ua-66 service manual(v2.0)

30
UA-66 Urine Analyzer Service Manual

Upload: ledangtuanbk

Post on 11-Mar-2015

1.002 views

Category:

Documents


166 download

TRANSCRIPT

UA-66 Urine Analyzer

Service Manual

UA-66 Urine Analyzer Service Manual(2.0)

- i -

Copyright

© 2005-2010 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.

For this Operator’s Manual, the issue date is 2010-04.

Intellectual Property Statement

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns

the intellectual property rights to this Mindray product and this manual. This manual may refer to

information protected by copyright or patents and does not convey any license under the patent rights

or copyright of Mindray, or of others.

Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the

information in this manual in any manner whatsoever without the written permission of Mindray is

strictly forbidden.

Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative

work of this manual in any manner whatsoever without the written permission of Mindray is strictly

forbidden.

, , , , BeneView, WATO,

BeneHeart, are the trademarks, registered or otherwise, of Mindray in China and other countries.

All other trademarks that appear in this manual are used only for informational or editorial purposes.

They are the property of their respective owners.

Responsibility on the Manufacturer Party

Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray shall not be liable for

errors contained herein nor for incidental or consequential damages in connection with the

furnishing, performance, or use of this manual.

Mindray is responsible for safety, reliability and performance of this product only in the condition

that:

• all installation operations, expansions, changes, modifications and repairs of this product are

conducted by Mindray authorized personnel;

• the electrical installation of the relevant room complies with the applicable national and local

UA-66 Urine Analyzer Service Manual(2.0)

- ii -

requirements;

• the product is used in accordance with the instructions for use.

Upon request, Mindray may provide, with compensation, necessary circuit diagrams, calibration

illustration list and other information to help qualified technician to maintain and repair some parts,

which Mindray may define as user serviceable.

Note

This equipment is to be operated only by medical professionals trained and authorized by Mindray

or Mindray-authorized distributors.

Warning

It is important for the hospital or organization that employs this equipment to carry out a reasonable

service/maintenance plan. Neglect of this may result in machine breakdown or injury of human

health.

Warranty

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,

EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS

FOR ANY PARTICULAR PURPOSE.

Exemptions

Mindray's obligation or liability under this warranty does not include any transportation or other

charges or liability for direct, indirect or consequential damages or delay resulting from the

improper use or application of the product or the use of parts or accessories not approved by

Mindray or repairs by people other than Mindray authorized personnel.

This warranty shall not extend to:

� any Mindray product which has been subjected to misuse, negligence or accident;

� any Mindray product from which Mindray's original serial number tag or product identification

markings have been altered or removed;

� any product of any other manufacturer.

UA-66 Urine Analyzer Service Manual(2.0)

- iii -

Return Policy

Return Procedure

In the event that it becomes necessary to return this product or part of this product to Mindray, the

following procedure should be followed:

1. Obtain return authorization: Contact the Mindray Service Department and obtain a Customer

Service Authorization (Mindray) number. The Mindray number must appear on the outside of

the shipping container. Returned shipments will not be accepted if the Mindray number is not

clearly visible. Please provide the model number, serial number, and a brief description of the

reason for return.

2. Freight policy: The customer is responsible for freight charges when this product is shipped to

Mindray for service (this includes customs charges).

3. Return address: Please send the part(s) or equipment to the address offered by Customer

Service department

Company Contact

Manufacture: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial

Park, Nanshan, Shenzhen, P.R.China,518057

Phone: +86 755 26582479 26582888

Fax: +86 755 26582500 26582501

Safety Precautions

1. Meaning of Signal Words

In this manual, the signal words WARNING, CAUTION and NOTE are used regarding

safety and other important instructions. The signal words and their meanings are defined as

follows. Please understand their meanings clearly before reading this manual.

Signal word Meaning

UA-66 Urine Analyzer Service Manual(2.0)

- iv -

WARNING Read the statement following the symbol. The statement is alerting

you to an operating hazard that can cause personal injury.

CAUTION Read the statement following the symbol. The statement is alerting

you to a possibility of system damage or unreliable results.

NOTE Read the statement following the symbol. The statement is alerting

you to information that requires your attention.

Labels Used on the System

The labels attached to some UA-66 panels use symbols with the text to clarify the meaning of the

text. The table below explains the symbols on the labels.

