), and depth profiling. Angle resolved measure-ments possible on some spectrometers.
Chemical State Information: Limited, in favorable cases if sufficient energy resolution is available from the analyzer.
Destructive: Generally not for inorganic materials, unless sputter depth profile analysis is used. Some sensitive samples may be damaged by the electron beam, organic
-
iNtRODuctiON 11
materials and polymers are particularly susceptible to damage during analysis.
Depth Resolution: 0.5 to 5 nm depending on the sample. With some types of analyzer this can be controlled by tilting the sample.
Lateral Resolution:
-
index
AAbbreviations, 193195Aberrations. See Lens aberrationsAcronyms, 193195Adhesion, 2, 70, 128, 152, 154,
156,186Adventitious carbon, 1, 69, 72,
152AES. See Auger electron
spectroscopyAL. See Attenuation lengthAlloy
catalysis, 157passivation, 149phase boundaries, 141144phase characterization, 139141
Analyzer, 3, 8, 18cylindrical mirror analyzer, 7,
76, 90, 112116, 171hemispherical sector analyzer, 7,
8990, 116120retarding field analyzer, 112transfer function, 36
Ancillary equipmentdescription, 122123ion gun, 123125secondary electron detector, 123
Angleazimuthal, 116, 174emission, 20, 23, 41, 7678, 113,
120, 173, 176, 178grazing, 87, 177, 184
incidence, 36, 41, 42, 44, 46, 81lapping, 68, 151scattering, 43, 46, 48, 54, 180
Angle-resolved ultraviolet photoelectron spectroscopy (ARUPS), 173174
Angle-resolved X-ray photoelectron spectroscopy (ARXPS), 173
Anode, 59, 60, 95, 96, 98, 171, 172Aqueous corrosion, 151153Area of emission, 93ARUPS. See Angle-resolved
ultraviolet photoelectron spectroscopy
ARXPS. See Angle-resolved X-ray photoelectron spectroscopy
Atomic orbitals, 197198Atomic number, 9, 16, 17, 18, 23,
26, 28, 29, 34, 42, Attenuation length (AL), 1924,
34, 36, 38, 40, 72, 77, 78Auger electron emissioncancer therapy, 7differential/direct display
method, 3033energy of, 2728mechanism of, 2425notation, 2527quantification, 3438trends in Auger energies, 2829yield, 3334
-
216 iNDEX
Auger electron spectroscopy (AES)
capabilities of, 1011chemical state information, 45components of, 89definition, 3history of, 57instrumentation, 7, 10pronunciation, 5range of, 5types of analysis, 65
Auger, Pierre 56Auger process, 4, 25, 26Axial astigmatism, 109110Azimuthal rotation, 86
BBackscattered electron (BSE)
elastic scattering, 14, 24, 4549imaging with, 4445inelastic scattering, 14, 24, 39,
5054types of measurement, 43
Backscattered electron detection (BSD), 14, 125.See also EBSD
Ball milled crater, 151Beam deflection and scanning,
110112Bias, 115, 116, 123Bias supply, 95BIS. See Bremsstrahlung
isochromat spectroscopyBohn, G. K., 7Bremsstrahlung, 14, 5456, 58Bremsstrahlung isochromat
spectroscopy (BIS), 175Bright field illumination, 185, 188Brightness, 9293, 102BSD. See Backscattered electron
detectionBSE. See Backscattered electron
CCancer therapy, Auger electron
emission, 7Capacitance, 190
Catalyst, 126, 127, 156160, 162, 189
Catalysismodel catalysts, 159160poisoning and deactivation,
158159surface composition, 157158
Cathodoluminescencecharacteristics, 6061definition, 15
CERN. See European Organization for Nuclear Research
Chamberanalysis, 69, 70, 123, 147, 181preparation, 9, 70, 126, 141, 145,
147Chemical force interaction, 190Chemical state(s), 1, 4 29, 57, 60,
65, 73, 83, 130, 131, 136, 155, 160, 170, 173, 202
Chemical state determination, 7576
Chemical state information, 45, 10
Chromatic aberration, 93, 109CMA. See Cylindrical mirror
