Tutorial on the Ortec 935 Quad Constant Fraction Discriminator. Kevin Carnes Fall, 2004. Threshold Monitor, -200mV to -10V. Threshold Adjust, -20mV to -1000mV. This is the input pulse used for simulating the operation of the Ortec 935. This shows how the input signal is split, - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Kevin CarnesFall, 2004

Page 2: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Threshold Adjust, -20mV to -1000mV

Threshold Monitor,-200mV to -10V

Page 3: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

This is the inputpulse used for simulating the operation of the Ortec 935.

Page 4: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

This shows how theinput signal is split, one half attenuatedto 20% of its originalheight, and the otherhalf delayed. The two are then combined to producethe signal thatcan be seen at theCF Monitor.

Page 5: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Three CF Monitorsignals producedwith different delays.

Page 6: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

tdext = T + 1.0 ns, W1 outtdext = T – 0.7 ns, W1 in

Constant FractionMonitor

The equations below arefrom the 935 manual. Theygive the amount of delay produced from the externalcable to achieve a specificinternal delay T. Note thatthe value depends on whether or not the internaljumper W1 is in or out.

Page 7: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

This is an actual scopetrace of the CF Monitoroutput from the 935with a reasonably longdelay cable.

Page 8: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

CF Monitor scope trace with much shorter delay cable.

Page 9: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Walk Adjust,+/- 15 mV

Walk Adjust TestPoint, reads actual voltage

Page 10: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Zero crossing pointof simulated CFMonitor, with longcable (lg) and shortcable. Walk adjustis set to + 10 mV and – 10 mV.

Page 11: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Width Adjust, <4ns to>400ns, depending oninternal jumpers, updateor block

Page 12: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Scope trace of 935 output into 50 ohms, other two outputs in this channelunterminated.

Page 13: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Scope trace of 935output into 50 ohms, second outputterminated in 50 ohms, third output unterminated.

Page 14: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Scope trace of 935output into 50 ohms,both of the other outputs terminated in50 ohms.