Serial Number

Manufacturer

Date of Manufacture

In Vitro Diagnostic equipment

Warning: risk of personal injury or equipment damage

~ Alternating current (AC)

Graphics

All graphics, including screens and printout, are for illustration purposes only and must not be

used for any other purpose.

UA-66 Urine Analyzer Service Manual(2.0)

- I -

Table of Contents

1 Introduction ........................................................................................................ 1

1.1 Using the manual .............................................................................................................1

1.2 Instrument Description .....................................................................................................1

1.3 Specifications ...................................................................................................................1

2 Operating System .............................................................................................. 3

2.1 Introduction.......................................................................................................................3

2.2 Initialization Phase ...........................................................................................................4

2.3 Self-Diagnosis Program ...................................................................................................5

3 General Maintenance......................................................................................... 7

3.1 Replace the Fuses ...........................................................................................................7

3.2 General Cleaning .............................................................................................................7

3.3 Daily Cleaning ..................................................................................................................7

3.4 Disinfection.......................................................................................................................8

3.5 Periodic Cleaning .............................................................................................................9

3.6 Threshold Adjustment ......................................................................................................9

3.7 Upgrade Software Version .............................................................................................12

4 Troubleshooting ............................................................................................... 13

4.1 E01-System Memory failure...........................................................................................13

4.2 E02 System Data Failure ...............................................................................................13

4.3 E03 Feed Table Failure ..................................................................................................13

4.4 E04 Calibration Error......................................................................................................13

4.5 E05 Feed Table Polluted................................................................................................13

4.6 E10 Stray Light Error......................................................................................................14

4.7 E11 Positioning Error .....................................................................................................14

4.8 E12 Strip Placement Error .............................................................................................14

4.9 E21 Printer Paper End ...................................................................................................14

4.10 E22 Printer Handle.........................................................................................................14

4.11 Troubleshooting Chart....................................................................................................14

5 Illustrated Parts ................................................................................................ 16

5.1 Electric Schematic..........................................................................................................16

5.2 Final Assembly ...............................................................................................................17

5.3 Upper Case Assembly....................................................................................................18

5.4 Lower Case Assembly....................................................................................................19

5.5 Feed Table Assembly .....................................................................................................20

5.6 Processor Board ............................................................................................................22

UA-66 Urine Analyzer Service Manual(2.0)

-1-

1 Introduction

1.1 Using the manual

This manual provides servicing information for the UA-66 Urine Chemistry Analyzer. The

instrument may be operated at 110VAC 50/60Hz or 220VAC 50/60Hz.

1.2 Instrument Description

The UA-66 Urine Analyzer is a semi-automated, bench top instrument designed to “read” U-11

Urine Reagent Strip for urinalysis. The instrument is initially configured to read U-11 Para.

Reagent Strips. U-11 Para. Reagent Strips are also qualified for use.

The instrument use a high-light white LED as the light source to illuminate the reacted pads,

and use 3 photoelectric sensor at 470nm, 540nm and 625nm to read the reflectance from the

reacted pads.

All the control signals and analog I/Os is processed directly by a Hitachi H8S/2350 MPU. This

makes the instrument simple and reliable.

The instrument uses a Flash ROM as system non-volatile memory to store the system software,

user options and test reports. System software can be upgraded using a personal computer via

RS-232C port. Up to 1000 test reports can be stored in the instrument. The reports can be

recalled via date and sample No.

Reports may be sent to the internal micro thermal printer.

1.3 Specifications

Parameter Detail

Power Required 220VAC±10% 50/60Hz

110VAC±10% 50/60Hz

Line Leakage Current ≤0.5 mA

Dimensions Width: 360 mm (14.2 in.)

Depth: 300 mm (11.8 in.)

Height: 100 mm (3.94 in.)

Weight 3.8 Kg

Ambient Operating

Temperature range

10 ℃ to 30 ℃

Ambient Operating

Humidity range

<70%

Testing Type Reflectance Photometer

Light Source LED

UA-66 Urine Analyzer Service Manual(2.0)

- 2 -

Parameter Detail

Wavelength 470nm, 540nm and 625nm

Work Flow 60 seconds( single step mode )

60 seconds( slow batch mode )

30 seconds( fast batch mode )

Printer EPSON micro thermo printer, width 57 mm

Display Liquid Crystal Display with back light, 240*128 dots

Interface RS-232C, DB9 female

Testing items Leukocytes, Urobilinogen, Bilirubin, Nitrite, pH, Specific

Gravity, Blood, Glucose, Protein, Ketone, Ascorbic Acid

Usable Strip U-10 Para. & 11 Para

Fuse T 250V, 2A

Table 1.1

UA-66 Urine Analyzer Service Manual(2.0)

-3-

2 Operating System

2.1 Introduction

This section describes the operating system functions of UA-66 Urine Analyzer.

