validation of sterilization equipment.pdf

51
VALIDATION OF STERILIZATION EQUIPMENTS Aseptic Area Validations May 2-3, 2002 İstanbul Hilton Suat Kumser Pfizer İlaçları Ltd. Şti. e - mail: [email protected] Turkish Pharmaceutical & Chemical Industry Research and Development Foundation Slide: 1/51

Upload: chefi84

Post on 03-Jan-2016

185 views

Category:

Documents


8 download

DESCRIPTION

Aseptic Area Validation

TRANSCRIPT

Page 1: Validation of Sterilization Equipment.pdf

VALIDATION OF STERILIZATION EQUIPMENTS

Aseptic Area ValidationsMay 2-3, 2002 İstanbul Hilton

Suat KumserPfizer İlaçları Ltd. Şti.e- mail: [email protected]

Turkish Pharmaceutical & Chemical Industry Research and Development Foundation

– Slide: 1/51

Page 2: Validation of Sterilization Equipment.pdf

Content:

• Definition of Sterilization and Depyrogenation• Microbiological aspects of Sterilization and

Depyrogenation, Lethality calculation, • D- Value, FH & F0 Values• Z- Value and use of microbiological indicators.

– Slide: 2/51

Page 3: Validation of Sterilization Equipment.pdf

Content:- Dry Heat Ovens- Dry Heat Sterilization Tunnels- Steam Sterilizator (Autoclaves) 1. Design Qualification2. Installation Qualification3. Operational Qualification4. Performance Qualification4.1. Thermodynamical aspects of Sterilization3.2. Temperature Distribution and Heat Penetration studies.

– Slide: 3/51

Page 4: Validation of Sterilization Equipment.pdf

Definitions:

• 1. Sterilization: Validated process used to render a product free of living microorganisms including bacterial endospores.

• 2. Depyrogenation: Removal or inactivation of bacterial endotoxin.

– Slide: 4/51

Page 5: Validation of Sterilization Equipment.pdf

Sterilization Only:

• The cycle is designed to assure that the probability of survival of the native microflora is no greater than one cell in one million units of the commodity. (10-6 probability of nonsterility)

• Dry Heat Sterilization, Theoretical requirement: 170 0C, 32 min.

• Steam Sterilization Theoretical requirement: 121 0C, 15 min.

– Slide: 5/51

Page 6: Validation of Sterilization Equipment.pdf

Sterilization - Overkill

• The overkill approach provides assurance of sterilization well in excess of the 10-6 probability of non- sterility. For example an FH provided by an overkill cycle may produce a 12- log reduction of a biological indicator that exhibits a high resistance to dry heat.

– Slide: 6/51

Page 7: Validation of Sterilization Equipment.pdf

Sterilization & Depyrogenation

• Applies to the cycles where the purpose is both sterilization and depyrogenation. Whenever depyrogenation is a desired end point, relatively high temperatures and/or extended heating times are necessary. Thus, microbial lethality delivered by these cycles provides a margin of safety far in excess of a 10- 6

probability of nonsterility. • Dry Heat Depyrogenation Theoretical requirement:

250 0C-30 min.

– Slide: 7/51

Page 8: Validation of Sterilization Equipment.pdf

D - Value : Time required for one log (or 90%) reduction of microorganism population at base

temperature.

Microbial Death Curve

0,00000010,000001

0,000010,0001

0,0010,01

0,11

10100

100010000

1000001000000

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Minutes at Base Temp.

Log

num

b. o

f Sur

vivo

rs

D-Value= 1.0 min.

– Slide: 8/51

Page 9: Validation of Sterilization Equipment.pdf

Determination of Z - Value:

• Determine the D - value of an organism at min. three different temperatures.

• Construct a Thermal Death Curve by plotting the logarithm of the D- Value versus temperature.

– Slide: 9/51

Page 10: Validation of Sterilization Equipment.pdf

Z-Value: Death Rate Constant

Assesment of Z Value

0,1

1

10

100

120 130 140 150 160 170

Temperature 0C

Log

D Va

lue Z Value = 20 0C

D130 0C : 10 min.

