ultrasound therapy systems service mmanual

39
SERVICE M MANUAL Ultrasound Therapy Systems ISO 13485 CERTIFIED Ultrasound Therapy System Models: 2776 - Serial numbers 1000 and above 2782 - Serial numbers 1000 and above 2792 - Serial numbers 1000 and above ®

Upload: others

Post on 03-Feb-2022

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Ultrasound Therapy Systems SERVICE MMANUAL

SERVICE MMANUALUltrasound Therapy Systems

ISO 13485 CERTIFIED

Ultrasound Therapy SystemModels:2776 - Serial numbers 1000 and above2782 - Serial numbers 1000 and above2792 - Serial numbers 1000 and above

®

Page 2: Ultrasound Therapy Systems SERVICE MMANUAL
Page 3: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy SystemsTABLE of CONTENTSForeword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

1- Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

2- Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

3- Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5

Intelect/Genisys Ultrasound Therapy System . . . . . . . . . . . . .4

Intelect/Genisys Ultrasound Symbol Definitions . . . . . . . . . . .5

4- Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6

5- Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14

5.1 Intelect/Genisys Ultrasound Software Error Messages .7-9

5.2 Intelect/Genisys Ultrasound Diagnostics . . . . . . . . . . . .10

5.3 Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11

5.4 Leakage Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11

5.5 Unit Startop and Fan Testing . . . . . . . . . . . . . . . . . . . . .11

5.6 Ultrasound Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

5.7 Ultrasound Applicator Identification Test . . . . . . . . . . . . .12

5.8 Ultrasound Applicator Output Test . . . . . . . . . . . . . . . . .13

5.9 Ultrasound Duty Cycle Test . . . . . . . . . . . . . . . . . . . . . .14

6- Removal & Replacement . . . . . . . . . . . . . . . . . . . . . . . .15-20

6.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15

6.2 Top & Bottom of Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . .15

6.3 Fan Removal & Replacement . . . . . . . . . . . . . . . . . . . . .16

6.4 Power Supply Removal & Replacement . . . . . . . . . . . . .17

6.5 Ultrasound Board Removal & Replacement . . . . . . . . . .18

6.6 LCD Display Removal & Replacement . . . . . . . . . . . . . .19

6.7 Control Board Removal & Repalcement . . . . . . . . . . . . .20

7- Ultrasound Applicator Calibration . . . . . . . . . . . . . . . . . .21

8- Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22-27

9- Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-33

10- Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34

©2004 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without express written consent from the Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by the Chattanooga Group of EncoreMedical, L.P.

Page 4: Ultrasound Therapy Systems SERVICE MMANUAL
Page 5: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

1

FOREWORDRead, understand and follow the Safety Precautions and information contained in this manual.

This manual contains the necessary safety, and field service information for those Field Service Technicians,approved by Chattanooga Group, to perform field service on the Intelect/Genisys Ultrasound Therapy Systems.

At the time of publication the information contained herein was current and up to date. However, due tocontinual technological improvements and increased clinical knowledge in the field of electrotherapy, as well asChattanooga Group’s policy of continual improvement, Chattanooga Group reserves the right to make periodicchanges and improvements to their equipment and documentation without any obligation on the part ofChattanooga Group.

It is the sole responsibility of field technicians to stay informed and trained in the latest technology utilized inthe Intelect/Genisys Ultrasound Therapy Systems by Chattanooga Group. From time to time, as significantimprovements are incorporated, Service Bulletins will be produced and made available on our web site(www.chattgroup.com) in lieu of reprinting a complete manual prematurely. These Service Bulletins will provideupdated service information and technology improvements to the Intelect/Genisys Ultrasound Therapy Systemsfor use by certified service technicians.

“Certified Service Technician” Definitions;

1. Level I- Those Field Service Technicians that have successfully completed the minimal training required byChattanooga Group in basic service techniques.

2. Level II- Those Field Service Technicians that have successfully completed Level I Training as well asLevel II training as required to perform specific troubleshooting and repair techniques and procedures.

3. Level III- Those Field Service Technicians that have successfully completed Levels I & II Training as well asLevel III Advanced Training as required to perform all necessary Troubleshooting and Repair techniques. The Technician having successfully completed the three levels of training and coupled with experience should have the ability to train other technicians in Level I and Level II Training with the necessary Training Materials from Chattanooga Group.

4. Temporary- Chattanooga Group, at its discretion and based on known experience of the technician, may grant a “Temporary Certification” to a field technician for particular troubleshooting and repair of a specific system requiring immediate attention. This “Temporary Certification” in no fashion acknowledges the training level of a technician as defined above. This “Temporary Certification” is utilized only in unique situations for a specific unit for a specific service technique only and is documented as such.

Due to the complex nature of the technology utilized by Chattanooga Group, the recommended troubleshootingtechniques are to determine “Bad Board” and board replacement only. No board component leveltroubleshooting is recommended nor will information or parts be supplied by Chattanooga Group. Any boardcomponent level troubleshooting performed will be at sole risk and liability of the Service Technicianperforming such troubleshooting techniques.

This equipment is to be used only under the prescription and supervision of a licensed medical practitioner.

This equipment is to be serviced only by a “Certified Service Technician”.

Page 6: Ultrasound Therapy Systems SERVICE MMANUAL

1.1 Precautionary Symbol DefinitionsThe precautionary instructions found in this manual areindicated by specific symbols. Understand these symbols and their definitions before operating or servicing this equipment. The definitions of these symbols are as follows;

A. CAUTION

Text with a “CAUTION” indicator will explain possible safety infractions that have the potential to cause minor to moderate injury or damage to equipment.

B. WARNING

Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause serious injury and equipment damage.

C. DANGER

Text with a “DANGER” indicator will explainpossible safety infractions that are imminently hazardous situations that would result in death or serious injury.

D. EXPLOSION HAZARD

Do not use this equipment in the presence of flammable anesthetics. This symbol is also prominently displayed on the serial number plate of the unit.

E. DANGEROUS VOLTAGE

Text with a “Dangerous Voltage” indicator serves to inform the Service Technician of possible hazards resulting in the electrical charge retained by faulty power supplies.

F. CORROSIVE HAZARD (NiMH Battery)Text with a “Corrosive Hazard” indicator will explain possible safety infractions if the chemical components of this product are exposed to air, skin or other materials.

G. NOTE:

Throughout this manual “NOTE” may be found. The Notes are helpful information to aid in the particular area or function being described.

1.2 Safety PrecautionsRead, understand and follow all safety precautions found in this manual. Below are general safety precautions that must be read and understood before attempting any service techniques on these units. Throughout this manual specific safety precautions willbe found. Read, understand and follow all safety precautions.

• Read, understand and practice the precautionary and operating instructions. Know the limitations and hazards associated with using any ultrasound device. Observe the precautionary and operationaldecals placed on the unit.

• DO NOT operate the Intelect or Genisys when connected to any unit other than Chattanooga Group devices. Do not operate the unit in an environment of short-wave diathermy use.

• The Ultrasound modality should be routinely checked before each use to determine that all controls function normally; especially that the intensity control does properly adjust the intensity of the ultrasonic power output in a stable manner. Also, determine that the treatment time control does actually terminate ultrasonic power output when the timer reaches zero.

• Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous exposure to ultrasonic energy.

• DO NOT use sharp objects such as a pencil pointor ballpoint pen to operate the buttons on the control panel as damage may result.

• Operate, transport and store this unit in temperatures between 59 °F and 104 °F (15 °C and 40 °C), with Relative Humidity ranging from30%-60%.

• Inappropriate handling of, and subjecting the ultrasound applicator to physical abuse, may adversely affect its characteristics.

• Inspect Sound Head for cracks, which may allow the ingress of conductive fluid before each use.

• Inspect all cables, leads and associated connectors before each use.

Intelect®/Genisys® Ultrasound Therapy Systems

2

1- SAFETY PRECAUTIONS

! CAUTION

WARNING!

!

! CAUTION

DANGER

Page 7: Ultrasound Therapy Systems SERVICE MMANUAL

2- THEORY of OPERATION2.1 Overview

The Intelect/Genisys Ultrasound Therapy Systems are comprised of several PC board assemblies housed within acommon enclosure. These assemblies each support a distinct function in the product. The basic elements are UserInterface, Control Board, Ultrasound Board, Ultrasound Applicator, and Power Supply Circuits.

