The world leader in serving science
Steve Seddio
WDS spectrometer technologies:Not all WDS spectrometers are created equally
![Page 2: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/2.jpg)
2
What is WDS?
nλ = 2d sinθ
![Page 3: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/3.jpg)
3
• The result:
• Best energy resolution
• Best peak-to-background
Why WDS?
EDS
WDS
![Page 4: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/4.jpg)
4
The Challenge of WDS
• Hypothetical sample in an SEM
• For microanalysis, sample is normal to the electron beam
![Page 5: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/5.jpg)
5
The Challenge of WDS
• X-rays are emitted from the excitation volume in a
3-dimensional hemispherical wavefront
![Page 6: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/6.jpg)
6
The Challenge of WDS
• Get diverging X-rays to diffract off of a diffractor and then be
counted by a detector at a meaningful intensity
• Flat diffractors:
• Simple geometry
• X-rays interact with diffractor with different θ
• X-rays continue to diverge after diffraction → low count rates at detector.
![Page 7: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/7.jpg)
7
Meeting the Challenge
• All WDS spectrometers consist of
• Diffractor
• Proportional counter (a.k.a., detector)
• Flowing P10 gas (90% Ar, 10% CH4)
• Sealed Xe
• Some WDS spectrometers include an X-ray optic near sample
• There are two types of spectrometers that have been
developed to meet the WDS challenge
• Rowland circle WDS
• Parallel beam WDS
![Page 8: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/8.jpg)
8
Meeting the Challenge: Intensity
• For WDS, the count rate is a function of
• X-ray generation• Accelerating voltage
• Beam current
• Vacuum
• Sample composition
• Sample preparation
• Solid Angle and Optic • Diffractor size
• Diffractor distance from sample
• Size of optic
• Reflectance and/or transmittance of optic
• Other• Diffractor “reflectance”
• Type of gas in detector
• Pressure of gas in detector
• Transmittance of detector window
• Ambient temperature
• Presence / transmittance of spectrometer window
![Page 9: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/9.jpg)
9
Rowland Circle WDS: The Classic Solution (c. 1882)
• Curved diffractors focus the diffracted X-rays on to detector
• Requires complex motion so that the sample, diffractor, and
detector remain on a circle of fixed radius
![Page 10: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/10.jpg)
10
Rowland Circle WDS: The Classic Solution
![Page 11: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/11.jpg)
11
Rowland Circle WDS: The Classic Solution
• To maintain RC geometry, diffractor must move away from
sample for low-energy X-rays
• The result is a decrease in solid angle → low intensity (c/(s×nA))
for low energy X-rays
![Page 12: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/12.jpg)
12
Rowland Circle WDS: The Classic Solution
• Diffractor rotation affects solid angle
![Page 13: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/13.jpg)
13
Rowland Circle WDS: Solid Angle
102030405060
Solid
Angle
θ
Solid angle as a function of θ
![Page 14: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/14.jpg)
14
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Rowland Circle WDS: SEM Solid Angle
Mo/B4C
2d = 200
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
Moon or Sun
![Page 15: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/15.jpg)
15
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Mo/B4C or C/W
2d = 145
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 16: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/16.jpg)
16
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Ni/C or C/W
2d = ~100
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 17: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/17.jpg)
17
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Ni/C or Cr/Sc or C/W
2d = ~80
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 18: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/18.jpg)
18
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
W/Si or C/W
2d = 60
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 19: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/19.jpg)
19
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
W/Si
2d = 45
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 20: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/20.jpg)
20
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
TAP
2d = 25.757
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 21: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/21.jpg)
21
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
PET
2d = 8.742
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 22: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/22.jpg)
22
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
LiF (200)
2d = 4.027
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 23: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/23.jpg)
23
X-ray Energy (keV)
0.1 1 10
Solid
Ang
le (
sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
LiF (220)
2d = 2.848
Rowland Circle WDS: SEM solid angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
210 mm SEM
SAdiff = 661 mm2
![Page 24: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/24.jpg)
24
X-ray Energy (keV)
0.1 1 10
Solid
Angle
(sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Rowland Circle WDS: SEM Solid Angle
Solid angle as a function of energy
210 mm SEM
SAdiff = 661 mm2
Be B O Al Si Ti Fe Cu Sr
LiF (200)2d = 4.027
PET2d = 8.742
TAP2d = 25.757
Mo/B4C2d = 200
Cr/Sc2d = ~80
![Page 25: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/25.jpg)
25
Rowland Circle WDS: The Electron Microprobe (EPMA)
• Electron microscope with typically 5 RC-WDS spectrometers
• Spectrometers can concurrently analyze the same or different elements
• 2 or 4 diffractors in each spectrometer
![Page 26: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/26.jpg)
26
X-ray Energy (keV)
0.1 1 10
Solid
Angle
(sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Rowland Circle WDS: EPMA Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
160 mm EPMA
SAdiff = 704 mm2
210 mm SEM
SAdiff = 661 mm2
![Page 27: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/27.jpg)
27
X-ray Energy (keV)
0.1 1 10
Solid
Angle
(sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Rowland Circle WDS: EPMA Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
160 mm EPMA
SAdiff = 704 mm2
210 mm SEM
SAdiff = 661 mm2
160 mm EPMA
SAdiff = 1320 mm2
![Page 28: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/28.jpg)
28
X-ray Energy (keV)
0.1 1 10
Solid
Angle
(sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Rowland Circle WDS: EPMA Solid Angle
Solid angle as a function of energy
Be B O Al Si Ti Fe Cu Sr
160 mm EPMA
SAdiff = 704 mm2
210 mm SEM
SAdiff = 661 mm2
160 mm EPMA
SAdiff = 1320 mm2
LiF (200)2d = 4.027
PET2d = 8.742
TAP2d = 25.757
Ni/C2d = ~100
W/Si2d = 60
W/Si2d = 45
![Page 29: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/29.jpg)
29
Rowland Circle WDS: The Classic Solution
• Pros
• Excellent energy resolution
• Excellent peak-to-background
• Cons
• Complicated spectrometer geometry
• Best results when the chamber is designed for the spectrometer.