D150 0C : 1.0 min.

– Slide: 10/51

Page 11: Validation of Sterilization Equipment.pdf

Z-Value:

• In general, for Dry Heat sterilization, Z- Value may be assumed as 20 0C. And for Steam Sterilization as 10 0C.

• However, it will be appropriate to verify for the biological indicators when they are used to measure the integrated lethality of a dry heat or steam sterilization cycle.

– Slide: 11/51

Page 12: Validation of Sterilization Equipment.pdf

LETHALITY RATE:

Also defined as : • FH For Dry Heat Sterilization• Fo For Steam Sterilization• The equivalent sterilization time spent

at the base temperature. • Tb : 170 oC (For Dry Heat Sterilization)• Tb : 121 oC (For Steam Sterilization)

– Slide: 12/51

Page 13: Validation of Sterilization Equipment.pdf

LETHALITY CALCULATION“Patashnik Method”

Lethality Rate : 10 (T-Tb)/Z

FH = ∆t x Lethality Rate∆t : Cycle timeT : Actual Cycle temperatureTb : Base Temperature Z : Microbial Death Rate Constant

– Slide: 13/51

Page 14: Validation of Sterilization Equipment.pdf

Example: Determination of FH of a 3 min. dry heat sterilization cycle at 175 0Ct = 3 min T = 175 0C FH = 4 x 10 (175-170)/20

Tb = 170 0C FH = 5.31z = 20 0CSterilization at 175 0C for 3 min. is equivalent to 5.31 min. at 170 0C .

LETHALITY CALCULATION

– Slide: 14/51

Page 15: Validation of Sterilization Equipment.pdf

Lethality in Dry Heat Sterilization

Time Temperature Lethality Rate(min) (0C) min. at 170 0C

5 105 0,000610 110 0,001015 120 0,003220 135 0,017825 150 0,100030 165 0,562335 170 1,000040 172 1,258945 174 1,584950 174 1,584955 174 1,584960 175 1,778265 165 0,562370 150 0,100075 140 0,031680 130 0,010085 110 0,001090 105 0,0006

– Slide: 15/51

– Σ of Lethal Rates : 10.1912

–FH = ∆t x Σ of Lethal Rates

–∆t = 5 min.

–FH = 5 x 10.192

–FH = 50.961 min. at 170 0C .

Page 16: Validation of Sterilization Equipment.pdf

PART-1

DRY HEAT STERILIZATION

AND DEPYROGENATION VALIDATION

– Slide: 16/51

Page 17: Validation of Sterilization Equipment.pdf

DRY HEATSTERILIZATION & DEPYROGENATION

• Dry heat is often the agent of choice for sterilizing items which will tolerate high temperatures. Dry heat sterilization processes are generally less complicated than steam processes, although higher temperature and/or longer exposure times are required because microbial lethality associated dry heat is much lower than that for saturated steam at the same temperature.

– Slide: 17/51

Page 18: Validation of Sterilization Equipment.pdf

Thermodynamical Aspects of Heating Process: 1. Convection Heating Process: • The heat transfer through a medium

by motion of it ‘s parts. Natural convection is a result of differences in density caused by temperature gradients in the fluid mass.

• Forced convection heating is effected by the action of a mechanical device.

– Slide: 18/51

Page 19: Validation of Sterilization Equipment.pdf

Thermodynamical Aspects of Heating Process:

2. Conduction Heating Process: • Conduction is accomplished ether by a

molecular interaction from higher energy level to a lower energy level or by free electrons.

• Thus, the ability of solids to conduct heat varies directly with the free electron concentration. Pure metals are best conductors and non- metals are the poorest.

– Slide: 19/51

Page 20: Validation of Sterilization Equipment.pdf

Thermodynamical Aspects of Heating Process:3. Radiant Heating Process:• Radiant heating is the process which

energy flows from high temperature body to a lower temperature.

• The geometry of both source and the exposure section of the unit will affect the uniformity of the radiation density in a unit.