2.2 Power Supply Circuits

A universal input power supply provides all parts of the system with 12 volts DC. The supply is connected to the mainsat all times when the cord is attached. The power switch on the top of the unit, switch the 12V output of the powersupply to the different components in the system by way of the Power Distribution Board. The 12V supply is regulatedlocally at each PC board as required.

2.3 Control Board

The Control Board serves just as its name implies. It controls the operation of the ultrasound board, user interface andoptional accessories. The control board communicates to the ultrasound board through a proprietary 8-bit parallel bus.The control board drives the display. The control board reads the menu buttons. The control board also reads theamplitude and the contrast control on the system. Sound output is generated by the control board and routed to aninternal speaker.

2.4 Ultrasound Board and Applicator

The ultrasound board generates the 1 or 3.3 MHz output to drive the ultrasound Sound Head. The ultrasoundboard is accessed much like an I/O port by the control board. It can provide current and voltage informationabout the ultrasound output of the board. The calibration data for the Sound Head is passed through theultrasound board from the applicator up to the control board. By storing the calibration data in the applicatorthere is no calibration necessary for the ultrasound board. Calibrated Chattanooga Group Ultrasound Applicators,designed specifically for the Intelect/Genisys Ultrasound Therapy Systems, can be connected and operated toprovide accurate coupling and output.

2.5 User Interface and Accessories

The LCD display panel provides the operator visible feedback in the way of menu choices. Pressing the menubuttons makes selections from the menus. The control board interprets these user inputs and respondsaccordingly. Audible feedback is given as well for events such as keypresses and end of treatment.

The Optional Battery mounts in the system bottom and supplies 24 VDC to the power supply. The Power Supplydistributes the power required to the system as needed. The charging circuit for the battery is incorporated on theControl PCB. An icon on the display reflects the available amount of charge in the battery. When the system isconnected to an approved power outlet and a battery is installed in the system, the system will charge the batteryto full capacity. The incorporated charger will not overcharge the battery.

Intelect®/Genisys® Ultrasound Therapy Systems

3

Page 8: Ultrasound Therapy Systems SERVICE MMANUAL

3- NOMENCLATURE3.1 Intelect/Genisys Ultrasound Therapy SystemThe nomenclature graphics below, Figure 3.1, indicatesthe general locations of the exterior components of theUltrasound Therapy System.

Know the components and their functions beforeperforming any operation on or service to the UltrasoundTherapy System.

FIGURE 3.1

1. On/Off Switch- Turns system On and Off.2. Fan Grill- Allow exhaust of air circulated by system fan.3. LCD Contrast Knob-Adjusts contrast of LCD Display.4. Plynth- Allows for wall mounting or sitting on a flat surface.5. Ultrasound Applicator- Used to administer Ultrasound

Therapy to patient.6. Applicator LED Indicator- Flashes 5 times upon connecting

applicator to system. Illuminates constantly when applicatoris coupled to patient treatment area.

7. Ultrasound Intensity Button- Used to increase anddecrease ultrasound output intensity.

8. Ultrasound Applicator Hook- Used to store applicator whennot in use.

9. Ultrasound Applicator Connector- Used to connect theUltrasound Applicators to the System

10. Treatment Start Button- Press to start therapy session.

11. Treatment Pause Button- Press to pause therapy session.Press again to continue session.

12. Treatment Stop Button- Press to completely stop therapysession.

13. Multi-Control Button- Press the respective icon on thisbutton to change parameters and to scroll through ClinicalResources Library

14. Treatment Time Button- Press to adjust treatment time upor down.

15. Clinical Resources Library Button- Press to access theClinical Resources Library.

16. LCD Display- Displays necessary information for operationand maintenance.

17. Mains Power Cord- Connects system to required electricaloutlet.

Not Illustrated- Battery Access Cover located beneath theplynth.

Intelect®/Genisys® Ultrasound Therapy Systems

4

12

3

4

5

6

7

8

9

1011

12

13

14

15

16

17

Page 9: Ultrasound Therapy Systems SERVICE MMANUAL

5

3- NOMENCLATURE3.2 Intelect/Genisys Ultrasound Symbol Definitions

The graphic symbols listed below can be found on the Ultrasound Therapy System and within the software.

Become familiar with these symbols and their definitions prior to any operation, mainteneance or repairs of the system.

Intelect®/Genisys® Ultrasound Therapy Systems

On/Off Switch- Flashes when plugged intopower outlet. Illuminates steady when systemis On

Stop Treatment- Press to stop therapy session.

Pause Treatment- Press to pause therapy sessionand retain time remaining. Press again to continuetherapy session

Start Treatment- Press to start therapy session

Ultrasound Applicator- Indicates connector for theultrasound applicators.

Clinical Resources- Press to access the ClinicalResourses area of the system software.

Increase- Found in two locations on the system,press to increase session time and/or ultrasoundintensity respectively.

Decrease- Found in two locations on the system,press to decrease session time and/or ultrasoundintensity respectively.

Charge Level- This icon indicates the chargelevel of the battery, if installed.

Battery Charging- Indicates the system is chargingthe battery when plugged into power outlet.

Frequency Press to change frequency to 3.3 MHz or1MHz

Intensity Display- Press to display “Watts” or“Watts per cm2“ for the output of the ultrasound.

Intensity Control- Press the Increase orDecrease symbol to adjust ultrasound intensityto the prescribed level.

Duty Cycle- Press to select and display the DutyCycle of the Ultrasound as prescribed. Optionsavaliable are: 10%, 20%, 50, or Continuous (100%).

LCD Display Contrast- Located on the Fan Grill, thisindicates the control to increase or decrease theLCD Display contrast.

Treatment Time- Press the Increase or Decreasesymbol to adjust ultrasound treatment time to theprescribed level.

Head Warming- Press to turn Off or On the HeadWarming feature of the system.

Back- Press to return back one screen on the LCDDisplay.

Scroll Up- Press to scroll up when making menuselections

Scroll Down- Press to scroll down when makingmenu selections

Accept & Return- Press to accept menuselection and to return to the Home Screen

Page 10: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

6

4- SPECIFICATIONS4.1 Intelect/Genisys Ultrasound Therapy System

Physical Specifications

A. Dimensions1. 28.8 cm (11.3”) Width

32.8 cm (12.9”) Depth16.3 cm (6.4”) Height

2. 2.3 kg (5.07 lbs) with 5 cm2 Applicator and Plynth3. Case Material: Polycarbonate Plastic4. Optional Battery Weight: 0.85 kg (1.87 lbs)

B.OutputsOne Ultrasound Output

C. Ultrasound Applicator Sizes Available1. 1 cm2

2. 2 cm2

3. 5 cm2 (Standard)4. 10 cm2

D. Power Input Requirements1. 100-240VAC, 50/60 Hz, 75W

4.2 Intelect/Genisys Ultrasound Specifications

Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MHz, +/- 5%; 3.3 Mhz, +/- 5%

Duty Cycles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10%, 20%, 50%, Continuous

Pulse Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100Hz

Pulse Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1mSec, +/-20%2mSec, +/-20%5mSec, +/-20%

Output Power10cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-20 Watts at 1MHz

0-10 Watts at 3.3 MHz5cm2 Crystal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-10 Watts, 1 and 3.3 MHz2cm2 Crystal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-4 Watts, 1 and 3.3 MHz1cm2 Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-2 Watts 3.3 MHz Only

Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 2.5 w/cm2 in continuous mode,. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-3 w/cm2 in pulsed modes

Output accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +/- 20% above 10% of maximum

Temporal Peak to Average Ratios: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2:1, ± 20%, at 50% Duty Cycle5:1, ± 20%, at 20% Duty Cycle9:1, ± 20%, at 10% Duty Cycle

Beam Nonuniformity Ratio. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.0 : 1 maximum

Beam Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Collimating

Effective Radiating Areas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10cm2 Crystal - 8.5cm2, +/- 1.5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 cm2 Crystal - 4.0cm2, +/- 1.0. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2cm2 Crystal - 1.8cm2, +0.2/-0.4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1cm2 Crystal - 0.8cm2, +0.2/-0.4

Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes

Ultrasound Head Warming Feature . . . . . . . . . . . 50% of maximum output until Crystal Temp reaches approx. 38 °C (100 °F)then 10% of maximum or Off to maintain Temperature

Page 11: Ultrasound Therapy Systems SERVICE MMANUAL

5- TROUBLESHOOTING5.1 Intelect/Genisys Ultrasound Software Error Messages

A. The information provided below is intended to aid in defining the Software Error Messages of the UltrasoundTherapy System. Once a particular Error Message is defined the information will also list probable causes and possible remedies. Once the problem area is determined subsequent diagnostics for verification will be necessary to determine “Bad Board”. All diagnostic tests will be to validate a “Bad Board” only. No component level troubleshooting information is or will be provided by Chattanooga Group for field troubleshooting of board components.Error messages in the range of 100 to 199 are

primarily user definable and remedied by following the instructions given by the unit. Error meesages in the ranges of 200- 299 and 300-399, require the attention of a field technician certified by Chattanooga Group in order to perform the necessary diagnostics and properly replace any necessary components.