The input device of the instrument is a keyboard with 6 keys, see Table 2.1.

Key Function

FUN Used to show or close a pop-up menu or mark a specimen during testing.

▲ Used to select the previous menu or option, or increase the selected digit

while a number is being edited.

▼ Used to select the next menu or option, or decrease the selected digit

while a number is being edited.

► Used to select the next digit while editing an integer, to feed paper at the

main menu or to select the menu in the next column while multi column

being displayed.

ESC Used to stop current operation.

Used to exit from current menu to upper menu level.

START Used to begin a reagent test or to make new settings accepted.

Table 2.1

The output devices include a 240x128 liquid crystal display (LCD) and a micro thermal printer.

The display indicates all the necessary information of the current operation, and the printer only

outputs the result of testing.

All the accessible functions of the instrument are organized in a multi level menu structure, see

Table 2.2.

UA-66 Urine Analyzer Service Manual(2.0)

- 4 -

Single Mode

Slow Batch Mode Test

Fast Batch Mode

Reports

Strip

Critical Value

Grade

Displayer

Date Realtime Clock

Time

Option

Shade

Printer

Diagnosis

9600

19200

38400

Serial Port

57600

Options

Serial No

Table 2.2

2.2 Initialization Phase

After the power is on, the instrument shows the logo immediately on the display and begins to

perform a self diagnosis and calibration process including the following steps:

1. Check the system ROM and RAM. If incorrect adding sum of the system ROM or failure of

any 1 bit of the system RAM was detected, an E01 error occurs. In this case, the software will

terminate after the error message being displayed.

2. The feed table tries to find the position sensor. If failed, an E03 error occurs. In this case, the

software will terminate after the error message being displayed.

3. The instrument samples the intensity of stray light from the environment. If the intensity is

greater than 100 AD value, an E10 error occurs. In this case, press START to continue.

4. The instrument samples the reflecting intensity of the calibration chip and compares the

values with the original values. If the difference greater than 10%, an E04 occurs.

If any error reported during the progress, refer to 4 Troubleshooting.

Once the self-diagnosis and calibration process completed, the instrument will show the system

main menu including the logo, the software version, operating items, date and time.

UA-66 Urine Analyzer Service Manual(2.0)

-5-

2.3 Self-Diagnosis Program

The instrument equips a self-diagnosis program. To run the self-diagnosis program, plug the

RS-232 short connector into the RS-232 port on the rear panel of the instrument and then power

on the instrument. The self-diagnosis program performs a step-by-step check as following.

1. Display Diagnosis

The self-diagnosis program turns the dots of the display on and off column by column first,

then turns the dots on and off row by row. During this period, any defective dots can be found

easily.

If the instrument can’t reach this step, the main board maybe defective.

2. Keyboard Diagnosis

The operator presses all the 6 keys one by one. Once a legal key input is read, the

corresponding key name is displayed on the display. When all the 6 keys are correctly entered,

the self-diagnosis program goes to the next step.

If any key cannot be read, the keyboard is defective.

3. Real Time Clock Diagnosis

The self-diagnosis program shows the current date and time.

If the date or time is not correct, the real time clock need to be correct or the real time clock

chip maybe defective.

If it does not run correctly, the real time clock chip maybe defective.

Press START to go to the next step.

4. Feed Table Diagnosis

Press ▲ to make the feed table to move 2.5mm inward. Don’t do this if the feed table tip is

inside the instrument. Press ▼ to make the feed table to move 2.5mm outward. Don’t do this if

the feed table is near the feed position. If the feed table does not move, the motor or the motor

driver on the main board maybe is defective.

Press ► to make the feed table to find the position sensor. If E03 occurs, the position sensor

is defective.

Press START to go to the next step.

5. Analog Circuit Diagnosis

The instrument samples the reflecting intensity of the calibration chip once 1 second, and

shows the difference between the current values and the original values.

Press START to go to the next step.

6. Printer Diagnosis

Open the printer cover on the top of the upper case.