– Slide: 20/51

Page 21: Validation of Sterilization Equipment.pdf

Dry Heat Sterilization Equipment Validation

• Batch SterilizersDry Heat Ovens

• Continuous SterilizersSterilization Tunnel

• The basics of the Batch and Continuous sterilizers are mainly the same. Since the continuous (Tunnel) sterilizer validation is more complicated, the topics will concentrate on the Convection continuous process qualification.

– Slide: 21/51

Page 22: Validation of Sterilization Equipment.pdf

Batch & Continuous Processing–1. BATCH PROCESSING:

–Washing–Sterilization Dry Heat Oven (Double Door)

– Filling–Manuel Transfer

–Manuel Transfer–1. CONTINUOUS PROCESSING:

–Washing –Filling

–Continuous Transfer –Continuous Transfer

–Washing –Sterilization (Tunnel) –Filling

– Slide: 22/51

Page 23: Validation of Sterilization Equipment.pdf

• Facility layout, decision of batch or continuous process.

• Utility requirements and specifications.• Pressure differential requirements. • Required capacity of the sterilizer.• Type of materials to be sterilized.• Any requirements for presterilization.

Dry Heat Sterilization Validation1. Design Qualification: – Slide: 23/51

Page 24: Validation of Sterilization Equipment.pdf

Dry Heat Sterilization Validation2. Installation Qualification:

• The equipment should comply with the original purchase specifications.

• Exceptions should be appropriately documented.

• The structural installation like; Leveling, insulation, and air flow requirements should meet manufacturer’s specifications.

– Slide: 24/51

Page 25: Validation of Sterilization Equipment.pdf

• All utility connections such as electrical and HVAC should meet the design specifications.

• Materials of construction of both the sterilizer and the facility should meet the design specifications.

Dry Heat Sterilization Validation2. Installation Qualification: – Slide: 25/51

Page 26: Validation of Sterilization Equipment.pdf

The following pieces of equipment should be calibrated by removing or in situ:

• Temperature sensors and recording devices• Temperature Controllers (in situ)• Pressure gauges• Belt speed controller and recorder• Cycle set point switches• Velometers

Dry Heat Sterilization Validation2. Installation Qualification- CALIBRATIONS:

– Slide: 26/51

Page 27: Validation of Sterilization Equipment.pdf

• The actual operational performance of the electro/mechanical components should be verified and documented.

• Electrical Logic: Ensure that each step is in the correct sequence and it’s repeatable.

• Cycle Set Point Adjustability: Limit Switch sequencing should be verified.

Dry Heat Sterilization Validation3. Operational Qualification: – Slide: 27/51

Page 28: Validation of Sterilization Equipment.pdf

• Overload interlocks: Should not allow excess commodity build- up during processing.

• Gasket Integrity: Zone to zone leak rate should be within the limits at all panel gaskets.

• Air Balance Ability: Check that, the baffle/linkage mechanisms can be adjusted for balance.

Dry Heat Sterilization Validation3. Operational Qualification: – Slide: 28/51

Page 29: Validation of Sterilization Equipment.pdf

• Blower Rotation: Check that the blowers rotated in the specified direction and speed.

• Vibration Analysis: Check the dynamic balancing of the blowers to minimize the vibration in each phase.

• Air Balance: Check that the ∆P is positive with respect to the preparation side of the tunnel.

Dry Heat Sterilization Validation3. Operational Qualification: – Slide: 29/51

Page 30: Validation of Sterilization Equipment.pdf

Tunnel Sterilizer Pressure Differential

– Slide: 30/51

Page 31: Validation of Sterilization Equipment.pdf

• Heater Elements: Check that all the heater elements are properly operating.

• Belt Speed: Check that the belt and belt speed recorder are operable.

• HEPA Filters: Verify the integrity of the filters.

Dry Heat Sterilization Validation3. Operational Qualification: – Slide: 31/51

Page 32: Validation of Sterilization Equipment.pdf

• In a conductive dry heat sterilization and depyrogenation process, significant variations may occur depending on the load configuration.

• Initial load temperature, specific heat of the load components, and the load variations should be tested for delta temperature and slowest to heat zone.