B. Once a particular PCB has been determined as bad, refer to the appropriate Removal and Replacement Section for the board affected for proper instructions for replacement of the board.

Intelect®/Genisys® Ultrasound Therapy Systems

7

CODE No. TYPE MESSAGE PROBABLE CAUSES POSSIBLE REMEDIESIcon- Sound Head

overtemp1. Sound Head has overheated to a

temperature that could cause damage to the crystal.

A. Turn Off Head Warming feature.B. Allow Sound Head to cool before

continuing therapy.C. Connect a known good Applicator to

the system and try again. If system works properly, re-calibrate Applicator or replace with new Applicator.

Icon- No applicator plugged into system

1. Applicator not plugged into system connector.

2. Applicator connector not completely seated in system connector.

3. Applicator Cable disconnected from Applicator PCB.

4. Applicator cable damaged.

5. Applicator not calibrated

6. Bad Ultrasound Board

A. Plug Applicator into system.

A. Make certain Applicator connector is completely seated in system connector.

A. Connect Applicator Cable to Applicator PCB

A. Replace with new Applicator.

A. Calibrate Applicator.

A. Replace Ultrasound Board

100 WARN_US_UNPLUGGEDTechnician attempted tocalibrate an ultrasoundapplicator but none isplugged into system.

1. Applicator not plugged into system connector.

2. Applicator connector not completely seated in system connector.

3. Applicator Cable disconnected from Applicator PCB.

4. Applicator cable damaged.

A. Plug Applicator into system.

A. Make certain Applicator connector is completely seated in system connector.

A. Connect Applicator Cable to Applicator PCB

A. Replace with new Applicator.

101 WARN_US_APPL_BECAME_UNPLUGGEDApplicator came unpluggedfrom system whileadministering therapy.

1. Applicator not plugged into system connector.

2. Applicator connector not completely seated in system connector.

3. Applicator Cable disconnected from Applicator PCB.

4. Applicator cable damaged.

A. Plug Applicator into system.

A. Make certain Applicator connector is completely seated in system connector.

A. Connect Applicator Cable to Applicator PCB

A. Replace with new Applicator.

Page 12: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

8

5- TROUBLESHOOTING5.1 Intelect/Genisys Software Error Messages (Continued)

CODE No. TYPE MESSAGE PROBABLE CAUSES POSSIBLE REMEDIES

200 ERR_EEPROMError accessing the internalEEPROM used to storesystem configurationsettings and protocols.

1. Bad Control Board A. Replace Control Board.

201 ERR_APPL_SAVE_CAL_DATAError attempting to savedata to applicator EEPROMduring applicatorcalibration.

1. Bad Applicator

2. Bad Applicator Cable

3. Applicator PCB Connector Bad

4. Bad Ultrasound Board

A. Replace with known good Applicator.

A. Replace with known good Applicator.

A. Replace Ultrasound Board.

A. Replace Ultrasound Board.

202 ERR_APPL_NOT_CALLED_OKGeneral error attempting tocalibrate an applicator.

1. Bad Applicator

2. Bad Applicator Cable

3. Applicator PCB Connector Bad

4. Bad Ultrasound Board

A. Replace with known good Applicator.

A. Replace with known good Applicator.

A. Replace Ultrasound Board.

A. Replace Ultrasound Board.203 ERR_APPL_SAVE_

TIMEOUTError attempting to savedata to applicator EEPROMduring applicator calibration

1. Bad Applicator

2. Bad Applicator Cable

3. Applicator PCB Connector Bad

4. Bad Ultrasound Board

A. Replace with known good Applicator.

A. Replace with known good Applicator.

A. Replace Ultrasound Board.

A. Replace Ultrasound Board.

Page 13: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

9

CODE No. TYPE MESSAGE PROBABLE CAUSES POSSIBLE REMEDIES300 CRIT_ERR_US_BOARD

Some type of criticalultrasound board error hasoccurred.

1. Ultrasound Board to Control Board Connecting Header loose or disconnected.

2. Bad Ultrasound Board.3. Bad Control Board.

A. Reseat Connecting Header into Control Board.

A. Replace Ultrasound Board.A. Replace Control Board.

301 CRIT_ERR_US_BOARD_NOT_DETECTEDNo ultrasound board isdetected in unit.

1. Ultrasound Board to Control Board Connecting Header loose or disconnected.

2. Bad Ultrasound Board.3. Bad Control Board.

A. Reseat Connecting Header into Control Board.

A. Replace Ultrasound Board.A. Replace Control Board.

302 CRIT_ERR_US_RESETUltrasound board has resetfor some unknown reason.

1. Ultrasound Board to Control Board Connecting Header loose or disconnected.

2. Bad Ultrasound Board.3. Bad Control Board.

A. Reseat Connecting Header into Control Board.

A. Replace Ultrasound Board.A. Replace Control Board.

303 CRIT_ERR_US_BOARD_READError reading fromultrasound board.

1. Ultrasound Board to Control Board Connecting Header loose or disconnected.

2. Bad Ultrasound Board.3. Bad Control Board.

A. Reseat Connecting Header into Control Board.

A. Replace Ultrasound Board.A. Replace Control Board.

304 CRIT_ERR_US_BOARD_WRITEError writing to ultrasoundboard

1. Ultrasound Board to Control Board Connecting Header loose or disconnected.

2. Bad Ultrasound Board.3. Bad Control Board.

A. Reseat Connecting Header into Control Board.

A. Replace Ultrasound Board.A. Replace Control Board.

305 CRIT_ERR_US_BOARD_CALIBRATION_ERRORError calibrating applicator

1. Ultrasound Board to Control Board Connecting Header loose or disconnected.

2. Bad Ultrasound Board.3. Bad Control Board.

A. Reseat Connecting Header into Control Board.

A. Replace Ultrasound Board.A. Replace Control Board.

5- TROUBLESHOOTING5.1 Intelect/Genisys Software Error Messages (Continued)

Page 14: Ultrasound Therapy Systems SERVICE MMANUAL

5- TROUBLESHOOTING5.2 Intelect/Genisys Ultrasound Diagnostics

A. General1. The following information is intended to aid in

troubleshooting the major components of theUltrasound Therapy Systems to “Board Level”only. These tests are OEM standard testingprocedures and methods used at the factorybefore shipment of any Ultrasound TherapySystem.

2. Due to the complex nature of thetechnology utilized by Chattanooga Group,the recommended troubleshootingtechniques are to determine “Bad Board”and board replacement only. No boardcomponent level troubleshooting isrecommended nor will information or partsbe supplied by Chattanooga Group. Anyboard component level troubleshootingperformed will be at sole risk and liability ofthe Service Technician performing suchtroubleshooting techniques.

B. Special Tools, Fixtures & Materials Required1. Certain tests require the use of special Tools

and/or Fixtures. These will be listed at theparticular test where they are required. Testingwith any other special tool or fixture other thanthose stated could give erroneous readings ortest results. Always perform the tests exactly asstated to ensure accurate results.

2. Any special tools or fixtures required can beobtained through Chattanooga Group, ServiceDepartment.

3. Scope and other standard test equipmentsettings will be listed for each test performed toaid in performing the test to OEM standards andensure proper readings.

4. The troubleshooting and repair of the UltrasoundTherapy Systems should be performed only byauthorized technicians trained and certified byChattanooga Group.

C. Equipment Required1. Oscilloscope and Probes

2. Electrotherapy Service Technician Interface(ESTI) Diagnostics Kit.

3. Digital Multimeter

4. Milliohm Meter

5. Dielectric Withstand (Hi-Pot) and groundresistance tester.

NOTE:Adjust Dielectric Withstand tester to indicate fault with120k Ohm Load across the output when at specified testvoltage.

6. Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter.