Rise up and put down the handle, the status of the handle will be displayed on the display. If it

is not correct, the handle sensor is defective.

Remove the paper from the printer then insert paper tip under the roller, and the status of the

paper end sensor will be displayed on the display. If it is not correct, the paper end sensor is

UA-66 Urine Analyzer Service Manual(2.0)

- 6 -

defective.

Install the paper correctly, and put down the handle, then press FUN , a test page will be

printed out. If the roller does not scroll, the roller motor or the motor driver is defective. Notice

the full width horizontal line at the bottom of the test page. If any dots lost, the thermal head of

the printer is defective.

Press START to end printer diagnosis.

This step is the last step of the self-diagnosis program.

UA-66 Urine Analyzer Service Manual(2.0)

-7-

3 General Maintenance

3.1 Replace the Fuses

Power off the instrument and pull out the line cord from the instrument. Screw off the fuse

holders on the rear panel anti-clockwise using a driver. Replace the broken fuses with the

identical fuses and reinstall the holders back to the rear panel.

3.2 General Cleaning

Keep the exterior of the instrument free from dust at any time. If needed, the exterior may be

cleared using a damp cloth and a mild detergent. Do not use any type of solvent on any part of the

instrument. Do not use any type of lubrication on the instrument. Use of oil, grease, silicon spray,

or any other lubrication may cause damage.

3.3 Daily Cleaning

The feed table and feed table insert must be kept clean to obtain the best result with the OP151

instrument / reagent strip system. At the end of each day use, the instrument‘s white plastic

calibration chip, located on the feed table insert, and the reagent strip holding channel should be

cleaned. To clean the feed table insert, perform the followings:

1. Turn the instrument on. The feed table will move out from the “storage” position to the “load”

position.

2. Grasp the front end of the insert on each side with the thumb and one finger. Apply a counter

pressure with the palm of the hand against the feed table and gently pull the insert outward.

After the insert is completely exposed, lift the front end of insert up and off the table.

Figure 3.1 3. Clean the insert using a cotton-tipped applicator stick or other suitable absorbent material

wetted with distilled water. Thoroughly swab the strip holding channel area until it is clear.

UA-66 Urine Analyzer Service Manual(2.0)

- 8 -

Inspect the surface of the white plastic calibration chip for dust, foreign material, scratches or

scuffs. If necessary, gently wipe the calibration chip with a cotton-tipped applicator stick

wetted with distilled water. Allow the calibration chip to air dry before using. Make sure no

foreign particles remain on its surface. If the calibration chip cannot be cleaned or is

scratched or scuffed, contact the customer service for replacement.

NOTE: Do not use any material that will scratch the feed table insert or the white

plastic calibration chip. Do not use solvents of any kind to clean the calibration chip.

4. Wipe the feed table surface and feed table mounting groove with a damp cloth. Wipe the feed

table dry with an absorbent tissue.

NOTE: Residual water or solution on the feed table or feed table insert will cause

damage to the electronic components inside the instrument.

5. To replace the feed table insert in the feed table:

Hold the feed table insert in the same manner as when it was removed (holding it by the end

opposite the calibration chip). With the front end angled upwards, insert the calibration chip

end about midway onto the feed table. Slide the insert in while lowering the front end until the

insert is laying flat against the table and resistance is felt. Then place the thumb against the

front edge of the insert and two fingers against the underside of the feed table. Apply a

counter-pressure to the feed table while pressing the insert firmly inward. The insert is

properly positioned in the feed table when the front edge is flush with the retain rails of the

feed table.

3.4 Disinfection

The only part of the OP151 instrument that comes into contact with the urine specimen is the

feed table insert. To disinfection the insert, perform the followings:

1. Remove, clean and dry the feed table insert as instructed in 3.3 Daily Cleaning.

2. Prepare one of the following solutions for use in disinfection:

a. 2% glutaraldehyde solution - CidexTM (use full strength) is a commercially available

product that is acceptable for use. Refer to the labeling accompanying this disinfection

product for complete instructions on its use.

b. 0.05% sodium hypochlorite solution – dilute 1 mL of 5% sodium hypochlorite( e.g., Clorox®

Bleach ) into 99 mL of water for a 1:100 dilution.

c. Isopropyl alcohol (70% - 85%), undiluted.