analyzerCoating(s), 136, 152, 154156,
182, 184Cold field emission, 99102, 109Collision(s), 3, 15, 16, 19, 20, 23,
34, 39, 45, 50, 54, 180Complementary analysis
techniques, 147148ComptesRendus, 6Concentric hemispherical analyzer
(CHA). See Hemispherical sector analyzer (HSA)
Condenser lens, 103104, 107Contamination
atmospheric, 79, 126, 141particulate, 2, 67, 79, 131132,
153, 186surface, 8, 9, 31, 49, 66, 67, 72,
73, 84, 127, 129
-
iNDEX 217
Corrosionaqueous, 151153due to contamination, 153154oxide layers, 149151passivation, 149stress, 140
Cross-sectionanalyzer, 112Auger emission, 19, 34electron interaction, 16, 19ionization, 33, 36, 37, 58, 59, 72,
116lens, 105, 106scattering, 37, 47sample, 132, 133, 151, 186X-ray emission, 59
Crossover, 42, 96, 98, 108Cylindrical mirror analyzer
(CMA), 7, 76, 90, 112116, 171
DDark field illumination, 185Data recording systems, 155156Depth profile(s), 5, 8, 10, 11, 65,
8388, 123, 133, 135, 137, 151, 154, 161, 173, 175, 1812, 184
Depth resolution, 86Detection limit, 56, 73, 145, 167,
179, 181, 182, 183, 205Diamond-like carbon (DLC),
154155Differential display method, 3033Direct display method, 3033DLC. See Diamond-like carbonDrift, 93, 102, 103, 121, 122, 161d-SIMS. See Dynamic secondary
ion mass spectrometryDynamic secondary ion mass
spectrometry (d-SIMS), 181182
EEAL. See Effective attenuation
length
EDS (EDX), 14, 5658, 125, 147148, 168, 174175, 185, 205
EELS. See Electron energy loss spectrum
Effective attenuation length (EAL), 19
Elastic interaction, 16Elastic peak electron spectroscopy
(EPES), 50Elastic scattering, 18, 19, 23, 43,
4549, 114, 180Electrochemical, 153, 186, 190Electron detector(s), 8, 9, 14, 32,
38, 43, 44, 66, 80, 114, 115, 118, 121, 123, 125
Electron emitter Auger spectrometers, 101103field emitters, 9799LaB6, 9697properties, 9293Schottky, 92, 97, 99104, 109thermionic emitters, 9397thoria coated, 94, 97
Electron energy loss spectrum (EELS), 4750, 5254, 176, 202
Electron guns beam deflection and scanning, 110112
electron emitter, 92103electron optics, 103110parameters of, 92
Electron microscopysecondary, 187188transmission, 188189
Electron opticscondenser lens, 103104electrostatic lens, 105108lens aberrations, 108110magnetic lens, 106108objective lens, 104105
Electrons as input radiation electron detection, 176178electron-stimulated desorption,
178
-
218 iNDEX
high-resolution electron energy loss spectroscopy, 176177
inverse photoemission spectroscopy, 175176
ion detection, 178low-energy electron diffraction,
177photon detection, 174176reflection high-energy electron diffraction, 177178X-ray spectroscopy, 174175
Electron spin, 198Electron trajectory, 98, 105, 107,
110, 113118Electron-stimulated desorption
(ESD), 178Electron stimulated desorption ion
angular distribution (ESDIAD), 178
Electron volt (eV), definition, 3Electrostatic lens, 105108Electrostatic octopole stigmator,
110Ellipsometry, 170171, 202Emission angle effects, 41, 7678Emitted electrons, 1824Emitted electrons spectrum, 1824Energy spread, 50, 93, 102, 109,
176EPES. See Elastic peak electron
spectroscopyESD. See Electron-stimulated
desorptionESDIAD. See Electron stimulated
desorption ion angular distribution
European Organization for Nuclear Research (CERN), 6
Excitation volume, primary, 1617, 19, 23, 24, 36, 37, 59, 61
Extraction field, 93, 94, 98, 99, 100, 102
FFailure analysis, steel, 144148FIB. See Focused ion beam
Field-assisted thermionic emission, 99