Dry Heat Sterilization Validation4. Performance Qualification: – Slide: 32/51

Page 33: Validation of Sterilization Equipment.pdf

Temperature Distribution: • External monitoring and recording instruments

shall be calibrated before and after the OQ/PQ studies (3 point calibration, ± 0.5 0C tolerance).

• Uniformity of the temperature distribution in case of Min&Max. loading should be verified by using Thermocouples with 3 replicates.

• T/C (Thermocouple) placement shall be documented on a diagram.

Dry Heat Sterilization Validation4. Performance Qualification: – Slide: 33/51

Page 34: Validation of Sterilization Equipment.pdf

Temperature Distribution: • Min. 10-12 T/C ‘s shall be used and they

should not be inserted in the load. Data should be recorded during the whole cycle at 1 min. intervals.

• At least one T/C shall be placed adjacent to the equipment temperature controller.

• Location of the “cold spot” should be determined and documented.

Dry Heat Sterilization Validation4. Performance Qualification: – Slide: 34/51

Page 35: Validation of Sterilization Equipment.pdf

Heat Penetration- Acceptance Criteria: • Thermocouples should be inserted into the load. • At least three biological indicators and T/C’s shall be

placed around the cold spot. • External T/C readings should comply with

manufacturer’s specifications (with Max ± 3 0C difference)

• Biological indicator inactivation results should assure 6-log reduction for Bacillus Subtilis and 3-log reduction for endotoxin. Lethality calculation should verify the Equivalent FH value for defined cycle.

Dry Heat Sterilization Validation4. Performance Qualification: – Slide: 35/51

Page 36: Validation of Sterilization Equipment.pdf

HEAT PENETRATION STUDIES

– Slide: 36/51

Page 37: Validation of Sterilization Equipment.pdf

HEAT PENETRATION STUDIESLOAD CONFIGURATION

a13 b13

b14 a14

7

a15 b15

b16 a16

8

a17 b17

b18 a18

9

a7 b7

b8 a8

4

a9 b9

b10 a10

5

a11 b11

b12 a12

6

a1 b1

b2 a2

1

a3 b3

b4 a4

2

a5 b5

b6 a6

3

END MIDDLE START

S7 S4 S1

S8 S5 S2

S9 S6 S3

–S1-S9 :Thermocouples

–a1-a18: Bio-Indicators (Bacillus Subtilis)

–b1-b18:Endotoxin challenged containers.

– Slide: 37/51

Page 38: Validation of Sterilization Equipment.pdf

PART-2

STEAM STERILIZATION VALIDATION

– Slide: 38/51

Page 39: Validation of Sterilization Equipment.pdf

STEAM STERILIZATION

Various types of steam sterilzers are commercially available;

• Saturated Steam• Water Immersion• Water Cascade System• Air-Steam Mixtures• Gravity Air Displacement (unpacked materials

sterilization)• Vacuum air Displacement (Packed materials)

– Slide: 39/51

Page 40: Validation of Sterilization Equipment.pdf

• Microbiological aspects of Steam Sterilization and Dry Heat Sterilization are basically the same;

• D- Value is determined in the same way• Z- Value = 10 0C• Tb = 121 0C• Lethality (F0 ) can be calculated in the same

way.

STEAM STERILIZATION

– Slide: 40/51

Page 41: Validation of Sterilization Equipment.pdf

Time Temperature Lethality Rate(min) (0C) min. at 121 0C

5 100 0,00806 103 0,01607 106 0,03208 109 0,06309 112 0,126010 115 0,251011 118 0,501012 121 1,000013 121 1,000014 121 1,000015 118 0,501016 115 0,251017 112 0,126018 109 0,063019 106 0,032020 103 0,016021 100 0,0080

Lethality in Steam Sterilization

– Slide: 41/51

Page 42: Validation of Sterilization Equipment.pdf

Steam Sterilization Validation1. Design Qualification :

• Facility layout.• Utility requirements and specifications. • Required capacity of the sterilizer.• Type of materials to be sterilized(Liquids, wrapped ,hollow or porous materials)

• Requirement for Gravity and/or Prevacuumcycles.