UA-66 Urine Analyzer Service Manual(2.0)

-9-

3. Fill a tall, narrow container to a depth of 4 inches (10cm) with the disinfection solution.

4. Immerse the feed table insert into the disinfection solution. So the white plastic calibration

chip must be above the liquid level.

NOTE: Do not allow the disinfection solution to come in contact with the white plastic

calibration chip, located on the feed table insert, or with any other portion of the feed table or

instrument.

5. Allow the feed table insert remain in the solution for 10 minutes.

6. Remove the feed table insert and rinse thoroughly with distilled water to remove any residual

solution.

7. Wipe the feed table insert in dry with a absorbent tissue.

8. Replace the feed table insert in the feed table, as instructed in 3.3 Daily Cleaning.

3.5 Periodic Cleaning

If the feed table insert becomes excessively dirty and cannot be adequately cleaned as

described in 3.3 Daily Cleaning, the following procedure should be used:

1. Remove the feed table insert from the table, as instructed in 3.3 Daily Cleaning.

2. Wipe the strip holding channel of the feed table insert with 0.1N sodium hyndroxide (NaOH)

using a cotton-tipped applicator stick or other suitable absorbent material. Thoroughly swab

the channel area until it is clean.

NOTE: Do not wipe the white plastic calibration chip with the NaOH. Do not use any

material that will scratch the feed table insert. Do not allow the NaOH to come in

contact with the white plastic calibration chip or with any other portion of the feed table

or instrument.

3. Rinse the feed table insert channel thoroughly with water to remove any residual NaOH.

4. Wipe the feed table surface and feet table mounting groove with a damp cloth. Wipe the feed

table dry with an absorbent tissue.

5. Replace the feed table insert in the feed table, as instructed in 3.3 Daily Cleaning.

3.6 Threshold Adjustment

NOTE: The adjustment to the thresholds will affect the test result. Only trained service

personnel can do the adjustment.

The threshold of each parameter can be modified in “the customer backdoor” of the

instrument. To enter the customer backdoor, pressing START while power on the instrument, a

password edit box will be shown:

PW:::: 000000

Figure 3.2

Using ▲ , ▼ and ► to enter the password “011119”, press START to enter the threshold

UA-66 Urine Analyzer Service Manual(2.0)

- 10 -

adjusting menu, see Figure 3.3. The instrument will print a threshold list in the mean time.

ThresholdThresholdThresholdThreshold

LEULEULEULEU PHPHPHPH BILBILBILBIL

NITNITNITNIT BLDBLDBLDBLD GLUGLUGLUGLU

UROUROUROURO SGSGSGSG ASCASCASCASC

PROPROPROPRO KETKETKETKET

Figure 3.3

At Figure 3.3 select the parameter to be adjusted using ▲ , ▼ and ► , press START to modify

the threshold list. Figure 3.4 shows the case while the thresholds of LEU being selected.

Threshold>LEUThreshold>LEUThreshold>LEUThreshold>LEU

- 0.662

±±±± 0.576

+ 0.501

2+ 0.402

3+

Figure 3.4

Press ▲ , ▼ to select the threshold to be modified, press START to pop up an edit box, see

Figure 3.5. Modify the value using ▲、▼、► , press START when done.

New Value:::: 0.662

Figure 3.5

At Figure 3.4, press FUN to popup a box, see Figure 3.6.

Load Default Test

Figure 3.6

Select “Load Default” and press START to recover all thresholds of current parameter to the

values while the instrument is being shipped. That means all the modification to the

thresholds of the current parameter made by the customer will be lost.

Select “Test” and press START to perform a specimen (for calibrating) test that will help the

operator to decide the value of a particular threshold. The test procedure show step by step as

following:

UA-66 Urine Analyzer Service Manual(2.0)

-11-

1. Press START to begin the test.

Press START

When ready

Figure 3.7

2. Dip the strip into the specimen.

Dip the strip

████████████████████

Figure 3.8

3. Place the strip onto feed table.

Place the strip

████████████████████

Figure 3.9

4. Wait for the reaction time.

Timing............

████████████████████

Figure 3.10

5. Read the reagent strip, and display the result.

0.476

Figure 3.11

The operator can decide how to adjust the threshold according to the test result above.

While modification of one parameter is done, press ESC to return to Figure 3.3. Before this

the instrument asks the operator to keep the changes, see Figure 3.12. Select “Yes” to keep the

changes, “No” to abandon the changes.