NOTE:The Ohmic Industries UPM DT 10 or UPM DT 100Ultrasound Power Meters are the only power meters onthe market that will meet the strict specifications set forthfor Chattanooga Group Ultrasound Products. It isabsolutely necessary for the Field Technician to haveavailable and use the Ohmic Industries UPM DT 10 orUPM DT 100 Power Meter when performing any service orcalibration of any Chattanooga Group Ultrasoundproducts.

7. Degassed Water (< 5 ppm) for Ultrasound PowerMeter.

Recipe(s) for Degassed Water1) Boil Distilled Water for 30 minutes. Place

water in a nonporous container andimmediately cover with cellophane. Allow tocool to room temperature of approximately21°C (70°F). May be refrigerated to aidcooling time.

2) Bring Distilled Water to a boil. Place thecontainer under vacuum for 5 to 10 minutes.

8. Dissolved Oxygen Test Kit. Used to test oxygenlevel of degassed water.

NOTE:Two liter softdrink bottles are ideal storage and transportcontainers for degassed water as they are designed tokeep oxygen out. Do not allow aeration of degassed waterduring transport or filling of the power meter.

Do not use Tap Water or Distilled water in the UltrasoundPower Meter. Use only Degassed Water in order to obtaincorrect test results. The chart below illustrates the oxygencontent of Degassed, Tap and Distilled Water.

9. Ultrasound Applicators (Intelect/Genisys Accessories).

Intelect®/Genisys® Ultrasound Therapy Systems

10

WATER TYPE ppm of Oxygen

Degassed(per recipe 1 or 2)

Less than 5 ppm

Tap Water Up to 35 ppm

Distilled Water Up to 20 ppm

Page 15: Ultrasound Therapy Systems SERVICE MMANUAL

5.3 Visual InspectionA. General

Visually inspect the Ultrasound Therapy System. Avisual inspection can, to an experienced Technician,indicate possible abuse of the system and/orinternal problems.

5.4 Leakage TestsConduct all necessary leakage tests as required per “Chapter 7 Electrical Equipment” of the 1999, or later, edition of the NFPA (National Fire Protection Association) “Health Care Facility” standards. See Figure 5.1.

5.5 Unit Startup and Fan TestingA. Equipment Required

1. N/A

B. Test1. Place unit face up on work surface.

2. Connect power cord to unit and plug into proper power receptacle.

3. Turn system on. Adjust intensity up and press Start. Fan will begin to run after the audible beep tones.

4. Place hand at the back of unit to verify fan is blowing out. See Figure 5.2.

C. Test Results1. Unit will not Start= Unit Failed Test

a) Possible Bad Power Supply

2. Main menu does not display= Unit Failed Test

a) Possible Bad Control Board

b)Possible Bad Power Supply

1). Visually check power button LED. If flashing, Power Supply is good, replace Control Board. If not flashing, replace Power Supply.

3. Fan not blowing outward= Unit Failed Test

a) Fan Blowing Inward

1) Fan wired wrong- Rewire or replace fan

b)Fan not blowing

1) Possible Bad Fan

2) Possible Bad Power Supply

Intelect®/Genisys® Ultrasound Therapy Systems

11

5- TROUBLESHOOTING

WARNING!UNIT FAILING DIELECTRIC WITHSTAND AND/OR LEAKAGETESTS COULD INDICATE SERIOUS INTERNAL SYSTEMPROBLEMS.DO NOT PLACE UNIT BACK INTO SERVICE! SEND UNIT TOFACTORY FOR REPAIR! DO NOT ATTEMPT TO REPAIR INTHE FIELD!

FIGURE 5.1

FIGURE 5.2

VERIFY FAN IS BLOWING OUT

Page 16: Ultrasound Therapy Systems SERVICE MMANUAL

5.6 Ultrasound TestsA. Equipment Required

1. Degassed Water. Refer to page 10 for Degassed Water Recipes.

2. Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter.

3. Dissolved Oxygen Test Kit. Used to test oxygen level of degassed water.

4. Known good Ultrasound Applicator

5.7 Ultrasound Applicator Identification TestNOTE:Use any Ultrasound Applicator for this test.

A. Ultrasound Applicator Identification Test Procedures1. Without Applicator installed, turn unit on.

2. Look at the display center. The Applicator not detected icon should appear. See Figure 5.3.

3. Plug Applicator into Applicator port on the front R.H. corner of the system. Applicator LED should flash five (5) times. See Figure 5.4.

4. Look at the display center. The Applicator detected icon should appear. See Figure 5.5.

B. Ultrasound Applicator Identification Test Results1. Unit operates as described in steps 2 and 4.

a) Unit passed test.

2. “NO CAL” icon displays center of system display.a) Applicator not calibrated.

b) Possible bad Applicator. Re-test with known good Applicator.

c) Possible bad Ultrasound PCB.

3. Not detected icon displayed after ten seconds of Applicator being plugged in.

a) Possible bad applicator. Re-test with known good Applicator.

b) Possible bad Ultrasound Board.

Intelect®/Genisys® Ultrasound Therapy Systems

12

5- TROUBLESHOOTING

FIGURE 5.3

FIGURE 5.4

FIGURE 5.5

Page 17: Ultrasound Therapy Systems SERVICE MMANUAL

5.8 Ultrasound Applicator Output TestPerform this test using all available Applicators for the unitbeing tested.

A. Ultrasound Applicator Output Test Procedures1.Set up Ohmic Instruments UPM DT 10 or UPM DT

100 Ultrasound Power Meter per Operator’sInstructions and fill test reservoir with DegassedWater.

2.Place an Applicator into the Power Meter retainer.As per Ohmic Manual, make certain the Applicator’saluminum face is submerged in the degassed waterand centered directly over the S.S. cone.

3. “Zero” the Power Meter.4. Turn the Ultrasound Therapy System on.5. Press the button until “Continuous” is

displayed in the Duty Cycle area of the display.6. Press the button until “Watts” appears in the

intensity area of the display.7. Increase Intensity for the applicator being tested

according to Figure 5.6 8. Press “Start”.9. Compare Power Meter readings, displayed on

power meter for all the settings for each respective applicator being tested to the values in Figure 5.6.

10. Press the button until “3.3 MHz” is diplayed in the Frequency area of the display. Repeat test and compare readings to Figure 5.6.

B. Ultrasound Applicator Output Test Results1. Output ranges fall within the specified ranges as

listed in Figure 5.6.a) Unit passed test.

2. Readings fall outside specified ranges of Figure 5.6.a)Possible bad Degassed Water in Power Meter.b)Possible use of Power Meter other than Ohmic

Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter.

c)Possible bad or out of calibration Applicator.d)Possible bad Ultrasound Board.

Intelect®/Genisys® Ultrasound Therapy Systems

13

5- TROUBLESHOOTING

WARNING!USE ONLY DEGASSED WATER IN POWER METER FORTESTING ULTRASOUND APPLICATORS. USE OF OTHERTYPES OF WATER WILL CAUSE FALSE TEST RESULTS.SEE PAGE 10 FOR DEGASSED WATER RECIPES.

DO NOT AERATE WATER WHEN FILLING POWER METER.

FIGURE 5.6

APPLICATOR OUTPUT SPECIFICATIONS

APPLICATOR SIZEPOWER SETTING

(WATTS)OUTPUT RANGE

1 cm21

2

0.8 - 1.2

1.6 - 2.4

2 cm2

1

2

4

0.8 - 1.2

1.6 - 2.4

3.2 - 4.8

5 cm2

1

2

5

10

0.8 - 1.2

1.6 - 2.4

4.0 - 6.0

8.0 - 12.0

10 cm2

1

5

10

15

20

0.8 - 1.2

4.0 - 6.0

8.0 - 12.0

12.0 - 18.0

16.0 - 24.0

Page 18: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

14

5- TROUBLESHOOTING5.9 Ultrasound Duty Cycle TestThis test is performed using only the 5 cm2 Applicator.

A. Ultrasound Duty Cycle Test Procedures1. Set up Ohmic Instruments UPM DT 10 or UPM DT

100 Ultrasound Power Meter per Operator’s Instructions and fill test reservoir with Degassed Water.

2. Place an Applicator into the Power Meter retainer. Asper Ohmic Manual, make certain the Applicator’s aluminum face is submerged in the degassed water and centered directly over the S.S. cone.

3. “Zero” the Power Meter.