Field emission, 99Field emitters, 97103Filament, 91, 97Focused ion beam (FIB), 125126,
132, 179Fracture stage, 9, 66, 70, 126, 141,
147
GGrid, 95Gun
electron, 10, 59, 90, 91, 92112, 114, 115, 120, 122, 157, 161, 172, 175
ion, 8, 11, 66, 69, 86, 123125, 128, 175, 181, 183
HHandling techniques, 67Harris, L.A., 6HDA. See Hemispherical
deflection analyzerHeat, 14, 6162Hemispherical deflection analyzer
(HDA). See Hemispherical sector analyzer (HSA)
Hemispherical sector analyzer (HSA), 7, 8990, 116120
High pressure gas cell (HPGC), 126
High-resolution electron energy loss spectroscopy (HREELS), 50, 168, 176177, 202
HPGC. See High pressure gas cellHREELS. See High-resolution
electron energy loss spectroscopy
HSA. See Hemispherical sector analyzer
IImage potential, 100Imaging analysis, 7882
-
iNDEX 219
IMFP. See Inelastic mean free pathInelastic collisions, 15, 19, 23, 34,
54Inelastic interaction, 16, 39, 49Inelastic mean free path (IMFP),
3, 1923Inelastic scattering, 18, 34, 5054In situ sample preparation, 5, 66,
6970, 126, 147148, 149,160, 178
Instrumentationanalyzer, 112120ancillary equipment, 122125electron detector, 121electron guns, 92112optional equipment, 125126sample stage, 121122vacuum system, 9092vibration isolation, 122
Insulator analysis, 8182, 157Interaction of electrons
analytical techniques, 1415Auger electron emission, 2438backscattered electrons, 4354bremsstrahlung, 5456cathodoluminescence, 6061depth, 1518emitted electrons, 1824heat, 6162secondary electrons, 3843X-ray emission characteristics,
5660International Standard 18115, 4,
19, 4950International Standards
Organization (ISO) TC 201, 207208
Inverse photoemission spectroscopy (IPES), 168, 175176. 202
Ions as input radiationdynamic SIMS, 181182electron detection, 179ion detection, 179185low-energy ion scattering, 183
medium-energy ion scattering, 183184
particle-induced X-ray emission, 178179
photon detection, 178179Rutherford backscattering,
184185static SIMS, 180181
Ion-scattering spectroscopy (ISS), 46, 168, 179, 180, 183, 203,
IPES. See Inverse photoemission spectroscopy
ISO TC 201, 207208ISS. See Ion-scattering
spectroscopy
KKimball Physics Inc., 209Kramers formula, 55Kratos Analytical, 90
LLander, J.J., 6Lanthanumhexaboride (LaB6),
9697LEED. See Low-energy electron
diffractionLens, 95, 96, 103112, 115, 116,
119, 120, 172, 187Lens aberrations
axial astigmatism, 109110chromatic aberration, 109spherical aberration, 108109
Lifetime, 93, 95, 102Line scan 8283, 134, 151, 160Low-energy electron diffraction
(LEED), 6, 7, 13, 15, 112, 159, 168, 177, 203
Low-energy ion scattering, 168, 180, 183, 203
MMagnetic induction, 190Magnetic lens, 106108Magnetic octopole stigmator, 110
-
220 iNDEX
MALDI. See Matrix-assisted laser desorption/ionization
Manufacturers, 8990 Materials analysis
catalysis, 156160metals, 139154nanomaterials, 160164surface coatings, 154156
Materials characterizationanalytical techniques, 168electrons as input radiation,
174178ions as input radiation, 178185microscopy, 185191photons as input radiation,
169174Matrix-assisted laser desorption/
ionization (MALDI), 168, 174Medium-energy ion scattering,
183184, 203Meitner, Elise (Lise), 5Metal analysis
alloy phases characterization, 139141
complementary analysis techniques, 147148
corrosion, 148154failure analysis in steel, 144147phase boundary segregation,
141144in situ sample preparation,
147148Micro-area profiling, 86Microscopy
electron, 187189optical, 185187scanning probe, 189191
Midlands Surface Analysis Ltd., 209