Validate Modifications???? Yes No

Figure 3.12

At Figure 3.3, press FUN to popup “Load Default” box. Press START to recover all the

thresholds of all parameters to the values when the instrument was shipped. That means all the

thresholds made by the customer will be lost.

At Figure 3.3, press ESC to show a confirm box, See Figure 3.13. Select “Yes” to save the

modified thresholds of all parameters to the flash ROM.

UA-66 Urine Analyzer Service Manual(2.0)

- 12 -

Save modifications???? Yes No

Figure 3.13

3.7 Upgrade Software Version

The system software is stored in Flash ROM, and a new version of the system software can be

downloaded from a personal computer via RS-232 cable. To upgrade the software, do the

followings:

1. Connect the instrument and a personal computer via a RS-232 cable.

2. Run the upgrade tool “update.exe” at a personal computer, see Figure 3.14.

Figure 3.14

3. Select the serial port being used.

4. Select “57600”, “odd”.

5. Click “File” to open the binary file of the new version software.

6. Click “Transmit” then power on the instrument to start the transmission.

7. Power off the instrument to end the entire procedure when the transmission ends,

UA-66 Urine Analyzer Service Manual(2.0)

-13-

4 Troubleshooting

4.1 E01-System Memory failure

Occurs when the instrument was turned on. If the system ROM or RAM cannot pass an initial

self-diagnosis, this error occurs.

The only way is replacing the main board.

4.2 E02 System Data Failure

Occurs if the system calibration data stored in the flash ROM was lost.

The instrument must be re-calibrated by trained service personnel.

4.3 E03 Feed Table Failure

Occurs if the instrument does not detect the table position sensor within a specified number of

motor steps. Possible reasons are:

♦ The position sensor is defective;

♦ The motor driver on the main board is defective;

♦ The motor is defective;

To determine the exact reason, run the self-diagnosis program. If the feed table can not move

in/out when ▲ / ▼ pressed, the motor driver on the processor board or the stepper motor is

defective. If the feed table can move in/out when ▲ or ▼ pressed, but cannot find the feed

position when ► pressed, the position sensor is defective.

4.4 E04 Calibration Error

Occurs during the initial self-diagnosis after the instrument is powered on.

The reflectivity of the plastic calibration chip was found 5% lower than its origin value. Possible

reason is that the plastic calibration chip was polluted or heavily scratched. Do the followings:

♦ Clean the plastic calibration chip as instructed in 3.3 Daily Cleaning.

♦ Clean the plastic calibration chip as instructed in 3.5 Periodic Cleaning.

♦ Replace the feed table insert with a new one.

4.5 E05 Feed Table Polluted

Occurs during the initial self-diagnosis after the instrument is powered on. A reflectivity higher

than 5% was detected while scanning the strip hold channel on the feed table insert.

If any foreign material is in the reagent strip holding channel, remove it.

If the feed table insert is heavily polluted by the urine specimen, clean it as instructed in 3.3

Daily Cleaning. If this doesn’t take effect, clean the feed table insert as instructed in 3.5 Periodic

Cleaning.

UA-66 Urine Analyzer Service Manual(2.0)

- 14 -

4.6 E10 Stray Light Error

Occurs when a background higher than 100 AD value was detected. This check is done before

and after every regent pad is tested. Possible reason is that the location of the instrument doesn’t

match the environment factors. Move the instrument to a place with no direct sunlight and no

working strong light source nearby and try again.

4.7 E11 Positioning Error

Occurs when a test is done and the feed table reaches the most outside position, a position

error greater than 10 pulses was detected. It’s normal if the error occurs once several months. If it

occurs frequently, the feed table assembly needs to be adjusted or even to be replaced.

4.8 E12 Strip Placement Error

Occurs when the strip is not placed correctly on to the feed table, especially the tip of the strip

does not touch the end of the strip holding channel.

This error has no relationship to any parts defective.

4.9 E21 Printer Paper End

Occurs when the thermal printer is out of paper. Install a new paper roll.

4.10 E22 Printer Handle

Occurs when the handle of the micro thermal printer is not in work situation.

4.11 Troubleshooting Chart

Symptom Probable Cause Corrective Action

The instrument

is dead.

1. The software is dead.

2. Line fuse is broken.

3. Power supply is

defective.

4. Main board is defective.

1. Power off the instrument

and power it again.

2. Replace the line fuse.

3. Replace the power supply.

4. Replace the main board.

Display is dead. 1. Display is defective.