• X-rays measured with large θ have low intensities
• Typically relies on an optical microscope to ensure sample is at proper
working distance
Not commonly available to the SEM user.
Requires a horizontal geometry
![Page 30: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/30.jpg)
30
Parallel Beam WDS: A Modern Approach
• Parallel beam WDS spectrometer
• Collimating optic is located near (~ 20 mm) from sample
• Grazing incidence
• Polycapillary
• Hybrid
• Parallel X-ray beam incident on diffractor
• No Rowland circle geometry needed
• Diffractor is flat
![Page 31: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/31.jpg)
31
Parallel Beam WDS: A Modern Approach
![Page 32: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/32.jpg)
32
X-ray Energy (keV)
0.1 1 10
Solid
Angle
(sr)
0.00
0.01
0.02
0.03
0.04
0.05
0.06
Parallel Beam WDS: A Modern Approach
Polycapillary
Be B O Al Si Ti Fe Cu Sr
![Page 33: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/33.jpg)
33
X-ray Energy (keV)
0.1 1 10
Solid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Parallel Beam WDS: A Modern Approach
Solid Angle ≠ IntensityGrazing incidence
Polycapillary
Hybrid
Be B O Al Si Ti Fe Cu Sr
![Page 34: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/34.jpg)
34
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Parallel Beam WDS: A Modern Approach
Grazing incidence
Polycapillary
Hybrid
Be B O Al Si Ti Fe Cu Sr
![Page 35: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/35.jpg)
35
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Parallel Beam WDS: A Modern Approach
Grazing incidence
Polycapillary
Hybrid
210 mm SEM
Be B O Al Si Ti Fe Cu Sr
![Page 36: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/36.jpg)
36
Parallel Beam WDS: A Modern Approach
• Pros
• Excellent energy resolution
• Excellent peak-to-background
• Excellent intensity
• Cons
• Cannot accommodate a slit to modestly improve energy resolution with
a large intensity cost
![Page 37: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/37.jpg)
37
Head-to-Head Comparison: SEM and EMP
• Identical accelerating voltage and beam current
• PB-WDS with hybrid optic
• 1 spectrometer
• Sealed Xe detector
• Flat diffractors
• 160 mm RC-WDS
• 5 spectrometers
• 3 detectors with 0.1 atm P10 flow
• 2 detectors with 1 atm P10 flow
• Curved diffractors (Johann and Johansson)
X-ray Energy (keV)
0.1 1 10E
ffective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Seddio and Fournelle 2015
![Page 38: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/38.jpg)
38
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Head-to-Head Comparison: SEM and EMP
LiF (200)2d = 4.027
LiF (220)
2d = 2.848
![Page 39: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/39.jpg)
39
Head-to-Head Comparison: SEM and EMP
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
LiF (200)2d = 4.027
![Page 40: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/40.jpg)
40
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Head-to-Head Comparison: SEM and EMP
PET2d = 8.742
LiF (200)2d = 4.027
![Page 41: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/41.jpg)
41
Head-to-Head Comparison: SEM and EMP
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
TAP2d = 25.757
PET2d = 8.742
![Page 42: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/42.jpg)
42
Head-to-Head Comparison: SEM and EMP
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
TAP2d = 25.757
![Page 43: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/43.jpg)
43
Head-to-Head Comparison: SEM and EMP
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Ni/C2d = ~100
W/Si2d = 60
![Page 44: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/44.jpg)
44
Head-to-Head Comparison: SEM and EMP
X-ray Energy (keV)
0.1 1 10
Effective S
olid
Angle
(sr)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Ni/C2d = ~100
Mo/B4C2d = 200
![Page 45: Wds technology june 2015](https://reader031.vdocuments.site/reader031/viewer/2022020103/58eccb731a28ab113c8b4619/html5/thumbnails/45.jpg)
45
Conclusions
• RC-WDS
• Excellent solution when microscope is designed primarily for WDS
• Yields impractically low intensities on SEMs
• PB-WDS
• Unrivaled low energy X-ray intensity
• High energy X-ray intensities consistent with the best RC-WDS
intensities on an SEM