University of Windsor Radiation Safety Bulletin Contamination Monitoring University of Windsor Radiation Safety Bulletin – Contamination Monitoring December 3, 2007 December 3, 2007 The University of Windsor’s Radiation Safety Polices and Canadian Nuclear Safety Commission’s (CNSC) consolidated licence stipulates that all licensed radioisotope laboratories develop and utilize a laboratory specific radioactivity contamination monitoring program. As such, all radioisotope facilities must be monitored for contamination and detailed records maintained supporting the contamination detection program. Methods of Measurement: Direct: Measurements are taken using a portable radiation detection instrument to detect both fixed and removable contamination. Steps: 1. Monitor the locations listed on your plan by slowly moving the detector face toward the surface being monitoring. 2. If contamination is detected, stop and obtain a measurement. Record the reading (Bq). 3. Determine if the reading is above the licence criteria. If so, clean the area. 4. Remonitor the location for each cleaned area and record the final measurement (Bq) after decontamination. Indirect: Measurements are indirectly taken by measuring the amount of radiation which is transferred to a disposable wipe. Steps: 1. Using uniform and constant pressure, wipe approximately 100cm 2 of each area on your plan with a filter paper, wipe, or cotton swab which is lightly moistened with alcohol or water. 2. Count the wipes using a liquid scintillation counter or using a contamination monitor if the wipe is equal or less than the area of the detector. 3. Determine if the readings observed are above the licence criteria. If so, clean the area. 4. Remonitor the location for each cleaned area and record the final measurement (Bq) after decontamination.

Upload: others

Post on 01-Nov-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: University of Windsor Radiation Safety Bulletin

University of Windsor Radiation Safety Bulletin Contamination Monitoring

University of Windsor Radiation Safety Bulletin – Contamination Monitoring 

De

ce

mb

er

3,

20

07

D

ec

em

be

r 3

, 2

00

7

 The University of Windsor’s Radiation Safety Polices and Canadian Nuclear Safety Commission’s (CNSC) consolidated  licence  stipulates  that all  licensed  radioisotope  laboratories develop and utilize  a  laboratory  specific  radioactivity  contamination  monitoring  program.  As  such,  all radioisotope  facilities must be monitored  for contamination and detailed  records maintained supporting the contamination detection program.   Methods of Measurement:  Direct: Measurements are taken using a portable radiation detection instrument to detect both fixed and removable contamination.   Steps: 

1. Monitor  the  locations  listed  on  your  plan  by  slowly moving  the  detector  face  toward  the  surface  being monitoring. 

2. If  contamination  is  detected,  stop  and  obtain  a measurement. Record the reading (Bq).  

3. Determine if the reading is above the licence criteria. If so, clean the area. 

4. Re‐monitor the location for each cleaned area and record the final measurement (Bq) after decontamination. 

 Indirect: Measurements  are  indirectly  taken  by measuring  the  amount  of  radiation which  is transferred to a disposable wipe.   

Steps: 1. Using  uniform  and  constant  pressure,  wipe 

approximately 100cm2 of each area on your plan with a filter  paper,  wipe,  or  cotton  swab  which  is  lightly moistened with alcohol or water.  

2. Count  the  wipes  using  a  liquid  scintillation counter or using a contamination monitor  if the wipe  is equal or less than the area of the detector. 

3. Determine if the readings observed are above the licence criteria. If so, clean the area. 4. Re‐monitor the  location  for each cleaned area and record the  final measurement  (Bq) 

after decontamination.  

Page 2: University of Windsor Radiation Safety Bulletin

University of Windsor Radiation Safety Bulletin Contamination Monitoring

University of Windsor Radiation Safety Bulletin – Contamination Monitoring 

De

ce

mb

er

3,

20

07

D

ec

em

be

r 3

, 2

00

7

 Location, Location, Location: The locations which are to be monitored must be numbered on a plan of the radioisotope work area.  These  surfaces  should  include working  surfaces  (i.e.  benches,  counter  tops,  and  fume hoods)  along  with  non‐working  surfaces  (i.e.  instruments,  door  handles,  and  phones).  In addition, a few random locations should also be included.   Frequency: All radioisotope laboratories actively using radioactive materials must be monitored at least on a weekly basis to detect surface contamination.   If no radioisotopes have been used since the previous survey, permit holders are not required to conduct contamination monitoring until the next usage. However,  it should be recorded  in the “Radioisotope Inventory Record” log book that no radioactive work has been conducted.   Decontamination: Any area that is found to have non‐fixed contamination exceeding the regulatory criteria must be cleaned and re‐monitored.  Please refer to your University of Windsor Radioisotope Permit for more information or visit our website (www.uwindsor.ca/ccc).   

Radionuclide EQ 

(MBq) 

Wipes Controlled 

area (Bq/cm2) 

Wipes Public area (Bq/cm2) 

C‐14  100  300  30 H‐3  1000  300  30 P‐32  0.01  300  30 S‐35  100  300  30 

Please see permit for specific information  Monitoring Records: The  results  of  all  wipe  tests must  be  kept  within  the  log‐book  for  up‐to  three  years.  The “Radioisotope  Inventory  Record”  should  be  kept within  the  licensed  area  and  available  for inspection  by  either  the  RSO  or  CNSC  Inspectors.  Please  see  appendix  for  an  example  of  a monitoring record sheet.     

Page 3: University of Windsor Radiation Safety Bulletin

University of Windsor Radiation Safety Bulletin Contamination Monitoring

University of Windsor Radiation Safety Bulletin – Contamination Monitoring 

De

ce

mb

er

3,

20

07

D

ec

em

be

r 3

, 2

00

7

  Calculations:  

  Example Calculation:  

 

Formula

Removable activity (Bq/cm2):

ValueNNBE60AF

Detector Efficiencies:Perkin‐Elmer Tri‐Carb© 2900 liquid scintillation analyzer 

Isotope EfficiencyC‐14 95%H‐3 58%P‐32 98%S‐35 95%

N ‐  NB / E x 60 x A x (F)

DescriptionTotal count rate in counts per minute (CPM) measured directly or on the wipeNormal background count rate (CPM) from the survey insturment or ont the blankInstrument efficiency factor (%) for the radioisotope being measuredSec/minArea wiped (not to exceed 100 cm2) or the area of the detectorCollection factor of the wipe (assume 10%)

Formula

Removable activity (Bq/cm2):

N ‐  NB / E x 60 x A x (F)

N 157NB 90E 98%

60 60A 100F 10%

Removable activity (Bq/cm2): 0.113945578

Liquid Scintiallation Calculations

Isotope: P32

Page 4: University of Windsor Radiation Safety Bulletin

PersonalWork A

reaResults (Bq/cm

2)Survey M

eter User: M

fg., Model, 

Serial No, &

 Probe Type

University of W

indsor Radiation Safety Program ‐ Laboratory Survey Log

Date

Time

Surveyor's Nam

eIsotope U

sedSurvey Type

* For more inform

ation, please refer to the University of W

indsor's Radiation Safety Program website (w

ww.uw

indsor.ca/radiation)

Page 5: University of Windsor Radiation Safety Bulletin
Page 6: University of Windsor Radiation Safety Bulletin
Page 7: University of Windsor Radiation Safety Bulletin
Page 8: University of Windsor Radiation Safety Bulletin
Page 9: University of Windsor Radiation Safety Bulletin
Page 10: University of Windsor Radiation Safety Bulletin