4. Turn the Ultrasound Therapy System on.

5. Press the button until “Watts” appears in the intensity area of the display.

6. Increase Intensity ( ) to maximum.

7. Press “Start”.

8. Press the button until “10%” is displayed in the Duty Cycle area of the display.

9. Compare Power Meter readings to Figure 5.7 for each of the following Duty Cycle settings listed in thechart.

10. Press the button until “3.3 MHz” is displayed inthe Frequency area of display. Repeat test and compare readings to Figure 5.7.

B. Ultrasound Duty Cycle Test Results1. Duty Cycles fall within the specified ranges as listed

in Figure 5.7.

a) Unit passed test

2. Readings fall outside specified ranges of Figure 5.7.

a) Possible bad Degassed Water in Power Meter.

b) Possible use of Power Meter other than Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter.

c) Possible bad or out of calibration Applicator.

d) Check Ultrasound Board.

e) Check Control Board.

FIGURE 5.7

WARNING!USE ONLY DEGASSED WATER IN POWER METER FORTESTING ULTRASOUND APPLICATORS. USE OF OTHERTYPES OF WATER WILL CAUSE FALSE TEST RESULTS. SEEPAGE 10 FOR DEGASSED WATER RECIPES.

DO NOT AERATE WATER WHEN FILLING POWER METER.

DUTY CYCLE SPECIFICATIONS

APPLICATOR SIZE DUTY CYCLE OUTPUT RANGE

5 cm2

10%

20%

50%

100% (Continuous)

0.8 - 1.2

1.6 - 2.4

4.0 - 6.0

8.0 - 12.0

Page 19: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

15

6- REMOVAL & REPLACEMENT6.1 General

A. Tools & Equipment Required1. #1 Phillips Screwdriver2. Flat Blade Screwdriver

6.2. Top & Bottom of UnitA. Removing Top from Bottom

1. Place unit face down on a grounded work surface.NOTE:Use a clean, soft cloth or cushion underneath the display lensto prevent scratching or damage to the lens.

2. Remove the plynth from the bottom of the system. See Figure 6.1.

3. Remove the Mains Power Cord from the system. See Figure 6.1.

4. Remove Battery Cover with the Flat Blade Screwdriver and remove Battery from the system if equipped. See Figure 6.2.

5. Remove the Rear Fan Grill from the system to gain access to the two rear mounting screws. See Figure 6.2.

6. Remove the two front feet from the system to gain access to the two front screws. See Figure 6.3.

7. Remove the four mounting screws from the bottom housing using the #1 Phillips Screwdriver. See Figure 6.3.

8. Carefully separate the upper and lower housings of the system. It may be necessary to pry the housings apart at the Applicator Hook. See Figure 6.4.

9. Disconnect the Power Supply and Fan Harnesses from the Control PCB of the system. See Figure 6.4.

B. Installing Top to Bottom1. With top laying face down, carefully place the bottom

housing in position over the top housing.2. Connect the Power Supply and Fan Harnesses to

the Control PCB onto their respective connectors.3. Make certain the front insert and the Applicator Hook

are in position, then lightly press housings together to properly seat them.

4. Secure housings together with the four phillips head screws.

NOTE:Do not overtighten the screws. Overtightening will damage thethreads of the brass inserts.

5. Follow Steps 1 - 7 in 6.2, A above in reverse order tocomplete reassembly of system.

WARNING!UNPLUG THE UNIT FROM THE POWER SOURCEAND/OR REMOVE BATTERY BEFORE ATTEMPTINGANY REMOVAL OR REPLACEMENT PROCEDURESTO PREVENT ELECTRICAL SHOCK.

FIGURE 6.1

FIGURE 6.2

FIGURE 6.3

FIGURE 6.4

REMOVEPLYNTH

REMOVE MAINSPOWER CORD

REMOVE BATTERYCOVER AND BATTERY

REMOVEFAN GRILL

REMOVEFRONT FEET

REMOVEFOUR SCREWS

APPLICATORHOOK

CAREFULLY PRY APARTHERE IF NECESSARY

DISCONNECT POWERSUPPLY AND FAN

HARNESSES

Page 20: Ultrasound Therapy Systems SERVICE MMANUAL

6.3 Fan Removal & Replacement

A. Fan Removal1. Separate Top from Bottom. Refer to 6.2, part A.

2. With #1 Phillips screwdriver, remove the 2 screws securing the Fan. See Figure 6.5.

3. Remove Fan from the unit.

4. Remove Fan Baffle from fan. See Figure 6.5.

B. Installing Fan1. Route Fan Harness underneath fan as shown in

Figure 6.6.

2. Install Fan Baffle onto fan as shown in Figure 6.6.

3. Place new Fan, Part Number 27266, into system base.

4. Make certain the Fan Baffle sits onto System Housing Baffle Tabs as shown in Figure 6.7.

5. Secure fan to housing with the 2 mounting screws. See Figure 6.8.

NOTE:Do not overtighten the screws. Overtightening will damage thethreads of the brass inserts.

6. Reassemble the Top and Bottom Housing. Refer to 6.2, part B.

Intelect®/Genisys® Ultrasound Therapy Systems

16

WARNING!UNPLUG THE UNIT FROM THE POWER SOURCEBEFORE ATTEMPTING ANY REMOVAL ORREPLACEMENT PROCEDURES TO PREVENTELECTRICAL SHOCK.

6- REMOVAL & REPLACEMENT

FIGURE 6.5

FIGURE 6.6

REMOVE SCREWS

FIGURE 6.7

FIGURE 6.8

FANBAFFLE

FANBAFFLE

ROUTEHARNESSAS SHOWN

BAFFLETABS

SECURELY TIGHTENMOUNTING SCREWS

Page 21: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

17

6- REMOVAL & REPLACEMENT

6.4 Power Supply Removal & Replacement

A. Power Supply Removal1. Separate Top from Bottom. Refer to 6.2, part A.

2. Discharge Power Supply Capacitor, C4, with multimeter.

3. With #1 Phillips screwdriver, remove the 2 screws securing the Power Supply. See Figure 6.9.

4. Lift Power Supply and disconnect the Power Supply Harnesses from the Mains Disconnect. See Figure 6.10.

5. Remove Power Supply from system.

B. Installing Power Supply1. Route Power Supply PCB Harness of new Power

Supply, Part Number 27265, as shown in Figure 6.11.

2. Connect Power Supply Mains Harness to Mains Disconnect as shown in Figure 6.10

3. Make certain the Power Supply sits onto alignment pins as shown in Figure 6.12.

5. Secure Power Supply to housing with the 2 mounting screws. See Figure 6.9.

NOTE:Do not overtighten the screws. Overtightening will damage thethreads of the brass inserts.

6. Reassemble the Top and Bottom Housing. Refer to 6.2, part B.

FIGURE 6.9

FIGURE 6.10

FIGURE 6.11

FIGURE 6.12

REMOVE SCREWS

BLUE

REDGREEN WITHYELLOW STRIPE

ROUTE HARNESS AS SHOWN

POWER SUPPLYALIGNMENT PINS

WARNING!UNPLUG THE UNIT FROM THE POWER SOURCEBEFORE ATTEMPTING ANY REMOVAL ORREPLACEMENT PROCEDURES TO PREVENTELECTRICAL SHOCK.

DANGERIF POWER SUPPLY HAS FAILED HI-POT TEST,

IT WILL RETAIN HIGH VOLTAGE IN THE POWER SUPPLY CAPACITOR C4. DISCHARGE

CAPACITOR USING A MULTIMETER PRIOR ANYSERVICE OF THE POWER SUPPLY

Page 22: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

18

6.5 Ultrasound Board Removal & Replacement

A. Ultrasound Board Removal1. Separate Top from Bottom. Refer to 6.2, part A.

2. With #1 Phillips screwdriver remove the 4 screwssecuring the Ultrasound Board to the unit. See Figure 6.13.

3. Carefully remove Ultrasound Board from the Control Board/Ultrasound Board Connector. See Figure 6.14.

4. Remove Control Board/Ultrasound Board Connector from the Control Board. Be careful not to bend or damage any of the pins. See Figure 6.14.

B. Installing Ultrasound Board1. Install Control Board/Ultrasound Board Connector

into new Ultrasound Board, Part Number 27269.

NOTE:Install the connector with the long pins inserted into UltrasoundBoard. Make certain connector is completely seated againstUltrasound Board. See Figure 6.15.

2. Place new Ultrasound board as shown in Figure 6.16 over Control Board and align connector pins up with Control Board Connector.

3. Carefully press the connector and board into place until the connector is completely seated into Control Board.

4. Secure Ultrasound Board with the 4 retaining screws.

NOTE:Do not overtighten the screws. Overtightening will damage thethreads of the brass inserts.