Model catalysts, 159160Mounting techniques, 6869Multilayer characterization
digital methods usage, 136139profiling and imaging, 135136superlattice structures, 133135
NNanocone imaging, 162164Nanomaterials
description, 160nanocone imaging, 162164phosphorus doping in Si nanowires, 161162
Near-field scanning optical microscope (NSOM), 190191
Noise, 56, 73, 93, 102, 103, 112, 138, 142
NSOM. See Near-field scanning optical microscope
OObjective lens, 104105, 107, 111,
187Omicron Nano Technology
GmbH, 89, 90Optical microscopy, 185187Optional equipment, 125126Orbital orientation, 198Orbital shape, 198Orbital size, 197198Overvoltage, 33, 58Oxide layer(s), 20, 52, 75, 77, 84,
129, 149151, 158
PPalmberg, P. W., 7Particle-induced X-ray emission
(PIXE), 178179, 203Passivation, 149, 153Phase boundary segregation,
141144Phase contrast technique, 186Phosphorus doping, Si nanowires,
161162Photons as input radiation
electron detection, 171174ellipsometry, 170171ion detection, 174Raman spectroscopy, 170reflection absorption infrared spectroscopy, 169170
-
iNDEX 221
ultraviolet photoelectron spectroscopy, 173174X-ray photoelectron spectroscopy, 171173
Photon-stimulated desorption (PSD), 174
Physical Electronics, 86, 89, 90, 209
PIXE. See Particle-induced X-ray emission
Plasmons, 15, 16, 52, 204Primary excitation volume, 1617,
19, 23, 24, 36, 37, 59, 61Proximal excitation of X-rays,
5960PSD. See Photon-stimulated
desorption
QQualitative analysis, 14, 28, 65,
7073, 128, 171Quantification, 32, 3438, 56, 65,
7375, 82, 182, 183, 191, 202sensitivity factors, 3738table, 75using formula, 3537using ratios, 37
Quantitative analysis, 7375Quantum numbers, 26, 197200Quasi-elastic scattering, 16
RRAIRS. See Reflection absorption
infrared spectroscopyRaman spectroscopy, 168, 169,
170Raster(ed), 38, 44, 72, 110, 111,
124, 132RBS. See Rutherford
backscatteringREELS. See Reflection electron
energy loss spectroscopyReflection absorption infrared
spectroscopy (RAIRS), 169170, 204
Reflection electron energy loss spectroscopy (REELS), 15, 50, 5253, 168, 176, 204
Reflection high-energy electron diffraction (RHEED), 177178
Relative peak intensity, 7678Resolution,
analyzer,66, 7273, 7576, 114120, 163
angular, 65, 120depth, 11, 8588, 133energy, 10, 65, 76, 77, 84, 112,
114120, 130, 135, 139lateral/spatial, 3, 4, 10, 11, 16,
24, 42, 44, 57, 59, 65, 82, 101, 112118, 121, 122, 127, 129, 136, 157, 171, 202
Retard ratio, 119Retarding field analyzer (RFA), 112RFA. See Retarding field analyzerRHEED. See Reflection high-
energy electron diffractionRichardson equation, 94Rotation (of sample), 8688, 121,
133Roughness, 36, 38, 8587, 170Rutherford backscattering (RBS),
184185, 204
SSample
charging, 8, 8182, 125 handling techniques, 67fracture, 66, 70, 126, 141142,
145148mounting techniques, 6869parking capability, 126preparation methods, 6768properties, 6667in situ preparation, 6970stage, 121122, 161storage, 68transport, 68
SCA. See Spherical capacitor analyzer
-
222 iNDEX
Scanning Auger microscope, 4, 14, 92
Scanning Auger spectrometer, 8, 44, 65, 8991
Scanning Probe Microscope (SPM), 159
Scanning probe microscopy (SPM), 189191
Schottky emitters, 92, 97, 99104, 109
Secondary electron detector (SED), 8, 10, 32, 38, 41, 66, 78, 90, 123, 179, 187
Secondary Electron Microscope, 38, 44, 92, 103
Secondary electron microscopy (SEM), 14, 187188
Secondary electronsdescription, 14, 19, 38energy distribution, 18, 4041origin of, 24, 3940yield, 4143
Secondary ion mass spectrometry (SIMS)