– Slide: 42/51

Page 43: Validation of Sterilization Equipment.pdf

The following pieces of equipment should be calibrated by removing or in situ:

• Pressure Gauges• Timing Devices• Temperature Recording Devices• Verification of safety Systems and Devices

(EN- 61010 Part- 1 and EN- 61010- 2-041)

Steam Sterilization Validation2. Installation Qualification- CALIBRATIONS:

– Slide: 43/51

Page 44: Validation of Sterilization Equipment.pdf

• The actual operational performance of the electro/mechanical components and utilities should be verified and documented.

• Clean Steam Generator(Free from non- condensables EN- 285)

• Air Filtration Systems and compressed air• Power Source• Heat Exchanger, Cooling Water

Steam Sterilization Validation3. Operational Qualification: – Slide: 44/51

Page 45: Validation of Sterilization Equipment.pdf

Temperature Distribution: • External monitoring and recording instruments

shall be calibrated before and after the OQ/PQ studies (3 point calibration, ± 0.5 0C tolerance).

• Uniformity of the temperature distribution in case of Min&Max. loading should be verified by using Thermocouples with 3 replicates.

• T/C placement shall be documented on a diagram.

Steam Sterilization Validation4. Performance Qualification: – Slide: 45/51

Page 46: Validation of Sterilization Equipment.pdf

Temperature Distribution: • At least one T/C shall be placed located in the

steam exhaust line or adjacent to the equipment temperature controller.

• Min. 10-12 T/C ‘s shall be used and they should not be inserted in the load. Data should be recorded during the whole cycle at 1 min. intervals.

• Location of the “cold spot” should be determined and documented.

Steam Sterilization Validation4. Performance Qualification: – Slide: 46/51

Page 47: Validation of Sterilization Equipment.pdf

Heat Penetration- Acceptance Criteria: • Thermocouples should be inserted into the load. • At least three biological indicators and T/C’s shall be

placed around the cold spot. • External T/C readings should comply with

manufacturer’s specifications (with Max ± 1 0C difference)

• Biological indicator (bacillus stearothermophilus) results should ensure the 6-log reduction and Lethalitycalculation should verify the Equivalent F0 (15 min. at 121 0C) value for defined cycle.

Steam Sterilization Validation4. Performance Qualification: – Slide: 47/51

Page 48: Validation of Sterilization Equipment.pdf

AIR REMOVAL TEST:• The ability of the pre-vacuum autoclaves to

effectively remove the air and non-condensable gases should be tested. If the air is not effectively removed, air pockets will occur in the chamber and sterilization conditions will not be attained.

• Bowie-Dick or DART Test pack, the uniformity of the colour change on the indicator sheet should be checked. (3.5 min. at 134 0C )

Steam Sterilization Validation4. Performance Qualification: – Slide: 48/51

Page 49: Validation of Sterilization Equipment.pdf

LEAK RATE TEST:• The presence of air prevents proper penetration of

of the load by steam and thus inhibits sterilization.• Air leaking from outside into the chamber at the

end of sterilization cycle will contaminate the load. • A leak rate equivalent to a rate of change in

pressure of 1 mm Hg/min. over a period of 10 min. after stabilization is the maximum permitted rate.

Steam Sterilization Validation4. Performance Qualification: – Slide: 49/51

Page 50: Validation of Sterilization Equipment.pdf

STERILIZATION VALIDATION-GENERALCHANGE CONTROL AND REVALIDATION

• Any changes to the sterilization equipment and/or related utilities should be evaluated by a Change Control Procedure.

Typical Changes Requiring Revalidation• Any changes in operating cycle (i.e:temperature ,

time, belt speed, chamber pressure)• Change in load configuration.• Change in sterilized materials. • Major maintenance work on critical

instruments/elements or utilities.

– Slide: 50/51

Page 51: Validation of Sterilization Equipment.pdf

…………..The END…………..

Thanks For Your Attention,Any Questions Please?

………………………………….

– Slide: 51/51