5. Reassemble the Top and Bottom Housing. Refer to 6.2, part B.

6- REMOVAL & REPLACEMENT

FIGURE 6.13

WARNING!UNPLUG THE UNIT FROM THE POWER SOURCEBEFORE ATTEMPTING ANY REMOVAL ORREPLACEMENT PROCEDURES TO PREVENTELECTRICAL SHOCK.

FIGURE 6.14

FIGURE 6.15

FIGURE 6.16

SCREW SCREW

SCREWSCREW

REMOVE US BOARDFROM CONNECTOR

REMOVE CONNECTORFROM CONTROL BOARD

INSTALL CONNECTOR TOUS BOARD AS SHOWN

ALIGN CONNECTOR PINSWITH CONTROL BOARD

CONNECTOR

Page 23: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

19

6.6 LCD Display Removal & Replacement

A. LCD Display Removal1. Separate Top from Bottom. Refer to 6.2, part A.

2. Remove Ultrasound Board. Refer to 6.7, part A.

3. Remove the 4 LCD retaining screws. See Figure 6.17.

4. With a flat blade screwdriver, gently push in and up on each end of the LCD retaining brackets to free the locking tabs from the Control Board. See Figure 6.17.

5. Set retaining brackets aside for reassembly.

6. Carefully remove the LCD from the Control Board being careful not to damage the LCD connector pins.See Figure 6.18.

B. Installing LCD Display1. Reinstall the two LCD Retaining Brackets onto new

LCD Display, Part Number 27264. See Figure 6.19.

2. Place LCD Display over Control Board and align the LCD Connector pins with the top row on the Control Board LCD Connector. See Figure 6.18.

3. Secure LCD Display with the 4 retaining screws. See Figure 6.17.

NOTE:Do not overtighten the screws. Overtightening will damage thethreads of the brass inserts.

4. Reinstall the Ultrasound Board. Refer to 6.7, part B.

5. Reassemble the Top and Bottom Housing. Refer to 6.2, part B.

6- REMOVAL & REPLACEMENT

FIGURE 6.17

WARNING!UNPLUG THE UNIT FROM THE POWER SOURCEBEFORE ATTEMPTING ANY REMOVAL ORREPLACEMENT PROCEDURES TO PREVENTELECTRICAL SHOCK.

FIGURE 6.18

FIGURE 6.19

2 SCREWSEACH END

RELEASE RETAININGBRACKET TABS

REMOVE LCD FROMCONTROL BOARD

INSTALL RETAININGBRACKETS TO LCD

Page 24: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

20

6.7 Control Board Removal & Replacement

A. Control Board Removal1. Separate Top from Bottom. Refer to 6.2, part A.

2. Remove Ultrasound Board. Refer to 6.7, part A.

3. Remove LCD from Control Board. Refer to 6.6, part A.

4. Remove the 7 retaining screws securing the control board to the housing. See Figure 6.20.

5. Using a flat blade screwdriver, carefully push locking tabs while lifting the Control Board to release ControlBoard from Housing. See Figure 6.21.

6. Remove the LCD Contrast knob for installation on the new Control PCB. See Figure 6.22.

B. Installing Control Board1. Place new Control Board, Part Number 27249, over

the locking tabs and push Control PCB down until locking tabs capture the Control PCB.

2. Secure Control PCB to housing with 7 retaining screws. Refer to Figure 6.20.

NOTE:Do not overtighten the screws. Overtightening will damage thethreads of the brass inserts.

3. Install LCD Display Contrast Knob. See Figure 6.22.Make certain Knob is completely seated on potentiometer.

4. Install LCD to Control PCB. Refer to 6.6, part B.

5. Install Ultrasound Board. Refer to 6.7, part B.

6. Re-assemble Top to Bottom. Refer to 6.2, part B.

6- REMOVAL & REPLACEMENT

FIGURE 6.20

WARNING!UNPLUG THE UNIT FROM THE POWER SOURCEBEFORE ATTEMPTING ANY REMOVAL ORREPLACEMENT PROCEDURES TO PREVENTELECTRICAL SHOCK.

FIGURE 6.21

FIGURE 6.22

RETAINING SCREWS

CONTRAST KNOB

PUSH TABS IN ANDPULL UP ON BOARDSIMUTANEOUSLY

Page 25: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

21

7.1 GeneralA. Tools and Equipment Required

1. Ultrasound Therapy System being serviced.

2. All Ultrasound Applicators for the unit being serviced.

3. Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter, set to “watts”.

4. Degassed Water. Refer to page 10 for Degassed Water Recipes.

B. Ultrasound Applicator Calibration Procedures1. Perform the following on all Ultrasound Applicators

for the unit being serviced at least annually.

2. With the system on, press the Clinical Resources button once. See Figure 7.1.

3. Simultaneously press and hold the Treatment Time and Intensity Down buttons for approximately 2 seconds. See Figure 7.2. The calibration procedures screen should display.

4. Press the icon once to highlight “Calibrate Applicator”.

5. Press the icon once to select.

6. Place the Ultrasound Applicator being calibrated into the Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter and set meter to “watts”. See Figure 7.3.

7. Follow the instructions on the LCD Display.

8. When calibration is complete, place another applicator on the Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter and repeat steps 1 through 6 above.

7- ULTRASOUND APPLICATOR CALIBRATION

WARNING!USE ONLY DEGASSED WATER IN POWER METER FORCALIBRATING ULTRASOUND APPLICATORS.

USE OF OTHER TYPES OF WATER WILL CAUSE FALSEREADINGS AND BAD TEST RESULTS.

SEE PAGE 10 FOR DEGASSED WATER RECIPES.

USE OF OTHER BRANDS OR TYPES OF TOOLS,EQUIPMENT, FIXTURES, MATERIALS AND SUPPLIESOTHER THAN THOSE SPECIFICALLY LISTED IN “A. Toolsand Equipment Required” ABOVE WILL GIVE BAD TESTAND CALIBRATION RESULTS.

IF PROPER EQUIPMENT IS NOT AVAILABLE OR CAN NOTBE OBTAINED, SEND THE ULTRASOUND APPLICATORSTO THE FACTORY FOR CALIBRATION.

FIGURE 7.3

FIGURE 7.1

FIGURE 7.2

SET METERTO WATTS

Page 26: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

22

8- PARTS

1 & 2(See Page 24)

3 & 4

Power Supply Connector

5

5

6

7 (Optional)

8

9

Battery Connector

10

5

12

Fan Connector

11(See Pages26 - 27)

Ultrasound Assembly

Page 27: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

23

8- PARTS

ITEMNUMBER

PART NUMBER

DESCRIPTION QTYREQ’D

12

2740227247

Top Assembly (Grey)- See page 24 for breakdownTop Assembly (Blue)- See Page 24 for breakdown

11

34

2736127363

Rear Vent (Grey)Rear Vent (Blue)

11

5 27142 M3 x 6mm S.S. Pan Head Screw 46 27365 Right Foot 17 27267 Battery Pack (Optional) 18 27410 Battery Compartment Cover Assembly 19 27253 Plynth 1

10 27363 Left Foot 111 27409 Bottom Assembly- See Pages 26-27 for breakdown 112 27373 Baffle 1

Ultrasound Assembly

Page 28: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

24

8- PARTS

Ultrasound Top Assembly

1 & 1A

2 & 2A

3 & 3A

4(See Page 25)

5

5

6

7

ITEMNUMBER

PART NUMBER

DESCRIPTION QTYREQ’D

11A

2755227553

Top Moulding Assembly (Grey)Top Moulding Assembly (Blue)

11

22A

2737027262

Power Button (Grey)Power Button (Blue)

11

33A

2736827561

Keymat (Grey)Keymat (Blue)

11

4 27249 Control PCB Assembly- See Page 25 15 27142 M3 x 6mm S.S. Pan Head Screw 156 27588 PCB Connector- Ultrasound PCB to Control PCB 17 27269 Ultrasound PCB Assembly 1

Page 29: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

25

8- PARTS

Ultrasound Control PCB Assembly

ITEMNUMBER

PART NUMBER

DESCRIPTION QTYREQ’D

1 27268 Control PCB 12 27012 Contrast Knob 13 27264 Transportables LCD 14 27261 LCD Spacer 2

1

2

3

4

4

Page 30: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

26

8- PARTS

Ultrasound Bottom Assembly

1

2

3

4

5

4

6

9

8

7

Page 31: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems

27

8- PARTS

Ultrasound Bottom Assembly

ITEMNUMBER

PARTNUMBER

DESCRIPTION QTYREQ’D

1 27136 M4 x 35mm S.S. Pan Head Screw 22 27266 Fan 13 27367 Fan Seal 14 27142 M3 x 6mm S.S. Pan Head Screw 25 27265 Power Supply 16 27277 Universal Snap-In Inlet 17 27256 Applicator Hang-Up 18 27252 Base 19 27259 Ultrasound Connector Infill 1

Page 32: Ultrasound Therapy Systems SERVICE MMANUAL

28

Inte

lect

®/G

enis

ys®

Ult

raso

un

d T

her

apy

Sys

tem

s9-

SCH

EMAT

ICS

US+

US+

US-

US-

DG

ND

_APP

LD

GN

D_A

PPL

VCC

_APP

LVC

C_A

PPL

SDA_

APPL

SDA_

APPL

SCL_

APPL

SCL_

APPL

MC

LRM

CLR

11 22

C6

C6

1.0u

F/16

V1.