dynamic, 181182static, 180181
Segregationgrain boundary, 2, 66, 70, 128,
145149phase boundary, 141144surface, 127, 128
SED. See Secondary electron detector
SEM. See Secondary electron microscopy
Semiconductordevice(s), 78, 84, 85, 128132,
188, 190distribution of chemical species,
129131materials, 15, 61multilayer characterization,
133139, 186
particulate contamination, 131132, 186
Sensitivity factor(s), quantification, 3738, 59, 7375
Shell(s), 4, 16, 33, 57, 198199SIMS. See Secondary ion mass
spectrometrySmall feature(s), 8, 65, 7882, 83,
84, 128, 163Source
electron, 3, 4, 8, 38, 56, 92103, 104, 114, 116, 178
ion, 5, 173, 181, 182virtual, 98, 102, 103, 116X-ray, 4, 9, 56, 59, 66, 123, 125,
172SPECS GmbH, 90Spectrometer, electron, 310, 13,
1415, 18, 24, 27, 32, 35, 39, 41, 43, 46, 65, 66, 75, 78, 84, 89126, 202
Spectroscopy methodologieschemical state determination,
7576emission angle effects, 7678qualitative analysis, 7073quantitative analysis, 7375
Spectroscopists notation, 26, 197Spherical aberration, 92, 93,
108109Spherical capacitor analyzer
(SCA). See Hemispherical sector analyzer (HSA)
Spherical sector analyzer (SSA). See Hemispherical sector analyzer (HSA)
Spin-orbit splitting, 199SPM. See Scanning probe
microscopySputter/Sputtering, 5, 10, 11, 83,
86cleaning, 69, 73danger of, 82, 99
-
iNDEX 223
depth profile, 68, 133135, 151, 161, 169
yield, 86SSA. See Spherical sector analyzerStage, sample, 121122, 161Standardization, surface analysis,
207Static secondary ion mass
spectrometry(s-SIMS), 180181
Steel analysis, 44, 58, 7072, 74, 139141, 144146, 148153,
Steel, failure analysis, 144147Storage, 68Subshells, 198199Superlattice structures, 133135Surface analysis
AES contribution areas, 2chemical interactions, 1definition, 23properties, 12standardization, 207vs. thin film techniques, 202205
Surface coatingsdata recording systems, 155156diamond-like carbon, 154155
Surface composition, 14, 127128, 149, 153, 157158
Surrey, University of, 57, 59, 209
TTEM. See Transmission electron
microscopyTemperature, 6162, 9395,
97, 99, 100, 102, 109, 121, 122, 126, 140, 145, 147151, 158160, 190
Thermionic emitters, 9397Thermo Fisher Scientific, 90, 209Thin film vs. surface analysis,
202205
Tilt (sample), 8, 11, 42, 78, 87, 88, 121, 162
Tracy, J.C., 7Transmission electron microscope,
50, 54, 128129, 157, 168Transmission electron microscopy
(TEM), 15, 188189Transport, sample methodology,
68
UUHV. See Ultra-high vacuumUltra-high vacuum (UHV), 7, 9,
10, 65, 70, 9092, 123, 141, 147, 149, 157, 171
Ultraviolet photoelectron spectroscopy (UPS), 168, 173174, 205
Universal curve(s), 21, 23UPS. See Ultraviolet photoelectron
spectroscopy
VVacuum system, 9092VELS. See Vibrational energy loss
spectroscopyVertical interconnect access (VIA),
85VIA. See Vertical interconnect
accessVibrational energy loss
spectroscopy (VELS), 176Vibration isolation, 122Virtual source, 98, 102, 103, 116
WWDS 14, 56, 168, 174Wehnelt electrode, 96
XX-ray emission, 25, 55, 168,
178179, 203
-
224 iNDEX
characteristics, 5660proximal excitation, 5960
X-ray mirror, 88X-ray notation, 2527, 57X-ray photoelectron spectrometer,
7, 90, 103, 105, 119, 120, 183X-ray photoelectron spectroscopy
(XPS), 4, 9, 10, 25, 26, 56, 5960, 66, 168, 171173, 205
X-ray spectroscopy, 14, 5759, 174175, 205. See also EDS, EDX, WDS
YYield
Auger electron, 3334, 57, 148backscattered electron, 43, 44ion, 181, 182secondary electron, 4143, 81sputter, 86, 133X-ray, 57, 148
ZZalar rotation, 86