2. Main board is defective.

1. Replace the display.

2. Replace the main board.

Feed table

doesn’t move.

1. The motor is defective

2. The main board is

defective.

1. Replace the feed table

assembly.

2. Replace the main board.

Reports can’t

be printed.

1. The printer option is set

to “not print”.

2. The printer is out of

1. Set the print option to

“print”.

2. Install a new paper roll.

UA-66 Urine Analyzer Service Manual(2.0)

-15-

Symptom Probable Cause Corrective Action

paper.

3. The handle of printer is

not put down.

4. The printer is defective.

5. The connect board is

defective.

3. Put down the handle.

4. Replace the printer.

5. Replace the connect

board.

6. Replace the main board.

No RS-232

communication.

1. The port selection of

terminal program is not

correct.

2. The baud rate selection

of terminal program does

not match the setting of

the instrument.

3. The RS-232C cable is

defective.

4. The main board is

defective.

1. Select another port and try

again.

2. Change the baud rate

selection of the terminal

program.

3. Replace the main board.

UA-66 Urine Analyzer Service Manual(2.0)

- 16 -

5 Illustrated Parts

5.1 Electric Schematic

Figure 5.1

UA-66 Urine Analyzer Service Manual(2.0)

-17-

5.2 Final Assembly

Figure 5.2

Final Assembly

No. Description Specification Number Required

1 Model Label 1

2 Lower Case Assembly 1

3 Upper Case Assembly 1

4 Brazier Head Telescopic Screw M3×10 7

5 RS-232 Label 1

6 Rating Label 1

Table 5.1

UA-66 Urine Analyzer Service Manual(2.0)

- 18 -

5.3 Upper Case Assembly

Figure 5.3

Upper Case Assembly

No. Description Specification Number Required

1 Upper Case 1 2 Printer Cover 1 3 Keyboard 1 4 Connect PCB 1 5 Brazier Head Telescopic Screw M3×6 2

6 Insulation Washer φ3 2

7 Thermal Paper 57mm×30mm 1

8 57mmThermal Printer M-T153 1

9 Brazier Head Telescopic Screw M2×6 3

10 Flat Washer 2 3

11 Brazier Head Telescopic Screw M3×6 4

12 LCD Module 1

13 Insulation Washer φ3 4

Table 5.2

UA-66 Urine Analyzer Service Manual(2.0)

-19-

5.4 Lower Case Assembly

Figure 5.4

Lower Case Assembly

No. Description Specification Number Required

1 Lower Case 1

2 Nut 1

3 Ring 1

4 Bolt 1

5 Philips Brazier Head Screw M3×8 1

6 Feed Table Assembly 1

7 Socket Head Screw M4×8 4

8 Switch Power Supply T-50B 1

9 Socket Head Screw M3×5 2

10 Philips Brazier Head Screw M3×6 2

11 Plastic Foot 4

12 Pin 4

13 Processor PCB 1

14 Philips Brazier Head Screw M3×6 4

15 Philips Brazier Head Screw M3×6 1

16 Screw 2

17 Nut 2

UA-66 Urine Analyzer Service Manual(2.0)

- 20 -

No. Description Specification Number Required

18 Serial Port Cable 1

19 Socket Head Screw M3×5 2

21 Power Input With Fuse Holder 0171-2-CW 1

22 Philips Chamfer Head Screw M3×8 2

23 Nut M3 2

24 Fuse 6183.15 2

25 Power Switch R130-22C-11P 1

Table 5.3

5.5 Feed Table Assembly

Figure 5.5

UA-66 Urine Analyzer Service Manual(2.0)

-21-

Feed Table Assembly

No. Description Specification Number Required

1 Sensor Assembly 1

2 Feed Table 1

3 Pillar 4

4 Motor Assemble Board 1

5 Gear 1

6 Leading Track 2

7 Box 1

8 Box Cover 1

9 Step-Motor KP4P15G 1

10 Philips Brazier Head Screw M3×8 4

11 Philips Brazier Head Screw M4×35 4

12 Flat Washer 4 4

13 Position Sensor ST-150 1

14 Signal PCB 1

15 Brazier Head Telescopic Screw M3×6 4

16 Insulation Washer φ3 4

Table 5.4

UA-66 Urine Analyzer Service Manual(2.0)

- 22 -

5.6 Processor Board

Figure 5.6

P/N: 0103-20-61108(2.0)