0uF/

16V

11 22

C5

C5

0.1u

F0.

1uF

11

2233

Q4

Q4 M

MBT

3904

MM

BT39

04

11 22

C4

C4

10nF

10nF

1122

R3R3 10K

10K

1122

R7R7

10K

10K

1111

PIN

3PI

N3

100P

AD10

0PAD

VDD

VDD

11

GP5

/T1C

KI/O

SC1/

CLK

ING

P5/T

1CKI

/OSC

1/C

LKIN

22

GP1

/AN

1/C

IN-//V

REF/

ICSP

CLK

GP1

/AN

1/C

IN-//V

REF/

ICSP

CLK

66G

P4/A

N3/

T1G

/OSC

2/C

LKO

UTG

P4/A

N3/

T1G

/OSC

2/C

LKO

UT33

GP3

/MC

LR/V

PPG

P3/M

CLR

/VPP

44

GP0

/AN

0/C

IN+

/ICSP

DAT

GP0

/AN

0/C

IN+

/ICSP

DAT

77

GP2

/AN

2/T0

CKI

/INT/

CO

UTG

P2/A

N2/

T0C

KI/IN

T/C

OUT

55

VSS

VSS

88

U1U1

PIC

12F6

75-II/

SNPI

C12

F675

-II/SN

11 22

RT1

RT1

100K

100K

11 22

R4R4 42.2

K 11%

42.2

K 11%

33

2244

66

7755

1188

K1K1

G6K

-22F-

YY5VD

CG

6K-22

F-YY5

VDC

2211

33

D1

D1

BAS1

6_N

LBA

S16_

NL

F1F1

11 22

C3

C3

0.1u

F0.

1uF

1122

R5R5 220

220

11 22

R6R6 10K

10K

1111

PIN

2PI

N2

100P

AD10

0PAD

L2L2

220u

H22

0uH

1111

PIN

1PI

N1

1111

PAD

1PA

D1

50PA

D50

PAD11

11

PAD

2PA

D2

50PA

D50

PAD11

11

PAD

3PA

D3

50PA

D50

PAD11

11

PAD

4PA

D4

50PA

D50

PAD

1122

3344

5566

7788

991010

JP1

JP1

Tra

nsd

ucer

Tra

nsd

ucer

Ca

se CC

onne

cted

Ca

se CC

onne

cted

Nex

t tto

PPin

11N

ext tt

o PPi

n 11

Ther

ma

l ccon

nect

Ther

ma

l ccon

nect

1-

NN/C

1-

NN/C

2,4-

UUS

+2,

4-

UUS+

6,8-

UUS

-6,

8-

UUS-

3-

MMC

LR3-

MM

CLR

5-

++5V

5-

++5V

7-

IIC2

DDAT

A7-

IIC

2 DD

ATA

9-

IIC2

CCLO

CK

9-

IIC2

CCLO

CK

10-

AAppp

licat

or BB

ody

10-

AApp

licat

or BB

ody

NO

T CC

ase

NO

T CC

ase

Con

nect

edC

onne

cted

2211

LED

1LE

D1

4433

55778811 22

66

T1T1

EFD

15EF

D15

C1

C1

10nF

10nF

C2

C2

10nF

10nF

11 22

R1R1 1K1K

11 22

R2R2 1K1K

C7

C7

10nF

10nF

C8

C8

10nF

10nF

C9

C9

330p

F/20

0V33

0pF/

200V

5 cm

2Ul

tras

ound

App

licat

orSc

hem

atic

Page 33: Ultrasound Therapy Systems SERVICE MMANUAL

29

Inte

lect

®/G

enis

ys®

Ult

raso

un

d T

her

apy

Sys

tem

s9-

SCH

EMAT

ICS

Cont

rol B

oard

- 272

68Sc

hem

atic

1 o

f 2

Page 34: Ultrasound Therapy Systems SERVICE MMANUAL

30

Inte

lect

®/G

enis

ys®

Ult

raso

un

d T

her

apy

Sys

tem

s9-

SCH

EMAT

ICS

Cont

rol B

oard

- 272

68Sc

hem

atic

2 o

f 2

Page 35: Ultrasound Therapy Systems SERVICE MMANUAL

31

Inte

lect

®/G

enis

ys®

Ult

raso

un

d T

her

apy

Sys

tem

s9-

SCH

EMAT

ICS

11 22

C1

C1

100p

F/10

0V10

0pF/

100V

1122

C5

C5

0.1u

F/50

V0.

1uF/

50V

11 22

C6

C6

0.1u

F/50

V0.

1uF/

50V 11 22

C18

C18

0.1u

F0.

1uF

11 22

C2

C2

100p

F/10

0V10

0pF/

100V

11

44 33

Q4

Q4

IRLR

110

IRLR

110

11

44 33

Q7

Q7

IRLR

110

IRLR

110

11 22

R10

R10

1.0M

1.0M

11 22

R2R2 4747

1122

R1R1 4747

11 22

R11

R11

1.0M

1.0M

1122

R9R9 2.2

2.2

1122

R12

R12

2.2

2.211 22

C3

C3

4.7u

F/25

V4.

7uF/

25V

11

TP1

TP1

TEST

POIN

TTE

STPO

INT

2211

33

D3

D3

BAS1

6BA

S16

1122

C9

C9

3300

pF/2

50V

3300

pF/2

50V

1122

L1L1

11.5

uH/2

A11

.5uH

/2A

11

33 22

Q8

Q8

2N70

022N

7002

44

33 55

99

1010 88

11

1212

K1K1 TX2-

224V

TX2-

224V

1122

C10

C10

1000

pF/2

50V

1000

pF/2

50V

1122

L2L2

2.8u

H/2A

2.8u

H/2A

11 22

R13

R13

10K

10K

11 22

R112

R112

00

1122

R18

R18 1.

00K

1.00

K

11 22

C13

C13

10nF

10nF

11 22

R7R7 100

100

11 22

R8R8 100

100

11 22

C14

C14

10nF

10nF

S1S1S1S1

S2S2S2S2

S3S3S3S3

1111

2222

3333

4444

5555

6666

7777

8888

J1J1 KMD

G-88

S-BBS

KMD

G-88

S-BBS

1122

R17

R17 14

.0K

14.0

K

1122

C21

C21

0.1u

F0.

1uF

1122

R21

R21

1.00

K1.

00K

1122

R20

R20

300K

_NL

300K

_NL

1122

C22

C22

0.1u

F0.

1uF

1122

R19

R19

1.5K

_NL

1.5K

_NL

11 22

C19

C19 0.

1uF

0.1u

F

11 22

C20

C20

0.1u

F_N

L0.

1uF_

NL

1122

L4L4

100u

H_N

L10

0uH_

NL

1122

R23

R23

0_N

L0_

NL

11 22

C23

C23

0.1u

F_N

L0.

1uF_

NL

1122

R22

R22

1.5K

_NL

1.5K

_NL

VDD

VDD

11

GN

DG

ND

22

ININ33

VCC

VCC

55

OUT

OUT

44

U2U2

TPS2

819D

BVTP

S281

9DBV

++ --

33 22

66

77 44

U4U4

LMH6

642M

ALM

H664

2MA

11 22

C16

C16

4.7u

F/25

V4.

7uF/

25V

FB1

FB1

HZ12

06HZ

1206

11 22

C15

C15

0.1u

F0.

1uF

+40

VAD

J+

40VA

DJ

VCC

VCC

VCC

VCC

GN

DG

ND

GN

DG

ND

VIN

VIN

GN

DG

ND

GN

DG

ND

GN

DG

NDG

ND

GN

D

VCC

VCC

-55V

-55V

GN

DG

ND

GN

DG

ND

VCC

VCC

US_G

ND

US_G

ND

US_G

ND

US_G

ND

IRF

IRF

1_3M

HZ1_

3MHZ

HEAD_CLOCK HEAD_CLOCK HEAD_DATA HEAD_DATA

DRA

IND

RAIN

VRF

VRF

HEAD

_DET

ECT

HEAD

_DET

ECT

SOUN

DHE

AD CC

ON

NEC

TOR

SOUN

DHE

AD CC

ON

NEC

TOR

1122

L3L3 100u

H10

0uH

11 22

C4

C4

33uF

/50V

33uF

/50V

44 33

55 66 881122

77

T1T1

ENG

0078

7EN

G00

787

11 22

C11

C11

4.7u

F/25

V4.

7uF/

25V

11 22

C12

C12

0.1u

F0.

1uF

GN

DG

ND

GN

DG

ND

R156

R156 00

R157

R157 00

11 22

C98

C98 NL

NL

GN

DG

ND

GN

DG

ND

GN

DG

ND

FB10

FB10

HZ12

06HZ

1206

FB9

FB9

HZ12

06HZ

1206

R151

R151 00

R152

R152 00

11 22

R160

R160

22_N

L22

_NL

VDD

VDD

11

GN

DG

ND

22

ININ33

VCC

VCC

55

OUT

OUT

44

U3U3

TPS2

819D

BVTP

S281

9DBV

GN

DG

ND

1122

C29

C29

0.1u

F0.

1uF

11 22

R27

R27 62

062

0

11 22

R25

R25

2.2K

2.2K

11 22

R26

R26

2.2K

2.2K

11 22

C25

C25

0.1u

F0.

1uF

11 22

C27

C27

0.1u

F0.

1uF

FBK

FBK

1313

LOC

K/IP

RGLO

CK/

IPRG

99EX

TLF

EXTL

F1010

VSS

VSS

1111

CLK

NC

LKN

1616

VDD

VDD

1414

REF

REF

1212

CLK

PC

LKP

1515

VDD

VDD

88

VSS

VSS

44

ADD

R1AD

DR1

77

XOUT

XOUT

66XI

NXI

N55

ADD

R0AD

DR0

33SD

ASD

A22

SCL

SCL

11

U5U5

FS61

31-00

1FS

6131

-001

11 22

C28

C28 47

pF47

pF

1122

R28

R28

620

620

1122

C17

C17

47pF

47pF

FB2

FB2

HZ12

06HZ

1206

11 22

C26

C26

4.7u

F/25

V4.

7uF/

25V

VCC

VCC

GN

DG

ND

VCC

VCC

GN

DG

ND

GN

DG

ND

GN

DG

ND

GN

DG

ND

GN

DG

ND

GN

DG

ND

GN

DG

ND

VCC

VCC

GN

DG

ND

GN

DG

ND

DUT

Y_C

YCLE

DUT

Y_C

YCLE

CLK

_REF

CLK

_REF

I2C

_DAT

AI2

C_D

ATA

I2C

_CLO

CK

I2C

_CLO

CK

11

TP14

TP14

TEST

POIN

TTE

STPO

INT

GN

DG

ND

33 44 55

66

U1B

U1B

74HC

27D

74HC

27D

11221313

1212

U1A

U1A

74HC

27D

74HC

27D

99

1010 1111

88

1414 77

U1C

U1C

74HC

27D

74HC

27D

GN

DG

ND

US GG

roun

dUS

GGro

und

11 22

R5R5 0.56

0.56

44

33 55

99

1010

88

11 1212

K2K2 TX2-

224V

TX2-

224V

GN

DG

ND

Ultr

asou

nd B

oard

Sche

mat

ic 1

of 3

(AM

P)

Page 36: Ultrasound Therapy Systems SERVICE MMANUAL

32

Inte

lect

®/G

enis

ys®

Ult

raso

un

d T

her

apy

Sys

tem

s9-

SCH

EMAT

ICS

Ultr

asou

nd B

oard

Sche

mat

ic 2

of 3

(Por

ts)

Page 37: Ultrasound Therapy Systems SERVICE MMANUAL

33

Inte

lect

®/G

enis

ys®

Ult

raso

un

d T

her

apy

Sys

tem

s9-

SCH

EMAT

ICS

Ultr

asou

nd B

oard

Sche

mat

ic 3

of 3

(V S

uppl

y)

Page 38: Ultrasound Therapy Systems SERVICE MMANUAL

Intelect®/Genisys® Ultrasound Therapy Systems10- WARRANTYChattanooga Group ("Company") warrants that the Ultrasound Therapy System("Product") is free of defects in material andworkmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If thisProduct fails to function during the two years warranty period due to a defect in material or workmanship, Company or the sellingdealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product isreturned to the Company or the dealer.All repairs to the Product must be performed by a service center certified by the Company. Any modifications or repairs performedby unauthorized centers or groups will void this warranty.The warranty period for applicators is one year (12 months).This Warranty Does Not Cover:Replacement parts or labor furnished by anyone other than the Company, the selling dealer or a certified Company servicetechnician.Defects or damage caused by labor furnished by someone other than Company, the selling dealer or a certified Company servicetechnician.Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonableand necessary maintenance or any use that is inconsistent with the Product User's Manual.

COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.

Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion maynot apply to you.To Obtain Service From Company or the selling dealer under this warranty:

1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to theCompany should be

sent to:4717 Adams RoadHixson, TN 37343 USATelephone: +1 423 870-7200FAX: +1 423 870-2046

and2. The Product must be returned to the Company or the selling dealer by the owner.

This warranty gives you specific legal rights and you may also have other rights which vary from state to state or location tolocation.The Company does not authorize any person or representative to create for it any other obligation or liability in connection with thesale of the Product. Any representation or agreement not contained in the warranty shall be void and of no effect.

THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

34

Page 39: Ultrasound Therapy Systems SERVICE MMANUAL

4717 Adams RoadP.O. Box 489Hixson, TN 37343 U.S.A.

+1-423-870-7200 OUTSIDE U.S.A

+1 423-870-2046 OUTSIDE U.S.A. FAX

www.chattgroup.com

27670A

0 4 1 3

More Trusted Products from Chattanooga GroupAdapta®

Massage Tables & ChairsA.E.R.™ Boot

Auto Edema Reduction BootA.E.R.™ Compression System

Hot and Cold Compression Therapyboo-boo pac™

Child Size Bear-Shaped Cold PackCambion®

Shock Dampening Foot Care ProductsCan-do®

Exercise Bands, Tubes and BallsCarpal-Trac™

Carpal Traction AccessoryCervical Traction System

Clinical Cervical TractionColPaC®

Chilling Units and Reusable Cold Therapy ProductsConductor Gel™

Highly Conductive Ultrasound GelContracture Products

Contracture Management Orthotic ProductsDURA-STICK® Electrodes

Self-Adhesive ElectrodesEMG Retrainer® and EMG Retrainer® IR

Dual Channel Surface EMGErgostyle®

Chiropractic TablesFlexi-PAC® I and II

Reusable Hot and Cold CompressesFluido DHT

®

Dry Heat Whirlpool Therapy UnitsGel Medex™

Gel Mattress OverlayHydrocollator®

Heating Units and HotPacs™

© 2004 Encore Medical, L.P.

Intelect®

Ultrasound and Electrotherapy ProductsMagneciser™

Exercise EquipmentMeasurement Instruments

Dynamometers, Goniometers, ect.Myossage®

Massage LotionNylatex®

Elastic WrapsOptiFlex®

Continuous Passive MotionOpti-Ice™

Cold Therapy SystemPara-Care®

Paraffin Wax UnitPillow Perfect™

Cervical Pillow LinePivotal Therapy System™

Orthotics for the SpinePresSsion®

Intermittent CompressionPron Pillo™

Positioning PillowSensaFlex™

Hot/Cold PackSPORT-PAC™

Soccer Ball Shaped Cold PackTherma-Wrap™

Hot and Cold CompressionTriton®

Treatment and Traction EquipmentTru-Trac®

Traction EquipmentWellness 1st™

Back Support

ISO 13485 CERTIFIED